1, 2017 January-March
Original Paper
Telangiectasia Detection in Wireless Capsule
Endoscopy Using the Color Slicing Technique
MIHAELA IONESCU1, C.T. STREBA2,3, C. CONSTANTIN VERE2,4,
A.G. IONESCU5, I. ROGOVEANU2,4
1
Department of Medical Informatics, University of Medicine and Pharmacy of Craiova, Romania
2
Research Center of Gastroenterology and Hepatology,
University of Medicine and Pharmacy of Craiova, Romania
3
Department of Scientific Methodology, University of Medicine and Pharmacy of Craiova, Romania
4
Department of Gastroenterology, University of Medicine and Pharmacy of Craiova, Romania
5
Department of Medical History, University of Medicine and Pharmacy of Craiova, Romania
ABSTRACT: Wireless capsule endoscopy represents a color imaging technology in the field of medical
endoscopy that is extensively used to detect lesions of the human digestive tract. It is the golden standard in
evaluating small bowel lesions, offering a set of digital snapshots difficult to get using other investigation methods. Its
major drawbacks are the time consumed for image analysis and the burden for the physicians that must spot and
classify lesions within more than 55000 images. This paper carries out a study on the detection of telangiectasia in
the small bowel, based on an adapted color slicing technique applied not only on unique frames, but on series of
successive frames, performing a global analysis suitable on partial sequences or entire wireless capsule endoscopy
movies. We have quantified the extracted features and determined a weighting algorithm to find telangiectasia
lesions. For frames containing potential lesions, we have determined features not only for the global image, but also
for the normal mucosa surrounding the lesion extracted from the image. This approach allows the physician to see
variations of parameters within a frame or a sequence that contains lesions. Experimental results prove that the
algorithm is effective in detecting telangiectasia patterns of different images, with an accuracy of 93.88%, reducing
thus the time spent for the analysis of the images acquired by wireless capsule endoscopy.
KEYWORDS: Wireless Capsule Endoscopy, telangiectasia detection, color slicing technique, medical image
analysis
10.12865/CHSJ.43.01.04 25
Mihaela Ionescu et al.- Telangiectasia Detection in Wireless Capsule Endoscopy Using the Color Slicing Technique
26 10.12865/CHSJ.43.01.04
Current Health Sciences Journal Vol. 43, No. 1, 2017 January-March
We have computed color, saturation and hue images, a more uniform illumination is required
for all original images, for potential lesions in regard to CCD [2].
extracted from the original frames following a Transmission means contribute to the amount
segmentation process, and for normal intestinal of noise in an image mainly due to the different
mucosa surrounding potential lesions within the interferences present in used channels.
same frame. The next step performed a Transmission using a wireless network tends to
quantification of all extracted features for the increase the noise, which is the case of the WCE
subsequent numerical analysis and the acquired images [13].
preliminary classification phase. A pre-processing stage meant to provide neat
images by reducing the noise is therefore optimal
Noise reduction
before specific image processing techniques are
Noise present in digital images may result
applied.
either from the image acquisition process (due to
physical sensing elements), or during Features identification, extraction and
transmission. A series of factors like analysis
environmental conditions or the quality of When it comes to telangiectasia detection, the
sensing elements may directly influence the main criterion used by physicians in making a
efficiency of image sensors [13]. Due to the diagnostic decision is the color appearance.
environment specific to the digestive tract and These lesions do not present active bleeding, so
also technical miniaturization restrictions of the they are not characterized by bright red colors,
physical equipment, WCE images are usually but more likely fall into a range of dark red to
corrupted and degraded by noise and blurs [14]. brown shades (Fig. 2).
Current versions of WCE use either From a digital point of view, this information
Complementary Metal Oxide Semiconductor is expressed through standard spaces (or models)
(CMOS) sensors or Charge Coupled Device that define a coordinate system and an associated
(CCD) sensors for image acquisition. Both subspace within this system, where every color is
technologies have advantages and drawbacks identified by a single point [13]. Thus, each color
when it comes to power consumption and is numerically expressed by the values of the
acquisition quality. CCD technology is coordinates in that specific system. In relation to
characterized by a high signal to noise ratio (i.e. a the most common color space-RGB (red, green,
high level of signal and low level of noise), blue)-telangiectasia lesions are characterized by a
despite the fact that, for CCD, the amount of dominant red color, therefore R has a high value
noise present in the acquired image is very (and thus reflects more light), compared to G and
sensitive to light levels and sensor temperature B. On the other hand, the surrounding normal
[1]. On the other hand, CMOS capsules have a intestinal mucosa has a low value of R [11,15].
better power management due to dedicated ASIC
imager chips, with dynamic light and exposure
adjustments but, in order to acquire good quality
Fig.2. a,b,c-Three images acquired by WCE, presenting normal intestinal mucosa, normal intestinal folds and
a telangiectasia lesion; d,e,f-Three color palettes, corresponding to previous WCE images
10.12865/CHSJ.43.01.04 27
Mihaela Ionescu et al.- Telangiectasia Detection in Wireless Capsule Endoscopy Using the Color Slicing Technique
However, from a human point of view, color The analysis and classification process was
is not interpreted as the sum of the tristimulus made based on a weighting algorithm. The
values (R, G, B) composing it, but is rather weights and thresholds were determined
described using its brightness and chromaticity experimentally. The main criterion of
(hue and saturation) [11,16]. Hue is a color telangiectasia detection is a significant number of
attribute describing the pure color (yellow, bits inside the sphere of interest, greater than a
orange or red), whereas saturation defines the predefined threshold, which is determined
degree to which a pure color is diluted by white according to the color palette of the entire
light. Bleeding regions are characterized by high sequence.
saturation and low brightness values, whereas
Results
normal tissue regions present low values for
After the analysis of half of the sequences
saturation and high values for brightness. Thus,
with telangiectasia lesions, we have concluded
saturation is another important feature that
that the threshold must be determined from the
differentiates potential telangiectasia lesions [17].
count of red pixels and the total number of
We have performed the separation of potential
neighboring frames indicating the same potential
telangiectasia lesions from the surrounding tissue
lesion. An experimental value of 84.6% from the
by highlighting the predefined range of colors,
radius was used in the classification algorithm.
using the color slicing technique. This technique
The variation of the saturation parameter
divides the color space in two non-overlapping
corresponds to the variation reported in literature
sections, according to a set of specific criteria
(higher values of the parameter for frames
[13]. The range of neighboring red shades that
containing potential telangiectasia lesions). The
must be detected in an image represents a section
general behavior of the hue parameter seems to
from the color space that must be separated from
be inverted unto the previous parameter. Lower
the rest. For the ease of computation, we will
values are determined for frames containing
consider this range as a sphere of interest within
potential telangiectasia lesions. The weight of
the color space, based on the Euclidian Distance
these parameters was computed based on the
(ED), as it is not very sensitive to variations in
current value extracted from a frame, combined
hue and saturation, but more to variations in
with the average difference between values
intensity [18]. The radius of this sphere was
corresponding to 10 successive neighboring
determined individually, for each sequence,
frames. Experimental values were used in the
based on the color analysis of all frames
classification algorithm: 2.2875 for saturation
included. This radius reflects the difference
and 1.1824 for hue.
between the average red values for all pixels in
Based on these weights, the remaining
the sequence and the pure red color.
sequences were analyzed using the classification
The quantification phase determines the
algorithm and achieved an accuracy of 93.88%,
numerical expression of all extracted features.
defined as the number of correct classifications of
Specific red shade colors of potential
the frames containing telangiectasia lesions,
telangiectasia lesions are extracted from the color
divided by the number of frames actually
palette. All frames are modified so that every
containing lesions. The algorithm has a
pixel with a color outside this range is replaced
sensitivity of 97.18% and a specificity of
by a black pixel. Then we count all non-black
99.71%.
pixels, thus we determine the number of pixels
that fall into the determined red shades range.
The quantification of the bleeding lesions is
Discussion
actually the expression of pixels whose color is a Automatic detection of telangiectasia lesions
shade of red, located in a specific range. in WCE videos must be guided by a series of
The red shades range is used also to segment aspects. Telangiectasia lesions are defined by
the original image into two complementary sub- reddish colors, different from the color palette
images: one contains only the potential lesion normally present in a lesion free WCE snapshot.
(only bits included in the sphere of interest), and Color may also be potentially altered by
the other contains the rest of the original image. intestinal fluids or food dregs. The illuminating
This segmentation is also used for the light source varies in intensity, and this may also
computation of saturation and hue parameters, affect the efficiency of lesion detection, as the
expressed as mean values for both the entire tissue closer to the light source is brighter; thus a
image and the extracted sections. descriptor robust to light variation is needed.
Two more major factors involved in a correct
28 10.12865/CHSJ.43.01.04
Current Health Sciences Journal Vol. 43, No. 1, 2017 January-March
automatic identification are: texture of based on different segments of the digestive tract,
telangiectasia lesions and location of the wireless and we have computed the differences related to
capsule, as the normal aspect of the mucosa in a pure red color. As a conclusion, dominant
several segments (esophagus, stomach, small yellow shades determine a larger red shades
intestine, and colon) may differ and may range, meaning also a higher rate of success for
influence the final result [7]. telangiectasia lesion detection. Due to these
That is why we have determined the range of arguments, we believe it is a good approach to
red shade colors corresponding to a potential perform a global analysis of the entire video file
telangiectasia lesion based on the color palette of prior to lesion identification, in order to take
the mucosa, modelling it as a sphere. However, correctly into account the color particularities of
the separation of telangiectasia lesions is not each film.
exact, thus pixels belonging to the normal Also, the global analysis and the computation
mucosa, but with more reddish colors, may be of individual parameters for lesions and non-
considered as pixels belonging to lesions. This lesions areas present a visual advantage for the
drawback may be addressed by modelling the examining physician. Fig. 3 shows a chart only
area of interest through various geometric for a sub-set of frames; for all parameters, except
elements, not only spheres, thus leading to the the red pixels count, the scales were adjusted in
extension of the area and to an enhanced order to have similar value ranges for all
flexibility needed to apply the algorithm to parameters and to allow a proper graphical
various patients presenting this type of lesions. comparison.
Prior to feature extraction, we have
determined the dominant colors of the frames,
Fig.3. Chart identifying a set of frames with a high content of red pixels (telangiectasia lesion present), and
the quantified features extracted from the same frames
Global charts comprising all numerical adjusting the red shades range according to the
parameters computed for all frames seem to be a color palette of the entire sequence. The color
good indicator for the examining physicians, slicing technique allows a proper definition of a
allowing them to focus their attention upon sphere of interest that regroups multiple shades
specific sequences from the entire films. of red-normally encountered in telangiectasia
lesions.
Conclusion Overall, the proposed algorithm reduces the
The quantification of features and the global number of images that are analyzed by the
analysis of sets of frames instead of individual physician, allowing him to concentrate on
frames turned out to be optimum, especially due potential lesion frames and to determine the
to taking into account the normal lesion-free diagnosis. Future work and improvements
frames in the neighborhood of one or more concern the use of several different spheres of
frames containing a lesion. interest, based on other potential lesions and the
Individuality and uniqueness of the patient’s extraction of new features related to the texture
characteristics are also taken into account by
10.12865/CHSJ.43.01.04 29
Mihaela Ionescu et al.- Telangiectasia Detection in Wireless Capsule Endoscopy Using the Color Slicing Technique
30 10.12865/CHSJ.43.01.04