Professional Documents
Culture Documents
August 1998
Issue 1
PRODUCT DESCRIPTION
The AMETEK Model 1200SE Series Dry Block Temperature The calibrators include an easy-to-read, high-contract LED
Calibrators are easy-to-use, portable, precision calibration in- display which indicates temperature in either oF or oC. Special
struments. The Model 1200SE Series calibrators provide fast, indicators are provided that alert the operator to current condi-
reliable temperature calibration of RTDs, thermocouples, tions, including a stability indicator that indicates when a stable
thermo switches, thermistors, plus other common tempera- temperature has been reached. Additional indicators are sup-
ture sensing devices. These units are ideal for high tempera- plied including SET, READ, HEAT, COOL, mA, HOLD and
ture applications such as kilns, boilers and incinerators. SLOPE.
Utilizing a compressed air cooling system, the cooling period A membrane-style keypad with tactile feedback is located on
is only 10-20 minutes. An integrated valve controllable from the front of the calibrator. Dedicated function keys are used to
the processor allows complete control of the cooling period in setup and operate the calibrator. Function keys include:
both manual and personal computer operations.
SET Let’s the user select any temperature
The Model 1200SE also features an ultra-stable thermocouple READ Calibrator will heat/cool to a selected SET
to ensure that the temperature is kept to within +1oF (+2.0oC) temperature and READ the actual tem-
o o o
at temperatures <1100 C and +2 F (+3.2 C) at temperatures perature
>1100C. This gives you a true value with minimal fluctuation. DISP HOLD Manually freeze a display as long as the
A special high temperature insertion tube is drilled to your key is depressed
exact specifications to maintain optimum contact between the SLOPE IN Used to store selected slope rates into
sensor and the dry block. memory
o
C/oF Select desired temperature readout
There are two models: the standard Model 1200SE and the scale
Model 1200SE-RS with RS232C serial data communications. MEMO Store or recall SET temperature or
The Model 1200SE-RS is supplied with RS232C serial data SLOPE rate (Up to three)
communications and calibration software allowing for auto- SLOPE Activates slope mode
mated temperature calibration from a remote personal com- SWITCH HOLD Used to determine set point of tempera-
puter. Options also exist that allow the user to convert RTD, ture switches
Thermocouple and transmitters into a temperature reading mA MEMO IN Used to display mA measurement from
on the personal computer. Units may be specified with a 0 to 24mA
separate heating/cooling tower. TEMP SET Used to increase or decrease tempera-
ture in 0.1 degree increments
SET ZERO Fixed memory for set temperature
Fast, Efficient Heating and Cooling Times Users select the calibration requirements needed including:
Model 1200SE Series calibrators feature fast heating and cool-
ing times. The instrument, including the insertion tube, reaches q Sensor
the maximum heating temperature (2192 oF/1200 oC) from q Heating Source
73.4oF (23oC) in just 35 minutes. q Scenario
q Reference Indicator
Instrument cooling time is also efficient. The instrument q External Reference Sensor
o o
reaches maximum cooling temperature (572 F/300 C) from q Input Device
o o
2192 F (1200 C) in 20 minutes. q Temperature Points/Procedure
Serial Data Interface Both calibration procedures and results are stored in a built-in
The Model 1200SE-RS calibrator is equipped with RS232C identification/tag number database. Procedures may be que-
serial data communications. Software is supplied with this ried by the user with results presented in a real certificate for-
option. Simply connect the calibrator with your personal com- mat. Certificate formats may be customized by the user includ-
puter, place sensor in thermowell, connect sensor to calibra- ing the ability to automatically number your certificate. The
tor, fill-in the “windows” displayed, and start your test. The certificate may be displayed at any time.
software allows you to print, save and depict actual test data in
a graphical format. Other features include input scaling which automatically con-
verts mA into oF/oC. You may also present information graphi-
Calibration Software cally and view in real-time the calibration trends being created.
Using the calibrator’s RS232 serial data port, the instrument Trend information may be stored for later analysis.
may be interfaced with a personal computer running AMETEK’s
AMECAL-T Plus application software. This WINDOWS-based Self Diagnostics
software is easy to use and provides the user with the ability to The Model 650SE Series calibrators feature extensive diag-
perform automated calibration with documented results and a nostics that help identify and isolate faults. Using the front
printed certification. display, the instrument indicates using error codes, diagnostic
conditions that may be impeding the instrument’s operability.
AMECAL-T Plus software presents the user with a hierarchy of Typical faults and their displays are:
operations that effectively guides you through the setup, cali-
bration and certification processes. Calibration is greatly sim- ___0 No RS232 or damaged interface
plified and may be completed in relatively short time periods. ___1 No RS232 contact with personal computer
Calibration requirements are selected by the user through icons, -100.0 RTD sensor open or short circuit
pulldown menus and extensive prompts. Users may select E0007 Program execution failure
heating sources such as liquid baths, ovens or dry-block cali- E0008 Memory access failure
brators. You may also select between an internal or external E0009 Timer malfunction
reference such as the AMETEK DTI Digital Temperature Indi-
cator so that extremely high resolution and accuracy are The instruments also perform a variety of self diagnostics at
achieved. Plus the AMECAL-T Plus software supplies you with startup including a semi-automatic switch test and hysteresis
a way to create and print customized certificate formats which test, validation tests for indicators and 4-20mA transmitter
comply with Quality Standards such as ISO 9000, GMP, HACCP measurement.
and others.
HEATING
Max. slope/minute 113oF/45oC 86oF/30oC 68oF/20oC
Min. slope/minute 1oF/1oC 1oF/1oC 1oF/1oC
COOLING
Max. slope/minute 41oF/5oC 48oF/9oC 68oF/20oC
Min. slope/minute 1oF/1oC 1oF/1oC 1oF/1oC
FUNCTIONAL SPECIFICATIONS PHYSICAL SPECIFICATIONS
Accessories Description
Note: All Model 1200SE Series Dry Block Temperature Calibrators are supplied with NIST Certification Data.
AMETEK is a leading global manufacturer of electrical and electromechanical products for niche
markets. Listed on the New York Stock Exchange (AME) since 1930, AMETEK’s annual sales are
approaching $1billion. Operations are in North America, Europe and Asia, with about one-third of
TEST AND CALIBRATION INSTRUMENTS sales to markets outside the United States.
Headquarters AMETEK Denmark A/S AMETEK Precision AMETEK Singapore Pvt. Ltd.
8600 Somerset Drive Gydevang 32-34 Instruments Europe GmbH 10 Ang Mo Kio Street 65
Largo, Florida 33773 Post Office Box 30 Rudolf-Diesel-Strasse 16 #05-12 TECHPOINT
Tel (727) 536-7831 DK-3450 D-40670, Meerbusch Singapore 569059
Tel (800) 527-9999 Allerod Denmark Germany Tel 65 484 2388
Fax (727) 539-6882 Tel 45 4816 8000 Tel 49 2159 9136 0 Fax 65 481 6588
Fax 45 4816 8080 Fax 49 2159 9136 39 E-Mail aspl@ametek.com.sg
Internet Addresses:
www.ametek.com Information within this document is subject to change without
www.chatillon.com notice.
ISO 9001
AMETEK is a registered trademarks of AMETEK, Inc. Manufacturer
Pub Code SS-CP-2230-0898
Issued 08/98 Copyright 1998, by AMETEK, Inc. Printed in U.S.A.