Professional Documents
Culture Documents
UC 212
Thermal Analysis Application
Application published in METTLER TOLEDO Thermal Analysis UserCom 21
Experimental details
As “raw data”, model free kinetics re- Figure 2. Comparison of MFK and AMFK: Left: Initial data for MFK and AMFK. Although MFK
and AMFK clearly yield different activation energy curves (above right), the predictions made
quires at least three DSC or TGA curves with the two procedures do not significantly differ (below right).
measured with different temperature
programs. the end of the reaction under investiga- the same way. In particular, the sample
First of all, it is always best to make a tion. In such cases, lower heating rates mass should be about the same for all
trial run in order to first get a general should be used. samples. And obviously the crucibles
impression of the “reaction profile”, As mentioned above, isothermal temper- used must not react with the sample.
that is, the temperature range, reaction ature programs can be used for AMFK. As Furthermore, it should be noted that the
enthalpy and mass loss. Relatively slow a rule, the temperatures used are below size of the hole in the cru-cible lid could
heating rates (e.g. 5 K/min) should be the peak temperature. The reason for influence reactions in which gases are
used for materials that release large this is that above the peak temperature liberated.
amounts of energy during the reaction the reaction is very rapid, which makes
(e.g. explosives). For safety reasons it is correct evaluation of the curves difficult. Evaluations
advisable to use high-pressure crucibles A fundamental problem with isothermal The typical revaluation procedure used
and low sample masses (< 5 mg). With DSC experiments is that at the beginning for MFK and AMFK is illustrated in the
other materials, heating rates of 10 or of the experiment the initial deflection example shown in Figure 1
20 K/min can be used for the trial run. and chemical reaction overlap. In this (for DSC measurements).
case, the DSC curve of a second run of
For MFK, a minimum of three experi- the same sample (which should then no Step 1 (Fig. 1 above left): Display of
ments must be performed at different longer show any reaction) measured un- DSC measurement curves. As already
heating rates. Four or five measurements der otherwise identical conditions can be mentioned, the peak maximum shifts
are how-ever recommended to improve used as the baseline. Just as with MFK, at to higher temperatures at increasing
the relia-bility of the calculated predic- least three measurements must be made heating rates. The reaction enthalpy
tions. The heating rates used should with different temperature programs for however is independent of the heating
increase in large steps such as 1, 2, 5, 10, AMFK. Four measurements are however rate and should therefore be constant
and 20 K/min. The use of smaller steps, also recommended for AMFK; dynamic within about ±10%. At first sight, this
for example 5, 6, 7, 8, 9 K/min, reduces measurements can be combined with does not appear to be the case in the
the accuracy of the results. At increasing isothermal measurements. With AMFK, heat flow curves displayed (the area
heating rates, chemical reactions take the rate at which the measurement under the reaction peak is clearly larger
place at higher temperatures. Accord- values are recorded should be set to the at high heating rates than at low heat-
ingly, the temperature range must match maximum value of 10 measurement val- ing rates). The reason for this is that
the heating rate. Particularly when high ues per second. the reaction enthalpy is evaluated from
heating rates are used, for example with As far as sample preparation is con- the curve plotted as a function of time,
curing reactions, (undesired) decompo- cerned, it is important to make sure that whereas in Figure 1 the curves are shown
sition reactions frequently occur toward all the samples are prepared in exactly as a function of temperature. It should
METTLER TOLEDO
TA Application No. UC 212 2
also be noted that the measured reac-
tion enthalpy depends strongly on the
type of baseline used. As far as possible
the same baseline type should be used
for all measurements. If the reaction
Thermal Analysis Application
METTLER TOLEDO
TA Application No. UC 212 3
Conclusions of view, the two approaches differ in the activation energies. However, with regard
Model free kinetics allows even complex numerical algorithms used. to predictions, MFK and AMFK give very
chemical reactions to be analyzed using similar results.
just a few DSC or TGA measurements and The main difference for the user is that
without making any assumptions about MFK can only evaluate heating curves, Publishing Note:
Thermal Analysis Application
e
a particular reaction model. The STAR but AMFK can in addition evaluate iso- This application has been published in
software has implemented two equivalent thermal or combinations of isothermal the METTLER TOLEDO Thermal Analysis
UserCom No. 21.
procedures (MFK and AMFK software and dynamic measurements. Both pro- See www.mt.com/ta-usercoms
options). From the mathematical point cedures usually yield somewhat different