You are on page 1of 11

d e n t a l m a t e r i a l s 3 4 ( 2 0 1 8 ) 1531–1541

Available online at www.sciencedirect.com

ScienceDirect

journal homepage: www.intl.elsevierhealth.com/journals/dema

Effect of high-intensity curing lights on the


polymerization of bulk-fill composites

Mindy M. Daugherty a,1 , Wen Lien b , Michael R. Mansell a ,


Douglas L. Risk b,2 , Daniel A. Savett b , Kraig S. Vandewalle c,∗
a United States Army, 36014 Wratten Loop, Ft. Hood, TX 76544, USA
b United States Air Force, 3650 Chambers Pass, Ft. Sam Houston, TX 78023, USA
c United States Air Force, 1615 Truemper St., Joint Base San Antonio — Lackland, TX 78236, USA

a r t i c l e i n f o a b s t r a c t

Article history: Objectives. The purpose of this study was to investigate the effect of high-irradiance light-
Received 2 April 2018 curing-units (LCUs) on the depth-of-cure (DoC) and degree-of-polymerization (DoP) of bulk-
Received in revised form fill composites (BFCs).
25 May 2018 Methods. The DoC of composites (Beautifil-Bulk, SHOFU; Filtek-Bulk-Fill, 3 M ESPE; Tetric-
Accepted 7 June 2018 EvoCeram-Bulk-Fill, Ivoclar; Sonic-Fill-2, Kerr; Venus-Bulk-Fill, Heraeus; Z250, 3M-ESPE)
were measured according to ISO-4049 using high-irradiance LCUs (FlashMax-P3, CMS-
Dental; SPEC3, Coltene) and conventional LCU (Paradigm, 3M-ESPE) for exposure times: 3/9-s,
Keywords: 3/20-s, and 10/20-s respectively. Using FTIR, the DoP per composite was measured at the bot-
High-intensity curing lights tom surface as a function of post-curing times for the LCUs at the same exposure times. Data
Depth of cure was analyzed with nonlinear regression and ANOVA/Tukey.
Degree of polymerization Results. Significant differences in DoC were found amongst the LCUs for the various expo-
Bulk-fill composites sure times. All BFCs failed to meet the DoC claimed by manufacturers and failed to satisfy
ISO-4049 with the high-irradiance LCUs with 3-s exposures. Standard irradiance and 20-s
exposures outperformed all other irradiance-exposure combinations for maximizing the
DoC and DoP of BFCs. A minimum of 15.3 J/cm2 radiant exposure was required to achieve
an adequate maximum polymerization rate. Venus Bulk exhibited the highest DoC and DoP
for any LCU-exposure-time combination.
Significance. Among the different combinations of BFCs and LCUs, DoC and DoP were always
increased with longer exposure time, but there exists a theoretical radiant-exposure limit
beyond which DoP or DoC remains unchanged. However, high DoC or DoP are not always
associated with one another. Thus, the exposure-reciprocity law must be approached
thoughtfully since irradiance and exposure can independently affect DoP and DoC.
© 2018 The Academy of Dental Materials. Published by Elsevier Inc. All rights reserved.


Corresponding author.
E-mail addresses: mindy.m.daugherty.mil@mail.mil (M.M. Daugherty), wen.lien.mil@mail.mil
(W. Lien), michael.r.mansell.mil@mail.mil (M.R. Mansell), drisk@amdpi.com (D.L. Risk),
daniel.a.savett.mil@mail.mil (D.A. Savett), kraig.s.vandewalle.civ@mail.mil (K.S. Vandewalle).
1
Present address: United States Army, 10590 Enduring Freedrom Dr., Ft. Drum, NY 13602, USA.
2
Present address: Private Practice, 2992 Waldorf Marketplace, Waldorf, MD 20603, USA.
https://doi.org/10.1016/j.dental.2018.06.005
0109-5641/© 2018 The Academy of Dental Materials. Published by Elsevier Inc. All rights reserved.
1532 d e n t a l m a t e r i a l s 3 4 ( 2 0 1 8 ) 1531–1541

LCU can properly initiate polymerization within the recom-


1. Introduction mended curing time. Today, the standard in most dental
practices is the LED (light-emitting diode) LCU. The LED LCU
Dentistry has witnessed significant innovation of resin-based
exploits the electronic nature between two dissimilar semi-
restorative materials in the last three decades. The success
conducting materials to produce light. They differ from the
of this ongoing evolution appears to have been promul-
quartz-tungsten-halogen (QTH) LCUs in which light produc-
gated from new treatment strategies based on scientific
tion is based on the scientific principle of electroluminescence
evidences and practical constraints imposed by our clinical
rather than incandescence [6]. In comparison to QTH LCUs, the
practices, patient demands, and empirical research. Espe-
advantages of LED LCUs are high photonic output in relatively
cially for restorative dentistry, a paradigm shift toward light
short exposure time, energy efficiency, and narrow spectral
cured materials and away from amalgams is driven partially
output [7,8].
by the post-handling of mercury and its environmental con-
The recent introduction of LED LCUs with “high power
cerns, partly by the gradual clinical success of RBCs, and by
and short exposure” has raised concerns regarding their
the increasing patient demand for esthetics. RBC is presently
role, effectiveness, and safety in operative dentistry [9,10].
a routine armamentarium for dental operative procedures
Questions have arisen among dental professionals as to
with approximately 146 million RBCs placed annually in the
whether the Bunsen–Roscoe concept of exposure reciprocity
United States [1,2]. Placing these RBCs is an intricate process
[11] can be applied to aid in the acceleration of dental photo-
that often poses many challenges such as adequate isola-
polymerization when using these new LCUs. There has been a
tion, etch-and-bond treatment of dentin and enamel surfaces,
discursive controversy around extrapolating the total energy
incremental or bulk placement, and adequate light curing
principle to curing of the dental RBCs [12–15]. The assump-
(photo-polymerization). Proper execution of these steps can
tion is that the photo-response of a photo-sensitive RBC only
minimize iatrogenic errors and can enhance the longevity of
depends on the total absorbed energy dose (radiant exposure
the restoration. However, the role of the dental LCU in photo-
or H, unit: J/cm2 ) rather than the two components, irradiance
polymerization has often been overlooked, perhaps because
and exposure time, which their product determines radiant
of its presumed simplicity in application relative to all of the
exposure [11]. Furthermore, these extrinsic factors can be
other challenging requirements when restoring a tooth.
complicated by the different manufacturers’ claims on the
Photo-polymerization is an important process because it
bulk-fill RBC depth of cure. Most are publicized to be bulk
serves as a “rate-limiting” factor, controlling the outcome of
placed and cured in 4 mm increments, and few are cured in
a RBC restoration. For example, a RBC requires a synergis-
5 mm increments [16]. Although reducing the time required
tic combination of intrinsic and extrinsic factors to achieve
to cure these materials has short-term, cost-saving implica-
photo-polymerization efficiency [3]. The former includes the
tions, the quality, integrity, and performance of the restoration
filler content/morphology, monomer ratios/chemistry, or pho-
remains the primary concern. Improperly polymerized RBCs
toinitiator concentrations/types, and the latter is involved
can result in insufficient monomer-to-polymer conversion,
with curing parameters such as exposure times, source-
a predisposing factor to clinical problems such as marginal
to-object distance, or standard-versus-high irradiance mode
discrepancy, secondary caries, and fracture, whereas a prop-
[2]. Although controlling, manipulation, and formulation of
erly cured composite will exhibit good physical properties in
intrinsic factors can have a direct and significant impact on the
strength, wear, and toughness [17].
final physical properties of a RBC, they are very much depen-
The present study aims to investigate the effect of
dent on the extrinsic attributes. It has been shown that the
high-irradiance LCUs on the depth-of-cure (DoC) and degree-
energy and spectral wavelengths of a LCU must match that of
of-polymerization (DoP) of bulk-fill RBCs. The null hypotheses
the RBC’s manufacturer photo-polymerization requirements
were:
[4,5]. Without proper delivery of the required photon ener-
gies by the LCU, the predictive course of photo-polymerization
reaction can be jeopardized in which suboptimal quantum (1) There are no significant differences in DoC and DoP
yields and monomer-to-polymer conversions are incurred and amongst various commercial bulk-fill RBC brands cured
cascaded across each of the photopolymerization kinetic steps by a LCU having a high irradiance (i.e., ≥2000 mW/cm2 )
such as free-radical initiation, chain-propagation, and chain- with either a short (≤10 s) or an ultra-short exposure
termination mechanisms. The aberrant products from these time (≤5 s) versus a LCU having a conventional irradiance
inchoate steps are ultimately reflected in the final outcomes (<2000 mW/cm2 ) and standard exposure time (20 s).
of the RBC’s physical properties and clinical performance [3]. (2) Given the same radiant exposure (energy dose), there is
Furthermore, of these two factors, the former is governed by no significant difference in DoC and DoP amongst various
the proprietary manufacturing formulations which are neither commercial bulk-fill RBCs.
changeable nor controllable by the end user. In contrast, the
latter is most susceptible to operator error, particularly involv-
2. Materials and methods
ing their light-delivery techniques. There are other extrinsic
factors like LCU’s beam profile, spectral emission, and irra-
2.1. LCU and bulk-fill RBC characteristics
diance, which are strictly set forth by the manufacturer as
well.
Three LED LCUs, having different irradiances, were studied:
When a curing light tip is placed over a resin-based
FlashMax-P3 (CMS Dental, Copenhagen, Denmark), Paradigm
composite, the dentist must be confident that the chosen
(3M ESPE, St. Paul, MN, USA), and SPEC3 (Coltene, Cuyahoga
d e n t a l m a t e r i a l s 3 4 ( 2 0 1 8 ) 1531–1541 1533

Falls, OH, USA). Table 1a shows the manufacturer salient char-

Standard = 1600 ± 10%


1200 ± −10% to +20%
acteristics for each of the three LCUs. These LCUs were used

and 3 K = 3000–3500
5000–6000
to polymerize five bulk-fill RBCs and one non-bulk-fill RBC in
different combinations. This study follows the ISO-4049 [18]

(mW/cm2 )
irradiance
Claimed
in defining a non-bulk-fill or conventional RBC as a RBC with
a manufacture’s claimed depth of cure of less than or equal
to 3 mm. Table 1b shows the manufacturer salient character-
istics for each of the RBCs. The standard or main mode of
FlashMax-P3 and of Paradigm was selected for exposure times
of 3/9 s and of 10/20 s respectively. For SPEC3, both the main

wavelength
(nm) peak
and 3K modes were applied to cure RBCs for the exposure

Claimed

455 ± 10

455–465
times of 3 s and 20 s respectively. To validate the manufacturer
claimed irradiance, an integrating sphere (sphere Ø = 15 cm

460
and entry port Ø = 19 mm; Lapsphere, North Sutton, NH, USA),
coupled to a calibrated spectrophotometer (USB4000-UV–VIS,

wavelength
(nm) range
Ocean Optics, Dunedin, FL, USA), was used to measure the

450–470
Claimed

430–480

430–490
irradiances of the three LCUs. The measured irradiance was
determined by dividing the integration of the spectral radiant
power (mW/nm) as a function of wavelength (nm) over the
entire wavelength range by the active area of the LCU tip. The

Time to fully
diameter of the LCU tip was measured with a digital microme-
ter (±0.1 mm). Respectively, the light guide diameter were 0.72,

2h

2h
charge
0.78, and 0.91 cm for SPEC3, FlashMax-P3, and Paradigm; for

2–3 h
Paradigm and SPEC3, the outer rim thickness of the light guide
(0.08 cm) was negated. Also, irradiance measurements were
made using a 4 mm diameter aperture at the entrance of the
integrating sphere to determine the estimated amount of irra- Battery type
diance delivered to a 4 mm diameter cylindrical specimen. The

Li-Ion

Li-Ion

Li-Ion
outputs of the 3 LCUs were measured 3 times each, after which
the mean and standard deviation were calculated. For each
combination of LCU’s irradiance mode and exposure time, the
Continuous 1 s or 3 s

between each 1 s or
radiant exposure (J/cm2 ), a measure of photon dose, was deter-

3 K: 1 s, 2 s, 3 s, and
ortho = continuous

between each 3 s
mode = 120 s and
with 0.5 s pause

mined from multiplying the measured irradiance (mW/cm2 )


mode options

tack-cure mode
Other curing

3 s with pause
3 s exposure

by the exposure time.


Continuous

exposure
2.2. Depth of cure (DoC)

The DoC was measured according to the ISO-4049 [18]. Sam-


ples were made for each composite brand using stainless steel
5 s, 10 s, 15 s, or 20 s
Standard (main)

(SS) molds. All SS molds were 4 mm in diameter but varied in


exposure time

5 s, 10 s, or 15 s
curing mode

height. The determination for which mold height to be used


was based on each RBC manufacturer’s claimed DoC and on
1 s or 3 s

the formula, mold height = 2 × (manufacturer’s claim DoC) + 2,


as specified by the ISO-4049 [18]. Six RBC brands (5 bulk-fills
and 1 conventional) and three LCU brands were evaluated.
Cordless or corded

Table 1b showed the manufacturer claimed DoC for each of


Power option

the RBC brands. Once a proper mold was selected and placed
Cordless
Table 1a – LCU brands and features.

onto a Mylar strip on a glass microscope slide, it was filled


Cordless

with the test RBCs, prepared in accordance with the man-


ufacturer’s instructions. Care was taken to exclude any air
bubble. Next, another Mylar strip was placed over the sample,
followed by a second microscope slide. After finger pressure
LED

LED

LED

was used to displace excess materials, the second microscope


slide was removed. Then, the exit aperture of the LCU light
(manufacturer)

tip was positioned concentrically to the mold opening and


SPEC 3 (Coltene)
& light source

(CMS Dental)

Paradigm (3M

directly against the Mylar surface. Immediately after curing,


FlashMax P3

the sample was removed from the mold. Any remaining soft
Product

ESPE)

uncured RBC was manually scraped away using a plastic spat-


ula. The maximum height of the remaining hard RBC was
measured three times with a digital caliper (±0.1 mm). The
1534 d e n t a l m a t e r i a l s 3 4 ( 2 0 1 8 ) 1531–1541

three measurements were averaged and divided by two to

Universal

Bis-EMA (bisphenol-A polyethylene glycol diether dimethacrylate), Bis-GMA (bisphenol-A diglycidyl ether dimethacrylate), Bis-MPEPP (bisphenol-A polyethoxy-dimethacrylate), DMA (dimethacrylate),
Universal
yield DoC. Based on the aforementioned ISO-4049 method [18],

A2
Shade

the DoC was measured in triplicate (n = 3) for each of the vari-

IV A
A2

A2
ous RBC + LCU combinations. For each combination, a new and
un-polymerized RBC was cured by using a LCU that was either
of the two high-irradiance LCUs (FlashMax-P3 or SPEC3) or a

40–3000 nm (mean
conventional LCU (Paradigm) for the exposure times of 3/9 s,

20–5000 nm
10–3500 nm
(20 nm/4–11 nm)

EBADMA (ethoxylated bisphenol-A-dimethacrylate), S-PRG (surface modified pre-reacted glass), TEGDMA (triethylene glycol dimethacrylate), UDMA (urethane dimethacrylate).
3/20 s, or 10/20 s respectively.
Cluster/non-
aggregated
Filler size

25–400 nm

550 nm)
2.3. Degree of polymerization (DoP)
N/A

A Fourier transform infrared attenuated total reflection (FTIR-


ATR) spectrometer (Spotlight 400, PerkinElmer, Waltham, MA,

Ba-Al-F-Si glass, YbF3, SiO2


USA) was used to measure the DoP per RBC brand at the
Zirconia/silica bottom surface (specimen thickness = 4 mm, Ø = 10 mm) as a
Zirconia/silica, YbF3

(monomer, glass, &

function of post-curing times for Paradigm, FlashMax-P3, and


pre-polymer filler
Filler

YbF3), spherical
(F-B-Al-Si glass)

SPEC3. RBC was placed following manufacturer’s instruction


Ba-Al-Si glass,
SiO2 , Ba-glass

mixed oxide

inside a SS mold (height = 4 mm, Ø = 10 mm) with direct con-


S-PRG filler

tact to the ATR crystal. Care was taken to exclude any air
bubbles. Next, a Mylar strip was placed over the sample, fol-
lowed by a microscope slide. After finger pressure was used to
displace excess materials, the microscope slide was removed.
Then, the exit aperture of the LCU light tip was positioned con-
wavelength

centrically to the mold opening and directly against the Mylar


440–490

400–500

400–520

400–500

400–500
Curing

surface. Depending on which LCU brand was applied, each


(nm)

N/A

new un-polymerized RBC sample received an exposure time


of either 10 or 20 s for Paradigm, either 3 or 9 s for FlashMax-
P3, or either 3 or 20 s for SPEC3, which produced 6 different
Bis-MPEPP,TEGDMA

Bis-GMA, TEGDMA,

combinations of RBC brand, LCU brand, and exposure time.


12-dodecane-DMA
Bis-GMA, UDMA,

Bis-GMA, UDMA,
UDMA, EBADMA

For each combination, real-time recording of the infrared (IR)


Bis-GMA, DMA

spectrum (2500–600 cm−1 ) began immediately after curing and


UDMA and 1,
chemistry

continued at steady intervals for 7 min (0, 15, 35, 55, 70, 90,
Bis-EMA

Bis-EMA
Matrix

105, 120, 180, 240, 300, 420 s), then again after 24 h. At each of
the collecting intervals, 4 scans (8 cm−1 resolution per scan)
were averaged to produce an IR spectrum, and 14 IR spectra
(including the baseline uncured spectrum) per each combi-
nation were recorded. The DoP (%) was determined by the
53–54
74.5

58.4

68.5

38.0
60.0
Filler wt% &

following equation:

  A1636  
vol.%

76–77
Table 1b – Resin-based composite brands and features.

87.0

76.5

83.5

65.0
82.0

A1609 polymer
DoP = 100 1 −  A1636 
A1609 monomer
4.0

5.0

5.0

4.0

4.0
2.5

where A1636 and A1609 represented the heights of the IR absorp-


tion intensities at 1636 cm−1 (C C of aliphatic bond stretch)
Ivoclar Vivadent
Manufacturer

Heraeus Kulzer

and 1609 cm−1 (C C of aromatic bond stretch)


and claimed

 respectively
 A1636   A1636 
DOC (mm)

and where / represented the


3M-ESPE

3M ESPE

A1609 polymer A1609 monomer


SHOFU

absorption-peak ratio of the aliphatic to aromatic stretch


Kerr

of the polymer normalized by the absorption-peak ratio of


aliphatic to aromatic stretch of the monomer.
Resin-based composite

Tetric EvoCeram Bulk Fill

2.4. Statistics and data analysis


Z250TM (Control)

Data were analyzed with nonlinear regression


  and
Beautifil Bulk

Sonic FillTM 2
FiltekTM Bulk

d(DoP)
Venus-Bulk

ANOVA/Tukey (˛ = 0.05). The maximum DoP rate, dt


,
®

max
defined as the maximum rate at which DoP changed with
respect to post-curing times, was calculated from the
following [19–22]:
d e n t a l m a t e r i a l s 3 4 ( 2 0 1 8 ) 1531–1541 1535

Fig. 1 – Effect of three commercial LCUs’ radiant exposures (J/cm2 ) on six RBC brands’ DoCs. Each bar represents an average
DoC of three samples per RBC brand and per radiant exposure (n = 3) from which for each of the radiant exposures, a mean
DoC of the six RBC brands’ DoCs (n = 18) was calculated and depicted as black squares on the figure.

(1) Each DoP-post curing time curve per RBC brand & radi- gorized: standard, short, and ultra-short exposure times to
ant exposure (n = 36) was fitted with a sigmoidal function, be 20 s, 9–10 s, and 3 s respectively. Using these 3 different
LCU brands and their various curing combinations, 6 radiant
a
DoP (t) = −(t−t0 )
exposures (7.1, 9.0, 12.3, 21.4, 24.5, and 36.5 J/cm2 ) were cal-
1+e ⁄b
culated from the 6 combinations of irradiances and exposure
where t = time, t0 = time at which sigmoidal reached maximum times (2.378 × 3, 3.001 × 3, 1.226 × 10, 2.378 × 9, 1.226 × 20, and
derivative, and a & b were the fitting constants. 1.827 × 20 J/cm2 ) respectively.
d(DoP(t))
(2) Once a, b, and t0 were found, the term, dt
, repre-
senting the rate of polymerization (Rp ), was used to obtain
  3.2. Influence of radiant exposure on DoC and DoP
d(DoP) d(DoP(t))
the maximum Rp or dt
value by solving dt
at t0 ,
a
  max
Figs. 1 and 2 show how successive radiant exposures can
which is 4b .
Then, for each set of 6 RBCs’ maximum Rp values that impact the DoC and DoP measurements of the 6 RBC brands.
were attained from exposing to the same radiant exposure, a Comparing amongst radiant exposures (black dashed line),
mean maximum Rp value per radiant exposure was calculated. the same upper-case letter per radiant exposure is not signifi-
These mean maximum Rp values were plotted as a function cantly different than each other (p > 0.05). Comparing amongst
of radiant exposure (J/cm2 ). RBC brands within each radiant exposure, the same lower-
case letter per RBC brand is not significantly different than
each other (p > 0.05). These differences indicated that the DoC
3. Results and DoP of bulk-fill RBCs were dependent more on material
brand and less on radiant exposure for high values of radiant
3.1. Irradiance exposure conditions.
Comparison between the insets on Figs. 1 and 2 showed the
Significant differences (p < 0.05) in mean irradiances mean irradiance values to conduct DoC measurements were
were found amongst the 3 LCU brands. LCU rankings of much higher than those of DoP measurements. These differ-
mean irradiances from highest to lowest were: SPEC3 (3 s ences in mean irradiance values can be best clarified from
mode = 3001 ± 8 mW/cm2 ), FlashMax-P3 (2378 ± 22 mW/cm2 ), the mathematical construct of irradiance, which is the rate
SPEC3 (main mode = 1827 ± 5 mW/cm2 ), and Paradigm of energy passage per unit area, orthogonal to the direction of
(1226 ± 0.3 mW/cm2 ). For this study, we categorized: (1) energy transport. This mathematical ratio demonstrates irra-
SPEC3 (3 s mode) and FlashMax-P3 as high irradiance diance is inversely proportional to the energy transport area.
(≥2000 mW/cm2 ) group and (2) SPEC3 (main mode) and Consequently, the smaller the active area, the larger the irra-
Paradigm as the conventional irradiance (<2000 mW/cm2 ) diance value. For example, the mean irradiance delivered to
group. The consensus is any LCU, having >1500 mW/cm2 each DoC specimen (see Fig. 1 inset) was calculated via divid-
irradiance, is considered high intensity [23]. We also cate- ing the total spectral radiant power output by the specimen’s
1536 d e n t a l m a t e r i a l s 3 4 ( 2 0 1 8 ) 1531–1541

cross sectional area, which was 0.13 cm2 for Ø = 0.40 cm. Here, go through the origin to ensure the assumption that at zero
the specimen’s cross sectional area was smaller than the exit deliverance of radiant exposure, the mean maximum Rp was
aperture of the LCU’s light tip and was limiting the amount also zero. To estimate the minimum radiant exposure, we
of photo-energy that could be transported from the light tip. used Ilie and Durner approach [24], where lines of best fit
Therefore, the specimen’s cross sectional area was used as were inserted through the first and last points of the approxi-
the “normalizing” active area to calculate irradiance. In con- mate function. The point, obtained from the projection of the
trast, the mean irradiance delivered to each DoP specimen (see best fit lines’ intersection, was determined to be the mini-
Fig. 2 inset) was calculated via dividing the total spectral radi- mum radiant exposure that is required to achieve an adequate
ant power output by the cross sectional, active area per LCU’s maximum Rp . The intersection of these lines is considered
light tip. Here, the specimen’s cross sectional area (Ø = 1 cm) to estimate the inflection point of the curve, approximat-
was slightly larger than the LCU’s light tip. Hence, the area of ing the glass transition state, which describes the motions
energy transport was adopted from the cross sectional active of monomeric segments and polymeric chains as they tran-
area of the LCU’s light tip. Because the total active area for the sit from a viscoelastic matrix to an elastoplastic network [24].
DoC specimen was smaller than total active area of the DoP A minimum of 15.3 J/cm2 is required to lengthen the RBCs’
specimen, given the same amount of exposure time, the irradi- glass-transition kinetics after which Rp ascends to its asymp-
ance (total spectral radiant power per unit area) received by the totic limit, where it remains steady and independent of radiant
DoC specimens were relatively more than the DoP specimens. exposure when approaching infinity. Alternatively, the x-axis
Even though there are large differences between the insets value of the inflection point could also be determined by taking
on Figs. 1 and 2, Fig. 3 shows radiant exposure, a product of the first derivative of the sigmoidal function, which resulted a
irradiance and exposure time, is more correlative to DoC and minimum radiant exposure of 9.0 J/cm2 . Comparing the two
DoP than the irradiance parameter by itself. When different minimum radiant exposure values, the method of Ilie and
irradiances in combination with various exposure times are Durner is more conservative.
converted to radiant exposures, Fig. 3 illustrates that in gen- As shown in Fig. 5, irrespective of various radiant exposures
eral, the higher the radiant exposures ascended, the faster the supplied, Venus-Bulk showed the highest DoC (p < 0.05), while
rate of change in DoC or DoP detected, but then the growth Sonic Fill 2 exhibited the lowest DoC (p < 0.05) when compared
leveled off at high radiant exposures. to all other bulk-fill RBC brands in this study. This further
Fig. 4 illustrates the mean maximum Rp plotted against demonstrates material brand dependence. However, DoP val-
its respective radiant exposure (here, the time of dark-cure ues of Venus-Bulk and Sonic Fill 2, measured at the bottom
was immediately measured beginning from the cessation of surface of a 4-mm specimen thickness, are not significantly
radiant exposure to post one day). Each point in Fig. 4 rep- different than each other and are ranked the highest (p < 0.05)
resents the mean of 6 RBCs’ maximum Rp (5 bulk-fill and 1 in comparison to all other bulk-fill RBC brands, whose DoP val-
conventional) per radiant exposure. The relationship between ues were also measured at an equivalent depth of 4 mm (Fig. 5).
mean maximum Rp and radiant exposure was best fitted using Interestingly, even though Sonic-Fill-2 has a high DoPs, its DoC
a sigmoidal nonlinear regression. The curve was forced to

Fig. 2 – Effect of three commercial LCUs’ radiant exposures (J/cm2 ) on six RBC brands’ DoPs. Each bar represents an average
DoP of one day of dark-cure measurements per RBC brand and per radiant exposure (n = 12) from which for each of the
radiant exposures, a mean DoP of the six RBC brands’ DoPs (n = 72) was calculated and depicted as black squares on the
figure.
d e n t a l m a t e r i a l s 3 4 ( 2 0 1 8 ) 1531–1541 1537

Fig. 3 – DoP or DoC in relationship with radiant exposure.

is one of the lowest significant values in comparison with the RBCs. The DoP of bulk-fill RBCs under the curing combina-
other five RBC brands. tion of standard irradiance & exposure is not significantly
The ISO-4049 states that DoC measurement should not be different than the high irradiance & short exposure combina-
below 0.5 mm from the values claimed by the manufacturer tion. However, the curing combination of standard irradiance
of the tested product using the recommended curing times. & exposure did significantly outperform the other two com-
In this study, Beautifil Bulk, Venus Bulk, Tetric EvoCeram Bulk, binations (standard irradiance & short exposure and high
and Z250 all met the standard and were within the range of irradiance & ultra-short exposure).
the manufacturers’ claimed DoC when using the control LCU,
Paradigm (1226 mW/cm2 ) for 20 s. Filtek Bulk and SonicFill 2
each claim 5 mm DoC, and both fell below the ISO-4049 stan-
dard at 3.6 mm and 3.2 mm respectively. 4. Discussion

Within the limits of this study, our null hypotheses were


3.3. Influence of irradiance and exposure combination rejected since: (1) significant differences were found for the
on DoC and DoP DoC and DoP amongst the various commercial bulk-fill RBC
brands cured by an LCU having a high irradiance and either
Fig. 6 demonstrates that light curing protocol with low irradi- a short or an ultra-short exposure time versus an LCU having
ance & long exposure combination significantly outperforms a conventional irradiance and standard exposure time; and,
all other combinations for maximizing the DoC of bulk-fill (2) there were significant differences in DoC and DoP amongst

Fig. 4 – Relationship between mean maximum Rp and radiant exposure.


1538 d e n t a l m a t e r i a l s 3 4 ( 2 0 1 8 ) 1531–1541

Fig. 5 – Comparison amongst various bulk-fill RBC brands against themselves and a non-bulk-fill RBC (Z250, Control). Each
black bar represents a mean DoP, which was calculated from averaging DoPs (n = 72) of the six radiant exposures per RBC
brand; and, each gray bar represents a mean DoC, which was calculated from averaging DoCs (n = 54) of the six radiant
exposures per RBC brand. The same case letters across columns are not significantly different than each other (p > 0.05).

various commercial bulk-fill RBCs when using the same radi- these studies are dependent on one crucial premise that deliv-
ant exposure (energy dose). ering a minimum radiant exposure not only can adequately
cure but also can generate clinically relevant strength.
For comparison, this study has found that as a “rule of
4.1. Influence of radiant exposure on bulk-fill RBCs thumb”, based on the mean maximum polymerization rate
as a function of radiant exposure (Fig. 4), a minimum radiant
There is still much to learn about the complex interactions exposure of 14 J/cm2 is required to adequately cure a bulk-fill
between photopolymerization and light curing parameters RBC. Furthermore, our findings provided supportive evidence
(e.g., radiant exposure, irradiance, and exposure duration). to the claims [14,15,30] that the Bunsen–Roscoe reciprocity law
Today, there is no adoption of a standard that recommends a does not apply under conditions involving high irradiance and
minimal radiant exposure beyond which this threshold should short exposure duration. In practical terms, we have observed
be adequate to meet the energetic demands of any photo- that the range of irradiance-and-exposure combination during
cured RBC, resulting in clinically relevant properties. However, which the greatest sensitivity (i.e., maximum polymerization
the current consensus [25] is that there exists a minimum rate) exists varies according to the brand of bulk-fill RBC. For
amount of radiant exposure and exposure duration that are example, two different RBC brands in this study (Venus-Bulk
resin-based material specific and that must be given along and Beautifil-Bulk at 24.5 J/cm2 ), given equal amounts of radi-
with the correct wavelengths to attain acceptable outcomes. ant exposure, produced different DoP (see Fig. 2).
Furthermore, “not all LCUs are created equal” [25]. Although bulk-fill RBCs are claimed to have improved DoC
Historically, clinical practice suggests a minimum radiant (on average up to 4 mm), saved time, and reduced the poten-
exposure of 12–24 J/cm2 to adequately cure a 2 mm thick RBC tial for entrapment of air and containments, they are like
[26–28]. This evidence is based on surveying the QTH LCUs any RBCs, not perfect [31]. Their photo-energetic demands
and conventional RBCs since LED technology and bulk-fill for polymerization are complex. Caution must be exercised to
RBC were not available at the time. For example, Rueggeberg ensure that when restoring with these bulk-fill RBCs, an ade-
et al. found that a combination of at least 400 mW/cm2 and quate amount of radiant exposure is delivered. The radiant
60 s (equating to a radiant exposure of 24 J/cm2 ) was required exitance, if not controlled properly, can produce deleterious
to adequately achieve a 2 mm DoC [26]. Comparatively, the events, consequently leading to increased risk of premature
ISO 10650:1999 recommended the measured irradiance of the RBC failures. Furthermore, having an improved DoC neither
QTH LCU to be above 300 mW/cm2 , which under an exposure buffers for a reduction in curing times nor provides a sense of
duration of 40 s, this equated to a 12 J/cm2 radiant exposure complacency for the clinicians to ignore standard curing pro-
[27]. Similarly, Fan et al. was able to show that for most con- tocols by which the false assumption is that RBC with greater
ventional RBCs, a radiant exposure of one half smaller than DoC are impervious to radiant exposure effects or equates to
24 J/cm2 (or 12 J/cm2 ) was capable to satisfy the minimum, improved DoP. For instance, this is best observed in Fig. 5,
1.5 mm DoC requirements, in accordance to the ISO-4049 [28]. where there is no observable relationship between DoC and
As LED technology evolved over the years, data have shown DoP.
that a QTH LCU having a comparable irradiance output of a LED Additionally, our findings support the views of a previous
LCU requires at least twice the amount of exposure duration study by Ilie and Stark, which were: (1) short exposure and
to achieve comparable biaxial flexure strength [29]. In general,
d e n t a l m a t e r i a l s 3 4 ( 2 0 1 8 ) 1531–1541 1539

Fig. 6 – Effect of irradiances and exposures on the mean DoP and DoC of five bulk-fill RBCs and one non-bulk-fill RBC (Z250,
control). The same case letter across the columns is not significantly different than each other (p > 0.05). The blue and red
bars refer to the DoP and DoC axes respectively.

high irradiance resulted in significantly lower DoC for bulk-fill first viewpoint is based on the assumption that RBC reacts to
RBCs as compared to long exposure and low irradiance; (2) an light curing in a viscoelastic manner during which the network
increase in radiant exposure resulted in an increase in DoC; formation of a RBC can be modeled by a superposition of two
and (3) the minimum radiant exposure to achieve a 4 mm DoC exponentials [24]. This double exponentials function describes
was in the range of 6–24 J/cm2 [32]. the transitional kinetics from a gel phase to a glassy phase
[24]. The second viewpoint is based on the curing process and
4.2. Fitting function its categorization into two stages, primary cure and post-cure
[15]. And respectively, the fitting function accounts for these
Past studies have employed several fitting approaches stages by separating them into two time-dependent factors:
to describe the kinetic behavior of photopolymerization exponential and logarithmic [15]. The difference between the
[15,19–22,24]. Of those approaches, there are two fitting func- two viewpoints is the latter ensures that at time zero, when
tions, reflecting two different points of view, of which each photo-irradiation is absent, DoP values are best fitted by start-
argues in favor of its empirical evidence on providing clarifi- ing at the origin.
cation to the onset of RBC network formation. Both of these In this study, a sigmoidal function was used to fit the
fitting functions involve the summation of two reactions and kinetic curve of DoP versus post-cure time and also to cor-
claim that a function, composed of a double component, can relate the mean maximum Rp against radiant exposure. The
better approximate the physical features of network forma- significance of the sigmoidal curve is threefold [19]: (1) at the
tion than the traditional, one component fitting function. The immediate onset of post-cure time, the slow-to-rise, nonlinear
1540 d e n t a l m a t e r i a l s 3 4 ( 2 0 1 8 ) 1531–1541

response of the kinetic curve describes an induction period at perhaps, the curing protocol may need to consider having an
which oxygen is known strongly to inhibit free-radical photo- appropriate amount of radiant exposure such that its radiant
polymerization; (2) once the oxygen consumption has reached temperature is negligible to affect pulpal irritation. Further
its peak as post-curing time extends, free radicals cascade study on the effect of radiant exposure and pulpal temperature
with less impediment, culminating in exponential growth of rise is required.
polymerization; and, (3) when post-curing time further length-
ens, the overall reaction slows as the result of several possible
reaction kinetics: (a) progressive depletion of free radicals and 5. Conclusion
reactive double bonds and (b) rising rate of increasing molec-
ular weight of the chains that have not yet been terminated, The clinical applications of this study emphasize the impor-
causing restrictions to the chains’ mobility (Trommsdorff or tance of material and equipment selection. It is imperative
gel effect) [19]. that clinicians understand the properties of composite mate-
Naturally, there are pros and cons to each of the afore- rials chosen, whether standard or bulk-fill, and modify the
mentioned fitting approaches. It may suffice to point out here placement procedure as necessary. Just as importantly, clin-
that the application of a sigmoidal fitting function has a dis- icians should be familiar with the chosen LCU and its features
tinctive mathematical advantage. Unlike an exponential or and realize that each brand of LED LCU is designed to be
logarithmic fitting function, the sigmoidal fitting function for used differently. There is not a “one size fits all” dental LCU
the standard curve of DoP versus time will always yield a available that one can operate the same way for all material
derivative with an absolute maximum from which the rate types.
of polymerization per time (Rp ), as well as the maximum Rp , The DoC and DoP are governed by three interrelated but
can be calculated and can then be plotted as a function of independent fundamental light-curing parameters: radiant
DoP. Another option is to use, the fitting function consisted exposure, irradiance, and exposure duration. Bulk-fill compos-
of the two time-dependency factors, to model our data, but ites cured with high-irradiance and short/ultra-short exposure
our measurements were solely based on the post-cure phe- time may not provide adequate DoC and DoP, which can lead
nomenon. We did consider the fitting approach that uses only to undesirable clinical properties. Irrespective of bulk-fill RBC
the logarithmic time-dependency factor, which was intended brands and LCUs, long exposure time always produced highest
for modeling the post-cure stage, to approximate our results. DoC and DoP. However, there exists a theoretical radiant expo-
However, the disadvantage of the logarithmic fitting function sure limit above which DoP or DoC plateaus. High DoC does
is that the Rp cannot be approximated from the first derivative not always associate with high DoP. Cautions must be taken
of a logarithmic function. Further study on which of the fitting when exercising exposure reciprocity law since irradiance and
approaches are best to characterize the photopolymerization exposure can independently affect DoP and DoC.
kinetics may be necessary.
references
4.3. Limitations

This study has several limitations. First, the DoP and DoC [1] American Dental Association. ADA 2005-06 survey of dental
measurements do not yield information about the mechanical services rendered; 2007. Chicago, IL.
properties of the bulk-fill RBCs in response to high irradiance [2] Price RB, Shortall AC, Palin WM. Contemporary issues in
and short exposure. However, based on past studies, which light curing. Oper Dent 2014;39(1):4–14.
have demonstrated that a direct correlation existed between [3] Leprince JG, Palin WM, Hadis MA, Devaux J, Leloup G.
hardness and DoP [33], this study approximates that as the Progress in dimethacrylate-based dental composite
technology and curing efficiency. Dent Mater
extent of DoP advances during post-cure, bulk-fill RBC hard-
2013;29(2):139–56.
ness must also increase. Second, more studies are needed to [4] Price RB, Felix CA. Effect of delivering light in specific narrow
explain the phenomena of contractional stress during and bandwidths from 394 to 515 nm on the micro-hardness of
after polymerization under the influence of high-irradiance resin composites. Dent Mater 2009;25(7):899–908.
and short exposure. Earlier studies have found that greatly [5] Nomoto R. Effect of light wavelength on polymerization of
increased radiant exposures are linked to significant rises in light-cured resins. Dent Mater J 1997;16(1):60–73.
[6] Jandt KD, Mills RW. A brief history of LED
polymerization stress but do not affect DoP significantly after
photopolymerization. Dent Mater 2013;29(6):605–17.
a certain threshold has been met [34,35]. Third, the rise of pul-
[7] Leonard DL, Charlton DG, Roberts HW, Cohen ME.
pal temperature when teeth are exposed to prolonged high Polymerization efficiency of LED curing lights. J Esthet
irradiance is an important concern, which this study did not Restor Dent 2002;14(5):286–95.
assess. A recent study has shown that consecutive curing [8] Vandewalle KS, Roberts HW, Andrus JL, Dunn WJ. Effect of
cycles under various irradiance protocols have the potential light dispersion of LED curing lights on resin composite
to generate enough heat to surpass the 5.5 ◦ C temperature polymerization. J Esthet Restor Dent 2005;17(4):244–54.
[9] Kutuk ZB, Gurgan S, Hickel R, Ilie N. Influence of extremely
threshold, inducing irreversible pulpal damage [36]. Based on
high irradiances on the micromechanical properties of a
our data, we have found that there exists a threshold after nano hybrid resin based composite. Am J Dent
which the amount of radiant exposure plays less of a signifi- 2017;30(1):9–15.
cant role in DoP amplification. Since supplying more radiant [10] Feng L, Carvalho R, Suh BI. Insufficient cure under the
exposures than the threshold augments no additional benefits condition of high irradiance and short irradiation time. Dent
and serve only to increase unnecessary pulpal temperature, Mater 2009;25(3):283–9.
d e n t a l m a t e r i a l s 3 4 ( 2 0 1 8 ) 1531–1541 1541

[11] Bunsen RW, Roscoe HE. Photochemische untersuchungen. [25] Roulet JF, Price R. Light curing – guidelines for practitioners –
Ann Phys 1859;108(2):193. a consensus statement from the 2014 symposium on light
[12] Halvorson RH, Erickson RL, Davidson CL. Energy dependent curing in dentistry held at Dalhousie University, Halifax,
polymerization of resin-based composite. Dent Mater Canada. J Adhes Dent 2014;16(4):303–4.
2002;18(6):463–9. [26] Rueggeberg FA, Caughman WF, Curtis Jr JW. Effect of light
[13] Peutzfeldt A, Asmussen E. Resin composite properties and intensity and exposure duration on cure of resin composite.
energy density of light cure. J Dent Res 2005;84(7):659–62. Oper Dent 1994;19(1):26–32.
[14] Hadis M, Leprince JG, Shortall AC, Devaux J, Leloup G, Palin [27] ISO 10650. Dental equipment—powered polymerization
WM. High irradiance curing and anomalies of exposure activators. Geneva, Switzerland: International Organization
reciprocity law in resin-based materials. J Dent for Standardization; 1999.
2011;39(8):549–57. [28] Fan PL, Schumacher RM, Azzolin K, Geary R, Eichmiller FC.
[15] Selig D, Haenel T, Hausnerová B, Moeginger B, Labrie D, Curing-light intensity and depth of cure of resin-based
Sullivan B, et al. Examining exposure reciprocity in a resin composites tested according to international standards. J
based composite using high irradiance levels and real-time Am Dent Assoc 2002;133(4):429–34.
degree of conversion values. Dent Mater 2015;31(5): [29] Rueggeberg FA, Cole MA, Looney SW, Vickers A, Swift EJ.
583–93. Comparison of manufacturer-recommended exposure
[16] Van Ende A, De Munck J, Lise DP, Van Meerbeek B. Bulk-fill durations with those determined using biaxial flexure
composites a review of the current literature. J Adhes Dent strength and scraped composite thickness among a variety
2017;19(2):95–109. of light-curing units. J Esthet Restor Dent 2009;21(1):43–61.
[17] Demarco FF, Corrêa MB, Cenci MS, Moraes RR, Opdam NJ. [30] Leprince JG, Hadis M, Shortall AC, Ferracane JL, Devaux J,
Longevity of posterior composite restorations: not only a Leloup G, et al. Photoinitiator type and applicability of
matter of materials. Dent Mater 2012;28(1):87–101. exposure reciprocity law in filled and unfilled photoactive
[18] ISO-4049. Dentistry—polymer-based filling, restorative and resins. Dent Mater 2011;27(2):157–64.
luting materials. Geneva, Switzerland: International [31] Leprince JG, Palin WM, Vanacker J, Sabbagh J, Devaux J,
Organization for Standardization; 2009. Leloup G. Physico-mechanical characteristics of
[19] Decker C, Decker D. Photoinitiated radical polymerization of commercially available bulk-fill composites. J Dent
vinyl ether-maleate systems. Polymer 1997;38(9):2229–37. 2014;42(8):993–1000.
[20] Trujillo M, Newman SM, Stansbury JW. Use of near-IR to [32] Ilie N, Stark K. Curing behaviour of high-viscosity bulk-fill
monitor the influence of external heating on dental composites. J Dent 2014;42(8):977–85.
composite photopolymerization. Dent Mater [33] Ferracane JL. Correlation between hardness and degree of
2004;20(8):766–77. conversion during the setting reaction of unfilled dental
[21] Pereira SG, Telo JP, Nunes TG. Towards a controlled restorative resins. Dent Mater 1985;1(1):11–4.
photopolymerization of dental dimethacrylate monomers: [34] Calheiros FC, Braga RR, Kawano Y, Ballester RY. Relationship
EPR studies on effects of dilution, filler loading, storage and between contraction stress and degree of conversion in
aging. J Mater Sci: Mater Med 2008;19(9):3135–44. restorative composites. Dent Mater 2004;20(10):939–46.
[22] Nunes TG, Ceballos L, Osorio R, Toledano M. Spatially [35] Asmussen E, Peutzfeldt A. Polymerization contraction of
resolved photopolymerization kinetics and oxygen resin composite vs. energy and power density of light-cure.
inhibition in dental adhesives. Biomaterials Eur J Oral Sci 2005;113(5):417–21.
2005;26(14):1809–17. [36] Runnacles P, Arrais CA, Pochapski MT, Dos Santos FA, Coelho
[23] Shortall AC, Felix CJ, Watts DC. Robust spectrometer-based U, Gomes JC, et al. In vivo temperature rise in anesthetized
methods for characterizing radiant exitance of dental LED human pulp during exposure to a polywave LED light curing
light curing units. Dent Mater 2015;31(4):339–50. unit. Dent Mater 2015;31(5):505–13.
[24] Ilie N, Durner J. Polymerization kinetic calculations in dental
composites: a method comparison analysis. Clin Oral
Investig 2014;18(6):1587–96.

You might also like