Pump Maintenance

Cleaning or Exchanging the Outlet Ball Valve


You should change the outlet ball valve as part of your regular maintenance routine –
or when the valve is leaking internally. Typically, if the pressure ripple is unstable,
you should run a leak test to determine if the outlet ball valve needs to be replaced.
1. Remove the capillary from the valve.
Outlet ball valve parts 2. Use a 14 mm wrench to remove the valve.
3. Remove the plastic cap and gold seal.
4. Clean the ball valve in an ultrasonic bath. (Or, replace with a new valve).
5. Replace the gold seal and cap.
6. Use a 14 mm wrench to reinstall the valve.

Outlet Ball Valve


Description Part No.

Outlet ball valve, (quaternary, isocratic) G1311-60012


Outlet ball valve for binary pump, 400 bar G1312-60012
Gold seal, outlet 5001-3707
Outlet caps, 4/pk 5062-2485
Binary pump outlet valve SS sieve, 10/pk 5063-6505

Tips & Tools


The outlet ball valve of the binary
pump has an additional sieve
(5063-6505).