Professional Documents
Culture Documents
Wei-Jen Lee Zhenyuan Zhang Shiuan-Hau Rau Tammy Gammon Ben Johnson James Beyreis
Fellow, IEEE Student Member, IEEE Student Member, IEEE Senior Member, IEEE Life Fellow, IEEE Fellow, SFPE
Energy Systems Research Center John Matthews & Associates Thermon WHMS, UL.Inc
University of Texas – Arlington P. O. Box 322 P. O. Box 609 Life Member, NFPA
Arlington, TX 76019 Cookeville, TN 38502 San Marcos, TX 78666 Northbrook, IL 60062
Abstract -- Electrical arcs emit extremely high intensity light Though eye shields or glasses can block UV radiation, high intensity
across a wide spectrum that can damage eye delicate structures, visible light during the arc flash incident can still damage the eyes.
such as the cornea or the retina. Developing industry safety Proper selection and use of eye protection is the best way to
standards to mitigate light hazards is a critical component of arc mitigate the hazard posed by the bright light generated during an arc
flash safety. flash. Accurate measurement of light intensity is the first step in
Accurate measurement of light intensity is the first step in understanding the level of eye protection needed. However, due to
understanding the level of eye protection needed. Traditionally, the dynamic nature of an arc flash blast, it is difficult to provide a
light measurement devices are an assembly of light detecting precise description and reliable measurement of light intensity
resistors made from cadmium sulfide (CdS) or silicon (Si) cells. during the arc flash event.
However, their response times are inadequate for capturing the This paper discusses a systematic light measurement technique to
peak light intensity or the dynamic changes during the initial overcome the limitations of traditional approaches. The following
arc flash. In addition, the brightness of an arc flash is greater sections below will: 1) discuss light levels capable of causing harm
than the measurable light levels from traditional light to the eyes and the mechanisms of injury; 2) explain the selection of
measurement devices. Furthermore the light sensitivity of the components, circuit design, and filter characterization and
traditional sensors does not match the Commission
system integration; and 3) present sample test results.
Internationale de l'Éclairage (CIE) luminosity function
V (λ ) curve, which defines the human eye’s sensitivity to bright
II. BASICS OF ARC FLASH LIGHT MEASUREMENT
light as a function of wavelength under typical ambient light
conditions. A number of research projects were conducted by the U.S.
This paper discusses a novel light measurement technique to Armed Forces on flash blindness in 1960s. The research was fueled
overcome the limitations of traditional approaches. This design by the need to know how well service personnel, especially pilots,
is capable of measuring the light intensity of the arc flash as could perform after a nuclear blast. The recovery time of flash
perceived by human eyes. These results can be used to evaluate blindness depends on the light intensity, direction, duration, and size
the potential of an arc flash to harm the eyes. of the flash; pupil size; age of the subject; individual variation; and
the spectrum of light radiated by the flash. Flash blindness is the
Index Terms—light measurement, arc flash, ambient light temporary loss of vision when the retina is subject to high-intensity
sensor, neutral density filter, operational amplifier light. A reduction in visual acuity can last a few minutes to a few
days. Contributing factors are the glare, the afterimage, and the
I. INTRODUCTION bleaching of the photochemical substances within the rods and cones
An arc flash is a flashover of electric current through air from one of the retina. Scotomatic glare occurs when extremely bright light
exposed live conductor to another conductor or ground in electrical fatigues the retina. The retina’s sensitivity is reduced, and therefore,
equipment. The most obvious arc flash hazard is excessive heat so is its ability to respond normally to light [5] [6].
exposure and serious burn injuries. In addition to burn injuries, the 2.1 Light Spectrum Range
blast pressure can cause permanent hearing loss and intense light The electromagnetic spectrum is the frequency range of all types
can cause blindness [1]. of electromagnetic radiation. Radiation over the range of
The intense light generated by an arc flash can be dangerous; eye wavelengths between 380 nm and 760 nm is detected by the human
injuries account for about 4.2 percent of all arc flash-related injuries eye and perceived as visible light. From previous tests, in an arc
[2] [3]. Temporary blindness due to sudden bright light, known as flash explosion, the radiated light falls in the visible, UV, and IR
flash blindness, is a common occurrence. High-intensity light in the spectrums, and is predominately in the range between 400 to 600 nm
visible, infrared (IR), and ultraviolet (UV) spectrums can damage [7]. Figure 1 shows the spectrum band of visible light, as well as the
important structures such as the cornea or retina. entire electromagnetic spectrum.
Due to these potentially harsh conditions, work is being
conducted to establish industry standards for the appropriate
personal protective equipment when electrical hazards are present
[4]. Personal protective equipment (PPE) is required by OSHA and
NFPA standards to protect workers from hazards in the workplace.
Those standards include headgear requirements, which involve a
non-conductive helmet, a UV-blocking shield and safety glasses.
100
80
Eo = 1Lux
S is sensitivity, and S = 10 μ A -20
200 400 600 800 1000 1200 1400 1600 1800 2000
Logarithmic output is advantageous in measuring light. Linear Wavelength (nm)
output detectors such as phototransistors cannot detect the small Figure 6. ND4 (1) transmittance corresponding to wavelength
variations in intensity at low levels of illumination. With a
logarithmic response, the small changes in light levels can induce a
detectable response in the sensor, whereas at higher light intensities,
only relatively large variations need to be considered.
3
100 Since the input resistance of the DAQ isless than infinite, the
output voltage of the sensor will be altered due to the loading effect.
For example, the reported input resistance of the DAQ system was
80
6680Ω in one of the laboratories. Consequently as shown in Figure
9, the output voltage of the sensor circuit is much lower than the
60 open circuit value. Though one can recover the actual number if the
input resistance of the DAQ is given, additional complexity was
% Transmittance
0.3
20
0.25
0
0.2
-20
200 400 600 800 1000 1200 1400 1600 1800 2000
Voltage (V)
0.15
Wavelength (nm)
Figure 7. ND4 (2) transmittance corresponding to wavelength 0.1
The response curves illustrate that bothND4 (1) and ND4 (2) have
relatively uniform attenuation in the visible light spectrum. 0.05
However, each ND filter has unique transmittance characteristics.
For example, the two tested ND4 filters have the same spectral 0
response, but have different transmittance rates in the same
spectrum range. In the range of visible light, ND4 (1) and ND4 (2) -0.05
have attenuation factors of 7.257 and 6.452, respectively. To ensure 0 50 100 150 200 250
the accuracy of the ND filter utilized in light measurement tests, the Time (ms)
frequency response of every neutral density filter was characterized Figure 9. Sensor measurement with manufacturer-suggested circuit
in the laboratory. Before adopting a light measurement system, each (actual light intensity is 16.57 million lux)
filter was characterized to identify the transmittance rate. The final
measured light intensity also needed to be calibrated to actual 4.2.2 Improved Design with Buffer Circuit
values. To eliminate the loading effect, a buffer amplifier was designed
Based on the published light intensity measurements (higher than to transfer the circuit from high output impedance to low output
1 million lux), two ND8 and one ND128 were used to extend the impedance.
brightness measurement range of the sensor up to 1 billion lux. In order to have a good performance, the buffer circuit should
have high input impedance and low output impedance. In the
4.2 Circuit Design improved design, a high input impedance operational amplifier (OP
Amp) was used to perform the core function of the buffer circuit.
4.2.1 Sensor Manufacturer- Suggested Circuit
The improved light measurement system circuit is shown in Figure
Since the selected sensor uses current I out as output, it can be 10.
directly transformed into voltage by employing a load resistor RL ,
which is convenient for data collection through a data acquisition
system (DAQ). Figure 8 provides the manufacturer-suggested
operating circuit. A 100 k Ω resistor is adopted as the load
resistance in the circuit.
4
4
3.5
2.5
Voltage (V)
1.5
Calorimeters
1
0.5
-0.5
0 50 100 150 200 250 300 350 400 450 500
Time (ms)
Figure 11. Sensor measurement with improved circuit
2
Voltage (V)
1.5
0.5
-0.5
0 50 100 150 200 250 300 350 400 450 500 Figure 18. Using laser pointer to align light measurement system
Time (ms)
V. RESULTS
2.5
2
Voltage (V)
1.5
0.5
Figure 19. Snap shot of arc flash from high speed video
0
6
As expected, the figure shows that light intensity increases when [12] Wei-Jen Lee, Tammy Gammon, Zhenyuan Zhang, Ben Johnson, and
the measurement distance is reduced. In addition, the value of the Sue Vogel, “IEEE/NFPA Collaboration on Arc Flash Phenomena
Research Project”, IEEE-PES, T&D Expo, Orlando, FL. May 2012
bolted fault current also significantly affected the arcing light
[13] Light Sensors, [Online]. Available: http://www.electronics-
intensity. According to test records, the light intensity at 3m utorials.ws/io/io_4.html
exceeded 200 million lux when the bolted fault current reached [14] Christine Rüth, Andreas Vogler, Wilhelm Karsten, “Ambient Light
40kA at 14.3kV. Sensors General Application Note” , 06/29/2006
[15] Construction and Characteristics of CdS Cells, [Online]. Available:
http://www.selcoproducts.com/pdfs/Photocell%20Contstruction%20an
d%20Characteristics.pdf
[16] Christine Rüth, Andreas Vogler, “Application Note for High accuracy
Ambient Light Sensor”, 8/18/2006
[17] J.C. Das, “Arc Flash Hazardanalysis And Mitigation”, John Wiley &
Sons, Inc., pp 306-314, 2012
[18] SFH5711 Datasheet, http://www.osram-os.com/
[19] Hanke, Rudolph (1979). Filter-Faszination. Monheim/Bayern. pp. 70
ff. ISBN 3-88324-991-2
[20] Camera Lens Filters, [Online]. Available:
http://www.cambridgeincolour.com/tutorials/camera-lens-filters.htm
Figure 20. Light intensity measured in 2.7kV HOA test with 10kA,
20kA and 30kA short circuit current, and calorimeters installed on
27 inches
VI. CONCLUSION
Many arc flash tests have been performed in different labs (with
different DAQ systems), the proposed light measurement system
proves to be durable, reliable and precise. The novel system
presented in this paper overcame the limitations of traditional
approaches. Since the selected light sensor closely matches the
V (λ ) photopic curve, the measurement results could realistically
represent the arcing light as perceived by human eyes. Moreover
with the ND filter, the measurable brightness level range of the
sensor system was greatly improved. The measurement results from
the proposed design provide a foundation for light hazard analysis
and industry standards development.
REFERENCES
[1] Chet Davis, Conrad St. Pierre, David Castor, Rober Luo, Satish
Shrestha, “Practical Solution Guide to Arc Flash Hazards”, ESA, Inc., ,
pp 1-14, 2003
[2] NFPA/IEEE Research and Testing Planning Committee (RTPC), Final
Report, July 28, 2005.
[3] T. Gammon, unpublished analysis of OSHA records, “electric arc” key
term, released between March 2005 and Sept. 2008.
[4] NFPA 70E-2014
[5] J. D. Teresi, “A review of research on flash blindness,” USNRDL-TR-
68-76, Radiological Defense Laboratory, San Francisco, California,
July 1, 1968.
[6] M. Luckiesh, L. L. Holladay “Glare and Visibility”, U Transactions of
the Illuminating Engineering Society 1925; 20: 221–252.
[7] R. A. Wilson, R. Harju, J. Keisala, S. Ganesan, “Tripping with the
speed of light: arc flash protection,” Proceedings of the 60th Annual
Conference for Protective Relay Engineers, pp. 226-238, College
Station, Texas, Mar. 27-29, 2007
[8] Electro Magnetic Spectrum and light, [Online]. Available: http://9-
4fordham.wikispaces.com/Electro+Magnetic+Spectrum+and+light
[9] Crawford (1949) The scotopic visibility function, Proceedings of the
Physical Society, B62, 321-334.
[10] CIE Proceedings (1951), Vol. 1, Sec 4; Vol 3, p. 37, Bureau Central de
la CIE, Paris, 1951
[11] Wyszecki, G., & Stiles, W. S. (1982). Color Science: concepts and
methods, quantitative data and formulae. (2nd ed.). New York: Wiley.
7