Professional Documents
Culture Documents
Estimation of PAM From The Oscillometric Cuff Pressure - Comparision of Dif Tecniques
Estimation of PAM From The Oscillometric Cuff Pressure - Comparision of Dif Tecniques
DOI 10.1007/s11517-010-0694-y
ORIGINAL ARTICLE
Received: 10 June 2010 / Accepted: 7 October 2010 / Published online: 2 November 2010
International Federation for Medical and Biological Engineering 2010
Abstract Mean arterial pressure (MAP) is determined in deviation of paired differences (3.7 mmHg). These values
most automated oscillometric blood pressure devices, but from the peak of the largest oscillometric pulse were -1.3
its derivation has been little studied. In this research, dif- and 6.2 mmHg, respectively. Determining MAP from a
ferent techniques were studied and compared with the model of the oscillometric pulse waveform had the smallest
auscultatory technique. Auscultatory systolic and diastolic differences from the manual auscultatory technique.
blood pressure (SBP and DBP) were obtained in 55 healthy
subjects by two trained observers, and auscultatory MAP Keywords Blood pressure Cuff pressure Mean arterial
was estimated. Automated MAP was determined by six pressure (MAP) Non-invasive blood pressure (NIBP)
techniques from oscillometric cuff pressures recorded Oscillometric pulse
digitally and simultaneously during manual measurement.
MAPs were derived from the peak and foot of the largest
oscillometric pulse, and from time domain curves fitted to 1 Introduction
the sequence of oscillometric pulse amplitudes (4th order
and three versions of the 6th order polynomial curve). The Measurement of blood pressure is one of the most common
agreement between automated and auscultatory MAPs was and important clinical and diagnostic measurements made
assessed. Compared with the auscultatory MAP, the auto- by family doctors and hospital physicians [3]. Accurate
mated MAP from the baseline cuff pressure at the peak of blood pressure measurement is very important, with pub-
the 6th order polynomial curve had the smallest mean lished data estimating that with a 5 mmHg overestimation,
paired difference (-1.0 mmHg), and smallest standard millions of people would receive an inaccurate diagnosis of
hypertension and take unnecessary anti-hypertension
medication, as well as increase unnecessary costs to
D. Zheng A. Murray healthcare providers [13]. A systematic error of 5 mmHg is
Regional Medical Physics Department, Freeman Hospital, also the absolute maximum systematic error allowed in
Newcastle University, Newcastle upon Tyne, UK assessing blood pressure devices [12].
Systolic and diastolic blood pressures (SBP and DBP)
J. N. Amoore
Department of Medical Physics, Royal Infirmary of Edinburgh, are important in assessing cardiovascular status, but there is
Edinburgh, UK also clinical interest in mean arterial pressure (MAP),
defined as the average pressure throughout the cardiac
S. Mieke
cycle. MAP is considered to be the driving pressure for
Physikalisch-Technische Bundesanstalt, Abbestrasse 2-12,
10587 Berlin, Germany perfusion of most vital organs, which should be at least
60 mmHg [4]. MAP has physiological and clinical
D. Zheng (&) importance, and can be used as a predictor of cardiovas-
Cardiovascular Physics and Engineering Research Group,
cular risk [21]. For women in pregnancy, MAP has been
Regional Medical Physics Department, Freeman Hospital,
Newcastle University, Newcastle upon Tyne NE7 7DN, UK reported as a better predictor for pre-eclampsia than SBP
e-mail: dingchang.zheng@ncl.ac.uk and DBP [7]. MAP is also useful in intensive care, where it
123
34 Med Biol Eng Comput (2011) 49:33–39
can often be directly measured from invasive arterial changes in the oscillometric pulses in the cuff pressure, and
pressure, for assessing hemodynamic variables to guide hence destroy the smoothness of the oscillometric wave-
treatment [10]. form envelope, leading to difficulties in accurate BP
The true MAP is calculated from the invasive blood determination. Therefore, developing a technique which
pressure curve, but invasive measurements are generally can reliably determine MAP is important.
not clinically acceptable. Traditionally, non-invasive MAP The aim of this study was to develop different tech-
is calculated by using an empirical formula, in which MAP niques to estimate automated MAPs from oscillometric
approximately equals DBP plus 33% of the pressure dif- waveforms and to compare these estimations of MAP with
ference between SBP and DBP [11]. The value of ‘33%’ is the auscultatory derived method.
referred to here as the manual mean weight for calculating
MAP. The gold standard for non-invasive SBP and DBP
measurement has been readings taken by a trained observer 2 Methods
using a mercury sphygmomanometer and the Korotkoff
sound technique [17]. MAP derived from this classic 2.1 Subjects
manual technique is regarded as a useful measurement.
Mean arterial pressure is also determined in most auto- Fifty-five normal healthy subjects, with no known cardio-
matic, non-invasive blood pressure measurement devices vascular disease, were studied. There were 39 male and 16
(NIBP) using the oscillometric technique [8, 18], even female subjects, with ages in the range 20–74 years. The
where it is not displayed. Those automated devices are detailed subject information including age, sex, height,
used frequently in many health care institutions [3]. weight and arm circumference are summarized in Table 1.
Technically, automated oscillometric devices analyse the This study received ethical permission, and all subjects
pressure pulse changes (oscillometric pulses) induced in a gave their written informed consent.
pressurized cuff wrapped round the limb during deflation.
These changes are caused by the pulse radiating down the 2.2 Blood pressure measurement protocol
artery producing pressure changes in the cuff, which are
expected to be greatest at MAP [18]. This feature has been Blood pressure measurements were undertaken in a quiet
used to determine MAP in these oscillometric blood pres- room. The subject was seated in a chair with their feet on
sure devices. Manufacturers of automated devices then the floor and with their arm supported at heart level. There
devise their own algorithms by adding additional infor- was a 5-min rest period before formal recording to allow
mation, such as characteristic ratios of the pulse amplitude cardiovascular stabilization.
to the maximum pulse amplitude, to estimate SBP and Figure 1 shows the schematic representation of the
DBP after using MAP in the calculation [8, 24]. measurement protocol. Auscultatory SBP and DBP were
A characteristic of the oscillometric pulses (typically recorded under controlled conditions with a reliable, accu-
amplitude) is used to construct the oscillometric waveform rate and clinically validated electronic sphygmomanometer
envelope which characterizes the variation in oscillometric
pulse amplitude with cuff pressure. Ideally, the oscillo-
metric waveform envelope is a smooth curve with a distinct
peak from which the MAP can be determined. However, Table 1 General data information for the subjects studied
the oscillometric waveform envelope is constructed from Subject information
samples (at the pulse rate) of the changing cuff pressure. In
No. of subjects 55
the majority of oscillometric NIBP devices, often with fast
No. of male 39
deflation rates up to 10 mmHg/s, the number of detectable
No. of female 16
oscillations is reduced in comparison with that from the
No. of 148
recommended deflation rate of 2–3 mmHg/s. Interpolation
measurements
is then necessary to compensate for long periods between
oscillometric pulses. Furthermore, the oscillometric wave- Minimum Maximum Mean SD
form envelope is not always a neat bell-shaped curve with a Age (years) 20 74 39 16
distinct peak [2]. Rather than a distinct peak the oscillo- Height (cm) 152 188 172 9
metric waveform envelope may have a plateau. To further Weight (kg) 46 103 71 11
add to the challenge of determining the MAP, the oscil- Arm circumference (cm) 22 38 28 2
lometric measurement is compromised by movement and SBP (mmHg) 90 140 119 13
other forms of artefact, such as respiratory disturbance DBP (mmHg) 53 90 75 10
[1, 14, 19, 22]. These disturbances are associated with
123
Med Biol Eng Comput (2011) 49:33–39 35
123
36 Med Biol Eng Comput (2011) 49:33–39
2.3 Mean arterial blood pressure determination cuff deflation rate, with a limited number of detectable
oscillometric pulses.
Auscultatory MAP was estimated in each recording from
auscultatory DBP plus 33% of the pressure change from
2.4 Data analysis
DBP to SBP, with supplementary data for the use of 40% in
the formula [5]. Six different techniques were then devel-
For all the blood pressure recordings, SPSS software
oped using Matlab 7.1. (MathWorks Inc., USA) to deter-
package (SPSS Inc., USA) was employed to determine the
mine the corresponding automated MAPs from the cuff
effect of different techniques on MAP estimation, and also
pressure at a specific time determined from oscillometric
to perform regression analysis between the auscultatory
pulse pressure. Details are illustrated in Fig. 3 and descri-
MAPs and automated MAPs from the six different oscil-
bed below:
lometric techniques, obtaining the square of the correlation
• MAP1 and MAP2 were the cuff pressures at the peak coefficients (R2). Bland–Altman analysis (Systat Software
and foot, respectively of the largest oscillometric pulse Inc., USA) was also performed to assess the agreement
[25]. They are the traditional oscillometric techniques between the automated MAPs and auscultatory MAPs,
for MAP estimation, but can be influenced by the with their paired differences and SD of differences calcu-
variation of oscillometric waveform caused by artefact lated. A value of P \ 0.05 was considered statistically
or noise. significant. Finally, the automated mean weights calculated
• MAP3 and MAP4 were the cuff pressures at the peak of from the automated MAPs were compared with the clas-
the pulse which was closest to the peak of the 4th and sical manual mean value of 33%.
6th order fitted polynomial curve, respectively. The
polynomial curve fitting technique was used here
because it has the power to interpolate between 3 Results
the sampled pulses and to smooth out noise in the
oscillometric waveform. Furthermore, because the The overall mean MAPs from all the recordings for the
oscillometric waveform envelope is similar to a bell- auscultatory and oscillometric techniques are given in
shaped curve, an even number of orders above 2 is Table 2. ANOVA analysis showed that different tech-
required to construct a curve with flat features at the niques resulted in significant differences in MAP estima-
beginning and the end. tion (P \ 0.001). The mean paired differences between the
• MAP5 was obtained from MAP4 after correcting for oscillometric techniques and the auscultatory technique
the amplitude difference between the 6th order poly- were significant (all P \ 0.01). Technique 6 with curve
nomial curve and the recorded oscillometric pulse, such fitting improved the paired differences from -1.3 and
that when the polynomial curve lay above the oscillo- -3.4 mmHg (classic technique, without curve fitting) to
metric pulse, the correction added was positive. This -1.0 mmHg. These paired differences were associated
amplitude difference can be caused by respiratory with differences between the manual and automated mean
effects. weights of 4% (33–29%) and 9% (33–24%) for the classic
• MAP6 was the baseline cuff pressure associated with techniques to 2% (33–31%) for technique 6.
the peak of the 6th order fitted polynomial curve, where Figure 4 shows the comparison of MAPs obtained
the baseline was drawn through the foot of all pulses. from the auscultatory and different oscillometric tech-
This technique may be able to correct the effect of fast niques. The linear regression analysis showed that the
123
Med Biol Eng Comput (2011) 49:33–39 37
correlations between the six calculated automated oscillometric pulses (see Fig. 3) with technique 2 associ-
MAPs and the auscultatory MAP were significant (all ated with the lower cuff pressure as it is the pressure at the
P \ 0.001), with the R2 of 0.71, 0.73, 0.84, 0.84, 0.84 foot, rather than the peak of the oscillometric pulse. Sim-
and 0.86, respectively, increasing from techniques 1 to 6. ilarly, the MAPs from technique 3, 4 and 5 were higher
Bland–Altman analysis showed that technique 6 than from technique 6 although the polynomial curve fit-
improved the SD of paired differences best from 6.2 and ting technique was applied to all of them. This difference
5.9 mmHg (classic technique, without curve fitting) to again reflects the amplitude of oscillometric pulse with
3.7 mmHg. technique 3–5, whilst technique 6 used the cuff pressure at
When the mean weight of 40% rather than 33% was the baseline (see Fig. 3). Although the mean MAP differ-
used to calculate the auscultatory MAP, all mean paired ences compared with the auscultatory technique were less
difference in comparison with the auscultatory MAP fell by than 5 mmHg, the systematic mean difference between
an identical 2.9 mmHg with no change to the differences different oscillometric techniques was up to 5.5 mmHg.
between oscillometric techniques. There were small dif- The International Standard requires a maximum systemic
ferences in the SD of paired differences, but not to the difference of 5 mmHg [12]. Furthermore, in a patient
order between oscillometric techniques, with MAP6 still population, the mean differences between the oscillometric
having the lowest variability. techniques and the auscultatory technique can be expected
to be much higher.
Next, the significant differences between MAPs from the
4 Discussion oscillometric techniques and the auscultatory technique
resulted in the automated mean weights calculated from
We have shown that the automated MAPs calculated from automated MAPs being different for all the six techniques
the oscillometric waveform using different measurement studied. Technique 6 produced the smallest mean difference
algorithms were different. The technique associated with (-2%) against the classically assumed mean value of 33%.
the peak of the 6th order polynomial model of the oscil- It is generally accepted that the automated MAP from the
lometric pulse waveform yielded the smallest mean paired oscillometric waveform gives a fairly accurate estimation in
difference in comparison with the auscultatory technique. comparison with the intra-arterial MAP [24], our finding
Technique 1 yielded a higher mean than that from tech- allows us to question whether 33% is a useful mean weight
nique 2. This difference reflects the amplitude of the for achieving accurate MAP estimation. It has been reported
123
38 Med Biol Eng Comput (2011) 49:33–39
123
Med Biol Eng Comput (2011) 49:33–39 39
formula is superior to the standard formula. Catheter Cardiovasc 23. Verrij EA, Vincent HH, Bos WJ (2008) Rule of thumb to
Interv 63:419–425 calculate mean arterial pressure at the brachial artery level.
21. Sesso HD, Stampfer MJ, Rosner B, Hennekens CH, Gaziano JM, J Hypertens 26:1043–1045
Manson JAE, Glynn RJ (2000) Systolic and diastolic blood 24. Yamakoshi K, Shimazu H, Shibata M, Kamiya A (1982) New
pressure pulse pressure and mean arterial pressure as predictors of oscillometric method for indirect measurement of systolic and
cardiovascular risk in men. Hypertension 36:801–807 mean arterial pressure in the human finger. Part 2: correlation
22. Van Horn RN, Kahlke RJ, Taylor LA, Dorsett TJ (2001) Non- study. Med Biol Eng Comput 20:314–318
invasive blood pressure performance: a reproducible method for 25. Zheng D, Murray A (2008) Estimation of mean blood pressure
quantifying motion artifact tolerance in oscillometry. Biomed from oscillometric and manual methods. Comput Cardiol
Instrum Technol 35:395–414 35:941–944
123
Copyright of Medical & Biological Engineering & Computing is the property of Springer Science & Business
Media B.V. and its content may not be copied or emailed to multiple sites or posted to a listserv without the
copyright holder's express written permission. However, users may print, download, or email articles for
individual use.