You are on page 1of 8

Med Biol Eng Comput (2011) 49:33–39

DOI 10.1007/s11517-010-0694-y

ORIGINAL ARTICLE

Estimation of mean arterial pressure from the oscillometric cuff


pressure: comparison of different techniques
Dingchang Zheng • John N. Amoore •

Stephan Mieke • Alan Murray

Received: 10 June 2010 / Accepted: 7 October 2010 / Published online: 2 November 2010
 International Federation for Medical and Biological Engineering 2010

Abstract Mean arterial pressure (MAP) is determined in deviation of paired differences (3.7 mmHg). These values
most automated oscillometric blood pressure devices, but from the peak of the largest oscillometric pulse were -1.3
its derivation has been little studied. In this research, dif- and 6.2 mmHg, respectively. Determining MAP from a
ferent techniques were studied and compared with the model of the oscillometric pulse waveform had the smallest
auscultatory technique. Auscultatory systolic and diastolic differences from the manual auscultatory technique.
blood pressure (SBP and DBP) were obtained in 55 healthy
subjects by two trained observers, and auscultatory MAP Keywords Blood pressure  Cuff pressure  Mean arterial
was estimated. Automated MAP was determined by six pressure (MAP)  Non-invasive blood pressure (NIBP) 
techniques from oscillometric cuff pressures recorded Oscillometric pulse
digitally and simultaneously during manual measurement.
MAPs were derived from the peak and foot of the largest
oscillometric pulse, and from time domain curves fitted to 1 Introduction
the sequence of oscillometric pulse amplitudes (4th order
and three versions of the 6th order polynomial curve). The Measurement of blood pressure is one of the most common
agreement between automated and auscultatory MAPs was and important clinical and diagnostic measurements made
assessed. Compared with the auscultatory MAP, the auto- by family doctors and hospital physicians [3]. Accurate
mated MAP from the baseline cuff pressure at the peak of blood pressure measurement is very important, with pub-
the 6th order polynomial curve had the smallest mean lished data estimating that with a 5 mmHg overestimation,
paired difference (-1.0 mmHg), and smallest standard millions of people would receive an inaccurate diagnosis of
hypertension and take unnecessary anti-hypertension
medication, as well as increase unnecessary costs to
D. Zheng  A. Murray healthcare providers [13]. A systematic error of 5 mmHg is
Regional Medical Physics Department, Freeman Hospital, also the absolute maximum systematic error allowed in
Newcastle University, Newcastle upon Tyne, UK assessing blood pressure devices [12].
Systolic and diastolic blood pressures (SBP and DBP)
J. N. Amoore
Department of Medical Physics, Royal Infirmary of Edinburgh, are important in assessing cardiovascular status, but there is
Edinburgh, UK also clinical interest in mean arterial pressure (MAP),
defined as the average pressure throughout the cardiac
S. Mieke
cycle. MAP is considered to be the driving pressure for
Physikalisch-Technische Bundesanstalt, Abbestrasse 2-12,
10587 Berlin, Germany perfusion of most vital organs, which should be at least
60 mmHg [4]. MAP has physiological and clinical
D. Zheng (&) importance, and can be used as a predictor of cardiovas-
Cardiovascular Physics and Engineering Research Group,
cular risk [21]. For women in pregnancy, MAP has been
Regional Medical Physics Department, Freeman Hospital,
Newcastle University, Newcastle upon Tyne NE7 7DN, UK reported as a better predictor for pre-eclampsia than SBP
e-mail: dingchang.zheng@ncl.ac.uk and DBP [7]. MAP is also useful in intensive care, where it

123
34 Med Biol Eng Comput (2011) 49:33–39

can often be directly measured from invasive arterial changes in the oscillometric pulses in the cuff pressure, and
pressure, for assessing hemodynamic variables to guide hence destroy the smoothness of the oscillometric wave-
treatment [10]. form envelope, leading to difficulties in accurate BP
The true MAP is calculated from the invasive blood determination. Therefore, developing a technique which
pressure curve, but invasive measurements are generally can reliably determine MAP is important.
not clinically acceptable. Traditionally, non-invasive MAP The aim of this study was to develop different tech-
is calculated by using an empirical formula, in which MAP niques to estimate automated MAPs from oscillometric
approximately equals DBP plus 33% of the pressure dif- waveforms and to compare these estimations of MAP with
ference between SBP and DBP [11]. The value of ‘33%’ is the auscultatory derived method.
referred to here as the manual mean weight for calculating
MAP. The gold standard for non-invasive SBP and DBP
measurement has been readings taken by a trained observer 2 Methods
using a mercury sphygmomanometer and the Korotkoff
sound technique [17]. MAP derived from this classic 2.1 Subjects
manual technique is regarded as a useful measurement.
Mean arterial pressure is also determined in most auto- Fifty-five normal healthy subjects, with no known cardio-
matic, non-invasive blood pressure measurement devices vascular disease, were studied. There were 39 male and 16
(NIBP) using the oscillometric technique [8, 18], even female subjects, with ages in the range 20–74 years. The
where it is not displayed. Those automated devices are detailed subject information including age, sex, height,
used frequently in many health care institutions [3]. weight and arm circumference are summarized in Table 1.
Technically, automated oscillometric devices analyse the This study received ethical permission, and all subjects
pressure pulse changes (oscillometric pulses) induced in a gave their written informed consent.
pressurized cuff wrapped round the limb during deflation.
These changes are caused by the pulse radiating down the 2.2 Blood pressure measurement protocol
artery producing pressure changes in the cuff, which are
expected to be greatest at MAP [18]. This feature has been Blood pressure measurements were undertaken in a quiet
used to determine MAP in these oscillometric blood pres- room. The subject was seated in a chair with their feet on
sure devices. Manufacturers of automated devices then the floor and with their arm supported at heart level. There
devise their own algorithms by adding additional infor- was a 5-min rest period before formal recording to allow
mation, such as characteristic ratios of the pulse amplitude cardiovascular stabilization.
to the maximum pulse amplitude, to estimate SBP and Figure 1 shows the schematic representation of the
DBP after using MAP in the calculation [8, 24]. measurement protocol. Auscultatory SBP and DBP were
A characteristic of the oscillometric pulses (typically recorded under controlled conditions with a reliable, accu-
amplitude) is used to construct the oscillometric waveform rate and clinically validated electronic sphygmomanometer
envelope which characterizes the variation in oscillometric
pulse amplitude with cuff pressure. Ideally, the oscillo-
metric waveform envelope is a smooth curve with a distinct
peak from which the MAP can be determined. However, Table 1 General data information for the subjects studied
the oscillometric waveform envelope is constructed from Subject information
samples (at the pulse rate) of the changing cuff pressure. In
No. of subjects 55
the majority of oscillometric NIBP devices, often with fast
No. of male 39
deflation rates up to 10 mmHg/s, the number of detectable
No. of female 16
oscillations is reduced in comparison with that from the
No. of 148
recommended deflation rate of 2–3 mmHg/s. Interpolation
measurements
is then necessary to compensate for long periods between
oscillometric pulses. Furthermore, the oscillometric wave- Minimum Maximum Mean SD
form envelope is not always a neat bell-shaped curve with a Age (years) 20 74 39 16
distinct peak [2]. Rather than a distinct peak the oscillo- Height (cm) 152 188 172 9
metric waveform envelope may have a plateau. To further Weight (kg) 46 103 71 11
add to the challenge of determining the MAP, the oscil- Arm circumference (cm) 22 38 28 2
lometric measurement is compromised by movement and SBP (mmHg) 90 140 119 13
other forms of artefact, such as respiratory disturbance DBP (mmHg) 53 90 75 10
[1, 14, 19, 22]. These disturbances are associated with

123
Med Biol Eng Comput (2011) 49:33–39 35

Oscillometric three recordings on 38 subjects and two recordings on 17


cuff pressure waveform Automated MAP
subjects. The overall mean and standard deviation (SD) of
to computer
auscultatory SBP and DBP over all subjects are also given
in Table 1.
Observer1
(SBP, DBP) During each recording, the cuff pressure was deflated at
Manual MAP
2–3 mmHg/s and recorded by a pressure sensor in a sep-
Observer2
(SBP, DBP)
arate recording system. As the subject was asked to keep
still during the whole measurement procedure, movement
Fig. 1 Overview of the blood pressure measurement protocol artefact and its effect to the cuff air system was minimized.
The cuff pressure was then digitally recorded to a data
(Accoson Greenlight 300) [9], simultaneously by two capture computer at a sample rate of 2000 Hz for sub-
trained observers using a dual-headed stethoscope. The sequent off-line analysis. The oscillometric pulses were
whole auscultatory measurement procedure followed the then extracted from the cuff pressure after segmenting each
guidelines recommended by the American Heart Associa- pulse and removing the baseline cuff pressure by using the
tion [17]. All measurement pairs agreed within 4 mmHg software previously developed by the PTB in Berlin,
and their average values were used for further analysis. On Germany, and checking the pulse extraction manually [16].
each subject, repeat recordings were performed after the The segmentation borders were at the feet of oscillometric
cuff pressure was released for 2 min. There was no signif- pulses. The peak of each oscillometric pulse was used for
icant difference on the auscultatory SBP and DBP between further analysis. Figure 2 shows typical examples of
repeat recordings (both P [ 0.15). The average value of deflating cuff pressures and the corresponding oscillomet-
auscultatory SBP and DBP from the repeat recordings was ric pulses as the cuff pressure was reduced. The oscillo-
calculated as the reference SBP and DBP for that subject. In metric waveform in Fig. 2a has a distinct peak, whilst there
total, 148 recordings were included from all subjects, with is a plateau in Fig. 2b.

Fig. 2 Typical examples of


cuff pressure and extracted
oscillometric pulse waveforms
with a distinct peak (a) and with
a plateau (b). Oscillometric
pressure was measured from the
oscillometric pulses, which
were extracted from the cuff
pressure after the removal of the
cuff deflation pressure

Fig. 3 Illustration of the


calculation of MAP from six
different techniques. See text
for the detailed definition. In
this example, MAP 5 overlaps
MAP4

123
36 Med Biol Eng Comput (2011) 49:33–39

2.3 Mean arterial blood pressure determination cuff deflation rate, with a limited number of detectable
oscillometric pulses.
Auscultatory MAP was estimated in each recording from
auscultatory DBP plus 33% of the pressure change from
2.4 Data analysis
DBP to SBP, with supplementary data for the use of 40% in
the formula [5]. Six different techniques were then devel-
For all the blood pressure recordings, SPSS software
oped using Matlab 7.1. (MathWorks Inc., USA) to deter-
package (SPSS Inc., USA) was employed to determine the
mine the corresponding automated MAPs from the cuff
effect of different techniques on MAP estimation, and also
pressure at a specific time determined from oscillometric
to perform regression analysis between the auscultatory
pulse pressure. Details are illustrated in Fig. 3 and descri-
MAPs and automated MAPs from the six different oscil-
bed below:
lometric techniques, obtaining the square of the correlation
• MAP1 and MAP2 were the cuff pressures at the peak coefficients (R2). Bland–Altman analysis (Systat Software
and foot, respectively of the largest oscillometric pulse Inc., USA) was also performed to assess the agreement
[25]. They are the traditional oscillometric techniques between the automated MAPs and auscultatory MAPs,
for MAP estimation, but can be influenced by the with their paired differences and SD of differences calcu-
variation of oscillometric waveform caused by artefact lated. A value of P \ 0.05 was considered statistically
or noise. significant. Finally, the automated mean weights calculated
• MAP3 and MAP4 were the cuff pressures at the peak of from the automated MAPs were compared with the clas-
the pulse which was closest to the peak of the 4th and sical manual mean value of 33%.
6th order fitted polynomial curve, respectively. The
polynomial curve fitting technique was used here
because it has the power to interpolate between 3 Results
the sampled pulses and to smooth out noise in the
oscillometric waveform. Furthermore, because the The overall mean MAPs from all the recordings for the
oscillometric waveform envelope is similar to a bell- auscultatory and oscillometric techniques are given in
shaped curve, an even number of orders above 2 is Table 2. ANOVA analysis showed that different tech-
required to construct a curve with flat features at the niques resulted in significant differences in MAP estima-
beginning and the end. tion (P \ 0.001). The mean paired differences between the
• MAP5 was obtained from MAP4 after correcting for oscillometric techniques and the auscultatory technique
the amplitude difference between the 6th order poly- were significant (all P \ 0.01). Technique 6 with curve
nomial curve and the recorded oscillometric pulse, such fitting improved the paired differences from -1.3 and
that when the polynomial curve lay above the oscillo- -3.4 mmHg (classic technique, without curve fitting) to
metric pulse, the correction added was positive. This -1.0 mmHg. These paired differences were associated
amplitude difference can be caused by respiratory with differences between the manual and automated mean
effects. weights of 4% (33–29%) and 9% (33–24%) for the classic
• MAP6 was the baseline cuff pressure associated with techniques to 2% (33–31%) for technique 6.
the peak of the 6th order fitted polynomial curve, where Figure 4 shows the comparison of MAPs obtained
the baseline was drawn through the foot of all pulses. from the auscultatory and different oscillometric tech-
This technique may be able to correct the effect of fast niques. The linear regression analysis showed that the

Table 2 The mean MAP


Measurement technique Mean MAP Mean paired SD of paired Mean weight
calculated from the manual
(mmHg) difference to difference calculated from
auscultatory method and from
auscultatory (mmHg) the MAPs (%)
six different oscillometric
MAP (mmHg)
techniques. Their mean and SD
of paired differences, and the Auscultatory 89.3 – – 33
mean weights calculated from
the MAPs are also presented Automated oscillometric 1 88.0 -1.3 6.2 29
technique 2 85.9 -3.4 5.9 24
3 91.5 2.1 4.1 38
4 91.0 1.6 4.1 37
5 90.9 1.6 4.1 37
6 88.3 -1.0 3.7 31

123
Med Biol Eng Comput (2011) 49:33–39 37

Fig. 4 Regression analysis


results of MAP from the
auscultatory and six
oscillometric techniques (top six
sub-figures). Bland–Altman plot
of MAP differences ±2SD
between the six oscillometric
techniques and the auscultatory
method (bottom six sub-figures)

correlations between the six calculated automated oscillometric pulses (see Fig. 3) with technique 2 associ-
MAPs and the auscultatory MAP were significant (all ated with the lower cuff pressure as it is the pressure at the
P \ 0.001), with the R2 of 0.71, 0.73, 0.84, 0.84, 0.84 foot, rather than the peak of the oscillometric pulse. Sim-
and 0.86, respectively, increasing from techniques 1 to 6. ilarly, the MAPs from technique 3, 4 and 5 were higher
Bland–Altman analysis showed that technique 6 than from technique 6 although the polynomial curve fit-
improved the SD of paired differences best from 6.2 and ting technique was applied to all of them. This difference
5.9 mmHg (classic technique, without curve fitting) to again reflects the amplitude of oscillometric pulse with
3.7 mmHg. technique 3–5, whilst technique 6 used the cuff pressure at
When the mean weight of 40% rather than 33% was the baseline (see Fig. 3). Although the mean MAP differ-
used to calculate the auscultatory MAP, all mean paired ences compared with the auscultatory technique were less
difference in comparison with the auscultatory MAP fell by than 5 mmHg, the systematic mean difference between
an identical 2.9 mmHg with no change to the differences different oscillometric techniques was up to 5.5 mmHg.
between oscillometric techniques. There were small dif- The International Standard requires a maximum systemic
ferences in the SD of paired differences, but not to the difference of 5 mmHg [12]. Furthermore, in a patient
order between oscillometric techniques, with MAP6 still population, the mean differences between the oscillometric
having the lowest variability. techniques and the auscultatory technique can be expected
to be much higher.
Next, the significant differences between MAPs from the
4 Discussion oscillometric techniques and the auscultatory technique
resulted in the automated mean weights calculated from
We have shown that the automated MAPs calculated from automated MAPs being different for all the six techniques
the oscillometric waveform using different measurement studied. Technique 6 produced the smallest mean difference
algorithms were different. The technique associated with (-2%) against the classically assumed mean value of 33%.
the peak of the 6th order polynomial model of the oscil- It is generally accepted that the automated MAP from the
lometric pulse waveform yielded the smallest mean paired oscillometric waveform gives a fairly accurate estimation in
difference in comparison with the auscultatory technique. comparison with the intra-arterial MAP [24], our finding
Technique 1 yielded a higher mean than that from tech- allows us to question whether 33% is a useful mean weight
nique 2. This difference reflects the amplitude of the for achieving accurate MAP estimation. It has been reported

123
38 Med Biol Eng Comput (2011) 49:33–39

that a weight of 41% is more accurate than the traditional References


33% for the calculation of MAP from invasive SBP and
DBP after investigating the pulses measured invasively 1. Amoore JN, Geake WB, Scott DHT (1998) The effects of pulse
rate, artifact and pulse strength on oscillometric non-invasive
from 150 patients during cardiac catheterization [15]. Bos blood pressure measurements. J Clin Eng 23:103–109
et al. [5] also found that, in comparison with the intra- 2. Amoore JN, Lemesre Y, Murray IC, Mieke S, King ST, Smith
arterially measured MAP, using the mean weight of 40% FE, Murray A (2008) Automatic blood pressure measurement: the
avoided the underestimation of MAP calculated from the oscillometric waveform shape is a potential contributor to the
differences between oscillometric and auscultatory pressure
empirical formula. Razminia et al. [20] reported the use of measurements. J Hypertens 26:35–43
the traditional 33%, but with an added correction for heart 3. Beevers G, Lip GY, O’Brien E (2001) ABC of hypertension:
rate (HR), where MAP was expressed as: MAP = DBP ? blood pressure measurement: Part I-sphygmomanometry: factors
[0.33 ? (HR 9 0.0012)] 9 (SBP-DBP). It is accepted that common to all techniques. BMJ 322:981–985
4. Berne RM, Levy MN (1998) The arterial system. In physiology.
there is still uncertainty with calls for more invasive clinical Mosby Inc, St Louis, pp 415–428
studies to evaluate the mean weight for MAP estimation 5. Bos WJ, Verrij E, Vincent HH, Westerhof BE, Parati G, van
[6, 23], and whether the mean weight changes with different Montfrans GA (2007) How to assess mean blood pressure prop-
clinical conditions, such as the age and blood pressure erly at the brachial artery level. J Hypertens 25:751–755
6. Chemla D, Brahimi M, Nitenberg A (2007) Thumb-rule for the
range, which could influence the shape of the oscillometric proper assessment of mean blood pressure at the brachial artery
waveform [2]. These values would provide useful infor- level: what should be changed? J Hypertens 25:1740–1741
mation for validating automated blood pressure devices for 7. Cnossen JS, Vollebregt KC, de Vrieze N, ter Riet G, Mol BW,
clinical use. Franx A, Khan KS, van der Post JA (2008) Accuracy of mean
arterial pressure and blood pressure measurements in predicting
More interestingly, we have also shown that the vari- pre-eclampsia: systematic review and meta-analysis. BMJ
ability of MAP measurement differences from different 336:1117–1120
oscillometric techniques in comparison with the ausculta- 8. Geddes LA, Voelz M, Combs C, Reiner D, Babbs CF (1982)
tory technique can be improved with curve fitting algo- Characterization of the oscillometric method for measuring
indirect blood pressure. Ann Biomed Eng 10:271–280
rithms. The technique, associated with the peak of the 6th 9. Graves JW, Tibor M, Murtagh B, Klein L, Sheps SG (2003) The
order polynomial model of the oscillometric pulse wave- Accoson Greenlight 300TM, the first non-automated mercury free
form, was shown to result in the lowest variability. It blood pressure device to pass the International Protocol for blood
yielded the smallest SD of paired differences (3.7 mmHg) pressure measuring devices in adults. Blood Press Monit 9:13–17
10. Guyton AC (1986) Textbook of medical physiology, 7th edn.
when referenced to those calculated from the classic W.B. Saunders, Philadelphia
empirical formula. The smoothing of the polynomial curve 11. Ira S (1996) Human physiology, 5th edn. Wm C Brown Pub-
fitting of the peaks accounted for its reduced SD. The fitted lishers, Iowa, pp 414–415
curve smoothes out random and noisy variations in the 12. ISO 81060-2 (2009) Non-invasive sphygmomanometers —Part 2:
Clinical validation of automated measurement type
oscillometric waveform. Having a good quality extracted 13. Jones DW, Appel LJ, Sheps SG, Roccella EJ, Lenfant C (2003)
oscillometric waveform is important for improving the Measuring blood pressure accurately: New and persistent chal-
estimation of automated MAP. In addition, although it is lenges. JAMA 289:1027–1030
recognized that the interpolation errors would be greater 14. Lazarashvili LT (1993) Interference resistance in automated
oscillometric sphygmomanometers. Biomed Eng 27:139–151
with fewer oscillometric peaks detected in a single mea- 15. Meaney E, Alva F, Moguel R, Meaney A, Alva J, Webel R
surement, this model-based technique with interpolation (2000) Formula and nomogram for the sphygmomanometric
takes account of the peak of the fitted curve lying between calculation of the mean arterial pressure. Heart 84:64
two oscillometric pulses, hence correcting for the effect of 16. Mieke S, Seemann R, Riedel W, Murray A, Amoore JN (2009) A
simulator to test automated non-invasive sphygmomanometers.
cuff deflation (or inflation) between heartbeats. This is IFMBE Proceedings 25/VII, pp 356–358
increasingly important because clinical demands lead to 17. Pickering TG, Hall JE, Appel LJ, Falkner BE, Graves J, Hill MN,
reduced determination time and hence increased cuff Jones DW, Kurtz T, Sheps SG, Roccella EJ (2005) Recommen-
deflation rate, with a reduced number of detectable oscil- dations for blood pressure measurement in humans and experi-
mental animals. Part 1: Blood pressure measurement in humans.
lometric pulses. A statement for professionals from the subcommittee of profes-
In conclusion, different techniques for automated MAP sional and public education of the American Heart Association
estimation from the oscillometric waveform produced dif- Council on high blood pressure research. Hypertension
ferent results, and the technique associated with the peak of 45:142–161
18. Ramsey M (1979) Noninvasive automatic determination of mean
the 6th order polynomial curve was shown to have the arterial pressure. Med Biol Eng Comput 17:11–18
smallest differences in comparison with the manual au- 19. Ramsey M (1991) Blood pressure monitoring: Automated oscil-
scultatory technique. lometric devices. J Clin Monit Comput 7:56–67
20. Razminia M, Trivedi A, Molnar J, Elbzour M, Guerrero M,
Acknowledgments Dr Dingchang Zheng is funded by the Engi- Salem Y, Ahmed A, Khosla S, Lubell DL (2004) Validation of a
neering and Physical Sciences Research Council (EPSRC). new formula for mean arterial pressure calculation: the new

123
Med Biol Eng Comput (2011) 49:33–39 39

formula is superior to the standard formula. Catheter Cardiovasc 23. Verrij EA, Vincent HH, Bos WJ (2008) Rule of thumb to
Interv 63:419–425 calculate mean arterial pressure at the brachial artery level.
21. Sesso HD, Stampfer MJ, Rosner B, Hennekens CH, Gaziano JM, J Hypertens 26:1043–1045
Manson JAE, Glynn RJ (2000) Systolic and diastolic blood 24. Yamakoshi K, Shimazu H, Shibata M, Kamiya A (1982) New
pressure pulse pressure and mean arterial pressure as predictors of oscillometric method for indirect measurement of systolic and
cardiovascular risk in men. Hypertension 36:801–807 mean arterial pressure in the human finger. Part 2: correlation
22. Van Horn RN, Kahlke RJ, Taylor LA, Dorsett TJ (2001) Non- study. Med Biol Eng Comput 20:314–318
invasive blood pressure performance: a reproducible method for 25. Zheng D, Murray A (2008) Estimation of mean blood pressure
quantifying motion artifact tolerance in oscillometry. Biomed from oscillometric and manual methods. Comput Cardiol
Instrum Technol 35:395–414 35:941–944

123
Copyright of Medical & Biological Engineering & Computing is the property of Springer Science & Business
Media B.V. and its content may not be copied or emailed to multiple sites or posted to a listserv without the
copyright holder's express written permission. However, users may print, download, or email articles for
individual use.

You might also like