You are on page 1of 4

Genucom, KT-1000, and Stryker knee laxity

measuring device comparisons


Devicereproducibility and interdevice comparison in
asymptomatic subjects
CARL L. HIGHGENBOTEN,*† MD, ALLEN AND
JACKSON,‡ EdD,
NEIL B. MESKE,* MS

*
From Orthopaedic Consultants, Dallas, Texas, and ‡ North Texas State University, Denton,
Texas

ABSTRACT com Knee Analysis System (FARO Medical Technologies


Inc., Champlain, NY)-can be used to test anterior-pusteriur
Generally, three devices (the Genucom Knee Analysis knee laxity. All three test devices and protocols are to a
System, the MEDmetric KT-1000 Arthrometer, and the large degree unique, but they have some common traits.
Stryker Knee Laxity Tester), which have been pre- When a knee laxity measuring device is attached to the leg
sented in the literature, may be used to assess clinically that is to be tested, an anterior-posterior drawer test is
the amount of knee laxity in an objective manner. This conducted and the anterior-posterior displacements are re-
study compared the reproducibility of the anterior and corded in millimeters. One of the primary purposes of such
posterior knee laxity values for each and made direct test devices is to aid in the diagnosis of disruptiun of the
comparisons of the results obtained. Thirty asympto- anterior cruciate and/or posterior cruciate ligaments and
matic subjects were given a test-retest protocol on the determination of improvement from surgery attd treat-
both legs with all three devices. Devices were tested in ment. Markolf and AmStUtZ6 have stated that future devel-
a counterbalanced order. Repeatability of test values
opments in sports medicine should lead to htandardized
within devices was variable, but all were acceptable, methods of quantifying knee stability.
the lowest being r 0.74. Analysis of variance (AN-
=
The Stryker Knee Laxity Tester is operatiuttally the least
OVA), and correlational analysis revealed that device- complicated of the test devices. It includes a patient pusi-
specific anterior and posterior laxity values were pro- tioning seat, a force applicator, and a measuring instrument
duced. We concluded that each commercially available for recording the millimeters of displacement resulting from
knee laxity testing device can provide reproducible the anterior-posterior drawer test. Boniface et al.l measured
quantitative measurements of knee laxity; however, the displacements of uninjured knees and knees with ACL
due to differences in device sensitivities and functional
disruption. Their tests were conducted with the knee joint
design, numerical results from one device cannot be at a 20° flexion angle and 89 N of applied force. The injured
generalized to another device. knees had a mean anterior displacement of 8.1 mm, and the
uninjured knees had a mean displacement of 2.5 mni.
The MEDmetric KT-1000 Arthrometer is a more complex
Recent literature reports the use of various testing devices device than the Stryker Knee Laxity Tester and alluws a
in order to obtain objective knee laxity measurement. 14 more sensitive test analysis. It also measure5 attteriur-pus-

These devices-the Stryker Knee Laxity Tester (Stryker, terior knee laxity displacements resulting trom atr anterior-
Kalamazoo, MI), the MEDmetric KT-1000 Arthrometer posterior drawer test. Daniel et a1.1 2 compared uninjured
(MEDmetric Corporation, San Diego, CA), and the Genu- knees and knees with ACL disruption using the arthrometer.
They also conducted their tests at a knee joint angle of 20°
of flexion and used 89 N of force. Their repurted mean
t Address correspondence and repnnt requests to Carl L Highgenboten,
MD, Orthopaedic Consultants, 7777 Forest Lane, Building C, Suite 106, Dallas, anterior displacement values were 13.0 man for ACL injured
TX 75230 knees and 5.65 mm for uninjured knees. Sherman et al.8
743
744

conducted a comparison of the KT-1000 and the University &dquo;test administration with two levels.&dquo; The dependent vari-
of California at Los Angeles (UCLA) instrumented clinical ables are the anterior and posterior laxity measurements.
knee testing apparatus. The results indicated that both For the Stryker Knee Laxity Tester and the MEDmetric
devices, while providing device-specific data, were able to KT-1000 Arthrometer, an anterior-posterior drawer test was
correctly classify 90% to 95% of patients with disrupted performed at 20° of flexion and 89 N (20 pounds) of force.
ACLs. The Genucom test was also performed at 20°, but the system
The Genucom Knee Analysis System, a computerized tool requires a force of 93.45 N (21 pounds), which was the only
for the evaluation of the knee, is the most complicated of deviation in test protocol between devices. Standard testing
the test devices. It combines a computer, a biomechanical procedures for each test device, as provided by their vendors,
digitizer, an electrogoniometer, a dynometer, and an inkjet were followed. Detailed descriptions of test protocols using

printer. It provides a variety of test protocols, which theo- the Stryker and the KT-1000 are available in the literature.’-
3, 5,
retically allows a total description of knee ligament stability. 8 However, the Genucom protocol has not been described;
Highgenboten and Jackson5 have reported that the device therefore, the essential steps in conducting a test on the
gives reproducible test-retest results for a variety of tests. device are given as follows:
Of particular relevance is an anterior-posterior drawer test 1. The subject’s permanent data file was established using
conducted on uninjured subjects. The knee joint angle was the system’s computer and stored on disk.
30° of flexion and the force was 93.45 N. The mean anterior 2. The subject’s test leg was secured with the system’s
displacement was 5.67 mm. thigh restraints, and an electrogoniometer was attached to
The purpose of the present investigation was to compare the lower leg.
the numerical results of anterior-posterior knee laxity test- 3. Seven anatomical landmarks (tibial tubercle, tibial
ing using the three test devices mentioned. Specifically, crest, medial femoral condyle, lateral femoral condyle, me-
reproducibility coefficients were estimated for each device dial tibial plateau, lateral tibial plateau, and the inferior
and a comparison of test results was made to determine if pole of the patella) were physically marked and digitized,
significant differences existed, and if they did, to determine and the distance between them recorded.
their meaning. 4. A soft tissue compensation test in three planes was
performed to compensate for possible movement that can
occur within the test apparatus without affecting the dis-
METHODS
tance between the digitized bony landmarks.
5. The anterior-posterior drawer test was performed at
Subjects 20°.
6. The subject’s data was stored permanently on disk.
Both knees of 30 subjects having no prior history of knee 7. A hard copy of each test was printed, from which data
injury abnormality provided a sample of 60 knees. Ages
or
for statistical analysis was obtained.
of the 9 males and 21 females ranged from 16 to 44 years.
Each volunteer signed an informed consent that provided
information concerning the voluntary nature of their partic- Data analysis
ipation, the purpose of the study, and the anticipated bene- To estimate reproducibility, intraclass correlation proce-
fits from its successful completion.
dures, which incorporate an ANOVA model, were used to
determine the relationship between trials and if significant
Procedures differences existed between trials. A 3 (test device) by 2
(test) ANOVA with repeated measures on both factors was
Subjects received two complete tests on each leg. A complete conducted to determine if significant differences existed
test using all three devices was performed on one leg and between test devices and if the test device interacted with
then the opposite leg; then, each leg received a second set of
repeated tests. Correlations were also calculated between the
complete tests. Intertest reliability was estimated between test device results to determine their relationships.
the two complete tests. During testing with each of the three
devices, three repeated measurements, or trials, were re-
corded. Intertrial reproducibility was estimated from the RESULTS
three repeated measurements. The tests administered to
each subject were performed in succession on the same day. Reproducibility
The order of the devices used in testing was counterbalanced
across subjects, resulting in a Latin Square design, which As indicated in the Procedures, three trials or repeated
prevents order effects in the results. Thus, each subject was measurements wereadministered within a test or device
given two complete tests using each of three devices on each session. Intraclass correlations indicated that all three de-
leg, with three repeated measurements per device during vices demonstrated excellent reproducibility for anterior and
each test session. Anterior and posterior displacements were posterior laxity values across the three trials within a ses-
recorded in millimeters for each trial. Therefore, the inde- sion. The values are given in Table 1. Trend analysis indi-
pendent variables are &dquo;test device with three levels&dquo; and cated two significant trial effects, but the magnitude of the
745

TABLE 1 DISCUSSION
Descriptive statistics and reliability estimates of laxity values
Intradevice reproducibility
As noted by Sherman et al.’, in order to obtain reproducible
(i.e., repeatable numerically) quantitative data from any of
these knee laxity measuring devices, the test administrator
must become proficient in the correct use and application
of the device and in the testing protocol. A routine testing
protocol should be followed each time for each patient tested.
In this study, the test administrator spent in excess of 40
hours becoming proficient with each of the three testing
devices and the three test protocols. The manufacturers of
TABLE 2 the Genucom System offer a testing and certification pro-
Correlation matrices of knee laxity test device results gram. The test administrator for this study has been rated
as excellent on the anterior-posterior drawer test by this
certification program. Hanten and Pace4found reproducible
results with the KT-1000 during repeated trials and repeated
testers. Based on our data in Table 1, we conclude that all
three devices can provide reproducible test results.

lnterdevice comparisons
The Genucom produces anterior laxity values higher than
a p < 0.0] .
either the KT-1000 or the Stryker, and the KT-1000 pro-
trend duces significantly higher anterior laxity values than the
was not greater than 0.3 mm and was considered
clinically unimportant. Complete test or device sessions were Stryker. As Table 1 demonstrates, a higher degree of inter-
individual variability will be detected by the Genucom due
given on a test-retest basis. Table 1 provides the statistical
to the larger standard deviations it produces for knee laxity
results relevant to the completed test session reproducibility
measurements.
analysis. R was variable between devices but acceptable, the The posterior laxity values of the Genucom and the Stry-
lowest being the Stryker’s anterior laxity value of 0.74. The
ker were not significantly different from one another, but
trend analysis indicated no significant differences between
each was significantly different from the posterior laxity
test sessions.
values obtained from the KT-1000. The reason for this
difference may be understood by looking at the functional
Device comparison design of the three devices. The Genucom fixes the femur.
The Stryker attempts to restrict movement of the femur
The 3 (test device) by 2 (test) ANOVA with repeated meas-
with the use of a VELCRO (VELCRO USA Inc., Manches-
ures revealed a significant difference between the device
ter, NH) strap. The KT-1000 does not attempt to fix the
measures for anterior (F 2,118 62.77; P < 0.01) and
=

femur. Sherman et al.’ found in his study of the KT-1000


posterior (F 2,118 17.26; P < 0.01) laxity values. The test
=
and another knee laxity measuring device that the KT-1000
administration and interaction effects were not significant.
demonstrated twice the magnitude in laxity values for both
A post hoc analysis (P 0.01) was conducted to examine
=

normal and ACL deficient knees when a force of 89 N was


the device main effect. The two degrees of freedom were
applied. However, the difference in magnitude decreased to
partitioned within the ANOVA to provide two single degree less than 2 mm when the standard force of 200 N was
of freedom contrasts. These results indicated that the Gen-
ucom produces significantly higher anterior laxity values (F
applied with the other device. The only difference between
devices was the fact that the KT-1000 did not fix the femur,
1,59 31.3; P < 0.01) than the Stryker or the KT-1000. The
=

while the other device did. They concluded that each device
KT-1000 produced significantly higher anterior laxity values
produced device-specific measurements due to differences in
(F 1,59 130.31; P < 0.01) than the Stryker. For posterior
=

functional design.
laxity, the Stryker and Genucom values were not signifi- Statistical analysis of the results of our study (Table 2)
cantly different, but they were significantly higher (F 1,59 indicates that each device produces anterior and posterior
=
45.7; P < 0.01) than the KT-1000 values. Table 2 shows laxity measurements that are peculiar for each device and
the results of a between-device correlation. While several of
related to sensitivity and functional design of the device.
the correlations were significant (P < 0.01), they tended to
range from moderate to extremely low values. These low to Clinical
moderate correlations, along with the ANOVA results, in-
significance
dicated that the anterior and posterior values were specific For the clinical practitioner, device-specific results do not
to the device used in the test. matter. The practitioner should be aware of the following:
746

1. The devices will provide reproducible results when pacity allows storage of patient data for subsequent compar-
appropriate practice and test experience have been achieved. ison of later tests to determine success of treatment and
2. Specific laxity values and results for one device cannot rehabilitation.
be generalized to another. Physicians must determine their needs in knee laxity
3. Past studies have demonstrated the Stryker Knee Lax- testing. If measurement of anterior-posterior laxity is all
that is needed, then the Stryker or KT-1000 will be accept-
ity Tester and the MEDmetric KT-1000 will discriminate
able. However, if a more sophisticated analysis of knee
between normal knees and ACL disrupted knees.1-3,8a
Currently, literature is not available supporting the ability ligament status is needed, then the Genucom and its asso-
ciated expenses may be justified.
of the Genucom to accomplish that goal. Oliver and
Coughlin’ have demonstrated, however, that the Genucom REFERENCES
test results will have a high degree of agreement for the
anterior-posterior drawer test at 30° with clinical examina- 1 Boniface JR, Fu FH, Ilkhanipour K. Objective anterior cruciate ligament
tion for subjects with an abnormal knee history. In addition, testing Orthopedics 9 391-393, 1986
2. Daniel DM, Malcom LL, Losse G, et al: Instrumented measurement of
the Genucom gives reproducible laxity values. Thus, it is anterior laxity of the knee. J Bone Joint Surg 67A: 720-726, 1985
highly probable that it can differentiate between normal and 3 Daniel DM, Stone ML, Sachs R, et al. Instrument measurement of anterior
knee laxity in patients with acute anterior cruciate ligament disruption. Am
ACL disrupted knees. J Sports Med 13 . 401-407, 1985
A final point that should be mentioned in discussing these 4 Hanten WP, Pace MB Reliability of measuring anterior laxity of the knee
instrumented knee laxity test devices is the cost factor. The joint using a knee ligament arthrometer. Phys Ther 67. 357-359, 1987
5 Highgenboten C, Jackson AW, Meske NB Measurement of knee laxity
costs of these devices vary widely, with the Stryker at $900, using the Genucom analysis system Med Sci Sports Exerc 21. S81, 1989
6 Markolf KL, Amstutz HC. The clinical relevance of instrumented testing of
the KT-1000 at $2900, and the Genucom at $49,000. While ACL insufficiency. Experience with the UCLA clinical knee testing appara-
the Stryker and KT-1000 are relatively simple mechanical tus Clin Orthop 223 198-207, 1987
7 Oliver JH, Coughlin LP. Objective knee evaluation using the Genucom
devices with a small cost, the Genucom is a complex com- Knee Analysis System. Clinical implications Am J Sports Med 15 . 571-
puterized system which allows a full battery of knee laxity 578, 1987
8 Sherman OH, Markolf KL, Ferkel RD: Measurements of anterior laxity in
tests for analysis of total knee ligament stability, including normal and anterior cruciate absent knees with two instrumented test
varus-valgus, pivot shift, and recurvatum. Its computer ca- devices Clin Orthop 215
. 156-161, 1987

You might also like