Professional Documents
Culture Documents
Purpose Despite the ubiquitous use of manual goniometry in measuring objective outcomes
of hand surgery and therapy, there are limited data concerning its accuracy or repeatability
for wrist motion. The purpose of this study was to evaluate the accuracy and reliability (both
inter- and intra-rater) in measuring wrist flexion and extension using 3 manual goniometric
alignment techniques (ulnar, radial, and dorsal–volar) in cadaveric upper extremities, using
fluoroscopic verification of posture as a gold standard. In addition, we sought to assess the
accuracy and reliability of the dorsal–volar technique for measurement of radioulnar deviation.
Methods External fixators were applied to 10 cadaveric wrists with intramedullary cannulated
rods in the radius and third metacarpal for gold-standard fluoroscopic verification of posture.
Manual goniometric measurements with each technique were captured by 2 raters (a hand
surgeon and a hand therapist) for reliability measurements and by a single rater for accuracy.
Wrists were positioned at angles of maximum flexion, extension, and radial and ulnar
deviation for reliability testing and at preselected angles across the range of motion for
accuracy testing. At each position, wrist angle was measured with a 1° increment goniom-
eter, and fluoroscopic angles were measured digitally. Intraclass correlation coefficients and
root mean square values were calculated for all combinations, and analysis of variance was
used to test differences between techniques.
Results No technique was statistically less accurate than any other (6° to 7°). Each method
was found to have high intra-rater reliability. For measurement of wrist flexion and exten-
sion, the dorsal–volar technique demonstrated the greatest inter-rater reliability, as compared
to ulnar and radial, respectively.
Conclusions Although each measurement technique demonstrated a similar degree of accuracy
and intra-rater reliability, the dorsal–volar technique demonstrates the greatest level of
inter-rater reliability for measurement of wrist flexion and extension. This information is
important clinically, as measurements are regularly exchanged between hand surgeons and
therapists as a basis for decisions regarding patient care. (J Hand Surg 2009;34A:1422–
1428. © 2009 Published by Elsevier Inc. on behalf of the American Society for Surgery of
the Hand.)
Key words Accuracy, continuity of care, manual goniometer, reliability, wrist range of motion.
From the Hospital for Special Surgery, New York, NY. Corresponding author: Scott W. Wolfe, MD, Hospital for Special Surgery, Department, 535 East
70th Street, New York, NY 10021; e-mail: wolfes@hss.edu.
SpecialthankstoMichaelSheehanandClaraHilario,withoutwhomthisprojectwouldnothavebeen
possible. Thanks to Hologics (Bedford, MA) for the use of their fluoroscopy unit and to Synthes (West 0363-5023/09/34A08-0007$36.00/0
Chester, PA) for their generous donation of 10 external fixation devices. doi:10.1016/j.jhsa.2009.06.002
Received for publication November 28, 2007; accepted in revised form June 2, 2009.
No benefits in any form have been received or will be received related directly or indirectly to the
subject of this article.
R
ANGE OF MOTION
sal barometer of success for surgical and non-
surgical treatment of wrist injuries. These mea-
surements, most often captured by manual goniometry,
serve as a common language between surgeons, pa-
tients, and therapists. To facilitate the successful treat-
ment of wrist injuries and to improve effective com-
munication between physicians and therapists, it is
important that measurements be both accurate and re-
liable between examiners. These parameters were eval-
uated in cadaveric specimens for 3 goniometric tech-
niques and between 2 experienced hand professionals.
There are currently 3 popular methods of measuring
wrist flexion and extension with a manual goniometer:
placement of the device along the ulnar, radial, or
dorsal–volar surfaces of the wrist. For radial and ulnar
deviation, the American Society for Hand Therapists
recommends placement of the goniometer on the dorsal
surface; the authors could find no other suggested
method in the literature.1 LaStayo et al. examined the
inter- and intra-rater reliability of these 3 techniques for
measurement of flexion and extension.2 Their study
compared measurements taken by physical therapists
with varying backgrounds and found the dorsal–volar
technique of measurement to be the most reliable. Re-
liability of passive and active measurements of ROM
between therapists was also assessed by Horger et al.3
This study, which compared therapists with different
levels of experience and specialization, found that more
specialized raters demonstrated a higher degree of intra-
rater reliability; however, no measurement technique
was specified. Several other studies of wrist ROM and
measurement tools have been conducted, dating back to
1928.4 – 6 To our knowledge, no study has examined the
accuracy of these techniques as compared to a gold FIGURE 1: Cadaveric wrist with external fixator positioned
standard of skeletal alignment. In addition, no compar- obliquely to the frontal and sagittal planes.
ison has been made of the reliability of each goniomet-
ric technique between a surgeon and a therapist, and
there is no published standard among hand surgeons METHODS
concerning use of either technique. Finally, the validity Ten above-elbow cadaveric wrists (Anatomy Gifts
and accuracy of manual goniometric measurements as a Registry, Glen Burnie, MD, and Science Care, Phoenix,
gold standard for evaluation of electronic methods of AZ) were used to determine the accuracy and reliability
wrist position (electrogoniometer, motion analysis, of each manual goniometric technique. Specimens were
data gloves) have not been established. The purpose positioned in neutral forearm rotation (in reference to
of this study was to determine the accuracy and the medial and lateral epicondyles of the humerus) and
reliability of manual goniometry for measurements pinned with a 2.0-mm (0.079-in) K-wire, driven
of static wrist position. Our primary hypothesis was through the proximal radius and ulna. We applied a
that the dorsal–volar technique would be the most bridging single-bar external fixator (Synthes USA, Pa-
accurate method to determine wrist flexion and ex- oli, PA) across the wrist in an oblique plane (45° with
tension. Our secondary hypothesis was that this tech- respect to the sagittal and frontal planes, with pin clus-
nique would exhibit the highest degree of inter- and ters in the index metacarpal and mid-diaphysis of the
intra-rater reliability. radius (Fig. 1). To detect the central axis of the radius
FIGURE 2: Fluoroscopic images of the intramedullary rods in position. A Posteroanterior view. B True scaphopisocapitate lateral
projection.
and third metacarpal under fluoroscopy (GE Medical each measurement, the axis of the goniometer was
Systems, Fairfield, CT), a cannulated rod was inserted centered over the axis of wrist rotation—the center of
in both intramedullary canals by the following proce- the capitate.7,8 To determine this point, we used the tip
dure. The long finger of each specimen was removed at of the radial and ulnar styloid as surface landmarks and
the metacarpophalangeal joint and a 1.6-mm (0.063-in) determined that the center of the capitate in the coronal
guidewire was advanced under fluoroscopic guidance plane was positioned at the midpoint of the line con-
through the third metacarpal, the capitate, and into the necting the 2 styloid tips, as verified under fluoroscopy.
radius, with the wrist in neutral position. A cannulated For each measurement technique, the axis of the goni-
rod was then advanced over the guidewire to contact the ometer was aligned with this midpoint. In the radial and
radial pin cluster of the external fixator (verified with ulnar techniques, the proximal arm of the goniometer
fluoroscopy). The radial articular surface was identified was aligned in parallel with the radius or ulna along its
under fluoroscopy to measure its distance from the outer border, and the distal arm of the goniometer was
radial pin cluster. With the guidewire left in place, the aligned in parallel with the third metacarpal (Figs. 3C, 3D).
cannulated rod was removed and cut to fit the measured For the dorsal–volar technique, the distal arm was
distance between the radial articular surface and the aligned with the third metacarpal, and the proximal arm
distal-most external fixator pin of the radial cluster. The 2 was aligned centrally on the forearm. With this tech-
segments of the rod were again advanced over the nique, the goniometer was placed on the dorsal surface for
guidewire sequentially, seating the proximal rod in flexion, radial deviation, and ulnar deviation and on
the distal radius and then withdrawing the distal the volar surface for extension (Figs. 3A–D, 4A–B).
portion into the capitate (Figs. 2A, 2B). Measurements of flexion and extension were cap-
Manual goniometric measurements of wrist exten- tured with all 3 goniometric techniques, and mea-
sion, flexion, radial deviation, and ulnar deviation were surements of radial and ulnar deviation were mea-
performed by a hand surgeon and a hand therapist, sured exclusively with the dorsal–volar technique.1
using previous references for guidance with regard to Fluoroscopy was used as a gold standard for deter-
positioning of the goniometer.7,8 A prestudy power mination of actual wrist position. Images were captured
analysis indicated power ⬎0.8 to detect differences in digitally, and the angle created by the intramedullary
both reliability and accuracy between techniques. A 1° rods was measured using the Adobe Photoshop (Adobe
resolution manual goniometer (Orthologic, Tempe, AZ) Systems Inc, San Jose, CA) angle measurement tool
was used for all measurements. After familiarizing (Fig. 5).9 For any measurement in one plane, it was
themselves with published techniques for measurement, important that the wrist be aligned at a neutral angle in
the therapist and surgeon reached consensus to stan- the unmeasured plane. To remain consistent between
dardize the alignment procedures of each technique and specimens and between measurements, 2 radiographic
were allowed to practice each technique repeatedly over alignment techniques were used. When measuring ra-
a 10-minute period before data collection.1–3 During dioulnar deviation in the coronal plane, the wrist was set
FIGURE 3: Demonstration of the A dorsal, B volar, C radial, and D ulnar measurement techniques as used on a cadaveric
specimen instrumented with a bridging single-bar external fixator.
in neutral posture in the sagittal plane by aligning the mally overlapped, assuring a consistent true scapho-
intramedullary rods under fluoroscopic control. Images pisocapitate lateral projection as described by Yang et al.10
were taken and verified using Adobe Photoshop. The To evaluate accuracy, wrists were each positioned at
specimen was then rotated into a coronal projection approximately 0°, 15°, 30°, and 45° of extension and
until the radioulnar joint space was maximized, ensur- flexion; 0° and 10° of radial deviation; and 0°, 10°, 20°,
ing a consistent posteroanterior (PA) view. Similarly, and 30° of ulnar deviation, as measured by the manual
when measuring in the sagittal plane, the intramedullary goniometer, allowing a 5° range at each position to
rods were aligned in the coronal plane under fluoros- confound any attempt by the raters to guess wrist po-
copy and verified using Adobe Photoshop. The speci- sition. The wrist was then imaged with fluoroscopy at
men was then rotated along the longitudinal (Z) axis each set position, with both PA and lateral exposures of
until the scaphoid tubercle and pisiform were maxi- the wrist, to ensure neutral rotation in the unmeasured
FIGURE 4: Demonstration of the dorsal technique for measurement of A ulnar and B radial deviation as used on a cadaveric
specimen instrumented with a bridging single-bar external fixator.