You are on page 1of 7

SCIENTIFIC ARTICLE

Accuracy and Reliability of Three Different Techniques


for Manual Goniometry for Wrist Motion:
A Cadaveric Study
Timothy I. Carter, BA, Brian Pansy, BE, Aviva L. Wolff, BS, Howard J. Hillstrom, PhD, Sherry I. Backus, BS,
Mark Lenhoff, BS, Scott W. Wolfe, MD

Purpose Despite the ubiquitous use of manual goniometry in measuring objective outcomes
of hand surgery and therapy, there are limited data concerning its accuracy or repeatability
for wrist motion. The purpose of this study was to evaluate the accuracy and reliability (both
inter- and intra-rater) in measuring wrist flexion and extension using 3 manual goniometric
alignment techniques (ulnar, radial, and dorsal–volar) in cadaveric upper extremities, using
fluoroscopic verification of posture as a gold standard. In addition, we sought to assess the
accuracy and reliability of the dorsal–volar technique for measurement of radioulnar deviation.
Methods External fixators were applied to 10 cadaveric wrists with intramedullary cannulated
rods in the radius and third metacarpal for gold-standard fluoroscopic verification of posture.
Manual goniometric measurements with each technique were captured by 2 raters (a hand
surgeon and a hand therapist) for reliability measurements and by a single rater for accuracy.
Wrists were positioned at angles of maximum flexion, extension, and radial and ulnar
deviation for reliability testing and at preselected angles across the range of motion for
accuracy testing. At each position, wrist angle was measured with a 1° increment goniom-
eter, and fluoroscopic angles were measured digitally. Intraclass correlation coefficients and
root mean square values were calculated for all combinations, and analysis of variance was
used to test differences between techniques.
Results No technique was statistically less accurate than any other (6° to 7°). Each method
was found to have high intra-rater reliability. For measurement of wrist flexion and exten-
sion, the dorsal–volar technique demonstrated the greatest inter-rater reliability, as compared
to ulnar and radial, respectively.
Conclusions Although each measurement technique demonstrated a similar degree of accuracy
and intra-rater reliability, the dorsal–volar technique demonstrates the greatest level of
inter-rater reliability for measurement of wrist flexion and extension. This information is
important clinically, as measurements are regularly exchanged between hand surgeons and
therapists as a basis for decisions regarding patient care. (J Hand Surg 2009;34A:1422–
1428. © 2009 Published by Elsevier Inc. on behalf of the American Society for Surgery of
the Hand.)
Key words Accuracy, continuity of care, manual goniometer, reliability, wrist range of motion.

From the Hospital for Special Surgery, New York, NY. Corresponding author: Scott W. Wolfe, MD, Hospital for Special Surgery, Department, 535 East
70th Street, New York, NY 10021; e-mail: wolfes@hss.edu.
SpecialthankstoMichaelSheehanandClaraHilario,withoutwhomthisprojectwouldnothavebeen
possible. Thanks to Hologics (Bedford, MA) for the use of their fluoroscopy unit and to Synthes (West 0363-5023/09/34A08-0007$36.00/0
Chester, PA) for their generous donation of 10 external fixation devices. doi:10.1016/j.jhsa.2009.06.002
Received for publication November 28, 2007; accepted in revised form June 2, 2009.
No benefits in any form have been received or will be received related directly or indirectly to the
subject of this article.

1422 䉬 ©  Published by Elsevier, Inc. on behalf of the ASSH.


RELIABILITY IN MEASURING WRIST RANGE OF MOTION 1423

(ROM) is regarded as a univer-

R
ANGE OF MOTION
sal barometer of success for surgical and non-
surgical treatment of wrist injuries. These mea-
surements, most often captured by manual goniometry,
serve as a common language between surgeons, pa-
tients, and therapists. To facilitate the successful treat-
ment of wrist injuries and to improve effective com-
munication between physicians and therapists, it is
important that measurements be both accurate and re-
liable between examiners. These parameters were eval-
uated in cadaveric specimens for 3 goniometric tech-
niques and between 2 experienced hand professionals.
There are currently 3 popular methods of measuring
wrist flexion and extension with a manual goniometer:
placement of the device along the ulnar, radial, or
dorsal–volar surfaces of the wrist. For radial and ulnar
deviation, the American Society for Hand Therapists
recommends placement of the goniometer on the dorsal
surface; the authors could find no other suggested
method in the literature.1 LaStayo et al. examined the
inter- and intra-rater reliability of these 3 techniques for
measurement of flexion and extension.2 Their study
compared measurements taken by physical therapists
with varying backgrounds and found the dorsal–volar
technique of measurement to be the most reliable. Re-
liability of passive and active measurements of ROM
between therapists was also assessed by Horger et al.3
This study, which compared therapists with different
levels of experience and specialization, found that more
specialized raters demonstrated a higher degree of intra-
rater reliability; however, no measurement technique
was specified. Several other studies of wrist ROM and
measurement tools have been conducted, dating back to
1928.4 – 6 To our knowledge, no study has examined the
accuracy of these techniques as compared to a gold FIGURE 1: Cadaveric wrist with external fixator positioned
standard of skeletal alignment. In addition, no compar- obliquely to the frontal and sagittal planes.
ison has been made of the reliability of each goniomet-
ric technique between a surgeon and a therapist, and
there is no published standard among hand surgeons METHODS
concerning use of either technique. Finally, the validity Ten above-elbow cadaveric wrists (Anatomy Gifts
and accuracy of manual goniometric measurements as a Registry, Glen Burnie, MD, and Science Care, Phoenix,
gold standard for evaluation of electronic methods of AZ) were used to determine the accuracy and reliability
wrist position (electrogoniometer, motion analysis, of each manual goniometric technique. Specimens were
data gloves) have not been established. The purpose positioned in neutral forearm rotation (in reference to
of this study was to determine the accuracy and the medial and lateral epicondyles of the humerus) and
reliability of manual goniometry for measurements pinned with a 2.0-mm (0.079-in) K-wire, driven
of static wrist position. Our primary hypothesis was through the proximal radius and ulna. We applied a
that the dorsal–volar technique would be the most bridging single-bar external fixator (Synthes USA, Pa-
accurate method to determine wrist flexion and ex- oli, PA) across the wrist in an oblique plane (45° with
tension. Our secondary hypothesis was that this tech- respect to the sagittal and frontal planes, with pin clus-
nique would exhibit the highest degree of inter- and ters in the index metacarpal and mid-diaphysis of the
intra-rater reliability. radius (Fig. 1). To detect the central axis of the radius

JHS 䉬 Vol A, October 


1424 RELIABILITY IN MEASURING WRIST RANGE OF MOTION

FIGURE 2: Fluoroscopic images of the intramedullary rods in position. A Posteroanterior view. B True scaphopisocapitate lateral
projection.

and third metacarpal under fluoroscopy (GE Medical each measurement, the axis of the goniometer was
Systems, Fairfield, CT), a cannulated rod was inserted centered over the axis of wrist rotation—the center of
in both intramedullary canals by the following proce- the capitate.7,8 To determine this point, we used the tip
dure. The long finger of each specimen was removed at of the radial and ulnar styloid as surface landmarks and
the metacarpophalangeal joint and a 1.6-mm (0.063-in) determined that the center of the capitate in the coronal
guidewire was advanced under fluoroscopic guidance plane was positioned at the midpoint of the line con-
through the third metacarpal, the capitate, and into the necting the 2 styloid tips, as verified under fluoroscopy.
radius, with the wrist in neutral position. A cannulated For each measurement technique, the axis of the goni-
rod was then advanced over the guidewire to contact the ometer was aligned with this midpoint. In the radial and
radial pin cluster of the external fixator (verified with ulnar techniques, the proximal arm of the goniometer
fluoroscopy). The radial articular surface was identified was aligned in parallel with the radius or ulna along its
under fluoroscopy to measure its distance from the outer border, and the distal arm of the goniometer was
radial pin cluster. With the guidewire left in place, the aligned in parallel with the third metacarpal (Figs. 3C, 3D).
cannulated rod was removed and cut to fit the measured For the dorsal–volar technique, the distal arm was
distance between the radial articular surface and the aligned with the third metacarpal, and the proximal arm
distal-most external fixator pin of the radial cluster. The 2 was aligned centrally on the forearm. With this tech-
segments of the rod were again advanced over the nique, the goniometer was placed on the dorsal surface for
guidewire sequentially, seating the proximal rod in flexion, radial deviation, and ulnar deviation and on
the distal radius and then withdrawing the distal the volar surface for extension (Figs. 3A–D, 4A–B).
portion into the capitate (Figs. 2A, 2B). Measurements of flexion and extension were cap-
Manual goniometric measurements of wrist exten- tured with all 3 goniometric techniques, and mea-
sion, flexion, radial deviation, and ulnar deviation were surements of radial and ulnar deviation were mea-
performed by a hand surgeon and a hand therapist, sured exclusively with the dorsal–volar technique.1
using previous references for guidance with regard to Fluoroscopy was used as a gold standard for deter-
positioning of the goniometer.7,8 A prestudy power mination of actual wrist position. Images were captured
analysis indicated power ⬎0.8 to detect differences in digitally, and the angle created by the intramedullary
both reliability and accuracy between techniques. A 1° rods was measured using the Adobe Photoshop (Adobe
resolution manual goniometer (Orthologic, Tempe, AZ) Systems Inc, San Jose, CA) angle measurement tool
was used for all measurements. After familiarizing (Fig. 5).9 For any measurement in one plane, it was
themselves with published techniques for measurement, important that the wrist be aligned at a neutral angle in
the therapist and surgeon reached consensus to stan- the unmeasured plane. To remain consistent between
dardize the alignment procedures of each technique and specimens and between measurements, 2 radiographic
were allowed to practice each technique repeatedly over alignment techniques were used. When measuring ra-
a 10-minute period before data collection.1–3 During dioulnar deviation in the coronal plane, the wrist was set

JHS 䉬 Vol A, October 


RELIABILITY IN MEASURING WRIST RANGE OF MOTION 1425

FIGURE 3: Demonstration of the A dorsal, B volar, C radial, and D ulnar measurement techniques as used on a cadaveric
specimen instrumented with a bridging single-bar external fixator.

in neutral posture in the sagittal plane by aligning the mally overlapped, assuring a consistent true scapho-
intramedullary rods under fluoroscopic control. Images pisocapitate lateral projection as described by Yang et al.10
were taken and verified using Adobe Photoshop. The To evaluate accuracy, wrists were each positioned at
specimen was then rotated into a coronal projection approximately 0°, 15°, 30°, and 45° of extension and
until the radioulnar joint space was maximized, ensur- flexion; 0° and 10° of radial deviation; and 0°, 10°, 20°,
ing a consistent posteroanterior (PA) view. Similarly, and 30° of ulnar deviation, as measured by the manual
when measuring in the sagittal plane, the intramedullary goniometer, allowing a 5° range at each position to
rods were aligned in the coronal plane under fluoros- confound any attempt by the raters to guess wrist po-
copy and verified using Adobe Photoshop. The speci- sition. The wrist was then imaged with fluoroscopy at
men was then rotated along the longitudinal (Z) axis each set position, with both PA and lateral exposures of
until the scaphoid tubercle and pisiform were maxi- the wrist, to ensure neutral rotation in the unmeasured

JHS 䉬 Vol A, October 


1426 RELIABILITY IN MEASURING WRIST RANGE OF MOTION

FIGURE 4: Demonstration of the dorsal technique for measurement of A ulnar and B radial deviation as used on a cadaveric
specimen instrumented with a bridging single-bar external fixator.

sures of the wrist were captured under fluoroscopy to


ensure neutral alignment in the plane orthogonal to the
plane of measurement and for computerized measure-
ment of the wrist angle. At each position, the raters
measured all wrists with one technique, in a randomized
order. Each cadaver was then measured again in a
separate randomized order, using the same technique in
a second trial without repositioning the wrists. The
process was repeated for each position (full flexion,
extension, radial deviation, and ulnar deviation) and for
each technique. Measurements of radial and ulnar de-
viation were performed with the dorsal–volar technique
only.
Intraclass correlation coefficients were completed to
determine inter- and intra-rater reliability, using the
JMP6 software package (SAS, Cary, NC). To deter-
FIGURE 5: Screen shot of digital image during measurement mine accuracy, the root mean squared error (RMSE)
of the angle created by the intramedullary rods. and mean deviation were calculated between comput-
erized measurements and manual goniometric values
plane. This enabled computerized verification of wrist for each technique over the range of positions. Statisti-
position in the plane orthogonal to the axis of the wrist cally significant differences between mean deviations
angle. All 10 wrists were measured by 1 rater with one were determined using analysis of variance, using an
technique. This was then repeated for each goniometric alpha level of p ⬍ .05.
technique, and the sequence of wrists was randomized
for each technique to confound any potential measure- RESULTS
ment recall. When measuring the neutral position, the RMSE for
To evaluate the inter- and intra-rater reliability of accuracy of the ulnar, radial, and dorsal–volar tech-
each technique, wrists were positioned at angles of full niques were 4°, 5°, and 4°, respectively (Table 1). No
flexion, full extension, full radial deviation, and full statistically significant difference in accuracy could
ulnar deviation. After positioning, PA and lateral expo- be detected between techniques. For measurement of

JHS 䉬 Vol A, October 


RELIABILITY IN MEASURING WRIST RANGE OF MOTION 1427

TABLE 1. In Vitro Accuracy and Reliability of


(falsely accepting the null hypothesis). It is possible that
Wrist Goniometric Measurement Methods one technique was more accurate than the others, but it
could not be statistically proven in the absence of a
Ulnar Radial Dorsal–Volar great number of specimens. Nonetheless, all accuracy
RMSE overall 7° 7° 8° measurements were within single digits, which we feel
RMSE neutral 4° 5° 4° is an appropriate margin of error; differences between
Mean deviation (p ⫽ .2) 5° 6° 6° techniques of 1° or 2° would not be clinically signifi-
Intra-rater reliability cant. The decision to perform these experiments in
Rater 1 0.8 0.8 1 cadaveric wrists was predicated on the need to mini-
Rater 2 1 0.9 1 mize confounding variables that could jeopardize mea-
Inter-rater reliability 0.5 0.3 0.9
surements of accuracy and reliability. The study re-
quired a gold standard for wrist posture, which was
achieved by the use of intramedullary alignment rods.
Furthermore, were it possible to accurately determine
flexion, extension, radial deviation, and ulnar devia- wrist position with fluoroscopy on live volunteers, the
tion, the range of RMSE values was 7° to 8°; p ⫽ .3 assessment would have exposed human subjects to un-
(Table 1). The mean deviation between fluoroscopic necessary radiation.
values and each goniometric method ranged from 5° The LaStayo et al. study suggests that, among phys-
to 6° and was also not statistically significant; p ⫽ .2 ical therapists, the dorsal–volar technique for measure-
(Table 1). ment of wrist flexion and extension demonstrates the
Each method demonstrated a high degree of intra- highest degree of reliability.2 The results of our study
rater reliability, with the radial technique for measure- support the reliability of the dorsal–volar technique
ment of wrist flexion and extension demonstrating the between a surgeon and a hand therapist who used and
lowest intraclass correlation coefficients values (0.8 and practiced a standardized technique. Contrary to the lit-
0.9). Between raters (inter-rater reliability), however, erature, our data indicate that disagreement between
the dorsal–volar technique for measurement of wrist raters of different training backgrounds can be con-
flexion and extension was the most reliable as com- trolled by use of a standardized dorsal–volar tech-
pared to ulnar and radial techniques: 0.9, 0.5, and 0.3, nique.3
respectively (Table 1). We were unable to reject the null hypothesis, that
DISCUSSION there was no difference in accuracy between the 3
techniques for measurement of wrist position. Each
In this study, we demonstrated that hand surgeons and
technique was accurate to within 5° of wrist position
therapists can attain accuracy within 7°, using the man-
when compared to the gold standard of fluoroscopic
ual goniometer and standardized techniques of wrist
measurement. Although no statistically significant dif-
measurement. This is important because the manual
ference was discovered, the 5° of inaccuracy inherent to
goniometer is ubiquitous in orthopedic and hand sur-
each technique supports the use of a 1°-increment go-
gery. Quantifying wrist range of motion allows the
surgeon to determine patient progress and is a particu- niometer clinically. Often manual goniometers are
larly vital communication tool between hand surgeons, marked with 5° increments. This inaccuracy can poten-
hand therapists, and other health care providers. In tially compound the examiner’s error, resulting in up to
addition, electronic techniques to measure wrist motion 10° of inaccuracy.
(electrogoniometry, data gloves, and external motion We were able to prove our secondary hypothesis,
analysis systems) are increasingly used in both clinical that the dorsal–volar technique demonstrates the highest
and research arenas; verification of the accuracy and degree of reliability for measurement of wrist flexion
reliability of manual goniometry is necessary as a and extension. Based on these findings, the authors
means to validate these instruments and techniques. To recommend the dorsal–volar measurement technique
our knowledge, no study has examined the accuracy of for assessment of wrist ROM. This method promotes
manual goniometry, the repeatability of different tech- the greatest agreement between raters, as is necessary in
niques of measurement, or the inter-rater reliability be- communication between physicians and between phy-
tween surgeon and hand therapist. sician and therapist, and it is consistent with the cur-
The use of cadaveric specimens limited the sample rently accepted use of the dorsal–volar technique for
size of our study, raising the possibility of a beta error radial and ulnar deviation.

JHS 䉬 Vol A, October 


1428 RELIABILITY IN MEASURING WRIST RANGE OF MOTION

REFERENCES 6. Edgar D, Finlay V, Wu A, Wood F. Goniometry and linear assess-


ments to monitor movement outcomes: Are they reliable tools in
1. Casanova JS. Clinical assessment recommendations. 2nd ed. Garner, burn survivors? Burns 2009;35:58 – 62.
NC: American Society for Hand Therapy, 1992:62– 63. 7. Neu CP, Crisco JJ, Wolfe SW. In vivo kinematic behavior of the
2. LaStayo PC, Wheeler DL. Reliability of passive wrist flexion and radio-capitate joint during wrist flexion-extension and radio-ulnar
extension goniometric measurements: a multicenter study. Phys Ther deviation. J Biomech 2001;34:1429 –1438.
1994;74:162–174. 8. Youm Y, Flatt AE. Kinematics of the wrist. Clin Orthop Relat Res
3. Horger MM. The reliability of goniometric measurements of 1980;149:21–32.
active and passive wrist motions. Am J Occup Ther 1990;44:342– 9. Guyot L. Comparison between direct clinical and digital photogram-
348. metric measurements in patients with 22q11 microdeletion. Intl
4. Boone DC, Azen SP. Normal range of motion of joints in male J Oral Maxillofac Surg 2003;32:246 –252.
subjects. J Bone Joint Surg 1979;61A:756 –759. 10. Yang Z, Mann FA, Gilula LA, Haerr C, Larsen CF. Scaphopisocapi-
5. Cobe H. The range of active motion at the wrist of white adults. tate alignment: criterion to establish a neutral lateral view of the
J Bone Joint Surg 1928;10:763–764. wrist. Radiology 1997;205:865– 869.

JHS 䉬 Vol A, October 

You might also like