SPECIFICATIONS SPECIFICATIONS

DEFIBRILLATOR SYSTEM

GENERAL DEFIBRILLATOR ECG VIA DEFIBRILLATION ELECTRODES RECORDER (OPTION)


• Size: 31 (h) x 36 (w) x 25 (d) cm • Waveform: Star™ biphasic truncated • Mode: Differential • Printout modes: Manual or automatic
• Weight: 7.6 kg incl. defibrillation electrodes exponential waveform • Frequency response (-3 dB): Display, 2 to 25 (selectable)
and battery. • Energy settings: AED: #1=150 J, #2=150 J, Hz; "Filter" 2 to 30 Hz • Printout time: Selectable:
9 kg incl. paddles, battery, pacer and recorder #3=150 J or as set in configuration menu • Impedance range: 20 to 2000 Ohm measured at 15, 20, 30 sec, continuous
• Power: 100-240 VAC, 50/60 Hz, max. 270W • External paddles and defibrillation electrodes: 110 kHz • Speed: 25 mm/sec
• Temperature: Operating: 0 to 40°C. Selectable: 2, 5, 7, 10, 20, 30, 40, 50, 60, 80, • Pace pulse limitation: Slew rate limitation • Paper: Thermosensitive,
Storage: -10°C to +60°C (<168h) 100, 150, 200, 270 J without marker 50 mm (w) x approx. 25 m (l)
• Humidity: Operating: 30 to 95% RH, Internal defibrillation electrodes: 1, 2, 3, 4, 5, • Mode: ECG delayed by 8 sec
ECG VIA ECG ELECTRODES
non-condensing. 6, 7, 8, 9, 10, 15, 20, 30, 40, 50 J • Weight: 370 g
Storage: 30 to 95% RH, non-condensing • Energy availability : Auto disarm after 30 sec • Mode: Differential
MEMORY CARD (OPTION)
• Operating and storage pressure: 860 kPa • Charging time (max. energy) at 15-35°C : • Frequency response (-3 dB): Display, 0.25 to
to 1060 kPa < 5 sec running on mains supply 25 Hz; Recorder, 0.25 to 30 Hz; "Monitor" 0.5 • Size: 256 KByte - 16 MByte
• Classification: Class I/Internally powered < 7 sec running on battery to 30 Hz; "No filter" 0.05 to 105 Hz at 0.5 • Data storage time: Min. 1 year
equipment • Synchronization : Internal generation of QRS mVpp input • Type: SRAM 68 pin
• Water protection: Drip-proof equipment (IP22) sync • Impedance: 20 to 2000 Ohm at 110 kHz • Conforming to JEIDA 4.0
• Mode of operation: Continuous monitoring • Synchronization shock delay: < 60 msec from • Pace pulse limitation: Discriminator with • Card Standard PCMCIA 1.0
top of QRS marker. Max. pace rate 180 b/min • Storage capacity (approximate number
• Synchronization marker: "S" with a vertical of sequences of 28 sec):
ECG FROM MONITOR
line 256k=35; 512k=70; 1M=140; 2M=280;
• Classification: Ext. paddles: BF. • Dynamic range: +/-5 V (1 V/mV signal) 4M=570; 8M=1140; 16M=2280
Int. defibrillation electrodes: CF • Max ECG signal delay in monitor: 5 msec

BATTERY
EXT. INTERFACE FOR REMOTE SERVICE (OPTION)
• Sensor: Nellcor reusable finger sensor
Cardio-Aid® 200
ECG • Saturation: 1-100 % SpO2. TM
• Type: Built-in, quick-charge NiCd, 12 V 2.0 Ah
• Operating time at 20°C: 2 h or 50 shocks
Accuracy: (70-100 %) ± 3% STAR BIPHASIC
• Lead selector: Selectable: I, II, III, P, Ext
• Pulse Rate: 20-300 bpm.
• QRS volume: Selectable 0 to 5 of 270 J
Accuracy: (20-250) ± 3 bpm
• Gain: x0.25, x0.5, x1, x2, x4, auto (default • Charging time for fully depleted battery: 1 h
setting) • Life time: Min. 18 months at normal usage EXT. INTERFACE FOR REMOTE SERVICE (OPTION)
• CMRR (Common mode rejection ratio): (i.e. up to one charge cycle per day)
• External interface via PCMCIA port:
Typically 110 dB (50/60 Hz) (with notch filter)
at the cardiac arrest • Differential input: +/-5 mV
PACER (OPTION) PCMCIA to RS 232 adaptor card 2 port
• Parity: none
• Overload recovery time: < 4 sec typically • Mode: VVI demand or VVO fixed
• Data bits: 8
• CMDR (Common mode dynamic range): +/-5 V • Pulse type: Rectangular, constant current
• Stop bit: 1
• Differential dynamic range: +/-360 mV • Pulse duration: 40 msec +/- 10%
• Flow control: XON/XOFF
• Classification: CF, defi-proof • Pulse amplitude: Variable 20-200 mA
• Baud rate: 9600
• QRS detection trigger level: Typically 200 µV +/- 10%. Default 0 mA
LCD DISPLAY • Heart rate alarm limits (Manual/Pacer mode, • Pulse adjust step size: +/-1 mA (+/-5 mA after
adjustable in step for 1 b/min). 1 sec)
• Type: Black and white LCD display with CCFL ECG manual alarm upper limit: Max. 300 b/min. • Pacing rate: Variable 30-180 p/min +/- 5%.
backlight ECG manual alarm lower limit: Min 30 b/min. Default 60 p/min
• Size: 120 (w) x 90 (h) mm Heart rate, alarm delay: Low and high limit: • Pace rate adjust step size: +/-1 p/min
• Resolution: 320 x 240 pixels < 10 sec. (+ 5-20 p/min after 1 sec)
• Mode: Non-fade, erasing bar HR display: Steps of 1 b/min. • Output voltage: Max. 350 V
• ECG window: 4.0 sec
• ECG speed: 25 mm/sec

INNOMED MEDICAL
CE marked according to Council Directive
Medical Developing and Manufacturing Inc.
93/42 EEC of 14 June 1993 concerning medical
Budapest, H-1146, Szabó József u. 12.
devices.
Phone: (36-1) 460-9200 • Fax: (36-1) 460-9222
Specifications subject to change.
Website: www.innomed.hu • E-mail: innomed@innomed.hu
99752-A

INNOMED MEDICAL