Professional Documents
Culture Documents
Bleasesirius Anesthetic Machine: Technical Manual
Bleasesirius Anesthetic Machine: Technical Manual
Anesthetic Machine
Technical Manual
073-0227-00/REV. D1
2
BleaseSirius
Anesthetic Machine
With 700/900 Series Ventilators
Technical
Manual
MODIFICATIONS LABEL
ECO A REV B REV C REV D REV E
3
4
Important
For all users and Service Personnel, refer to the User Manual before operating
the machine.
All rights reserved. The information contained in this publication may not be
used for any other purpose than that for which it was originally supplied.
This publication may not be reproduced in part or in whole without the written
consent of Spacelabs Healthcare.
5
This page is intentionally left blank
6
Contents
Chapter 1-Technical Description............................................................... 19
1.1 Description .........................................................................................................21
1.1.1 General ................................................................................................21
1.1.2 The Frame............................................................................................21
1.1.3 Pneumatic Assembly ............................................................................21
1.1.4 The Monitor Shelf .................................................................................21
1.2 Specification ......................................................................................................24
1.2.1 Machine Dimension..............................................................................24
1.2.2 Work Surface Dimensions....................................................................24
1.2.3 Monitor Shelf Dimensions ....................................................................24
1.2.4 Maximum Loading ................................................................................24
1.3 Pneumatics ........................................................................................................25
1.3.1 Gas-Specific Color Specifications ........................................................25
1.3.2 Gas Supply Combinations....................................................................25
1.3.3 Common Gas Outlet ............................................................................25
1.4 Technical/Performance Specification ..............................................................26
1.4.1 Controls ................................................................................................26
1.4.2 Ventilator ..............................................................................................26
1.4.3 Alarms / Indicators ...............................................................................26
1.4.4 Regulator Safety Valve Settings...........................................................26
1.4.5 Electrical...............................................................................................27
1.4.6 Supplies ...............................................................................................28
1.4.7 Environmental ......................................................................................28
7
Contents
8
Contents
9
Contents
10
Contents
11
Contents
Chapter 7
12
Contents
13
Figures
15
Figures
16
Figures
17
List of Schematics
18
BleaseSirius
Anesthetic Machine
Chapter 1
Technical Description
19
Technical Description
20
Technical Description
1.1 Description
1.1.1 General
The BleaseSirius anesthetic machines contain all the pneumatic circuitry,
controls, monitoring, ancillaries and storage required to control, distribute and
mix medical gases and anesthetic agents in order to deliver them to a patient
system.
In US markets, use a cord fitted with a NEMA 5-15 hospital grade plug to connect
the BleaseSirius to the mains supply. Connection of equipment to the auxiliary
mains socket outlets may increase leakage currents to values exceeding the
allowable limits.
21
Technical Description
R
Q
A
P
B
O
C
D
E
N G
M H
I
L
J
K
22
Technical Description
Figure 1 Key
A Monitor shelf
B Ventilator
C Vaporizer
D Cylinder/pipeline gauges
E Pneumatic unit - (behind gauge panel)
F Handle
G Oxygen flush
H Common gas outlet (or option A.C.G.O.)
I Drawer
J Frame
K Brake Pedal
L Absorber
M Bag Arm Link Pipe
N Adjustable Bag Arm
O Main On/Off Switch
P Suction Controller
Q Flow Control Valves with Hypoxic Guard
R Flowblock Assembly
S Auxiliary Flowmeter
23
Technical Description
1.2 Specification
Height Area
854mm / 33.6” 98612.2mm2 / 152.8”2
Height Area
1486mm/58.5” 175833mm2 / 272.54”2
24
Technical Description
1.3 Pneumatics
25
Technical Description
1.4.1 Controls
Oxygen flow 150ml/m to 10 l/m Simplex/ Cascade
Nitrous oxide flow 0ml/m to 12 l/m Simplex/Cascade
MED AIR flow 0ml/m to 15 l/m
1.4.2 Ventilator
A Blease700/900 ventilator is built
into the BleaseSirius.
26
Technical Description
1.4.5 Electrical
The auxiliary sockets are numbered 1 through 4, top to bottom and are rated
as in the following tables. If the machine has the switching sockets option,
sockets 3 and 4 will switch On and Off when the Main On/Off (Figure 1, Key
O) is switched On and Off, rather than with the Mains On/Off switch (Figure 2).
230V 110V
Outlet Outlet
Socket Breaker Fuse Socket Breaker Fuse
Rating Rating
1 2A 3A 5A 1-4 2A 3A 4A
2-4 1A 2A 3.15A
27
Technical Description
1.4.6 Supplies
O2, MED AIR, N2O pipeline Nominal pressure 400 kPa/58.0 psi,
minimum 275 kPa/39.8 psi, maximum
482 kPa/69.9 psi
Auxiliary pneumatic outlets MED AIR or O2 - 400 kPa/58.0 psi at
zero flow. 80 l/m max. flow
Auxiliary Oxygen Outlet 0-15 lpm
1.4.7 Environmental
Operating Temperature 5ºC-40ºC (41ºF-104ºF)
oxygen cell operates to specification
10ºC-40ºC (50ºF-104ºF)
28
BleaseSirius
Anesthetic Machine
Chapter 2
Overview
29
Overview
30
Overview
When the user changes a value such that it does not comply with the required
difference, the box surrounding the feature being adjusted and the value of
the feature causing limiting to be enforced is highlighted red. The value being
adjusted could either be the PEEP level or the relevant pressure control.
If the user does not make the change before end of time allowed for adjustment,
the value will set to the nearest acceptable value.
31
Overview
2.1.1.2 Trigger
The level of negative flow caused by the patient’s attempt to breathe before the
ventilator initiates flow to assist the patient with the spontaneous breath.
Range: 1 to 15 l/m all modes (SIMV + PSV and Pressure Support).
2.1.1.5 Oxygen
A fuel cell sensor can be connected to allow monitoring of 0% to 110% O2, with
alarms of 18% to 110%. The sensor can be calibrated.
32
Overview
2.1.1.14 Compliance
System compliance in ml/cmH2O
Patient compliance in ml/cmH2O
33
Overview
Effective TV = TV - TV due to FG
TVE = TV - TVFG
Therefore,
Effective TV = 600 - 167 = 433mL
2.2.2 Compliance
Compression of gas in the dead space within the breathing system reduces the
tidal volume delivered to the patient. In an ideal ventilator, the Set TV would
be the volume of gas that is delivered to the patient’s lungs. This cannot be
achieved because the anatomy of the patient is unknown. However, the Set
TV can be accurately delivered from the catheter mount, thus reducing Set TV
errors to a minimum.
34
Overview
The patient airway flow sensor head must be in the patient circuit
in order to carry out compliance compensation.
The ventilator delivers a breath of known volume to the breathing circuit, records
the pressure (cmH2O) achieved and verifies that a leak is not present.
Volume (mL)
Pressure (cmH20) = dead space compliance (C3)
This value is retained in memory until the ventilator is switched off or retested.
Running the ventilator on a patient with a set 500ml TV gave a peak pressure
of 20cmH2O.
35
Overview
The 735ml is the actual ventilator output into the breathing circuit to give
500ml at the catheter mount. The value must be recalculated every time volume
controlled ventilation starts.
8
new TV = 500 * (1+ 25 - 8
)= 735mL
2.2.3 Compensation
36
Overview
PRESSURE Pressure
SUPPORT O O ALWAYS ALWAYS Meas Meas Meas
Limit
SIMV-VC + Meas/ Meas/ Meas/ Pressure
PSV O P ALWAYS ALWAYS
Set Set Set Limit
SIMV-PC + Meas/ Meas/
PSV O O ALWAYS ALWAYS
Set O Set
Ptot
37
Overview
38
Overview
39
Overview
S T V Y
Z
U
W
M X
R
M O P Q
K
G
L N
J
H
A B C D E F
40
Overview
The flow and pressure are measured and monitored by the microprocessor
feedback system.
Expiration occurs when the flow control valve is closed and the expiratory
solenoid opens and releases the gas from the bellows assembly.
For expiration, the expiratory solenoid is opened which releases the gas from
the bellows assembly.
41
Notes
42
BleaseSirius
Anesthetic Machine
Chapter 3
Planned Maintenance
43
Planned Maintenance
44
Planned Maintenance
Only authorized technical engineers should service the machine, since they
have the appropriate training and qualifications in the use of high-pressure
medical devices. Servicing should take place in a clean and controlled
environment taking particular care to prevent contaminants entering the unit
while disassembled.
45
Planned Maintenance
• Bodocs Seals
• Backbar Seals
• Absorber Stop
+ O2 Blanking
Plug
• Backbar Dzus
Springs
• AGSS Probe
Seal
• AGSS Float
Seal
• Pipeline Filter
• Pipeline ‘O’ Ring
• Bellows Base/
Cover
• Ped Bellows
• Bellows Base/
Pop off
• Absorber Upper
Seal
• Absorber ‘O’
Rings for Insp/
Exp + APL
• Pop off Valve
• Absorber Gauge
• Absorber
Control Seal
• Absorber
Bottom Canister
46
Planned Maintenance
• Cylinder
Regulator
Service Kit
• ‘O’ Ring for
Filter
• Battery
Note:
Includes annual
kit, PN 14000511
47
Planned Maintenance
48
Planned Maintenance
4. Refit the outer part of the connector and tighten the two small grub
screws.
5. Repeat for all three pipelines.
6. Note each gas type has a unique fitting and can not be incorrectly
mixed.
7. Refit the pipeline and check for leaks.
49
Planned Maintenance
50
Planned Maintenance
51
Planned Maintenance
A - Cover
B - Bellows
A C - Pop-off valve
D - Valve seat
E - Fixing screws
F - Bellows base
G - O-ring
G D
E
52
Planned Maintenance
3.6 Absorber
3.6.1 Valve Covers
When removing or refitting any of these valves, it is important to
ensure the metal valve disk and seat are not damaged.
1. Remove the expiratory valve by twisting the clear plastic valve cover.
2. Replace the ‘O’ Ring seal on the valve cover.
3. Replace the expiratory valve on the absorber.
4. Remove the inspiratory valve cover.
5. Replace the ‘O’ ring seal.
6. Replace the valve cover.
7. Remove the APL valve.
8. Replace the ‘O’ ring seal.
9. Replace the APL valve.
3.6.2 Manometer
1. Remove the manometer gauge by pressing the quick release lever at the
base of the gauge.
2. Replace the ‘O’ ring seal.
3. Refit the gauge.
53
Planned Maintenance
3.6.5.1 Two Valve Version Only (on units before Feb 2009)
1. Ensure all gas supplies are disconnected.
2. See Section 5.1 for removal of top surface and front cover.
3. Once the front cover is removed, you can access the 6 screws that secure
the top of the ventilator casing.
4. The front display panel will hinge forward a little. This will allow the top
cover to be hinged back.
5. The filler cover is in the inlet manifold which can be seen from the front
of the unit.
6. Remove the cover.
7. The filter can then be
unscrewed.
8. Replace the filter and cover
and make sure the new ‘O’
ring is fitted.
9. Replace the screws in the
top cover and front display
panel.
10. Reconnect the O2 pipeline
only and turn on the system.
11. Check that there are no Figure 13 - Pneumatic Module
leaks.
12. Turn the systems OFF.
13. Fit the front cover and top surface.
54
Planned Maintenance
Disassembly
Assembly
Adjustment
Turn the adjusting screw clockwise to increase and counterclockwise to
decrease outlet pressure setting. To reduce pressure, first reduce to a
pressure less than that desired, then increase to the desired outlet pressure.
Turn the adjusting screw using a 1/8-inch hex key.
55
Planned Maintenance
3.7 CLEANING
Cleaning of external surfaces is possible using water or isopropyl alcohol.
Cleaning of internal surfaces and parts should not be required and is not
recommended.
DO NOT use any lubricants that are not compatible with oxy-
gen or medical gases. The use of lubricants not compatible
with oxygen may pose a risk of fire or explosion.
Open the cylinder valve slowly to prevent the risk of fire or explosion due to
oxygen pressure shocks. The relief valve is a safety device and must only be
removed for servicing. When servicing is complete, a new relief valve must be
fitted. Do not attempt to adjust or tamper with the relief valve.
56
Planned Maintenance
Adjusting Screw
Spring Locknut
Slip Ring
Diaphragm Assembly
Actuator Assembly
Seal Ring
Valve Cartridge
‘O’ Ring
Body
Mounting Spacer
Relief Valve
57
Notes
58
BleaseSirius
Anesthetic Machine
Chapter 4
System Checks
59
System Checks
60
System Checks
4.1 Overview
This chapter explains the systems checks that must be used on all BleaseSirius
anesthetic machines whenever any of the following procedures has been
performed:
• Installation
• Planned Maintenance (should be performed annually)
• Service
• Repair
Use the service check sheet, shown on page 64, to record the results.
61
System Checks
Parts Required
Description Part Number Quantity
62
System Checks
63
Hospital: . SiriusSN: . Customer
PrintName: .
64
Address: . FrontPanelSW:. Signature: .
........ Date: .
City:.ST:. BavControlerSW:.
....... Representative
Contact: .. Model#: . PrintName: .
Signature: .
PhoneNumber: . AbsorberSN: . Date: .
Visual Inspection High Pressure Outlets (= or > 80LPM) Ventilator
Description Pass Fail N/A Description Pass Fail N/A Description Pass Fail N/A
Cylinder Yokes O2 Shraeder (if fitted) Compliance test
Pipeline Inlets O2 diss (if fitted) Systems check
Pipeline Hoses Air Shraeder (if fitted) CALBK: flow control cal
Absorber Rods Hypoxic Guard/Flow Control Accuracy CALBK: inspiratory valve cal
Hp Outlet (if fitted) Description Pass Fail N/A CALBK: block and pressure zeros
Backbar Flowmeter backlight (non-EFM) Touchscreen calibration
O2 Flush Assy Min. O2 flow (130 - 170ml) Fresh gas calibration
Common Gas Outlet N2O = 1LPM (O2 = 21 - 30 %) VTI = 250 ±25ML
Flowmeter Assy / Control Knobs N2O = 6LPM (O2 = 21 - 30 %) VTI = 500 ±50ML
Drawers N2O = 12LPM (O2 = 21 - 35 %) VTI = 750 ±75ML
Castors O2 = 3LPM (O2 = 21 - 30 %) PEEP = 10 ±1.5cmH2O
Pipeline Leak Test O2 = 1.5LPM (O2 = 21 - 30 %) Pressure = 10 ±2cmH2O
Description Pass Fail N/A O2 = 0.4LPM (O2 = 21 - 30 %) Pressure = 30 ±3cmH2O
O2 O2 & N2O = 3LPM (O2 = 47 - 53 %) Pressure High Alarm
N2O (if fitted) O2 & N2O = 6LPM (O2 = 47 - 53 %) Alarms
Air (if fitted) O2 & Air = 3LPM (O2 = 57 - 63 %) Description Pass Fail N/A
Cylinder Leak Test O2 & Air = 6LPM (O2 = 57 - 63 %) Apnea
Description Pass Fail N/A Backbar Sustained pressure
System Checks
Check that the following components are free of damage and are securely
mounted to the machine and/or fittings.
1. Cylinder Yokes
Verify the yoke block, index pins and bridges are free of damage and that the
tee screw freely threads into the yoke bridge.
2. Pipeline Inlets
Verify that the pipeline inlets are free of damage and inspect the threads for
damage as well. Connect pipeline hoses to each inlet and ensure that the hoses
connect freely.
3. Pipeline Hoses
Verify that pipeline hoses are free of deterioration and damage. Connect both
ends of the hoses and pull them to check for a secure connection.
4. Absorber Rods
Verify that the rods are free of damage, secure and not bent.
6. Backbar
Verify the backbar and backbar valves are free of damage. Check that the
o-rings on the valves are free of damage.
65
System Checks
10. Drawers
Verify that the drawer(s) are free of damage. Open and close the drawers to
check for a smooth operation.
11. Castors
Verify that the castors are free of damage. Check that when the castors brakes
are not engaged, the machine can move freely in any direction. Listen for noise.
When the brake is engaged, ensure that the machine cannot be moved.
66
System Checks
6. Connect the appropriate pipeline hose to the pipeline test shut-off valve
assembly, and then connect it to the hospital supply.
7. Turn the pipeline test shut-off valve “ON” to allow flow.
8. Verify that the appropriate gauge registers.
9. Wait 15 seconds for the system to pressurize and shut the pipeline test
shut-off valve off.
10. Disconnect the pipeline hose from the hospital supply and verify that the
appropriate digital gauge is still registering.
11. Verify that the pipeline gauge does not fall 50 kPa, half of a division in
three minutes.
12. If the leak test is passed, disconnect the pipeline test shut-off valve from
the machine and verify that the pipeline gauges value drops to zero.
13. Repeat steps 5-12 for all additional pipelines fitted.
67
System Checks
68
System Checks
4. Disconnect the 6 mm plug from the regulator test point, then connect a
pressure measuring device to the test point.
9. Repeat steps 3-8 for all regulators. (Note: When testing outputs or making
adjustments on N2O regulators, ensure that an O2 supply is connected to
the O2 pipeline inlet.)
69
System Checks
70
System Checks
Only order parts that are required for testing, depending on which high pressure
outlet is fitted to the machine.
71
System Checks
72
System Checks
73
System Checks
5. Turn the power switch to the “ON” position. Set of flow of 0.3 L/min on the
O2 flow control valve.
6. Ensure that a pressure of 150 mmHg is reached within 7 seconds, then
turn the power switch to the “OFF” position.
74
System Checks
75
System Checks
76
System Checks
77
System Checks
78
System Checks
79
System Checks
80
System Checks
81
System Checks
82
System Checks
Figure 37 - Exhaust
Port Location
83
System Checks
10. Perform the “Inspiratory Valve Cal” by selecting this option on the
touchscreen. Once complete, reconnect the drive gas hose onto the drive
gas fitting.
84
System Checks
20. Check that the inspired volume (VTI) on the flow measuring device is
250 ± 25 ml. Once complete, switch the ventilator back to adult mode and
use the adult flow sensor for the remaining checks.
21. Increase the tidal volume to 500 ml. Wait for the ventilator to stabilize.
85
System Checks
22. Check that the inspired volume (VTI) on the flow measuring device is
500 ±50 ml.
23. Increase the tidal volume to 750 ml. Wait for the ventilator to stabilize.
24. Check that the inspired volume (VTI) on the flow measuring device is
750 ±75 ml.
86
System Checks
87
System Checks
41. Decrease the oxygen lower limit to 18% and the higher limit to 19%. Verify
that the Oxygen High alarm is displayed on the ventilator.
42. Disconnect the O2 sensor, then verify that the O2 Sensor Error alarm is
displayed on the ventilator.
43. Disconnect the 700/900 ventilator AC power cord. Verify that the Mains
Fail alarm is displayed on the ventilator and that the green LED light turns
red.
44. Disconnect the AC power cord. Verify that the mains fail alarm is
displayed, the battery icon is displayed on the ventilator and that the
green LED light turns red.
45. Press Setup > Configuration > System Information. Display the second
System Information screen and verify that the second battery voltage
(Switched) is greater than 11V. If not greater than 11V, ensure the unit
has had the battery continually charging for a minimum of 8hrs before
verifying. If the voltage is still low after 8 hours of continuous charge,
replace the battery.
46. Slightly dismount the absorber. Verify that the Absorber Not Fitted alarm is
displayed on the ventilator.
47. Remount the absorber.
88
System Checks
89
System Checks
6. Connect a pressure sample tee to the CGO taper and the CGO boot to
the other end of the pressure sample tee. (Note: For ACGO units, connect
a 22 mm reusable hose from the other end of the pressure sampling tee
to the AGSS auxiliary port and set the switch to ACGO.)
7. Ensure that the pressure sampling tee is connected to a gas analyzer and
ensure that the gas analyzer is also connected to an active scavenger.
8. Ensure that the bag-to-vent switch on the absorber is set to Bag. Open
the APL valve by turning it fully counter-clockwise.
9. Turn the power switch to the “ON” position.
10. Set a flow of 10 L/min on the O2 flow control valve to flush the system of
any residual gas. Reduce the flow to 4 L/min after one minute.
11. Set the vaporizer dial to 1.0% and wait a few minutes for the vaporizer
concentration to stabilize.
90
System Checks
12. Verify that the reading on the gas analyzer is between 0.8 – 1.2%.
13. Set the vaporizer dial to 2.5% and wait a few minutes for the vaporizer
concentration to stabilize.
14. Verify that the reading on the gas analyzer is between 2.0 – 3.0%.
15. Set the vaporizer dial to 5.0% and wait a few minutes for the vaporizer
concentration to stabilize.
16. Verify that the reading on the gas analyzer is between 4.0 – 6.0%.
17. If a Sevoflourane vaporizer is being tested, set the vaporizer dial to 8.0%
and wait a few minutes for the vaporizer concentration to stabilize.
18. Verify that the reading on the gas analyzer is between 6.4 – 9.6%.
19. If a Desflurane vaporizer is being tested, set the vaporizer dial to 10.0%
and wait a few minutes for the vaporizer concentration to stabilize.
20. Verify that the reading on the gas analyzer is between 8.0 – 12.0%.
21. Set the vaporizer dial to 18.0% and wait a few minutes for the vaporizer
concentration to stabilize.
22. Verify that the reading on the gas analyzer is between 14.4 – 21.6%.
23. Set the vaporizer dial to “0”.
24. Set the O2 flow control valve to minimum flow and turn the switch to the
“OFF” position.
91
System Checks
92
BleaseSirius
Anesthetic Machine
Chapter 5
Detailed Repair
Procedures
93
Detailed Repair Procedures
94
Detailed Repair Procedures
1. Remove the 4 screws shown in Figure 45 and lift the top cover off.
95
Detailed Repair Procedures
96
Detailed Repair Procedures
Figure 49 - Front Cover Fixing Screws (on units after Jan 2009)
97
Detailed Repair Procedures
98
Detailed Repair Procedures
99
Detailed Repair Procedures
100
Detailed Repair Procedures
Test
Points
Fuses
101
Detailed Repair Procedures
Regulators
102
Detailed Repair Procedures
5.2.1 Introduction
The hypoxic guard prevents the operator from administering a mixture of gases
that contain less than 21% oxygen. To achieve this, the oxygen and nitrous
oxide control valves are linked via a set of gears. Any attempt to reduce the flow
of oxygen or to increase the nitrous oxide flow to a level which would result in
a oxygen concentration below 21%, results in the nitrous oxide valve closing or
the oxygen valve opening to maintain the 21% minimum oxygen concentration.
103
Detailed Repair Procedures
2. Remove the two M4 screws each side of the flow meter while supporting
the flow meter.
104
Detailed Repair Procedures
3. Carefully disconnect the power supply cable from the back light inverter
on the rear of the flow meter (marked BL1 in Figure 55).
BL1
Connector
4. Mark the positions of the pipes entering the flow meter, and remove them
from their fittings.
105
Detailed Repair Procedures
106
Detailed Repair Procedures
Gas Output
N2O Supply
If only Simplex Gas tubes are fitted, the pipes running from
the top to the bottom will not be fitted.
107
Detailed Repair Procedures
If it is necessary to recalibrate the block, then use the following method. It should
be noted that this operation can take some time for the inexperienced engineer
to master and should only be used if a replacement block is not available. If it
proves impossible to achieve a satisfactory result, then the precision valves are
most likely to have become damaged and a new block is the only solution.
108
Detailed Repair Procedures
1. Ensure that the Air and Nitrous oxide valves are shutting fully and not
overly tight. Follow the procedure, “Setting Valve shut off stop” (Section
5.4).
3. Next, set the Oxygen Basal flow, Fit the oxygen gear to the valve and
screw it fully on, but DO NOT tighten it. It must be free running.
4. Slide the oxygen stop arm onto the valve shaft and push it up to the
oxygen gear just fitted. You will notice there is a groove around the gear
into which the small drive ‘peg’ on the back of the stop should fit.
5. Turn the stop arm clockwise until it hits the stop screw. Turn the oxygen
gear counterclockwise until you feel the drive ‘dog’ hit the end of the
groove. (This should give about one full turn of the gear from its fully back
position.)
6. There must be a small clearance between the gear face and the back
face of the stop arm to ensure that the only point of contact is the ‘drive
dog.’ A 0.005-inch feeler gauge can be used or the thickness of a piece of
standard writing paper is a useful alternative (Figure 75).
7. Turn on the machine so that you have an oxygen supply to the flow meter.
8. Hold the stop arm against the stop screw in a clockwise direction and the
gear turned fully counterclockwise and hitting the ‘drive peg’.
109
Detailed Repair Procedures
9. Turn the shaft of the valve so that oxygen starts to flow. Adjust the flow
to the prescribed Basal flow level (still holding the gear and stop arm in
position). When this is achieved, tighten the two grub screws on the stop
arm.
10. The stop screw height must now be adjusted. Turn the valve a full turn
counterclockwise and check that it just clears the top of the stop screw. If
not, adjust the stop by screwing it in till the arm JUST clears and tighten
the lock nut. Check that the arm hits the stop when turned clockwise and
clears the top when turned counterclockwise.
11. The ‘Idler gear’ must now be meshed to link the two valves. Turn the
stop arm clockwise to the stop and hold, and turn the oxygen gear fully
counterclockwise, (as in Step 8). This should give you the ‘basal flow’
showing on the oxygen flow meter. Turn the nitrous oxide flow up to
approximately 400 ml and, while holding the stop and gear in position,
carefully insert the idler gear and its bush onto it’s shaft. Meshing it’s teeth
with the oxygen and nitrous oxide gears, then fit the idler gear nut.
12. Carry out a full check of the hypoxic guard calibration as described in
Chapter 4.
1. Remove the front cover (Section 5.1). Turn all flow controls to minimum.
Turn the machine on and allow basal flow on the oxygen to stabilize for a
few minutes.
2. There are two grub screws holding the oxygen stop arm to the valve
shaft. Hold the stop arm tight against the stop screw and slightly slacken
the two screws. Turn the valve shaft until the basal flow is within limits and
retighten the grub screws.
110
Detailed Repair Procedures
3. Check that when the valve is opened that at its next complete revolution,
the stop arm clears the top of the stop screw and that on closing, it
engages firmly with the stop screw. If not, the screw may be adjusted by
slackening its lock nut and screwing the stop in or out to achieve this.
4. Carry out a full check of the hypoxic guard calibration as described in
Chapter 4.
Small variations from the specified Basal flow can usually be rectified using the
above method. If this does not give a satisfactory result, it is possible that, if the
flow had been shut off completely, the valve seat may have been damaged. In
these cases, you may need to replace the block with a new calibrated unit.
To check these valves, first remove the front cover (Section 5.1).
111
Detailed Repair Procedures
Block
Idler Gear
O2 Valve N2O Valve
Bush & Nut AIR Valve
O2 Stop
Stop Collar
O2 Valve
112
Detailed Repair Procedures
Flowmeter
Block
Springs
Filters
Flow Tube
Inserts
Figure 62 - Flow Block Showing Flow Tube Filters and Tube Lower Inserts
113
Detailed Repair Procedures
Note Clearance
Between O2 Gear
and Stop Arm
114
Detailed Repair Procedures
1. Remove the smaller of the flow tubes for the gas to be inserted (Removal
of Flow Tubes, Section 5.10).
2. Remove the spring and lower insert from the block (Figure 63).
3. Ensure that all flow control valves are set minimum and turn on the
machine.
4. Pour a small amount of leak detector fluid such as ‘Snoop’ into the insert
bore of the block (in the absence of ‘Snoop’ plain water is effective) if the
valve is not sealing there will be bubbles visible in the fluid, if there are no
bubbles then the valve is sealing correctly.
5. If there are any bubbles, then the valve stop must be adjusted. The stop
collars of the two valves differ slightly in that the nitrous oxide valve collar
is also the drive gear. Both versions have two Allen grub screws holding
them to the valve shaft.
6. To adjust the valve stop, slacken the grub screws slightly and gently turn
the valve shaft till no bubbles are present and tighten the collar grub
screws.
7. Slightly open and close the valve a couple of times and check that when
the valve is shut there are no bubbles visible.
8. Carefully dry out the insert bore and open the valve to ensure that any
fluid is blown out. The glass flow tube, the insert and spring can then be
replaced.
If, due to damage or wear, it is deemed necessary to replace the hypoxic guard
valve block, you can order a calibrated unit which will be supplied with a test
sheet. This test sheet gives the pressures to which the secondary regulators
should be set.
115
Detailed Repair Procedures
11. Disconnect the pressure gauge and refit the blanking plug.
12. Check the pressure of the secondary nitrous oxide regulator by repeating
steps 3 to 10 for the nitrous oxide regulator.
116
Detailed Repair Procedures
Regulator
Adjustment
2. Adjust the pressure by pulling the adjusting knob to release it. Turn it
clockwise to increase the pressure and counter-clockwise to reduce it. Set
the pressure by adjusting the pressure UP to the desired level to give the
most stable setting. Once the pressure is correct, push the knob towards the
regulator body to lock it in position.
117
Detailed Repair Procedures
7. Locate the secondary regulator test points and gas locations (Figure 67).
118
Detailed Repair Procedures
119
Detailed Repair Procedures
Important!
Monitor the flow to ensure that it remains at 0.5 L/min because the
flow changes when the set screw is adjusted.
120
Detailed Repair Procedures
121
Detailed Repair Procedures
Adjusting Screw
Spring Locknut
Slip Ring
Diaphragm Assembly
Actuator Assembly
Seal Ring
Valve Cartridge
‘O’ Ring
Body
Mounting Spacer
Relief Valve
Figure 70 - Primary Regulator Assembly
122
Detailed Repair Procedures
3. Release the internal spring tension at the bonnet hex nut using a 1/8”-inch
Allen key. Use a crescent wrench to remove the bonnet (Figure 70).
123
Detailed Repair Procedures
4. Remove and discard the diaphragm assembly and the actuator assembly
(Figure 73).
124
Detailed Repair Procedures
6. Fit the regulator parts from the Service Kit (Figure 75).
125
Detailed Repair Procedures
8. Place the upper spring rest and the spring into the bonnet as shown in
Figure 77.
126
Detailed Repair Procedures
11. Fit the actuator assembly into the valve cartridge as shown in Figure 80.
12. Fasten the bonnet to the threads of the regulator body (Figure 81). Ensure
that the assembly stays in place.
127
Detailed Repair Procedures
13. Fit the relief valve to the regulator body (Figure 82). Ensure that an O-ring
is fitted on the relief valve.
128
Detailed Repair Procedures
If you find a bypass leak on any Backbar valve, replace the valve.
1. Loosen the M4 screw on the under side of the backbar, below the valve to
be replaced.
129
Detailed Repair Procedures
2. Use the screw to push the valve upwards to free it from its seals, then
remove the M4 screw completely and withdraw the valve.
3. Lightly lubricate the ‘‘O’-ring seals on the new valve with ‘Fomblin’ Oxygen
safe grease and insert it into the backbar, replace the M4 retaining screw
in the backbar and tighten.
130
Detailed Repair Procedures
4. The block is held in place by three M5 screws. Remove these screws and
withdraw the block from the machine.
5. Replacement is the reverse of this procedure.
Fixing Fixing
Screw 2 Screw 1
131
Detailed Repair Procedures
Fixing
Screws
Air
Regulator
O2
Regulator N2 O
Regulator
132
Detailed Repair Procedures
The CGO block is located inside the main work surface of the machine and
can be seen from the rear of the machine. There are two pipes connected to it
and these pipes are different diameters (6 mm & 8 mm).
The 8 mm pipe carries the mixed gases and vapors from the flow meter and
back bar. The 6 mm pipe is the high pressure supply to the Oxygen flush
control.
O2 Flush
Limit
Adjustment
6mm Pipe
to Flush
Valve
Safety
8mm Valve
Pipe from Adjustment
Backbar
133
Detailed Repair Procedures
5.10.1 Oxygen Flush Adjustment on units sold before July 2010 (Units
sold after July 2010 are not adjustable.)
1. To adjust the Oxygen flush flow, first remove the sealing compound from
the adjusting screw on the CGO block.
2. Set the flow rate to 45 L/min). Turn the adjuster clockwise to reduce the
flow and counter-clockwise to increase the flow.
3. When the flow has been set, seal the adjuster with silicone sealant or
something similar.
Common Gas Outlet Safety Valve Adjustment on units sold before July
2010 (Units sold after July 2010 are not adjustable.)
1. To adjust the safety valve, turn the adjuster clockwise to increase the
pressure or counter-clockwise to reduce the pressure.
2. If the valve is leaking, remove the block and the valve adjuster. The valve
seat and disc bellow can then be accessed and cleaned.
134
Detailed Repair Procedures
Fixing
Screws
8mm Pipe
from Backbar
Safety
Valve
To
6mm Pipe Absorber
to Flush
Valve
1. Remove the rubber gas feed connector from the CGO port on the front of
the machine.
2. Remove the rear panel (see section 5.1).
3. The block is held in place by three M5 screws. Remove these screws and
the block can be removed from the machine.
4. Replacement is the reverse of this procedure
135
Detailed Repair Procedures
3. Remove the (3) screws where indicated to remove the ACGO assembly.
136
Detailed Repair Procedures
5. Remove the 8 mm tubing, indicated in Figure 94, that is routed to the flow
sensor block assembly.
6. Remove the 6 mm tubing from the top 4-6 mm increaser routed to the
8 mm/6 mm tee.
7. Remove the 8mm stem clamped to the fresh gas hose routed to the
absorber manifold.
137
Detailed Repair Procedures
8. Remove the microswitch, indicated in Figure 97. Note the wire orientation
when reinstalling the microswitch.
9. Note the orientation of the tubing, then reinstall the ACGO assembly by
reversing the removal procedure.
138
Detailed Repair Procedures
10. After the ACGO assembly is installed, turn the ACGO switch to “Absorber”
and turn the BleaseSirius on/off switch to the “On” position. Bypass the
pre-use test.
11. Set the Bag/Vent Switch on the absorber to “Vent” and ensure the
ventilator is functioning correctly.
12. Turn the ACGO switch to “ACGO” and ensure ACGO is displayed on the
ventilator screen.
139
Detailed Repair Procedures
1. Ensure the machine is powered off and that no pipelines or cylinders are
connected to the machine.
2. Remove the front cover of the machine to gain access to the switch.
3. Fit the two microswitches to the switch mounting bracket.
140
Detailed Repair Procedures
4. Place the vent wires in the correct orientation and secure them with the
(4) screws.
141
Detailed Repair Procedures
6. Fit (4) 4-6mm increasers into the NC valves and (2) 6-5mm reducers into
the 4-6mm increasers.
142
Detailed Repair Procedures
7. Connect the 5mm tubing to the 6-5mm reducers and the 6mm tubing to
the 4-6mm increasers. See the tubing diagram on the next page.
143
Detailed Repair Procedures
144
Detailed Repair Procedures
8. Tighten the (2) screws, indicated in Figure 105, to secure the switch
assembly.
1. Turn the power switch on and ensure the machine powers up.
2. Set a flow of 2 L/min on all gases. Verify that flow is available.
3. Turn the power switch off. Verify that all flows drop to zero and that the
machine powers off.
4. Turn the power switch back on and ensure that all flows are restored.
5. Perform a successful leak and compliance test.
145
Detailed Repair Procedures
5.12 Absorber
5.12.1 General
Original CAS Absorber starting - PN 12205XX
Introduced with the original Sirius system in 2003 and offered with the choice of
soda lime capacities and CO2 bypass fitted as standard.
Distinguishing features:-
146
Detailed Repair Procedures
Distinguishing features:-
147
Detailed Repair Procedures
Introduced in 2009 with re-routed gas flow path for improved moisture handling
capabilities and is, to date, the current specification. Offered with the choice of
soda lime capacities and CO2 bypass option:
Distinguishing features:-
148
Detailed Repair Procedures
Interchangeable Parts
All the ‘bolt on’ and accessory parts of the CAS absorbers are mechanically
interchangeable among the different models with most differences being color
only.
For 122005** and 122006** series models, besides the color differences, most
main body parts, are interchangeable including bottom moldings, central body
plate, central seal etc. The top molding assembly and manometer are not
interchangeable between MRi and non-MRi units.
Non-Interchangeable parts
The 122009** models do not share any main body (top/bottom moldings, central
body plate, central seal etc.) Interchangeable parts with the 122005** and
122006** models. Also for the 122009** models, the top molding assembly
and manometer are not interchangeable between MRi and non-MRi units.
149
Detailed Repair Procedures
Casting
PN 13600036
Gasket
PN 13600069
The three components are held in place in the main molding with 8 M3 screws.
Do not over-tighten these screws or the threaded inserts in the molding may be
damaged.
If the pipes are removed from the casting (they are different diameters so they
cannot be interchanged before refitting), the mating surface of the casting
should be lightly smeared with a silicone sealant to ensure a gas-tight seal to
the casting.
150
Detailed Repair Procedures
151
Detailed Repair Procedures
The absorber pin is not adjustable. Test the micro switch at the hole to the right
of the short absorber rod to remove the “ABSORBER NOT FITTED” alarm. If
faulty, replace the micro switch
152
Detailed Repair Procedures
153
Detailed Repair Procedures
154
Detailed Repair Procedures
155
Detailed Repair Procedures
2. Remove the gasket to expose the fifteen M4 screws which hold the top
molding to the center plate.
156
Detailed Repair Procedures
Switching
Magnet
Valve Arc
Screws
Seals
3. To replace seals, remove the two M3 screws from the valve arc.
157
Detailed Repair Procedures
158
Detailed Repair Procedures
Bottom Swivel
A) Check that the bottom swivel is
securely held by the two screws to
the central plate of the absorber.
B) Check that the screw threads, where
the top angled casting is screwed B
onto the upright stainless steel tube,
do not show. The stainless steel tube
should be free to turn in the bottom
swivel.
C
C) Check for any build-up of grease or
powder on the bottom swivel to the
upright stainless steel tube.
160
Detailed Repair Procedures
161
Top Swivel
162
Detailed Repair Procedure
Using the ventilator installed on the BleaseSirius, complete the pre-use system
test, then complete the system test. The system test specifically checks the bag
arm and associated components for leaks and blockages.
163
Detailed Repair Procedure
164
Detailed Repair Procedures
The Drawer will slide off the end of the runners - take extra
care.
165
Detailed Repair Procedures
166
Detailed Repair Procedures
167
Detailed Repair Procedures
168
Detailed Repair Procedures
169
Detailed Repair Procedures
170
Detailed Repair Procedures
Electrical
171
Detailed Repair Procedures
5.19 Ventilator
After the ventilator has been serviced or repaired, perform the ventilator
Checkout Procedure described in Chapter 4.
1. Disconnect the gas supplies and ensure the electrical supply is turned off.
2. Remove the Top molding, front cover and the main rear cover (see
section 5.1).
3. Remove the six screws around the edge of the ventilator case top.
The front of the ventilator will now hinge forward a little to allow the top to open
up completely.
172
Detailed Repair Procedures
E A
G
C
F
173
Detailed Repair Procedures
174
Detailed Repair Procedures
175
Detailed Repair Procedures
176
Detailed Repair Procedures
1. Disconnect the connector on the left, as shown, and on the lower right.
Disconnect the two wires to the pressure switch.
2. Disconnect the ribbon wires at the other end and not from this board.
3. Remove the two nylon nuts in the middle at either side.
4. Disconnect the four corner quick clips.
177
Detailed Repair Procedures
Figure 138
Figure 139
178
Detailed Repair Procedures
179
Detailed Repair Procedures
3. Remove the three screws from the bottom of the front molding (Figure 120)
and lift out the front panel assembly.
4. To remove the
board, disconnect
connectors as
shown.
180
Detailed Repair Procedures
181
Detailed Repair Procedures
182
Detailed Repair Procedures
2. Remove the two nuts as shown, which in turn releases the valves.
183
Detailed Repair Procedures
5.19.10 Removal of Gas Inlet Module (FR 140026) (up to March 2009)
2. Disconnect the outlet hose and the two connectors to the pressure switch.
184
Detailed Repair Procedures
185
Detailed Repair Procedures
186
Detailed Repair Procedures
If you replace any of the following components: BAV controller, display board,
inspiratory blocks or flow valves, you must complete the following on-screen
installation routine.
187
Detailed Repair Procedures
188
Detailed Repair Procedures
If all is correct, the ventilator will bleep and state DONE on the screen. After a
few seconds, you will be returned to the normal run screen.
If an error occurs, the calibration will be aborted. The only reason for this error
is that there was not a step change between one calibration point and the next.
189
Detailed Repair Procedures
In very low flow applications, it is often necessary to return the sample gas to
the breathing system to prevent the bellows from slowly collapsing.
It is important that the ventilator knows where the return point is, so that it
can make the correct compliance correction. The instructions to follow vary,
depending on which option is selected here.
If the return is to atmosphere/AGSS, the user will be told to remove the sample
line before pre use test.
If the return is to the breathing circuit, the user will be told to ensure the analyzer
is switched on and that the sample line is connected.
On some models or for some installations, this option can be turned off by
selecting disable and then setting a fixed compliance correction. The value
of this compliance correction should be selected to give a good approximation
of a ‘normal set-up,’ i.e., measure the system with a normal absorber full of
soda lime and the most frequently used breathing system in place. Look at
the measured system compliance displayed on the main screen. Then set this
value for the disabled value.
Next, four boxes are displayed as a confirmation screen. Touch each of the
four boxes as near to the centre as possible. When an acceptable reading is
obtained, the box will show ‘DONE.’ When all four boxes are ‘DONE,’ the screen
will be calibrated and will display ‘passed’.
190
Ventilator Passwords
CLEAR
Select: Defaults>Save Hospital Defaults>CLEAR
191
Ventilator Passwords
Fresh Gas Select this option to calibrate the fresh gas flow
reading on the ventilator.
192
Ventilator Passwords
Enable Pre-Use Test Select this option to set a fixed compliance value
to be set from 0.2 to 10.0 mL/cmH20 in 0.2
increments.
193
Ventilator Passwords
FAN FAIL Select this option to turn the fan fail alarm
on and off. For stand alone versions of the
ventilator, this option will remain unchecked.
For integrated ventilators this will be checked.
194
Ventilator Passwords
Flow Valve Calibration Select this option to calculate the min and
max flow through the flow valves. Flow Valve
Calibration needs to be performed only when the
flow valves are replaced.
Block and Pressure Select this option to give the zero point to the
Zeros processor for flow and pressure. Always perform
the Block and Pressure Zeros when the inspiratory
block, BAV controller or Front panel is replaced.
Front Panel Zero Select this option to zero the flow and pressure
on the Front Panel. Always perform Front Panel
Zero when the inspiratory block or front panel is
replaced.
You choose a different language on this menu, if a language pack has been
installed.
195
Ventilator Passwords
196
Ventilator Passwords
Select this option to manually adjust the monitored Vti and Vte. The Zero value
should not be changed.
197
Detailed Repair Procedures
4. Run the ventilator in volume control with the sensor at the patient using
the following settings:
BPM 12
Volume 500 ml
I:E 1:2.0
Pressure Limit 50
PEEP OFF
5. Once the volume has stabilized, verify that the measured volume on the
flow analyzer attached to the inspiratory port is 450 - 550 ml. The Vte and
Vti readings on the ventilator (displayed in the red box) should represent
the measured delivered volume +/- 5ml. If an adjustment is necessary,
the gain factor can be adjusted for both Vte and Vti to align the ventilators
displayed measured volume with the actual measured delivered volume.
198
Detailed Repair Procedures
Test Procedure
1. Run the ventilator using the settings noted in Step 4 of the previous
section. Ensure that the displayed and measured values are accurate with
one another and both are 500 ml +/- 50 ml.
2. Set the tidal volume to 250 ml and ensure both readings are
250 ml +/-25 ml.
3. Set the tidal volume to 750 ml and ensure both readings are
750 ml +/-75 ml.
199
Detailed Repair Procedures
5.29 Reprogramming
The software for the Blease700/900 series of ventilator consists of two parts,
the control software and the monitor software.
The version of software installed on the ventilator can be seen on the screen
by selecting Setup, Configuration, System Information. The display software
will be displayed next to the FRONT PANEL software and will be in the format
V700900 x.ox followed by the date. The control software will be displayed next
to control board software and will be in the format BAV040-700900-V9_5x_
followed by the date.
200
Detailed Repair Procedures
3. Open VentFlash.
201
Detailed Repair Procedures
The language file version number must match the front panel
software version number. Otherwise, only group 1 languages
will be available.
202
Detailed Repair Procedures
When a ventilator model upgrade package is purchased, these files will be sent.
All three files must be placed into one folder on your PC.
This file will only work on the ventilator with matching serial
number.
203
Detailed Repair Procedures
204
BleaseSirius
Anesthetic Machine
Chapter 6
EFM Option
205
EFM Option
206
EFM Option
The needle valves and hypoxic guard components of EFM are identical to the
traditional mechanical flowmeter components. These components are installed
in a new block allowing the EFM flow sensors for each gas to be fitted.
An SD Card reader slot is provided on the EFM for software upgrades and
service.
A
D
Figure 149 Key
A Total Combined Flow Tube
B B Flow Sensor Display
C Flow Controls
D SD Card Reader
C (on the side)
207
EFM Option
208
EFM Option
209
EFM Option
The hypoxic guard block itself is modified to fit the EFM flow sensors that replace
the mechanical flow tubes.
210
EFM Option
211
EFM Option
212
EFM Option
213
EFM Option
1. Remove the top shelf cover, the front cover and the backbar cover from
the BleaseSirius system. See Section 5.1.2 in this manual for instructions.
2. Remove the four hex screws (two on each side) that secure the EFM to
the machine.
3. Disconnect the ribbon cable and the power connector from the EFM.
4. Remove the light pipe from the EFM circuit board.
5. Disconnect the four gas flow tubes from the back of the EFM.
6. Remove the EFM from the machine. Place it on a clean surface with the
display side down.
7. Using a 3mm hex key, disconnect the banjo fitting from the side of the
hypoxic block.
8. Remove the four retaining screws
(2.5mm hex key) from the back of
the EFM and gently hinge the display
screen assembly (Figure 158).
214
EFM Option
215
EFM Option
All persons handling ESD must be properly grounded via a 1mW resistive
grounded wrist strap.
Cover all ESD bench tops with grounded conductive mats and connect all
work surfaces and equipment to earth ground.
1. Remove the top shelf cover, the front cover and the backbar cover from
the BleaseSirius system. See Section 5.1.2 in this manual for instructions.
2. Remove the four hex screws (two on each side) that secure the EFM to
the machine.
3. Disconnect the ribbon cable and the power connector from the EFM.
4. Remove the light pipe from the EFM circuit board.
5. Disconnect the four gas flow tubes from the back of the EFM.
6. Remove the EFM from the machine. Place it on a clean surface with the
display side down.
7. Using a 3mm hex key, disconnect the banjo fitting from the side of the
hypoxic block.
8. Remove the four retaining screws (use a 2.5mm hex) from the back of the
EFM and gently hinge the display screen assembly.
9. Carefully remove the three circuit board connectors.
216
EFM Option
217
EFM Option
218
EFM Option
219
EFM Option
220
EFM Option
Note: The EFM Board only functions with EFM Software Versions that are
compatible with the EFM Board Version. The Service Application will dis-
play an error message if the EFM software update is not compatible with
the EFM Board Version. You cannot install incompatible software.
4. Select Software Update to install the user software on the EFM Flash
drive. This installation takes less than 1 minute.
Note: If the touch screen does not function, reboot the system.
5. After the software is installed, you must configure the Flow Tubes
as described in the next section. Do not exit this application before
configuring the flow tubes!
221
EFM Option
222
EFM Option
Note: If the touch screen does not function, reboot the system.
5. Remove the SD card when you have completed the required diagnostics
and store it in a secure place. Place a new label over the EFM SD Card
Reader.
223
EFM Option
224
EFM Option
1. Select Copy Log. A new log file is created with the filename
log_000000000X.txt. Note that each time you select Copy Log, the last
digit of this filename increments. Therefore, the most recent log file has
the highest number in its filename. The file is copied to the /efm/log folder
on the SD card. This process takes less than one minute.
2. When the log is copied, remove the SD card from the EFM reader.
3. Connect a SD card reader to your laptop, then insert the SD card into the
reader.
4. Retrieve the log from the /efm/log folder on the SD card. Typically, there is
only one log file, but there can be more than one depending on the issue.
If a StackTrace_0000000000X.txt is also in the /efm/log folder, retrieve
that file as well.
5. Email the most recent log file (or all log files) and the StackTrace file (if
available) to Spacelabs Technical Support.
Technical Support may also instruct you to delete the log and collect fresh
data. To delete the log, select Delete Log, then select Yes to confirm the
action.
You may be instructed to recover the flash drive. When you select
Recovery, the flash drive is erased. Select Yes to confirm the action.
Under normal conditions, you should not use this feature.
6. Select the Exit tab. When you see the message, “OK to turn off the
system,” power off the system.
225
EFM Option
226
EFM Option
227
EFM Option
229
EFM Option
Schematic 3 - Power
230
EFM Option
231
EFM Option
Schematic 5 - Sensors
232
EFM Option
233
EFM Option
Schematic 8 - SD Card
235
EFM Option
Chapter 7
Error Codes
237
Error Codes
238
Error Codes
This list includes error codes which were assigned to the 64180 controller
board, but which are not applicable to the MPC68302 controller board. These
are described as N/A. It also includes codes which only apply to 6x00 serial
vents with control potentiometers - these are indicated 6x00 only.
The following table includes only the error codes that the Blease700/900 series
of ventilators will generate, with their common cause and corrective action.
239
Error Error Name Description common cause Corrective action
240
Code
20 E_ADC_FLOWZERO_LOW Drive Flow zero is too Drive flow pressure transducer (PT1) has lower Use CALBK to reset Zero. If error still remains replace
low output then expected inspiratory block
21 E_ADC_FLOWZERO_HIGH Drive Flow zero is too Drive flow pressure transducer (PT1) has Higher Use CALBK to reset Gain. If error still remains replace
high output then expected inspiratory block
23 E_ADC_TEST0 Internal i/o test line 0 Multiplexer U23 is not functioning Replace BAV controller
failed
24 E_ADC_TEST1 Internal i/o test line 1 Multiplexer U22 is not functioning Replace BAV controller
failed
25 E_ADC_TEST2 Internal i/o test line 2 Multiplexer U21 is not functioning Replace BAV controller
failed
26 E_ADC_TEST3 Internal i/o test line 3 Multiplexer U20 is not functioning Replace BAV controller
failed
27 E_ADC_PPRESZERO_LOW Patient pressure zero is Patient pressure transducer (PT2) has lower Use CALBK to reset Zero. If error still remains replace
too low output then expected inspiratory block
28 E_ADC_PPRESZERO_HIGH Patient pressure zero is Patient pressure transducer (PT2) has higher Use CALBK to reset Gain. If error still remains replace
too high output then expected inspiratory block
29 E_ADC_DPRESZERO_LOW Drive pressure zero is Drive pressure transducer (PT3) has lower output Use CALBK to reset Zero. If error still remains replace
too low then expected inspiratory block
Error Codes
30 E_ADC_DPRESZERO_HIGH Drive pressure zero is Drive pressure transducer (PT3) has higher Use CALBK to reset Gain. If error still remains replace
too high output then expected inspiratory block
31 E_ADC_5VANEG The -5V supply is out of The -5v supply generated on the BAV controller Replace BAV controller
specification is low
32 E_ADC_5VA The 5V Analog supply is The +5vA supply generated on the BAV controller Replace BAV controller
out of specification is low
34 E_ADC_5VALARM The 5V Alarm supply is The +5v Alarm supply generated on the Power If the supply atTP7 on the power supply is low -
out of specification supply board is low Replace the PSU. If the supply is good then the
monitoring circuit on the controller could be faulty -
Replace BAV controller
35 E_ADC_5VD The 5V Digital supply is The VCC supply generated on the BAV controller Replace BAV controller
out of specification is low
37 E_TEST_LINK The 8 way test link A link has been left on JP3 Remove the link
header is incorrect
104 E_PARAM_CHKSUM Stored Config & Calibration/ serial numbers have not been stored Use SpacelabsTerm to store serial number. Use
Calibration Data invalid CALBK to store calibration.
BleaseSirius
Anesthetic Machine
Chapter 8
Spare Parts
241
Spare Parts
242
Spare Parts
Internal Tubing
Oxygen 8mm 53500027 (white) 53500061 (green)
Nitrous Oxide 53500041 (blue) 53500041 (blue)
Please order tubing by meter lengths.
4mm
Air 5mm 53500042 (black) 53500043 (Yellow)
Moldings
Grey Moldings White Moldings up White moldings from
to Feb 2009 Feb 2009
Bottom Wall CGO 13600031 14200027 14200120
Bottom Wall ACGO 14200026 14200121
Top Cover 13400059 13401114 14200123
Rear Cover 13600030 13600255 14200122
Top Shelf 13600130 13600254 13600254
Drawer 13600035 13600536 13600536
13600537 13600537
Electrical
Power Sockets
UK 80200039
EUROPE 80200040
USA / CANADA 80200082
243
Spare Parts
4A Fuse US 81308015
Fuse Board 110V 13600284
Absorber detect switch ER2275458
O2 sensor board (base molding) 13600217
O2 Sensor cable Curly 940453
Ext peep connector (base molding) 13600316
On / Off Switch knob 14200003
Switch segment PXB_291 SI0112
Switch segment ZB2-BE102 70600006
Switch segment ZB2-BE1016 54200059
Reed Switch Assy 13600100
Pneumatics
CGO Assy FR136027
Alarm module 3 gas 13400005
Alarm whistle 11901909
On/Off switch assy 14200003
Alarm valve position 1, 3 and 4 53000010
Alarm valve position 2 53000011
Flush tap valve (clipard) 53000025
DISS Outlet 13602110
O2 Flush button label (green) 13600260
244
Spare Parts
Yokes
Yoke bridge 13500127
Yoke block O2 13500317
Yoke block N2O 13500318
Yoke block Air 13500319
Transfer manifold 13600063
Transfer manifold nut SM1681
Transfer manifold washer SM1690
245
Spare Parts
Absorber CAS II
Absorber Top No bypass white (NEW) FR122030
Absorber Top With bypass white (NEW) FR122031
Absorber Top No bypass white (OLD) FR122028
Absorber Top With bypass white (OLD) FR122029
Absorber 2kg No bypass White 12200902
2Kg outer canister no drain 12200228
1Kg outer canister no drain 12200229
Inner canister 12200221
Central Seal (middle) 12200556
Upper Seal (top) 12200557
Cas Top seal 12200235
Cas plate seal top 12200535
CAS outer canister seal SI0821
Rear seal Black 12200432
Rear seal White 12200533
Absorber plate seal lower 12200433
Absorber manifold gasket (base molding) 13600069
246
Spare Parts
Backbar
Dzus Spring (complete) 930201
Dzus Spring (spring only) SI212
Backbar Assembly (complete) 14200103
Backbar valve 12001003
247
Spare Parts
Ventilator
BAV Power Supply FR136029
BAV Display PCB FR101026
BAV Controller Board Type 3 FR101027
BAV Switchmode Power Supply FR136025
BAV Inspiratory Block FR137025
248
Spare Parts
Bellows
Bellows Assembly Complete (white) 12500011
Adult Bellows Canister Cover 83035
Pop Off Valve Assembly (new) 12500014
Bellows Base (white) with pop off valve 12500012
Adult Bellows 83034
249
Spare Parts
Drawers
Drawer slider 54600085
drawer lock identical key 55000026
drawer lock cam 55000027
Frame
Front Castor 13600337
Rear Castor 13600336
White V bracket 13600261
Accessories / Random
patient flow sensor 10110090
Flow Sensor Caps (pkg of 50) 54100063
patient flow sensor tubing 1.5m 10110219
patient flow sensor tubing 2m 10110218
patient flow sensor tubing 3m 10110102
patient flow sensor tubing 6m 10110103
Fresh gas flow sensor tubing 10110104
Fresh gas flow sensor 10110105
Suction controller 11400060
Suction Filter 11400063
Suction Filter housing 11400064
Suction filter housing oring 11400065
Fomblin grease ST7014
Silicoset 151 ST3201
250
Spare Parts
PM Kit
One Year PM Kit (BleaseSirius) 14000511
Four Year PM Kit (BleaseSirus) 14000512
One Year PM Kit (Sirius 3000,2000) 13600530
Four Year PM Kit (Sirius 3000,2000) 13600531
251
Spare Parts
NOTES
252
BleaseSirius
Anesthetic Machine
Chapter 9
Notices and Important
Information
253
Notices and Important Information
254
Notices and Important Information
If the ventilator is suspected of being worn, defective or otherwise unfit for use,
it should under no circumstances be used.
255
Notices and Important Information
9.4 Disclaimer
Opening of the control unit by unauthorized personnel automatically voids all
warranties and specifications. The prevention of tampering is solely the user's
responsibility; the manufacturer assumes no liability for any malfunction or
failure of the ventilator if the control unit is opened.
The contents of this manual are not binding. If any significant difference is found
between the product and this manual, please contact Spacelabs Healthcare for
further information.
256
Notices and Important Information
9.7 CE Mark
The product is labeled with the CE mark and number of Notified Body.
257
Notices and Important Information
258
Notices and Important Information
This handbook contains important hazard information. You must read this
hazard information before using the Blease700/900 Series Ventilators.
Warning Notices
Caution Notices
Relevant or helpful
259
Notices and Important Information
The following statements are made to comply with the requirements of IEC
60601-1 and IEC 60601-2-13
260
Notices and Important Information
261
Notices and Important Information
19. The following equipment is mandatory, please refer to the relevant user
manuals for instructions in usage; blood oxygen monitor, CO2 alarm,
blood pressure alarm, ECG alarm, anesthesia agent concentration.
20. The machine is to be equipped with an anesthetic gas scavenging
transfer and receiving system, complying with ISO 8835-3 or
applicable local standard before being put into service.
262
Notices and Important Information
9.9.3 Cautions
Anesthetic Machines
• Do not leave gas cylinder valves open if the pipeline supply is in use and
the system master switch is turned ON. Pressures from both supplies
may become equal and, if simultaneously used, cylinder supplies could
be depleted, leaving an insufficient reserve supply in case of pipeline
failure.
• The hypoxic guard control system only ensures that oxygen-nitrous
mixtures will have a minimum oxygen concentration. HYPOXIC
MIXTURES MAY BE DELIVERED IF GASES OTHER THAN OXYGEN,
NITROUS OXIDE OR MED AIR ARE USED, OR WHEN OPERATING
AT LOW OXYGEN FLOW RATES. When using carbon dioxide, as
an additional gas, make sure the proportions of all gases are carefully
adjusted in accordance with accepted clinical practice. Gas mixtures
within the breathing system must be monitored when using these gases.
• Leaking gases and vapors (downstream of the flow control valves and
Oxygen Flush valve) may deprive the patient of metabolic gases and
anesthetic agent may pollute the atmosphere. Tests that detect leaks
must be performed frequently. If detected, leakage must be reduced to
an acceptable level.
• Do not use the anesthesia system if the hypoxic guard control system
does not operate within permitted ranges. Using an incorrectly operating
control system may result in incorrect gas mixtures and injury to the
patient.
• When occluding the breathing system for test purposes, do not use
any object small enough to slip completely into the system. Objects in
the breathing system can interrupt or disrupt the delivery of breathing
system gases, possibly resulting in injury to the patient. Before using
the breathing system on a patient, always check the breathing system
components for foreign objects.
• Do not place materials weighing more than 25kg/55.1lbs on the
bottom shelf, or more than 25kg/55.1lbs on the upper monitor shelf.
Overloading may cause damage to the shelves or cause instability.
(Refer to Specification on page 30 for more details).
• Secure any equipment placed on the shelves.
263
Notices and Important Information
• To avoid stripping threads, do not use tools on the yoke gate T screws.
Use only one cylinder gasket per yoke. Using more than one gasket
could cause cylinder gas leakage.
WARNING:
If an alarm condition cannot be resolved, do not continue to
use the system.
264
Notices and Important Information
• Do not obstruct the drive gas exhaust. An obstruction may lead to the
ventilator not functioning correctly.
Warnings
1. USE OF EQUIPMENT
Incorrect use of the equipment described herein may result in injury to the
patient. Read this manual before operating the machine. You must be familiar
with the machine and its functions before using it on a patient.
3. PRESSURE LIMITING
Pressure limiting is a safety feature and is also used in adult and pediatric mode
to provide ventilation (Pressure Control). It should not be used for any other
purpose, such as creation of micro-breaths or emulation of specialized high
frequency/low volume ventilation systems, otherwise injury to the patient may
result.
5. AUDIBLE ALARM
An audible alarm indicates an anomalous condition that may result in damage
to the equipment or injury to the patient. The cause of each alarm should be
investigated and any necessary measures taken to remove the alarm condition.
265
Notices and Important Information
8. EXPLOSION HAZARD
An explosion hazard exists if this equipment is used with flammable anesthetic
agents such as ether or cyclopropane. To avoid the risk of explosion, use this
equipment only with anesthetic agents that comply with the requirements on
non-flammable anesthetic agents in the IEC standard ‘Particular Requirements
for the Safety of Anesthetic Machines’.
9. DRIVING GAS
Use of a driving gas other than oxygen or MED AIR may damage the ventilator
and will cause inaccurate operation, resulting in potential injury to the patient.
Do not use any other driving gas.
266
Notices and Important Information
19. SIMV
The SIMV modes use a flow Trigger to sense the patient attempt to breath.
Therefore, SIMV modes will be unable to work with patient sensor placed at the
absorber.
267
Notices and Important Information
3. BATTERY LIFE
To preserve battery life, never store the ventilator with its battery discharged.
Do not store or use the ventilator in close proximity to heat sources of any kind.
4. POWER FAILURE
In the event of a mains electrical power failure when the ventilator is running
from its internal battery, do not remove the mains plug from the mains supply
as this would prevent the immediate resumption of normal operation when the
mains power is restored.
5. ELECTROMAGNETIC INTERFERENCE
Excessive electronic noise caused by poorly regulated devices may interfere
with the proper functioning of the ventilator. To avoid this, do not connect the
ventilator to the same mains supply outlet into which an electrocautery unit is
connected.
6. EXHAUST GAS
The driving gas is discharged through the port on the rear panel of the
ventilator. This port must be completely free of any obstruction and should have
nothing connected to it. The exhaust gas is oxygen or MED AIR and does not
contaminate the environment.
7 BELLOWS EXHAUST
A negative or positive pressure applied to the EXHAUST port of the bellows
assembly results in positive pressure in the patient breathing system. The
scavenging system must therefore not generate more than 5cmH2O positive
or negative pressure when connected to the ventilator. The use of an AGSS to
EN 740:1999 is recommended. Do not connect a PEEP valve to the EXHAUST
268
Notices and Important Information
port of the bellows base. This will increase the pressure inside the bellows
and cause it to detach from the base, resulting in a serious malfunction. Any
problem arising from an improperly functioning scavenging system is solely the
responsibility of the user.
9. CLEANING AGENTS
The ventilator surfaces are not scratch-resistant. Do not use abrasive cleaning
agents otherwise damage to the surfaces will result.
Chemical decontaminants or liquid sterilization agents will damage the sensor
and must NOT be used for cleaning or sterilizing. If autoclaving the patient
sensor, the autoclave must only be used with distilled water.
269
Notices and Important Information
270
Notices and Important Information
1. Use of Equipment
Incorrect use of the equipment described herein may result in injury to the
patient. Read this manual before operating the machine. You must be familiar
with the equipment and its functions before using it on a patient.
2. Absorber Malfunction
Injury to the patient may result if a faulty absorber is used. If there is any
malfunction, or if the absorber does not pass all pre-use tests, do not use
the absorber. If the malfunction cannot be rectified, call an authorized service
engineer or return the absorber to the supplier.
3. Spontaneous Breathing
If the APL valve is not in the OPEN position, it will not operate until the pressure
reaches as much as approximately 65cmH2O. This may result in injury to the
patient. Set the APL valve to the OPEN position to allow spontaneous breathing.
4. Soda Lime
Soda lime is caustic. Observe the manufacturer’s instructions for correct
handling and storage. When handling, always wear suitable eye, face and
hand protection as crushed soda lime can degrade to a fine dust which may
be harmful by inhalation. To minimize the levels of soda lime dust, ensure that
the soda lime is not crushed and that regular cleaning of the absorbers and
breathing circuits is carried out. To prevent injury to the patient, always protect
the patients face using a facemask.
5. Filling Canisters
Under-filling of canisters can lead to inefficient CO2 absorption. Overfilling
can result in poor canister sealing. Fill the reusable canisters only to the
recommended level.
6. Condensate
The condensate in the bottom of the circuit is caustic. Drain the condensate
periodically. Accidental splashes should be rinsed immediately with water.
7. Absorber Orientation
Use the absorber only in the upright position. Use in any other orientation may
affect the gas flow or damage the absorber.
271
Notices and Important Information
1. Cleaning
The surfaces of the absorber may be damaged by caustic substances. Do not
use caustic substances such as trichlorethylene for cleaning.
2. Sterilization
Sterilization will damage the manometer. Remove this item before sterilizing the
Absorber. The PEEP valve should be separated.
This seal is located above the upper soda lime canister, attached to the black
molding within the body of the absorber.
Please ensure all seals are correctly replaced when re-assembling the absorber.
272
BleaseSirius
Anesthetic Machine
Chapter 10
Schematics
273
Schematics
274
1 2 3 4 5 6 7 8
D D
FRESH GAS TP13
C17 +4VREF +5 New Block FRESH GAS.SCH FGO
R9 R92 2.2K
100nF
12K +5V Fresh Gas Out
+4VREF R146 R148
SGND TP10 4.7K 4.7K
R10 -5V R37 2.2K FGZ
Fresh Gas In FGZ TP17
12K FGAZ
TP7 +5 -5
+4V R98 2.2K R149 R147
4.7K 4.7K
R11
8
2 TP18 JP2
PFAZ
12K 1 TRANSDUCERS SK2
R21 R86 2.2K 1 2
3 New Block TRANSDUCERS.SCH 1
4.7R 3 4
U4A 6
4
FGAZ Fresh Gas In R95 2.2K PP 5 6
LTC2051 -4VREF 2 11
Q3 +5V Patient Pressure 7 8
7
R3 MMBFJ175 +4VREF R96 2.2K TP15 TP16 9 10
C15 -5 DF 3 10
100nF 12K SGND Drive Flow 11 12
8
-5V R93 2.2K C83 13 14
C81 4
Q4 -4VREF Drive Pressure 470pF 15 16
R14 470pF 9
BC847B PFAZ Patient Flow In 17 18
4.7K 5
19 20
21 22
TP11 23 24
TP8 DE9 SOCKET
DP 25 26
-4V
PATIENT FLOW HEADER 13X2
C -5 New Block PATIENT FLOW.SCH TP20 C
R91 2.2K PFO
+5V Patient Flow Out
+4VREF R140 2.2K
SGND
-5V R139 2.2K TP21
L4 Patient Flow In PFZ PFZ
+
C22 100uH
47uF
PL1
5
9
4 11
+5 8
R144 3 10
2.2K 7
2
6
J4 R113 R136 R114 R151 TP9 TP19 1
CALIBRATION PLUG 47K 47K 47K 47K FGI PFI
R150 DE9 PLUG
47K +6VB D3
Schematics
2
4
6
2
4
6
3 220uF
B 2 3 B
ACM4532-801 C96 +4VREF
24
11
12
14
2
23
1
13
24
11
12
14
2
23
1
13
1 R2 2
C10 L1 100nF
1
3
5
1
3
5
1
10K 100nF
SI
SI
CS
A1
A0
SO
CS
A1
A0
SO
C2
WP
WP
3-PIN MOLEX +
220uF TP12 R90
SCLK
SCLK
HOLD
HOLD
18 7 18 7 C11 ACM4532-801 TEMP 10K
GND Vin+ GND Vin+ C6
JP1 10nF
100nF TP5
U15E U17E SK1 HEADER 3X2 X-Valve Daughter Board
X9251 X9251 2x3 BOTTOM ENTRY +6VB
R89
TP6 TP4 2.2K TH1
-t
+6VA TP14 TP2 TP1 NTC THERMISTOR
C13 +5 +5V Regulated +5V Output +6VA
10nF 0V 0V 0V
U3 U2 U1
J2
C9 1 8 1 8 1 8 3-PIN MOLEX
150uF (BOOST) VIN OUTPUT INPUT OUTPUT INPUT
1
2 7 2 7 2 7
CAP+ OSC SENSE FB SENSE FB 2
+
J3 JP3
R138 R99 3
3 6 3 6 3 6 FL9 10K FL11 100R
GND LV SD 5V TAP SD 5V TAP 2 1
1 2 R127
4 5 4 5 4 5
CAP- VOUT GND ERR GND ERR 3 R131 100R
2-PIN MOLEX
4 C82 2K +6VB -5
LM2660 + LP2951 + LP2951 C12 RJ45 100nF
10nFx5nF R137 10nFx5nF
A + D2 100nF 10K A
MBRA140T3
275
1 2 3 4 5 6 7 8
276
PFAZ
D +4VREF R130 D
R109 +5
FL5 47R +4VREF
10nFx5nF 560K FL10
R115 10nFx5nF
+5
U17A R111 13K
1
2
1
2
REF+ X9251 REF+ R129
200K C91 47R C92
nc 10nF nc 10nF
8
R122 PT3 U15A
110K PXL0075DN R123 X9251
PT4
PXL12X5DN 10 64.9K
10
3 5 R134 3 5
1.2K R103
R121 R112 8 9
OUT+ OUT- 1K R143 OUT+ OUT- 120K R128
1K
9
120K 120K
8
8
R110 2 R152 2 R116 R76 C69
1K 1K
nc 100K 1 nc 1 110K 22nF +5
REF- 3 REF- C84 3
6
4
6
4
C85 U19A 6 22pF 6 R62 R68 R75
4
4
18K 82K 68K
8
C88
FL7 R125 FL8 100nF 100nF LT1366
10nFx5nF -5 10nFx5nF -5
47R +4VREF
10nF R106
47R C86
G=(1+R2/R1)+2*R2/R5 10nF -5 R101
R153 IF R1=R3, R2=R4 1.5M
16
C 1K R132 C
R142 1K
47R 17 U15B
FGAZ
+4VREF X9251
+5
FL6 FL4 +4VREF
10nFx5nF R108 10nFx5nF +
15
120K R100
+5
R104 C68 82K
1
2
1
2
U17B R105 R66 C67
REF+ 120K C95 REF+ 47R C65 10nF
nc X9251 15K 10nF nc 10nF 4.7uF
15
PT5 R107
PXL0075DN PT2
110K PXL0075DN R79
17 64.9K
20
3 5 R135 5 3
680R
R124 R71 22 21
OUT+ OUT- 1K R141 OUT- OUT+ R70 1K
120K
16
R119 120K U15C
8
8
2 R133 2 R85 X9251 C78
62K 1K R67
nc 1 nc 1 120K 110K 22nF
REF- 3 REF- C70 3
6
4
6
4
C87 U18A 6 22pF U13A R73 6 R87 R88 R84
4
4
1K 18K 82K 68K
22pF OP213FS 7 OP213FS 7 6
R117 Fresh Gas In R78
5 C73 5 7
Schematics
Patient Pressure
1K C93 U18B -4VREF 1K 22pF U13B 5
C80 C77
-4VREF 22pF OP213FS OP213FS U14B
FL12 FL3 100nF 100nF LT1366
10nFx5nF -5 10nFx5nF -5
+4VREF
R118 R81
47R C89 47R C75
10nF 10nF R97
B R145 B
1.5M
4
1K R64 R82
+4VREF 1K
+5
FL2 47R 3 U15D
10nFx5nF +5 X9251
+
5
1
2
+5V R94
REF+ C64 C76
nc 10nF C79
PT1 10nF 4.7uF 82K
PXL02X5DN
U17C
-5V
X9251
20
3 5 R77
R63 680R
R69 22 21 -5
OUT+ OUT- 120K R65
1K 120K
8
2 R72 R61
1K C58
nc 1 62K
22nF
REF- 3
6
4
C71 U12A 6 R44 R54 R45
4
18K 82K 68K SGND
22pF OP213FS 7 6
R74 C72 5 7
Drive Flow
1K 22pF U12B 5
C53 C62
-4VREF OP213FS U7B +4VREF
FL1 R83 100nF 100nF LT1366
10nFx5nF -5 +4VREF
47R
C74
A 10nF -4VREF A
R40 R39
1.5M 130K -4VREF
+
R80 C47
1K C43
10nF 4.7uF Title
Pressure Interface Board -Transducers
D D
R30
+5 +5
2.4K +5V
+5
1
R31 C40 SGND
3.6K 100pF D5A
8
2 BAV199
3
Patient Flow In R53 R43
1 1.5K
1.5K
3
3
-5V
U9A
4
R28 TLE2072 D5B -5
27K BAV199
R42
2
R29 -5 Q7
47R 10K
R27 BC847B
Q8
15K BC847B
1
D7A R41
+5 C39 BAV199 10K
10nF
8
2 R38
1 1.2K
3 U8A
C C
4
LTC2051
14
R26 -5 -5
27K R32 13 R33
C38 1.2K
10nF 11K R52
Q5D 27K
12
MAT-04 PFZ
C42
100pF C55
9
2
Q5C +5 100pF
MAT-04 Q5B R51
27K
MAT-04 6
8
8 10 5 7 3 1 2 7
1 5
1
C54 Q5A 3 U11B
390pF MAT-04 R50 LT1366
4
27K C20 +5
TLE2072 C35 U6A
6
100nF
U10A 470nF R5 LT1366
18K
8
-5 R20 3 3
10K+10K
2 3 R48 1
1.5K R58 R8 Patient Flow Out
2
Schematics
7 4 Substrate 11 -5
5 Q6 R4
B U9B BSS83 39K B
R57 TLE2072 Q5E or SST213
27K C14
R59 MAT-04
R56 R55 100nF
47R 15K 15K -5 +5
8
R49
7 33K 3
3
6 1
C57 2 5 4
10nF +4VREF
U10B
4
TLE2072 U17D
6 X9251
7 R60 R120
5 15K U7A 100K
U8B LT1366
-5
LTC2051
C37
R46 C56 100nF
27K 10nF C61
100pF
A A
Title
Pressure Interface Board -Patient Flow
277
1 2 3 4 5 6 7 8
278
D D
+5
+5V
+5
SGND
C52 C31 C49 C50 C25 C59 C33 C51
10nF 10nF 10nF 10nF 10nF 10nF 10nF 10nF
-5V
-5
-5 +4VREF
C41 C16 C36 C18 C44 C19 C48 C46 C28 C60
10nF 10nF 10nF 10nF 10nF 10nF 10nF 10nF 10nF 10nF
C FGZ C
14
Q2D
MAT-04
13
R16
12
5.6K
9
2
Q2C C23 C24
MAT-04 Q2B C26 100nF 22nF
MAT-04 100pF
8 10 5 7 3 1 6
7 R36 R13 R6 5
C27 Q2A 5 7
18K 82K 68K Fresh Gas Out
MAT-04 TLE2072 6
R7 U6B
6
U5B
390pF 100K C34 LT1366
2 3 100nF
D4B R23 R25
+5 BAV199 R18
1.5K 43K
160k
Schematics
Iref =93uA
8
R17 2
Fresh Gas In C32
1 4 Substrate 11 22nF
3.6K
3
U5A
4
R24 TLE2072 Q2E
B 27K B
+5 MAT-04
-5 R19 -5
R12 R22 15K
8
47R 15K 3
1
2 R35
+4VREF
4
U11A 43K
C30 LT1366
10nF
6 -5
7 R34
5 15K
U4B
LTC2051
R15 C29 C45
27K 10nF 100pF
A A
Title
D7
D6
D5
D4
D3
D2
D1
D0
A0 12 13 D0 A0 12 13 D0 A0 11 13 D0 RESET#
A0 D0 A0 D0 A0 DQ0 RESET# GND VCC PCDATA
A1 11 14 D1 A1 11 14 D1 A1 10 14 D1 PCDATA
A1 D1 A1 D1 A1 DQ1
A2 10 15 D2 A2 10 15 D2 A2 9 15 D2 VCC
50 A13
49 A12
48 A11
47 A10
46 A9
45 A8
44
43 A7
42 A6
41 A5
40 A4
39 A3
38 A2
37 A1
36 A0
35
34
33
32
31
30
29
28
27
26
A2 D2 A2 D2 A2 DQ2 R3
A3 9 17 D3 A3 9 17 D3 A3 8 18 D3 U5 PRS_OVRLD
A3 D3 A3 D3 A3 DQ3 PRS_OVRLD
A4 8 18 D4 A4 8 18 D4 A4 7 19 D4 WAIT# APNOEA
A4 D4 A4 D4 A4 DQ4 WAIT# APNOEA
A9
A8
A7
A6
A5
A4
A3
A2
A1
A0
D9
D8
Vss
A5 7 19 D5 A5 7 19 D5 A5 6 20 D5 A14 51 25
Vcc
A13
A12
A11
A10
D15
D14
D13
D12
D11
D10
P4.7
D A5 D5 A5 D5 A5 DQ5 A14 P4.6 D
A6 6 20 D6 A6 6 20 D6 A6 5 21 D6 A15 52 24 R63
A6 D6 A6 D6 A6 DQ6 A15 P4.5
A7 5 21 D7 A7 5 21 D7 A7 4 22 D7 A16 53 23 R61
A7 D7 A7 D7 A7 DQ7 VCC BUSCLK A16 P4.4
A8 27 A8 27 30 A8 29 BUSCLK A17 54 22
A8 A8 A17 A8 A17 Vss GND
A9 26 A9 26 1 A9 28 32 VCC A18 55 21 R62
A9 A9 A18 A9 Vcc VCC R23 A18 P4.3 DIN
A10 23 32 A10 23 A10 24 R4 A19 56 20 R67
A10 Vcc VCC A10 A10 VCC A19 A19 P4.2 BUSY
A11 25 A11 25 32 A11 27 57 19 R65
A11 A11 Vcc VCC A11 GND Vss P4.1 DOUT
A12 4 A12 4 16 A12 3 16 58 18 R64 ABS DETECT
A12 A12 GND GND A12 Vss GND WAIT P4.0 ABS DETECT
A13 28 16 A13 28 59 17 SHDN211
A13 Vss GND A13 P6.1 P9.5 SHDN211
A14 29 A14 3 29 WR# 1 2 60 16 EN211
A14 A14 WE DNC RY/BY U11 P6.2 P9.4 EN211
A15 3 A15 31 24 RD# 17 1 61 15 RxD1
A15 A15 OE DNC J2 1 8 PHI RXD1 RxD1
A16 2 31 WR# A16 2 22 12 31 WR# 2 MR WDO 62 14 RxD0
A16 WE A16 CS1 VCC NC WE 2 7 VCC STBY RXD0 RxD0
A17 30 24 RD# 26 25 RD# VCC RST 63 13 TxD1
A17 OE NC OE 1 3 6 RESET TXD1 TxD1
A18 1 22 68128/628512 UC6 30 23 J13 GND WDI C3 64 12 TxD0
A18 CE NC CE 2 4 5 NMI TXD0 TxD0
CS1# PFI PFO XT2 65 11
CS1# GND Vss Vss GND
GND 66 10
VCC EXTAL VPP
U2 A18 SB24 67 9 DTR1
GND XTAL PB.7 DTR1
A0 12 13 D0 68 8 CTS1
A0 D0 A17 SB25 C4 VCC Vcc PB.6 CTS1
A1 11 14 D1 R44 AS# 69 7 DTR0
A1 D1 VCC GND AS# AS PB.5 DTR0
A2 10 15 D2 SB26 1K R11 RD# 70 6 CTS0
A2 D2 VCC R18 RD# RD PB.4 CTS0
A3 9 17 D3 WR# 71 5 M/R#
A3 D3 WR# HWR PB.3 M/R#
A4 8 18 D4 LWR# 72 4 MUX_AD2
A4 D4 R17 LWR# LWR PB.2 MUX_AD2
A5 7 19 D5 73 3 MUX_AD1
A5 D5 MD0 PB.1 MUX_AD1 VCC
A6 6 20 D6 C37 R12 74 2 MUX_AD0 R13
A6 D6 R21 MD1 PB.0 MUX_AD0
A7 5 21 D7 75 1
A7 D7 MD2 Vcc VCC 3.3k
A8 27
A8 JP2 R105 R20 R59 R19
A9 26 1 C35 C55 C56 VPP
AVcc
VREF
P7.0
P7.1
P7.2
P7.3
P7.4
P7.5
P7.6
P7.7
AVss
P8.0
P8.1
P8.2
P8.3
P8.4
Vss
PA.0
PA.1
PA.2
PA.3
PA.4
PA.5
PA.6
PA.7
A9 VPP
A10 23 32 VPP 2 USER
1
A10 Vcc VCC GND VPP VCC C14 C15
C A11 25 C
A11 LOCATE CLOSE
76
77
78
79
80
81
82
83
84
85
86
87
88
89
90
91
92
93
94
95
96
97
98
99
100
4
5
6
7
VCC EXPVOLB
16
EXPVOLB
U15 DRVPRSB R5
GND VCC U10A DRVPRSB
A16 1 15 GND FLWPRSB
Y0
Y1
Y2
Y3
A Y0 FLWPRSB
A17 2 14 OXYB
VCC
UC10 B Y1 VPP CONTROL OXYB R6
A18 3 13 FRESHGASB
C Y2 FRESHGASB
12 LIMSET
VCC Y3 LIMSET
A
B
E
11 R10
GND Y4 VCC C5
4 10
E1 Y5 U16A
2
3
1
CS0# 5 9 DA1 SB22
Schematics
GND
A19 E3 Y7 2 13 SB23
A20 ANAL_IN GND
A20 12 ANAL_IN
8
ENCODER SW
VCC GND
CS4# IRQN
B CS4# 3 IRQN B
6 11
DCLK R100
8 CS
CS
U10B
14 12 AGND GND
A Y0
13 11
B Y1 5
ENCODER SW
10
Y2 4 9
15 9
E Y3 7 10
GND
VCC
GND
VCC
PSU+
R156 R157 R158 R159 R160 R161
47K 47K 47K 47K 47K 47K
728&+6&5((1,17(5)$&( LS1
U46
AD7873ARV
J17 1 16
VCC DCLK DCLK SPEAKER
2 15
1 X+ CS CS
3 14
2 Y+ DIN DIN R69
4 13
3 X- BUSY BUSY
5 12
4 Y- DOUT DOUT R66
6 11
4 HEADER GND PENIRQ PENIRQ
7 10
VBAT VCC FET4
8 9
AUX VREF
A A
VCC R104
VCC GND
C68
Title
0.1U %/($6(/9'6',63/$<,17(5)$&(
GND GND UC11 UC13 UC15 GND
Size Number Revision
CAPACITORS ANNOTATED UCn ARE TO BE LOCATED CLOSE TO THE VCC PIN OF THE RESPECTIVE IC's
279
1 2 3 4 5 6 7 8
280
AVCC
UC30
J10 10 way IDC male J11 10 way IDC male
AGND
10
9
8
7
6
5
4
3
2
1
10
9
8
7
6
5
4
3
2
1
4
U25 C11 TP4 RFF8
12 13 3 R48 R45 FRESHGAS
C1+ V+ rfi FRESHGAS
17 FRESHGASB 1 U30A 100K 82K
C10 V- FRESHGASB
2 R144
11 C38
14 Vcc VCC GND GND NOT FITTED
C1-
15
C2+
11
C9 R46 R47
C12 10 AGND
GND GND 200K
D 16 C13 AGND D
C2- RFF19 RFF21 RFF20 RFF22
TxD1 7 2 TxD1
TxD1 T1in T1out rfi TP5 RFF6
DTR1 6 3 DTR1 R52 R49
DTR1 T2in T2out rfi 5 PATPRS
TxD0 20 1 TxD0 rfi PATPRS
TxD0 T3in T3out rfi PATPRSB 7
DTR0 21 28 DTR0 PATPRSB U30B
DTR0 T4in T4out rfi 6 R145
8 9 RFF25 C39
SB21 R1out R1in
CTS0 5 4 CTS0
CTS0 R2out R2in rfi
CTS1 26 27 CTS1 R50 R51
CTS1 R3out R3in rfi
RxD1 22 23 RxD1 RFF26 AGND
RxD1 SB18 R4out R4in rfi
RxD0 19 18 RxD0 RFF7
RxD0 R5out R5in rfi TP6 R75 R53
24 25 10 DRVPRS
EN SHDN rfi DRVPRS
R35 DRVPRSB 8 * *
DRVPRSB U30C
VCC RFF23 RFF24 9 R146
GND C40
R36 *
EN211 R54 R55
EN211
SHDN211 AGND
SHDN211 *
R37 TP7 RFF5
12 R130 R76 FLWPRS
rfi FLWPRS
R38 FLWPRSB 14
FLWPRSB U30D
GND 13 R147
C41
R77 R78
AGND
VCC
1 2 4 16
1A 1Y VCC 1,2EN Vcc AVCC
3 SB20 UC31
U28A 1Z 2 3
4
7 6 1A 1Y AGND
2A 2Y 1 RFF4
C 5 1B TP8 R134 R131 C
4 2Z U26A SB19 3 EXPVOL
1,2EN 6 5 rfi EXPVOL
8 2A 2Y EXPVOLB 1 U31A
16 GND 7 8 EXPVOLB
VCC Vcc 2B GND 2 R148
GND C42
GND
9 10 VCC
3A 3Y
11
11 12 R132 R133
U28B 3Z 3,4EN AGND
14 14 10 11
UC28 4A 4Y 3A 3Y AGND
13 9
12 4Z 3B
3,4EN U26B RFF3
14 13 D9 TP9 R138 R135
4A 4Y 5 TMPSNS
GND 15 BAR43S rfi TMPSNS
MC3487 4B 7
UC26 U31B
MC3486 VCC GND 6 R149
RFF17 RFF15 C43
RFF18 RFF16 GND
MUX_AD2
422RxD MUX_AD2 R43 R40 R136 R137
rfi MUX_AD1
422RxD# MUX_AD1 +12V
rfi MUX_AD0 22K 10K
MUX_AD0
422DTR
rfi U33
422DTR#
rfi 10 3 13
R143 COM NO0 U31D
ANAL_IN 8 R74
ANAL_IN U31C R151 11 1 DA1 12
RFF14 9 ADDC NO1 DA1
rfi
14 R39
Schematics
GND 10 15 10K 1 J8
C45 ADDB NO2 13 470R
RFF13 2 Buffered D/A Output
rfi
9 2 R139
ADDA NO3 R101 2 way 0.1"
AVCC AGND 1M AGND
RFF12 16 14
rfi
R42 R41 V+ NO4
D8 AGND R102
422CTS UC33 6 5 AGND
RFF11 BAR43S INH NO5 AGND
rfi
B 422CTS# +12V B
8 12
422TxD AVCC GND NO6 U35
422TxD# 7 4 8 1
V- NO7 I.C. I.C.
GND 7 2
R58 I.C. IN
R107 6 3 C36
PRS_OVRLD O/P VOUT NR
PRS_OVRLD O/P 10K 5 4
RFF9 AGND TRIM GND
PRS_OVRLD R153 C46
PRS_OVRLD rfi 0.1uF MAX6325
AVCC OFFSET ADJUST
VR1 AGND 0.1uF
AGND VCC
TP2
AGND R106 1M AGND
D2 APNOEA O/P
APNOEA O/P
RFF10 R95 AGND
APNOEA R152
APNOEA rfi 15M
RFF2 U32C
BAVREF R84 R85 10 TP1
AGND VCC BAVREF rfi R92 RFF27
8 R142
AGND 5 OSGND
D6 BAR43S R86 9 R94 rfi SENS-
C53 C32 OXYB 7 56K
OXYB U32B
AVCC UC32 J7 AVCC 6 TS2
VCC 1K0 C77
R71 R70
4
AGND 1 C51 220uF
AGND
2 C44 0.01uF
LIMSET 3 10uF 16V CALIBRATION TP3
LIMSET R154 3 AGND C52
U32A 1 LIMSET O/P R166 AGND VR2
LIMSET O/P 4 220uF
2 2.2K 100K
5 RFF28
R155 RFF1 U32D R140 R103 R141 NOT FITTED
R87 R88 OSOP
12 rfi SENS+
5 way 0.1" Header rfi
11
14 R93 1M 0.1% 825K 0.1% 61.9K 0.1%
CGO DET R167 TS1
AGND CGO DET R89 13 R150
47K AGND C54 C33 100M
A R171 A
BAGCGO R73 R72 0R
AGND GND
R168 AGND
BAG/VENT
BAG/VENT AGND Title
100K %/($6(/9'6',63/$<,17(5)$&(
Size Number Revision
D0 ... D7
SB5 WAIT#
WAIT#
RESET#
RESET#
SB1
SB2 WR#
WR#
SB3 LWR#
GPIO0 LWR#
RD#
D0
D1
D2
D3
D4
D5
D6
D7
RD#
D SB4 AS# D
AS#
M/R#
M/R#
CS5#
CS5#
VCC
GND A0 ... A20
A0
10n 10n 10n COREVDD
A1
GND A2
VCC GND
U24
32
31
30
29
28
27
26
25
24
23
22
21
20
19
18
17
16
15
14
13
12
11
10
9
8
7
6
5
4
3
2
1
R34 SB16 VCC
MD0 MD0 2 17 MA0 U23
I/O0 A0
R33 SB10 MD1 3 18 MA1
MD1 I/O1 A1
MD2 4 19 MA2
BS#
CS#
VSS
DB0
DB1
DB2
DB3
DB4
DB5
DB6
DB7
DB8
DB9
VSS
RD#
AB0
AB1
AB2
R32 I/O2 A2
M/R#
SB11
DB10
DB11
DB12
DB13
DB14
DB15
WE1#
WE0#
MD2 MD3 5 20 MA3 33 128 A3
GPIO0
WAIT#
I/O3 A3 COREVDD AB3
IOVDD
RESET#
MD4 7 23 MA4 MA3 34 127 A4 C61 C58 C59 C60
RD/WR#
R29 SB12 MD3 I/O4 A4 MA3 AB4 GND
MD5 8 24 MA5 MA4 35 126 A5
R30 SB13 I/O5 A5 MA4 AB5 Locate decoupling capacitors close to VCC pins of U19
MD4 MD6 9 25 MA6 MA2 36 125 A6
I/O6 A6 MA2 AB6
R27 SB14 MD7 10 26 MA7 MA5 37 124 A7
MD5 I/O7 A7 MA5 AB7
MD8 33 27 MA8 MA1 38 123 A8
R26 SB15 I/O8 A8 MA1 AB8
MD6 MD9 34 28 MA9 MA6 39 122 A9
I/O9 A9 MA6 AB9
R31 SB9 MD10 35 16 MA10 MA0 40 121 A10
MD7 I/O10 A10 MA0 AB10
MD11 36 15 MA11 MA7 41 120 A11
R25 SB8 I/O11 A11 MA7 AB11
MD8 MD12 38 1 MA10 42 119 A12
I/O12 VCC VCC MA10/GPIO1 AB12
C R24 SB7 MD13 39 6 MA8 43 118 A13 C
MD9 I/O13 VCC 0.01uF ON EACH MA8 AB13
MD14 40 21 MA11 44 117 A14 50MHz CRYSTAL OSCILLATOR MODULE
R28 SB6 I/O14 VCC VCC PIN TO GND MA11/GPIO2 AB14
MD10 MD15 41 22 MA9 45 116 A15
I/O15 VSS MA9/GPIO3 AB15
11 37 46 115 A16 VCC COM1 VCC
NC VSS GND VCC IOVDD SED1354 AB16
12 14 47 114 A17 SB17 1 8
NC RAS RAS# AB17 NC VCC
32 13 48 113 A18
NC WE WE# AB18 C34
29 30 49 112 A19
OE UCAS UCAS# AB19
42 31 50 111 A20 4 5
VSS LCAS LCAS# AB20 GND OP
51 110
GND VSS IOVDD VCC GND
HM5116160 MD7 52 109 GND
MD7 VSS GND
GND MD8 53 108 BUSCLK
MD8 BUSCLK BUSCLK
MD6 54 107
MD6 TESTEN GND
MD9 55 106 SUSPEND#/GPO
MD9 SUSPEND/GPO
MD5 56 105
MD5 CLKI
MD10 57 104 BACKLIGHT
MD10 VSS GND
MD4 58 103 CONTROL ?
MD4 VRTC/GPIO11
MD11 59 102
MD11 HRTC/GPIO10
MD3 60 101
MD3 DACRS1/GPIO9
MD12 61 100
Schematics
MD12 DACRS0/GPIO8
MD2 62 99
MD2 DACWR#/GPIO7
MD13 63 98
MD13 DACP0/GPIO6
MD1 64 97 COREVDD
MD1 COREVDD COREVDD
J4
1
B 2 B
MD14
MD0
MD15
VSS
FPFRAME
FPLINE
LCDPWR
MOD/DRDY/FPSHIFT2
FPSHIFT
VSS
FPDAT0
FPDAT1
FPDAT2
FPDAT3
FPDAT4
FPDAT5
FPDAT6
FPDAT7
IOVDD
DACRD#/GPIO4
BLANK#/GPIO5
DACCLK
VSS
FPDAT8
FPDAT9
FPDAT10
FPDAT11
FPDAT12
FPDAT13
FPDAT14
FPDAT15
VSS 3
U43 75LVDS84 4
MD14
65
66
67
68
69
70
71
72
73
74
75
76
77
78
79
80
81
82
83
84
85
86
87
88
89
90
91
92
93
94
95
96
44 41 5
MD0 D0 Y0M
45 40 6
MD15 GND GND VCC GND GND D1 Y0P
47 7
D2
48 39 8
D3 Y1M
1 38 9
D4 Y1P
LCDPWR 3 10
LCDPWR D5
4 35 11
D6 Y2M
FPSHIFT
6 34 12
D7 Y2P
7 13
D8
9 33 14
D9 CLKOUTM
10 32 15
D10 CLKOUTP
12 16
D11
13 37 17
D12 LVDSVCC
VCC 15 18
D13
16 31 19
D14 LVDSGND
2
4
6
8
10
12
14
16
18
20
22
24
26
28
30
18 36 20
J5 D15 LVDSGND
19 42
DB31CON D16 LVDSGND
20 29 HRS DF19G-20
1
3
5
7
9
D17 PLLVCC
11
13
15
17
19
21
23
25
27
29
31
22 28
D18 PLLGND
23 30 GND
D19 PLLGND
25
A D20 A
26 5
CLKIN GND
GND 27 11
SHTDN GND Title
2 17 %/($6(/9'6',63/$<,17(5)$&(
VCC GND
8 24
VCC GND
21 46
VCC GND Size Number Revision
GND B
LVVXH
281
1 2 3 4
282
5V DC SUPPLY
VCC
PSU+ 1 3
PSU+ INPUT OUTPUT
C21 C22
220uF 35V C1 C26
POWER C23 10uF 16V 0.1uF
GND TAB
GND
J3 10uF 25V 0.1uF
U41
1 GND LM2931 GND
4
2
3689
D 2 D
EP1 EP2 EP3 EP4
3
*1'
4 GND VCC EARTH POINTS
9 5 +12V R164 R80 GNDGND R79
J6
,19(57(5 6 1K BATONCHG GND
,19(57(5 7 GND R81 SENS- 1
FET12
%$77(5< 8 LED2 LED1 LED1 IRFR9024N 2
+12VINV SENS+ 26(1625
FET11 LED LED BAT ON LED 3
C75
R57 BAG/VENT 4 %$*9(17'(7(&7
2N7002 LED2
LED3 BAT LOW CGO DET 5 &*2'(7(&7
10K R165 AGND 6
C76 LED LED3
GND 1K ABS DETECT 7 $%6'(7(&7
PWR FAIL BACKLIGHT
8
FET2 GND GND 9 %$&./,*+7
LCDPWR BATLOW
GND LCDPWR 2N7002 10
PWRFAIL
INVERTER SUPPLY ENCO DIS 11 (1&2'(5',6$%/(
J1 12
+12VINV GND
1 GND 13 6,5,866:,7&+
6833/<
2 Q2 SIRIUS OFF 14
3.3V SUPPLY FOR LCD + CONTROLLER
C 3 U39 ZTX650 15 C
*1' GND PCDATA
4 3 1 COREVDD PSU+
INPUT OUTPUT COREVDD R96 CON15
%/212)) 5 R170
6
%5,*+7 100R
7
GND
8 C7 C17 C8
9 R169 LP8340CDT-3.3 10uF 16V R98
2
10uF 16V 0.1uF
10 10K
10K U42 7812CT +12V
GND
IN OUT
GND GND GND
GND
3 1
OUTPUT INPUT
GND J16 34 WAY HEADER
C66 C67
C64
GND
GND TAB
422RxD 422RxD#
START5
B R112 422RxD 1 2 422RxD# B
0.1uF 10uF 25V 422DTR
100uF 16V GND 3 4 422DTR
2
4
1k 422DTR# LIMSET O/P
422DTR# 5 6 LIMSET O/P
APNOEA O/P PRS_OVRLD O/P
R110 APNOEA O/P 7 8 PRS_OVRLD O/P
PSU+ PSU+
1k PSU+ 9 10 PSU+
GND GND GND 11 12 GND
422CTS 422CTS#
422CTS 13 14 422CTS#
U12A U14A 422TxD 422TxD#
R115 422TxD 15 16 422TxD#
1 14 1 14 PATPRS
ENCODER SW VCC VCC 10K 17 18 PATPRS
2 13 2 13 DRVPRS FLWPRS
DRVPRS 19 20 FLWPRS
12 12 FRESHGAS BATONCHG
FRESHGAS 21 22 BATONCHG
PWRFAIL TMPSNS
R111 C65 PWRFAIL 23 24 TMPSNS
R114 EXPVOL
39k 680NF 25 26 EXPVOL
3 2M2 3 GND 27 28 SIRIUS OFF
AGND MS29 MS30
D10 6 11 6 11 29 30
C63 8 8 31 32
DIODE MS33
100u 33 34 BAVREF
GND
5 5 Q1 FET10 Title
A %/($6(/9'6',63/$<,17(5)$&( A
4 9 4 9 IRFR9024N
ZTX650
7 10 GND 7 10
GND GND
Size Number Revision
ENCO DIS
R113 A4 LVVXH
GND GND GND
TP6
LS1 TP8 +6V SW-WAF
R32 R13 R14 NOTE: R57, D11 NOT FITTED
100R +26V D7 D8 R57 D11 GF1B PL2
SPEAKER
SBAT 22K 8K2 C43 GF1B GF1B VBAT 9
100n SW-WAF
U7 PL6
20
L3 D14 47R
74HC541 Q3 GF1B D12 GF1B
1 R58
100uH 1 STD12NF06L +24V
G1 2
D2 19 47R
G2 3
VCC
MBRD835 2 18 SBAT 3
R29 R59 R60 R61 U2 A1 Y1 4 SW-POL
GND 3 17
1.5K 1.5K 1.5K 1.5K A2 Y2 5 R24 R63 R64
Main Power 7 5 4 16 U5
IN SWITCH A3 Y3 6 100R 100R
3 2 5 15 Q2 TP7 LM341P5 100R
FB COMP A4 Y4
D 6 1 6 14 STD12NF06L 6-WAY 2.54mm KK D
FREQ.SYNC ON/OFF A5 Y5 R62
7 13 Pins 1-4 Flow Valves 3 1
A6 Y6 OUT IN
8 12 Pins 5-6 Inspire Valve 100R
GND
C19 R11 A7 Y7
9 11
GND
GND
10u C9 1K8 A8 Y8
C40 R12 TP10 C17
C16
4
1u 220u 1K5 LM2588S-ADJ D6 100n
TP9 10u
2
R31
10
ICCFLW
Q1 D21 ON-FLW 1
4K7 GF1B GND
BC847B BAT54
C10 C11 C12 R28 R26 ON-SOL GND
R30 220u 10K 10K R20 R21 R22
LK1 10K 4.7u 10u 2
OFFLOW
4
3
R73 R74 2R7 1R0 1K C15
10K 10K 10u 4
RL1 AD-ON
GND
G3VM-61D1
6
Inspire Valve Flow Valve AD-ST
ALMSIG ON-24V GND ICCSOL
GND 8
PD-ST
+24V +24V DC IN FROM PSU
TP4 10
Q12 R70 PD-ON
ALM+5V
1
2
R3 D3 Q14
TIP30A VBAT +6V 10WAY SW
680R BC847B
1R GF1B R1 D15
R4 R6 R7 5.1A
BATT CHARGER 1K 110K RT1 I= 1K5 GF1B C20
1R 3.0A D22 PL1
MFR300 100n 8EWS08SPBF 2
PD-ON
R72
BATT+ 4
C R8 4K7 PD-ST C
1 16 Main Power R23
CSO SINK 39K
2 15 SW-WAF GND 47K 6
CS- DRIV GND AD-ST
3 14 +6V
CS+ COMP
4 13 8
CL SENS AD-ON
5 12 BATT-
+VIN EN R19
6 11 10
4
3
2
1
GND BIAS R10 OFFLOW
7 10 C38 C2 100R
VI LO SLC 22u PL5 D13
8 9 R71 ON-24V 21
8
7
6
5
4
3
2
1
9
8
7
6
5
4
3
2
1
OCT OVR C7 360K BAT54 ON-24V
10
4-WAY 2.54mm KK 4K7
100n R9 DA-Lite
PL3 PL4 1000u ON-SOL 23
C8 R5 30K Work light ON-SOL
C6 U3 8-WAY 3.96mm KK 10-WAY 3.96mm KK GND
100n 1K D26
100n C39 Pins 1-7 Front Panel ON-FLW 22
Pins 1-4 BAV Board ON-FLW
Pins 8-10 PIB
Pins 5-6 +24V PSU R38 BAT54
24
Pins 7-8 Battery 22u SWBAT
100R TP3
GND C37
C5 R18
/OVRCH R53 TP2 U1 1n 1K +6V
10R 22u 24V Mains
LM2576T-ADJ
R75 9-14V Batt
TP5 GND D10 1 2 +6V 15
controller
INV-SUPPLY 1K R15 5 4 17
to
C35 C36
C14 ICCSOL 18
4.7u 4.7u R46 ICCSOL
TP1 GROUND R41 1 3 10µF
A TP2 MAIN POWER 2R7 ICCFLW 20 A
GND 13K VR1 ICCFLW
TP3 +6V LK2 10K
GND 26 W TRANS POWER
TP4 VBAT
TP5 INV-SUPPLY Title
TP6 +26V GND POWER SUPPLY BOARD
TP7 ALM+5V BLEASE MEDICAL
TP8 SBAT Size Number Revision
TP9 ICCSOL A3 10110376SCH 5
TP10 ICCFLW Date: 5-Jul-2006 Sheet 1 of 1
File: C:\10110376-4\Issue 4\10110376.DDB Drawn By: A.Atkins
1 2 3 4 5 6 7 8
283
1 2 3 4 5 6 7 8
284
A[1..19]
A[1..19]
D[0..15]
D[0..15]
A19
A18
A17
A16
A15
A14
A13
A12
A11
A10
A9
A8
A7
A6
A5
A4
A3
A2
A1
RP1
SPEN 1 2
D PL6 C2 3 4 D
U5 PL312 TXD C3 5 6
GND 1 2
C0 7 8
27
26
25
24
22
21
20
19
17
16
15
14
12
11
10
9
8
7
6
5
3
2
1
3 4
8
5 6
7 8 10K
A9
A8
A7
A6
A5
A4
A3
A2
A1
A23
A22
A21
A20
A19
A18
A17
A16
A15
A14
A13
A12
A11
A10
9 10 RP2
12.288MHz
CLK 100 HEADER 5X2 FC0 1 2
EXTAL
101 Auxiliary Serial Connector FC2 3 4
XTAL
98 DELAY 5 6
BUSCLK CLKO
RESET 92 VCC PL315 7 8
RESET /RESET
HALT 91
/HALT C3
BERR 94 GND 10K
R9 /BERR 10u 50V
C4 74 VCC
VCC VCC BUSW C2 RP3
4K7 73
GND DISCPU +
10u 50V C4
FRZ 72 + BCLR 1 2
/FRZ C1 4u7 63V
C[0..4] C0 104 U10 FRZ 3 4
C[0..4] /AS
C1 103 4u7 63V + 11 15 + NVHOLD 5 6
R/~W V+ V-
C2 106 10 13 NVWP 7 8
/UDS-A0 C1+ C2+
C3 105 12 14
/LDS-DS C1- C2-
C4 85 SCC21 5 2 PL312 TXD 10K
/DTACK T1I T1O DTR used to drive
PL1[1..26] 123 SCC25 18 1 PL314 DTR
PL1[1..26] /RMC-IOUT1 T2I T2O Autozero Valve RP4
122 MASTER SCC11 19 24 PL37
IAC T3I T3O
BCLR 86 30 D15 21 20 BERR 1 2
/BCLR D15 VCC T4I T4O
C5 90 31 D14 SCC20 6 7 PL311 RXD DB9 - Recorder AVEC 3 4
/BR D14 R1O R1I
VCC VCC 87 32 D13 SCC24 4 3 PL313 CTS C5 5 6
/BG D13 R2O R2I
C7 88 33 D12 SCC10 22 23 PL39 C7 7 8
/BGACK D12 R3O R3I
ICTL0 97 35 D11 17 16 DB25 - RS232
C /IPL0 D11 R4O R4I VCC C
U4D R11 ICTL1 96 36 D10 10K
R12 /IPL1 D10
SN74HC05N 1K2 ICTL2 95 37 D9 ADM238LJN PL3[1..26]
4K7 /IPL2 D9 PL3[1..26] RP5
FC0 132 38 D8 In IC Socket
FC0 D8
9 8 PL121 FC1 130 40 D7 SCC26 1 2
FC1 D7 R3 R4 Z1
FC2 129 41 D6 SCC24 3 4
FC2 D6 10K 10K 5.6V
AVEC 93 42 D5 SCC23 5 6
/AVEC-IOUT0 D5 R1
43 D4 SCC22 7 8
D4 VCC GND
6 5 ON_24V 69 45 D3 1K2
/DREQ-PA13 D3
U4C TEST0 70 46 D2 10K
/DACK-PA14 D2
SN74HC05N 71 47 D1 RESET 4 3 PL214 GND
/DONE-PA15 D1 RP6
48 D0
D0 U4B
CS0 128 C1 1 2
/CS0-IOUT2 R2 SN74HC05N
CS1 127 ICTL0 3 4
/CS1 VCC
TEST[0..4] CS2 125 1K2 ICTL1 5 6 VCC
TEST[0..4] /CS2
CS3 124 ICTL2 7 8
/CS3
VCC HALT 2 1
WATCHDOG 4K7
CS[0..3] U4A
CS[0..3] SN74HC05N RP7
10K VCC
R97 SCC12 1 2
U1 JP2 SCC13 3 4
RXD1
TXD1
RCLK1
TCLK1
/CTS1
/RTS1
/DCD1
BRG1
RXD2-PA0
TXD2-PA1
RCLK2-PA2
TCLK2-PA3
/CTS2-PA4
/RTS2-PA5
/DCD2-PA6
BRG2-SDS2-PA7
RXD3-PA8
TXD3-PA9
RCLK3-PA10
TCLK3-PA11
/CTS3-SPRXD
/RTS3-SPTXD
/DCD3-SPCLK
BRG3-PA12
/IACK7-PB0
/IACK6-PB1
/IACK1-PB2
TIN1-PB3
/TOUT1-PB4
TIN2-PB5
/TOUT2-PB6
/WDOG-PB7
PB8
PB9
PB10
PB11
U3
JP1 1 8 SPI-CLK 5 6
MC68302FC16C PB RST VCC 1
Schematics
DELAY 2 7 ON_24V 7 8
1 TD ST 2
52
80
82
81
51
79
50
76
53
54
55
56
58
59
60
61
63
64
65
66
49
78
77
68
3 6
108
109
110
111
113
114
115
117
118
119
120
121
2 TOL RST 3
VCC 4 5 10K
GND RST
HEADER 2 GND
FP[0..2] MAX1232CPA SWITCH TO VCC RP8
FP[0..2]
WATCHDOG GND CLK FORCE FLASH MODE SCC14 1 2
SCC10
SCC11
SCC12
SCC13
SCC14
SCC16
SCC20
SCC21
SCC22
SCC23
SCC24
SCC25
SCC26
CLK
B R13 B
GND SCC16 3 4
4K7
FP0 Input data VCC SPI-Din 5 6
FP1 Output data FC1 7 8
FP2 Output data enable
VCC
SPI-Din Calibrate Protection 10K
SPI-Din S2
SPI-Dout GND
SPI-Dout OVERLOAD_INT R5 GND
SPI-CLK PL249
SPI-CLK
R14 10K
DONE
4K7
INIT
VCC
SPI-ENA1
VCC 10 11 12 13
D1 BZX84-C4V7 D2
BAR42 R6 VCC R108 U4E U4F U7B
10
1
33K R106 10K
U2 MC74HC74ANG Pressure Switch 12 9
R7 20K R16 D Q
SD
6 1 6
Backup Vbat SCLK Q 10K R8
CD
100K 7 2 PL2[1..50] 11
X2 Dout PL2[1..50] VCC CLK
D3 + 8 3 3 12 13 PL235 10K
X1 Din CLK
BAR42 C5 X2 4 5 8
CD
GND CS Q
10u 50V 5 2 FLWSIG JP4
SD
Q D U41F
MAX6902 MC74HC14ADG HEADER 2 GND
13
4
ALTERNATIVE A/D CONVERTER GND C47
10n 2
1 8
+5VA VCC VCC +5VA VCC
2 7 U6 GND VCC
D D
U52C
ALARM7 Pressure High Display 9 U52A
8 1
10 3
PHIDSP
2
D17 U42F MC74HC132ANG
13 12 MC74HC132ANG
OVERLOAD_INT
U52D
BAR42 MC74HC14ADG
12
PRS_OVRLD
11
U52B 13
ALARM6 Pressure Low Display 4 U53D
6 MC74HC132ANG 12
5 11
PLODSP
13
D18 U42A MC74HC132ANG
1 2 MC74HC132ANG
U53A
BAR42 MC74HC14ADG
1
APNOEA
3 PL2[1..50]
C PL2[1..50] C
U54A 2
ALARM5 Power Fail LED 1 U53C U40C
3 MC74HC132ANG 9 8
+5VALM PWR_FAIL 2 8 10 R15 PL244
GND
10 9 390R
R73 U42B D20 MC74HC132ANG TP9
TP18 3 4 BAR42 R88 MC74HC132ANG MC74HC02ANG
8
160K
REF1 470K
LT1004CZ1.2 R72 47K 3 U32A +5VALM PL423
R74 MC74HC14ADG U51D
1 12 U53B
R69 R71 68K 2 11 4
100R
62K LMC662CN 13 6
R70 10K 5
GND
4
PL113 +25V + MC74HC132ANG
R87 C19 MC74HC132ANG
GND 560K 2u2 63V
ALARM4 Mute LED
ALARM3 Mute Audio U42C
GND D25
VCC U40B
5 6 5 BAR42
4 R89 PL243
R75
2M2 MC74HC14ADG 6 0R0
D19 U34A U40A
Schematics
BAR42 2 6 2 MC74HC02ANG
RC Q
1
+ C20 3
100u 25V U51C
1 9 MC74HC02ANG
C
B 8 U54C B
4 7 10 9
A Q
5 U42D 8
GND B MC74HC132ANG COMALM
9 8 10
3
VCC CLR
MC74HC132ANG
MC74HC14ADG
MC74HC4538ADG D22
ALARM[1..7]
U51B
4 BAR42
U42E
6
11 10 R77 5 U40D
RESET U54B
47K 4 U54D 11 PL1[1..26]
MC74HC132ANG PL1[1..26]
D15 6 12 13 PL17
MC74HC14ADG BAR42 1 5 11 12
TP10
ALARM2 Continuous Audio 13
R78 MC74HC132ANG MC74HC02ANG
47K MC74HC132ANG
R86
D16 +5VALM 1K0
BAR42 U51A
ALARM1 Pulse Audio 1 R84
3 R81 100K
2 100K
+5VALM U32B
5
ALARM[1..7] D21 MC74HC132ANG TP11
+ 7
BAR42 R79 R83
U34B (560K) 6
R76 14 10 NOT FITTED R80 R82 LMC662CN
VCC RC Q 100K 10M
100K C21 470K
C36 GND 2u2 63V R85
10n GND
100K
285
1 2 3 4 5 6 7 8
286
D D
U60B
U61B 12
TP17 GND 1,2EN
U56 12 10
GND 1,2EN 1Y
LM2937ET-5.0 11
1Z
9 14 10 11
+6V VCC GND 1A 2Y VCC 1A 1Y
1 3 PL5 15 13 9
+6V IN OUT VCC GND 2A 2Z GND 1B Backup
PL54 13
4 2Y
BATONCHG DS34C87TN 14 PL1[1..26]
3 VCC 2A PL1[1..26]
+ C22 15
GND
C40 100u 25V C38 TP16 2 GND GND 2B
PSU+ PL1
100n 100n 1 VCC
DS34C86TN GND
1
2
CON4 R90 PL12 PL2[1..50]
GND 2 PL2[1..50]
47K GND
GND 3
PL4 U59 PL14
4
PL12 Lamp Backlight 1 16 PL224 PL15
34 UPDOWN I1 O1 5
PL54 2 15 PL240 PL16 PL2
33 PHIDSP I2 O2 6
PL14 3 14 PL241 PL17 PL21
32 PLODSP I3 O3 7 1
PL16 4 13 PL242 PL18 PL22
TP15 31 INODSP I4 O4 8 2
U57 PL18 5 12 PL19 PL23
30 COMALM I5 O5 9 3
LP2950CZ-5.0 PL15 6 11 PL110 PL24
29 I6 O6 10 4
PL110 7 10 +5VALM PL25
+5VA 28 I7 O7 11 5
1 3 8 9 PL112 PL26
IN OUT +5VA 27 GND GND GND VCC +6V 12 6
PL326 PL113 +25V PL27
26 DS2003CN 13 7
C 25 AGND 14 +5VMAG 8 C
+ C23 D23 +6V PL29
GND
C39 10u 50V C41 24 15 9
PL423 GND VCC PL116 PL210
100n BAR42 100n 23 16 10
BATONCHG +6V PL211
22 17 11
2
PL321 PL118 PL212
21 R104 18 12
PL320 U60A PL119 PL213
20 10K 19 13
PL319 4 PL120 PL214
19 VCC 1,2EN 20 14
PL318 R103 PL121 PL215
18 21 15
PL317 10K PL122 PL216
17 22 16
422TxD# 422RxD 2 3 FP0 PL123 PL217
TP14 16 1A 1Y 23 17
422TxD 422RxD# 1 PL124 PL218
15 1B 24 18
422CTS# 5 PL125 PL219
14 2Y 25 19
U55 4VREF 422CTS 422CTS 6 GND PL220
4VREF 13 2A 26 20
8 1 422CTS# 7 PL221
VIN VOUT 12 GND 2B 21
7 5 GND VCC CON26 PL222
FDBK ERR 11 GND 22
3 R93 PSU+ DS34C86TN FP[0..2] +6V
SHTDN C42 10 FP[0..2] 23
22K1 0.1% PSU+ PL224
100n 9 R102 24
6 4VREF
TAP R92 8 PRS_OVRLD 10K 25
2 + PL3[1..26] AGND
GND
SENS C43 7 APNOEA PL3[1..26] 26
300R C24 PL26 R101 PL227
100n 6 27
LP2951 10u 50V 422DTR# 10K U61A PL228
5 28
4
R91 422DTR 422TxD 2 4 FP2 PL229
4 1Y 1,2EN 29
10K 0.1% 422TxD# 3 PL3 PL230
3 GND 1Z 30
Schematics
GND
SENS 150u 10V AGND PL246
GND 16 46
100n PL317 PL247
LP2951 17 47
PL318 RESET
4
L1 100uH +5VA 18 48
PL319 PL249
C50 C51 UC27A UC28A UC29A UC30A UC31A UC20A UC21A UC22A UC23A 19 49
PL320 PL250
+
150u 10V
+
150u 10V 100n 100n 100n 100n 100n 1n 1n 1n 1n 20 50
PL321
AGND 21
PL322 CON50
UC27B UC28B UC29B UC30B UC31B UC20B UC21B UC22B UC23B 22
4VREF
100n 100n 100n 100n 100n 1n 1n 1n 1n 23
-5VA
-5VA 24
+5VA
25
Power supply decoupling - adjacent to I.C of same number PL326
26
VCC
CON26
UC1 UC2 UC4 UC5 UC6 UC7 UC8 UC9 UC10 UC12 UC13 UC14 UC15 UC16 UC17 UC18 UC24 UC25
1n 1n 1n 1n 1n 1n 1n 1n 1n 1n 1n 1n 1n 1n 1n 1n 1n 1n
GND
VCC
UC26 UC3A UC3B UC3C UC33 UC34 UC36 UC37 UC38 UC39 UC40 UC41 UC42 UC43 UC44 UC45 UC46 UC47 UC48 UC49 UC50
D D
R46
ANALOGSIG
4K7 RESET
D26
BAR42
U24
1
OC GND
11
CLK ANALOGSEL
AGND
19 2 D0
AGND 1Q 1D
18 3 D1
2Q 2D
17 4 D2
D4 D5 D6 3Q 3D
16 5 D3
BAR42 BAR42 BAR42 4Q 4D
15 6 D4
5Q 5D
14 7 D5
6Q 6D
13 8 D6
7Q 7D
U20 12 9 D7
8Q 8D
PL318 Patient Pressure 13 3
X0 X
PL319 Drive Pressure 14 MC74HC574ADWG
C X1 U17 C
PL320 Flow Pressure 15 9
X2 C VCC
RP11 SWBAT 12 10 7 6
100K X3 B Y7 E3
+24V 1 11 9 5 R45 D8 D[0..15]
X4 A Y6 E2 GND D[0..15]
PL15 1 2 SBAT 5 10 4 1M BAR42
X5 Y5 E1 GND
PL113 3 4 +25V 2 11 U33A
X6 Y4
PL116 5 6 RP10 TEST3 4 6 12 2 6
+25V X7 INH Y3 RC Q VCC TP19
PL124 7 8 10K 13 3 C13
+24V Y2 C
1 2 7 14 2 47u 16V +
4
1M Q
5 6 13 3 3
X0 X CLR
7 8 14 MC74HC74ANG R44
+5VA X1
1
15 9 MC74HC4538ADG 10K
X2 C -5VA
1 2 12 10 PL317
+5VMAG X3 B C28
3 4 1 11
4
3 4 3 4 MC74HC4051ANG R43
100K R36 33K R37
5 6 5 6 PL14
+5VA
B 7 8 7 8 470K D13 B
AGND
RP14 RP15 AGND BAR42
U22 R35
10K 10K GND
PL3[1..26] R96 TEST1 Self-Test 1 13 3 C16
PL3[1..26] EXTFLW X0 X 10K PRS_OVRLD
14 1u0 63V Pressure Limit Set
10K R60 X1 R30
15 9 PL26
EXTPRS X2 C AGND TP4
+
PL1[1..26] U28B
+ 2 10K 1u0 63V U9B
AGND U28A TLC2272A D10
C11 C10 15 12 9
TLC2272A BAR42 GND C D Q
SD
X2 C R31 Q
PL27 1K0 CPP Set 12 10 2u2 63V R115 -5VA 13
X3 B 10K CLR
PL26 Pressure Limit Set 1 11 33K2 MC74HC74ANG
X4 A 0.1% TP3
13
287
1 2 3 4 5 6 7 8
288
D D
VCC
2
4
6
8
2
4
6
8
RP18 10K
10K RP17
RESET
PL1[1..26] D[0..15]
PL1[1..26] D[0..15]
1
3
5
7
1
3
5
7
U44
PL2[1..50] PL18 Adult On 2 18 D0 CS[0..3]
PL2[1..50] 1A1 1Y1 CS[0..3]
PL16 Adult Standby 4 16 D1 U18
1A2 1Y2
PL12 Paed On 6 14 D4 7 6 A[1..19]
1A3 1Y3 Y7 E3 A[1..19]
PL232 /PRSSEL 8 12 D5 9 5 CS2
1A4 1Y4 PWMLATCH Y6 E2
10 4
Y5 E1 GND
TEST1 11 9 D7 11
2A1 2Y1 Y4
13 7 D6 12
2A2 2Y2 DACCS Y3
PL14 Paed Standby 15 5 D3 13 3 A5
2A3 2Y3 DACSTRB Y2 C
PL110 Off 17 3 D2 14 2 A4
2A4 2Y4 ANALOGSEL Y1 B
15 1 A3
Y0 A
1
VCC 1G
19 MC74HC138ADG
2G
2
4
6
8
2
4
6
8
RP20 RP19 MC74HC244ANG
C 10K 10K C
U46
1
3
5
7
1
3
5
7
7 6
Y7 E3
U45 9 5 CS3
Y6 E2
PL112 /OVRCH 2 18 D8 10 4
1A1 1Y1 Y5 E1 GND
PL231 /MUTE 4 16 D9 11
1A2 1Y2 Y4
PL230 /TSTRST 6 14 D10 12
1A3 1Y3 Y3
PL233 /VOLSEL 8 12 D11 13 3 A3
1A4 1Y4 Y2 C
14 2 A2
Y1 B
PL229 /INTLOC 11 9 D12 15 1 A1
2A1 2Y1 Y0 A
PL236 DISCON 13 7 D13 A4
2A2 2Y2
PL250 15 5 D14 MC74HC138ADG PL2[1..50]
2A3 2Y3 PL2[1..50]
TEST2 17 3 D15
2A4 2Y4
+ C18 1
10u 50V 1G
19 U47
2G
2 18 STRB1 PL29
D1 Q1
MC74HC244ANG 3 17 STRB2 PL210
D2 Q2
GND 4 16 STRB3 PL211
D3 Q3
TEST1 TEST1 U49 5 15 STRB4 PL212
Schematics
D4 Q4
2 3 D15 6 14 STRB5 PL213
Q1 D1 D5 Q5
PL125 ON_PSU 5 4 D14 7 13 PL246
Q2 D2 D6 Q6
PL119 ON_BAT 6 7 D13 8 12 PL247
Q3 D3 D7 Q7
PL19 ON_LEK 9 8 D12 9 11
Q4 D4 D8 Q8
12 13 D11
CLRBAK Q5 D5
B 15 14 D10 1 B
EN_SOL Q6 D6 E1
16 17 D9 19
EN_FLW Q7 D7 E2
TEST2 19 18 D8
Q8 D8
TEST[0..4] GND MC74HC540ANG
TEST[0..4]
11
CLK
1
CLR
MC74HC273ANG
ALARM[1..7]
ALARM[1..7] Display Data
U50 U48
ALARM2 2 3 D2 D7 9 12 PL215
Q1 D1 8D 8Q
ALARM3 5 4 D3 D6 8 13 PL216
Q2 D2 7D 7Q
ALARM6 6 7 D6 D5 7 14 PL217
Q3 D3 6D 6Q
ALARM7 9 8 D7 D4 6 15 PL218
Q4 D4 5D 5Q
ALARM5 12 13 D5 D3 5 16 PL219
GND Q5 D5 4D 4Q
ALARM4 15 14 D4 D2 4 17 PL220
Q6 D6 3D 3Q
ALARM1 16 17 D1 D1 3 18 PL221
2
4
6
8
2
4
6
8
Q7 D7 2D 2Q
TEST3 19 18 D0 D0 2 19 PL222
Q8 D8 1D 1Q
RP16 RP9 11 11
10K CLK CLK
10K 1 1
CLR OC
VCC
1
3
5
7
1
3
5
7
MC74HC273ANG GND MC74HC574ANG
JP3 U63
2 18 D0
1 2 1A1 1Y1
4 16 D1
3 4 1A2 1Y2
6 14 D2
5 6 1A3 1Y3
A 8 12 D3 A
7 8 1A4 1Y4
4VREF
D C35 D
A[1..19] R51 100n R68 PL228
A[1..19]
20K 2K0 TP6
DACSTRB R67
DACCS VCC 20K
-5VA EXTPRS
D[0..15] C31 D24
D[0..15] Not Fitted
21
4
4
33p
U25 U43 PL3[1..26]
PL3[1..26]
1 2
GND OC R52
11 D7 6 3 1 6
CLK D7 RFBA
VCC
U36 D6 7 2 3 7 R66 PRSOUT PL34
D6 IOUTA U31A 10K
VREFA
D0 2 19 3 14 D5 8 LMC662CN 5 2K0 [FG Autozero 8500/6500]
1D 1Q P0 Q0 D5 U31B
D1 3 18 4 13 D4 9
2D 2Q P1 Q1 D4
8
D2 4 17 5 12 D3 10 LMC662CN
3D 3Q P2 Q2 D3
D3 5 16 6 11 D2 11
4D 4Q P3 Q3 D2 +5VA
D4 6 15 15 D1 14
5D 5Q TC D1
D5 7 14 7 D0 15 1
6D 6Q CEP D0 AGND AGND
D6 8 13 10
7D 7Q VCC CET
D7 9 12 2 A1 17
8D 8Q CLK M/ L +5VA
9 A2 18
PE B/ A
MC74HC574ADWG 1 19
8
MR LDAC
20 5
WR R65
MC74HC161ADG 16 24 3 7 FLWOUT PL31
VCC RESET IOUTB R53
23 1 6 2K0 [PF Auto zero 8500/6500]
EN_FLW RFBB U30B
2 10K
GND
VREFB
PWMDIVCLK U30A LMC662CN
U37 C32 LMC662CN
C CLK 33p R64 TP7 C
3 14 DAC8248FP
P0 Q0
5
4
22
4 13 20K
P1 Q1
5 12
P2 Q2 GND R54
6 11
P3 Q3 4VREF EXTFLW
15 20K C34
TC -5VA
7 100n
CEP R49 TP5
10
CET
2 100K
CLK R56
9 PL2[1..50]
PE R55 +5VA PL2[1..50]
1 56K
GND MR TP8
+5VA 100K
U26 MC74HC161ADG Meter Zero
1 +5VA R100 5
OC R63
11 10K 7 PL227
PWMLATCH CLK R58
U38 6 2K0
D8 2 19 3 14 100K U29B
8
1D 1Q P0 Q0 R57
D9 3 18 4 13 LMC662CN
2D 2Q P1 Q1 U29A -5VA
D10 4 17 5 12 3 47K C33
3D 3Q P2 Q2 R50
D11 5 16 6 11 1 100n
4D 4Q P3 Q3
D12 6 15 15 2 33K
5D 5Q TC
D13 7 14 7 C46 LMC662CN AGND
6D 6Q CEP R59 470n16V
D14 8 13 10 100K
7D 7Q CET
4
R10
Schematics
D15 9 12 2 0R0
8D 8Q CLK
9
MC74HC574ADWG PE
1
MR
15HZ
MC74HC161ADG
B -5VA B
R61
U8 1M PL1[1..26]
PL1[1..26]
3 14
P0 Q0
4 13
3 4 P1 Q1
GND 5 12 U39A
P2 Q2
U41B 6 11 11 10 2 U39C
P3 Q3 R62
MC74HC14ADG 15 1 1 2 8
TC U41E
7 3 10K 10 Flow Control Valve ON_FLW PL122
CEP MC74HC14ADG U41A
10 9 1KHz Pulse Width Modulation
CET MC74HC02ANG C30 MC74HC14ADG
2
CLK 220p MC74HC02ANG
9
PE
1
MR
U39D
MC74HC161ADG U39B GND 11
U41C ON_SOL
Valve Override Control 5 13 Dump Solenoid Valve PL123
PRS_OVRLD
4 5 6 12
U19A 6
1 MC74HC02ANG
EN_SOL MC74HC02ANG MC74HC14ADG
3
2
PWM
MC74HC132AN
VCC
R47
289
1 2 3 4 5 6
290
D D
VCC VCC
BUSCLK
R110
10K
CS[0..3]
12
25
37
51
63
75
89
100
CS[0..3] U19B
CS2 4 U19C U35
6 9
5 8 24
Vcc
Vcc
Vcc
Vcc
Vcc
Vcc
Vcc
Vcc
M0 PWM
A8 10 22
MC74HC132ANG M1
26
MC74HC132ANG PWRDWN
2 70
PGCLK1 Press. Bypass
C[0..4] 31 65 JP8
C[0..4] CS PWM0
C1 32 66
R/W PWM1 1
A1 33 67
A1 GAS0 2
A[1..19] A2 34 68
C A[1..19] A2 GAS1 C
A3 35 69 HEADER 2
A3 GAS2
D[0..15] GND
D[0..15]
D0 3 90
D0 ANSEL0
D1 4 91
D1 ANSEL1
D2 5 92
D2 ANSEL2
D3 99 94
D3 SPARE
D4 7
D4
D5 8 39
D5 I0
D6 9 40
D6 I1
D7 10 41
D7 I2
D8 13 42 JP6
D8 I3
VCC D9 14 43
D9 O0 2 1
D10 15 44
D10 O1 4 3
D11 16 45
D11 O2 6 5
D12 17 46
D12 O3 8 7
D13 18
D13 10 9
D14 19 53 I0
R109 D14 IO0
D15 20 54 I1 HEADER 5X2 VCC
4K7 D15 IO1
55 I2 GND GND
Schematics
R111 IO2
1K 56 I3
IO3
50 57 I4
DONE DONE IO4 R116
52 58 I5
SPI-ENA1 Program IO5 (2K2)
72 59 I6
SPI-Dout Din IO6
B 73 60 I7 NOT FITTED B
SPI-Din Dout IO7
74 I[0..7] U64
SPI-CLK CCLK
36 71
INIT INIT PWMTEST 3 4
93
R112 ALARM
1K I7
2 5
TMS
GND
GND
GND
GND
GND
GND
GND
GND
R113 1 6
XCS05VQ100
2K2
6
1
11
23
38
49
64
77
88
DS2401P
GND NOT FITTED
JP7
I3 I4
1 2
GND I2 I5
3 4
I1 I6
5 6
U19D I0 I7
7 8
12
GND 9 10
11 GND
11 12 GND
13
GND
HEADER 6X2
MC74HC132ANG
A A
Title
GND CONTROLLER BOARD ;LOLQ[3:0DQG,2
1
2
3
4
5
6
7
8
291
Schematics
292
1 2 3 4
J6
F5
L L
2 Wire links fitted in fuse
N
FUSE 1
VENTILATOR
D D
J1
F6
N N
1
L
2 FUSE
MAINS IN
J7
WHEN THERE IS NO TRANSFORMER F7
L L
ASSMBLY FITTED J3 (NO TX OUT) IS 2 Fuses and sockets not fitted
FUSE N
CONNECTED TO J4 (SKTS IN) 1
J2
WHEN THERE IS A TRANSFORMER
N SPARE
ASSMBLY FITTED J2 (TX OUT) IS 1
L
CONNECTED TO J1 (IN) OF THE 2 F8
N
TRANSFORMER ASSEMBLY AND J2 TX OUT
(ISOLATED OUT) OF THE FUSE
TRANSFORMER ASSEMBLY IS
CB1 J8
CONNECTED TO J4 (SKTS IN)
J3 1 2 L
CIRCUIT BREAKER 2 When 110v 3A breaker and 4A fuse
L N
1 1 When 230v 3A breaker and 4A fuse
N
C 2 C
SKT 1
3 F1
L
4
N
5 FUSE
NO TX OUT
CB2 J9
1 2 L
CIRCUIT BREAKER 2 When 110v 3A breaker and 4A fuse
N
1 When 230v 2A breaker and 3A fuse
J4 SKT 2
L
1 F2
N
2
L
3 FUSE
Schematics
N
4
SKTS IN CB3 J10
JUMPER1-2 1 2 L
CIRCUIT BREAKER 2 When 110v 3A breaker and 4A fuse
N
B 1 When 230v 2A breaker and 3A fuse B
Vent +ve JUMPER1-3
3
1
J5 SKT 3
SSR1
5 3 Switched Live F3
4 2 Live To CBs
3 PF240D25 1 Permenant Live FUSE
2
JUMPER1
4
2
1
Vent -ve CON3 CB4 J11
SSR IN 1 2 L
CIRCUIT BREAKER 2 When 110v 3A breaker and 4A fuse
JUMPER1-1 N
1 When 230v 2A breaker and 3A fuse
SKT 4
Live To CBs normally connected to Permanent Live
Connected to Switched Live when sockets are to F4
be switched on and off with trolley
FUSE
Title
A A
Sirius Power Distribution 110VAC
293
1 2 3 4
294
R6
1R8
FS2
D D
6 4
R3 1A-T FS2-2
10R
G1 2W RELAY+ 0 1
250V R4
LIVEIN
10R
SSLIVE
R3-1
2W G7L-1A
D1 D3 K1
J1
L VR1 1N4007 1N4007
2 TB11
IN N R1 250V C1 1u5
G1-2
1 140V C1-2
+ C3 + C4 + C5 D5 2
POWER IN 1
OUT
1N5359B
470u 470u 470u 24V
R2
VR1-1
C2 2u2 35V 35V 35V SOFT-START POWER
140V D2 D4
TO PART B
1N4007 1N4007
FS1 RELAY-
NEUTRAL
R4-1
1A-T R5
C C
1M
PART A
SOFT-START
Schematics
TB12 TB1
TXNEUTRAL
1 1 YELLOW
SOFT-START IN 2
TXLIVE LK1-4
2 YELLOW
PART B
TB7 TRANSFORMER
POWER FROM TB2 SECONDARY
B GREY 1 B
LK3-3
LK1-2
2 1 WHITE BLUE 1 4
1 2 PINK RED 2
LINKS3
TRANSFORMER 1 PRIMARY
LINKS1
TRANSFORMER 1 SECONDARY
J2
TB4
LSEC1 L
1
1 YELLOW NSEC1 N SEC 1 OUT
2
LK2 2 YELLOW LSEC2 L
LK2-4 NSEC2 N
3
SEC 2
ISOLATED
4 TB5 4
LK2-3 TB9
3
LK2-2 1 BROWN SECONDARYS OUT
2 GREY 1
2 BLACK
1 VIOLET 2
LINKS2
TB6
LK4
1
110V
LK4-2
TRANSFORMER 2 PRIMARY
RED 2 4
Size Number Revision
LINKS4 13600379SCH 1
TRANSFORMER 2 SECONDARY
A4
Date: 24-Nov-2006 Sheet 1of 1
File: D:\Soft-start&Transformer Board\Protel\13600375-1.Ddb
Drawn By:
1 2 3 4
1 2 3 4
R6
6R8
FS2
D D
6 4
R3 1A-T FS2-2
10R
G1 2W RELAY+ 0 1
470V R4
LIVEIN
10R
SSLIVE
R3-1
2W G7L-1A
D1 D3 K1
J1
C1 1N4007 1N4007
L VR1
2 NOT FITTED TB11
IN N R1 440V
G1-2
1 275V C1-2
+ C3 + C4 + C5 D5 2
POWER IN 1
OUT
1N5359B
470u 470u 470u 24V
R2
VR1-1
C2 2u2 35V 35V 35V SOFT-START POWER
275V D2 D4
TO PART B
1N4007 1N4007
FS1 RELAY-
NEUTRAL
R4-1
1A-T R5
C C
1M
PART A
SOFT-START
Schematics
TB12 TB1
TXNEUTRAL
1 1 YELLOW
SOFT-START IN 2
TXLIVE LK1-4
2 YELLOW
PART B
TB7 TRANSFORMER
POWER FROM TB2 SECONDARY
B GREY 1 B
LK3-3
LK1-2
2 1 WHITE BLUE 1 4
1 2 PINK RED 2
LINKS3
TRANSFORMER 1 PRIMARY
LINKS1
TRANSFORMER 1 SECONDARY
J2
TB4
LSEC1 L
1
1 YELLOW NSEC1 N SEC 1 OUT
2
LK2 2 YELLOW LSEC2 L
LK2-4 NSEC2 N
3
SEC 2
ISOLATED
4 TB5 4
LK2-3 TB9
3
LK2-2 1 BROWN SECONDARYS OUT
2 GREY 1
2 BLACK
1 VIOLET 2
LINKS2
TB6
LK4
1
230V
LK4-2
TRANSFORMER 2 PRIMARY
RED 2 4
Size Number Revision
LINKS4 13600380SCH 1
TRANSFORMER 2 SECONDARY
A4
Date: 24-Nov-2006 Sheet 1of 1
File: D:\Soft-start&Transformer Board\Protel\13600375-1.Ddb
Drawn By:
1 2 3 4
295
NOTES
296
297
Spacelabs Healthcare Ltd.
Spacelabs Healthcare Inc.
1 Harforde Court,
35301 SE Center Street
John Tate Road,
Snoqualmie, WA 98065
Hertford
USA
SG13 7NW. United Kingdom
Tel: +1 425-396-3300
Tel: +44 (0)1992 507700
Fax: +1 425-396-3301
Fax: +44 (0)1992 501213
North America: 800-287-7108
e-mail (enquiries): advsales@spacelabs.com e-mail (technical): advsupport@spacelabs.com
www.spacelabshealthcare.com
298