Professional Documents
Culture Documents
Title:
Operational Qualification Protocol for the Capsule Filling Machine.
Objective:
The objective of this protocol is to document the satisfactory operation of the Capsule Filling
Machine used for manufacturing at National Institute of Health
Signature: Signature:
Date: Date:
Quality Assurance
Approved By: ______________________ Approved By: ______________________
Signature: Signature:
Date: Date:
Qualification
Process PROPRIETARY INFORMATION:
Solutions, Inc. Unauthorized Duplication Is Prohibited
DOCUMENT NUMBER
DATE: 3/16/10 PAGE: 1 of 36
REVISION NUMBER 1
Author: Michael Molloy ______________________ Revision DATE: N/A
Operational Qualification
OPERATIONAL QUALIFICATION
FOR THE
INSTALLED
AT
TABLE OF CONTENTS
1.0 Purpose3
2.0 Scope 3
3.0 Equipment Description 3
4.0 Responsibilities 4
5.0 Reference Documents 4
6.0 Definitions 5
7.0 Test Function Number One: OQ Signature Verification List (Attachment #1) 5
8.0 Test Function Number Two: Test Equipment (Attachment #2) 5
9.0 Test Function Number Three: Procedure List (Attachment #3) 6
10.0 Test Function Number Four: Training Verification (Attachment #4) 6
11.0 Test Function Number Five: Operator Controls and Indicators (Attachment #5) 6
12.0 Test Function Number Six: Functional Verification (Attachment #6) 7
13.0 Test Function Number Seven: Alarms, Safeties and Interlocks (Attachment #7) 7
14.0 Test Function Number Eight: Sequence of Operations (Attachment #8) 7
15.0 Test Function Number Nine: Maximum / Minimum Values (Attachment #9) 7
16.0 Deviation Reporting Log (Attachment #10) 8
17.0 Operational Qualification Protocol Review (Attachment #11) 8
18.0 Summary Report Requirements 8
19.0 Attachment List10
DOC. #
OPERATIONAL QUALIFICATION PROTOCOL
Capsule Filling Machine
National Institute of Health, Building 10 Room 1n249
Bethesda MD 20892
Page 3 of 36
1.0 Purpose
The objective of this protocol is to verify that the Capsule Filling Machine, described in Section 3.0,
operates in accordance to manufacturers’ specification, and in accordance with Current Good
Manufacturing Practices (cGMP’s).
2.0 Scope
This Operational Qualification pertains to the verification of the operation of the equipment described
in Section 3.0 of this protocol installed in the National Institute of Health, Building 10 room 1N249,
Bethesda MD, 20892. This protocol is specific to the Capsule Filling Machine and related
instrumentation and equipment. This protocol is specific to the Capsule Filling System, and related
instrumentation and equipment, and does not address other facility, or facility related equipment and
systems.
The filler is an alternate movement capsule filling machine provided with a rotary stage that provides
clockwise axis centered slip stick motion to transport capsules thru 8 process stages. The stages are:
1) Base
2) Motorization
3) Moving Parts
4) Protective hood
5) Electric cabinet and control panel
6) Size-change parts for process changeover
7) Manual controls
8) Powder feed unit
9) Powder layer and leveling control
10) Pellet or tablet dosing unit
11) Powder or pellet dosing unit
12) Defective capsule selection unit
13) Capsule closing unit
DOC. #
OPERATIONAL QUALIFICATION PROTOCOL
Capsule Filling Machine
National Institute of Health, Building 10 Room 1n249
Bethesda MD 20892
Page 4 of 36
4.0 Responsibilities
4.1 Qualification Process Solutions, LLC.
4.1.1 It is the responsibility of QPS to write this protocol.
4.1.2 It is the responsibility of QPS to execute this protocol.
6.0 Definitions
6.1 Critical Instrumentation – Critical instruments are defined as those whose performance will
affect the operation of the system and/or the quality attributes of the processed materials.
Critical instruments are calibrated and certified using standards that are traceable to NIST
(National Institutes of Standards and Testing).
7.0 Test Function Number One: OQ Signature Verification List (Attachment #1)
7.1 Objective
To identify and document all personnel involved in executing this protocol.
7.2 Procedure
Verify that each person who executes this protocol signs Attachment #1. Each person must be
identified by name, title, company, signature, initials and date on the attached form.
7.3 Acceptance Criteria
Signatures from all personnel involved in execution of this protocol shall be present.
8.2.2 Verify that the test equipment/instruments are calibrated and the reference standards
are traceable to National Institute for Standards and Technology (NIST) or other
approved standard.
8.2.3 Complete the applicable attachment, indicating the equipment description, ID/serial
number, calibration date, re-calibration date
8.3 Acceptance Criteria
All equipment required for execution shall be calibrated with copies of calibration records
attached.
11.0 Test Function Number Five: Operator Controls and Indicators (Attachment #5)
11.1 Objective
To verify the controls and indicators operate as specified by manufacturer recommendations.
11.2 Procedure
11.2.1 Manually operate the control switches listed in the applicable attachment.
11.2.2 Visually verify and document the actual responses.
11.2.3 Document any discrepancies in Attachment #10, Deviation Reporting Log. If any
variances/deviations are present, complete the appropriate deviation/variance form.
11.3 Acceptance Criteria
All operator controls and indicators tested shall conform to the expected response.
DOC. #
OPERATIONAL QUALIFICATION PROTOCOL
Capsule Filling Machine
National Institute of Health, Building 10 Room 1n249
Bethesda MD 20892
Page 7 of 36
13.0 Test Function Number Seven: Alarms, Safeties and Interlocks (Attachment #7)
13.1 Objective
To verify and document the operation and functionality of the alarms, safeties and interlocks.
13.2 Procedure
13.2.1 Follow the procedures in the applicable attachments.
13.2.2 Visually verify and document the actual responses as pass or fail against the expected
results.
13.2.3 Document all operational parameters used during execution.
13.2.4 Document any discrepancies in Attachment #10, Deviation Reporting Log. If any
variances/deviations are present, complete the appropriate deviation/variance form.
13.3 Acceptance Criteria
All safeties and interlock requirements must conform to the expected responses.
14.1 Objective
To verify and document the operational sequence of the system.
14.2 Procedure
14.2.1 Execute the procedures in the order listed.
14.2.2 Observe the expected operations and write “Pass” if the results were as expected, and
“Fail” if the expected operation did not take place.
14.2.3 Document any discrepancies in Attachment #10, Deviation Reporting Log. If any
variances/deviations are present, complete the appropriate deviation/variance form.
14.3 Acceptance Criteria
The operating sequence occurs as expected.
15.0 Test Function Number Nine: Maximum / Minimum Values (Attachment #9)
15.1 Objective
To verify the maximum and minimum allowable values of changeable setpoints.
15.2 Procedure
15.2.1 Display the changeable setpoint on the applicable attachment and test maximum and
minimum values.
DOC. #
OPERATIONAL QUALIFICATION PROTOCOL
Capsule Filling Machine
National Institute of Health, Building 10 Room 1n249
Bethesda MD 20892
Page 8 of 36
15.2.2 Visually verify and document the actual results of the functions that appear on the
operator interface.
15.2.3 Document any discrepancies in Attachment #10, Deviation Reporting Log. If any
variances/deviations are present, complete the appropriate deviation/variance form.
15.3 Acceptance Criteria
The actual results of the changeable setpoint limits shall conform to the expected responses.
16.2 Procedure:
16.2.1 Document all deviations found during execution using the Deviation Report Form that
is with the Deviation Reporting Log. Make copies of the form for each logged
Deviation.
16.2.2 Deviations caused by Out of Specification (OOS) Conditions. If during the evaluation
of equipment parameters against specifications, the equipment fails to meet
predetermined acceptance criteria, a formal deviation report investigates the cause. In
these cases, the cause must be known prior to continuing the protocol execution since
a true failure in equipment performance is likely to affect the completion of the
protocol. The protocol deviation report determines 1) if the acceptance criteria are
correct and current with user requirements and manufacturer's ratings, or 2) if there is
an immediate and obvious explanation for the failure. If a true failure is documented,
then appropriate root cause analysis and corrective actions ensue and are agreed to
between the protocol executors and the department owners and QA. The protocol
deviation report documents the conclusion to this investigation and any appropriate
retesting and an impact assessment on the portions of the protocol executed prior to
the documented failure.
16.2.3 Completed Deviation Report Forms are submitted to QA for signature acceptance.
This section identifies the personnel conducting the Operational Qualification. Each person who will be
executing this protocol shall complete an entry on the page.
Department/
Print Name Title Signature Initials Date
Company Name
Comments: ________________________________________________________________________
__________________________________________________________________________________
__________________________________________________________________________________
__________________________________________________________________________________
Acceptance Criteria: Signatures from all personnel involved in execution of this protocol shall be
present.
This attachment identifies the instruments, equipment, and materials that are not part of the equipment or
system being qualified, but are used in this qualification.
Yes/No
Yes/No
Yes/No
Yes/No
Yes/No
Yes/No
Yes/No
Yes/No
Comments: ______________________________________________________________________________
_______________________________________________________________________________________
_______________________________________________________________________________________
_______________________________________________________________________________________
Acceptance Criteria: All test equipment required for execution shall be calibrated with copies of calibration
records attached.
List the Standard Operating Procedures (SOP) that are required for execution of this protocol.
Comments: ________________________________________________________________________
__________________________________________________________________________________
__________________________________________________________________________________
__________________________________________________________________________________
Acceptance Criteria: A list of written procedures that are used to operate the equipment is provided above.
If procedures are in draft form, attach a copy to the executed protocol.
List the Standard Operating Procedures (SOP) used to train personnel for operating the equipment.
Training
SOP # Procedure Title Documented Initials/Date
(Y/N)
Operation of the Capsule Filling
Yes/No
Machine
Maintenance of the Capsule Filling
Yes/No
Machine
Comments: ________________________________________________________________________
__________________________________________________________________________________
__________________________________________________________________________________
__________________________________________________________________________________
Acceptance Criteria: A list of written procedures that are used to train personnel to operate the
equipment is provided above.
Press the control keys to verify that the equipment/system operations function as indicated. Note:
document any failing results.
machine normally.
Pushbutton normally. Red
Red
Machine Start Starts the machine ____________ the
Pass/Fail
Pushbutton normally machine normally
Indicates machine ___________ machine
Lamp Machine
operating. Lamp operating. Lamp Pass/Fail
Running
Green Green
Indicates no tablets ___________ no
LED No Tablets Pass/Fail
LED Red tablets LED Red
Indicates powder __________ powder
LED Powder Mixer
mixer is selected. mixer is selected. LED Pass/Fail
(MAN/AUTO) On
LED Green Green
Powder Mixer __________ either
Selects either Manual
MAN/AUTO Manual or Auto Pass/Fail
or Auto Control
Selector Control
LED Powder Mixer Indicates powder _________ powder
Pass/Fail
Enabled mixer on. LED Green mixer on. LED Green
Powder Loading Enables Powder ____________
Pass/Fail
Enable Selector Loading Powder Loading
______________ run
Hourly Production Logs run time when
time when machine is Pass/Fail
Counter machine is operating
operating
DOC. #
OPERATIONAL QUALIFICATION PROTOCOL
Capsule Filling Machine
National Institute of Health, Building 10 Room 1n249
Bethesda MD 20892
Page 17 of 36
______________ run
Logs run time when
Hour Counter time when machine is Pass/Fail
machine is operating.
operating.
_____________ /
Engage/Disengage Engages / disengages
_____________ Pass/Fail
Guards Selector guards
guards
Comments: ________________________________________________________________________
DOC. #
OPERATIONAL QUALIFICATION PROTOCOL
Capsule Filling Machine
National Institute of Health, Building 10 Room 1n249
Bethesda MD 20892
Page 18 of 36
__________________________________________________________________________________
__________________________________________________________________________________
__________________________________________________________________________________
Acceptance Criteria: The actual results are the same as the expected results or explained.
1. Manually fill hopper with capsules. No 1. Capsules flow into the machine
dosing units installed. Start is initiated until stop pushbutton is
pumps/aspirator. Initiate machine start. pressed.
2. Verify capsule minimum level operates
correctly by positioning sensor to simulate 2. LED M is lit.
Pass/Fail
no capsules.
3. Verify operation of the Sort/Unopened 3. Closed capsules will be ejected
capsule system. Manually insert 4 into the unopened capsule
unopened capsules in the feed channels. container.
Operate the machine
4. Adjust machine speed to 6000 Caps/hr 4. Verify machine is at minimum
feed of 6000 caps/hr
Powder Feed System
1. Load upper powder hopper ¾ full. Turn 1. Observe powder flow into lower
the manual feed selector on the control powder hopper via plexiglass fill
panel. tube
2. With tube full of powder calibrate
capacitive sensor per IMA 35064M04 2. Observe indicator light on
(Instruction manual) capacitive sensor turns on.
3. Utilize “Jog” to start and stop machine Pass/Fail
until product bowl is filled.. 3. Observe product bowl.
4. Place feed selector in “auto” and
observe powder coming out of dosators 4. Confirm proper compactness and
and a “tablet falls into capsule bottom. verify dosator piston projects .5mm
Adjust dosators in accordance with IMA from dosating unit.
35064M04 (Instruction manual).
5. Adjust powder layer in accordance with 5. Observe powder layer for proper
IMA 35064M04 (Instruction manual). height and homogeneity.
DOC. #
OPERATIONAL QUALIFICATION PROTOCOL
Capsule Filling Machine
National Institute of Health, Building 10 Room 1n249
Bethesda MD 20892
Page 24 of 36
Pellet System
Comments: _______________________________________________________________________
Acceptance Criteria: The actual results are the same as the expected results or explained.
CONTROL PANEL
EMERGENCY
STOP
LOW VOLTAGE
Fault LED B
DOC. #
OPERATIONAL QUALIFICATION PROTOCOL
Capsule Filling Machine
National Institute of Health, Building 10 Room 1n249
Bethesda MD 20892
Page 26 of 36
DC POWER
SUPPLY FAULT
LED B AND LED C
MACHINE
INVERTER FAULT
LED D
MOTOR
OVERLOAD
FAULT LED D
HANDWHELL
ENGAGED
ALARM LED E
LOW PRESSURE
ALARM LED F
POWDER MIXER
EXCEEDS LIMITS
ALARM LED L
INSUFFICIENT
CAPSULE LEVEL
ALARM LED M
DOC. #
OPERATIONAL QUALIFICATION PROTOCOL
Capsule Filling Machine
National Institute of Health, Building 10 Room 1n249
Bethesda MD 20892
Page 28 of 36
INSUFFICIENT
CAPSULE LEVEL
TIMEOUT ALARM
LED M
INSUFFICIENT
PELLET LEVEL
ALARM LED N
With no or low pellet level and Insufficient pellet level LED N flashing
Pass/Fail
machine running LED will flash. RED
INSUFFICIENT
PELLET LEVEL
TIMEOUT ALARM
LED N
INSUFFICIENT
TABLET LEVEL
ALARM LED S
With no or low tablet level and Insufficient tablet level LED S flashing
Pass/Fail
machine running LED will flash. RED
POWDER
LOADING
MANUAL ALARM
LED X
MAXIMUN JOG
RUNTIME ALARM
LED Z
GUARDS
DISABLED
ALARM LED G
GUARDS OPENED
ALARM LED G
TABLET UNIT
RUNNING
ALARM LED A
Comments: _______________________________________________________________________
Acceptance Criteria: The actual results are the same as the expected results or explained.
Setup equipment to run with the appropriate operational parameters for a typical Production Cycle. Run
equipment. Inspect cycle run data and verify it meets acceptance criteria.
Comments: ________________________________________________________________________
Acceptance Criteria: The actual results are the same as the expected results or explained.
Meets Acceptance Criteria? ( ) Yes ( ) No
Setup equipment to run with the appropriate operational parameters for a typical Production Cycle. Run
equipment. Inspect cycle run data and verify it meets acceptance criteria.
Comments: ________________________________________________________________________
Acceptance Criteria: The actual results are the same as the expected results or explained.
Change the system setpoints and verify the maximum and minimum values allowed. Note: document any
failing results.
Comments: ________________________________________________________________________
__________________________________________________________________________________
__________________________________________________________________________________
__________________________________________________________________________________
Acceptance Criteria: The actual results are the same as the expected results or explained.
Document all deviations found during execution. Note: Make additional copies of this attachment as
necessary.
DEVIATION LOG
Deviation
Description Resolved (Y/N) Initials & Date
Number
Yes / No
Yes / No
Yes / No
Comments: _______________________________________________________________________
_________________________________________________________________________________
_________________________________________________________________________________
Acceptance Criteria: A list of deviations resulting from the execution of this protocol shall be recorded.
Executor/Discovered By Step
Description of Deviation
Investigation / Findings
Statement of Impact
Resolution
Approvals
Role Name (Print) Signature Date
Executor
Reviewer
NIH QA Officer
DOC. #
OPERATIONAL QUALIFICATION PROTOCOL
Capsule Filling Machine
National Institute of Health, Building 10 Room 1n249
Bethesda MD 20892
Page 36 of 36
Qualification Process Solutions, LLC signature below verifies this Operational Qualification protocol has been
reviewed, for completeness and for accuracy of calculations. A separate final report will be written and
attached to this protocol. The final report will discuss any deviations, conclusions, and/or recommendations
resulting from the execution of the protocol.
Name (print):_____________________________________
Title: ____________________________________________