6.1.1 Mounting position
‘Mounting location
q
I
Preferably install the sensor in an ascending pipe, and ensure a sufficient distance to the
next pipe elbow:h > 2 DN
To prevent measuring errors arising from accumulation of gas bubbles in the measuring
tube, avoid the following mounting locations in the pipe:
‘Highest point of a pipeline.
« Directly upstream ofa free pipe outlet in a down pipe.
25xDN _ 22xDN7 Electrical connection
‘The measuring device does not have an internal circuit breaker.
» For this reason, assign the measuring device a switch or power-circuit breaker so that
the power supply line can be easily disconnected from the mains.
> Although the measuring device is equipped with a fuse, additional overcurrent
protection (maximum 10 A) should be integrated into the system installation.
7.1.3 Terminal assignment
‘Transmitter:
supply voltage, input/outputs
‘The terminal assignment of the inputs and outputs depends on the individual order
version of the device. The device-specific terminal assignment is documented on an
adhesive label in the terminal cover.
Supply voltage Input/output 1 Input/output 2 Input/output 3
1) 2m fo 26m [27m [ome fae foe [oe
Desvice-specific terminal assignment: adhesive label in terminal cover
7.2.1 Connecting the transmitter
Cable entry for supply voltage
2° Cable entry for signal transmission, mput/output 1 and 2
3 Cable enty for input/output signal transmission; Optional: connection of external WLAN antenna,
conection of remote display and operating module DKX0O or service plug‘Proline Promag # 300
Power consumpeion
ers are
Mas. 20W (ace power)
CN
1 ae. 400 mA (260)
1 ae 200 ma (120, 50760
0, 50/60)
7 Teale cp atic st aac eee
1 Depening ane dr ests, he conigurtn retain the device memarjerinthe
gobi ata memory (teAGM DA)