Professional Documents
Culture Documents
*Research supported by Science and Engineering Research Council
within the Agency for Science, Technology and Research through the II.CHALLENGES OF VTA
Biomedical Engineering Programme under Grant 132 148 0014).
Cailin Ng is with the NUS Graduate School of Engineering and
Integrative Sciences, National University of Singapore (e-mail:
A. TM Angle
cailin.ng@u.nus.edu). Another challenge of the current VTA design is that the
Wenyu Liang and Kok Kiong Tan are with Department of Electrical and human tympanic membrane is usually at an oblique angle.
Computer Engineering, National University of Singapore.
Chee Wee Gan and Hsueh Yee Lim are with Department of Although the success rate of the VTA on a mock membrane
Otolaryngology, National University of Singapore. is very high at 98%, it may not be indicative of the actual
B. TM Variability
The human TM is a thin cone-shaped membrane that
separates the external ear from the middle ear. It is made up
of three layers of tissue: the outer cutaneous layer, the fibrous
middle layer and a mucous inner membrane. It is held in
place by a thick cartilage ring. The TM is approximately
0.08mm thick, and 10mm in diameter. However, the TM
thickness and its strength and flexibility varies between
individuals. In [16], the thickness distribution of six human
TMs is taken using a high-resolution optical coherence
success rate when performed on humans. This is because the tomography setup. It was found that the thickness varies
strongly across each membrane and a great deal of inter-
specimen variability can be seen. Hence, the VTA needs to
be able to adapt to different TM thickness. The current VTA
design inserts the Tiny Tytan grommet by incising the TM
using a 0.76mm diameter needle-cutter, and pushing a 1.5mm
diameter Tiny Tytan grommet into the incision. Since the
diameter of the incision is much smaller than that of the
width of the grommet, a considerable force is needed to push
the grommet through the small incision for a successful
insertion. With the needle-cutter, the VTA might be able to
achieve a high success rate with the mock membrane,
however, if a thicker mock membrane is used, the success
rate drops, which is not acceptable.
III. METHOD
experiment setup in the laboratory utilized an imperfect TM
model [15]. The TM3.model
Figure is formed
VTA and its controlby slightly stretching a
system A. Toolset Design
mock membrane over the opening of an 8mm diameter Our objective was to redesign the VTA such that it can be
cylinder. The cylinder is shaped regularly, hence the mock used on a bigger range of TM angles and to increase its
membraneforms a vertical plane. During the experiments, success rate on different type and thickness of mock
the VTA approaches the mock
Figure 1. Working process membrane
of the VTA horizontally, membrane. Since the fundamental problem is that the incision
forming a near 90o approach with the mock membrane. In made by the needle-cutter is too small and the grommet slides
reality, the angular placement of the TM within the ear canal away from the incision if the TM is oblique, the idea of
ranges between 45o to 60o for adults [13], and an extremely cover-cutter was conceptualized. The needle-cutter is
oblique position for neonate and infants [14]. An experiment designed such that the needle-like cutter is contained within
was done to measure the success rate of VTA with needle- the hollow of the holder and passes through the bore of the
cutter on mock membrane at different angle. The results, as grommet when protruded, hence the diameter of the cutter is
shown in Fig. 4, are not satisfactory. The success rate of VTA limited to the diameter of the inner bore of the grommet,
with needle-cutter falls to 40% with 50 o mock membrane and which is 0.76mm. To overcome this limitation, the cover-
it is unable
Figure 6. toSuccess
insertrate
theofgrommet if themembrane
VTA on different angle isangles
lower than cutter is designed such that it is on the exterior of the holder
50o. This is because of a few reasons: and positioned so that its cutting blade is on the same side of
Figure 7.
Geometrical shape of the grommet not suited for 0 o z- the sliding motion. The cover-cutter is a cylinder that
direction insertion for very oblique TM, sheathes the holder, as shown in Fig 5a, and is used to make
an incision in the TM. The front portion of the cover-cutter is
Grommet sliding away from the incised hole because a partial cylinder with a sharp blade tip, which measures
of the angle of the oblique surface, and 1.6mm arc length. As shown in Fig 5c, the incision made by
High speed insertion causing the grommet to tilt the cover-cutter is a crescent-shaped slit that follows the
excessively on the oblique surface, hence it is not in contour of the round inner flange of the Tiny Tytan grommet.
the correct position for insertion. To support the grommet such that the grommet lies axially,
an inner holder with a conical tip as shown in Fig 5b
TM angle is an important factor as statistics have shown substitutes the needle-cutter. The holder acts as pusher to
that OME is more prevalent in toddlers aged 3 and below [4], insert the grommet into the incision. The conical tip allows
and for this age group, their TM is likely to be very oblique. the grommet to tilt and weave into the incision in any
To make VTA useful as a treatment tool for OME, it has to direction.
overcome the oblique TM issue.
As opposed to the z-direction insertion by the needle-
cutter method, the cover-cutter method makes use of the
oblique angle of the TM as well as the inner holder with a
conical tip to assist in the sliding of the grommet for Thus, a control system that can adjust the insertion depth
insertion. When an object is pushed against an oblique accordingly is needed to improve success rate and safety.
surface, it naturally slides in the direction of the slope. Since
To do this, we make use of the onboard force sensor in
the human eardrum is typically inclined at an angle of around
the VTA to get the force reading during insertion, and noticed
30o - 60o, the cover-cutter method makes use of this slope to
that there is a sharp drop in the force reading when the
create a sliding action to weave the grommet into the
grommet has been pushed into the incision, as shown in Fig.
incision. We apply a topical anesthetic gel (Lidocaine 2%) on
7. Therefore, it is possible to control the insertion process
the grommet to ease the pain as well as to aid the sliding and
such that once there is a sharp drop in the force reading
weaving-in motion of the grommet from the lubricating effect
indicating that the grommet has been successfully inserted,
of the anesthetic gel. Coupled with a crescent-shaped slit that
the VTA stops immediately and retracts. Hence, an adaptive
follows the contour of the flange of the grommet, it will be
insertion control algorithm is designed and implemented so
able to weave into the slit easily with minimal effort. The
cover-cutter method created three degree of freedom (DOF) that it can adapt to the different TM angles as shown in (1). δ
motion, even though the motor only had 1 DOF in the z- is set as 0.9 for a fast response. Since the tympanic cavity has
direction. With a wider cutting blade and a 3-directional a transverse diameter of 3-5mm [9], for safety purpose, the
motion of the grommet, the success rate of the procedure is maximum insertion depth, d max is set to 2.5mm for less
increased. TM angles, which was previously impossible with oblique and normal angle TM, and 3.5mm for very oblique
the needle-cutter, is now achievable with the cover-cutter. TM.
The working process of the VTA with cover-cutter design v=¿ (8)
is very similar to that of the needle-cutter and involves the
following five steps as illustrated in Fig 6. f max ( k ) =max (f ( k ) , f max ( k−1 ) ) (9)
Initialization: surgeon starts up the VTA program to where v = velocity of motor; vinsert = insertion velocity;
initialize VTA and controller. Surgeon then inserts the
distal end of the toolset into the ear canal of patient
0< δ ≤1; f = force and f max = max force;
and slowly approaches the TM until it is close but not C. Different cover-cutter setups according to TM angle
yet touching.
Touch detection: surgeon steps on the foot pedal to
start the VTA operation. The device’s motor will
actuate the toolset move forward slowly until the
force sensor detects a touch.
Myringotomy: an actuator punches out the cover-
cutter to make an incision on the TM. After an
incision is made, the cover-cutter is retracted.
Grommet insertion: the motor will move the toolset
forward for insertion. Because the TM is oblique, and
the inner holder has a slanted tip, the grommet tilts
slightly in the direction of the TM’s slope and slides
into the incision previously made by the cover-cutter.
Grommet release: upon sensing a successful
insertion by the force sensor, the motor will withdraw
the toolset. A buzzer will sound, signaling the end of
the procedure. The surgeon can then pull the toolset
out of the ear canal.
Figure 7. Force diagrams comparing (a) with motion control, and (b) without motion
Figure 6. Working process of VTA with cover-cutter control
With the cover-cutter toolset, the insertion is facilitated by From the table, it is clear that S1 setup has an advantage
the weaving-in slide of the grommet into the incision with an in terms of lower insertion force for less oblique angles and
oblique TM. However, the anatomy of TM is different S2 offers higher success rate for more oblique angle.
between individuals. Hence, it is possible for a patient whose However, the boundaries of when the surgeon should switch
TM falls outside of the normal TM range of angles (30 o – from S1 to normal or from normal to S2 is not clear, as there
60o). For a TM which is less oblique and forms a near 90 o isn’t a single point where both objectives of maximizing
with respect to the central axis of the ear canal, there will be success rate and minimizing insertion force is met. One
very little sliding action; and for TM which is very oblique possible way to solve this is to use multi-objective
such that the angle is less than 45o, the grommet will tilt too optimization. As there are two objectives here, the most
much until the sliding becomes unstable and the grommet common employed method, the weighted sum approach will
misses the incision hole. A possible solution is to have a be used, as shown below:
different toolset for three different cases: less oblique TM, k
normal TM and very oblique TM. The distal end of the
U =∑ wi F i(x ) (3)
holder where the grommet is placed is made to be angled at i=1
30o such that when the grommet is positioned on the holder, it
is tilted. Depending on the angle of the TM, the holder can be where F i is the objective function, w i is a weighting factor
rotated such that the angle that grommet makes with the TM for the i th objective function. This approach requires the
is always around the optimal angle, as shown in Fig 5c. To
further investigate the range of angles for the different setups, selection of the weights,w i, which are general gauges of the
mock membrane holders with degrees from 90 o to 30o were importance for each objective function, and it is up to the
printed using a 3D printer and fixed to a platform stage decision-maker’s preference. In this case, the success rate is
equipped with a force sensor. Similar to [6] and for the same viewed to be more important than the insertion force, hence
reasons, a membrane made of PE material was chosen for the the weight of the success rate, w s is set at 0.6, while the
mock membrane. For each angle, experiments were weight of the insertion force, w if is set at 0.4. Since the aim
performed on all three setups, normal holder (normal setup); is to maximize success rate and minimize insertion force, the
slanted holder at an opposite angle to the membrane (S1 cost function is obtained as shown below:
setup); slanted holder at the same angle as the membrane (S2
setup). The success rate and the median insertion force as Maximize U n ( θ )=w s ,n F s , n ( θ )−w f , n F f , n (θ) (4)
measured by the force sensor on the stage are shown in Table
1. where F s is a function of success rate and F f is a function of
insertion force. The insertion force is normalized for each
angle and the cost function is then applied to the three IV. RESULTS
different setups. After utilizing a curve fitting toolbox from
The comparison between the success rate and force on the
Matlab, it was found that the rational function fits the cost
membrane between needle-cutter and cover-cutter is
function with the lowest root mean square error. Hence, the
performed. The same experimental setup and the same
cost function of the three setups are as below.
working process with the motion control for both needle-
0.78 θ−31.05 cutter and cover-cutter was used so that only the merits of the
U s1 ( θ )= (5) mechanical design of the toolset is considered. Different
θ−16.66
membrane materials were also sourced for this experiment as
−0.0044 θ 2+1.03 θ−32.34 one of the challenge of VTA is TM variability. As explained
U N ( θ )= (6) earlier, the TM thickness and its strength and flexibility
θ−33.49 varies between individuals. Therefore, it is necessary to test
24.42 θ−573.3 the performance of the needle-cutter and cover-cutter on
U s2 ( θ )= 2 (7) mock membrane of different thickness and materials of
θ −64.22θ+2103 different tensile strength. Besides the mock membrane made
Figure 10. Peak force comparison for the four types of membranes
Fig. 8 shows the cost function of the three setups. Since of PE material that was used earlier, another three different
the aim is to maximize the cost function, the pareto front is materials were used, as shown in Table II. Mock membrane 1
chosen as the solution that maximizes U (θ), as shown in the is made of soft PVC material and is very thin. Mock
green highlighted line. Therefore, the Pareto optimal solution membrane 2 is made of PE material with normal thickness.
is to use S1 setup for 90 ° ≤θ<78 ° , normal setup for Mock membrane 3 is also made of PE material but is stronger
than Mock membrane 2. Mock membrane 4 is significantly
78 ° ≤ θ< 42° and S2 setup for 42 ° ≤θ< 30° . thicker than the others as it is made up of 2 layers, with the
first layer made of PE and the second layer made of latex the neonatal TM angle is near horizontal, but the cover-
adhesive. This wrap is made to be self-adhesive that is cutter’s success rate drops drastically for angle less than 40 o.
activated by compression, hence the second layer consist of a More work needs to be done to facilitate an even larger range
series of protrusions in a regular pattern. Mock membrane 4 of TM angles. Secondly, the cover-cutter design requires
was chosen although it is much thicker than the rest as the some knowledge of the TM angle. Hence, a device that is
human TM has an array of collagen fibers in the radial and able to capture and create a 3D image of the desired grommet
circumferential direction, and the protrusions on mock insertion site will be helpful to minimize judgement errors
membrane 4 bear a slight resemblance to it. The experiments that might result in lower success rate. The current 3D
were repeated for membrane angles 90o to 30o, and for the modeling techniques are quite advanced, but to be able to
four different membrane materials. capture a 3D image of an 8mm by 8mm TM might be
challenging. To add to the difficulty, the device has to go
The results as shown in Fig. 9 and 10 exhibit a vast
through a narrow and tortuous ear canal before it can be near
improvement in the success rate for oblique membrane angles
enough to the TM to get a good image, hence it has to be very
and across all different types of membrane. Angles that was
small.
previously unsuccessful with the needle-cutter, is now
possible using the cover-cutter. The success rate of the As with all handheld devices, the surgeon must feel
needle-cutter is also greatly affected by the thickness and comfortable to hold the device while performing the
material of the mock membrane, whereas it is less so for the procedure. Although the automated system offers some
cover-cutter. This might be because the needle-cutter requires advantages over the manual system, the introduction of
a larger force to push the grommet into a small incision. The sensing and motorized actuation elements in the automated
insertion depth required may also have exceeded the system is both costly and increases the weight and bulkiness
maximum insertion depth for safety purpose. The insertion of the hand-held device. Hence, device miniaturization is a
force of cover-cutter is lower across the board, with up to challenge that must be overcome.
78% improvement in force reading as compared to the
needle-cutter. However, due to the wider cutting edge of the
REFERENCES
cutter, the incision force of the cover-cutter is much higher
than that of the needle-cutter. Therefore, for a fair [1] Paradise JL, Rockette HE, Colborn DK, et al. “Otitis media in 2253
Pittsburgh area infants: prevalence and risk factors during the first two
comparison, the overall peak force, f peak is considered. The years of life.” Pediatrics 99, 318– 333 (1997)
average improvement in peak force is around 14%. [2] Ilechukwu, G.C., et al. “Otitis Media in Children: Review Article.”
Open Journal of Pediatrics 4, 47-53 (2014)
[3] Vaile L. Williamson T. Waddell A. Taylor G. “Interventions for ear
V. CONCLUSION discharge associated with grommets (ventilation tubes).” Cochrane
Database Syst Rev 2, CD001933 (2006)
The VTA with its motorized mechanical system and a [4] Auernger P et al. Trends in OME in America. Pediatrics 112, 514-520
sensing system offers an automated electromechanical (2003)
solution that is able to complete the procedure in a short time [5] DiMaggio C et al, “A retrospective cohort study of the association of
anesthesia and hernia repair surgery with behavioral and developmental
efficiently and precisely. To overcome its existing challenges,
disorders in young children.” J Neurosurg Anesthesiol 21, 286-91
the cover-cutter design with 3 different setups was (2009)
conceptualized. The insertion control of the VTA was also [6] K. K. Tan, W. Liang, S. Huang, L. P. Pham, H. Y. Lim and C. W. Gan.
improved by adding in adaptive control to improve success “Design of a surgical device for office-based myringotomy and
grommet insertion for patients with otitis media with effusion.” Journal
rate and to minimize force on the membrane. Experimental of Medical Devices, Vol. 8 / 031001 (2014)
results on mock membrane showed that the cover-cutter [7] Christopher J. Payne and Guang-Zhong Yang, “Hand-held medical
design has great improvement over needle-cutter design in robots” Annals of Biomedical Engineering, Vol 42, No. 8, August
terms of success rate and insertion force. Very oblique angles 2014, pp. 1594-1605
[8] Franklin L Rimell et al. “Single-pass tympanostomy tube insertion
that the needle-cutter failed is now possible with the cover- device for use in the uncooperative child” The Triological Society 2014
cutter. The cover-cutter also proved to be less susceptible to presentation at COSM
variability in the types of mock membrane as it maintained a [9] Franklin L Rimell et al. “Single stage tympanostomy tube insertion:
high success rate for the 4 different types of mock membrane design issues on a new surgical approach” American Academy of
Otolaryngology 2014 presentation
used, unlike the needle-cutter. This paper has shown that the [10] K Belani et al. “A single-pass tympanostomy tool – Further experience
mechanical design of the toolset plays a significant role in the in children” Poster presentation at Society for Pediatric Sedation
performance of the device. By making use of the natural 5/2015
sliding motion of the grommet on the oblique surface, the [11] Akio Shino, Tetsuo Ishii, Mikiko Takayama, “Measurement of the
strength of human tympanic membrane using a force-sensing
cover-cutter design has made it possible for VTA to myringotomy knife” Otol Jpn 7(2) 1997: 97-101
overcome its existing challenges, and moved one step closer [12] Michael Bergin et al. “Measuring the forces of middle ear surgery;
from a lab prototype towards clinical trials. evaluating a novel force-detection instrument” Otology & Neurotology
35: 77-c83 (2014)
[13] Jonathan P.Fay, Sunil Puria, Charles R. Steel, “The discordant
VI. FUTURE WORK eardrum” Proceedings of the National Academy of Sciences vol 103
19743-19748 (Dec, 2006)
The cover-cutter design offers huge advantages over the [14] Charles D. Bluestone, Jerome O. Klein, 2007. “Otitis Media in Infants
needle-cutter, but it is not without its shortcomings. Firstly, and Children”, 4th edition, BC Decker
[15] Cailin Ng, Wenyu Liang, Wenqiang Wu, Chee Wee Gan, Kok Kiong
Tan, "Pain mitigation approach in office-based procedure: case
study", 42nd Annual Conf of the IEEE, pp. 601-606, IECON 2016
[16] Van der Jeught et al, “Full-field thickness distribution of human
tympanic membrane obtained with optical coherence tomography,”
JARO, 14(4):483-494, 2013.