Spesifikasi ventilator E 360 T – Newport

Brand : Newport - Covidien Page 1 of 4


Type/Model : E 360 T
Made in : USA
Control and Features
Patient Selection : Pediatric/infant – Adult
Breath types/Modes : Volume Control (VC)
Pressure Control (PC)
Volume Target Pressure Control (VTPC)
Biphasic Pressure Release Ventilation (BPRV)
Assist/Control Mandatory Ventilation (A/C MV)
Synchronized Intermittent madatory ventilation (SIMV)
Spontaneus (SPONT)
Non-invasive Ventilation (All breath types/modes)
: Pressure Support (PS)
Spontaneous Breath Choices Volume Target Pressure Support (VTPS)
: All modes
Backup Ventilation : Automatic or Manual adjustment 1-19 for PC, VTPC, PS and VTPS
Slope/Rise Breaths
Pressure Support : 0 to 50 cmH2O/mbar
Pediatric/infant : 0 to 60 cmH2O/mbar
Adult : Automatic or Manual adjustment 5% to 55% peak flow for PS and VTPS
Expiratory Threshold
Tidal Volume : 5 to 1.000 ml
Pediatric/infant : 100 to 3.000 ml
Adult
Resp Rate : 1 to 150 b/min
(Frequency) : 1 to 80 b/min
Pediatric/infant
Adult : 1 to 100 L/min
: 1 to 180 L/min
Flow
: Square or descending ramp
Pediatric/infant
: Off, 0.1 - 2.0 seconds
Adult
: Delivers one sigh breath every 100 breath,
Flow Wave Pattern
sigh VT = 1.5 X VT setting
Pause
Sigh : 0 to 70 cmH2O/mbar
: 0 to 80 cmH2O/mbar
Pressure limit
Pediatric/infant : 0.1 to 3.0 seconds
Adult : 0.1 to 5.0 seconds
Inspiratory Time : Max inverse 4 : 1
Pediatric/infant
Adult : 0 to -5 cmH2O/mbar
I : E Ratio
Trigger (sensitivity) : 0.1 to 2.0 L/min
P-Pressure trigger : 0.1 to 2.0 L/min
Flow trigger : 0.21 to 1.00
Pediatric/infant
Adult : 0 to 30 cmH2O/mbar
: 0 to 45 cmH2O/mbar
FiO2 (oxygen concentration)
PEEP/CPAP (Pbase)
: 8 L/min max
Pediatric/infant
: 15 L/min
Adult
: 3 L/min
Leak Compensation (Automatic)
Pediatric/infant
Adult
manual inflation
Manual inflation : 5 seconds max
O2 (3 minute) : Delivers 100% oxygen for 3 minutes
Ideal Weight : 1 -375 Kg
Weight units : Kg or Lb
Volume units : mL or
mL/Kg Maneuver (tools for assessing lung dynamics)
PV Maneuver : Quasi-static pressure volume curve
NIF (MIP) : Maximum occlusion pressure
P0.1 : 100 m seconds occlusion pressure
Insp. Hold : 15 seconds max
Exp. Hold : 20 seconds max
Ventilation Standby : At power up, allows setting to be preset and circuit check
test to be perfomed prior to starting ventilation
Open Exhalation Valve : Off/on for Biphasic Pressure Release Ventilation
(BPRV) Volume Target : Off/on for Volume Target Pressure Control (VTPC)
Event History log : Records 1.000 event, alarm and setting, color coded
Quick Setup : Esstablishes new ventilation settings based on an
entered ideal body weight, patient type and mode
Circuit Check : Automatically test for leaks, compliance an resistance
RS-232 comm protocol : Communication protocol selection for remote
monitoring Save and Download : Allows saved screen images and event history file to
be
downloaded to a USB flash memory drive
Display Brightness : Adjustable display backlight
Calibrate Sensor : Exhalation flow and oxygen sensors
Date/Time : Adjust and format
Language Selection : For display message and screen text
Pressure units : cmH2O/mbar
Circuit Type Compensation : Heated exp. Limb, heated insp. Limb, HME or test
lung Altitude Compensation : 0-4.000 m/13.124 ft (200 m/656 ft increments)
Compliance Compensation : On/Off (volume control)
Monitored Parameters
Ppeak, Cdyn effective, Vte % variance, Exp. Flow
Pplat, Cstat, MVI, I:E Ratio, RSBI, RR spont
Pmean, RI, MVE spont, Time Constant, FiO2
Total PEEP, VTE Insp flow, Rrtot, RE
P0.1, NIF
Graphics
Waves : Pressure/time
Volume/time
Flow/time
Loops : Volume/Pressure
Flow/Volume
Trend Screens
VTE % Var/time, Ppeak/time
Rrtot/time, PEEP/time
MVE/time, RSBI/time
VTE/time, Pmean/time
Audible and Visual Alarms
Adjustable Alarm (via GUI) : Low MVE, High Rrtot
HighMVE, Apnea delay time
Low Paw, Disconnect (treshold %)
High Paw
Automatics Alarms : (Setting )out of range, Low and High FiO2
Pressure limit below Pbase, Low Paw below Pbase
Sustained high baseline pressure, Insp.time too
short I:E ratio invers violation, insp.time too long
Low and high baseline (PEEP) pressure
Alarm feature
Alarm silence : Mutes audible alarm for 120 seconds
Backup Vent : Backup ventilation supplied in response to low MVE alarm
O2 Sensor : O2 sensor error/O2 sensor disconnect
Flow sensor : Flow sensor error
Gas supply alarms : Loss of one gas supply/loss of both gas supplies
Power Down Alarms : Audible only
Power fail alarms : Loss of AC power/low internal battery
Device Alert : Indicator lights and messages are display
Check Vent Fan : Cooling fan failure
Suction Disconnect Function : Pre-silence alarms for 120 seconds, suspend ventilation
after a planed disconnect an senses reconnection to
resume ventilation
Alarm Reset : Clears visual indicator and messages
Alarm Tone : 3 tones
Display : Dual-view touchscreen (15" & 6,2")
Dimension
Width : 13.9 in. (35.3 cm)
Depth : 13.9 in. (35.3 cm)
Height : 12.1 in, (30.7 cm)
Weight : 38 lbs (17.3 Kg)
Environmental Operating
Relative humidity : 10% to 95% Rh non-condensing
Altitude : 0 to 13.124 feet (0 to 4.000 meters)
Pressure : 21 to 31 in.Hg (700 to 1.060 hPa)
Storage
Ambient temperature : -20 to 60 °C (-68 to 140 °F)
Relative humidity : 10% to 95% Rh non-condensing
Altitude : 0 to 18.000 feet ( 0 to 5.500 meters)
Pressure : 15 to31 in.Hg ( 500 to 1.060 hPa)
Power Requirement : 100 - 240 VAC, 250 VA max, 50/60 Hz (+/-10%)
2A for 125 VAC, 1A for 250 VAC
Internal Battery : Provides an average of 60 minutes of complete ventilator
function when new and fully charge
Gas supply requirement : Air and O2 supply inlet pressure 30 to 90 psig, 50 psig nominal
External Connection
Remote Alarm : 1/4 in. Jack; normally open for nurse call or remote system
RS-232 : Nine-pin D-shell, 38.400 baud; for use with central monitoring
system
External Alarm silence : external alarm silence
External Battery : Three-pin DIN input for external power, 10 VDC to 14 VDC
VGA : Output for external display monitor
USB : Output for connecting a data storage device
COM 1 : Input/output for connecting external touch-screen graphical user
interface
Patient Circuit Connection : Inspiratory and expiratory port connectors: 22 mm OD
Standard accessories include : Two exhalation flow sensors
Support arm and breathing circuit
hanger Support arm rail block
Air and oxygen hoses, 10 ft (3 m)
Two disposable bacteria filters (for patient ports)
Operating manual
Compressor
Humidifier
Breathing circuit (infant and adult)
Trolley
Test lung (infant and adult)
Description : MORTARA Electrocardiograph (EKG)
Brand : Mortara - USA
Model / Type : ELI 380

Feature and Specifications


Instrument Type : Multi-lead resting electrocardiograph
Input Channels : Simultaneous acquisition of all 12 or 15 leads
Standard Leads : Acquired I, II, III, aVR, aVL, aVF, V1, V2, V3, V4, V5, V6

12 Lead Alternate Lead Groups : Up to three additional groups can be labeled using any combination of 6 precordial
leads
with V1, V2, V3, V4, V5, V6, V7, V8, V9, V3R, V4R, V5R, V6R, and V7R lead labels
Three lead groups are available and can be labeled using any combination of 3 leads

15 Lead Groups with


V1, V2, V3, V4, V5, V6, V7, V8, V9, V3R, V4R, V5R, V6R, and V7R lead label
:
ERGO: Backlit, 17” color touchscreen display, 180° rotation and 120° tilt Standard:
Waveform Display : Backlit, 17” color LCD with 1280 x 1024 resolution, 120° tilt
Input Impedance
Input Dynamic Range
Electrode Offset Tolerance : Meets or exceeds the requirements of ANSI/AAMI/IEC 60601-2-25
Common Mode Rejection

Patient Leakage Current


Chassis Leakage Current : Meets or exceeds the requirements of ANSI/AAMI ES 60601-1

40,000 samples/s/channel used for pacemaker spike detection

;
Digital Sampling Rate
1,000 samples/s/channel used for recording and analysis
: Best 10 automatic capture of the 10 seconds of data with the least amount of
noise from the last 5 minutes of full disclosure; Alternate lead placement
selection with default for pediatric, right-sided, posterior and any combination
Standout Clinical Features of user-defined precordial lead labeling

:
Smart (210 x 280 mm), perforated Z-fold thermal cued paper with full grid; 250
Paper : sheets stored in paper tray
Thermal Printer : Computer-controlled dot array; 1 dot/ms horizontal, 8 dots/mm vertical
Thermal Printer Speeds : 5, 10, 25 or 50 mm/s
Gain Settings : 5, 10 or 20 mm/mV
Report Print Formats : Standard or Cabrera: 3+1, 3+3, 6, 6+6 or 12
channel Rhythm Print Formats : 3, 6, 8 or 12 channel with configurable lead
groups
Glass keyboard with alphanumeric keys, soft-key menu, dedicated function keys, and
Keyboard
: touchpad pointing device
Frequency Response : 0.05 to 300 Hz
High-performance baseline filter; AC interference filter 50/60 Hz; low-pass filters 40
Hz,
Filters
: 150 Hz, or 300 Hz
A/D Conversion : 20 bits (1.17 microvolt LSB)
Device Classification : Class I, Type CF defibrillation-proof applied parts
ECG Storage : Internal storage up to 500 ECGs
AM12M Compatibility : For use with Quickels Electrode Suction Pump Decapus III or
greater Weight : ERGO: 27 lbs (12 kg) including one battery (without paper)
Dimensions : 15.5 (width) x 20 (depth) x 5.5 in (height) (39 x 51 x 14 cm)

Power Requirements : Universal AC power supply (100-240 VAC at 50/60 Hz) ; internal rechargeable
lithium-ion
battery with support for a second optional battery