Technical Description
Ident. Nr. 0303 7606 E
PROVOTEC GmbH
In den Dorfwiesen 14 | 71720 Oberstenfeld
Telefon: +49 (0) 7062 – 92 55 0
Fax: +49 (0) 7062 – 22 68 5
e-Mail: info@provotec.de
Internet: www.provotec.de
Content
page
Content ....................................................................................................................... i
NOTE ......................................................................................................................... a
Document Effectivity................................................................................................ a
Mechanical - Electrical Warning.............................................................................. b
Radiation Warning.................................................................................................... b
To the User................................................................................................................ b
Improvement Recommendations ............................................................................ b
1 Equipment Description .................................................................................... 1
1.1 General description.................................................................................................... 1
1.2 Intended Purpose ...................................................................................................... 1
1.3 Features .................................................................................................................... 1
1.4 Specifications ............................................................................................................ 2
1.4.1 Model versions and physical characteristics....................................................... 2
1.4.2 Attenuation Equivalent of the table top ............................................................... 3
1.4.3 Equipment Classification .................................................................................... 3
1.4.4 Power requirements ........................................................................................... 3
1.4.5 Total weight........................................................................................................ 3
1.4.6 Transport and stock environment ....................................................................... 3
1.4.7 Operating environment....................................................................................... 3
1.4.8 Nameplate.......................................................................................................... 3
1.4.9 Standards .......................................................................................................... 4
1.5 Optional Equipment ................................................................................................... 4
1.6 EC Conformity ........................................................................................................... 4
1.7 Equipment components ............................................................................................. 5
1.8 Disposal..................................................................................................................... 5
2 Installation ........................................................................................................ 6
2.1 Unpacking Functional Check ..................................................................................... 6
3 Functional description..................................................................................... 7
3.1 Control Elements and Indicators ................................................................................ 7
3.2 Brakes ....................................................................................................................... 7
3.3 Brake for Table top .................................................................................................... 7
3.4 Hand grips (optional).................................................................................................. 7
3.5 Fixing mechanism (optional) ...................................................................................... 7
3.6 Deflector (optional) .................................................................................................... 8
3.7 Side edge guard (optional)......................................................................................... 8
3.8 Bumper profile (optional)............................................................................................ 8
3.9 Protection foil (optional) ............................................................................................. 8
3.10 Center stop (optional - not shown -) ........................................................................... 8
3.11 Emergency Stop Switch and Power On indicator ....................................................... 8
3.11.1 Status of indicator lamp...................................................................................... 9
3.11.2 Acoustic status message.................................................................................... 9
3.11.3 Acoustic and indicator lamp message if drive blocked.......................................10
3.12 Hand Control ............................................................................................................11
3.13 Table top UP and DOWN movement ........................................................................12
3.14 Column synchronization............................................................................................12
3.15 Adjustment and storage of preferences table heights................................................12
3.16 Reset error and error log...........................................................................................12
3.17 Labels on head and foot end of table top ..................................................................13
4 Replacement of Components........................................................................ 14
4.1 Introduction...............................................................................................................14
4.2 Necessary tools and equipment................................................................................14
4.3 Replacement of rollers and ball bearings for floating table top ..................................15
01.12.2010 PROVOTEC GmbH, In den Dorfwiesen 14, 71720 Oberstenfeld, Germany i
0303 7606 E Prognost XPE, Technical Description
4.3.1 Rollers and ball bearings for longitudinal movement of table top .......................15
4.3.2 Rollers and ball bearings for transverse movement of table top ........................16
4.4 Table top brakes .......................................................................................................17
4.4.1 Table top brake removal and adjustment ..........................................................17
4.4.2 Replacement of brake bowden cable, pressure spring or brake bolts................18
4.4.3 Table top-Brake release pedal bar ....................................................................19
4.5 Elevating column ......................................................................................................20
4.6 Wheels .....................................................................................................................23
4.6.1 Brake wheels ....................................................................................................23
4.7 Rear side wheels ......................................................................................................23
4.8 Emergency Stop Switch and Power-On indicator......................................................24
4.9 Hand Control ............................................................................................................24
4.10 Power supply cord ....................................................................................................24
4.10.1 Plug of power supply cord .................................................................................24
4.10.2 Fuse..................................................................................................................25
4.11 Replacement of fixing mechanism ............................................................................25
4.11.1 Mounting of fixing mechanism position-disk ......................................................26
4.12 Replace of control electronic components.................................................................27
4.12.1 Replace drive controller.....................................................................................28
4.12.2 Replace drive power electronic .........................................................................28
4.12.3 Setting of drive power electronic line voltage input ............................................28
5 Maintenance ................................................................................................... 29
5.1 Introduction...............................................................................................................29
5.2 Safety Information.....................................................................................................29
5.3 Technical Safety Information.....................................................................................29
5.4 Maintenance schedule ..............................................................................................29
5.4.1 General .............................................................................................................29
5.4.1.1 Cleaning........................................................................................................30
5.4.1.2 Disinfection ...................................................................................................30
5.4.2 User’s daily maintenance before and during operation......................................30
5.4.3 Monthly Checks ................................................................................................30
5.4.3.1 Quality Control ..............................................................................................31
5.4.4 Technical safety checks and maintenance ........................................................31
6 Combination with other equipment .............................................................. 31
7 Description of Symbols, Labels and Abbreviations.................................... 32
7.1 Symbols....................................................................................................................32
7.2 Labels.......................................................................................................................33
7.3 Abbreviations ............................................................................................................34
8 Maintenance check list .................................................................................. 35
Appendix
Spare parts list........................................................................................ 0303 7606 P
Schematics.............................................................................................. 0303 7804 B
NOTE
The information contained in this document conforms to the
configuration of the equipment as of the date of manufacture.
Revisions to the equipment subsequent to the date of manufacture
will be addressed in service updates distributed to the PROVOTEC
Technical Service Organization.
Document Effectivity
Revision No. Date List of effective Comments
pages
18/06/2007 Original Issue
Radiation Warning
The component of the equipment described within this Document is part of
a system for the intended generation of X-rays for medical diagnosis.
X-rays generate a potential risk for both patients and operators.
For this reason the application of X-rays for a given medical purpose must
aim at the minimization of radiation exposition to any persons.
Those persons responsible for the application must have the specific
knowledge according to legal requirements and regulations and must
establish safe exposure procedures for these kinds of systems. Those
persons responsible for the planning and installation of this equipment
must observe the national regulations.
To the User
The user of this Document is directed to read and carefully review the
instructions, warnings and cautions contained herein prior to beginning
operation, installation or service activities. While you may have previously
operated equipment similar to that described in this Document, changes in
design, manufacture or procedure may have occurred which significantly
affect the present operation.
The installation and service of equipment described herein is to be
performed by authorized, qualified Provotec GmbH Customers.
Assemblers and other Customers not employed by nor directly affiliated
with Provotec GmbH technical services are directed to contact the local
Provotec GmbH office before attempting installation or service
procedures.
Improvement Recommendations
Users of this Document are encouraged to report errors, omissions and improvement
recommendations to PROVOTEC GmbH
1 Equipment Description
1.1 General description
The Prognost XPE is a mobile, elevating table patient table with floating table top for
general radiography in diagnostic human medicine use. It must only be used in
medically rooms.
Pre-positioning of the table will be performed through wheels. 2 brake wheels are
provided at the front side, which grants to fix position of the table.
The floating, flat table top of the table is locked in place in the longitudinal and
transverse directions by highly effective, foot-operated mechanical brakes.
Movement in both directions can be easily released by means of a conveniently
placed brake release pedal bar. The easy floating movement of the table top and its
wide range of displacement ensure convenient patient positioning directly above the
corresponding image receptor.
The table top height can be adjusted by a motorized elevation system from 575 to
865 mm table top height. 3 different working heights can be stored and memorized
for repeated table heights examination positions.
All control elements of the table are easily accessible from the front.
1.2 Intended Purpose
The Prognost XPE is a patient table for examinations with X-ray units (e.g. L/U-
arms) for general radiography in diagnostic human medicine use. It must only be
used in medically rooms.
1.3 Features
• Wide range of applications
• Motor driven elevation table for heights from 575 up to 840 mm with smooth
start and stop and convenient patient access
• Automatic stop at pre adjusted working height
• Short installation time due to completely assembled delivery
• High reliability
• Floating, flat table top of carbon fiber combines high radio transparency with
high rigidity
• Table top color is pure white, optional blue-violet
• Long hand grip at the rear side of the table top (optional)
• Short hand grips at front side of table top (optional)
• Large range of table top displacement for convenient patient positioning
• Short table top to image receptor distance
• Control elements are arranged for easy access
01.12.2010 PROVOTEC GmbH, In den Dorfwiesen 14, 71720 Oberstenfeld, Germany 1of 37
0303 7606 E Prognost XPE, Technical Description
1.4 Specifications
1.4.1 Model versions and physical characteristics
01.12.2010 PROVOTEC GmbH, In den Dorfwiesen 14, 71720 Oberstenfeld, Germany 3of 37
0303 7606 E Prognost XPE, Technical Description
1.4.9 Standards
DIN EN 60601-1 (1996-03) Medical electrical equipment,
general requirements for safety
DIN EN 60601-1-2 (2002) Medical electrical equipment,
general requirements for safety
Collateral Standard:
Electromagnetic compatibility – Requirements and tests
DIN EN 60601-1-3 (1995-03) Medical electrical equipment,
general requirements for safety;
Collateral Standard: general requirements
for radiation protection in diagnostic X-ray equipment
DIN EN 60601-1-6 (2005-06) Medical electrical equipment,
general requirements for safety;
Collateral standard: Usability
DIN EN 60601-2-32 (1995-11) Medical electrical equipment, Part 2:
Particular requirements for the safety of
associated equipment of X-ray equipment
1.6 EC Conformity
This unit is in conformance with the basic requirements of the MDD 93/42
ECC for medical products according article 11 and Annex VII.
PROVOTEC GmbH
In den Dorfwiesen 14 | 71720 Oberstenfeld
Telefon: +49 (0) 7062 – 92 55 0
Fax: +49 (0) 7062 – 22 68 5
e-Mail: info@provotec.de
Internet: www.provotec.de
13
10
8
12
10 5
6
9
4
14
1
2
1.8 Disposal
The Prognost XPE does not contain any toxicological components.
It consists mainly out of mechanical parts, some plastic pieces and a few electronic
components. For disposal the local and national regulations have to be kept.
In case of doubt contact PROVOTEC.
01.12.2010 PROVOTEC GmbH, In den Dorfwiesen 14, 71720 Oberstenfeld, Germany 5of 37
0303 7606 E Prognost XPE, Technical Description
2 Installation
2.1 Unpacking Functional Check
• Remove all shipping material.
• The table is shipped in the lowest table position.
• The table is shipped with a German standard type plug. If the plug does not
comply with your local standards it has to be replaced against an appropriate type
(refer to chapter 3.5).
• Insert the plug of the line cable into an appropriate power supply line.
The three color indicator lamp next to the STOP switch (Fig. 1-1, item 4) indicates
that line voltage is present. If it does not light up, make sure that STOP switch at
the front panel is released. To release turn CW
• Take Hand control and move table up for about 15 cm.
• Synchronize the elevating of the columns according to chapter 3.14
• Check function of the STOP switch. Depress switch, green lamp extinguishes.
Check also if elevation stops if STOP switch is activated during movement.
• Check that elevation stops at upper and lower end positions.
3 Functional description
3.1 Control Elements and Indicators
3.2 Brakes
In order to fix the patient table in all directions, it is fitted with 2 brake wheels at the
front, which have to be fixed with a fastener (Fig. 1-1/1) on the left or right side before
a patient gets on or off the table top.
1 1
The Prognost XPE has to be positioned above the floor mounted position-disk (Fig.
2-1a/2), so that the springy-pivot (Fig. 2-1a/3) locks in place into the position-disk
(Fig. 2-1a/2) by pressing the fastener according to (Fig. 2-1a/1).
Pressing the fastener according to (Fig. 2-1b/1) the springy-pivot will loosen (Fig. 2-
1a).
01.12.2010 PROVOTEC GmbH, In den Dorfwiesen 14, 71720 Oberstenfeld, Germany 7of 37
0303 7606 E Prognost XPE, Technical Description
The three color indicator lamp (Fig. 1-1/9) is placed near the emergency stop switch.
It indicates that the table elevation controller is ready for use.
Attention
If the emergency stop switch is activated and the Power On indicator is off,
the table input power line connector can be still under line voltage. The
table is disconnected from power line only by disconnecting the power
cord from power line receptacle connector.
Further status messages are displayed with red blinking cycles of indicator lamp
Status 2 and 4 :
Call Provotec authorized, qualified technical service, if after reset of status
message by pressing the release button (Fig. 1-1/4) the failure status is
indicated during next moving again.
With all indicator lamp messages (see chapter 3.11.1) an acoustic beeper message
is generated.
table 3.3
01.12.2010 PROVOTEC GmbH, In den Dorfwiesen 14, 71720 Oberstenfeld, Germany 9of 37
0303 7606 E Prognost XPE, Technical Description
Attention!!
If during moving one drive is blocked (e.g. collision of table top ), stop motion
and remove blocked reason
If drive is blocked and reason not found (e.g. internal blocked drive), stop table
top elevating and call Provotec authorized, qualified technical service
If one drive is blocked, an optical indicator lamp message and acoustic warning
sound take place (see table 3.2 and 3.3)
If after reset of indicator message with release button (Fig. 3-1/6) the drive blocked
reason is not removed, the drive in function will move only in direction of the blocked
drive, independently if up or down (Fig. 3-1/ 1,2 or 7,8) is commanded. This will
prevent inclination of table top.
If the difference between both drives is more than 1cm (0.394inch), moving of
drives will be disabled.
The indicator lamp shows cyclic red flashing and the acoustic beeper warning
will become also cyclic peep
9
LED
5
4
1 3
7
8
Figure 3-1
Note:
To move the table top up/down, it is necessary to depress at first the motion-
release button (Fig. 3-1/6) before any other motion button (Fig. 3-1/ 1,2,3 or 5).
To stop motion it is necessary to release at first the motion button (Fig. 3-1/
1,2,3 or 5) before the motion-release button (Fig. 3-1/6).
If not used place hand control into holding device (Fig. 1-1/5).
01.12.2010 PROVOTEC GmbH, In den Dorfwiesen 14, 71720 Oberstenfeld, Germany 11of 37
0303 7606 E Prognost XPE, Technical Description
For synchronization depress pushbutton „S“(Fig. 3-1/4) three times and then
pushbutton (Fig.3-1/5) in conjunction with the motion-release button (Fig. 3-1/6), to
drive the table completely down until both columns stops. The columns elevation
controllers are synchronized now.
Attention:
Never make column synchronization with patient in position on table top!
For selection of a stored position, press key 1, 2 or 3 in conjunction with the motion-
release button (Fig. 3-1/6) until table top stops.
Attention!
Before deleting an error message, make sure that the error reason is found and
solved!
Reset of critical errors, e.g. Internal limit switch error, or column difference more than
1cm (0.394inch):
Depress pushbutton “S” (Fig. 3-1/4) five times and then the pushbutton (Fig. 3-1/9).
Attention
When the table top is completely moved forwards, backwards, to head or
foot end, the maximum permissible load on the outside edges are:
01.12.2010 PROVOTEC GmbH, In den Dorfwiesen 14, 71720 Oberstenfeld, Germany 13of 37
0303 7606 E Prognost XPE, Technical Description
4 Replacement of Components
4.1 Introduction
Before service, it is absolutely essential to thoroughly read this chapter.
ATTENTION:
The replacement of components may be performed only by PROVOTEC Service
personal or agents expressly authorized by us.
Installation work must never be performed with the line voltage connected.
REAR
HEAD
FOOT
FRONT
4.3 Replacement of rollers and ball bearings for floating table top
4.3.1 Rollers and ball bearings for longitudinal movement of table top
Figure 4-1
The floating table top moves longitudinally on four rollers (Fig. 4-2, item 3) in the
support rails of the table top, and is guided on its side by four ball bearings (Fig. 4-2,
item 2). All rollers and ball bearings are attached to the intermediate frame
(Fig. 4-2, item 1).
To replace longitudinal roller(s) or ball bearing(s), the table top (Fig. 4-1, item 1) has
to be removed. To remove the table top and replace longitudinal roller(s) or ball
bearing(s), perform the following:
• The table top must be removed toward the foot or head end from the table base.
Remove the two end-stops (Fig. 4-1, item 2) each at the foot or head end.
The table top is heavy and should be lifted by at least two persons.
Failure to comply may result in Customers injury or equipment damage.
• Release the table top-brake with the foot-pedal (Fig. 1-1, item 2) and remove the
table top by sliding it toward the foot or head end from the table base.
• As necessary, replace roller(s) or ball bearing(s).
• Installation of the table top is the reverse of this procedure.
1
5
2 2
3
3
6
4
Figure 4-2
01.12.2010 PROVOTEC GmbH, In den Dorfwiesen 14, 71720 Oberstenfeld, Germany 15of 37
0303 7606 E Prognost XPE, Technical Description
4.3.2 Rollers and ball bearings for transverse movement of table top
The floating table top moves transversely on four rollers (Fig. 4-4, item 3) in the
support rails of the intermediate frame (Fig. 4-2, item 1), and is guided on its side by
four ball bearings (Fig. 4-4, item 1+2). All rollers and ball bearings are attached to the
lower frame of the table (Fig. 4-4, item 4).
1
1
3
4
2
Figure 4-3
To replace transverse roller(s) or ball bearing(s), the table top with intermediate
frame has to be removed.
To remove the table top with intermediate frame and replace transverse roller(s) or
ball bearing(s), perform the following:
• Unscrew two nuts each (Fig. 4-3, item 2) from the inside of the head and foot end
cover plates (Fig. 4-3, item 1) and remove the head and foot end cover plates.
• Remove the bumper assembly (Fig. 4-3, item 3) from the head and foot end each
of the intermediate frame by removing 2 screws (Fig. 4-3, item 4).
1
Figure 4-4
• Release the brake and remove the table top with intermediate frame assembly by
sliding it toward the front from the table base.
• As necessary, replace roller(s) or ball bearing(s).
• Installation of the table top with intermediate frame assembly is the reverse of this
procedure.
Figure 4-5
• Remove the bumper assembly (Fig. 4-5, item 2) according to chapter 4.3.2.
• Release the brake and remove the table top with intermediate frame assembly by
sliding it toward the front from the table base (see chapter 4.3.2).
10mm
Figure 4-6
• Unscrew the 2 cylinder-head screws (Fig. 4-6, item 1) and take away the table
top-brake.
• Installation of table top brake is the reverse of this procedure.
• Pre-adjust brake unit according to Fig. 4-6.
• Adjust the brake according to chapter 4.4.2
01.12.2010 PROVOTEC GmbH, In den Dorfwiesen 14, 71720 Oberstenfeld, Germany 17of 37
0303 7606 E Prognost XPE, Technical Description
2
3
Figure 4-7
• Remove cover (Fig. 4-8, item 1), lift the foot pedal so that you can take out the
bowden cable out of the holder (Fig. 4-8, item 4).
• Unscrew nut (Fig. 4-8, item 3) and 2 screws (Fig. 4-7, item 3).
Pull out the brake bolt (Fig. 4-7, item 2) and unscrew bowden wire.
• Replace pressure spring (if broken or less spring tension) or brake bolt
(if brake pad is used up).
• Installation of bowden cable and table top brake is the reverse of this procedure.
1
2 3 4
Figure 4-8
• The brake bolt (Fig. 4-7, item 2) has be pre-adjusted according to Fig. 4-6 and
adjusted using the screw (Fig. 4-8, item 2) and the locknut (Fig. 4-8, item 3). If
these adjustments do not lead to the required brake force, the block (Fig. 3-8,
item 4) has to be adjusted, too.
NOTE
Adjust the brake bolts. When the brake release pedal bar is depressed,
the distance between the brake bolt and the table top panel must be at least
4 mm without load.
The braking force of approx. 220 N has to be achieved.
Use dynamometer for check.
18 of 37 PROVOTEC GmbH, In den Dorfwiesen 14, 71720 Oberstenfeld, Germany 01.12.2010
Prognost XPE, Technical Description 0303 7606 E
Figure 4-9
The brake release pedal bar (Fig. 4-9, item 4) is foot-actuated and releases the
brakes of the floating table top.
To replace the brake release pedal bar, perform the following:
• Remove the cover plate (Fig. 4-8, item 1) and the cover holder (Fig. 4-9, item
1).
• Loosen one screw each (Fig. 4-9, item 3) in the right and left brake block (Fig.
4-9, item 2).
• Pull the brake release pedal bar (Fig. 4-9, item 4) out of the brake blocks (Fig.
4-9, item 2) by pulling to the front.
• Installation of the brake release pedal bar is the reverse of this procedure.
01.12.2010 PROVOTEC GmbH, In den Dorfwiesen 14, 71720 Oberstenfeld, Germany 19of 37
0303 7606 E Prognost XPE, Technical Description
1
2
Figure 4-10a 4
On right side: remove the cover holder which takes the electronic (Fig. 4-10b, item 2)
and disconnect from controller the connection cables (Fig. 4-10b, item 3).
1
3
Figure 4-10b
Figure 4-10
• Remove the table top with intermediate frame according to chapter 4.3.1.
Figure 4-11
• Unscrew 4 countersunk screws (Fig. 4-11, item 1) and remove profile plate
(Fig. 4-11, item 2).
Figure 4-12
01.12.2010 PROVOTEC GmbH, In den Dorfwiesen 14, 71720 Oberstenfeld, Germany 21of 37
0303 7606 E Prognost XPE, Technical Description
4
3
Figure 4-13
• Lay table on the frontside. Make sure that it doesn’t fall down!
• Unscrew 4 screws (Fig. 4-13, item 1) from bottom profile plate (Fig. 4-13, item
3).
• Remove the profile plate with the elevating column.
• Remove elevating column by unscrewing 4 countersunk screws (Fig. 4-13,
item 4).
• Unplug cable from motor drive (Fig. 4-13, item 2)
• Installation of the elevating column is the reverse of this procedure.
ATTENTION:
Make sure that the columns move parallel!
Measure the distance between the 2 columns in upper and lower position.
Tolerance has to be smaller than 2mm.
Wrong adjustment causes high abrasion and can cause malfunction.
Synchronization has to be adjusted according to chapter 3.14!
4.6 Wheels
4.6.1 Brake wheels
The Prognost XPE has 2 wheels on the front side with a central locking mechanism
for both wheels. To replace the wheels perform the following procedure:
4
5
2
3
6
Figure 4-14
ATTENTION:
Remember the direction of the red marking (Fig. 4-14, item 1) on the wheel pivot.
The red marking of both front wheels must face to the foot end side of the table,
which is the right side with look from the front side.
• Install the new wheel. Use Omni fit for the thread to secure the screws.
• Insert connecting bar and add the cover plate
• Add brake lever (horizontal) and tighten screw (Fig, 4-14, item 2).
• Use the same procedure for the wheel at the other front side.
01.12.2010 PROVOTEC GmbH, In den Dorfwiesen 14, 71720 Oberstenfeld, Germany 23of 37
0303 7606 E Prognost XPE, Technical Description
Attention
Line voltage is still present on some parts of the table,
even when the emergency stop-switch is activated!
To disconnected the table completely from the power line,
remove the power line cable!
ATTENTION:
It is only allowed to use Power cord plugs with protected earth contacts!
4.10.2 Fuse
or
3 2
1
Figure 4-15
01.12.2010 PROVOTEC GmbH, In den Dorfwiesen 14, 71720 Oberstenfeld, Germany 25of 37
0303 7606 E Prognost XPE, Technical Description
Make sure that no electric or others installation were implemented in the floor
position
3
1
1
2
Figure 4-16
To open the controller box, remove the outer cover with a srewdriver (Fig. 4-10a).
7
Figure 4-17
6 5 4
3
Description to Fig. 4-17:
1: line filter 5: indicator LED cable
2: drive power electronic 6: internal power cord
3: drive controller 7: drive cables
4: internal hand control cable 8: line transformer
jumper JP1 line
voltage setting
Figure 4-18
01.12.2010 PROVOTEC GmbH, In den Dorfwiesen 14, 71720 Oberstenfeld, Germany 27of 37
0303 7606 E Prognost XPE, Technical Description
Make sure that power line voltage selector Jumper JP1 has correct settings
according to actual line power voltage at installation room of Prognost XPE
Figure 4-19 shows setting of Jumper JP1 for 230Vac / 50Hz line voltage
Figure 4-20 shows setting of Jumper JP1 for 115Vac / 60Hz line voltage
Attention!!
5 Maintenance
5.1 Introduction
This chapter provides a maintenance schedule to ensure proper and safe operation
after installation.
Before operating this X-ray unit, users have to make themselves acquaint
with all control elements and their functions.
The preventative maintenance procedures required to ensure the
operational integrity and safety of the equipment are listed in the following
paragraphs. It is the owner/user's responsibility to perform preventative
maintenance at the specified intervals or arrange for such service with an
authorized service representative
Maintenance has to be recorded.
Personal must make itself familiar with all warnings, placed on the device.
They are necessary for the safety and ensure the correct operation.
In the event of a malfunction, do not longer use the mobile table and notify
immediately PROVOTEC service or the expressly authorized service provider.
5.4.1.1 Cleaning
NOTE
Do not use water for cleaning. Water causes short circuits in the electrical portion
of the system, and corrosion in the mechanical components.
Do not use corrosives, solvents or abrasive cleaning materials.
Clean painted and plastic surfaces only with a cloth and common household
cleaners and wipe surfaces with a clean, dry, lint-free cloth.
5.4.1.2 Disinfection
All components, including accessories and connecting cables, may be disinfected
only by wiping with a disinfectant solution. Please contact PROVOTEC if detailed
information is needed concerning appropriate disinfection materials.
The equipment must be well covered with plastic sheets during room disinfection.
NOTE
For safety reasons, no spray disinfectant may be used.
Recommendations and instructions regarding the use of disinfectants and methods
of disinfection may be derived from the most recent rules and guidelines concerning
disinfection and explosion safety.
5.4.2 User’s daily maintenance before and during operation
• Check that brake wheels are locked if brake lever is turned into locking position
(Fig. 1-1/1).
• Check fixing and easy movement of table top by using the brake foot-pedal
(Fig. 1-1/2).
• Check if there are visible differences in columns heights for elevation of table
top.
• If yes perform synchronization according to 3.14.
• Check that the power supply cord shows no damages. In case of visible
damages do not insert the cable connector into the receptacle wall out let and
inform your service provider immediately.
Check of motion-release function
• Depress the motion-release button (Fig. 3-1/6) without pressing any other of
the motion buttons (Fig. 3-1/ 1,2,3 or 7,8).
No motion may start.
• Depress one of the motion buttons (Fig. 3-1/ 1,2 or 7,8) without pressing the
motion-release button (Fig. 3-1/6).
No motion may start.
If table height motion starts during this check, put the table out of order and
call service.
For schedule refer to attached yearly technical safety check and maintenance check
list.
Copy this list and perform checks and maintenance. Add necessary remarks and add
list to system documentation files.
• For combination with other equipment contact PROVOTEC for compatibility tests
and release.
01.12.2010 PROVOTEC GmbH, In den Dorfwiesen 14, 71720 Oberstenfeld, Germany 31of 37
0303 7606 E Prognost XPE, Technical Description
CE-marking
Protective earth
EMC interference
Position in longitudinal
Position in transversal
Labels brake
FDA Label release/activate
Identification of
actual line
voltage setting Warning label at controller
box (background yellow)
Near line fuse
01.12.2010 PROVOTEC GmbH, In den Dorfwiesen 14, 71720 Oberstenfeld, Germany 33of 37
0303 7606 E Prognost XPE, Technical Description
7.3 Abbreviations
mm millimeter
cm centimeter
kg kilogram
C
° Degree centigrade
hPa Hektopascal
DIN German Industrial Standard
EN European Norm
CE CE-marking
V Voltage
A Ampere
Hz Frequency
int.2min/18min Intermittent duty 2min ON, 18min Off
01.12.2010 PROVOTEC GmbH, In den Dorfwiesen 14, 71720 Oberstenfeld, Germany 35of 37
0303 7606 E Prognost XPE, Technical Description
* Die unter „Hinweise“ aufgeführten Fig. Nummern sind der “Technical Description” zu entnehmen.
* you will find the figures in “Technical Description “
Kunde Service-Techniker
customer Service-technican
Kunden-Nr. Unterschrift
Customer-Nr. Signature
Datum
Date
01.12.2010 PROVOTEC GmbH, In den Dorfwiesen 14, 71720 Oberstenfeld, Germany 37of 37
Prognost XPE
Prognost XPEL
PROVOTEC GmbH
In den Dorfwiesen 14 | 71720 Oberstenfeld
Telefon: +49 (0) 7062 – 92 55 0
Fax: +49 (0) 7062 – 22 68 5
e-Mail: info@provotec.de
Internet: www.provotec.de
03037636P Prognost XPE, Spare parts list
14
1
15
13 16
17
3
18
30 12
29
19
4
5 6 20
21
7
22
8 23
24
9
25
26
10 27
11 28
35
36
31
32
34
33 30
37 38
03010401 TUBE 4
10
21000905 HEXAGON NUT 4
03010301 AXLE 4
24
03010302 RETAINING RING 4
03010303 SPACER 8
34 03030104 POSITION-DISK 1
37 03033010 CPU 1
ORDERING: NOTE:
To order parts the following details are necessary: If a part cannot be found in the Parts List,
EXAMPLE:
DEVICE TYPE: PROGNOST XPE include the Catalogue No. of the unit, the
MODEL/ID: 0303 _ _ _ _ (refer to nameplate) serial No. and detailed descriptions or
SERIAL NUMBER: S303 _ _ _ _ (refer to nameplate) photos and part in question. Please order
PARTS LIST NO.: 03037636P common nuts, screws and washers if
IITEM NO.: 1
PART NO.: 03030110 required.
DESCRIPTION : COLLISION PROTECTION DEVICE
QUANTITY: 2 (as required)
EDITED: February 2007, revised 15/07/2008