MAINTENANCE CALIBRATION

The pH probe will exhibit indications when Use fresh buffer for calibration. Perform a
cleaning is required. The probe will require two-point calibration using buffer solutions in
longer stabilization time, measures erratic accordance with ASTM method D1293.
reading and, having difficulties to perform Schedule periodic calibration according to the
calibration. Clean the sensor when these key time frame for initial and subsequent
indications occur. It should be done calibration.
periodically throughout the operation for
optimum performance. 1. Rinse the sensor thoroughly with clean
water to avoid cross contamination.

2. Immerse the probe in the First Point (pH


Cleaning
7.00 buffer solution) and momentarily stir
with sensor for proper even contact.
Initial cleaning:
Rinse off impurities from the probe thoroughly 3. Choose the Second Point pH buffer. For
with clean water. Scrub the probe gently with best results, make sure the operating pH
soft toothbrush or cloth. Rinse off loose range in the process is as nearest and in
deposit from scrubbing. between pH 7.00 and the Second Point.

Inorganic scale deposit:


Immerse the probe’s tip in dilute hydrochloric Example 1
acid HCl 2% to 5% for 5 minutes. Repeat the
steps for initial cleaning. The pH range of the process:
pH 6.00 - 7.50
User Guide
Oils and fats: First Point : pH 7.00

pH/ORP
Immerse the probe with soap solution for 5 Second Point : pH 4.01
minutes. Rinse off the excess soap on sensor
before installation.
Example 2

The pH range of the process:


STORAGE pH 10.50 - 12.50

First Point : pH 7.00


Second Point : pH 10.01
 When the pH probe is not in operation,
clean the sensor thoroughly to remove
impurities. pH probe requires storage
with liquid solution.

 Use buffer solution of pH 4.01 or tap


water as storage liquid. Immerse the
probe in a container for short term
storage.

 Fill the storage cap with buffer solution of


pH 4.01 or tap water before screwing the Leadtec Instruments (Asia) Sdn Bhd (1220327-P)
cap on the sensor for long term storage.
B-8-810, Kelana Square Block B, 17, 100, Jalan Pusat Perniagaan 1,

 Position the sensor in an upright


Jalan SS7/26, Kelana Square, 47301 Pusat Perniagaan Raja Uda, 12300
Petaling Jaya, Selangor, Butterworth,Pulau Pinang,
Malaysia. Malaysia.
position.
Phone : (+60)3 - 7611 0701 Phone : (+60)4 - 3200 701
Fax : (+60)4 - 7611 0702 Fax : (+60)4 - 3233 327 www.leadtec-asia.com
PREPARATIONS FOR USE DIMENSIONS WIRING DIAGRAM
1. Inspect the completeness of the sensor that

C-Series
includes a storage cap with KCl solution.
The cap should remain intact with the
probe.

2. Unpack the sensor from the box and


inspect for any physical damage. Report
any defect(s) promptly to your supplier or
any Leadtec representative.

3. Remove and store the storage cap for


future use. Rinse off excessive remains
from the sensor and storage cap.

4. The pH probes works better after


conditioning it to the measurement liquid or SPECIFICATIONS
pH buffer. Immerse the probe into the
measurement liquid or pH buffer for about
Type of Junction Double Junction
30 minutes before calibrating.
E-Series
5. The conductive layer covering the pH cable Measuring Range 0.00 – 14.00 pH
must be removed in order to isolate the pH ±2000 mV
circuit connection from the Reference.
Otherwise, the sensor will display incorrect
Temperature Range -5oC to +105oC
readings.
-5oC to +140oC
REF

Resistance ≤200 MΩ
PH
Conductive layer not removed Reference Electrode Ag-AgCl

REF 3.8M KCl

PH Maximum Pressure 10 bars

Conductive layer removed INSTALLATION Materials Ryton; Teflon; Glass

1. Install the sensor with a minimum gradient 2. Remove and store the sensor cap for Thread 3/4” NPT
of 30° above from the horizontal line. future use.
This is important for accurate 3. Rinse the probe and proceed with Cable Type Low Noise Coaxial,
measurement over time. installation on fitting. Dual Shielded