The pH probe will exhibit indications when Use fresh buffer for calibration. Perform a
cleaning is required. The probe will require two-point calibration using buffer solutions in
longer stabilization time, measures erratic accordance with ASTM method D1293.
reading and, having difficulties to perform Schedule periodic calibration according to the
calibration. Clean the sensor when these key time frame for initial and subsequent
indications occur. It should be done calibration.
periodically throughout the operation for
optimum performance. 1. Rinse the sensor thoroughly with clean
water to avoid cross contamination.
pH/ORP
Immerse the probe with soap solution for 5 Second Point : pH 4.01
minutes. Rinse off the excess soap on sensor
before installation.
Example 2
C-Series
includes a storage cap with KCl solution.
The cap should remain intact with the
probe.
Resistance ≤200 MΩ
PH
Conductive layer not removed Reference Electrode Ag-AgCl
1. Install the sensor with a minimum gradient 2. Remove and store the sensor cap for Thread 3/4” NPT
of 30° above from the horizontal line. future use.
This is important for accurate 3. Rinse the probe and proceed with Cable Type Low Noise Coaxial,
measurement over time. installation on fitting. Dual Shielded