. Switch on the instrument by pressing ON/Off key and give warm up for 10minutes
Conductivity cell may be cleaned by using warm50% detergent solution.
Take the Potassium Chloride solution 1413 µs/cm
Platinum plates of the conductivity cell must be completely dipped in
supernatant solution while taking readings
Press the Calibration key to enter Conductivity calibration .
Dip the electrode in the standard conductivity solution say
Swirl gently and allow the meter reading to stabilize and wait till the display shows READY
Press enter key to conform the calibration .
Clean the electrode with distilled water and wipe with dry tissue paper .
The instrument is calibration and ready for sample measurement