Form 72 Recommended Scope of Accreditation: (For Testing Laboratories)
Form 72 Recommended Scope of Accreditation: (For Testing Laboratories)
1988
(ii) Fly-ash Fineness by-Blaine IS:1727-1967 100m2/kg to ±6.7 m2/kg @ 359
Air Permeability 800m2/kg m2/kg
Method
Soundness by- IS:1727-1967 0.01% to 5% ±0.012 % @ 0.07
Autoclave method %
Particles Retained IS:1727-1967 5% to 50% ±3.84 % @ 27.1%
on 45micron (wet
sieving)
(iii Concrete Compressive IS 516:1959 5 N/mm2 to 80 ±2.15N/mm2 @
) Strength of Cubes N/mm2 21.5 N/mm2
Concrete Core IS 516:1959 5 N/mm2 to 80 ±2.50N/mm2 @
N/mm2 28.7 N/mm2
(iv Fresh Slump Test IS:1199-1959 1 mm to 250 mm ±10.32 mm @ 110
Signature, Date & Name of Lab Signature, Date & Name of Signature, Date & Name of Lead
Representative Assessor (s) Assessor
FORM 72
RECOMMENDED SCOPE OF ACCREDITATION
(For Testing Laboratories)
) Concrete mm
(v) Burnt Clay Bricks Compressive IS 3495 (Part 3 N/mm2 to 60 ±8.8 N/mm2 @85.9
/Pulverized Fuel Strength 1):1992 N/mm2 N/mm2
Ash Lime Water absorption IS 3495 (Part 1 % to 30 % ±1.39% @14.5 %
Bricks/Burnt 2):1992
Efflorescence IS 3495 (Part Qualitative NA
Clay Fly Ash
3):1992
Bricks / Heavy Dimensions IS 1077:1992 L: 4000 mm to ± 10.89 mm @
Duty Burnt Clay Length IS 12894:2002 5000 mm 4624 mm
Buildings Bricks Width W: 2100 mm to
Height 2400 mm ± 10.17 mm
H: 1300 mm to @2198 mm
1600 mm ±9.53 mm @1410
mm
Bulk Density IS 2180:1988 0.1g/cm3 to 5g/cm3 0.101 g/cm3@2.66
(Heavy Duty g/cm3
Bricks)
Dimensions IS 2180:1988 30mm to 310mm ± 5.06 mm @188
(Heavy Duty mm
Bricks)
(vi) Precast Dimensions IS 15658:2006 ±3.87 mm@200
Concrete Length Annex-B mm
Block for Paving ± 3.50 mm@159
Width 40 mm to 300 mm mm
Thickness ± 2.14 mm @80
mm
Visual Inspection IS 15658:2006 Qualitative NA
Water Absorption IS 15658: 2006 1 % to 15 % ± 0.77% @5.1%
Annex-C
Compressive IS 15658: 2006 5 N/mm2 to 80 ±5.33 N/mm2
Strength Annex-D N/mm2 @50.5 N/mm2
Signature, Date & Name of Lab Signature, Date & Name of Signature, Date & Name of Lead
Representative Assessor (s) Assessor
FORM 72
RECOMMENDED SCOPE OF ACCREDITATION
(For Testing Laboratories)
Signature, Date & Name of Lab Signature, Date & Name of Signature, Date & Name of Lead
Representative Assessor (s) Assessor
FORM 72
RECOMMENDED SCOPE OF ACCREDITATION
(For Testing Laboratories)
IV)-1963 1500KN 18.34KN
Signature, Date & Name of Lab Signature, Date & Name of Signature, Date & Name of Lead
Representative Assessor (s) Assessor
FORM 72
RECOMMENDED SCOPE OF ACCREDITATION
(For Testing Laboratories)
Laboratory: Shreya Test House Date(s) of Visit:
(Laboratory Division of Shreya Realty Pvt. Ltd.)
Discipline: Mechanical Testing
S. No Product(s) / Material Specific tests Test Method / Range of Uncertainty of
of test performed Standard against Testing/ Measurement
which tests are Limits of ()UM at value
performed detection
Group I. Building Materials
(viii) Fine Aggregate Sieve analysis IS:2386 (Part I)-
0 to 100 % ±4.2 % @ 95.6 %
(Gradation) 1963
Total Deleterious IS:2386 (Part I&II)- 0.1 % to 20.0 ±0.44 % @ 2.68
Materials 1963 % %
Materials finer than
75 micron IS:2386 (Part I)- 0.1 % to 20.0 ±0.32 % @ 1.90
Clay Lumps 1963 % %
Coal & Lignite
IS:2386 (Part II)- 0 % to 5 % ±0.15 % @ 0.46
1963 0 % to 5 % %
IS:2386 (Part II)- ±0.12 % @ 0.32
1963 %
IS:2386 (Part III)-
Water absorption 0.1 % to 10 % ±0.16% @ 1.70%
1963
IS:2386 (Part III)-
Specific gravity 1.0 to 4.0 ±0.04 @ 2.60
1963
IS:2386 (Part III)- 1.0 kg/l to 4.0 ±0.038 kg/l @
Bulk density
1963 kg/l 1.73 kg/l
IS:2386 (Part III)- ±1.02 % @ 4.60
Bulking of Sand 5 % to 35 %
1963 %
IS:2386 (Part V)- ±0.26 % @ 2.28
Soundness 0.5% to 20 %
1963 %
(ix) Bitumen Products Specific gravity IS:1202-1978 0.9 to 2.0 ±0.028 @1.04
Softening point IS:1205-1978 30 °C to 200°C ±1.18°C @ 52.2
°C
Penetration IS:1203-1978 1-400 ±4.01 @ 66.80
Ductility IS:1208-1978 5 cm to 100 cm ±3.37 cm @
78.20 cm
Marshal Stability ASTM D 6927, 5 KN to 20 KN ±1.33 KN @
Value MS2:2018 13.66 KN
Marshal Flow ASTM D 6927, 2 mm to 4mm ±0.83 mm @ 4.9
MS2:2018 mm
(x) Bitumen mix material Bitumen content(% IRC-SP-11:1984 1 % to 10% ±0.11 % @ 5.29
by mass) %
Signature, Date & Name of Lab Signature, Date & Name of Signature, Date & Name of Lead
Representative Assessor (s) Assessor
FORM 72
RECOMMENDED SCOPE OF ACCREDITATION
(For Testing Laboratories)
(Laboratory Division of Shreya Realty Pvt. Ltd.)
Discipline: Mechanical Testing
S. No Product(s) / Specific tests Test Method / Range of Testing/ Uncertainty of
Material of test performed Standard against Limits of detection Measurement
which tests are ()UM at value
performed
Group II. Soil
(i) Soil Moisture
IS:2720 (Part-2)-
Content/Water 0.1 % to 30 % ±0.67% @ 12.6 %
1973
Content
MDD: 1.0g/cm3to ±0.040 g/cm3@
IS:2720 (Part-7)-
Light compaction 3.0 g/cm3 1.93g/cm3
1980
OMC: 1 % to 20% ±1.08 % @ 10.3 %
MDD: 1.0g/cm3to ±0.040 g/cm3@
IS:2720 (Part-8)-
Heavy compaction 3.0 g/cm3 1.93g/cm3
1983
OMC: 1 % to 20% ±1.08 % @ 10.3 %
IS:2720 (Part-5)-
Liquid Limit 20% to 70% ±1.31 % @ 24.5%
1985
IS:2720 (Part-5)-
Plastic Limit 10% to 30% ± 1.43 %@ 17.0%
1985
California Bearing IS:2720 (Part-16)-
1% to 80% ± 1.19% @ 9.7 %
Ratio (CBR) 1987
IS:2720 (Part-40)-
Free Swell Index 0 to 200 % ±4.02 % @ 28.0%
1977
Sand Content IS:6186 :1986 0.50 % to 15 % ±0.29 % @ 2.00 %
In Place Dry
Density of Soils by IS:2720 (Part 28) 1.0g/cm3to 3.0 ±0.031 g/cm3@
Sand Replacement -1974 g/cm3 2.144 g/cm3
Method
In Place Dry
IS:2720 (Part 29)- 1.0g/cm3to 3.0 ±0.030g/cm3@
Density of Soils by
1975 g/cm3 1.816 g/cm3
Core Cutter Method
Signature, Date & Name of Lab Signature, Date & Name of Signature, Date & Name of Lead
Representative Assessor (s) Assessor
FORM 72
RECOMMENDED SCOPE OF ACCREDITATION
(For Testing Laboratories)
Laboratory: Shreya Test House Date(s) of Visit:
(Laboratory Division of Shreya Realty Pvt. Ltd.)
Discipline: Chemical Testing
S. Product(s) / Specific tests Test Method / Range of Uncertainty of
No Material of test performed Standard against Testing/ Measurement
which tests are Limits of ()UM at value
performed detection
1.1 Atmospheric Pollution
(i) Ambient Air Respirable Suspended 3.61 µg/m3 @
IS: 5182 10 µg/m3-1000
Quality Particulate Matter (PM 108.78 µg/m3
(Part23):2017 µg/ m3
10)
6 µg/m3–500 0.648 µg/m
Sulphur Dioxide (SO2) IS-5182 (Part-2):2017
µg/m3 @18.81 µg/m3
Nitrogen Dioxide 6 µg/m3–750 2.225 µg/m3 @
IS: 5182 (P-6)-2017
(NO2) µg/m3 36.99 µg/m3
Particulate matter, SHR/SOP/AIR-01 10 µg/m3–500 2.141 µg/m3 @
PM2.5 (Dated: 01.04.2019) µg/m3 56.60 µg/m3
STH/SOP/AIR-
02/Indophenol ±4.328 µg/m3
Ammonia (NH3) 1-500 µg/m3
Method (Dated : @21.26 µg/m3
01.04.2019)
IS: 5182 (Part-9)- 20 µg/m3-1000 3.714 µg/m3 @
Ozone (O3)
2014 µg/m3 34.80 µg/m3
IS 5182 (Part 0.01 µg/m3 -10 0.080 µg/m3 @
Lead (as Pb)
22):2004 µg/m3 0.824 µg/m3
IS 5182 (Part 1 ng/m3-50 0.220 ng/m3 @
Nickel (as Ni)
22):2004 ng/m3 2.12 ng/m3
(ii) Noise Ambient Noise Levels IS:9989-2001 30 dB(A) to 2.12 dB(A)
100 dB(A) @72.02 dB(A)
(iii) Point Source Noise Source Noise Level IS 4758:2017 30 dB(A) to 3.53 dB(A) @ 75
Level ,Leq (DG Set) 130 dB(A) dB(A)
(iv) 0.28 mg/Nm3 @
IS:11255 (Part- 10 mg/Nm3-
Stack Emission Particulate matter (PM) 70.92 mg/Nm3
1):2009 2000 mg/Nm3
1.16 mg/Nm3@
IS:11255 (Part- 5 mg/Nm3-250
Sulphur Dioxide (SO2) 16.89 mg/Nm3
2):2003 mg/Nm3
3.08 mg/Nm3 @
Oxides of Nitrogen 5 mg/Nm3-500
IS:11255 (P-7):2012 113.77 mg/Nm3
(NOx as NO2) mg/Nm3
Carbon Monoxide 0.002 % @ 2.08
IS: 13270: 2003 2 % to 20%
(CO) %
0.007 % @ 8.72
Carbon Dioxide (CO2) IS: 13270: 2003 2 % to 20%
%
0.006 % @ 7.48
Oxygen (O2) IS:13270: 2003 2 % to 20 %
%
Signature, Date & Name of Lab Signature, Date & Name of Signature, Date & Name of Lead
Representative Assessor (s) Assessor
FORM 72
RECOMMENDED SCOPE OF ACCREDITATION
(For Testing Laboratories)
Laboratory: Shreya Test House Date(s) of Visit:
(Laboratory Division of Shreya Realty Pvt. Ltd.)
Discipline: Chemical Testing
S. No Product(s) / Material Specific tests Test Method / Range of Uncertainty of
of test performed Standard against Testing/ Measurement
which tests are Limits of ()UM at value
performed detection
1.1 Pollution & Environment
(i) Soil 0.7 g/cc – 2.5 0.050 g/cc @1.34
Bulk Density USDA Method:2014
g/cc g/cc
5 mg/kg - 500 1.50 mg/kg @75
Calcium STH/SOP/SOIL-1
mg/kg mg/kg
Cation Exchange IS 2720 (Part 1.0 meq/lit -25 0.830 meq/lit
Capacity (CEC) 24):2006 meq/lit @12.10 meq/lit
2.0 mg/kg - 1.0 mg/kg @45
Chloride STH/SOP/SOIL-2
300 mg/kg mg/kg
Colour STH/SOP/SOIL-3 Qualitative NA
Electrical 1 mS/cm-20 5.80 mS/cm
IS 14767: 2000
Conductivity mS/cm @322.6 mS/cm
5 mg/kg-250 0.245 mg/kg @
Magnesium STH/SOP/SOIL-4
mg/kg 28.02 mg/kg
0.298 %
Moisture IS 2720 (Part 2):2010 0.5 %-40 %
@5.776%
0.5 mg/kg-500 0.061 mg/kg
Nickel USEPA 3050 B: 1996
mg/kg @7.40 mg/kg
IS 2720 (Part 0.141 % @ 2.998
Organic Matter 0.2 % - 10 %
22):2010 %
pH IS: 2720 (P-26)-1987 1-14 0.080 @8.33
1 kg/hac – 100 1.33 kg/hac
Phosphorus STH/SOP/SOIL-5
kg/hac @39.07 kg/hac
1.0 kg/hac – 1.00 kg/hac @
Potassium STH/SOP/SOIL-6
200 kg/hac 65.0 kg/hac
1 mg/kg – 200 5.81 mg/kg
Sodium STH/SOP/SOIL-7
mg/kg @80.39 mg/kg
Sodium Absorption
STH/SOP/SOIL-8 Qualitative NA
Ratio (SAR) Value
Signature, Date & Name of Lab Signature, Date & Name of Signature, Date & Name of Lead
Representative Assessor (s) Assessor
FORM 72
RECOMMENDED SCOPE OF ACCREDITATION
(For Testing Laboratories)
Signature, Date & Name of Lab Signature, Date & Name of Signature, Date & Name of Lead
Representative Assessor (s) Assessor
FORM 72
RECOMMENDED SCOPE OF ACCREDITATION
(For Testing Laboratories)
Signature, Date & Name of Lab Signature, Date & Name of Signature, Date & Name of Lead
Representative Assessor (s) Assessor
FORM 72
RECOMMENDED SCOPE OF ACCREDITATION
(For Testing Laboratories)
Laboratory: Shreya Test House Date(s) of Visit:
(Laboratory Division of Shreya Realty Pvt. Ltd.)
Discipline: Chemical Testing
S. No Product(s) / Specific tests Test Method / Standard Range of Uncertainty of
Material of performed against which tests are Testing/ Measurement
test performed Limits of ()UM at value
detection
1.1 Water
(i) Water for pH IS:3025 (Part-11) :1996 2 - 12 0.04 @ 8.12
Construction Neutralization value
IS: 3025 (Part-23):2003 0.1 - 50 ml 0.60 ml @22.4 ml
Purpose with 0.02N H2SO4
Neutralization value
IS: 3025 (Part-22):2009 0.1 - 10 ml 0.20 ml @ 4.5 ml
with 0.02N NaOH
Chloride (as Cl) IS:3025 (Part-32) :2003 5 - 2000 mg/l 18.2 mg/l @1612 mg/l
Sulphate (as SO4) IS: 3025 (Part-24):2003 1 - 400 mg/ l 3.501 mg/l @126 mg/l
Organic Solids IS: 3025 (Part-18):2012 5 - 200 mg/l 4.20 mg/l @96 mg/l
52.00 mg/l @ 1918
Inorganic Solids IS: 3025 (Part-18):2012 5 - 3000 mg/l
mg/l
Suspended Matter IS:3025 (Part-17) :2012 5 - 1000 mg/l 3.50 mg/l @ 76 mg/l
pH IS:3025 (Part-11):1996) 2 - 12 0.028 @7.77
Total Hardness (as 4.08 mg/l @145.02
IS: 3025 (Part-21):2009 2 – 5000 mg/l
CaCO3) mg/l
Chloride (as Cl) IS:3025 (Part-32):2003 5 – 2000 mg/l 6.23 mg/l @ 224 mg/l
1.35 mg/l @ 36.64
Calcium (as Ca) IS:3025 (Part-40):2003 2-1000 mg/l
mg/l
Magnesium (as Mg) IS:3025( Part-46):2003 2-1000, mg/l 2.7 mg/l @31.6 mg/l
Total Alkalinity 1.24 mg/l @27.21
IS: 3025 (Part-23):2003 2 - 1000 mg/l
(as CaCo3) mg/l
0.01 mg/l @0.317
Fluoride (as F) IS: 3025 (Part- 60):2008 0.1 - 15 mg/l
mg/l
0.84 mg/l @23.04
Drinking Sulphate (as SO4) IS: 3025 (Part-24):2003 1 - 400 mg/ l
(ii) mg/l
Water
Total Dissolved 5 - 10000 11.20 mg/l @334.8
IS:3025 (Part-16):2006
Solids mg/l mg/l
APHA 3111D Direct
Aluminum (Al) Nitrous Oxide-Acetylene 0.02-50 mg/l 0.08 mg/l @0.49 mg/l
Flame Method :2017
0.1 – 50.0
Boron IS:3025 (Part-57):2005 0.06 mg/l @2.5 mg/l
mg/l
APHA 3111B, Air 0.05 – 50
Cadmium (Cd) 0.01 mg/l @ 3.0 mg/l
Acetylene Method:2017 mg/l
APHA 3111B, Air
Copper (Cu) Acetylene Flame 0.03 -50 mg/l 0.01 mg/l @1.5 mg/l
Method:2017
Signature, Date & Name of Lab Signature, Date & Name of Signature, Date & Name of Lead
Representative Assessor (s) Assessor
FORM 72
RECOMMENDED SCOPE OF ACCREDITATION
(For Testing Laboratories)
Laboratory: Shreya Test House Date(s) of Visit:
(Laboratory Division of Shreya Realty Pvt. Ltd.)
Discipline: Chemical Testing
S. Product(s) / Specific tests Test Method / Standard Range of Uncertainty of
No Material of test performed against which tests are Testing/ Measurement
performed Limits of ()UM at value
detection
Iron (Fe) APHA 3500 Fe B1, 10 0.02 mg/l -5 0.020 mg/l @1.0 mg/l
Phenanthroline Method : mg/l
2017
Lead (Pb) APHA 3111 B Air Acetylene 0.01 mg/l-25 0.02 mg/l @ 3 mg/l
Method: 2017 mg/l
Manganese (Mn) APHA 3111 B Air Acetylene 0.06 mg/l-25 0.22 mg/l @5.06 mg/l
Method: 2017 mg/l
Nickel (Ni) APHA 3111 B Air Acetylene 0.02 mg/l – 50 0.02 mg/l @ 3.0 mg/l
Flame Method: 2017 mg/l
Total Chromium (as APHA 3111 B Air Acetylene 0.03 mg/l – 25 0.05 mg/l @ 2 mg/l
Cr) Method: 2017 mg/l
Zinc (Zn) APHA 3111 B Air Acetylene 0.03 mg/l – 50 0.02 mg/l @3.0 mg/l
Method: 2017 mg/l
Ground Water IS: 3025 (P-22)-1986(RA- 9.2 mg/l @105.6 mg/l
Acidity 0.5 - 500 mg/l
(iii) 2009)
Aluminum (Al) APHA 3111 D: 2018 0.02-50 mg/l 0.08 mg/l @0.49 mg/l
Boron (B) IS:3025 (Part-57): 2005 0.1-50 mg/l 0.06 mg/l @ 2.5 mg/l
APHA 3111B, Direct Air 0.01 mg/l @3.0 mg/l
0.003
Cadmium (Cd) Acetylene Flame Method :
-50 mg/l
2017
1.35 mg/l @36.64
Calcium (Ca) IS: 3025 (Part-40) :2003 2-1000 mg/l
mg/l
Chloride (as Cl) IS:3025 (Part-32): 2003 5 – 2000 mg/l 6.23 mg/l @224 mg/l
1-10,000 20 µmhos/cm@ 1500
Conductivity IS: 3025 (Part-14):2013
µmhos/cm µmhos/cm
APHA 3111B, Air Acetylene 0.01 mg/l @1.5 mg/l
Copper (Cu) 0.03-50 mg/l
Method :2017
Dissolved Oxygen IS: 3025 (Part-38):2003 0.1- 10 mg/l 0.29 mg/l @6.86 mg/l
IS:3025 (Part-60):2008 0.01mg/l@0.317 mg/l
Fluoride (as F) 0.1 - 15 mg/l
APHA 4500 H+B 0.02 mg/l @1.0 mg/l
Iron (Fe) 0.02-5.0 mg/l
Electrometric Method:2017
APHA 3111B, Air Acetylene 0.02 mg/l@ 3.0 mg/l
Lead (Pb) 0.01-25.0 mg/l
Method:2017
Magnesium (as Mg) IS:3025 (Part-46):2003 2-1000 mg/l 1.24 mg/l@27.21mg/l
0.06 mg/l–25 0.22 mg/l @5.06 mg/l
Manganese (Mn) APHA 3111B:2017
mg/l
APHA 3111B,Air Acetylene 0.02 mg/l–50 0.02 mg/l @3.0 mg/l
Nickel (Ni)
Method:2017 mg/l
Signature, Date & Name of Lab Signature, Date & Name of Signature, Date & Name of Lead
Representative Assessor (s) Assessor
FORM 72
RECOMMENDED SCOPE OF ACCREDITATION
(For Testing Laboratories)
Signature, Date & Name of Lab Signature, Date & Name of Signature, Date & Name of Lead
Representative Assessor (s) Assessor
FORM 72
RECOMMENDED SCOPE OF ACCREDITATION
(For Testing Laboratories)
IS:3025 (Part-60):2008 0.01mg/l@0.317 mg/l
Fluoride ( F) 0.1 - 15 mg/l
Signature, Date & Name of Lab Signature, Date & Name of Signature, Date & Name of Lead
Representative Assessor (s) Assessor