Professional Documents
Culture Documents
Safety precautions
Warnings
• Disconnect equipment from power line before cleaning
• If liquid has accidentally entered the system or its parts, disconnect the power cord from
the power supply and have the equipment serviced by authorized service personnel.
• Regular preventive maintenance should be carried out annually.
• Do not use unspecified cleaners, materials or chemicals as they may damage device
surfaces, labels, or cause equipment failures.
• Do not pour or spray any liquid directly on cables or leadwires or permit fluid to seep into
connections or openings.
• Never use conductive solutions, solutions that contain wax, or wax compounds to clean
devices, cables or leadwires.
• Do not immerse any part of the device in liquids or allow liquid to enter the interior.
• The decision to sterilize must be made per your institution’s requirements with an
awareness of the effect on the integrity of the cable or leadwire.
• Clean the surface of the probe before and after each patient use.
• CABLE/SENSOR AFTER CARE
− Do not immerse sensors or patient cables in water, solvents or cleaning solutions.
− Do not reuse sensors intended for single patient use.
− Do not sterilize sensors or patient cables by irradiation, steam, or ethylene oxide.
− Clean the surface of the probe before and after each patient use.
− Allow sensor and cable to dry completely after cleaning. Moisture and dirt on the
connector can affect the measurement accuracy.
− If a probe is damaged in any way, discontinue use immediately.
− Inaccurate SpO2 data can result if a sensor is past its useful life.
− A damaged sensor or a sensor soaked in liquid may cause burns during
electrosurgery.
• Since calibration gas contains anesthetic agents, always ensure sufficient ventilation of
the room during calibration.
• Inaccurate readings due to:
− use of unapproved accessories
− reuse of single-use accessories
− affect on pressure strong scavenging suction
• The user may only perform maintenance procedures specifically described in this
manual.
• On the gas module indicates that airway gases should be calibrated every six months in
normal use and every two months in continuous use.
9-1
B40 Patient Monitor
• Do not autoclave any part of the system with steam or sterilize with ethylene oxide.
• Do not disinfect or open the water trap cartridge. Do not touch the water trap membrane.
The hydrophobic membrane is damaged if any cleaning is attempted other than rinsing
with water.
Cautions
• PACKAGING DISPOSAL - Dispose of packaging material, observing applicable waste
control regulations.
• Do not use or store equipment outside the specified temperature, humidity, or altitude
ranges.
• DISPOSAL - At the end of its service life, the product described in this manual, as well as its
accessories, must be disposed of in compliance with the guidelines regulating the
disposal of each product. If you have any questions concerning disposal of a product,
please contact GE or its representatives.
Overview
For safe and reliable function and operation of the monitor, regular care has to be carried out
according to the instructions in this manual and to the maintenance procedures described in
the "Technical Reference Manual”.
If the monitor does not function as it should and troubleshooting cannot solve the problem,
contact your service or sales representative. Do not perform any other cleaning or
maintenance procedures than those described in the monitor manuals.
NOTE: Refer to the accessory package for detailed cleaning instructions.
Preventive maintenance
Daily and between the patients
• Wipe the monitor and module surfaces.
• Wipe the ECG trunk cable and leadwires, NIBP cuff and cables and SpO2 sensors. Avoid
excessive use of liquids.
• Change or sterilize all airway and invasive patient accessories.
• Clean, disinfect or sterilize reusable temperature probes.
• Empty the D-fend water trap whenever half full.
• Check that all accessories, cables and monitor parts are clean and intact.
• Clean the device as described in the “Cleaning” part of this section.
• When you start monitoring, check that the module is firmly in place, the accessories are
intact and properly connected, and that you have selected desired parameters to be
displayed in digit and waveform fields.
• Check/note the following points regarding different parameters.
9-2
Cleaning and care
Regular checks
If the monitor does not work as described, see section "Troubleshooting." first.
Pulse oximetry
• After connecting SpO2 cable and sensor, check that the red light is lit in the sensor .
Temperature
• After connecting Temp cable and sensor, check that the message ‘Performing temp test:’
will display in Temperature digital field for several seconds.
IBP
• After connecting IBP cable and transducer, check that the message ‘InvBP’s not Zeroed’
will display in message field.
Airway gas
• Make sure the acquisition module is connected to the monitor.
• After installing the module, ‘Calibrating gas sensor’ will display in gas waveform field for
about 1 minutes.
Entropy
• Make sure the acquisition module is connected to the monitor.
• Connect the Entropy sensor cable to the module, and check that the “No sensor”
message apprears.
9-3
B40 Patient Monitor
Every 12 months
Planned maintenance check
The annual check according to detailed instructions of the "Technical Reference Manual"
requires trained service personnel and appropriate testing tools and equipment.
Power interruption
NOTE: Always use the monitor with batteries inserted. Otherwise all trend data and
temporary settings are lost if the power cable is detached from the mains.
If the monitor is turned off, patient data and the latest user-made settings remain in the
monitor memory for 15 minutes even if the mains power is interrupted. If not, contact service
personnel. After 15 minutes, patient data is lost and the monitor returns to the user default
settings.
9-4
Cleaning and care
Changing fuses
1. Remove the power cord if used.
2. Remove the fuse holder by pulling the holder out with screwdriver.
3. If a fuse is blown, replace it with a fuse of the correct type and rating.
NOTE: Only a qualified medical personnel allow to operate.
Cleaning
The appropriate cleaning procedure depends on where and how the part or accessory is used
and on the patient's condition.
Monitor casing
• Wipe with mild hospital detergent solution. Make sure not to leave any liquid spills on any
metal part.
• Let dry completely before connecting to power source.
• Check ventilation holes and clean if necessary.
Modules
• Wipe the front panel as any monitor casing.
• Do not wash or immerse module in any liquid.
The internal sampling system of the airway module does no cleaning nor sterilization. The
water trap functions as a bacteria filter and there is no reverse flow back to the patient. If the
measuring chamber is suspected to be contaminated (for example, gas zero error), the airway
module should be serviced by authorized service personnel.
Display
• Wipe all splashes immediately with a dry cloth.
• Wipe the LCD display after use with a cloth moisturized with mild detergent solution.
9-5
B40 Patient Monitor
ECG cables
• Wipe the cables with mild detergent solution.
• Avoid excessive use of liquids.
• Disinfect when necessary.
• Allow the product to dry completely after cleaning.
Pulse oximetry sensors
The GE Healthcare pulse oximetry sensors are latex-free. Take possible patient allergies into
account also when selecting the cleaning agent.
1. Detach the sensor from the patient and the monitor.
2. Wipe the sensor with mild detergent solution. Allow it to dry completely before use.
Sensors can be disinfected with chlorite compounds.
The sensors may be sterilized using an ethylene oxide mixture at 50 to 60°C / 120 to 140°F.
NOTE: After ethylene oxide sterilization, sensors must be well aerated in a ventilated place.
Temperature probes
• Clean with mild detergent solution and rinse with water.
• Disinfect or sterilize when necessary.
NIBP cuff hose
• The cables and hose can be wiped with mild detergent solution.
• Disinfect when necessary.
NIBP cuff
The NIBP cuffs listed in the “Supplies and Accessories” catalog are latex-free. Take possible
patient allergies into account also when selecting the cleaning agent.
• Clean only when necessary. Wash the cuff in mild detergent solution. Do NOT use alcohol.
Airway adapter
• Replace the single use adapter after each patient.
A reusable adapter can be disinfected with glutaraldehyde or alcohol.
If you want to clean the adapter before use: submerge the adapter in 70% alcohol solution for
30 seconds and rinse carefully with water.
Make sure that all traces of alcohol or detergent are rinsed away or dried before connecting to
the patient.
Sampling line
• Do not reuse the sampling line. Reusing a cleaned sampling line may affect measurement
results.
NOTE: Do not wash or disinfect calibration gas sampling lines.
9-6
Cleaning and care
Water trap
WARNING Do not disinfect or open the water trap cartridge. Do not touch the water
trap membrane. The hydrophobic membrane is damaged if any cleaning is
attempted other than rinsing with water.
The water trap is based on a hydrophobic membrane, which prevents water and secretions
from entering the measuring chamber. Condensed water and saliva are collected into a
washable container.
• Empty the container whenever it is more than half full.
• In anesthesia: Replace the water trap every two months or when the text 'Replace D-
Fend' appears.
• In critical care: Replace the water trap every 24 hours, for each new patient, or when the
text 'Replace D-Fend' appears.
• When taking a new water trap into use, mark the date on the appropriate label on the
water trap cartridge.
• The water trap cartridge is disposable. Do not wash or reuse the cartridge.
To lengthen the lifetime of the monitor and minimize downtime:
• After washing or disinfecting the water trap container, make sure there is no alcohol nor
detergent left when used again. Traces of alcohol or other organic cleaning solutions may
affect measurement.
• Do not force air or oxygen through the water trap.
• Do not allow smoke and dust to enter the water trap.
• While administering nebulized medication, disconnect the gas sampling line from the
patient circuit for 30 minutes.
If the message ‘Sample line blocked’ alarm occurs:
• Replace the sampling line.
• Empty the water trap container. It may be full.
9-7
B40 Patient Monitor
To remove the water trap, push the trigger above the water trap. The water trap is spring-
loaded. The message ‘Check D-fend’ is displayed.
• Detach the container from the water trap cartridge by pulling it carefully downward.
• Empty and clean the container.
• Attach the container tightly back into the cartridge.
• Push the whole unit into its housing on the front panel until the latch is set.
• Press the Normal Screen key to restart monitoring. Check that the message ‘Check D-
fend’ disappears.
Other accessories
See the accessory package for instructions for how to clean and check them. Do not reuse
single-use disposable accessories.
9-8
10 Troubleshooting