You are on page 1of 48

ANTERIOR OCULAR MICROSCOPY 18

PART 1 BIOMICROSCOPY
S. Zantos and 1. Cox

18.1 INTRODUCTION subject of biomicroscopy of the eye, the


first books being published in the early
The transparent nature of the ocular tissues
1920s by Koby, Vogt and Koeppe. Other
enables non-invasive examination of the liv­
noteworthy treatises were later published
ing eye that is unique in clinical work. The
by Butler (1924), Berliner (1943) and Dog­
slit lamp microscope combines a clever illu­
gart (1949). This was a period of discovery­
mination system with a convenient stere­
the various slit lamp techniques for illumi­
omicroscopic viewing system. Its great
nating and observing the ocular tissues
variety of attachments make it an extremely
were discussed meticulously, and the nor­
versatile instrument for examining the
mal and abnormal appearances of the eye
human eye.
were beautifully described and illustrated,
The first binocular corneal microscope
with detail similar to that seen using mod­
was constructed by H. Aubert in 1891.
em instrumentation.
Improvements were made by S. Czapski,
A significant advance in slit lamp evolu­
and in particular by Alvar Gullstrand who,
tion came in 1965 with the introduction of
in 1911, presented his design of an illumi­
instrumentation capable of photographing
nation system that made slit lamp micros­
copy practical by providing a sufficiently
bright and focused beam of light to the ~ye.
.",¥~
A subsequent modification, the so-called
;';'.
Kohler illumination system, served to
enhance the homogeneity of the slit beam
projected on the eye, enabling optical sec­
tioning of the eye tissues, and it became
the basis of today's instruments. Over the
next 40 years numerous improvements
were made by Henker, Koeppe, Vogt,
Koby, Poser, Goldmann, Littman and oth­
ers, culminating in instrumentation that
resembled and operated similarly to
today's slit lamps (Fig. 18.1). Concurrently,
numerous publications appeared on the Figure 18.1 A typical modern slit lamp.

Contact Lens Practice. Edited by Montague Ruben and Michel Guillon.

Published in 1994 by Chapman & Hall, London. ISBN 041235120 X

360 Anterior ocular microscopy


optical sections of the eye. This coupled 18.2 INSTRUMENTATION
with the development of numerous attach­
Many well-designed and well constructed slit
ments (pachometer, goniolens, Hruby fun­
lamps are available today. The main features of
dus lens, applanation tonometer, filters,
slit lamps can be seen in Figures 18.2, 18.3 and
polarizers, fluorophotometer, endothelial
18.4. The three essential components of all
lens, video, etc.), have made the modern
instruments are the slit illumination system,
slit lamp microscope an extraordinarily
the stereomicroscope viewing system and the
useful instrument in research and in clini­
mechanical system for adjusting position and
cal practice. Its application to contact lens
height of the instrument and of the patient.
work is extensive.
Detailed reviews of these are available in sev­
With regard to biomicroscopic examina­
eral textbooks (e.g. Brandreth, 1978),and famil­
tion of the eye, it is worth adding that in
iarization is best done using the actual
recent years two new instruments have been
instrument. However, in selecting a brand of
investigated in the search for higher magni­
slit lamp, we recommend that particular atten­
fication and resolution of tissue structures:
tion be given to certain features.
the specular microscope and the confocal
(tandem scanning) microscope. Both instru­
ments can be used for the living eye, though
the confocal microscope is presently in the
research and development phase, and not yet
used in clinical practice. The specular micro­
scope is widely available for clinical work,
having been extensively used for evaluation
of the corneal endothelium before and after
cataract surgery and intra-ocular lens
implantation; in contact lens research, the
instrument has been used for investigating
long-term changes in the endothelium, as
well the effects of lens wear on the corneal
epithelium. Where the specular microscope
reveals cellular patterns only at tissue layers
differing in refractive index (e.g. epithelium­
to-air; endothelium-to-aqueous), the confo­
cal microscope reveals tissue structure at any
layer. Moreover, it does so with considerably
more detail, for example, stromal keratocytes
and cell nuclei can be seen in uiuo with the
confocal microscope. Numerous fine review
articles exist on the optical principles and
applications of the specular microscope Figure 18.2 The slit lamp illumination system. Slit
(Koester & Roberts, 1990; Mayer, 1984) and beam width and height are adjusted by the
the confocal microscope (Cavanagh et al., knurled rings at the front, while the knobs on the
1990; Masters & Kino, 1990). Still, the slit rear of the housing allow coloured filters to be
introduced into the light beam. The diffusing
lamp microscope remains the instrument of
filter can be seen at the top of the illumination
choice for clinical practice, and its main housing ready to be flipped into place in front of
features, applications, and procedures will the prism. Slit beam offset is adjusted by turning
now be discussed. the prism housing laterally.
Instrumentation 361

Figure 18.3 The stereomicroscope viewing sys­


tem. The eyepieces are adjustable to compensate
for the viewer's refractive error and pupillary
distance. Magnification is adjusted by turning the
handle located on either side of the lens housing.

Figure 18.4 The mechanical system. Lateral move­


18.2.1 THE SLIT LAMP ILLUMINATION ments are controlled by the joystick, while vertical
SYSTEM adjustments are made with the electric switch
located in front of the joystick. This adjustment
1. High illumination is required to enhance may be manual on many slit lamps. Illumination
detection of subtle conditions in the eye intensity can be altered using the rheostat located
on the left of the joystick.
tissues when using optic section or indi­
rect illumination techniques (e.g, for
assessing corneal thickness or for evaluat­
ing changes in corneal transparency). A 18.2.2 THE STEREOMICROSCOPE VIEWING
slit illumination up to 600 000 lux is desir­ SYSTEM
able, and a halogen or xenon light source The optical design of today's slit lamps is
is preferred to a tungsten lamp for reasons quite good, so that the practitioner need not
of longevity. colour, temperature and low be too concerned with the resolution, bright­
heat generation. ness, and stereoscopic results. Detailed dis­
2. The ability to displace (offset) the slit cussion oithese items is given in other books
beam sideways is invaluable for perform­ (e.g. Muller; Schmidt, 1975). From a more
ing indirect illumination techniques (e.g. practical standpoint, the magnification, focus
sclerotic scatter, retro-illumination), and it adjustment and parfocality, and the working
is necessary that a control device for doing distance are more important considerations:
this be present, conveniently and finely
adjustable, and steady. A click-stop posi­ 1. The magnification should be easily
tion is useful for re-setting the beam to its changeable, and it should preferably
normal position. cover the range approximately 5X to 30X
3. A diffusing filter and a cobalt blue filter before one has to change eyepieces to
should always be incorporated into the achieve higher magnification.
instrument, and other filters are desir­ 2. It is advisable to check periodically that
able, e.g. green (red-free) and polarizer. the object of regard remains in focus as
362 Anterior ocular microscopy
the magnification is changed through
the range available. A well-constructed
focusing rod accompanying the slit lamp
can be invaluable in checking the parfo­
cality of the slit illumination system and
the microscope viewing system. It is also
worthwhile to check for absence of
vertical prismatic effect between the
two eyepiece views as the objective
magnification is changed.
3. The working distance of the microscope
(from object plane to the front lens of the
microscope) should be approximately Figure 18.6 Diffuse illumination. Low magnifica­
110 mm, long enough to allow for tion overview of the whole eye.
manipulation of the eye or of accessory
attachments, but not too long so as to
require an uncomfortable ann position
during such manipulations.

18.2.3 lHE MECHANICAL SYSTEM

Accurate and convenient positioning of the


instrument in relation to the patient's eyes is
vital to efficient biomicroscopy:

Figure 18.7 Diffuse illumination. Medium magni­


fication view of the edge of a soft lens.

Diffusing
Filier
1. It is preferable that the height adjust­
ment of the slit lamp and the joystick
for lateral adjustment and focus be
combined, or be near enough to each
other so that one hand can control all
Illumination these functions, leaving the other hand
System
free to manipulate the patient's eyelids
Observation or an accessory device, or to manoeu­
System vre the illumination system or the
microscope.
2. It is important that the friction in the
Figure 18.5 Diffuse illumination. The diffusing
sliding stage be not so loose that slip­
filter in front of the illumination system spreads a page occurs as the instrument is moved,
wide beam of unfocused, scattered light uni­ and not so tight that the joystick might
formly across the field of view. be jerky or excessively tilted from its
Instrumentation 363

Normal to
Cornea

Illumination Actual
System View

Observation
System

Figure 18.8 Optic section. The illumination system sections the cornea with a thin, beam of light,
focused co-incidentally with the axis of the observation system.

vertical position as the instrument is In addition to the slit lamp microscope,


moved during the examination. illumination system and mechanical stage,
3. The well-constructed instrument will the construction of the patient's headrest
have the slit illumination system and and of the instrument table are obviously
microscope system nearly perfectly important to the ease of performing biomi­
coaxial, so that as the former is rotated croscopy. As with other ophthalmic instru­
(swung) from one side to the other, the ments, the comfort and adjustability of
slit beam will not shift significantly these functions should be tried for suitabil­
sideways in the field of view, and ity to the practitioner's personal preference
parfocality can be maintained. The before purchase.
focusing rod can be used to check this, Reviews of the specifications of the vari­
but it must be perfectly constructed ous commercially available instruments have
and positioned in the coaxial slot to be been published elsewhere, and the reader is
of any value. referred to these articles for comparison of
364 Anterior ocular microscopy
instrument features (Stockwell, 1983 Marto­
nyi, 1989;).

18.3 SLIT'LAMP ILLUMINATION AND


OBSERVATION TECHNIQUES

The ability to detect and diagnose the vari­


ous anterior segment conditions which may
occur depends to a large extent on the ability
of the observer to correctly adjust and posi­
tion the illumination system of the slit lamp.
In many cases, subtle changes to anterior
segment structures can only be detected and Figure 18.10 Optic section. High magnification
thoroughly assessed by using a combination view of infiltrates invading the peripheral cornea.
of illumination techniques. Slit lamp illumi­ Note that the optic section reveals the infiltrates to
nation can be categorized into four main be concentrated towards the anterior stroma.
groups:
1. Diffuse illumination. Within each of these categories are several
2. Direct illumination. techniques which are defined and described
3. Indirect illumination. in the following section according to the
4. Filtered illumination. adjustment and positioning of both the illu­
mination and observation systems. This
differentiation is important for the inex­
perienced observer to ensure that the opti­
mal observation technique is used for
viewing each particular condition or struc­
ture. However, it should be understood that,
in reality, several different methods of illu­
mination may be simultaneously present in
the field of view at anyone moment during
the slit lamp procedure, providing the expe­
rienced observer with the necessary differen­
tial information without resorting to several
different physical adjustments of the slit
lamp. This speeds up the slit lamp examina­
tion and reduces patient fatigue and discom­
fort.

18.3.1 DIFFUSE ILLUMINATION

Diffuse illumination is so named because a


ground glass diffusing filter is placed in the
Figure 18.9 Optic section. Medium magnification focused light beam of the slit lamp. This
view of a central corneal scar. Note that the optic defocuses and scatters the light to give a
section allows the depth of the scar to be easily broad even illumination over the entire field
determined. of view of the observation system (Fig. 18.5).
Slit lamp illumination and obsertiation techniques 365
The angle of the illumination arm is not 18.3.2 DIRECT ILLUMINAnON
critical when the diffuser is in place and can
Direct illumination describes any illumina­
be anywhere from 10 to 70 degrees in relation
tion technique where the slit beam from the
to the observation arm. Brightness is con­
illumination arm and the optics of the obser­
trolled by both the slit width and the illumi­
vation system are focused co-incidentally on
nation rheostat of the slit lamp.
the object or area under view. The following
Diffuse illumination is typically used for
set-ups are considered to be direct illumina­
low magnification views of the anterior seg­
tion techniques:
ment, particularly for assessing conjunctival
redness, contact lens fitting performance, or 1. Optic section.
for initial gross scanning views of anterior 2. Parallelepiped.
ocular structures (Figs 18.6 and 18.7) 3. Broad beam.

Normal to
Cornea

Illumination Actual

System View

Observation
System

Figure 18.11 Parallelepiped. Essentially a broadened optic section. Note that the size of an object
within a parallelepiped can be estimated by comparing the width of the parallelepiped to its depth
(which is approximately 500 lAID).
366 Anterior ocular microscopy
4. Conical beam.
5. Specular reflection.
6. Oblique.

Optic Section
Optic section desaibes an illumination tech­
nique which utilizes an extremely thin beam
of light (e.g. 0.02-0.1 mm) to 'cross-section'
the corneal tissue in much the same way as
would be seen with a transverse histological
section. This provides the ability to assess
the depth of an object within the corneal Figure 18.13 Parallelepiped. High magnification
layers or the depth of the tear film between a view of endothelial folds caused by stromal
rigid lens and the anterior cornea. oedema. Note that the anterior portion of the
To set up an optic section, the illumination cornea is out of focus when the plane of focus is at
the endothelium. This demonstrates the short
arm is positioned on the side of the micro­
depth of field that accompanies higher magnifica­
scope that corresponds to that section of the tion views.
cornea to be viewed (i.e. temporal side for
temporal cornea) and is set at an angle
between 30 and 60 degrees to the observation
arm. Increasing the angle creates a wider refined while viewing the cornea through
section but may not allow the intersection of the microscope.
several objects separated by depth (Fig. 18.8). Optic section is used for viewing anything
The illumination rheostat is turned fully where a sense of depth will enhance the
on and the slit beam is narrowed as much as appreciation of the object's size and location,
possible while still providing sufficient illu­ such as a foreign body embedded in the
mination for viewing. The slit width can be cornea, corneal haziness and oedema, endo­
thelial bedewing or pigment cells, or the
clarity of the corneal epithelium (which
appears dark when it is clear and grey when
it is oedematous) (Figs. 18.9 and 18.10).

Parallelepiped illumination
A parallelepiped is perhaps the most com­
monly used form of direct illumination used
in slit lamp examinations. It is essentially the
same in setup as an optic section, except that
the slit beam is widened until it is approxi­
mately the same width as the apparent cor­
neal depth (e.g. 0.1-0.7 mm) (Fig. 18.11). This
view allows an appreciation of an object in
Figure 18.12 Parallelepiped. High magnification
view of the central cornea, the parallelepiped
true three dimensions, since its width,
being almost twice as wide as it is deep. Note the height and depth can be assessed simulta­
microcysts seen to the right of the parallelepiped neously.
in indirect retro-illumination. Parallelepiped is used to assess the corneal
Slit lamp illumination and observation techniques 367
endothelium, corneal scarring, corneal stain­ observation arms is not critical and should
ing, neovascularization, corneal infiltrates, be adjusted to provide the optimum view of
and corneal striae and folds (Figs. 18.12 and the object in question.
18.13). Broad beam illumination is particularly
useful for viewing corneal nerve fibres,
debris trapped beneath soft and rigid lenses,
Broad beam illumination
conjunctival abnormalities and pterygia, and
Broad beam illumination is the next logical larger scars and opacities within the cornea
'step from the optic section and parallelepi­ (Figs. 18.15 and 18.16).
ped, and refers to a similar setup, except that
the slit beam is now widened to somewhat
Conical beam illumination
broader than the apparent corneal depth (e.g.
1-5 mm) (Fig. 18.14). Light intensity should Conical beam is useful primarily for observ­
be reduced using the illumination rheostat. ing inflammatory cells and protein in the
The angle between . . the illumination and anterior chamber as a result of a uveal

Normal to
Cornea

Illumination
System

Observation
System

Figure 18.14 Broad beam. An extremely wide parallelepiped, used mostly for looking at large corneal
defects.
368 Anterior ocular microscopy
inflammatory response. The conical beam is
formed by setting up an optic section and
then reducing the height of the slit beam
until it is about 1-2 mm in height. Light
intensity should be maximized using the
illumination rheostat. Since the illumination
intensity of the conical beam is low, observer
sensitivity will be increased if the room
illumination is turned off. The anterior
chamber should then be scanned at low to
medium magnification from side to side.
Cells (inflammatory cells) will appear as
bright, white reflections as they pass through Figure 18.16 Broad beam. High magnification
the conical beam, while flare (protein) will view of corneal nerves.
appear as yellowish particles. These reflec­
tions will be more easily detected if viewed
against a dark background, it is therefore equals the angle of the observation axis
best to reduce light scattering by avoiding through one of the oculars (Fig. 18.18). At
having the conical beam strike the iris, and this specific angle the illumination beam is
by enhancing pupil dilatation by directing reflected from the smooth surfaces of the
the conical beam entering the pupil away anterior segment (tear film, contact lens
from the foveal area. Some observers suggest posterior surface, endothelium and ante­
oscillating the slit beam using the offset rior lens capsule) and provides a bright,
control to enhance the visibility of cells and mirror-like reflection. To achieve specular
flare (Fig. 18.17). reflection, set-up the slit lamp to form a
parallelepiped. Adjust the magnification to
the lowest available setting. Move the
Specular reflection
observation arm up to 20 degrees away
Specular reflection is a specific case of the from the illumination arm, and then begin
parallelepiped setup where the angle of the to move the illumination arm in the oppos­
incident slit beam to the corneal surface ing direction while observing the corneal
surface. At the point where specular reflec­
tion is achieved, a bright reflex will fill one
of the oculars. Note that specular reflection

r
cannot be achieved binocularly. Having
located the specular reflex, increase the
magnification of the observation system
stepwise to its highest setting to provide a
t clear view of the endothelial mosaic or
other object of interest.
Specular reflection is used for observing
the quality of the tear layer, since it allows
the lipid layer of the tears to be viewed.
Obviously, lens front surface wetting can
also be observed in this manner. It is also
Figure 18.15 Broad beam. High magnification important to note that specular reflection is
view of debris trapped beneath a soft lens. the only technique which allows observa­
Slit lamp illumination and observation techniques 369

Cells & Flare

Illumination
System

Observation
System

Figure 18.17 Conical beam. A narrow optic section with the height reduced to form a small oval beam.
Note that the dilated pupil is used as a dark field background to enhance observation of any cells or
flare in the anterior chamber.

tion of the corneal endothelial mosaic in Oblique illumination


vivo, along with guttae, folds and blebs.
However, it is often difficult for the inexpe­ Although infrequently used in contact lens
rienced observer to clearly view the endo­ practice, oblique illumination provides a
thelial reflex because the significantly unique view of contour changes which
brighter reflex from the tears causes visual would not be visible by other methods of
distraction. Increasing the overall angle illumination. As the name suggests, oblique
between the two arms of the slit lamp will illumination is achieved by setting up a
increase the separation between the two parallelepiped and then moving the illumi­
specular reflexes and minimize this inter­ nation arm away from the observation sys­
ference. For this same reason, viewing the tem until the angle between them is close to
endothelial specular reflex is easier for the 90 degrees. The illumination arm is adjusted
peripheral areas of the cornea than for the so that the light beam is almost tangential to
central region, since the cornea's greatest the object of regard. In this way, very small
thickness is in the periphery (Figs. 18.19 deviations in the topography of the iris, for
and 18.20). example, will cast large areas of shadow,
370 Anterior ocular microscopy

Normal to
Cornea

.Illumination
System
Observation
System
Figure 18.18 Specular reflection. Note that specular reflection can only be achieved when the angle of
the incident illumination to the normal of the cornea equals the angle of the observation system, and
that specular reflection is viewed through one eyepiece only.

clearly defining the high and low points of a where the focus of the illumination beam
surface which appears to be relatively flat does not coincide with the focal point of the
under other more frequently used methods observation system (Fig. 18.23). To achieve
of illumination. this the slit beam must be offset (i.e. dis­
Oblique illumination is used to examine the placed sideways). This can be achieved
uniformity of the iris surface, as well as any manually by turning the prism or mirror
changes in the corneal structure such as Fleis­ controlling the slit beam away from the axis
cher's ring, which may otherwise be hard to of the illumination arm, or, as is often done
assess using other methods of direct illumina­ by experienced observers, by achieving
tion. In contact lens practice, oblique illumina­ direct illumination and focusing the sli t
tion may also be used to investigate the beam to one side of the object to be viewed
elevation height of front surface deposits, lens and utilizing a different part of the micro­
edge lift, and the location of bifocal lens optic scope field of view to observe the object
zones (Figs. 18.21 and 18.22). which is now indirectly illuminated (Fig.
18.24).
18.3.3 INDIRECT ILLUMINATION
As with direct illumination, indirect illu­
mination can be divided into several specific
Indirect illumination refers to any technique illumination techniques. The following set­
Slit lamp illumination and observation techniques 371

Figure 18.21 Oblique illumination. Medium mag­


nification view of a pinguecula. Note that the
shadow cast by the pingecula when illuminated
by oblique illumination provides an indication of
the height of the growth.
Figure 18.19 Specular reflection. High magnifica­
tion view of endothelial guttae.

ups are considered indirect illumination


techniques:

1. Proximal.
2. Sclerotic scatter.
3. Retro-illumination (direct and indirect).

Figure 18.22 Oblique illumination. High magnifi­


cation view of the iris architecture. The radial
folds seen in this photograph were not visible
with frontal direct illumination. Note also that the
small areas of pigmentation do not cast a shadow,
indicating that they are not raised from the sur­
face of the iris.

Proximal
Proximal illumination is very similar to par­
allelepiped direct illumination, except that
the parallelepiped is offset slightly to one
Figure 18.20 Specular reflection. Low magnifica­
tion view of front surface wrinkling on a soft lens
side of the object being viewed. This allows
fitted too steeply. Note that the wrinkles are only the object and its immediate surrounding
visible in the area of specular reflection from the area to be illuminated by light scattered
tear film. through the cornea, uncovering subtle
372 Anterior ocular microscopy

Normal to
Cornea

Illumination

System

Observation
System

Figure 18.23 Indirect illumination. The illumination and observation system axes do not coincide in
indirect illumination. Note that in this case indirect illumination has been achieved by offsetting the
slit beam.

changes in the corneal transparency which Sclerotic scatter


may not have been visible using direct illu­
mination. One example of this is the obser­ Sclerotic scatter is a specific case of indirect
vation of the leading edge of a pterygium as illumination used to investigate any subtle
it moves onto the cornea. Direct illumination changes in corneal clarity occurring over a
will only display the pterygium itself, while relatively large area, such as the central cor­
proximal indirect illumination allows the neal clouding associated with steeply fitted,
changes in the corneal tissue in advance of low Dk rigid lenses. Sclerotic scatter is
the pterygium to be viewed and assessed. achieved by setting the slit lamp up for a
Proximal illumination can be used to view all wide angle (45 to 60 degrees) parallelepiped
objects and conditions within the cornea view of the central cornea, and then manu­
which are viewed with direct illumination. ally offsetting the slit beam until it is focused
Slit lamp illumination and observation techniques 373

Normal to
Cornea

Illumination
System

Observation
System

Figure 18.24 Indirect illumination. Note that in this case indirect illumination has been achieved by
utilizing the microscope field of view.

on the nasal or temporal limbus (Fig. 18.25). grey hazy area within a clear background.
Correct positioning of the slit beam is sig­ Sclerotic scatter can also be used to observe
nalled by a bright glow from the limbal corneal scars, foreign bodies and deposits
region directly opposite that being illumi­ within the corneal layers. It should be noted
nated. This glow results from the fact that the that widespread corneal oedema, such as that
light from the slit beam is being totally seen with soft lenses, will not be detected
internally reflected between the epithelial using sclerotic scatter because there is no
and endothelial layers of the cornea and then portion of non-oedematous cornea with
scattered as it hits the scleral tissue of the which to compare it (Figs. 18.26 and 18.27).
opposing limbus. Obviously any area of the
cornea which is altered in such a way as to
Retro-illumination (direct and indirect)
increase light scatter will interrupt the total
internal reflection of the slit beam and can be Retro-illumination refers to any indirect illu­
seen either through the observation system mination technique where light is reflected
at low magnification or by the naked eye as a from the iris, anterior lens surface or retina
374 Anterior ocular microscopy
and used to illuminate an object or area in direct illumination are seen as black or dark
the cornea from behind. It should be noted objects on a lighter background in direct
that in retro-illumination there is no light retro-illumination. Direct retro-illumination
being reflected directly from the object of is achieved by setting the slit lamp as if
regard. forming a direct illumination parallelepiped,
then offsetting the, slit beam until the
reflected light of the slit beam from the
Direct retro-illumination
surface of the iris is directly behind the
This is that specific situation where the object under observation when viewed
object of regard is illuminated entirely from through the eyepieces (Figs. 18.28, 18.29 and
behind (by light reflected from a deeper 18.30).
tissue), and is viewed against an illuminated Direct retro-illumination of objects within
background, causing the object to be seen in the central region of the cornea and lens/
shadow. Therefore, objects seen as white or vitreous can also be achieved by aligning the
opaque against a darker background in slit lamp centrally as if forming a direct

Normal to
Cornea

Illumination
System

Observation
System

Figure 18.25 Sclerotic scatter. Corneal illumination is achieved by directing the illumination beam to
the limbal region of the cornea and creating total internal reflection of the light beam within the cornea.
Slit lamp illumination and observation techniques 375
illumination optic section on the geometric ~? -­
axis of the eye. The slit lamp is focused in the
plane of the object under observation and i
then the slit beam is offset to pass just inside
the edge of the pupil. The slit beam will be
reflected from the retina/choroid and the
object under observation will be seen against
the red glow of the fundus. This procedure is
more easily performed with large or dilated
pupils.

Indirect retro-illumination
Figure 18.27 Scleroticscatter. Medium magnifica­
Like direct retro-illumination, indirect retro­ tion view of a pterygium on the cornea. Note that
illumination has the object of regard being the full extent of the encroachment of the ptery­
illuminated entirely from behind (by light gium is visible under this illumination.
reflected from a deeper tissue) but now it is
being viewed against a dark background.
Indirect retro-illumination can be achieved the full extent of changes in the corneal
using a similar setup to that used for direct layers, such as epithelial microcysts, vacu­
retro-illumination, except that the illumina­ oles, scars, degenerations and dystrophies,
tion system is offset such that the reflected and trauma. Direct retro-illumination from
image from the iris is not directly behind the the retina is also used for estimating the
object under observation (Fig. 18.31). It is extent of lenticular opacities and changes,
interesting to note that if several objects are and also the location of optic zones in bifocal
viewed in direct retro-illumination, such as contact lenses. It should be noted that, even
epithelial microcysts, those to the left and though stereoscopic perception is present,
right of the centrally viewed microcysts are the depth of anomalies or lesions cannot be
seen in indirect retro-illumination, assessed reliably when only retro-illumination
Retro-illumination is used to investigate is used; therefore, the optic section technique
is used to complement the retro-illumination
technique (see Figs. 18.32 and 18.33).

18.3.4 FILTERED ILLUMINATION

Slit lamps are typically fitted with two


coloured filters which can be introduced into
the light path of the illumination system - a
cobalt blue filter to aid in fluorescein obser­
vation and a green ('red-free') filter to
enhance the contrast of blood vessels.
Fluorescein stain is routinely added to the
tear film to aid in the observation of corneal
and conjunctival staining, rigid contact lens
Figure 18.26 Scleroticscatter. Medium magnifica­ fittings and qualitative assessment of the tear
tion view of cellular debris trapped behind a soft film itself (soft lenses require the use of
lens. special high-molecular-weight fluorescein).
376 Anterior ocular microscopy

Normal to
Cornea

Microscope
Field of View

.Illumination
System

Observation
System

Figure 18.28 Direct retro-illumination, The illumination beam is directed so that the axes of the light
reflected from the iris and the observation system coincide.

Following fluorescein instillation, illumi­ allows transmission of the green light emit­
nation of the tear film with cobalt blue light ted by the fluorescein but blocks most of the
of about 490 nm maximum causes a greenish blue light reflected from the ocular surface,
light with a maximum of about 520 nm to be greatly enhances the contrast between adja­
emitted by the excited fluorescein molecules. cent areas covered by fluorescein (Figs. 18.34
The brightness of the emitted light is an and 18.35). This allows a more accurate
indicator of the amount of fluorescein accu­ assessment of subtle variations in tear film
mulated in a local area, and hence variations thickness beneath a rigid lens, and areas of
in the brightness of the area under observa­ corneal staining which would otherwise go
tion can be used as an indication of the undetected. When fluorescein is instilled, the
relative thickness of the tear film or, in the cobalt blue filter is typically used with broad
case of epithelial staining, the severity of the beam illumination and maximum slit lamp
tissue damage, at that location. Addition of a intensity since the combination of blue and
yellow filter in front of the microscope objec­ yellow filters significantly decreases image
tive (e.g. Kodak Wratten No. 15), which brightness through the oculars. It should be
Procedure for the slit lamp examination of the anterior segment 377

50% of the overall available light energy.


This can be improved by using long pass
interference filters in place of the regular
cobalt blue filter (Fig. 18.36). Long pass inter­
ference filters typically have an extremely
sharp 'cut-off' around the 490 nm region
with essentially no transmission above that
value. Average light transmission is typically
above 75% for the short wavelength region.
These characteristics enhance the contrast
and brightness of the area under observation
above and beyond that seen with the typical
Figure 18.29 Direct retro-illumination. High mag­ cobalt blue filter.
nification view of limbal vessel spikes growing Cobalt blue filters used without fluores­
towards the centre of the cornea. Note that the cein and the yellow filter can also be used to
looping of the vessels is visible under the direct enhance the view of certain intracorneal
retro-illumination from the iris. anomalies such as Fleischer's ring (some­
times seen in keratoconic patients) and the
Hudson-Stahli line.
'Red-free' filters are actually greenish-blue
in colour and are so-named because they
block all wavelengths of light at the red end
of the visible spectrum. Red-free filters
enhance the contrast of any structure which
is red in colour, such as arteries and veins, or
stains such as rose bengal, by displaying
them as dark against the lighter green back­
ground of the sclera, lids and iris.

18.4 PROCEDURE FOR THE SLIT LAMP


EXAMINAnON OF THE ANTERIOR
SEGMENT
Examination of the anterior segment of the
potential contact lens patient with the slit
Figure 18.30 Direct retro-illumination. High mag­ lamp should be done in a systematic and
nification view of epithelial microcyst seen ill ordered fashion to ensure that all structures
direct retro-illumination from the retina. are investigated, not just those indicated by
the patient's history and symptoms. In the
case of the pre-fitting baseline examination,
noted that the majority of cobalt blue filters this is especially important, since this infor­
provided by the slit lamp manufacturer are mation will be the basis for all future deci­
not ideal in their spectral and overall light sions regarding patient management and
transmission characteristics. Typical glass care.
cobalt blue filters transmit unwanted longer The procedure outlined here is designed to
wavelengths which are not blocked by the flow smoothly from one group of structures
yellow filter, and often transmit less than to the next, allowing adequate observation of
378 Anterior ocular microscopy

Normal to
Cornea

Illumination
System
Observation
System

Figure 18.31 Indirect retro-illumination, The observation system is directed so that the axes of the light
reflected from the iris and the observation system do not coincide.

all the tissues in the minimum possible time. the appropriate mounting hole, illuminate
If however, a particular point of interest is the rod with a narrow slit beam and focus
noted, it behoves the observer to deviate each eyepiece of the microscope on the slit
from this routine and use any other illumina­ beam individually. Always adjust the eye­
tion and observation techniques appropriate pieces from positive to negative to avoid
to that condition or structure before return­ accommodative spasm. Ensure that the slit
ing to the outlined procedure. beam is in the centre of the field of view,
otherwise the illumination system is out of
18.4.1 SEITING UP TIiE SLIT LAMP
alignment and must be serviced.

It is essential that the illumination and


18.4.2 SEITING UP THE PATIENT
observation systems of the slit lamp be
focused coincidentally, or clear, direct obser­ Adjust the height of the slit lamp and chair to
vation of an object will not be possible and provide a comfortable height for the patient
the light beam will appear offset whenever so that their neck is not extended or their
an object is viewed. To achieve this, place the head tilted. Take note of the patient's outer
focusing rod supplied with the slit lamp in canthus relative to the canthus mark on the
Procedure for the slit lamp examination of the anterior segment 379

vertical bar of the slit lamp headrest, and if


necessary, adjust the height of the chinrest so
that they are aligned. Instruct the patient to
look straight ahead and use either the fixa­
tion target provided with the slit lamp or a
distant target to provide a fixation point for
the eye not under examination. Instruct the
patient not to focus on the fixation target
since this may induce fatique and conver­
gence. Set the illumination arm temporally at
about 45 degrees and open the slit wide.
Have the patient close their eyes and align
the light beam over either their right or left Figure 18.33 Indirect retro-illumination, High
globe (start with either the right or left eye, magnification view of debris trapped behind a
but try to maintain consistency in which eye soft lens. Note that the debris in the light beam is
is examined first). Insert the diffuser into the in direct illumination, while that seen against the
light beam. Select the lowest magnification. dark background of the pupil is in indirect retro­
illumination.

Figure 18.34 Filtered illumination. Low magnifi­


cation view of a rigid lens fitting with fluorescein
instilled in the tears. Note that only the cobalt
blue illumination filter was used in this photo­
graph.

Look through the observation system and


Figure 18.32 Indirect retro-illumination. Medium ensure that their eye is centred in the field of
magnification view of deposits on the anterior view. Now instruct the patient to open their
surface of a soft lens. Note that the deposits can be eyes.
seen in three different types of illumination. The
deposits in the light beam are under direct illumi­
nation, those seen against the dark background of 18.4.3 EXAMINATION ROUTINE
the pupil are in indirect retro-illumination, and
those illumination from behind by the light Initially, using low magnification (e.g. 5-8x),
reflected from the iris (which are just barely examine the gross appearance of the eye and
visible) are in direct retro-illumination. lids with diffuse illumination, paying par­
380 Anterior ocular microscopy
matted or missing lashes, any localized dis­
coloration or redness of the bulbar conjunc­
tiva/ clarity of the cornea and unusual
colourings of the iris. If a lens is worn, note
the centration and movement of the lens
both in primary gaze and upgaze.
Move the slit lamp and focus on the tem­
poral canthus. Decrease the light beam to a
wide parallelepiped and remove the diffuser.
Increase the magnification to a medium level
(1~15X) and begin scanning along the lower
lid towards the nasal canthus. Note the
Figure 18.35 Filtered illumination. Low magnifi­ cleanliness of the lashes (makeup, etc.), the
cation view of a rigid lens fitting with fluorescein patency of the meibomian glands, the
instilled in the tears. Note that a yellow Wratten amount of debris in the tear film and the size
filter has been placed over the objective lens of the of the lower tear meniscus. At the mid-point,
observation system in addition to the cobalt blue pull down on the lower lid and focus on the
illumination filter in this photograph. The
palpebral conjunctiva. Note its redness and
enhanced contrast of this filter combination is
obvious when compared to Figure 18.34. the presence or absence of papillae or fol­
licles. Have the patient look up and scan the
bulbar conjunctiva for any undue redness or
discoloration, move the slit up and scan the
inferior limbus, estimating vessel ingrowth
and severity of injection. Also note any
undue hazing or arcus senilis near the lim­
bus. Release the lid and swing the illumina­
tion arm to the nasal side of the microscope,
again to about 45 degrees. If the patient is
wearing soft toric lenses, this is an ideal time
to assess the lens orientation. Continue the
scan to the puncta and caruncle, noting anv
abnormalities or debris.
Continue back along the upper lid margin,
Figure 18.36 Filtered illumination. Low magnifi­ paying particular attention to the redness of
cation view of a rigid lens fitting with fluorescein the lid. Again, at the mid-point, have the
instilled in the tears. Note that in addition to the patient look down, and assess the superior
yellow Wratten filter over the objective lens of the bulbar conjunctiva while holding the lid up.
observation system, the cobalt blue illumination Complete the scan to the temporal canthus
filter was replaced by a long wavelength cut-off
interference filter in this photograph. This filter remembering to swing the illumination arm
combination has slightly enhanced contrast when back to the temporal side after passing the
compared to Figure 18.35/ but most importantly mid-point.
the flash reflexes have been eliminated, improv­ Direct the patient to look nasally and scan
ing the view of the central fitting pattern. the temporal quadrant of the bulbar conjunc­
tiva. Assess the injection and sheen of this
tissue since this is a possible indicator of
ticular attention to any apparent abnormali­ reduced tear volume. Have the patient look
ties such as redness of the lid margins, straight ahead once again and move the slit
Procedure for the slit lamp examination of the anterior segment 381
nasally to the limbus. Again assess the ness, corneal striae or folds, nerve fibres or
degree of vessel ingrowth and injection. If any other abnormalities or observable mark­
the patient is wearing soft lenses, estimate ers. When the point of specular reflection is
the amount of lens/cornea overlap and reached, observe the smoothness of the
observe whether there is any conjunctival endothelium, any guttae or, bumps and
compression by the lens edge. Continue to degree of polymegethism. Also assess the
scan the cornea, noting any peripheral haze presence and quality of the lipid layer in the
or arcus and the general clarity of the cornea. specular reflection from the tears. Continue
If lenses are worn, note any deposits on the the scan, remembering to move the illumina­
lens front surface. At the corneal mid-point, tion arm nasally after the corneal mid-point
swing the illumination arm nasally once is reached. Move up and then down to com­
more and continue the scan of the nasal plete a full scan of the entire cornea in a
cornea, limbus and conjunctiva. similar manner. If corneal staining was
Repeat this examination on the other eye. detected earlier, narrow the slit beam and
At this point if the patient is wearing soft assess the depth of the staining.
lenses, remove them. Instill a drop of fluores­ Finally, widen the slit to form a broad
cein into each eye. Introduce the cobalt blue beam, move the illumination arm nasally,
filter into the illumination system and open and reduce the magnification to low or
the beam to its full width. Place the yellow medium. Evert the upper eyelid by
Wratten filter in front of the observation approaching from the temporal side. Hold
system. Increase the illumination using the the everted lid and observe the upper palpe­
slit lamp rheostat. Using low to medium bral conjunctiva. Assess the level of injec­
magnification, scan the ocular surface and tion, smoothness, presence of papillae,
estimate the completeness of the tear cover­ follicles or concretions. Lower the eyelid and
age of both cornea and conjunctiva following repeat these observations on the other eye.
a blink. Look for any epithelial basement Record the findings, preferably on a record
membrane irregularities, particularly near card designed for this purpose. This saves
vessel ingrowth around the periphery of the time and acts as a prompt when needed. It is
cornea. If rigid lenses are worn, assess the useful to use grading systems such as the
fitting pattern of each lens. Scan the cornea example listed in Table 18.1. Such systems
for any apparent areas of staining. If any are provide quantitative information regarding
noted, increase the magnification and assess findings which would otherwise need to be
their density and size. Move back to low laboriously described. Record all informa­
magnification, and view the central cornea. tion, even that which seems normal, such as
Have the patient blink several times. Assess the visibility of corneal nerve fibres or vis­
the completeness of their blinking pattern. ibility of pigmented areas on the sclera.
Have the patient then hold their blink and Sketch any abnormalities and try to estimate
estimate the tear break up time. Repeat these size and location as well as shape. Careful
observations on the other eye. If rigid lenses record keeping will not only provide a base­
are worn, remove them. line against which to compare future find­
Decrease the width of the slit to that of a ings, but in this age of increased litigation
parallelepiped, remove the cobalt blue and proceedings against health care profession­
yellow filters and move the light beam to the als, they may prove extremely useful as sup­
temporal side. Increase the magnification to porting evidence when confronted by such a
medium or high (15-20X). Starting at the circumstance. Finally, it should be noted that
temporal limbus, scan the cornea slowly, the efficiency with which the entire biomi­
looking for irregularities in corneal thick­ croscopy procedure is carried out can be
Table 18.1 Expected slit lamp signs in contact lens wearers
Location Condition Type of lens wear
. -­ ----­ _.-~._,-
- -­
No Lens Wear Daily Wear Extended Daily Wear Daily Wear Extended
RGP Wear PMMA Soft Lens Wear
RGP Soft Lens
Palpebral Papillae Few papillae Few papillae Mild papillae Few papillae Mild papillae Mild papillae
conjunctiva and redness No redness No redness Mild redness Mild redness Mild redness Mild redness
Corneal Staining None 3 and 9 o'clock 3 and 9 o'clock 3 and 9 o'clock Chemical 5PK Dehydration
epithelium Foreign body Foreign body . Foreign body Dehydration Metabolic 5PK
Indentation Dimple veil-
from lens ad­ ing
herence
Microcysts None None Few" Few" Few" Numerous
Corneal Striae Absent Absent Possible" Probable" Possible" Possible"
stroma
Limbal vessel Absent < 0.5 mm < 1.0 mrn" < 1.5 mm" < 0.75 mm" < 1.0 mm"
ingrowth
Central cor- Absent Absent Possible" Present" Absent Absent
neal clouding
Corneal Increased Absent Absent Probable Present" Absent Probable"
Endothelium polymeg­
ethism
Folds Absent Absent Possible" Probable" Absent Possible"
" Indicates that the presence or absence of the condition is determined by lens Dk/L, patient metabolic needs, and/or the time of day of
the examination.
Procedure for the slit lamp examination of the anterior segment 383

Table 18.2 Suggested grading scale for recording or slit lamp signs
Sign/ Grade (0-4) Description
condition
Bulbar o No hyperaemia present. A few major vessels present but small vessels
conjunctival and capillaries are not filled and bulbar conjunctival region is essentially
injection clear and white.
1 Trace. Very slight hyperaemia of the bulbar conjunctival vessels. Several
minor vessels now apparent in bulbar region.
2 Mild. Hyperaemia of the bulbar conjunctiva and limbal region. Con­
junctival capillaries are apparent with filling of some limbal vessels.
3 Moderate. Hyperaemia of the bulbar conjunctiva obvious when viewed
with the naked eye; large numbers of capillaries and larger vessels of the
bulbar conjunctiva are apparent, although they may be localized to one
quadrant of the conjunctiva. Limbal vessels are filled and engorged,
although again, this may be localized to one area of the limbus.
4 Severe. Hyperaemia of the bulbar conjunctiva obvious when viewed
with the naked eye, all quadrants of the bulbar conjunctiva are signifi­
cantly injected with vessels of all sizes apparent. Limbal vessels are filled
and engorged around a majority of the limbus.
Palpebral o None. No papillae of follicles visible. Specular reflex from palpebral
conjunctival conjunctiva reveals the tissue surface to be smooth. Palpebral conjunc­
changes tiva appears clear and white, several major vessels may be apparent but
no capillaries.
1 Slight. No papillae or follicles visible and the specular reflex from the
palpebral conjunctiva reveals a slight unevenness across the tissue
surface. Palpebral conjunctiva has a slight pink hue, but no distinct
capillary engorgement is visible.
2 Mild. Small papillae or follicles visible and the specular reflex begins to
be appear slightly mottled due to the increased surface unevenness.
Palpebral conjunctiva has a light red hue; slight capillary engorgement
can be seen.
3 Moderate. Marked papillae and/or follicles across the majority of the
palpebral conjunctival surface. Specular reflection from the tissue is
broken up due to marked roughness of the tissue surface. Palpebral
conjunctiva has definite red hue, capillaries are engorged across the
majority of the lid.
4 Severe. Entire surface of palpebral conjunctiva is covered with large
papillae and/or follicles, specular reflection is essentially absent and is
only seen as bright pinpoints from the crests of the papillae. Palpebral
conjunctiva has a bright red hue; capillaries and larger vessels are less
distinct because of their engorgement and close proximity to each other.
Corneal 0 None. No fluorescein staining visible.
staining 1 Trace. Few localized discrete stipples, non-coalescing, in any corneal
area. Includes superficial foreign body staining.
2 Mild. Lightly coalescent punctate staining in any corneal area. Superfi­
cial, with no staining diffusion into the underlying stromal tissue.
3 Moderate. Densely coalescent punctate staining in any area of the
cornea. Slight diffusion of fluorescein into the underlying stroma after
several minutes.
4 Severe. Abrasion or erosion causing noticeable loss of the epithelial
substance over any area of the cornea. Marked and rapid diffusion of
fluorescein stain into the underlying stromal tissue.
384 Anterior ocular microscopy

Sign! Grade (0-4) Description


condition
Epithelial -0 None. No microcysts seen using retro-illumination,
microcysts 1 Trace. Fewer than approximately 50 microcysts over the central or
paracentral cornea. No overlying staining or surface anomaly.
2 Mild. More than approximately 50 microcysts over the central or
paracentral cornea. No overlying staining or surface anomaly.
3 Moderate. More than approximately 50 microcysts, tending to be coales­
cent. Surface anomaly manifesting itself as faint staining and/or dry
spots when fluorescein is used.
4 Severe. Very numerous, dense, coalescent microcysts, visible using
direct and retroillumination, with overlying staining and erosions.
Corneal o None. Normallimbal vessel dilation and advancement.
neovascular­ 1 Trace. Slight dilation and less than 1.5 mm of advancement into the
ization cornea in one quadrant only.
2 Mild. Slight dilation and less than 1.5 mm of advancement into the
cornea in more than one quadrant.
3 Moderate. 1.5 mm to less than 3.0 mm of advancement in any quadrant.
4 Severe. More than 3.0 mm of advancement in any quadrant.
Lens deposits o No surface deposits.
1 Presence of 5 or less, small «0.1 mm) individual deposits.
2 Presence of more than 5 small, individual deposits and/or one indi­
vidual deposit O.lmm to O.5mm in diameter.
3 Multiple deposits O.lmm to O.5mm in diameter or one individual
deposit larger than O.5mm in diameter.
4 Multiple deposits of O.5mm in diameter or larger.

enhanced if the practitioner knows in lamp (Fig. 18.38) (Khaw & Elkington, 1988).
advance the type of contact lens wear that the Before describing these two methods of achiev­
patient is using, since there are expected slit ing photographic slit lamp records in detail, it
lamp findings for the different types of lens is worth introducing a few terms for the
wear (Table 18.2). photographic neophyte.
Film speed or ASA rating is an indicator of
how much light a particular film type needs
18.5 SLIT LAMP PHOTOGRAPHY
to provide a correctly exposed image. 'Fast'
Since one of the main uses of the slit lamp is films are those which need lower amounts of
to enable assessment of the structures of the light for correct exposure, while 'slow' films
anterior segment as an aid to the detection need more light. ASA 1000 or ASA 400 films
and diagnosis of abnormal conditions, it is are considered fast, while ASA 25 or ASA 64
beneficial if a photographic record of any are considered slow. Other typical film
abnormalities can be made as an aid to speeds are ASA 100 and ASA 200. Fast films
identifying any changes over time. usually produce grainier images with lower
Ideally, a photographic slit lamp with an resolution than do slow films. Since the
inbuilt flash and camera mounts should be majority of changes to structures in the ante­
used for taking anterior segment photo­ rior segment are subtle, such as endothelial
graphs (Fig. 18.37). However, a less expen­ polymegethism or stromal striae, it is best to
sive alternative is to attach a camera and use the slowest film which still provides the
ordinary flash unit to a non-photographic slit correct exposure.
Slit lamp photography 385
the colour temperature of specific light
sources. Hence, films may be described as
'daylight' films or 'tungsten' films. Since
preferred films are often not matched to the
colour temperature of the light source
being used, specific filters can be used to
compensate for the difference in colour
temperature between film and light source
and provide a more realistic rendition of
the true colour of the subject being photo­
graphed. Daylight film should be used with
flash photography.
Figure 18.37 A dedicated photo slit lamp. Note Exposure describes the amount of light
the camera attached to the objective housing, and which is allowed to fall on the film. Expo­
the flash charging unit attached to the table. sure can be controlled by the shutter speed,
i.e. the fractions of a second that the camera
shutter is open, such as 1I60th or 1I125th.
Ideally, exposure times should be kept as
short as possible since even the slightest
eye movement will cause a blurred photo­
graph. Many of the anterior segment condi­
tions which clinicians would like to
photograph at high magnification cannot
be correctly exposed using the available
illumination of the slit lamp. Therefore,
dedicated photographic slit lamps have
in-built flash systems which provide suffi­
cient light to enable correct exposure with
very short shutter speeds.
Figure 18.38 A non-photo slit lamp adapted for Aperture size refers to the size of the iris in
photographic use. Note that the camera is the camera lens and, along with the shutter
attached to the eyepiece, and that the flash is
positioned to provide additional illumination. speed, controls the exposure of the film.
Aperture sizes are usually described in
'f-stops', where the larger the number, the
Film colour temperature describes the smaller the aperture size. It is important to
spectrum of light to which the film was note that large apertures also shorten the
designed to respond. All light sources emit a depth of field. However, in most commer­
range of wavelengths described by their cially available photo slit lamps, the camera
colour temperature. All these light sources is used without a lens, so that lens aperture is
may be described as 'white' in general terms, not a factor. In slit lamps where the camera
but, just as their colour temperatures differ, and lens are positioned behind the eyepiece
so does the composition of their emitted (e.g. the Holden-Zantos technique, described
spectra. Therefore, there are more blue wave­ later), changing the aperture setting on the
lengths in the apparently 'white' light emit­ camera lens changes the size of the field of
ted by a fluorescent light source than in the view but not the brightness. In this situation,
'white' light provided by an equivalent halo­ film exposure is controlled by camera shutter
gen light source. Colour films are matched to speed or flash intensity.
386 Anterior ocular microscopy
18.5.1 DEDICATED SLIT LAMP filters are used. Film exposure guidelines for
PHOTOGRAPHY photographic slit lamps have been published
elsewhere (Spivak, 1977; Zantos et al., 1980).
Slit lamps designed for photography have The camera mount for most photographic
two additional facilities - an in-built flash slit lamps is placed between the objective
system which allows a powerful beam of lens and the magnification tumbler of the
high intensity light to be generated and microscope; a right-angled prism or semi­
flashed through the slit beam as illumination reflecting mirror is put into the optical path­
for the photograph, and a camera mount way of the microscope to divert the images to
somewhere on the slit lamp microscope. the camera during photography but not dur­
The flash system usually uses a semi­ ing normal viewing. A bayonet-type mount
reflecting mirror in the illumination system is usually provided to which the camera
of the slit lamp to allow both the regular body (without lens) is mounted. The mount
incandescent/halogen light and the flash is designed in such a way that the image
tube generated light to follow identical path­ seen by the observer in the eyepieces is
ways through the slit aperture. Therefore, all focused simultaneously on the film plane of
structures illuminated by the regular light the camera, ensuring correct focus if the
source will be illuminated in the same pat­ microscope eyepieces have been correctly
tern by the flash tube. However, the inten­ adjusted for the observer. Drawbacks to this
sity of the flash will be very different, and type of mounting system is that the magnifi­
this is usually controlled
. by
. a switch on the cation of the camera image is fixed, and
flash generator housing. Typically there are relatively low (approximately 3X), since the
four settings for flash intensity, and the best light rays from the objective lens are diverted
one to use will vary according to the reflec­ before entering the magnification tumbler of
tance of the structure being viewed and the the microscope head. This limitation can be
slit lamp set-up chosen. overcome by mounting the camera and lens
Thus, a photograph of the sclera and bul­ over the eyepiece of the microscope, as
bar conjunctiva will require a low intensity described by Zantos and Holden. This tech­
flash setting, whereas a high magnification nique takes full advantage of the increased
shot of an optic section of the cornea will magnification available through the magnifi­
require a high flash intensity. The optimum cation tumbler of the microscope and alter­
flash settings for a particular photographic nate high powered eyepieces. A special
slit lamp are best determined by trial and adapter must be fabricated which clamps
error, and they will depend on the film ASA over the eyepiece and screws into the filter
rating and the structure being photographed. thread of the camera lens to utilize the
However, a good starting point is to use 200 Holden-Zantos technique, but this inconve­
ASA film and the lowest intensity flash set­ nience is far outweighed by the versatility
ting for diffuse and broad beam, low magni­ and enhanced convenience that variable
fication shots; and the highest flash intensity magnification and through-the-camera
setting for low magnification parallelepiped focusing bring to any photographic slit lamp.
and optic section photographs, and for all
medium to high magnification shots. It 18.5.2 PHOTOGRAPHY USING A
should be noted that good fluorescein photo­ NON-PHOTOGRAPHIC SLIT LAMP
graphs of corneal staining or rigid lens fitting
patterns can only be taken using a slit lamp While a dedicated photo slit lamp is the
equipped with a high output flash tube preferred method of taking photographs of
when the cobalt blue and yellow barrier the anterior ocular segment, quite useful
References 387
photographs can be obtained by using an ence with the balance and normal function­
auto-exposure camera mounted over the eye­ ing of the slit lamp. Cameras with remote
piece of a regular non-photographic slit heads are ideal for this use. Contemporary
lamp. The 35 mm camera used should have video cameras have sufficient gain to allow
an aperture priority, auto-exposure mode. good images to be captured with the incident
The major limitation of this method of pho­ light from the slit beam, negating the need
tography is the lack of flash, and hence a for additional light sources or low illumina­
relatively high speed film should be used tion 'surveillance-type' cameras. The other
(e.g. 400 ASA) in conjunction with the high­ parameter which should be considered is the
est light settings of the slit lamp. Ideally, the resolution capabilities of the camera. Black
slit lamp should have a halogen or other high and white cameras generally have higher
output light source, and the film colour tem­ resolution capabilities than colour cameras
perature should be matched to the light but, for the majority of clinical applications,
sources. Since the exposure is automatic and the advantages of a colour image far out­
controlled by shutter speed, a good estimate weigh the loss of resolution. It should be
of correct exposure can be judged by listen­ noted, however, that while the majority of
ing to the shutter open and close. For shutter colour cameras have sufficiently high resolu­
speeds of slower than 1I8th second, move­ tion to demonstrate lens fitting performance
ment of the eye will blur the photograph. and front surface deposits, oniy the most
Faster shutter speeds can be produced by expensive have the capability of capturing
increasing the light intensity of the slit lamp high resolution images such as the endo­
or by increasing the width of the slit beam. thelial mosaic. The interested reader IS
This technique is limited to white light directed to recent publications regarding
photographs, and cannot provide high mag­ clinical videography both with and without
nification images of optic sections, but it can slit lamps (Robboy & Hammack, 1987).
provide reasonable quality photographs of
many features of the anterior segment which
are of interest to the contact lens clinician. REFERENCES
Berliner, M.L. (1943) Biomicroscopy of the Eye. Paul
18.5.3 VIDEO PHOTOGRAPHY WITH TIfE B. Hoeber, Inc.
Brandreth, R.H. (1978) Clinical Slit Lamp Biomicros­
SLIT LAMP
copy, Blaco Printers, Inc.
The recent availability of relatively inexpen­ Butler, T.H. (1924) Focal illumination of the eye.
sive video cameras and recorders has made Br. J. Ophthal., 8(12), 561-90.
Cavanagh, H.D. et al. (1990) Confocal microscopy
the rather appealing concept of recording slit
of the living eye. CLAO L 16, 6~73.
lamp observations on video tape a clinical Doggart, J.H. (1949) Ocular Signs in Slit Lamp
reality. The major advantage which such a Biomicroscopy. C V. Mosby Co., St Louis.
system provides over 35 mm photography is Goldberg J.B. (1984) Biomicroscopyfor Contact Lens
the ability for immediate playback. This is Practice: Clinical Procedures. The Professional
particularly useful in describing the appear­ Press, Inc.
ance of many conditions such as lens depos­ Holden, B.A., Zantos, S.G. and Jacobs, K.J. (1978)
its or lens fitting to patients or colleagues. The Holden-Zantos technique for endothelial and
high magnification slit lamp photography. Bausch
Video cameras are usually attached to slit
& Lomb.
lamps using the same beam-splitters incor­ Khaw, P.T. and Elkington, A.R. (1988) Slit-lamp
porated for 35 mm photographic attach­ photography made easy by a spot metering
ments. Ideally the camera should be compact system. Br. J. Ophthalmol., 67, 63-6.
and light-weight to minimize any interfer- Koester, CJ. and Roberts, CW. {199m Wide-field
388 Anterior ocular microscopy
specular microscopy. In Non-invasive Diagnostic biomicroscope. j. Am. Optom. Assoc., 58(4),
Techniques in Opthalmology (ed. B.R. Masters), 290-2.
Springer-Verlag, New York, pp. 91-121. Schmidt, T.A.F. (1975) On slit-lamp microscopy,
Long, W.F. (1988) How to choose and use a Doc. Ophihalmol., 39, 117-53.
photoslitlamp. Review of Optometry, Septem­ Spivak, T.W. (1977) Photography through the
ber, 69-73. Nikon slit lamp biomicroscope. Rev. Opiom.,
Martonyi, CL. (1989) Photographic slit-lamp November, 56-60.
biomicroscopes. Ophthalmology - Instrument Stockwell, H.J. (1983) Considering the slit lamp and
and Book Issue, pp. 6-19. biomicroscope, The Optician, June 10, 16-20.
Masters, B.R. and Kino, G.S. (1990) Confocal Terry, J.E. and Kercheval, D.B. (1979) Interpretive
microscopy of the eye. Noninvasive Diagnostic biomicroscopy. J. Am. Optom. Assoc., 50, 79J-..­
Techniques in Opthalmology (ed. B.R. Masters), 803.
Springer-Verlag, New York, pp. 152-71. Zantos, S. G. (1984) Slit Lamp Examination of Con­
Mayer, D.J. (1984) Clinical Wide-Field Specular tact Lens Wearers. Bausch & Lomb.
Microscopy. Balliere Tindall. Zantos, S.G. and Pye, D.C. (1979) Clinical photog­
Muller, O. Ocular Examination with the Slit Lamp. raphy in optometric practice, Aust. j. Opiom.,
Carl Zeiss Publication I<3O-1l5-e. 62,279-85.
Nom, M.S. (1983) External Eye: Methods of Exami­ Zantos, S.G., Holden, B.A. and Pye, D.C. (1980)
nation. Scriptor Publisher ApS. A guide to ocular photography with the
Robboy, M.W. and Hammack, G.G. (1987) Video­ Nikon photo slitlamp. Aust. J. Optom., 63,26­
tape recording system for use with the slitlamp 32.
PART 2 REAL-TIME CONFOCAL MICROSCOPY OF THE IN VIVO HUMAN CORNEA

B.R. Masters, A.A. Thaer and a.-c. Geyer

18.6 ABSTRACT and devoid of any digital image processing


enhancements.
A new, non-applanating, real-time slit scan­
The primary instrument for the examina­
ning confocal microscope is described for the
tion of the cornea is the slit lamp. This
in vivo examination of the living human eye.
important instrument was developed in its
This confocal microscope produces real-time
present form by Gullstrand who developed
video images of the in vivo human cornea. In
the system to condense the illumination
contrast to other confocal microscopes
light into a slit aperture and then project
designed for in vivo ocular imaging, which
the image of the slit into the eye. The
are based on either the Nipkow disk (pin­
modern slit lamp uses a narrow slit which
holes) or the modified wide-field specular
removes the light scattered from adjacent
microscope (a photographic system which is
tissue and provides a degree of optical
not real-time), this new instrument produces
sectioning. However, the slit lamp pro­
images which are captured as single video
vides an oblique view across the cornea
frames which have superior contrast. This
and provides limited contrast of cellular
new real-time slit scanning confocal micro­
components.
scope produces en face, high contrast, high
A more recent development for imaging
resolution, images of superficial epithelial
the cells of the corneal endothelium is the
cells, wing cells, epithelial basal cells, corneal
specular microscope. If the illumination and
innervation, nuclei of stromal keratocytes,
the observation axes are adjusted correctly,
and the cell bodies of the stromal keratocytes
then the specular reflection from the bound­
in the posterior stromal region.
ary cf the corneal endothelium and aqueous
humor will show the cell outlines of the
18.7 INTRODUCTION
endothelial cells. These cells were first
observed and described by Vogt who used
This chapter presents the principles of in vivo the modified Gullstrand slit lamp to examine
coniocal microscopy and applications to the the human corneal endothelium. There are
observation of the in vivo human cornea. A several problems with the specular micro­
new, real-time, scanning slit confocal micro­ scope; the surface reflection from the tear­
scope is described. Examples are presented corneal interface degrades the contrast of the
of confocal microscopic examination of the endothelial cells. An edematous stroma will
normal human cornea which illustrate the also reduce contrast of the endothelial cells.
resolution and the contrast of single video The field of view is small with the standard
frames. The photographs are unprocessed specular microscopes. The development of a
Contact Lens Practice. Edited by Montague Ruben and Michel Guillon.

Published in 1994 by Chapman & Hall, London. ISBN 041235120 X

390 Anterior ocular microscopy


scanning wide-field specular microscope by nique, real-time scanning slit confocal
Koester provided high-contrast, wide-field microscopy, with the capability to over­
views of about one square millimeter of come the limitations of the slit lamp and
corneal endothelium. This microscope the specular microscope. In vivo real-time
scanned a narrow slit over a larger area of the scanning sli t confocal microscopy provides
endothelium and recorded the composite a new approach to the observation of the
image on film. human eye.
Both the slit lamp and the specular
microscope are limited in their ability to 18.8 CONFOCAL MICROSCOPY
observe cellular and subcellular details
throughout the full thickness of the cornea. This section provides a brief introduction to
In this chapter we introduce a new tech- the principles of confocal microscopy and

Figure 18.39 In vivo confocal microscopic en face image of human cornea. The U-matic tape was
stopped on a single video frame, using the pause mode, and a 35 mm film camera was used to
photograph the television monitor to produce the image. No digital or analog image enhancement or
processing was used. There was no frame averaging applied to the video images. The photograph
shows the raw data directly photographed from the screen of the television monitor. This figure shows
a single video frame of the superficial cells of the corneal epithelium. This cell layer, which is just below
the tear film, has the following characteristics: the cell nucleus appears as a highly reflecting bright oval
which is surrounded by a darker band; the focal plane is set to optically section the most superficial
cells which appear brightest; cells slightly below this focal plane appear darker. Some of these
superficial epithelial cells are in the process of desquamation. Objective 25X, NA 0.6.
Confocal microscopy 391

Figure 18.40 In vivo confocal microscopic en face image of human cornea. This optical section is in the
plane of the wing cells which are located between the superficial epithelial cells and the basal epithelial
cells of the human cornea. The image of these cells is characterized by the bright cell nuclei which are
devoid of the darker band which is observed in the superficial epithelial cells. Objective SOX, NA 1.0.

describes the basic types of confocal micro­ The confocal microscope has a scanning
scopes (Wilson, 1990; Pawley, 1990). First we mechanism such that the narrow field of
answer the question: What is a confocal illumination, which is identical to the
microscope? detection field, is scanned over the speci­
In 1957 Minsky filed a patent application men to generate a composite image of it.
for a confocal microscope which elucidated This is the principle of confocal micros­
the principles of operation upon which copy.
today's commercial confocal systems work. In a standard microscope - which is
A confocal microscope has two sets of pin­ operated in the epi-illumination mode ­
holes, or apertures; which are arranged in the lateral and axial resolutions are func­
conjugate, or confocal, planes. One pinhole tions of the wavelength of the light used for
or slit is placed in the illumination path the illumination and of the numerical aper­
and another set is placed in the detection ture of the microscope objective. In a confo­
path. These sets of conjugate apertures only cal microscope there is an increase in both
illuminate a narrow field of the specimen the axial and lateral resolutions of the
and detect the back-scattered and reflected microscope objective. While the increase in
light from the same narrow field of view. the axial resolution of a confocal micro­
392 Anterior ocular microscopy

Figure 18.41 In vivo confocal microscopic en face image of human cornea. The optical section is in the
plane of the basal epithelial cells. The image shows the bright cell borders and the darker cell interiors.
The cell borders appear thick since there is a high degree of invagination between the adjacent cells.
Water immersion microscope objective sox, NA 1.0.

scope is important, it is the increase in the degraded in sharpness and contrast as in a


lateral resolution which is of prime impor­ standard light microscope. The second
tance for imaging thick specimens. The advantage is that thick sections of living
increased lateral resolution gives the confo­ tissues and cells can be optically sectioned
cal microscope the unique ability to opti­ for observation and subsequent three­
cally section the specimen. It is important dimensional reconstruction.
to note that the axial resolution is about There are several types of confocal micro­
three times the lateral resolution. scopes. Some designs are based on a rotating
There are several advantages of the disk which contains many sets of pinholes
optically-sectioning capability of a confocal arranged in a spiral pattern. This design is
microscope as compared to the standard called a Nipkow disk-based microscope;
microscope. The first advantage is that such designs are further divided into
images are obtained with enhanced con­ tandem-scanning confocal microscopes and
trast; this is because the confocal micro­ one-sided Nipkow disk-based confocal
scope excludes light scattered from optical microscopes.
sections above and below the section in the The tandem-scanning confocal microscope
focal plane, therefore the focal plane is not uses two sides of the disk which contains
Confocal microscopes for in vivo clinical examination of the cornea 393

Figure 18.42 In vivo confocal microscopic en face image of human cornea. This is an optical section of
the basal epithelial cells similar to that shown in Figure 18.41. The wider field is obtained with the use
of a microscope objective with a lower magnification. Water immersion microscope objective 25X, NA
0.6.

two sets of conjugate pinholes. One set of They are also real-time confocal systems.
pinholes is in the illumination path, and the
other is in the detection path. As the Nipkow 18.9 CONFOCAL MICROSCOPES FOR IN
disk rotates, the full field of view is swept out VIVO CLINICAL EXAMINAnON OF THE
by both the illumination spots of light and CORNEA
the detec.tion spots. Since both sets of pin­
holes are conjugate to each other the system The advantages of confocal microscope
is a confocal microscope. It is called a imaging systems as compared to standard
tandem-scanning confocal microscope since microscopy include enhanced x, y and z
both sets of pinholes rotate in tandem. resolution and improved contrast of the
Another type of Nipkow disk-based images. These factors permit the optical
microscope uses the same set of pinholes sectioning of thick living tissue. The result­
on one side of the rotating Nipkow disk for ing images are sharp and free from signifi­
both the illumination path and the detec­ cant blur due to contributions of scattered
tion path. This is the basis of the so-called light from adjacent sections. In contrast to
one-sided, real-time, confocal microscope. slit lamp observations, and other standard
Other designs are based on conjugate slits. imaging techniques, confocal microscopy is
394 Anterior ocular microscopy

Figure 18.43 In vivo confocal microscopic en face image of human cornea. This optical section shows the
corneal nerves in the anterior stroma just below Bowman's membrane. The large bifurcating nerves are
easily observed. Objective 25X, NA 0.6.

a new paradigm for ocular visualization. Confocal microscopy has been used to
Confocal microscopy permits real-time observe living specimens and provides high
observation across the full thickness of the resolution, high contrast images with good
cornea (Lernp et al., 1986). The use of confo­ optical sectioning capability (Koester and
cal microscopic imaging systems permits Khanna, 1989; Koester and Roberts, 1990;
two-dimensional, high-resolu tion imaging Florakis et al., 1992; Koester et al., 1992;
of the cornea and ocular lens which is sharp Koester, 1993). Confocal microscopy has also
and in the plane of the cornea perpendicu­ been used to examine the ex vivo eye.
lar to the optic axis (Masters, 1990a; Mas­ Tandem-scanning confocal imaging of ill
ters and Kino, 1990) and in the plane of the vivo ocular tissue has been reported. A one­
cornea perpendicular to the optic axis. The sided Nipkow disk confocal imaging system
advances made possible with ocular confo­ has been used to observe the excised rabbit
cal microscopy are not limited to corneal cornea and ocular lens.
imaging. A major development in oph­ The in situ observation and visualization
thalmic imaging has been the adoption of of the ocular lens with confocal microscopy
confocal microscope imaging systems to is a major technological advance in ocular
image ocular structures (Beuerman et al., lens imaging. The feasibility of confocal
1992; Wegener et al., 1992; Masters, 1988). microscopy is to produce high contrast,
A new real-time scanning slit system for clinical examination of the cornea 395

Figure 18.44 In vivo confocal microscopic en face image of human cornea. This optical section shows the
corneal nerves in the anterior stroma just below Bowman's membrane. Objective 25x, NA 0.6.

high resolution images across 1.7 mm of ing cone of the microscope objective, limits
ocular tissue. The in situ observation of the its use in the clinic and tends to modify the
lens capsule, lens epithelium.. lens suture normal structure of the cornea.
and detailed structure of the lenticular
fibers provides a new approach to lens and 18.10 A NEW REAL-TIME SCANNING SLIT
cataract research.
SYSTEM FOR CLINICAL EXAMINATION OF
The development of the wide-field specu­ THE CORNEA
lar microscope by Koester was limited by the
low numerical aperture of the applanating The in vivo observation of the living cornea
cone objective (Florakis et al., 1992; Koester et by the technique of confocal microscopy pro­
al., 1992; Koester, 1993). Recent develop­ vides en face images of high contrast and
ments of a high numerical aperture for the resolution. Slit-based confocal systems col­
wide-field specular microscope have resulted lect more light from the eye as compared to
in a confocal microscope for the eye. How­ Nipkow disk pinhole confocal microscopes.
ever, this confocal microscope has two disad­ Therefore the light incident on the cornea
vantages - it is a photographic system (and can be reduced.
therefore not real-time) and the microscope We describe a new flying slit, real-time
objective flattens (applanates) the cornea. confocal microscope which has unique imag­
The corneal flattening, due to the applanat­ ing characteristics for in vivo human confocal
396 Anterior ocular microscopy

Figure 18.45 In vivo confocal microscopic en face image of human cornea. This optical section shows a
large corneal nerve in the anterior stroma. The bright oval objects are the nuclei of the stromal
keratocytes. Objective 25X, NA 0.6.

microscopy (Thaer et al., 1990; Masters and tion. Filters are inserted in the lamp
Thaer, 1993a,b,c). This system uses non­ housing to remove both short ultravio­
applanating microscope objectives and the let and infrared light. The light levels
illumination is provided by a halogen at the cornea are less than those from
lamp. the wide-field specular microscope.
This new real-time confocal microscope The light intensity is sufficiently low
has several unique and advantageous fea­ so that after 30 minutes of continuous
tures which differentiate it from other in vivo examination of a subject's cornea the
confocal imaging systems. subject has no after image from the
microscope.
1. The standard light source is a halogen Other in vivo confocal microscopes
lamp which has many advantages over for corneal examination use a mercury
the mercury or xenon arc lamps used in or xenon arc lamp as the source of
other confocal designs. The mercury or illumination. These sources are much
xenon arc lamps have the problem of brighter than the new real-time system
arc jitter which results in changing described in this chapter.
illumination intensity. The halogen 2. The microscope uses standard non­
lamp offers steady levels of illumina­ applanating microscope objectives with
A new real-time scanning slit system for clinical examination of the cornea 397

Figure 18.46 In vivo confocal microscopic en face image of human cornea. This optical section shows the
oval shaped nuclei of the stromal keratocytes. Objective 25X, NA 0.6.

RMS threads which are readily inter­ detector. Scanning of the image of the
changeable. This permits the use of sev­ slit over the plane of the cornea is
eral different microscope objectives and accomplished with the use of an oscillat­
provides the capability to vary the field ing two-sided mirror which is used for
of view and the magnification. The stud­ beth scanning and descanning. The
ies we describe in this chapter used a scanning bilateral system based on a
Leitz sox, NA 1.0 water immersion double-sided mirror was published over
objective. Other studies which required 24 years ago (Svishchev, 1969). However,
a larger field of view used a Leitz 2SX, there are alternative methods of scan­
NA 0.6 water immersion objective. ning and des canning which may have
Other in vivo confocai systems are advantages over the bilateral scanning
limited to a built-in objective which is system which is described here.
not removable. Some microscopes 4. The detection system consists of an
require custom designed microscope intensified video camera with video out­
objectives which are of high cost. put to a Sony U-Matic tape recorder. The
3. The confocal microscope is based on two PAL video format provides 62S lines.
sets of adjustable conjugate slits which The U-Matic video tape recorder pro­
are located in conjugate planes in front vides a high bandwidth. In parallel with
of the halogen source and in front of the the video recording there is a video
398 Anterior ocular microscopy

._~''::::::-
._-------~
-~-_._-

------­
_------­
_..

Figure 18.47 In vivo confocal microscopic en face image of human cornea. This optical section shows the
polygonal corneal endothelial cells. This image is 500 microns below the surface of the cornea (Figure
18.39). The focal plane is at the interface between the cornea and the aqueous humor. Objective 2Sx,
NAO.6.

monitor in order that -the operator can vides three-axis control. Immediatelv
observe the confocal images of the sub­ before the corneal examination, the eye
ject's eye in real time. was anaesthetized. The subject was
5. Hard copy of individual video frames is seated at the ocular examination table
obtained immediately with a video and a drop of methylcellulose gel was
printer. Individual video frames can also applied to the tip of the microscope
be digitized. These digital images can be objective; the microscope was slowly
registered and preprocessed by digital positioned to make contact between the
image processing techniques for further gel and the cornea. The tip of the micro­
analysis and quantification. However, in scope objective never touches or flattens
this chapter all the figures are photo­ the corneal surface; a layer of gel is
graphs of single video frames of the raw always present between the objective
data - no analog or digital enhancements and the eye. The z-axis position of the
were used. confocal microscope was controlled by
6. The real-time confocal microscope is manual movement of the microscope
mounted on a special ophthalmic head stage with a joystick. Alternatively, a
rest (Leitz). The microscope stand pro­ computer-controlled stepping motor
A new real-time scanning slit system for clinical examination of the cornea 399

Figure 18.48 In vivo confocal microscopic en face image of human cornea. This optical section shows the
polygonal corneal endothelial cells. This image is 500 microns below the surface of the cornea (Figure
18.39). The numerous small black particles are due to inflammation of the eye. The focal plane is at the
interface between the cornea and the aqueous humor. Objective 25x, NA 0.6.

scanned the focal plane Of the confocal section can be manually adjusted, or auto­
microscope across the full thickness of mane continuous scanning (Z-Scan attach­
the cornea. ment) oi the depth of the focal plane can be
7. The microscope is built from a modular initiated. Both scan time and scan depth
system. This design concept permits can be adjusted.
rapid modification of the system to suit The second configuration provides a quan­
different applications. The confocal titative measure of the reflected and back­
microscope can be used in several scattered light in a given focal plane. This
modes. Z-Scan attachment is used for the confocal
photometric recording of scattered light and
The first configuration is capable of obtain­ fluorescence intensity profiles through the
ing real-time (video rates) en face optical full thickness of the cornea. The Z-Scan can
sections across the full-thickness of the in also be used for quantitative photometric
vivo human cornea. The thickness of the measurements of fluorescence and/or scatter­
optical section can be modified by adjust­ ing in the anterior chamber.
ing the width of the two confocal slits in This configuration can also be used for
the microscope. The depth of the optical fluorescence measurements. Both autofluo­
400 Anterior ocular microscopy

Figure 18.49 In vivo confocal microscopic en face image of human cornea. This optical section shows the
polygonal corneal endothelial cells. This image is 500 microns below the surface of the cornea. The
numerous small black particles are due to inflammation of the eye. The focal plane is at the interface
between the cornea and the aqueous humor. This image shows a different field of view than in the.
previous figure. Objective 25X, NA 0.6.

rescence and fluorescence from externally The figures are typical raw data. The figures
applied fluorescent dyes or fluorogenic sub­ show the superficial cells (Fig. 18.39), the
strates can be measured (Thaer et al., 1990). wing cells (Fig. 18.40) and two fields of the
This configuration is suitable for the in vivo normal human basal epithelium (Figs. 18.41
measurements of cellular metabolism based and 18.42). High contrast photographs of
on the principles of non-invasive redox fluo­ corneal nerves are shown in Figs. 18.43­
rometry. 18.46. The corneal endothelium is shown in
Figs. 18.47-18.50.
The full three-dimensional reconstruction
18.11 RESULTS
of the ex vivo rabbit cornea is the only figure
The following in vivo human corneal confocal in this chapter which is not from the in vivo
microscopic images were made by photo­ human cornea (Fig. 18.51). The optical sec­
graphing the single still video frames on the tions which were used for the reconstruction
monitor with a camera. These figures are were obtained with a laser scanning confocal
photographs of single video frames. There microscope. The inclusion of this figure in
was no image processing or frame averaging. the chapter is to present the future of in vivo
Discussion 401

FiSU'" 18.50 In viv. eonfocel microKOpic ... f.<e image 01 human come" The image shows a ..pon of
the comeal endothelium with endothelial defects. The three black regions represent endothelial defects.

sufficient light collection efficiency so that


three-dimensional visualization of ocular single video frames show high contrast,
imaging. sharp images of the living in vivo human eye.
There is no need for video frame averaging
or digital image processing techniques to
18.11 DISCUSSION
enhance the contrast of the single video
A new, non-applanating, non-invasive real­ images. This is not the case for confocal
time, slit-scanning confocal microscope has microscopes designed on Nipkow disk (pin­
been developed for the observation of the in hole) configuration - they pass such low light
vivo human eye. This chapter illustrates the from the eye to the camera that frame averag­
use of the instrument for the observation of ing and digital image processing enhance­
the cornea. The high contrast and large light ments are required.
collection efficiency of the microscope are The light collection efficiency of this flying
due to the use of confocal slits in conjunction slit in vivo ocular confocal microscope is due
with a high numerical aperture objective. to the use of slits (instead of pinholes) and a
The microscope does not contact the cor­ microscope objective with a numerical aper­
nea nor does it applanate the surface of the ture of 1.0. These two design implementation
cornea: a polymer gel is applied in a layer results are bright, sharp, high-contrast, high­
between the ocular surface and the tip of the resolution confocal images of the in vivo
microscope objective. This instrument has
402 Anterior ocular microscopy

Figure 18.51 Three-dimensional volume reconstruction of the full thickness of the ex vivo rabbit cornea
shown in an isometric view. The endothelium is on the top of the figure and the epithelium is on the
bottom. The linear bright feature on the front face is a nerve fiber. The thickness of the reconstruction is
400 microns and the epithelial thickness is 40 microns. This figure demonstrates that thin optical
sections of living cornea can be reconstructed in a computer to yield a three-dimensional volume
visualization of the cornea.

human cornea. The images are obtained in the depth of focus on the z-axis. In addition,
real time and are stored on magnetic video a slit confocal system passes many times
tape (Sony V-Matic). more light from the subject's eye to the
The advantages of slits over confocal detector as compared to pinhole type confo­
microscope systems based on Nipkow disks cal systems. Another advantage of adjustable
containing pinholes are the following. Slits slits is that the scanned field of view and the
have the possibility of continually adjusting light intensity, incident on the field of view,
Discussion 403
can be easily varied. Finally, the light inten­ the basal epithelial cells. Jester et al. have
sity incident on the subject's eye is signifi­ written several papers in which they cat­
cantly less with a slit confocal microscope as egorically state their inability to image basal
compared to a Nipkow disk pinhole system. epithelial cells in the normal in vivo cornea
This new in vivo confocal instrument has (jester et al., 1990; 1991; 1992). They state,
unique advantages over other confocal sys­ 'Reflections from cells below the surface are
tems. The bright, high contrast confocal not normally resolved but can be imaged in
images of the wing cells and the basal cells certain disease states.' In another paper
demonstrate its unique optical characteris­ Jester et al. state, 'By in vivo confocal micros­
tics. The low reflectivity of the wing and copy, only the first layer of superficial cells
basal epithelial cells in the normal human can normally be imaged in the intact, living
cornea presents a low contrast object for eye.' They then state, 'In vivo confocal optical
confocal microscopy. The high rejection of sections do not detect any reflections from
stray light and narrow depth of field, coupled the wing and basal cell layers in the intact,
with the high numerical aperture microscope normal living eye.' Finally, in a recent paper
objective (NA 1.0), results in the ability of the published in Investigative Ophthalmology and
instrument to clearly image these cell layers Visual Science, Jester et al. wrote on their in
in the live, normal human cornea. vivo confocal observations of the normal cor­
The clear advantage of slit-scanning confo­ nea in the live rabbit, 'Below the superficial
cal microscopes (the new microscope dis­ epithelium, wing and basal epithelial cells do
cussed here, and the Koester wide field not appear to reflect light above background
specular microscope) for ophthalmic diag­ levels. Because the cornea is a transparent
nostics and basic eye research is best appre­ tissue and optically configured not to scatter
ciated when imaging the in vivo human basal light, it is to be expected that some structures
epithelium in the anterior cornea. The real­ of the cornea may not be visualized. . .
time confocal microscope described here pro­ under pathologic and ex vivo conditions,
vides high-contrast, high-resolution images of images of the wing and basal epithelial cells
both the wing and basal epithelial cells in the can be "detected.' Although they averaged
normal in vivo human eye. The modified wide­ video frames they were still unable to image
field specular microscope of Koester has the the normal in vivo basal epithelial cells. They
ability to image basal epithelial cells in the used a Nipkow disk confocal microscope
normal in vivo human eye. however, the produced by the Tandem Scanning Corpora­
Koester microscope uses a 35 mm film camera tion, Inc. Furthermore, they 'explain' their
as the detector and is therefore not real-time. inability to image basal epithelial cells in the
Ocular confocal microscopes designed on normal in vivo eye by stating that the normal
the slit-scanning principle provide sensitiv­ epithelium of the cornea is transparent and
ity and resolution that has not been possible should not reflect light. Our results on many
with Nipkow disk based microscopes. The human subjects, all made on the normal
advantages of a slit-scanning confocal system living cornea, are at variance with the results
over a Nipkow disk pinhole system are illus­ of Jester et al. Koester has also published
trated below. results indicating the ability of his wide-field
A Nipkow disk (pinhole) real-time confo­ specular microscope (slit confocal micro­
cal microscope (Tandem Scanning Corpora­ scope) to image the basal epithelial cells in
tion, Inc.) did not have the capability to the normal in vivo eye. Our results, and those
image basal epithelial cells in vivo in the of Koester, demonstrate the advantage of a
normal eye; however, in swollen or patho­ slit-system-based confocal microscope for
logical specimens it was capable of imaging the in vivo examination of the eye.
404 Anterior ocular microscopy
The in vivo real-time scanning slit confocal studied. Also, the migration and adhesion of
microscope can be utilized for two other cells to the intraocular lens could be investi­
types of imaging. The set of optical sections gated with this real-time confocal microscope.
of the full thickness of the cornea provides a In vivo studies of the cornea and the tear
new way to visualize the cornea. The set of film are current themes of investigation. A
optical sections can be constructed into a full major use of this new in vivo confocal micro­
three-dimensional visualization (Masters, scope is the investigation of the etiology and
1991a,b,c; Masters and Paddock, 1990; Mas­ treatment of the dry eye. The new imaging
ters, 1993; Masters and Farmer, 1993). This is capability permits new studies of tear secre­
illustrated in Fig. 18.51. Once the two­ tion, the renewal of the ocular surface and
dimensional optical sections are formed into the pathophysiologic conditions which
the three-dimensional image in the com­ result in the dry eye.
puter, it can be viewed from any angle and Another area of investigation is the
the transparency of the view varied. There­ anatomy and physiology as related to contact
fore the observer could look inside the cor­ lens wear. The real-time in vivo confocal
nea from any angle. microscope could follow the time-dependent
The second type of confocal imaging is changes on both the contact lens and the
redox imaging. This method is based on the cornea. The real-time system could follow
intrinsic fluorescence from the reduced pyri­ changes in corneal innervation, renewal of
dine nucleotides. Several papers illustrate the corneal epithelium, changes in the cor­
this method of non-invasive metabolic imag­ neal stroma and changes which occur in the
ing of the cornea (Masters, 1990b; Masters et corneal endothelium.
al.,1993). The effects of corneal laser surgery on the
migration and distribution of stromal kerato­
18.13 APPLICATIONS OF CONFOCAL
cytes is currently under investigation. The
MICROSCOPY TO CONTACT LENS SCIENCE
real-time in vivo confocal microscope is
capable of imaging the keratocyte nuclei, and
What are some of the applications of this therefore keratocyte densities prior to and
new real-time slit scanning confocal micro­ after laser refractive surgery. The investiga­
scope in basic eye research and in clinical tion of the time dependence of the post
ophthalmology? surgery stromal haze, as well as its quantita­
The instrument is currently configured for tive analysis with respect to size and light
real-time confocal microscopy of the cornea. scattering ability, is currently under way on
With the introduction of a microscope objec­ human subjects.
tive with a longer free-working distance it The use of the real-time in vivo confocal
would be possible to image the in vivo microscope for the diagnosis and examina­
human lens. Applications to the ocular lens tion of corneal dystrophies, corneal erosions,
include the investigation of the natural pro­ superficial punctate keratopathy, corneal
gression of cataracts in animal models and in inflammation and corneal infection are clini­
human subjects. The investigation of drugs cal applications in which new technology
which cause retardation of cataract develop­ will lead to better diagnosis, treatment and
ment could be monitored in vivo with the use understanding of ocular disease.
of the confocal microscope.
ACKNO~EDGEMENTS
The in vivo study of intraocular lenses in
humans could be performed. The effects of This work was supported by a grant from
laser surgery on the physical integrity and the NIH, National Eye Institute, EY-06958
optical quality of the intraocular lens could be (BRM). The authors acknowledge the
References 405
co-operation from The Institute for Medical drial redox state of rabbit cornea measured
Vision Aid and Helmut Hund GmbH. noninvasively with an optically sectioning
redox fluorometer microscope, in The Cornea:
Transactions of the World Congress on the Cornea
REFERENCES III, (ed. H. Dwight Cavanagh), Raven Press,
New York, 281-6.
Beuennan, RW., Chew, S.J., Pedroza, L. et al. Masters, B.R.. (199Oa) Confocal microscopy of ocu­
(1992) Early diagnosis of infectious keratitis lar tissue, in Confocal Microscopy, (ed. T. Wil­
with in vivo real-time confocal microscopy. son) Academic Press, London, 305-24.
Invest. Ophthalmol. Vis. Sci., 33 (4) suppl., 1234. Masters, B.R.. (199Ob) In vivo corneal redox fluo­
Florakis, G.T., Auran, J.D., Koester, CJ. et al. (1992) rometry, in Noninvasive Diagnostic Techniques in
High resolution scanning slit microscopy of the Ophthalmology, (ed. B.R Masters), Springer­
human corneal epithelium in vivo. Invest. Oph­ Verlag, New York, 223-47.
thalmol. Vis. Sci., 33 (4) suppl., 1234. Masters, B.R (1991a) Three-dimensional imaging of
Jester, J.V., Cavanagh, H.D. and Lemp, M.V. the living eye. Proceedings of the Electron
(1990) Confocal microscopic imaging of the liv­ Microscopy Society of America, 170-1.
ing eye with tandem scanning confocal micros­ Masters, B.R.. (1991b) Three-dimensional visualiza­
copy, in Noninvasive Diagnostic Techniques in tion of the living eye. Proceedings of the IEEE
Ophthalmology, (ed. B.R. Masters), Springer­ Engineering in Medicine and Biology Society.
Verlag, New York, 172-88. Computers in Medicine, 13 (3/5), 1129-30.
Jester, J.V., Andrews, P.M., Petroll, W.M.. et al. Masters, B.R (1991c) Two- and three-dimensional
(1991) In vivo, real-time confocal imaging. J. visualization of the living cornea and ocular
Electron Microscopy Technique, 18, 50-60. lens. Machine Vision and Applications (special
Jester JV, PetroD WM, Garana RMR. et al. (1992) issue on three-dimensional microscopy), 4,
Comparison of in vivo and ex vivo cellular 227-32. .
structure in rabbit eyes detected by tandem Masters, B.R.. (1993) Specimen preparation and
scanning microscopy. J. Microscopy, 165, 169­ chamber for confocal microscopy of the eye.
81. Scanning Microscopy, (in press).
Koester, CJ. (1993) Confocal Microscopy of the Masters, B.R. and Fanner, M.A. (1993) Three­
Cornea, in Ophthalmic and Visual Optics/ dimensional confocal microscopy and visual­
Noninvasive Assessment of the Visual System. ization of the in situ cornea. Computerized
Technical Digest, (Optical Society of America, Medical Imaging and Graphics, (in press).
Washington, DC), 3,132. , Masters, B.R and Kino, G.S. (1990) Confocal
Koester, CJ. and Khanna, S.M .. (1989) Optical microscopy of the eye, in Noninvasive Diagnos­
Sectioning with the Scanning Slit Confocal tic Techniques in Ophthalmology, (ed. B.R Mas­
Microscope: Applications in Ophthalmology ters), Springer-Verlag, New York, 152-71.
and Ear Research. New Methods in Microscopy Masters, B.R and Paddock, S.W. (1990) Three­
and Low Light Imaging, (ed. J.E. Wampler). dimensional reconstruction of the rabbit cornea
Proc. SPIE 1161, 378-89. by confocal scanning optical microscopy and
Koester, CJ. and Roberts, CW. (1990) Wide-field volume rendering. Applied Optics, 29, 3816-22
specular microscopy, in Noninvasive Diagnostic Masters, B.R. and Thaer, A. (1993a) In vivo confo­
Techniques in Ophthalmology, (ed. B.R Masters), cal microscopy of the human cornea. Biophysical
Springer-Verlag, New York, 99-121. Journal, 64 (2), AI08.
Koester, CJ., Auran, J.D., Rosskothen, H.D. et al. Masters, B.R and Thaer, A. (1993b) In Vivo Confo­
(1992) A contact microscope objective with high cal Microscopy of the Living Human Eye: A New
numerical aperture for clinical examination of Real-time Confocal System. Proceedings of the
the cornea. Invest. Ophthalmol. Vis. Sci., 33 (4) 1993 International Conference on Confocal
suppl., 1233. Microscopy and 3D Image Processing, Sydney,
Lemp, M.A., Dilly, P.N. and Boyde, A.. (1986) Australia, p. BO.
Tandem-scanning (confocal) microscopy of the Masters, B.R and Thaer, A. (1993c) Confocal
full-thickness cornea. Cornea, 4,205-9. microscopy of the human in vivo cornea, in
Masters, B.R. (1988) Effects of contact lenses on the Ophthalmic and Visual Optics/Noninvasive
oxygen concentration and epithelial mitochon- Assessment of the Visual System. Technical Digest,
406 Anterior ocular microscopy
(Optical Society of America, Washington, DC), Thaer, A., Geyer, O.-c. and Kaszli, F.A. (1990)
3,133-6. Fluorogenic substrate techniques applied to the
Masters, B.R., Kriete, A. and Kukulies, J. (1993) noninvasive diagnosis of the living rabbit and
Ultraviolet confocal fluorescence microscopy of human cornea, in Noninvasive Diagnostic Tech­
the in vitro cornea: redox metabolic imaging. niques in Ophthalmology. (ed. B.R. Masters),
Applied Optics, 34 (4), 592-6. Springer-Verlag, New York, 569-90.
Pawley, J.B. (1990) Handbook of Biological Confocal Wegener, A.R., Gaida, G., Massig, I.H. and Brei­
Microscopy, Plenum Press, New York. pohl, W. (1992) In-vivo microscopical investiga­
Svishchev, G.M. (1969) Microscope for the study tion of cornea and lens by confocal optics.
of transparent light-scattering objects in Invest. Ophthalmol. Vis. Sci., 33 (4) suppl., 1234.
incident light. Optics and Spectroscopy, 26, Wilson, T. (ed.) (1990) Confocal Microscopy, Aca­
171-2. demic Press, London.

You might also like