310 Flame lonization Detector (FID) Replacement procedures Replacing the detector weldment WARNING _ Before proceeding, turn off the oven and any heated zones and let them cool down. Turn off any detector gases at their supply. Then turn off the main power switch and unplug the power cord. 1. Remove the entire detector as described in this section. 2. Remove the collector assembly and the detector’s electrometer as described in the Removing the electrometer procedure in this section. 3. Remove the 1 1/4-inch brass hex nut from the detector. Collector assembly > ‘N ~ Electromeier Hex nut Figure 310-7 Replacing the detector weldment B of 24 Detectors Jun 2001 ‘Agilant 6800 Gas Chrmatograph Sarvice Marual Flame lonization Detector (FID) 310 Replacement procedures 4, Remove the thermal strap and weldment from the detector pallet. Disconnect the strap from the weldment (two M7 hex nuts), h Fexnt CSCS ex ut © Thermal strap J eK H y Heater/sensor assembh oo \ Detector weldment q 4 M7 hex nuts ‘ ; i = Figure 310-8 — Removing the detector block Replacing the collector assembly WARNING _ Before proceeding, turn off the oven and any heated zones and let them cool down. Turn off any detector gases at their supply. Then turn off the main power switch and unplug the power cord. 1. Remove the right side cover of the instrument. Caution Make sure you are properly grounded with an ESD strap before continuing. Jun 2001 Detectors 9 of 24 ‘Asin 6890 Gas Chromatograph Service Manual