About Configuration
Configuration is a two part process for most GC accessory
devices that require power and/or communication resources
from the GC. In the first part of the configuration process, a
power and/or communication resource is assigned to the
device. The second part of the configuration process allows
setting of any configuration properties associated with the
device.
Status
Time
Valve #
Cryo Focus Valve
2 Dimensional GC Valve
Front injector
Back injector
Sample tray
Instrument
In many cases you can move directly to the item of interest
by pressing [Config][device].
3 When the Configure Device Display opens, the cursor should
be on the Unconfigured field. Press the [Mode/Type] key and
follow the GC prompts to assign resources to the device.
4 After assigning resources, the GC prompts for you to power
cycle the GC. Turn the GC power switch off and then on.
When the GC starts, select the device just assigned the GC
resources for further configuration if needed. When accessed,
its mode should indicate Configured and the other settleable
configuration properties are displayed.
General Topics
Ignore Ready =
The states of the various hardware elements are among the
factors that determine whether the GC is Ready for analysis.
Under some circumstances, you may not wish to have a specific
element readiness considered in the GC readiness
determination. This parameter lets you make that choice.
For example, an inlet heater is defective but you don’t plan to
use that inlet today. By setting Ignore Ready = TRUE for that inlet,
you can use the rest of the GC. After the heater is repaired, set
Ignore Ready = FALSE or the run could start before the that inlet’s
conditions are ready.
Information displays
Some configuration displays include lines similar to these:
Unconfigured:
Accessory devices requiring GC power or communication must
be assigned these GC resources before they can be used. To
make this hardware element usable, first “Unlock the GC
Configuration” then go to the Unconfigured parameter and press
[Mode/Type] to install it. If the hardware element you are
configuring requires selection of additional parameters, the GC
asks for that selection. If no parameters are required, press
[Enter] at the GC prompt to install that element. You are
required to power the GC off and then power the GC on to
complete this configuration.
After restarting the GC, a message reminding you of this change
and its effect on the default method is displayed. If needed,
change your methods to accommodate the new hardware.
Oven
Slow oven cool down mode Reduces the oven fan speed during
the cooldown cycle.