CARESCAPE Modular Monitors

Software Settings Transfer Guide


Software version 2 host build v2.4

CARESCAPE Modular Monitors


English
2nd edition
2106778-006
© 2018 General Electric Company.
All rights reserved.
Due to continuing product innovation, specifications in this manual are subject to change without notice.
For technical documentation purposes, the abbreviation GE is used for the legal entity names, GE Medical Systems Information Technologies,
Inc. and GE Healthcare Finland Oy.

2 CARESCAPE Modular Monitors 2106778-006


2018-06-20
Contents
Intended use of this guide ............................................................................ 5
New settings in software version 2.4 ........................................................... 7
Extended setting: V Tach alarm priority .................................................... 7
New QT/QTc alarm priority ......................................................................... 7
New Ventilator disconnected alarm priority.............................................. 7
New settings: V Tach Criteria ..................................................................... 8
Changed setting: Minimum HR/min........................................................... 8
New SpO2 option ......................................................................................... 9
New PR option ............................................................................................. 9
New impedance respiration option............................................................ 9
Extended setting: HR for SVT...................................................................... 9
Extended WLAN security settings .............................................................. 10
Changed setting: Speed & Duplex ............................................................. 10
New ADT server connection option............................................................ 10
New E-LoFlo options ................................................................................... 10
New settings in software version 2.3 ........................................................... 11
Change in non-invasive and invasive blood pressure alarm
priorities ....................................................................................................... 11
New option: Snapshot sweep speed ......................................................... 14
New option: 12 lead print speed ................................................................ 15
Extended settings: Allowed alarm priorities .............................................. 15
New settings in software version 2.0.8 ........................................................ 17
New option: ECG lead off alarm priority .................................................... 17
Extended settings: PVC, SVC, Impedance Resp, ST and Temp alarm
priority.......................................................................................................... 17
Change in Art and ICP Invasive pressure labels ........................................ 18
Change in other Invasive pressure labels .................................................. 19
Extended settings: ECG Printout format .................................................... 19
New settings in software version 2.0.7 ........................................................ 21
Extended settings: Pacemaker Detection with PDM................................. 21
Extended settings: 12SL Display Format ................................................... 21

2106778-006 CARESCAPE Modular Monitors 3


Extended settings: New default value for Spirometry sensor
type .............................................................................................................. 21
Change in Menu structure (Diagnostic Tool).............................................. 22
New global options ..................................................................................... 22
New options in NICU software package with CARESCAPE Respiratory
module ......................................................................................................... 22

4 CARESCAPE Modular Monitors 2106778-006


Intended use of this guide
This guide provides information about new software settings that require manual
configuration after a software upgrade for the following CARESCAPE Monitors:
● CARESCAPE Monitor B450
● CARESCAPE Monitor B650
● CARESCAPE Monitor B850
In this document software version 2.0.6 refers to software version 2 with host build
2.0.6.
In this document software version 2.0.7 refers to software version 2 with host build
2.0.7.
In this document software version 2.0.8 refers to software version 2 with host build
2.0.8.
In this document software version 2.3 refers to software version 2 with host build 2.3.
In this document software version 2.4 refers to software version 2 with host build 2.4.
The settings are divided in the tables as follows:
● Settings listed in Chapter “New settings in software version 2.4” are available in
the software version 2.4 only.
● Settings listed in Chapter "New settings in software version 2.3" are available in the
software versions 2.3 and 2.4.
● Settings listed in Chapter “New settings in software version 2.0.8” are available in
the software versions 2.0.8, 2.3, and 2.4.
● Settings listed in Chapter “New settings in software version 2.0.7” are available in
the software versions 2.0.7, 2.0.8, 2.3, and 2.4.
NOTE: If you upgrade software from version 1, refer to Software Setting Transfer
Instruction, 2062973-011 for CARESCAPE B850 and B650 Monitor for new software
settings that were introduced in version 2.
For more detailed information regarding all settings, refer to the supplemental
information manual.
Use this guide together with the monitor’s service manual. The service manual has
useful service information and important safety precautions.

2106778-006 CARESCAPE Modular Monitors 5


6 CARESCAPE Modular Monitors 2106778-006
New settings in software version 2.4
Extended setting: V Tach alarm priority
Settings Selection Selection Default Menu Notes
in v2.0.8 in v2.4 value in
v2.4
V Tach High High High Alarm Setup > Arrhytmia Profile settings. Separate
alarm Medium Medium for each user profile.
priority Selection needs to be
Low Low
activated in Default
Informa- Setup > Care Unit
tional Settings > Parameters
Off (NICU > Parameter Alarms >
only) Allowed Priorities > ECG.
NICU only: To enable
the Off selection, the
selection needs to be
activated in Default
Setup > Care Unit
Settings > Parameters >
ECG.

New QT/QTc alarm priority


Setting Selection in Default Menu Notes
v2.4 value
QT/QTc alarm Medium Low Alarm setup > Alarm Profile setting. Separate for
priority Low Priorities each user profile.
Informa- Informational selection
tional needs to be activated in
Default Setup > Care Unit
Settings > Parameters >
Parameter Alarms > Allowed
Priorities > ECG.

New Ventilator disconnected alarm priority


Setting Selection in Default Menu Notes
v2.4 value
Ventilator Medium Medium Alarm setup > Alarm Profile setting. Separate for
disconnected High Priorities > Other each user profile.
alarm priority Parameters

2106778-006 CARESCAPE Modular Monitors 7


New settings: V Tach Criteria
Setting Selection Default value Menu Notes
in version
2.4
Event Duration Off Off Alarm Setup > Profile setting. Separate
5s Arrhytmia > Lethal for each user profile.
Alarms > V Tach This setting has been
10 s Criteria locked. To change
15 s the value, the setting
20 s must be unlocked
in Default Setup >
25 s
Locking settings >
30 s Parameters.
V Tach Criteria AND AND Alarm Setup > Profile setting. Separate
OR Arrhytmia > Lethal for each user profile.
Alarms > V Tach This setting has been
Criteria locked. To change
the value, the setting
must be unlocked
in Default Setup >
Locking settings >
Parameters.
V Tach Event Allowed Not allowed Monitor Setup > Profile setting. Separate
Duration Criteria Not Default Setup > Profile for each user profile.
allowed Settings > Alarm
Delays

Changed setting: Minimum HR/min


Setting Selection in Selection in Default Menu Notes
version 2.0.8 version 2.4 value
Minimum ICU, OR, With PDM/PSM: 100 Alarm Setup > Profile setting.
HR/min PACU, and 90-300 (adjustable) Arrhytmia > Lethal Separate for each
ED: 100 (not Alarms > V Tach user profile.
With combination
adjustable) Criteria This setting has been
monitoring:
NICU: 160 (not locked. To change
● For patients who
adjustable) the value, the setting
are 14 years or
older: 100-300 must be unlocked
(adjustable in in Default Setup >
steps of 10) Locking settings >
Parameters.
● For patients
younger than
14 years: Not
adjustable

8 CARESCAPE Modular Monitors 2106778-006


New SpO2 option
Setting Selection in Default Menu Notes
v2.4 value
SpO2 alarm 0s 5s Parameter Setup > SpO2 > Profile setting. Separate for
delay 5s Alarms > Priorities & Delays each user profile.
info Selection needs to be
10 s
activated in Default Setup
15 s > Profile Settings > Alarm
Delays.

New PR option
Setting Selection in Default Menu Notes
v2.4 value
PR alarm 0s 5s Parameter Setup > ECG > Profile setting. Separate for
delay 5s HR/PR Alarms > Priorities & each user profile.
Delays info Selection needs to be
10 s
activated in Default Setup
15 s > Profile Settings > Alarm
20 s Delays.

New impedance respiration option


Setting Selection in Default Menu Notes
v2.4 value
Impedance 0s 10 s Parameter Setup > Profile setting. Separate for
respiration 5s 15 s (NICU, Impedance Respiration each user profile.
alarm delay Pediatric, or > Alarms > Priorities & Selection needs to be
10 s Delays info
Infant) activated in Default Setup
15 s >Profile Settings > Alarm
20 s Delays.
25 s
30 s

Extended setting: HR for SVT


Setting Selection Selection Default Menu Notes
in v2.0.8 in v2.4 value
HR for SVT 0 - 150 0 - 220 120 Alarm setup > Arrhytmia Profile setting. Separate
/min > Atrial Alarms for each user profile.
NICU only.

2106778-006 CARESCAPE Modular Monitors 9


Extended WLAN security settings
Setting Selection Selection Default Menu Notes
in v2.0.8 in v2.4 value in
v2.4
WLAN None None None Monitor Setup > Service Platform setting. For all
security WEP WEP > Configuration > WLAN user profiles.
protocol Available for CARESCAPE
WPA-PSK WPA-PSK
Monitor B450 only.
WPA2-PSK WPA2-PSK
802.1x AEP
Custom

Changed setting: Speed & Duplex


Setting Default New default value in Menu Notes
value in version 2.4
version
2.0.8
Speed & auto 10 Mb half Monitor Setup > Service Platform setting. For all
Duplex > Configuration > user profiles.
Network > Wired This setting is
Interfaces configurable in
CARESCAPE Monitor B450
and CARESCAPE Monitor
B850. In CARESCAPE
Monitor B650 this is a
non-configurable setting.

New ADT server connection option


Setting Selection in Default Menu Notes
v2.4 value in v2.4
ADT server Enabled Disabled Monitor setup > Service > Platform setting. For all user
connection Disabled Configuration > Network profiles.

New E-LoFlo options


Setting Selection in Default Menu Notes
v2.4 value in v2.4
CO2 average 10 s 10 s Parameter setup > Gases > Profile setting. Separate for
20 s Setup each user profile.
Selection available only when
E-LoFLo module connected.
N2O noise 0-30 0-30 Parameter setup > Gases > Profile setting. Separate for
level 30-80 Setup each user profile.
Selection available only when
E-LoFLo module connected.

10 CARESCAPE Modular Monitors 2106778-006


New settings in software version 2.3
Change in non-invasive and invasive blood pressure alarm
priorities
Settings Default Default Menu Notes
value in value in v2.3
v2.0.8
NIBP alarm Escalating Medium Alarm Setup > Alarm Profile settings. Separate for
priority Priorities > Other each user profile.
Parameters Selection needs to be
activated in Default Setup
> Care Unit Settings >
Parameters > Parameter
Alarms > Allowed Priorities >
parameter group.
Art alarm Escalating Medium Alarm Setup > Alarm Profile settings. Separate for
priority Priorities > Other each user profile.
Parameters Selection needs to be
activated in Default Setup
> Care Unit Settings >
Parameters > Parameter
Alarms > Allowed Priorities >
parameter group.
Fem alarm Escalating Medium Alarm Setup > Alarm Profile settings. Separate for
priority Priorities > Other each user profile.
Parameters Selection needs to be
activated in Default Setup
> Care Unit Settings >
Parameters > Parameter
Alarms > Allowed Priorities >
parameter group.
FemV alarm Escalating Medium Alarm Setup > Alarm Profile settings. Separate for
priority Priorities > Other each user profile.
Parameters Selection needs to be
activated in Default Setup
> Care Unit Settings >
Parameters > Parameter
Alarms > Allowed Priorities >
parameter group.
PA alarm Escalating Medium Alarm Setup > Alarm Profile settings. Separate for
priority Priorities > Other each user profile.
Parameters Selection needs to be
activated in Default Setup
> Care Unit Settings >
Parameters > Parameter
Alarms > Allowed Priorities >
parameter group.

2106778-006 CARESCAPE Modular Monitors 11


Settings Default Default Menu Notes
value in value in v2.3
v2.0.8
CVP alarm Escalating Medium Alarm Setup > Alarm Profile settings. Separate for
priority Priorities > Other each user profile.
Parameters Selection needs to be
activated in Default Setup
> Care Unit Settings >
Parameters > Parameter
Alarms > Allowed Priorities >
parameter group.
LAP alarm Escalating Medium Alarm Setup > Alarm Profile settings. Separate for
priority Priorities > Other each user profile.
Parameters Selection needs to be
activated in Default Setup
> Care Unit Settings >
Parameters > Parameter
Alarms > Allowed Priorities >
parameter group.
RAP alarm Escalating Medium Alarm Setup > Alarm Profile settings. Separate for
priority Priorities > Other each user profile.
Parameters Selection needs to be
activated in Default Setup
> Care Unit Settings >
Parameters > Parameter
Alarms > Allowed Priorities >
parameter group.
Escalating Medium Alarm Setup > Alarm Profile settings. Separate for
ICP alarm Priorities > Other each user profile.
priority Parameters Selection needs to be
activated in Default Setup
> Care Unit Settings >
Parameters > Parameter
Alarms > Allowed Priorities >
parameter group.
CPP alarm Escalating Medium Alarm Setup > Alarm Profile settings. Separate for
priority Priorities > Other each user profile.
Parameters Selection needs to be
activated in Default Setup
> Care Unit Settings >
Parameters > Parameter
Alarms > Allowed Priorities >
parameter group.
RVP alarm Escalating Medium Alarm Setup > Alarm Profile settings. Separate for
priority Priorities > Other each user profile.
Parameters Selection needs to be
activated in Default Setup
> Care Unit Settings >
Parameters > Parameter
Alarms > Allowed Priorities >
parameter group.

12 CARESCAPE Modular Monitors 2106778-006


Settings Default Default Menu Notes
value in value in v2.3
v2.0.8
UAC alarm Escalating Medium Alarm Setup > Alarm Profile settings. Separate for
priority Priorities > Other each user profile.
Parameters Selection needs to be
activated in Default Setup
> Care Unit Settings >
Parameters > Parameter
Alarms > Allowed Priorities >
parameter group.
UVC alarm Escalating Medium Alarm Setup > Alarm Profile settings. Separate for
priority Priorities > Other each user profile.
Parameters Selection needs to be
activated in Default Setup
> Care Unit Settings >
Parameters > Parameter
Alarms > Allowed Priorities >
parameter group.
P1 alarm Escalating Medium Alarm Setup > Alarm Profile settings. Separate for
priority Priorities > Other each user profile.
Parameters Selection needs to be
activated in Default Setup
> Care Unit Settings >
Parameters > Parameter
Alarms > Allowed Priorities >
parameter group.
P2 alarm Escalating Medium Alarm Setup > Alarm Profile settings. Separate for
priority Priorities > Other each user profile.
Parameters Selection needs to be
activated in Default Setup
> Care Unit Settings >
Parameters > Parameter
Alarms > Allowed Priorities >
parameter group.
P3 alarm Escalating Medium Alarm Setup > Alarm Profile settings. Separate for
priority Priorities > Other each user profile.
Parameters Selection needs to be
activated in Default Setup
> Care Unit Settings >
Parameters > Parameter
Alarms > Allowed Priorities >
parameter group.
P4 alarm Escalating Medium Alarm Setup > Alarm Profile settings. Separate for
priority Priorities > Other each user profile.
Parameters Selection needs to be
activated in Default Setup
> Care Unit Settings >
Parameters > Parameter
Alarms > Allowed Priorities >
parameter group.

2106778-006 CARESCAPE Modular Monitors 13


Settings Default Default Menu Notes
value in value in v2.3
v2.0.8
P5 alarm Escalating Medium Alarm Setup > Alarm Profile settings. Separate for
priority Priorities > Other each user profile.
Parameters Selection needs to be
activated in Default Setup
> Care Unit Settings >
Parameters > Parameter
Alarms > Allowed Priorities >
parameter group.
P6 alarm Escalating Medium Alarm Setup > Alarm Profile settings. Separate for
priority Priorities > Other each user profile.
Parameters Selection needs to be
activated in Default Setup
> Care Unit Settings >
Parameters > Parameter
Alarms > Allowed Priorities >
parameter group.
P7 alarm Escalating Medium Alarm Setup > Alarm Profile settings. Separate for
priority Priorities > Other each user profile.
Parameters Selection needs to be
activated in Default Setup
> Care Unit Settings >
Parameters > Parameter
Alarms > Allowed Priorities >
parameter group.
P8 alarm Escalating Medium Alarm Setup > Alarm Profile settings. Separate for
priority Priorities > Other each user profile.
Parameters Selection needs to be
activated in Default Setup
> Care Unit Settings >
Parameters > Parameter
Alarms > Allowed Priorities >
parameter group.

New option: Snapshot sweep speed


Settings Available Selection Default Menu Notes
setting in v2.3 value in
v2.3
Snapshot Sweep 12.5 mm/s 12.5 mm/s Trends > View > Profile settings. Separate
Setup Speed Snapshot > Snapshot for each user profile.
25 mm/s
Setup
50 mm/s

14 CARESCAPE Modular Monitors 2106778-006


New option: 12 lead print speed
Setting Selection Selection Default value in v2.3 Menu Notes
in v2.0.8 in v2.3
12 Lead 50 mm/s 25 mm/s 25 mm/s Default Setup > Care unit setting.
Print Not 50 mm/s Care Unit Settings Separate for each
Speed possible > Parameters > 12 user profile.
to lead ECG
change.

Extended settings: Allowed alarm priorities


Setting Selection Selection Default Menu Notes
in version in version value in
2.08 2.3 version 2.3
NIBP alarm Escalating Escalating High Alarm Setup > Alarm Care unit settings.
priority Priorities > Other Separate for each user
High High Medium Parameters profile.
Medium Medium Selection needs to be
activated in Default
Low Low
Setup > Care Unit
or Settings > Parameters
> Parameter Alarms
High > Allowed Priorities >
Medium parameter group.

IP alarm Escalating Escalating High Alarm Setup > Alarm Care unit settings.
priority High Medium Priorities > Invasive Separate for each user
High Pressures profile.
Medium
Medium Selection needs to be
Low activated in Default
Low Setup > Care Unit
or
Settings > Parameters
High > Parameter Alarms
> Allowed Priorities >
Medium parameter group.
CO2 alarm Escalating Escalating High Alarm Setup > Alarm Care unit settings.
priority Medium Priorities > Other Separate for each user
High High Parameters profile.
Medium Medium Selection needs to be
activated in Default
or Setup > Care Unit
Settings > Parameters
High > Parameter Alarms
Medium > Allowed Priorities >
parameter group.
FiAA alarm Escalating Escalating High Alarm Setup > Alarm Care unit settings.
priority Medium Priorities > Other Separate for each user
High High Parameters profile.
Medium Medium Selection needs to be
activated in Default
or Setup > Care Unit
Settings > Parameters
High

2106778-006 CARESCAPE Modular Monitors 15


Setting Selection Selection Default Menu Notes
in version in version value in
2.08 2.3 version 2.3
> Parameter Alarms
Medium > Allowed Priorities >
parameter group.

16 CARESCAPE Modular Monitors 2106778-006


New settings in software version 2.0.8
New option: ECG lead off alarm priority
Settings Selection Selection Default Menu Notes
in version in version value
2.0.6 and 2.0.8
version
2.0.7
ECG lead Selection Low Low Alarm Setup > Alarm Profile setting. Separate
off alarm not Informa- Priorities > ECG for each user profile.
priority available tional The new selection needs
to be activated in Default
Setup > Care Unit
Settings > Parameters
> Parameter Alarms >
Allowed Priorities > ECG

Extended settings: PVC, SVC, Impedance Resp, ST and Temp


alarm priority
Settings Selection Selection Default Menu Notes
in version in version value
2.0.6 and 2.0.8
version
2.0.7
PVC alarm Escalating Escalating Escalating Alarm Setup > Alarm Profile setting. Separate
priority High High Priorities > ECG for each user profile.
Medium Medium
Low Low
Informa-
tional
SVC alarm Escalating Escalating Low Alarm Setup > Alarm Profile setting. Separate
priority High High Priorities > ECG for each user profile.
Medium Medium
Low Low
Informa-
tional
RR (Imped.) Escalating Escalating Escalating Alarm Setup > Alarm Profile setting. Separate
high/low High High Priorities > Other for each user profile.
alarm Parameters The new selection needs
priority Medium Medium
to be activated in Default
Low Low Setup > Care Unit
Informa- Settings > Parameters
tional > Parameter Alarms
> Allowed Priorities >
Other Parameters

2106778-006 CARESCAPE Modular Monitors 17


Settings Selection Selection Default Menu Notes
in version in version value
2.0.6 and 2.0.8
version
2.0.7
ST alarm High High Medium Alarm Setup > Alarm Profile setting. Separate
priority Medium Medium Priorities > ECG for each user profile.
Low Low The new selection needs
to be activated in Default
Informa- Setup > Care Unit
tional Settings > Parameters
> Parameter Alarms >
Allowed Priorities > ECG
Parameters
Temp Selection Escalating Escalating Alarm Setup > Alarm Profile setting. Separate
alarm not Informa- Priorities > Other for each user profile.
priority available tional Parameters The new selection needs
to be activated in Default
Setup > Care Unit
Settings > Parameters
> Parameter Alarms
> Allowed Priorities >
Other Parameters

Change in Art and ICP Invasive pressure labels


Settings Selection Selection Default Menu Notes
in version in version value
2.0.6 and 2.0.8
version
2.0.7
Art Art can be Art can be Multiple Parameter Setup > Profile setting. Separate
assigned to assigned to Invasive Pressure > Px > for each user profile.
one InvBP multiple In- Setup If you assign additional
channel at vBP chan- (Px = P1, P2, P3...) Art channels, the settings
a time nel simul- are copied from the
taneously original Art channel.
ICP ICP can be ICP can be Multiple Parameter Setup > Profile setting. Separate
assigned to assigned to Invasive Pressure > Px > for each user profile.
one InvBP multiple In- Setup If you assign additional
channel at vBP chan- (Px = P1, P2, P3...) ICP channels, the settings
a time nel simul- are copied from the
taneously original ICP channel.

18 CARESCAPE Modular Monitors 2106778-006


Change in other Invasive pressure labels
Settings Selection Selection Default Menu Notes
in version in version value
2.0.6 and 2.0.8
version
2.0.7
ABP available not N/A N/A You need to check the
available settings manually after
upgrading. *)
UAC available available N/A Parameter Setup > Profile setting. Separate
only in in OR, Invasive Pressure > Px > for each user profile.
NICU SW ICU,NICU, Setup Settings transferred in
package ED and (Px = P1, P2, P3...) NICU SW package only.
PACU SW
packages
UVC available available N/A Parameter Setup > Profile setting. Separate
only in in OR, Invasive Pressure > Px > for each user profile.
NICU SW ICU,NICU, Setup Settings transferred in
package ED and (Px = P1, P2, P3...) NICU SW package only.
PACU SW
packages

*) NOTE: Some of the ABP settings are transferred from the original Art and ABP
settings. To ensure proper settings configuration, GE recommends that you check
the settings manually after upgrading.

Extended settings: ECG Printout format


Settings Selection Selection Default Menu Notes
in version in version value
2.0.6 and 2.0.8
version
2.0.7
ECG 4x2.5 – 1 4x2.5 – 1 4x2.5 – 1 Default Setup > Care Unit Profile setting. Separate
Printout Rhythm Rhythm Rhythm Settings > Parameters > for each user profile.
format 4x2.5 – 3 4x2.5 – 3 ECG
Rhythm Rhythm
12 12
Rhythms Rhythms
25mm/s 25mm/s
12 12
Rhythms Rhythms
50mm/s 50mm/s
Cabrera 2x6
Rhythms
50mm/s
Cabrera
50mm/s

2106778-006 CARESCAPE Modular Monitors 19


20 CARESCAPE Modular Monitors 2106778-006
New settings in software version 2.0.7
Extended settings: Pacemaker Detection with PDM
Settings Selection Selection Default value Menu Notes
in version in version
2.0.6 with 2.0.7 with
PDM PDM 2.4
Pace- On Sensitive Sensitive (with PDM Parameter Setup > Profile setting.
maker Off Normal 2.4) ECG > Advanced Separate for each
detection (New) On (with PDM 2.3 or user profile.
Off older) New selection
Sensitive (with available only with
E-PSM(P)) PDM 2.4 and software
version 2.0.7.
Pace2 (with TRAM)

Extended settings: 12SL Display Format


Settings Selection Selection Default Menu Notes
in version in version value
2.0.6 2.0.7
12SL 4x2.5 - 1 4x2.5 - 1 4x2.5 - 3 Parameter Setup > ECG Profile setting. Separate
Display Rhythm Rhythm Rhythm > 12 Lead Analysis > for each user profile.
Format 4x2.5 - 3 4x2.5 - 3 Settings
Rhythm Rhythm
12 12
Rhythms Rhythms
Cabrera 2x6
Rhythms
Cabrera

Extended settings: New default value for Spirometry sensor


type
Settings Selection New default value in Menu Notes
in version version 2.0.7
2.0.6
Spirometry Adult All user profiles in NICU Parameter Setup Profile setting. Separate
sensor type Pediatric software package: > Spirometry/Gas for each user profile.
Pediatric Exchange
User profiles Infant and
Pediatric in ED, ICU,
OR, and PACU software
package: Pediatric
User profiles other than
Infant and Pediatric in
ED, ICU, OR, and PACU
software package: Adult

2106778-006 CARESCAPE Modular Monitors 21


Change in Menu structure (Diagnostic Tool)
Settings Selection Combined Default Menu Notes
in version menu in value in
2.0.6 version version
2.0.7 2.0.7
ACITIPI On/Off ACITIPI ACITIPI Parameter Setup > ECG Moved to other menu
ACS > 12 Lead Analysis > destination.
ACS On/Off Settings
Off Separate for each user
profile.

New global options


Settings New Default Menu Notes
available value
setting in
version 2.0.7
Citrix Server Support for Default Webmin > Configuration > Platform setting. For all user
version Citrix server server Citrix profiles.
versions version:
4.0/4.5 and 4.0/4.5
5.0.
Citrix Server Support Default IP Webmin > Configuration > Platform setting. For all user
address 2–4 for 4 Citrix address: Citrix profiles.
servers with Empty
4 different IP
addresses.
Alarm New Immediate Default Setup > Care Unit Care Unit setting. For all user
Activation for available Settings > Parameters > profiles.
PR selection: Allowed priorities > Other
Immediate Parameters
or Delayed
Print Inter- New Enabled Parameter Setup > ECG > 12 Profile setting. Separate for
pretations available Lead Analysis > Settings each user profile.
selection:
Enabled or
Disabled
Low-Pass New 150Hz Parameter Setup > ECG > 12 Same default for all user
Filter available Lead Analysis > Filter profiles.
selections: Not possible to change and
20Hz or 40Hz save as profile of unit settings.
or 100Hz or
150Hz

New options in NICU software package with CARESCAPE


Respiratory module
NOTE: Not available in the US and its territories.
NOTE: Not available in China.

22 CARESCAPE Modular Monitors 2106778-006


Settings Available New default Menu Notes
setting setting in
version 2.0.7
Alarm Limit New factory 2 Parameter Setup > Available only with
default for Spirometry > MV/Vent “E-sC(Ai)O(V) module NICU
Mvexp High Alarms SW” license.
Alarm Limit Profile setting. Separate for
each user profile.
New factory 30 Parameter Setup > Available only with
default for Spirometry > PAW Alarms “E-sC(Ai)O(V) module NICU
Ppeak High SW” license.
Alarm Limit Profile setting. Separate for
each user profile.
New factory 8 Parameter Setup > Available only with
default for Spirometry > PAW Alarms “E-sC(Ai)O(V) module NICU
PEEPe High SW” license.
Alarm Limit Profile setting. Separate for
each user profile.

2106778-006 CARESCAPE Modular Monitors 23