Professional Documents
Culture Documents
Isokinetic
Handbook
ISOKINETIC SOURCE SAMPLING
Isokinetic Handbook
Revision No: 7
Revision Date: February 2018
2
TABLE OF CONTENTS
Chapter 1: Introduction 5
System Description 7
Source Sampler Console 8
Electrical Subsystem 8
Thermocouple Subsystem 8
Vacuum Subsystem 9
External Vacuum Pump Unit 11
Probe Assembly 12
Probe Liner 13
Probe Heater 14
Modular Sample Case 16
Umbilical Cable with Umbilical Adapter 17
Glassware Sample Train 18
3
Method 3 – Gas Analysis for Dry Molecular Weight 46
Determine Dry Molecular Weight 48
Method 4 – Moisture Content of Stack Gas 49
Reference Method 4 51
Approximation Method 53
Calculating Stack Gas Moisture Content 55
Method 5 – Determination of Particulate Emissions 56
K-Factor Calculations 58
Method 5 Test Procedure59
Recommended Reading List for Isokinetic Sampling 70
Appendix A 86
Recommended Equipment for Isokinetic Sampling 87
Recommended Spare Parts 90
Equipment Checklist 92
Appendix B 93
Calibration Data Sheets 94
Appendix C 102
Stack Testing Field Data Sheets 103
Appendix D 113
Calculation Worksheets 114
4
Chapter 1
Introduction
The purpose of this manual is to provide a condensed understanding of the procedures established by the
United States Environmental Protection Agency (US EPA) in accordance with Reference Methods 1 through
5 – Determination of Particulate Emissions from Stationary Sources.
In addition to providing an easy-to-use guide for Methods 1 through 5, we have provided users with
reference information on system configuration, calibration procedures, maintenance and troubleshooting. Be
on the lookout for helpful tips and training aids throughout this manual.
We trust you will find this guide useful regardless of the sampling equipment you use and will use our
equipment for demonstrations and illustrations.
Isokinetic Sampling is the collection and measurement of a gas sample from a source at the same velocity
as the gas travels in the stack to provide a representative assessment of solid particulate matter that is in the
source.
To perform isokinetic testing, you must have a thorough understanding of the first five test methods
presented in Title 40 Part 60 Appendix A of the Code of Federal Regulations (40CFR60 App. A). While
Method 5 outlines the general sampling train operation protocol, Methods 1 through 4 prescribe techniques
that serve as a foundation for Method 5 sampling activities. Together, these methods outline the basic
protocols for determining particulate concentrations and mass emission rates.
US EPA Description
Method
Method 1 Determination of Sampling Location and Traverse Points
You can easily adapt the basic Method 5 sampling train to test for many other gaseous and particulate
emissions from stationary sources. Adapting basic test methods allow you to expand testing to include
5
parameters of interest such as metals, polychlorinated biphenyls (PCBs), dioxins/furans, polycyclic aromatic
hydrocarbons (PAHs), particle size distributions and an ever-increasing group of other pollutants.
While the different methods are designated by other US EPA method numbers, they actually are variations of
Method 5 procedures. Variations might include using different impinger solutions, organic resin traps,
different filter media, various sampling temperatures, or a range of other alternative procedures.
The manual and the automated source sampling consoles can be used for the following isokinetic test
methods and pollutants:
6
System Description
The first step to successful sampling is to familiarize yourself with the standard equipment. To illustrate the
necessary components of source sampling, we’ve included a diagram of the five main components
demonstrated on the Apex Instruments Isokinetic Source Sampling System, shown in Figure 1-1:
1. Source Sampler Console, which includes a differential pressure transducer (or dual-column
manometers), sample flow control valves with orifice flow meter, dry gas meter, and electrical
controls.
2. Sample Pump External Rotary Vane (or internal diaphragm pump) , including hoses with quick-
connect fittings and lubricator.
3. Probe Assembly includes a stainless-steel probe sheath, probe liner, tube heater, Type-S pitot tube,
stack and heater thermocouples, and an Orsat line.
4. Modular Sample Case includes a filter oven for the filter assembly, an impinger case for the
impinger glassware and electrical connections.
5. Umbilical Cable includes electrical and pneumatic lines to connect the Modular Sample Case to the
Source Sampling Meter Console.
7
Description of Component Parts
The Source Sampler Console is the operator’s control station. It monitors gas velocity and temperatures at
the sampling location and controls system sampling rate and system temperatures. Housed within the console
are the Electrical, Thermocouple, and Vacuum sub-systems.
Electrical Subsystem:
The electrical subsystem provides switched power to several circuits, including main power, pump power,
manometer zero, timer, probe heater and oven heater. Power rating should be chosen according to your
source of electricity. For example, the Apex Instruments XC-500 Sampling Systems can be configured for
120VAC/60Hz or 240VAC/50Hz electrical power.
Note: When the main power switch is on, the cooling fan should operate. Also, the pump has its own power
cord, which is plugged into the Source Sampler Console.
To activate the heaters in the filter compartment (Hot Box) and the
probe heater, turn on the switches labeled FILTER and PROBE.
Adjust the dials to approximately 120°C (248°F) and check the temperature display to verify that the heaters
are working. Allow time for the temperatures to stabilize and then verify operation of the circuits.
Thermocouple Subsystem:
The thermocouple subsystem displays, measures and provides feedback for the temperature controls that are
critical to isokinetic sampling operation. The Apex Instruments thermocouple system consists of Type-K
8
thermocouples, extension wires, male/female connectors, receptacles, a thermocouple selector switch, and a
digital temperature display with internal compensating junction.
Existing consoles offer automatic and digitally programmable temperature controllers for probe and filter
oven heat. The controllers receive temperature feedback signals to maintain temperatures within range of the
set point. The thermocouple electrical diagram is presented in the electrical schematic, found in the
appropriate operator’s manual.
Vacuum Subsystem:
The vacuum pump assembly provides the vacuum action for extracting the gas sample from the stack and
through the various components of the isokinetic source sampling system.
Typically, the vacuum subsystem consists of an external vacuum pump assembly (or internal diaphragm
pump), quick-connects, internal fittings, two (2) control valves (coarse and fine), an orifice meter, and a
dual-column inclined manometer (or pressure transducer).
A popular method of controlling flow is the dual-valve design that helps the operator obtain very precise
control over the sample flow rate through use of the Coarse Control Valve and the Fine Increase Valve.
● The Coarse Control Valve is a ball valve with a 90°
handle rotation from closed to full open. This valve
controls the flow from the SAMPLE inlet to the
Vacuum Pump inlet. Tips from a Stack Tester
● The Fine Increase Valve is a needle-type valve with
four (4) turns from closed to full open. The Fine
Increase Valve allows flow to re-circulate from the
pump outlet back to the pump inlet. The Fine Increase
Valve is used for precise vacuum control during leak
checks. By observing the orifice
reading (ΔH) on the front
You can zero the ΔH manometer before or during a sampling side of the manometer, you
run by flipping on the Manometer Zero switch found on the can quickly adjust the
front panel. This actuates the solenoid valves to vent the sample flow rate using the
pressure lines to atmosphere. Then, you can adjust the Fine Increase Valve so that
manometer’s fluid level using the knobs located at the bottom the sample is extracted under
of the manometer. isokinetic conditions.
Did you know? You can zero the pitot tube manometer by disconnecting the pitot lines
at the quick-connects on the Source Sampler Console.
9
Figure 1-2 below demonstrated the Apex Instruments Model XC-522 Source Sampler Console’s front panel
to provide an example of the layout of Source Sampling Console.
10
Source Sampler Console Front Source Sampler Console Rear
Product Highlight
Apex Instruments offers two pump styles: the E-
0523 lubricated rotary-vane pump and the
internal diaphragm housed inside the console.
Both pump assemblies are available in either
120VAC or 240VAC operation and customers
can choose from two enclosure options for the
external pump: a black polyethylene case with
molded handles and removable covers or an
aluminum open frame (seen to the right).
11
Tips from a Stack Tester
3. Probe Assembly
The main components of a Probe Assembly are:
● Probe Liner: 15.9 mm (5/8 in.) OD tubing made from either Borosilicate Glass, Quartz, Stainless Steel,
Inconel, or PTFE.
● Probe Heater: Removable rigid tube heater with coiled heating element, electric thermal insulation and
thermocouple, with a maximum recommended temperature of 260oC (500oF).
● Probe Sheath: 25.4 mm (1 in.) OD tube with quad-assembly attached that includes a replaceable,
modular S-type pitot tube, stack thermocouple and a 1/4 in. OD stainless steel tube to collect gas samples
for Orsat analysis.
● Small Parts Kit: Fittings to attach nozzle to Probe Assembly. Fittings include 5/8 in. bored union, nut
and ferrules.
Did you know? The most effective probe length in the stack is equivalent to 0.305 m
(1 ft) less than nominal length. Probe lengths vary from 0.914 m (3 ft) to 4.877 m (16
ft) nominal length.
12
Figure 1-4, right side, illustrates a standard Probe Assembly (top) and a Probe Assembly with the optional
50.8 mm (2 in.) Oversheath and Packing Gland (bottom).
The left part of the diagram shows the connection between the nozzle and probe using fittings from the small
parts kit.
Probe Liner:
Standard Probe Liners are constructed from 5/8 in. OD tubing and have #28 ball joints with an O-ring
groove. Available liner materials are borosilicate glass, quartz, stainless steel, Inconel, and PTFE. You will
need a ball joint adapter if you have a PTFE liner, straight liner, or liner with integrated nozzles. Figure 1-5
shows the potential Probe Liner configurations.
13
Important: Review maximum operating temperatures before selecting a Probe Liner material and
configuration. Table 1-3 shows the temperature limits for Probe Liner materials, while Table 1-4 reflects
temperature limits for various probe configurations.
Table 1-3: Maximum Stack Gas Temperatures for Probe Liner Materials
Maximum
Material Temperature
177°C (350°F)
PTFE Liners and Fittings
260°C (500°F)
Glass-Filled PTFE Fittings
Borosilicate Glass Liners 480°C (900°F)
Stainless Steel Liners 650°C (1200°F)
Quartz Liners 900°C (1650°F)
Inconel or C276 Alloy Liners 870°C (1600°F)
Maximum
Probe Assembly Configuration Temperature
480°C (900°F)
Stainless Steel Sheath and Glass Liner
Stainless Steel Sheath and Liner 650°C (1200°F)
Inconel or C276 Alloy Sheath and Liner 870°C (1600°F)
Inconel or C276 Alloy Sheath and Quartz Liner 870°C (1600°F)
15
4. Modular Sample Case
The Modular Sample Case is used for support, protection and environmental control of the glassware in the
sampling train. The Modular Sample Case consists of an insulated heated filter compartment (Hot Box) and
insulated impinger case (Cold Box).
Figure 1-6 illustrates the major components and accessory connections on the Modular Sample Case.
Hinged Probe
34.3cm Clamp Filter Oven
Umbilical Adapter Clamp
31.8cm
Amphenol Connector
24.1cm
Impinger
Oven Thermocouple Connector
Case
Aux. Power Riser
16
5. Umbilical Cable with Umbilical Adapter
The Umbilical Cable connects the Modular Sample case and Probe Assembly to the Source Sampler
Console. The Umbilical Cable typically contains the gas sample and pitot lines, as well as thermocouple and
power lines.
Below is an example of the configuration of the umbilical connections (using the Apex Instruments product
as an example):
● The primary gas sample line (blue, 3/8 in. i.d./12.7 mm (5/8 in. o.d.)) has a male quick-connect on the
outlet, at the opposite end, a 12.7 mm (5/8 in.) female quick-connect on the inlet.
● The two (2) pitot lines, (black and white, 6.35 mm (1/4 in.)), have female quick-connects to the Probe
Assembly and 6.35 mm (1/4 in.) male quick-connects to the Source Sampler Console. There is an
additional gas sample line for Orsat analysis (yellow, 6.35 mm (1/4 in.)), which also can be used as a
spare pitot line.
● Multiple thermocouple extension cables for Type-K thermocouples, which terminate with durable full-
size connectors. The connectors have different diameter round pins to ensure proper polarity, and will not
fully connect if reversed. Each thermocouple extension wire in the Umbilical Cable is labeled and color-
coded for temperature measurement of Stack, Probe, Oven (Hot Box), Exit (Cold Box), and Auxiliary
(spare).
● The AC power cable (for Filter Oven and Probe Assembly heaters) terminates with a circular, military-
style connector on each end.
● The Umbilical Adapter connects the outlet of the last impinger to the Umbilical Cable and contains the
exit thermocouple. This adapter serves to relieve strain between the Umbilical Cable and the glassware
train.
● The body of the circular connector is grounded. A line-up guide is placed on each connector’s end, and
the retainer threads should be engaged for good contact. Figure 1-8 illustrates the circular connector with
pins labeled.
● The Umbilical Cable is covered with a woven nylon mesh sheath to restrain the cable and reduce friction
when moving the cable.
17
How to Construct a Glassware Sample Train:
The sample glassware train contains the filter holder for collection of particulate matter, glass impingers for
absorption of entrained moisture, and connecting glassware pieces.
Figure 1-9 below illustrates the glassware of the US EPA Method 5 sampling train.
The order in which a typical US EPA Method 5 glassware train is constructed is as follows:
1. Cyclone Bypass (GN-1), Optional: Cyclone (GN-2) and Cyclone Flask (GN-3)
2. 3 in. Glass Filter Assembly (GNFA-3). Assembly consists of the Filter Inlet (GN-3S), Teflon Filter Disk
or “Frit” (GA-3T), Filter Outlet (GN-3B), Filter Clamp (GA-3CA), and Glass Fiber Filter (GF-3 Series).
3. Double “L” Adapter (GN-8), or alternate GN-8-18K with thermocouple assembly
4. 1st Impinger Modified Greenburg-Smith (GN-9A)
5. U-Tube (GN-11)
6. 2nd Impinger Greenburg-Smith with Orifice (GN-9AO)
7. U-Tube (GN-11)
8. 3rd Impinger Modified Greenburg-Smith (GN-9A)
9. U-Tube (GN-11)
10. 4th Impinger Modified Greenburg-Smith (GN-9A )
18
Chapter 2 Tips from a Stack Tester
Operating Procedures
A drop of penetrating oil
There are several steps to complete before testing for
helps keep the quick-
particulate matter, which include:
connects in good working
1. Set up and inspection of source sampling system, condition. Inspect the pitot
review of calibration records tube openings for damage
2. Test design or misalignment, and
replace or repair them if
3. Site preparation necessary.
4. Sampling equipment calibrations (see Chapter 3)
5. Assembling sampling equipment and accessories,
reagents, sample recovery equipment, and sample storage containers
6. Preliminary measurements of stack dimensions, gas velocity, dry molecular weight, and moisture.
2. Slide the Impinger Case (Cold Box) onto the Modular Sample Case’s heated filter oven (Hot Box),
using the steel slide guides. Check the fit and height of the Sample Case and Umbilical Adapter.
Adjust the steel slides to achieve the desired fit. Engage the spring latch that locks the Impinger Case
into place.
3. Inspect the Probe Liner and Probe Assembly. Wipe clean the quick-connects on the Probe Assembly.
19
4. Slide the Probe Liner into the probe sheath. The liner’s plain end (no ball joint) should come out
approximately 1.27 cm (1/2 in.) at the pitot tube end of the Probe Assembly.
5. Insert the Probe Assembly into the probe clamp that is attached to the Filter Oven and tighten. Then,
carefully insert the outlet ball of the Probe Liner through the hole into the Hot Box. The back of the
sheath should be even with the inside of the Hot Box. Next, plug the Probe Heater electrical plug into
the probe receptacle on the Hot Box.
6. To install a nozzle on the Probe Assembly, consult Figure 2-1. Slide the ferrule system onto the plain
exposed end of the Probe Liner. Substitute high-temperature braided glass cord packing for the
ferrule when stack temperatures are greater than 260°C (500°F) or use liner with integrated nozzle.
Apex Instruments’ Probe Assembly Spare Parts Kit (the bag taped to probe sheath) contains fittings for
two (2) different ferrule installation options: 1) Stainless Steel Single Ferrule and 2) Backer Ring with O-
Ring. The recommended configurations for different liner options are detailed below:
Stainless Steel Liner: Stainless Steel Single Ferrule or Backer Ring with O-Ring
Glass Liner: Backer Ring with O-Ring, or PTFE Single Ferrule (Optional), or Glass-filled PTFE
Single Ferrule (Optional).
Figure 2-1: Installation of Stainless Steel and Glass Probe Nozzle Connectors
7. Thread the 15.875 mm (5/8 in.) union onto the nut welded to the probe sheath. This is a compression
fitting that is tapered to seal the ferrule system inserted on the Probe Liner.
Important: Tighten the fitting until the liner has a leak-tight seal, but DO NOT OVERTIGHTEN.
8. Connect the glassware sampling train completely in the Filter Oven (Hot Box) and Impinger Case
(Cold Box) (see Figure 1-9 and the related “How to Construct a Glassware Sample Train” at the end
of Chapter 1). Tighten all joints using the Ball Joint Clamps. The final connection is the Umbilical
Adapter, which slides into the clamp on the outside of the Impinger Case. Again, do not load the
Filter Assembly with a filter, and do not fill the impingers because this is a “dry run” set-up.
9. To connect the Umbilical Cable to the Modular Sample Case, first connect the Umbilical Cable
circular connector plug to the receptacle on the side of the Filter Oven (see Figure 1-6). Next, connect
the labeled Umbilical Cable thermocouple plugs into the receptacles on the Filter Oven, Probe
Assembly, and Umbilical Adapter. Then, insert the Umbilical Cable sample line female quick-
20
connect into the Umbilical Adapter male quick-connect. Finally, insert the Umbilical Cable female
pitot line quick-connects into the Probe Assembly male quick-connects.
10. To connect the Umbilical Cable to the Source Sampler Console, first connect the Umbilical Cable
circular connector plug to the receptacle on the front panel of the Source Sampler Console. Then,
connect the labeled Umbilical Cable thermocouple plugs into the receptacles on the Source Sampler
Console front panel.
Next, insert the Umbilical Cable sample line male quick-connect into the Source Sampler Console
female quick-connect. Finally, insert the Umbilical Cable pitot line male quick-connects into the
Source Sampler Console female quick-connects (labeled + and −). The pitot lines are colored to
differentiate the positive and negative lines and ensure the connections are consistent between the
pitot tube and Source Sampler Console.
11. To connect the Vacuum Pump Assembly to the Source Sampler Console, first wipe the quick-
connects clean, then connect the pressure and vacuum hoses on the Vacuum Pump Assembly to the
pump connections located on the lower left of the Source Sampler Console front panel. Then, connect
the power cord of the Vacuum Pump Assembly to the receptacle labeled PUMP on the Source
Sampler Console.
12. Plug the Source Sampler Console into an appropriate electrical power source.
21
2. Test Design
3. Site Preparation
Preparing the site so that sampling equipment can be positioned correctly is frequently the most difficult part
of the sampling process. When the sample ports do not have a platform or catwalk, then scaffolding must be
erected to reach the sampling site. At many sites, operators must use their ingenuity to get the sampling
equipment to the sampling location.
When selecting the site for sample ports, keep in mind that the distance from the probe to the bottom of the
sample case is about 33 cm (13.5 in.). This means that when traversing the stack, the sampling equipment
needs 33 cm of clearance below the port level to avoid bumping into guardrails or other structures.
To calculate the clearance needed along the sample port plane, start with the effective probe length (which is
stack diameter plus port nipple length) and add at least 91 cm (36 in.) to accommodate the length of the
sample case (Filter Oven, Impinger Case, and probe clamp).
Figure 2-2 on the next page illustrates the clearance zones required.
22
Figure 2-2: Clearance Zones at Stack for Isokinetic Sampling Train
23
Apex Instruments Product Highlight
If you can’t find a solution for sampling train clearance problems, Apex
Instruments can provide one. We offer a Non-Rigid Method 5
sampling train with a separate and/or miniature heated Filter Box (SB-
2M), which allows you to put the Cold Box on the sampling platform,
where it is connected by the sample line and Umbilical Adapter (GA-
104).
24
Apex Instruments Product Highlight
There are two ways to mount the isokinetic sampling system (which includes the Filter Oven and Impinger
Case) on a stack:
1. Assemble a monorail system with lubricated roller hook above each sample port, or
2. Construct a wooden platform slide apparatus (where feasible).
25
Figure 2-5 illustrates an isokinetic sampling system mounted on a monorail system above a sample port with
a tee bracket system. Another way to do a monorail mounting is when there is an angle iron, with a hole or
an eyehook, welded to the stack.
Illustrated below is a complete stack set-up in two different configurations: Figure 2-7 shows the Hot
Box/Cold Box together (SB-1) and Figure 2-8 shows the Filter Oven and Impinger Case separated (SB-2M
and SB-3).
26
Figure 2-7: Stack Platform Set-up with Modular Sample Case on Monorail
27
Figure 2-8: Stack Set-up with Hot Box on Monorail Separated from Cold Box
Remember to use checklists when assembling the sampling equipment, reagents and auxiliary supplies for a
test. See Appendix A for checklists of recommended equipment, spare parts and reagents for isokinetic
sampling. Keep in mind that you may not need the entire list at the test site. Also, refer to Section 3 of US
EPA Method 5 for the list of reagents required to perform an isokinetic particulate test.
Before attempting to calculate the parameters needed for isokinetic sampling – such as probe nozzle size,
ΔH/Δp ratio (K factor), gas sample volume, etc. – you first must determine several preliminary values
described in Table 2-1 on the next page.
28
Tips from a Stack Tester
Did you know? Methods 1 through 4 are like building blocks, providing the
preliminary values needed to complete Method 5 sampling. Thus, when you perform a
sampling run using Method 5, you will need to know Methods 1 through 4 procedures
to collect the data needed for Method 5 sampling and calculations.
Through the rest of Chapter 2, we have provided concise and easy-to-use information
for performing US EPA Methods 1 through 5.
Method 1
Determining Sample and Velocity Traverse Points
Method 1 is the first step toward collection of a representative sample to measure a stack’s particulate
concentration and mass emission rate. Method 1 is applicable to gas streams flowing in ducts, stacks and
flues. Since the velocity and particle concentration in the stack are not uniform, you must traverse the cross-
section to obtain a representative sample.
The number of required traverse points depends on the shape - or straightness - of the stack or duct.
Straighter lengths of stack or duct produce flow streamlines that are more uniform, so fewer traverse points
are needed to obtain a representative sample. Conversely, if your sampling site is close to bends in the stack
or other flow disturbances, you will need to use more traverse points to obtain a representative sample.
We can identify appropriate sampling sites by measuring their distance from any type of flow disturbance in
the stack. Disturbances can be bends, transitions, expansions, contractions, stack exit to atmosphere, flames,
or the presence of internal installations such as valves or baffles.
To calculate distance between the sampling site and flow disturbances, first we measure the internal stack
diameter. Then, we use that length to determine how many “diameters” are in between the sampling site and
the flow disturbance.
Figure 2-9 on the next page depicts how stack diameters are used to measure distance from a flow
disturbance such as a bend. The example on the left shows an acceptable sampling site, which is eight (8)
diameters downstream from a bend and two (2) diameters upstream from the stack exit. The example on the
right is not an acceptable sampling site because it is too close to a flow disturbance (the bend in the stack).
30
Figure 2-9: Visualizing Stack Diameters from Flow Disturbances
Example: The dotted lines in Figure 2-10 represent a Distance A measurement of 1.7 duct diameters and a
Distance B measurement of 7.25 duct diameters. Looking at where the dotted lines intersect the solid
traverse points line, Distance A requires 16 traverse points, while Distance B requires 12 traverse points.
Always choose the higher of the two numbers. In this example, the sampling site requires 16 total traverse
points. The 16 points are arranged in two lines of eight points each which are 90° apart, forming a cross
shape.
32
Figure 2-10b: Example of How to Determine Number of Traverse Points (Non-Particulates)
● For circular stacks with diameters greater than 60 cm (24 in.), the minimum number of traverse
points required is twelve (12), or six (6) in each of two directions 90° apart (see Figure 2-11). This
applies when disturbances are eight (8) or more duct diameters upstream and two (2) or more
downstream.
● For circular stacks with diameters between 30 and 60 cm (12 and 24 in.), the minimum number of
traverse points required is eight (8), or four (4) in each of two directions 90° apart.
● For stacks less than 30 cm (12 in.) in diameter, refer to Method 1A for calculating traverse points.
● The minimum number of traverse points required for rectangular stacks is nine, or 3 x 3.
For rectangular stacks or ducts, first calculate an equivalent diameter using the following equation:
33
Determine the Location of Traverse Points
Circular Stacks
After determining the number of traverse points, you must calculate the location of each traverse point. The
method for locating the traverse points for circular stacks is as follows:
1. Divide the number of traverse points by four (4). The resulting number will give you the number of
concentric circles of equal area to use in your sample point matrix. In the example illustrated in
Figure 2-11, twelve (12) traverse points divided by four (4) equals three (3).
2. Bisect the circles twice, cutting them into quarters, as shown below.
3. Place the sample points in the centroid (center of mass) of each equal area, as shown in Figure 2-11.
35
Tips from a Stack Tester
36
Figure 2-13: Illustration of Marking Traverse Points on Probe Assembly
Rectangular Stacks
For rectangular stacks, the centroids for placing the traverse points are much easier to determine, as shown in
Figure 2-12.
37
Method 1A
Sample and Velocity Traverses,
Small Stacks or Ducts
Method 1A is the same as Method 1, except for the special provisions that apply to small circular stacks or
ducts where the diameter is between 4 and 12 inches (10.2 cm (4 in.) ≤ D ≤ 30.5 cm (12 in.)), or for small
rectangular ducts where the area is between 12.57 in. 2 and 113 in.2 (81.1 cm2 (12.57 in.2) ≤ A ≤ 729 cm2 (113
in. 2)).
Important: You must use a standard type pitot tube for the velocity measurements. Do not attach the tube to
the sampling probe.
Did you know? In small diameter stacks or ducts, the conventional Method 5 stack
assembly (which consists of a Type-S pitot tube attached to a sampling probe
equipped with a nozzle and thermocouple) blocks a significant portion of the duct’s
cross-section, resulting in inaccurate measurements. Therefore, for particulate matter
sampling in small ducts, measure the gas velocity either:
➢ Downstream of the sampling nozzle (for unsteady flow conditions), or
➢ In the same sample port alternately before and after sampling (for steady flow
conditions).
The procedure for determining sampling location, traverse points, and flow rate (preliminary or otherwise) in
a small duct is as follows:
1. Use Method 1 to locate traverse points for each sampling site and choose the highest of the four
numbers for the total traverse point number.
2. For measurements of particulate matter (PM) with a steady flow, or velocity with either a steady or
unsteady flow, select one sampling location and use the same criterion as described in Method 1.
3. For PM with a steady flow, conduct velocity traverses in the same port before and after PM sampling
to demonstrate steady state conditions, i.e., within ± 10% (vf/vi ≤ 1.10).
4. For PM with an unsteady flow, monitor velocity and sample PM at two separate locations
simultaneously. The velocity port should be downstream of the sampling port. See the location of the
two ports labeled in Figure 2-14 on the next page.
38
Figure 2-14: Set-up of EPA Method 1A Small Duct Sampling Locations
39
Method 2
Stack Gas Velocity and Volumetric Flow Rate
Method 2 is used to measure the average velocity and volumetric flow rate of the stack gas stream. To
determine the average gas velocity in the stack, you must measure the gas density and the average velocity
head with a Type-S (Stausscheibe or reverse-type) pitot tube. Note: Have a thorough knowledge of at least
Method 1 before proceeding, as some of the steps reference Method 1 procedures.
Did you know? Do not use Method 2 with measurement sites that are less than eight
stack diameters downstream or two stack diameters upstream from a flow disturbance.
Method 2 also cannot be used for direct measurement in cyclonic or swirling gas
streams (see section 11.4 of Method 1 to determine these conditions).
When faced with cyclonic or swirling gas streams, you must use alternative procedures
such as installing straightening vanes, calculating the total volumetric flow rate
stoichiometrically, or moving to another measurement site where the flow is
acceptable.
The
equation
for average gas velocity in a stack or duct is:
To calculate the average stack gas velocity (Vs), you must first obtain values for the molecular weight and
moisture (refer to Method 3 and Method 4 sections). The stack gas molecular weight dry basis (M d) is
corrected to the wet basis (Ms) using the moisture fraction (Bws) with the equation:
Method 3 Method 1
Selection of Traverse Points
Stack Gas Molecular Weight
Options:
Sample with Orsat analysis
Sample with Fyrite analysis
Assign 29.0 if air
Method 2
Assign 30.0 if combustion Stack Gas Velocity
Method 4
Stack Gas Moisture
Options:
Reference Method
Approx Method (Midgets)
Drying Tubes
Wet Bulb-Dry Bulb
Psychrometric chart
Previous experience
41
Using the Pitot Tube
Use a pitot tube connected to an inclined manometer to make velocity measurements in a duct. Alternatives
to the inclined manometer are a magnehelic pressure gauge or an electronic manometer, but you must
calibrate each of these devices periodically against an oil-filled inclined manometer (see Calibration in
Chapter 3).
The S-type pitot tube has a fixed coefficient of 0.84 if it is manufactured and maintained to meet the
geometric specifications of Method 2. You may also use a standard or P-type (Prandl) pitot tube with a
coefficient of 0.99 for these measurements.
Insert the S-type pitot tube into the stack with one leg (hole opening) pointing into the direction of gas flow,
as shown in Figure 2-16. The leg pointing into the flow stream measures impact pressure (P i) of the gas
stream, while the opposite leg (pointing away from the flow) measures wake pressure (Pw).
The velocity pressure (Δp) is the difference between the impact and wake pressures:
42
Determine Flow Rate
The procedure for determining flow rate (preliminary or other) in a stack gas stream is as follows:
1. Fill out the top section of a Velocity Traverse field data sheet.
2. Mark the pitot tube with the traverse points according to Method 1 (see “Pinpoint and Mark Traverse
Points” in the Method 1 section).
3. Assemble the apparatus for flow velocity measurement using one of these configurations:
a. Pitot tube with thermocouple, pitot and thermocouple extension lines, inclined manometer, and
temperature display device, or
b. Probe Assembly, Umbilical Cable, and inclined manometer on Meter Console.
4. Pitot Line Leak-Check: Conduct a pre-test leak-check of the
pitot and lines by blowing lightly into the positive (impact)
side of the pitot tube opening until at least 7.6 cm (3 in.) H 2O
Tips from a Stack Tester
registers on the manometer; then, close off the impact
opening. The pressure should remain stable for at least 15
seconds. Do the same, except suck lightly, for the negative
(wake) side.
5. Level and zero the manometer. If using a separate manometer,
If the pitot tube is dirty or
cup a hand or place a glove over the pitot opening to prevent
chemically contaminated,
wind from affecting the zero adjustment. If using the Source
attach a short piece of
Sampler Console, use the Zero Manometer switch.
flexible tubing to the pitot
Periodically check that the manometer is zeroed and leveled
leg for leak checking, and
between ports.
pinch it closed to hold the
6. Insert the pitot tube into the stack to a marked traverse point pressure.
and seal off the port opening with a rag or towel to prevent
ambient effects. Starting with the farthest points, measure the
velocity head and temperature and record the values on the
field data sheet. Allow the temperature reading to stabilize
before recording it.
7. Move to each
subsequent
Did you know? The S-type pitot tube is most often used in stack
traverse point, testing because it is:
reseal the port
➢ Compact and attaches easily to a Method 5 probe assembly,
and record the
velocity head and ➢ Relatively easy to manufacture,
temperature. ➢ Relatively insensitive to plugging in stack gas streams, and
Switch to the ➢ Relatively insensitive to yaw and pitch errors.
next port and
repeat traverse
sampling.
43
8. Conduct a post-test leak-check (which is mandatory to prove that no leakage occurred) as described in
Step 4 above, and record it on field data sheet.
9. Measure the static pressure in the stack. One reading is adequate.
10. Determine the barometric pressure at the level of the sample port.
11. Calculate the average stack temperature from the traverse readings and record it.
12. Calculate the average square root of velocity head by taking the square root of each velocity head reading
and averaging the square roots (sum the square roots and then divide by the number of traverse points).
Then, record it on the field data sheet.
44
After recording the static pressure (Pg), you must convert the value from mm H2O to mm Hg (in. H2O to in.
Hg) prior to inputting it into the velocity equation as Ps (absolute stack gas pressure). The density of mercury
is 13.6 times that of water, so the conversion equation is:
Correct the station’s barometric pressure reading by subtracting 0.832 mm Hg for every 100 m that the
weather station is above sea level (0.1 in. Hg for every 100 ft.). To get valid results, you also must know the
sampling site’s elevation.
where Pr = Barometric pressure at site ground level or at weather station, mm Hg (in. Hg)
A = Elevation at ground level or at weather station, m (ft. above sea level)
B = Elevation of the sampling site, m (ft. above sea level)
After calculating the average stack gas velocity (Vs, see the beginning of the Method 2 section), you can find
the volumetric flow rate. Begin by determining the area (As) for a circular stack with this equation:
Where π = 3.14159
45
Ds = Diameter of a circular stack
The formula to find the area of a rectangular stack is:
Now, you can calculate the stack gas volumetric flow rate (actual, standard, and dry standard) using the
following equations:
46
Method 3
Gas Analysis for Dry Molecular Weight
Method 3 is used to measure the percent concentrations of carbon dioxide (CO2), oxygen (O2), and carbon
monoxide (CO) if greater than 0.2%. Nitrogen (N2) is calculated by difference. With this data, you can
calculate the stack gas dry molecular weight, or density, and then apply that data to the equation for stack gas
velocity.
From the gas composition data, you can calculate the amount of excess air for combustion sources. In
jurisdictions where particulate emissions are regulated on a concentration basis, such as mg/m3, use the gas
composition data to correct the concentration results to a reference diluent concentration, for example 7% O 2
or 12% CO2. Note: Before beginning, you first must have a thorough knowledge of Method 1 procedures,
which are referenced in Method 3.
Analyze gas samples using either an Orsat or Fyrite analyzer. Figure 2-17 on the next page depicts the
options for sample collection and analysis.
47
Gas Analysis
Sampling Options
Options
Single-point
Grab
Sampling Orsat
Analyzer
Single-Point
Bag
Integrated
Sample
Sampling
Fyrite
Multi-Point Analyzer
Integrated
Sampling
Figure 2-18 on the next page illustrates an Orsat Analyzer connected to a bag sample collection enclosure.
For a more detailed discussion of gas analysis using an Orsat Analyzer, please refer to Apex Instruments’
Combustion Gas (ORSAT) Analyzer, Model VSC-33, User’s Manual and Operating Instructions, or the
operating instructions provided with the Fyrite Analyzer.
48
Figure 2-18: Illustration of Orsat Analyzer and Gas Sample Bag Container
49
Method 4
Moisture Content of Stack Gas
Method 4 is used to determine the moisture content of stack gas, which is needed to calculate emission data.
A gas sample is extracted at a constant rate from the source, moisture is removed from the sample stream,
and moisture content is determined either volumetrically or gravimetrically. Before using this method, you
should have a thorough knowledge of at least Methods 1, 5 and 6.
There are two separate procedures for determining moisture content in stack gases:
➢ The Reference Method, which gives accurate measurements of moisture needed to calculate
emission data, and
➢ The Approximation method, which gives a close estimate of percent moisture to aid in setting
isokinetic sampling rates before a pollutant emission sampling run.
The approximation method is only a suggested approach. There are acceptable alternatives for approximating
moisture content, for example:
● Wet bulb/dry bulb techniques (applicable to gas streams less than 100°C),
● Stoichiometric calculations (applicable to combustion sources),
● Condensation techniques,
● Drying tubes, and
● Previous experience testing at a stack.
50
Tips from a Stack Tester
Did you know? The Reference Method is almost always performed simultaneously
with a pollutant emission measurement run. You also can use the Reference Method
when you need to correct to a dry basis your continuous monitoring for pollutants such
as SO2, NOx or O2.
51
Figure 2-19: Set-up of Cold Box with Sample Frame and Probe Clamp
Part 2: Sampling
1. Assemble and set up the sampling train.
2. Turn on the probe heater and, if applicable, the filter heating system to temperatures of about 120°C
(248°F). Allow time for the temperatures to stabilize. Place crushed ice in the ice bath container
(Cold Box) around the impingers.
3. Optional: Leak-check the sampling train by disconnecting the probe from the first impinger or, if
applicable, from the filter holder. Plug the inlet to the first impinger (or filter holder) and pull a 380
mm (15 in.) Hg vacuum. It is not acceptable to have a leakage rate greater than 4% of the average
sampling rate, or 0.00057 m3/min (0.020 cfm), whichever is less. If the leakage rate exceeds the
52
allowable rate, either reject the test results or correct the Tips from a Stack Tester
sample volume as in section 12.3 of Method 5.
Following the leak check, reconnect the probe to the
sampling train.
4. Position the probe tip at the first traverse point. Sample
at a constant (± 10%) flow rate. Record data on a field
data sheet.
5. Traverse the cross-section, sampling at each traverse Make sure to wipe off
point for an equal period of time. moisture from the outside
6. Add more ice and, if necessary, salt to maintain a of each impinger before
temperature of less than 20°C (68°F) at the silica gel weighing it. Do not weigh
impinger exit. impingers with U-tubes
connected.
7. After the last traverse point of the cross-section, turn off
the sample pump, switch to the next sample port, and
repeat steps 5 and 6 above.
8. After completing sampling, disconnect the probe from the first impinger (or from the filter holder).
9. Mandatory: Leak-check the sampling train, as required at the end of the run (see step 3 above).
53
Method 4: Approximation Method
The Method 4 approximation method specifies use of midget impingers and a Source Sampler Console sized
for midget impinger trains, such as the one used for Method 6 sampling of SO2.
Did you know? Many stack testers perform preliminary moisture measurements for
input into their isokinetic calculations nomograph by using a full-size sampling train
and collecting approximately 0.283 sm3 (10scf) of gas sample. These runs take about
15 to 20 minutes.
Part 2: Sampling
1. Assemble and set up the sampling train.
2. Turn on the probe heater and, if applicable, the filter heating system to temperatures of about 120°C
(248°F). Allow time for the temperatures to stabilize. Place crushed ice in the ice bath container
(Cold Box) around the impingers.
3. Optional: Leak-check the sampling train by disconnecting the probe from the first impinger or, if
applicable, from the filter holder. Plug the inlet to the first impinger (or filter holder) and pull a 380
mm (15 in.) Hg vacuum. It is not acceptable to have a leakage rate greater than 4% of the average
sampling rate, or 0.00057 m3/min (0.020 cfm), whichever is less. If the leakage rate exceeds the
allowable rate, either reject the test results or correct the sample volume as in section 12.3 of Method
5. Following the leak check, reconnect the probe to the sampling train.
4. Position the probe tip well into the stack. Sample at a constant (± 10%) flow rate of twenty-one (21)
lpm until the dry gas meter registers about 30 liters, or until visible liquid droplets carry over from the
first impinger to the second. Record initial and final data on a field data sheet.
5. Add more ice and, if necessary, salt to maintain a temperature of less than 20°C (68°F) at the silica
gel impinger exit.
6. Mandatory: Leak-check the sampling train as described in Step 3 above. After the leak-check,
reconnect the probe to the sampling train.
Part 3: Sample Recovery
54
1. Disassemble the impinger glassware and weigh each impinger or drying tube to the nearest 0.5 g.
Record weight data on a field data sheet.
2. Verify constant sampling rate.
3. Calculate the stack gas moisture percentage (see the equations at the end of this section).
Part 2: Sampling
1. Assemble and set up the sampling train.
2. Turn on the probe heater and, if applicable, the filter heating system to temperatures of about 120°C
(248°F). Allow time for the temperatures to stabilize. Place crushed ice in the ice bath container
(Cold Box) around the impingers.
3. Optional: Leak-check the sampling train from the first impinger inlet or, if applicable, the filter
holder (see leak-check procedures described in the Midget Impinger Train section on the previous
page).
4. Position the probe tip well into the stack. Sample at a constant (± 10%) flow rate of less than twenty-
one (21) lpm (0.75 cfm) until the dry gas meter registers 0.283 m3 (10 cf). Record initial and final
data on a field data sheet.
5. Add more ice and, if necessary, salt to maintain a temperature of less than 20°C (68°F) at the silica
gel impinger exit.
6. Mandatory: Leak-check the sampling train (as detailed in Part 2, Step 3 on the previous page).
55
Equations for Calculating Stack Gas Moisture Content
To calculate the stack gas moisture content (Bws), first use the following equations to compute the sample gas
volume (Vm(std)) and gas moisture volume (Vwc(std)):
Method 5 involves withdrawing particulate matter (PM) isokinetically from the source and collecting it on a
temperature-controlled glass fiber filter. The PM mass is determined gravimetrically after the removal of
uncombined water. To perform Method 5, you must have a thorough understanding of Methods 1, 2 and 3.
To maintain an isokinetic sampling rate, you will need to calculate the appropriate probe nozzle size and the
optimal ratio of orifice tube pressure (ΔH) to stack gas velocity (Δp). This ratio is called K-factor.
Determine nozzle size and K-factor in one of the following ways:
➢ Calculate by hand or on a worksheet (see worksheets in Appendix D).
➢ Use a specially designed stack testing slide rule nomograph (M5A-1M or M5A-1), as shown in
Figure 2-21.
➢ Use a pre-programmed handheld calculator (M5A-C).
➢ Input figures into specialized spreadsheets for data collection and reduction (ISOCALC 2.0) using a
personal or laptop computer, as shown in Figure 2-21.
Figure 2-21: Stack Sampling Slide Rule and Laptop Computer with IsoCalc 2.2
57
➢ Ps: Stack gas pressure. Measure this before the sample run, or use barometric pressure if the static
pressure of the stack is very low (such as when sample ports are near the stack exit).
➢ ΔH@: Source Sampler Console orifice calibration factor. This is determined from the laboratory
calibration and should be readily available on-site (see Calibrations).
➢ Tm: Meter temperature. Temperature at the meter rises about 14°C (25°F) above ambient temperature
due to heat from the vacuum pump. Measure the ambient temperature at the Source Sampler Console
site.
➢ Pm: Meter pressure. Same as barometric pressure.
58
K-Factor Calculations
After selecting the appropriate nozzle from the Nozzle Set (shown in
Figure 2-22), calculate the K-factor ratio used to maintain an
isokinetic sampling rate at each traverse point for the sampling test
run. K-factor refers to the ratio of ΔH/Δp such that ΔH = KΔp. Use
the following equation:
59
Method 5 Test Procedure
Part 1: Pre-Test Preparation (Before Traveling to Site)
1. Check filters visually against light for irregularities, flaws or pinhole leaks. Label the filters on the
back side near the edge using numbering machine ink.
2. Desiccate the filters at 20° ± 5.6°C and ambient pressure for more than 24 hr. Then weigh them at
intervals of greater than 6 hours to a constant weight (less than 0.5 mg change from previous
weighing). Record results to the nearest 0.1 mg. During each weighing, do not expose the filter to the
laboratory atmosphere for more than 2 minutes, or to a relative humidity of more than 50%.
3. Optional: If you are measuring condensable or back-half particulate matter, run analytical blanks of
the deionized/distilled water to eliminate a high blank on actual test samples.
4. Clean the Probe Liners and Probe Nozzles internally by brushing first with tap water, then
distilled/deionized water, followed by reagent-grade acetone. Rinse the Probe liner with acetone and
allow to air-dry. Inspect visually for cleanliness and repeat the procedure if necessary. Cover the
Probe Liner openings to avoid contamination. Keep nozzles in a case to avoid contamination or
damage to the knife-edge.
Note: Special cleaning procedures may be required for other test methods (for example, metals or
dioxin).
5. Clean the Glassware (Filter Assemblies, Impingers and Connecting Glassware) internally by wiping
grease from the joints, washing with glass cleaning detergent, rinsing with distilled/deionized water,
and then rinsing again with reagent-grade acetone, before allowing to air-dry. To avoid
contamination, cover all exposed openings with parafilm, plastic caps, serum caps, ground-glass
stoppers or aluminum foil (do not use foil with metals).
Note: Special cleaning procedures may be required for other test methods (for example, metals or
dioxin).
60
1. Mark the Probe Assembly with heat-resistant tape or “Wite-Out” to denote the proper distance into
the stack or duct for each sampling point (See Method 1).
2. Insert the Probe Nozzle into the probe sheath union and hand-tighten the union fitting. Avoid over-
tightening to prevent cracking of the glass probe liner. Keep the nozzle tip and ball joint on the glass
probe liner covered until you complete the train assembly and are about to begin sampling. Secure the
Probe Assembly to the Sample Case by tightening the probe clamp.
3. Prepare each set of impingers for a sampling run:
a. Impingers 1 & 2: 100 ml water in each
b. Impinger 3: Empty
c. Impinger 4: 200 to 300 g of silica gel
TIP: You can prepare more than one sampling run at once by using multiple sets of glassware!
4. Weigh each impinger to the nearest ± 0.5 g using a top-loading electronic balance (BAL-1200) and
record initial weights on a field data sheet.
5. Assemble the impingers in the Cold Box with U-tubes, Double “L” Adapter, and the Sample
Case/Umbilical Adapter, using Ball Joint Clamps or Clips. See Figure 2-24.
61
Figure 2-25: Exploded View of Filter Assembly
6. Using tweezers or clean disposable surgical gloves, place the tared filter on the grooved (inlet) side of
the Teflon Frit (TFE) filter support in the Filter Holder. Check the filter for tears after placement, and
center it on the filter support. Assemble the Filter Holder (as shown in Figure 2-25), and tighten the
clamps around the Filter Holder to prevent leakage around the O-ring. Record filter number on the
field data sheet.
7. Using Ball Joint Clamps, connect the Filter Holder and Cyclone Bypass (GN-1) in the Hot Box to the
Probe Liner ball joint and to the “L” Adapter. Close the Hot Box doors and fasten shut.
8. Connect the Umbilical Cable electrical and pitot tube lines to the assembled sampling train and to the
Source Sampler Console. If applicable, also connect the Orsat line.
9. Place the assembled sampling train near the first sample port, either on the monorail or with other
support.
10. Turn on and set Probe and Hot Box heaters. Allow the Hot Box and Probe to heat for at least 15
minutes before starting the test, and then make periodic checks and adjustments to ensure the desired
temperatures. Check all thermocouple connections by dialing through each selection and noting
ambient or heated temperatures. Place crushed ice and a little water around the impingers.
62
11. Optional: Leak-check the sampling train (see Leak-Check Procedure for Isokinetic Sampling Trains
in Reference Method 4 on page 50, and Pitot Tube and Line Leak-Check Procedures in Method 2 on
page 41).
8. Record readings for the ΔH manometer and pump vacuum, as well as temperatures for stack gas,
DGM, filter box, probe, and impinger exit.
9. Record the ID numbers for the DGM, thermocouples, pitot tube, and Sample Box.
10. If simultaneously running a Method 3 gas bag collection, turn on the Orsat pump. Turn Orsat pump
off during port changes.
63
11. Traverse the stack cross-section for the same time period at each point without turning off the pump,
except when changing ports.
Important: Do not bump the probe nozzle into the stack walls.
a. Make periodic adjustments to maintain the Hot Box temperature (around the filter holder) at
the proper level.
b. Monitor the Δp (gas velocity head) during sampling of each point. If the Δp changes by more
than 20%, record another set of readings.
c. Periodically check the level and zero of the manometers and readjust if necessary.
d. Record DGM readings at the beginning and end of each sampling time increment, before and
after each leak-check, and when sampling is halted.
e. Take other readings (ΔH, temps, vacuum) at least once at each sample point during each time
increment to ensure the ΔH/Δp isokinetic ratio is maintained.
f. Add more ice and, if necessary, salt to maintain a temperature of less than 20°C (68°F) at the
silica gel impinger exit.
12. At the end of the sample run, turn off the Coarse Valve, remove the probe and nozzle from the stack,
turn off the pump and heaters, and record the final DGM reading.
13. Mandatory: Leak-Check the sampling train (See Method 4, page 50) at the maximum vacuum
achieved during the sample run. Record leak-check results on field data sheet.
14. Mandatory: Leak-Check the pitot lines (See Method 2, page 41). Record results on field data sheet.
15. Allow the probe to cool. Wipe off all external particulate material near the tip of the probe nozzle and
cap the nozzle to prevent contamination or sample loss.
Hint: Open the Hot Box doors to allow the filter holder to cool.
16. Before moving the sampling train to the clean-up site, disconnect the probe from the Cyclone Bypass
inlet and cover both ends. Avoid losing any condensate that might be present. Disconnect the Filter
Holder from the “L” Adapter and cap off the Filter Holder.
17. Disconnect the Umbilical Cable from the Sample Box and cover the last impinger outlet and first
impinger inlet. Disconnect the Cold Box from the Hot Box. Now, the Probe/Nozzle Assembly, Filter
Holder, and impinger case are ready for sample recovery.
18. Transfer the probe and filter-impinger assembly to an area that is clean and protected from the wind.
64
5. Use more silica gel in impinger 4, if necessary, but ensure that there is no entrainment or loss during
sampling.
Hint: Loosely place cotton balls or glass wool in the neck of the silica gel impinger outlet stem to
avoid spillage.
6. If you are using a different type of condenser (other than impingers), measure the amount of moisture
condensed either volumetrically or gravimetrically.
7. For moisture content, measure the impinger contents volumetrically before and after a sampling run.
Use a pre-weighed amount of silica gel in a shipping container. Then, after the run, empty the silica
gel back into the container for weighing at another time.
Figure 2-28: Recover Silica Gel for Weighing Figure 2-29: Determine Moisture Volumetrically
8. If you expect the total particulate catch to exceed 100 mg or more, or when water droplets are present
in the stack gas, use a Glass Cyclone between the probe and Filter Holder.
9. If you have difficulty maintaining isokinetic sampling due to high pressure drops across the filter
(with a high vacuum on the gauge), replace the filter.
10. Use a single train for the entire sampling run, except when:
a. Simultaneous sampling is required in two or more
separate ducts, or
Tips from a Stack Tester
b. Simultaneous sampling is required at two or more
different locations within the same duct, or
c. Equipment failure necessitates a change in trains.
In all other situations, obtain approval from the regulatory
agency before using two or more trains.
11. When using two or more trains, analyze separately the front- Whenever practical, make
half and (if applicable) impinger catches from each train, every effort to use
unless you used identical nozzle sizes on all trains. In this borosilicate glass or quartz
case, you may combine the front-half catches (and the probe liners. Metal liners
impinger catches) and perform one analysis of the front-half will bias particulate matter
catch and one analysis of the impinger catch. Consult with results higher than actual.
65
the regulatory agency for details concerning
the calculation of results when using two or Tips from a Stack Tester
more trains.
12. If using a flexible line between the first
impinger or condenser and the Filter Holder,
disconnect the line at the Filter Holder and let
any condensed water or liquid drain into the
impingers or condenser. Instead of trying to catch the probe
13. Do not cap off the probe tip too tightly while rinse with a glass funnel and sample
the sampling train is cooling down as this container (which can easily result in
would create a vacuum in the Filter Holder, sample loss), clamp an Erlenmeyer
which may draw water from the impingers into flask outfitted with a female ball joint
the Filter Holder. on the probe liner ball joint when
conducting the probe rinse procedure.
If the probe is short, one person can
Part 6: Sample Recovery perform the brushing and rinsing.
Note: Sample Recovery is an extremely important step
that requires a high level of precision. If sample loss
occurs, your results will be biased low, and if sample
contamination occurs, your results could be biased high.
1. Place 200 ml of acetone from the wash bottle you are using for clean-up in a glass sample container
labeled “Acetone Blank.”
2. Inspect the train prior to and during disassembly and note any abnormal conditions on the data sheet.
3. Container No. 1: Filter
a. Using a pair of tweezers (TW-1) and/or clean disposable surgical gloves, carefully remove the
filter from the Filter Holder and place it in its identified petri dish container. If necessary, fold
the filter so that the particulate matter cake is inside the fold.
b. Using a nylon bristle brush (DB-3) and/or a sharp-edged blade (LS-1), carefully transfer to the
petri dish any particulate matter (PM) and/or remaining pieces of filter or filter fibers that
adhered to the filter support or gasket.
Figure 2-31: Rinsing Probe Nozzle Figure 2-32: Brushing Probe Nozzle
67
Figure 2-33: Front Half of Acetone Rinse Samples
At the conclusion of each sampling run, it is prudent to calculate the stack gas moisture (for the next
sampling run), as well as the average isokinetic rate.
To calculate the stack gas moisture content (Bws), use the following equations to compute the sample gas
volume (Vm(std)) and gas moisture volume (Vwc(std)):
68
where K2 = 0.001335 m3/g (metric units), or 0.04715 ft3/g (English units)
Wf = Final weight of water collected, g
Wi = Initial weight of water collected, g
Next, calculate the average stack gas velocity. The equation for average gas velocity in a stack or duct is:
69
Bws = Proportion of water
vapor, by volume, in the gas stream
Tips from a Stack Tester
70
Recommended Reading List for Isokinetic Sampling
● Code of Federal Regulations. Title 40. Part 60, Appendix A. Office of the Federal Register. National
Archives and Records.
● Compliance Test Coordination and Evaluation. Workshop Manual. U.S. Environmental Protection
Agency. APTI 01-94a. 1994.
● Jahnke, J. A., et al. Source Sampling for Particulate Pollutants. Student Manual, APTI Course 450.
Edition 3.0. Raleigh, NC: North Carolina State University, 1995.
● Manual for Coordination of VOC Emissions Testing Using EPA Methods 18, 21, 25, and 25A. U.S.
Environmental Protection Agency. EPA 340/1-91-008. September 1991.
● Quality Assurance Handbook for Air Pollution Measurement Systems. Vol. 3. Stationary Source
Specific Methods, Section 3.4. U.S. Environmental Protection Agency. EPA-600/4-77-027b. 1988.
● Rom, J. J. Maintenance, Calibration, and Operation of Isokinetic Source Sampling Equipment.
Publication No. APTD-0576. Office of Air Programs. U.S. Environmental Protection Agency.
Research Triangle Park, NC. 1972
Chapter 3
Calibration and Maintenance
Establishing and adhering to a routine maintenance program is beneficial in two ways. First, properly
maintaining the isokinetic sampling system helps to ensure trouble-free operation. Secondly, carefully
documenting your maintenance and calibration system will help you obtain the most accurate results during
stack testing activities.
This chapter describes calibration, maintenance and troubleshooting procedures for the various subsystems
of the isokinetic sampling system.
71
Calibration Procedures
It is critical to create and maintain a regularly scheduled calibration and record-keeping protocol for your
stack testing program. If you fail to ensure proper calibrations, you will not be able to verify that the test was
conducted isokinetically and your stack emission test results will be meaningless.
The results of a particulate sampling test cannot be checked for accuracy because no independent technique
or test atmosphere exists to provide a standard or known particle concentration. Collaborative testing
conducted by the US EPA has determined that the interlaboratory standard deviation is ± 12.1%. Only
through careful calibration, maintenance and record keeping can you ensure that the data collected during the
stack test is representative of the site’s particle concentrations and mass emission rate.
2. Thermocouples (Stack, Probe, Filter Box, Impinger Exit, and Dry Gas Meter) and Digital
Temperature Indicator,
3. Pitot Tube,
Table 3-1 presents a summary of the components that require calibration and the equipment used to do so, as
well as acceptable calibration factor limits, recommended calibration frequency, and steps to take if your
system fails to meet calibration acceptance limits.
Dry Gas Meter 1. Wet Test Meter Yi = Y ± 0.05Y Semiannually Recalibrate, repair
Initial 5-point 2. Secondary or replace
Reference DGM
Post-test 3-point 1. Wet Test Meter Y = Y ± 0.05Yavg After each field Recalibrate at 5-
2. Reference DGM test points
3. Critical Orifices
Orifice Tube Measured during DGM ΔH@ = 46.7 ± 6.4 mm With DGM Repair or replace
calibration H2O (1.84 ± 0.25 in.
H2O)
Thermocouples and Certified Hg-in-glass Stack: ±1.5% °K After each field Recalibrate, repair
Digital Indicator thermometer in ice DGM: ±3°C test or replace
72
Probe: ±3°C
slush and boiling water
Filter: ±3°C
Exit: ±1°C
Pitot Tube 1. Standard pitot tube If part of Probe Quarterly, or after Recalibrate, repair
in wind tunnel and Assembly, calibrate each field test or replace
calculate Cp (pitot tube with assembly.
coefficient)
σ (avg. deviation) ≤
0.001 for side A & B
2. Measure with angle α1 ± 10° Quarterly, or after Recalibrate, repair
indicator to α2 ± 10° each field test or replace
demonstrate meeting β1 ± 5°
geometric β2 ± 5°
specifications and Z = ≤ 0.125”
assign Cp = 0.84 W = ≤ 0.031”
PA - PB ≤ 0.063”
0.188” ≤ DT ≤ 0.375”
Sampling Nozzles Micrometer with at Average of 3 inner Before each field Recalibrate,
least 0.025-mm (0.001- diameter use reshape, or
in.) scale measurements; ΔD ± resharpen when
0.1-mm (0.004-in.) dented or
corroded
Probe and Filter Box Gas thermocouple Capable of Initially Repair or replace,
Heater System maintaining 120°C ± and verify
14°C at 20-lpm flow calibration
rate
Conduct an initial (or full) calibration at five (5) selected flow rate (ΔH) settings.
Repeat the procedure once every 6 months, or if the results of a post-test 3-point calibration show
that the dry gas meter calibration factor (Y) has changed by more than 5% from the pre-test
calibration value.
73
When performing quarterly or post-test calibrations, use the abbreviated calibration procedure described in
Section 5.3.2 of Method 5, which includes three (3) calibration runs at a single intermediate flow rate (ΔH)
setting.
The dry gas meter (DGM) calibration factor (Y) is the ratio of the wet test meter’s measured volume to the
dry gas meter’s measured volume. The DGM calibration factor is part of calibrating the Source Sampler
Console.
The Orifice Tube calibration factor (ΔH@) represents the pressure drop across the orifice for a sampling flow
rate of 21.2 lpm (0.75 cfm), which is the standard sampling rate for solving the isokinetic equation and
setting up the nomograph (sets of equations) for testing.
Find the true Orifice Tube calibration factor with the following equation, where Km is the orifice calibration
factor:
ΔH@ = 0.9244/Km2
Described below are both the initial and intermediate calibration procedures. Prior to conducting a
calibration run, perform a leak-check on the portion of the sampling train from the pump to the Orifice Tube
in the Source Sampler Console, as described below.
74
Figure 3-1: Source Sampler Console Plumbing Diagram for Leak-Check
75
Initial or Semiannual Calibration of Dry Gas Meter and Orifice Tube:
Calibrate the Source Sampler Console (Dry Gas Meter and Orifice Tube) by connecting the Source Sampler
Console to a wet test meter or secondary reference dry test meter, according to the set-ups shown in Figures
3-2 and 3-3.
Conduct a series of five (5) calibration runs at differing flow rates (ΔH settings) that bracket the range of
expected sampling rates during particulate sampling tests.
If using a wet test meter as the calibration standard, it should have a meter correction factor of 1.000.
Alternatively, you may use a properly calibrated secondary reference dry gas meter to calibrate the Source
Sampler Console’s dry gas meter.
76
10. At the conclusion of the five calibration runs, calculate
the average Y value to obtain the dry gas meter Tips from a Stack Tester
calibration factor, as well as the average orifice pressure
(ΔH@). The tolerance for individual Y values is ± 0.02
from the average Y. The tolerance for individual ΔH@
values is ± 6.4 mm (0.25 in) H2O from the average ΔH@.
11. If the Y and ΔH@ are acceptable, record the values on a
label on the front face of the Console Meter. If you don’t obtain a value
of ± 6.4-mm (0.25-in) H2O
from the average ΔH@,
adjust the orifice opening
or replace the Orifice
Tube.
Figure 3-2: Set-up for Calibrating the Source Sampler Console Against Wet Test Meter
77
Figure 3-3: Illustration of Console Meter Calibration with Secondary Reference Dry Gas Meter
78
Post-Test Calibration of the Source Sampler Console:
Conduct a post-test, or 3-point, calibration of the Source Sampler Console after each test series or trip to and
from the field to ensure that the dry gas meter correction factor (Y) has not changed by more than 5%.
Conduct the post-test calibration check at the average ΔH and highest system vacuum observed during the
test series. The steps are as follows:
1. Before starting the calibration, fill out a meter calibration data sheet (it can be a computer
spreadsheet) as shown in Appendix C. Record the following:
a. Barometric pressure at the start of calibration,
b. Identification numbers for the Source Sampler Console and wet test meter (or secondary
DGM or critical orifice),
c. Date and time of calibration, and
d. Confirmation of acceptable leak checks on the Source Sampler Console.
2. Connect the outlet of the wet test meter to the inlet (SAMPLE) of the Source Sampler Console, as
depicted in Figure 3-2. Insert a valve between the wet test meter and the inlet of the Source Sampler
Console to adjust the vacuum to the desired level. If using a Critical Orifice, simply insert the male
quick-connect end of the critical orifice into the inlet (SAMPLE) of the Source Sampler Console.
3. Turn on the Vacuum Pump and adjust the Coarse and Fine Control Valves on the Source Sampler
Console until you obtain an orifice differential pressure (ΔH) reading equivalent to the average ΔH
observed during the test series. Set the calibration system vacuum to the highest vacuum observed
during the test series. Allow both meters to run like this for at least 15 minutes to let the meter
stabilize and the wet test meter to wet the interior surfaces.
4. If using a Critical Orifice, select an orifice with ΔH properties similar to the average ΔH observed
during the test series. Note that the vacuum will be independent of the vacuum observed during the
test series, due to the physics of the critical orifice requiring the vacuum to be greater than the
theoretical critical vacuum. Theoretical critical vacuum can be estimated at one-half (1/2) of
barometric pressure. It is recommended to sample at maximum vacuum with the Coarse Valve fully
opened and the Fine Valve fully closed.
5. Turn off the Vacuum Pump.
6. Record initial settings of ΔH, dry gas meter volume reading, wet test meter volume reading, dry gas
meter temperature, and wet test meter temperature.
7. Start the Vacuum Pump and quickly adjust the ΔH to the desired setting. Start the Elapsed Timer on
the Source Sampler Console at the same time that the pump is started.
8. Let the Vacuum Pump run until the dry gas meter indicates a volume of approximately 140 liters (5
cubic feet). Allow the calibration run to continue until the next minute elapses, and then stop the
Vacuum Pump and Elapsed Timer.
9. Record the final dry gas meter volume reading, wet test meter volume reading, dry gas meter
temperature, and wet test meter temperature. Calculate the dry gas meter and wet test meter volumes
by subtracting initial readings from final readings. Calculate the average dry gas meter and wet test
meter temperatures.
10. Calculate the meter correction factor Y. Repeat the calibration two (2) more times at the same ΔH
and system vacuum settings and calculate the average Y for the three runs.
79
11. Calculate the percent change in the meter correction factor Y.
12. If the dry gas meter Y values obtained before and after the test series differ by more than 5%, either
void the test series or perform calculations for the test series using the lower Y value (which would
give a lower sample volume, and therefore higher concentration values).
2. Calibration of Thermocouples
Apex Instruments suggests the following procedures for calibrating thermocouples and temperature display
readouts:
● Check thermocouples for calibration at three temperatures. For example, ice-point and boiling point
of water and ambient temperature. Thermocouples that are used at higher temperatures than boiling
water, such as the stack gas thermocouples, can be checked for calibration using a hot oil bath.
● Another more modern technique is to use a Thermocouple Simulator Source (M5C-22), as shown in
Figure 3-5. The M5C-22 can calibrate without external compensation or ice baths, with a temperature
range from 0° to 2,100°F in divisions of 100°F, resulting in 22 precise test points.
Figure 3-5:
Thermocouple Calibration Procedure: Thermocouple Simulator for
Temperature Display Calibration Check
1. Prepare an ice-water bath in an insulated container (such
as the Cold Box).
a. Insert the thermocouple and a mercury reference thermometer into the bath.
b. Allow the readings of both to stabilize and record the temperatures on a thermocouple
calibration data sheet, as shown in Appendix C.
c. Remove the thermocouple and allow it to stabilize at room temperature.
d. Reinsert the thermocouple into the bath and record another reading.
e. Repeat Steps c and d.
f. Calculate the average of the thermocouple readings and the average of the reference
thermometer readings. The averages should differ by less than 1.5% of the absolute
temperature (°K) for the stack thermocouple.
2. Place a beaker of distilled water on a hot plate, and then add a few boiling chips and heat to boiling.
a. Insert the thermocouple and a mercury reference thermometer into the beaker and repeat Step
1, b through f, above.
80
3. Set the thermocouple you are checking and a mercury reference thermometer side-by-side at ambient
temperature and allow the readings to stabilize.
a. Repeat Step 1, a through f, above.
4.
4.
Did you know? Apex Instruments provides both geometric and wind tunnel 4.
calibrations of type-S pitot tube assemblies for a reasonable fee. Contact us at (800) 4.
882-3214 to learn more. 4.
4.
4.
4.
4.
4.
Place a container of oil on a hot plate and heat to a temperature below the boiling point. DO NOT BOIL.
a. Insert the thermocouple and a mercury reference thermometer into the container and repeat
Step 1, b through f, above.
The temperature display requires additional calibration procedures. To check the linearity of the temperature
display, use a thermocouple simulator (such as Apex Model M5C-22 pictured in Figure 3-5). Connect the
simulator to the temperature display and record the display reading at each temperature setting on a
calibration data sheet.
Calibration
Of
S-type Pitot Tube
Option 1: Since few stack testers have access to a wind tunnel facility, the US EPA allows you to assign a
co-efficient of Cp = 0.84, if the pitot tube meets geometric specifications. Keep in mind that using the
agency’s standard co-efficient will likely bias your velocity and flow rate measurements high.
81
Option 2: When using a wind-tunnel calibration procedure, the pitot tube coefficient (C p) will typically
range from 0.77 to 0.82. This means that your subsequent measurements of stack gas velocity will likely be
between 2% and 8% lower due to increased accuracy associated with wind-tunnel calibration. Consequently,
your measurements of the stack gas volumetric flow rate and emission rate also will be lower. The
procedures for conducting a wind tunnel calibration are described in detail in US EPA Method 2.
82
4. Calibration of Sampling Nozzles
Inspect and calibrate probe nozzles in the field immediately before each use to verify that they were not
damaged in transport or shipment. We recommend the following procedure, illustrated in Figure 3-7:
1. Before starting the calibration check, fill out a probe nozzle calibration data sheet as shown in
Appendix C (or use a computer spreadsheet).
2. Using venier or dial calipers with at least 0.025 mm (0.001 in.) tolerance, measure the inside diameter
of the nozzle by taking three readings approximately 45-60° apart from one another, and record them.
3. Calculate the average of the three readings.
4. If readings do not fall within 0.1 mm (0.004 in.) of one another, the nozzle must be reshaped,
resharpened and recalibrated.
5. With a permanent marking tool, identify each nozzle with a unique number.
83
5. Initial Calibration of Probe Heater
Apex Instruments calibrates the probe heater assembly before shipping according to procedures outlined in
US EPA APTD-0576. Probes are constructed according to specifications given in US EPA APTD-0581,
which is the original 1971 document entitled “Construction Details of Isokinetic Source-Sampling
Equipment,” by Robert M. Martin. (Available from National Technical Information Service (NTIS) as
document PB-203 060.)
The procedure outlined in APTD-0576 involves passing heated gas at several known temperatures through a
probe assembly, and monitoring and verifying that the probe assembly is capable of maintaining 120°C ±
14°C, as shown in Figure 3-8.
84
6. Calibration of Pressure Sensors
The gauge-oil inclined manometer and a mercury-in-glass manometer are primary measuring devices, and
thus do not require calibration. When using a differential pressure gauge (magnehelic gauge), U-tube
manometer, or electronic manometer, they must be calibrated against a primary measuring device.
To check the calibration of differential pressure sensors other than inclined manometers, use the following
procedure:
1. Connect the differential pressure sensor to a gauge-oil inclined manometer as shown in Figure 3-9.
2. Vent the vacuum side to atmosphere, and place a pressure on each system.
3. Compare Δp (velocity head of stack gas) readings of both devices at three or more levels that span
the range, and record on a calibration data sheet.
4. Repeat Steps 1 through 3 for the vacuum side. Vent the pressure side and for the vacuum side, place a
vacuum on the system.
5. The readings at the three levels must correspond within ± 5% of the reference sensor.
To pressure source or 1 2 4 .6
vented to atmosphere OR
Magnehelic
Valve
To vacuum system or
vented to atmosphere
Manometer
85
Maintenance Procedures
For maintenance procedures, check your appropriate user’s manual for the equipment that you are using for
your stack test.
If you are using Apex Instruments equipment, call 1 (800) 882-3214 to speak with a member of our
Technical Services Group (TSG) to aid you in your maintenance needs. Additionally, you may use our
website www.apexinst.com to find a user’s manual for your specific equipment that needs maintenance.
86
Appendix A
Equipment Lists for Isokinetic Sampling
87
Recommended Equipment for Isokinetic Sampling
1 XE-0523 External Pump Assembly, Double Diaphragm Pump, SS Quick Connects, 5 FT Power Cord &
Hoses, 240V (120V option), Black UHMW Polyethylene Case
1 SB-2M Miniature Heated Filter Oven with 2 Access Doors & Probe Clamp, 240V (120V option)
1 SBR-10 Riser for SB-2M, Impinger Box Adapter, 10 IN Riser with Insulated Reservoir
1 SB-3 Impinger Box / Insulated Coolant Reservoir, Holds 4 Impingers
1 PS-3H 3 FT Method 5 Probe Sheath with Tube Heater, 120V (240V option)
1 PS-6H 6 FT Method 5 Probe Sheath with Tube Heater, 120V (240V option)
1 PS-9H 9 FT Method 5 Probe Sheath with Tube Heater, 120V (240V option)
1 PL-3S 3 FT SS Probe Liner, 5/8” o.d. with #28 Grooved Ball
1 PL-6S 6 FT SS Probe Liner, 5/8” o.d. with #28 Grooved Ball
1 PL-9S 9 FT SS Probe Liner, 5/8” o.d. with #28 Grooved Ball
3 PLN-3G 3 FT Glass Probe Liner, 5/8” o.d. with Unground #28 Grooved Ball
3 PLN-6G 6 FT Glass Probe Liner, 5/8” o.d. with Unground #28 Grooved Ball
3 PLN-9G 9 FT Glass Probe Liner, 5/8” o.d. with Unground #28 Grooved Ball
1 US-30-10 30 FT Split Umbilical Cable, Method 5, with 1/4” Pitot QCs, Stainless Steel Quick Connects
1 US-60-10 60 FT Split Umbilical Cable, Method 5, with 1/4” Pitot QCs, Stainless Steel Quick Connects
1 USL-15-SST 15 FT Unheated Sample Line, PTFE with SS over braid, #28 Socket Elbow Fitting Both Ends,
Includes TCA-06F
1 SB-8 Modular Sample Frame with Probe Clamp & Impinger Box Slides
1 GA-107-12 Sample Line Adapter (for SB-2M), Includes Bracket and 1/2” ID Sample Line Clamp
1 GA-100 Umbilical Adapter (Gooseneck), #28 Socket with Thermocouple, Mounting Bracket, 1/2” Male QC
1 NS-SET Set of 7 Stainless Steel Nozzles - Sizes 4,6,8,10,12,14,&16, Includes Case and Four 5/8” tube nuts
1 NG-SET Set of 7 Glass Nozzles - Sizes 4,6,8,10,12,14&16, Includes Case & Three 5/8” Glass Filled PFA
Ferrules
1 GN-DGS Basic Method 5 Glassware Set with Unground Joints & 3” Filter Assembly
1 GF-3TPG 82mm PTFE Coated Glass Fiber Filters - Hydrophobic, Low Absorption of Acid Gases such as SOX,
or NOX, 50/box (MFS PG60)
1 GF-3 3” Glass Fiber Filters, 100/box (Whatman 934AH-82.6mm)
1 PBX-S Modular Probe Brush Extension Kit, Nylon Brushes with Stainless Steel Extensions
1 NB-SET Nozzle Brush Set (sizes 3, 5, & 8) in Carrying Tube
1 P1000-6 Monorail , 6 feet, with Chain, L-Bracket & Hardware
2 P1000-9 Monorail , 9 feet, with Chain, L-Bracket & Hardware
2 P2751 Swivel Frame Trolley with Eye-bolt, 12” Chain & Quick Link
1 TX-XC5 Transport Case, XC500 Series
1 M5CO-SET Calibration Orifice Set, 5 Calibrated Critical Orifices & Spreadsheet Diskette for Method 5
88
It is also recommended to have additional Probe Assemblies to best suit your testing needs. Apex
Instruments, Inc. recommends a 4 foot and 8 foot Probe Assembly in addition to the standard 6 foot which is
included in the above listed system.
89
Apex Instruments recommended method 29 additional accessories.
90
Recommended Spare Parts for Isokinetic Sampling
91
Recommended Spare Parts for Isokinetic Sampling
(continued)
92
Equipment Checklist
93
Appendix B
Calibration Data Sheets
94
95
96
97
98
99
100
101
102
Appendix C
Stack Testing Field Data Sheets
103
104
105
106
107
108
109
110
111
112
113
Appendix D
Calculation Worksheets
114
FEDERAL REFERENCE METHOD 1
Sample and Velocity Traverses for Stationary Sources
INPUT PARAMETERS
Circular Stack:
Rectangular Stack:
115
Location of Sampling Points:
CIRCULAR
12.
RECTANGULAR
116
Plant ______________________________________ Date _________________________
INPUT PARAMETERS
Average of square root of velocity head (mm. H2O)1/2 or (in. H2O)1/2 = (√∆p)avg = _________________
CALCULATIONS
T s( avg )
v s = K p x C p x ( √ Δp )avg x
√ Ps M s
vs = Kp x ( ) x ( ) x
√ (
( )
)x( ) = __________________________m/s or ft/s
Qa = 60 x v s x A s
Qa = 60 x ( ) x( ) = _____________________acmm or acfm
117
Qd = 60 x (1 − Bws ) x v s x A s
Q d = 60 x ( 1 − ( )) x ( ) x ( ) = ____________________scmm or scfm
T std P
Qsd = 60 x (1 − B ws ) x v s x A s x x s
Ts P std
T std ( )
Q sd = 60 x ( 1 − ( )) x ( ) x ( ) x x
( ) Pstd
= ____________________dscmm or dscfm
118
FEDERAL REFERENCE METHOD 3
Gas Analysis for the Determination of Dry Molecular Weight
INPUT PARAMETERS
CALCULATIONS
M d = 0. 44 (% CO 2 ) + 0. 32 (% O 2 ) + 0.28 ( % N 2 + % CO)
M d = 0. 44 ( ) + 0 . 32 ( ) + 0 . 28 ( + ) = __________g/g-mole or lb/lb-mole
M s = M d (1 − B ws ) + 18.0 ( Bws )
(% O2 ) −( 0 .5 % CO )
% EA = x 100
0 . 264 ( % N 2 ) − (% O2 − 0 .5 % CO)
( ) −( 0. 5 ( ) )
% EA = x 100
0 . 264 ( ) − ( ( )− 0 .5 ( )) = ______________________________%
119
FEDERAL REFERENCE METHOD 4
Determination of Moisture Content of Stack Gases
INPUT PARAMETERS
Average DGM orifice pressure differential (mm H2O) or (in. H2O) = ∆H = ________________
CALCULATIONS
T
P bar + (13ΔH. 6 )
( )
V m( std ) = V m Y std x
Pstd Tm
T std
( )+ (13( . 6) )
V m( std ) = ( ) ( )
( )
Pstd ( )
= _____________dscm or dscf
V wc ( std ) = K 2 ( V f −V i )
120
Determination of Moisture Content of Stack Gases
(continued)
V wsg ( std ) = K 3 ( W f −W i )
V wsg ( std ) = K 3 ( )
= ______________scm or scf
( )+ ( )
B ws =
( )+ ( )+( ) = ________________________
%H 2 O = 100 x B ws
%H 2 O = 100 x ( ) = _______________________%
121
FEDERAL REFERENCE METHOD 5
Nozzle Size Selection Worksheet
Note: The most commonly used equation for estimating isokinetic sampling nozzle diameter is the following (assumes that
moisture fraction at dry gas meter equals zero.):
Dn( est) =
√ K 1 Qm P m
√ Ts Ms
T m C p ( 1 − Bws ) P s Δpavg
INPUT DATA
Ms = ( ) (1 − ( ) ) + 18 .0 ( ) = Ms = ________________
°C + 273° = K or °F + 460 = °R
122
Nozzle Size Selection Worksheet
(continued)
Ts Ms
v=
√ P s Δp avg
)x(
v=
√ (
( )x(
)
) = v = ________________
Dn( est) =
√ K 1 Qm P m
√ Ts Ms
T m C p ( 1 − Bws ) P s Δpavg
K1 x ( ) x( )x(
D n( est ) =
√ ( )x( ) x (1 − (
)
)) = Dn(est) = ________________
123
Nozzle Size Selection Worksheet
(continued)
K-FACTOR CALCULATION
ΔH M T P
K= = K 6 D 4n ΔH @ C2p ( 1 − B ws ) 2 d m s
Δp M s T s Pm
ΔH ( )( )( )
K= = K6 ( )4 ( ) ( )2 ( 1 − ( ) )2
Δp ( )( )( ) = _________________
CHECK CALCULATIONS FOR SUFFICIENT SAMPLE VOLUME AND ISOKINETIC RATE 90-110%
100 ( 1 − ( )) ( ) ( ) ( )2
Qm( std ) =
1039 ( ) = ________________
V m( std ) = Q m( std ) x θ
= ________________
124
Nozzle Size Selection Worksheet
(continued)
K4 ( )( )( )
%I i =
( )( )( ) ( ) (1 − ( )) = ________________
π 2
D
An = Nozzle Area 4 n = __________in2 or mm2
V m( tot )
%I f =
[ (
100 T s (avg ) K 3 V lc +
T m (avg ) ) (P bar +
ΔH
13. 6 )]
60 θ v s( avg) Ps A n
%I f =
[
100 ( ) K 3 ( ) +
( )
( )(
( )
( )+
( )
13 .6 )]
60 ( ) ( ) ( ) ( ) = ________________
125