Professional Documents
Culture Documents
Biometry by Suryakant Jha and Wangchuk Doma PDF
Biometry by Suryakant Jha and Wangchuk Doma PDF
Diagnostics
Biometry
Suryakant Jha
MS
Venu Eye Institute & Research Centre, Sheikh Sarai, New Delhi
www. dosonline.org l 43
Diagnostics
A-scan ultrasound biometry are currently in use. The first technique does not require contact with the globe, so
is contact applanation biometry. This technique requires corneal compression artifacts are eliminated. Compared
placing an ultrasound probe on the central cornea. While with ultrasonography, the PCI provides more accurate,
this is a convenient way to determine the AL for most reproducible AL measurement. However, it is difficult to
normal eyes, errors in measurement almost invariably obtain a measurement in the presence of a dense cataract
result from the probe indenting the cornea and shallowing or other media opacities, which limits the use of this
the anterior chamber. Since the compression error is technique.
variable, it cannot be compensated for by a constant. IOL
Another advantage of PCI over ultrasound biometry is
power calculations using these measurements will lead to
that the AL measurement is performed through the visual
an overestimation of the IOL power. In shorter eyes, this
axis since the patient is asked to fixate into the laser
effect is amplified. The second type is immersion A-scan
spot. In highly myopic or staphylomatous eyes, this can
biometry, which requires placing a saline filled scleral shell
be particularly advantageous since it can sometimes be
(Prager) between the probe and the eye. Since the probe
difficult to measure the true AL through the visual axis with
does not exert direct pressure on the cornea, compression
an ultrasound probe. PCI is also superior to ultrasound in
of the anterior chamber is avoided. A mean shortening of
the measurement of pseudophakic and silicone oil-filled
0.25–0.33mm has been reported between applanation
eyes. Lenstar LS 900 (Haag-Streit) is relatively a newer
and immersion AL measurements, which can translate into
optical biometer approved by the FDA in October 2009
an error of IOL power by approximately 1 D. In general,
that is designed to provide more accurate measurements
immersion biometry has been shown to be more accurate
for IOL power calculations than was previously possible.
than contact applanation biometry in several studies. The
Unlike IOL master, the Lenstar LS 900 captures all
main limitation with the A-scan ultrasound is the poor
biometric measurements on the patient’s visual axis. It is
image resolution due to the use of a relatively long, low-
more precise, combines axial length and keratometry, and
resolution wavelength (10 MHz) to measure a relatively
enables different formulae to be used, This is extremely
short distance. In addition, variations in retinal thickness
helpful in extremely short eyes, very long eyes with
surrounding the fovea contribute to inconsistency in the
posterior staphylomata, eyes containing silicone oil, and
final measurement.
pseudophakic eyes. It should be noted that the axial length
The Zeiss IOLMaster was the first device to use partial obtained from PCI may be slightly longer than that obtained
coherence interferometry (PCI) in clinical ophthalmology. from ultrasound. This is due to PCI measuring the distance
The technique of partial coherence interferometry from the corneal surface to the RPE while ultrasound
measures the time required for infrared light to travel to measures to the anterior retinal surface. Therefore, many
the retina. Because light travels at too high a speed to IOL measurement machines require refined IOL constants
be measured directly, light interference methodology is unique to their mechanism.
used to determine the transit time and thus the AL. This
Tips for accurate measurement of axial length (using repeat if the difference between eyes is greater than 0.3
applanation): mm, or if consecutive measurements differ by more than
0.2 mm.
• Ensure the machine is calibrated and set for the correct
velocity setting (e.g. cataract, aphakia, pseudophakia) Measuring the power of the cornea
• The echoes from cornea, anterior lens, posterior lens, Corneal power measurement is required by all the formulas
and retina should be present and of good amplitude used to calculate the power of the IOL to be implanted
• Misalignment along the optic nerve is recognised by an in cataract surgery. Traditionally, corneal power has been
absent scleral spike determined using instruments able to measure the anterior
corneal curvature. These include manual keratometers,
• The gain should be set at the lowest level at which a auto- keratometers, and Placido ring–based corneal
good reading is obtained topographers. Once the anterior corneal curvature has
• Take care with axial alignment, especially with a hand- been converted to the overall corneal power by means of
held probe and a moving patient (as described above) the keratometric index, these methods allow accurate IOL
power calculations for most patients.
• Do not push too hard – corneal compression commonly
causes errors Manual keratometry is the most commonly used method
to measure corneal curvature. Keratometry should be done
• Average the 5–10 most consistent results giving the before AL measurement, and for both eyes. The patient is
lowest standard deviation (ideally < 0.06 mm) seated with the chin positioned in the chin rest and the
• Errors may arise from an insufficient or greasy corneal head resting on the head band. The keratometer is focused
meniscus due to ointment or methylcellulose used on the central portion of the cornea to be examined using
previously. the focusing knobs. The instrument is now rotated to align
the (-) signs in the same vertical meridian and the (+)
Take note of eyes that are very short (less than 22 mm) or signs in the same horizontal meridian. This will determine
very long (more than 25 mm). Axial length errors are more the axis of the pre-existing astigmatism. The left drum is
significant in short eyes and a posterior staphyloma may be rotated to superimpose the (+) signs and the horizontal
present in a long eye. Look out for the unexpected result, measurement is read out. The right drum is now rotated
for example an axial length of 27 mm in a patient with a to superimpose the (-) signs and the vertical measurement
+4.00 D refractive error. Always measure both eyes and reading is recorded.
www. dosonline.org l 45
Diagnostics
Table 1: Range of axial length and preferred formula Using the appropriate formula
Axial length (mm) Formula The Hoffer Q, Holladay I, and SRK/T formulae are all
<20 mm Holladay II commonly used, but the SRK I and II regression formulae
are now regarded as obsolete. More recent formulae, are
20-22 mm Hoffer Q
the Holladay II or Haigis. Where audit software is in use,
22-24.5 mm SRK/T/Hoffer Q/Holladay (average) the personalisation of calculation constants can increase
24.5-26 mm Holladay I accuracy.
>26 mm SRK/T Difficult eyes
Tips for accurate manual keratometry: • Intumescent cataracts will yield a 0.15 mm longer
axial length resulting in a +0.4 to +0.5 hyperopia
• Calibrate and check the accuracy of the keratometer postoperatively.
• Use a dedicated single instrument that is known to be • For aphakic eyes being planned for ACIOL or scleral
accurate fixated IOL, the appropriate A constant must be used
• Do not touch the cornea beforehand and ensure a and the mode of the machine changed to compensate
good tear film for the change in speed of the sound waves.
• Adjust the eyepiece to bring the central cross-hairs into • In eyes with silicone filled vitreous, the sensitivity of
focus the system should be increased to visualize the retinal
echospike and the components of the eye must be
• Make sure that the patient’s other eye is occluded and measured separately to reach an accurate result.
that the cornea is centred
• After corneal refractive surgery, the K reading may not
• Take an average of three readings, including the axes truly reflect the corneal power. Hence the refractive
• If high or low results are encountered (< 40.00 D or > history method or the contact lens method must be
48.00 D), it is advisable to have a second person check used to obtain corrected K value.
the measurements • In eyes with high myopia, a B-scan examination is
• Repeat if the difference in total keratometric power recommended to rule out a posterior staphyloma or
between the eyes exceeds 1.50 D other retinal pathologies. Identification of the posterior
pole may be difficult. The problem is compounded in
• In a scarred cornea, use the fellow eye or average the unilateral cases.
results.
Some common mistakes:
The IOL Master (Carl Zeiss Meditec) is an automated
keratometer that calculates K readings by analyzing 6 light • Wrong A-constant selected
reflections projected onto the anterior cornea within a 2.3 • Wrong formula used
mm radius. IOL master cannot image the very center of the
cornea, which is also problematic in eyes that have had • Wrong K-readings entered by hand (90 degrees out)
previous refractive surgery; in these cases, assumptions • Biometry print-out stuck in wrong patient’s notes
regarding the steepness of the central cornea might be
invalid. • Incorrectly labelled IOL
within the first year of life, because growth of the eye will have a few well-trained and experienced members of staff
result in a large myopic shift if IOL has been implanted with who can get consistent results, than to have many people
intraoperative K and axial length measurements. When with limited training and experience. In biometry, 90 per
surgery is being performed after the age of two years, a cent of eyes should be within 1 dioptre of their intended
myopic shift of 4-6 D is expected depending upon the age. refraction. Try to identify any issues that are leading to
Under correcting the IOL power compensates for this. consistent errors. Because we must remember that in
current scenario, cataract surgery is not only a rehabilitative
Biometry in children is difficult and may require general
procedure but a refractive as well.
anesthesia. The lack of fixation in children who have
keratometry under general anesthesia may lead to inaccurate References
keratometry readings. Partial coherence inferometry (PCI) 1. Connell B, Brian G, Bond MJ. A case-control study of biometry
has been used in cooperative children with reliability and in healthy and cataractous Eritrean eyes. Ophthalmic
accuracy. PCI requires patient cooperation and thus may Epidemiol.1997;4(3):151–5.
not be a viable option in infants and young children. 2. Nemeth J, Fekete O, Pesztenlehrer N. Optical and ultrasound
measurement of axial length and anterior chamber depth
Intraocular Lens Power Calculation in Pediatric for intraocular lens power calculation. J Cataract Refract
Patients Surg.2003;29(1):85–8.
Previous studies showed that the SRK II, SRK T, and Holladay 3. Kielhorn I, Rajan MS, Tesha PM, Subryan VR, Bell JA.
Clinical assessment of the Zeiss IOLMaster. J Cataract Refract
I formulas had no significant difference in lens power Surg.2003;29(3):518–22.
predictability in children. However, there was increased
variability in postoperative refractive outcome in patients 4. Shammasa HJ. Slack Inc; 2004. Intraocular Lens Power Calculations.
younger than 2 years of age with all formulas. The Hoffer Q 5. Gale RP, Saha N, Johnston RL. National Biometry Audit II.Eye.
formula had a tendency to overestimate the IOL power and 2006;20:25–28.
showed the greatest degree of variability. Newer formulas 6. The Royal College of Ophthalmologists. Cataract Surgery Guidelines
such as the Holladay II formula were designed to increase 2004. [(accessed October 2006)]
the accuracy of the IOL power calculation. The Holladay 7. Tromans C, Haigh PM, Biswas S, Lloyd IC. Accuracy of intraocular
II formula incorporates measured anterior chamber depth, lens power calculation in paediatric cataract surgery. Br J Ophthalmol
lens thickness, and corneal diameter. However as per a 2001;85(8):939 –941.
recent study at L.V. Prasad Eye Institute, SRK II is the best 8. Wilson ME, Bartholomew LR, Trivedi RH: Pediatric cataract surgery
formula for calculation of IOL power in children. and intraocular lens implantation: practice styles and preferences of
the 2001 ASCRS and AAPOS memberships. J Cataract Refract Surg
Conclusion 29:1811--20, 2003
Departments should aim for consistency in their biometry 9. Plager DA, Kipfer H et al Refractive change in pediatric pseudophakia:
6 year follow up. J Cataract Refractive Surg 2002:28:810-815
and audit their results. Mistakes are easy to make, but
difficult (and sometimes expensive) to rectify. It is better to
www. dosonline.org l 47