CALIBRATION AND CERTIFICATION OF STORAGE TANK FOR CUSTODY
TRANSFER
1. Procedure
This procedure is used for all equipment and instrument used at custody transfer points.
The equipment and instruments are:
Thermometers
Hydrometers
Steel Tapes
Storage Tanks (upright cylindrical types)
Storage Tanks (horizontal cylindrical types)
Barges
Trailer Mounted Tanks (Tank Cars)
2. General Guidelines
These equipments are to be calibrated by an approved or accredited corporate body.
This corporate body shall be prequalified by the Director of Petroleum Resources
(Government Agency).
This validity period for each equipment shall be stated.
Calibration shall be local except when the facilities are not available and approval must
be given by D.P.R.
Corporate Bodies and Agencies accredition for each equipment type must be set out and
stand the reference.
3. Accreditation
This shall vary from job to job and nominated corporate body must formally apply for
permission from the Department of Petroleum Resources (Government Agency).
4. Principle
6. Applicable Standards
This shall be the API Standard 2550 or ASTM-D1220-85 under the title
“Method for Measurement and Calibration of Upright Cylindrical Tanks”
7. Calibration Equipment
As stated in the reference standards i.e. The essential equipment shall be the same with those
for taking measurement of tank, shall be above ground and shall always be maintained in good
condition. Both the tape for height measurement and that for circumference measurement
should themselves have been calibrated in accordance with DPR guide or
ECHO SOUND INSPECTION SERVICES
Nargis Bhaban, Ground Floor, Bandar, Chattogram
Tel: +88 031 800214, Mobile: +88 1796591435
Email: echosoundctg@gmail.com, info@metcbd.com
www.metcbd.com
Current certificate of recent calibration at 68°F in accordance with the National Bureau
of Standard Test Fee Schedule 202.404.
Accessory equipment shall be provided.
Equipment for determining temperature and gravity:
Sample Can: For withdrawing samples in the process of tank temperature measurement,
shall be the thief type as specified in ASTM-D4057.
Thermometer: Shall be the total immersion type of range -5 to 215 deg F and shall
conform with specification E1 described in ASTM book of Standard Part 18.
Hydrometer: For gravity measurement shall be the thermohydrometertype calibrated in
deg API and with range 0 to 100 deg API conforming to the specification E100 for ASTM
Book of Standards.
Calibration Procedure:
Tank clean and degassed.
Tank be filled completely with water prior to the time of strapping. Tanks with the
capacity above 500 bbl could be strapped at any condition of fill. (No discharge no fill)
For critical measurement the following must be noted.
Expansion and contraction of the tank shell under liquid head and temperature.
Effect of tilt on the tank shell.
Alignment of the tank bottom.
8. Presentation of result
Field Data: All data collected during strapping as illustrated in the form.
(f) Calibration Procedure: The tank shall be clean,degassed and completely filled with
water. The procedure outlined in the reference standard is followed to obtain the necessary
parameters. A calibration chart for the tanks critical parameters to be obtained are:
9. Presentation of Result
Field Data
(1) A diagrammatic representation
(2) Other data are liquid type and its level in the tank during strapping, cylinder shell
thickness, the head and the horizontal inclination of the tank