Professional Documents
Culture Documents
Frictional heat can be generated during seating of dental implants into a drill-prepared osteotomy. This in vitro
study tested the heat generated by implant seating in dense bovine mandible ramus. A thermocouple was
placed approximately 0.5 mm from the rim of the osteotomy during seating of each dental implant. Four
diameters of implants were tested. The average temperature increases were 0.0758C for the 5.7-mm-diameter
implant, 0.978C for the 4.7-mm-diameter implant, 1.48C for the 3.7-mm-diameter implant, and 8.68C for the 2.5-
mm-diameter implant. The results showed that heat was indeed generated and a small temperature rise
occurred, apparently by the friction of the implant surface against the fresh-cut bone surface. Bone is a poor
thermal conductor. The titanium of the implant and the steel of the handpiece are much better heat
conductors. Titanium may be 70 times more heat conductive than bone. The larger diameter and displacement
implant may act as a heat sink to draw away any heat produced from the friction of seating the implant at the
bone-implant interface. The peak temperature duration was momentary, and not measured, but this was
approximately less than 1 second. Except for the 2.5-mm-diameter implants, the temperature rises and
durations were found to be below those previously deemed to be detrimental, so no clinically significant
osseous damage would be expected during dental implant fixture seating of standard and large-diameter-sized
implants. A 2.5-mm implant may generate detrimental heat during seating in nonvital bone, but this may be
clinically insignificant in vital bone. The surface area and thermal conductivity are important factors in removing
generated heat transfer at the bone-implant interface. The F value as determined by analysis of variance was
69.22, and the P value was less than .0001, demonstrating significant differences between the groups
considered as a whole.
Key Words: dental implant, friction, compression, bone, heat, thermal, screw rotation, osteotomy,
temperature
I
rrigation to address heat generated by the
dental implant osteotomy drill is a highly be heat generated due to the friction of the implant
controversial topic. Many clinicians do not use fixture against the fresh-cut bone. The magnitude
sterile water or saline irrigation to cool the and time duration for causation of any detrimental
implant drill with no detrimental outcomes.1 osteotomy heat are uncertain. Most heat-genera-
However, there may be another source of heat tion studies on this were done on osteotomies and
generation that may have been overlooked by in vitro or on lower animals in nondental implant
clinicians, that is, frictional heat generated during circumstances.3
implant fixture seating into the osteotomy. Bone has a thermal conductivity between low-
The implant osteotomy is generally 0.5 mm conducting subcutaneous fat and high-conducting
smaller in diameter than the definitive implant that muscle.4 This relatively low-to-moderate thermal
is placed in the osteotomy. This undersizing is done conductivity property may cause the bone to retain
generated heat and induce osseous necrosis.4 This
to ensure that the implant is mechanically secure in
also causes bone to not readily absorb heat.
the bone to prevent any micro movement that may
The purpose of this study was to determine
what heat magnitude may be generated during
Private practice, Willimantic, Conn
* Corresponding author, e-mail: dffdds@comcast.net dental implant fixture seating into the osteotomy.
DOI: 10.1563/AAID-JOI-D-13-00261 The null hypothesis was that there is no heat
FIGURES 1–3. FIGURE 1. A segment of bovine ramus was used to simulate osseous conditions. FIGURE 2. A thermocouple was
used to measure any heat generated in the bone segment. FIGURE 3. The drill, thermocouple lead, and bone segment were
immersed in a water bath held at approximately 368C during the implant fixture seating and temperature increase
measurements.
produced during dental implant placement into Initiate implant placement torqueing to 50 Ncm
osteotomies. at 12 rpm noting starting temperature and peak
temperature
Evaluate data
MATERIALS AND METHODS
All sites were drilled in the usual sequence as
Clean bovine mandibular rami were secured from a recommended by the manufacturer to the desired
local university pathobiology service (University of test diameter that matched the test implant
Connecticut, Storrs, Conn). The bone was sectioned diameter. The sites were first drilled with the
to provide osseous segments for study convenience proprietary pilot drill and then the succeeding
(Figure 1). The bone sections were stored in saline incremental drills for that particular final implant
in a refrigerator at approximately 58C prior to study. size. The sequence of the 2.5 implant was a 1.5-mm
During the study testing, each segment was diameter drill and finally the 2.0-mm drill. For the
maintained in a saline bath at a range of 348C– 3.7 implant, the drill sequence was 2.0 and then 2.8
388C. All sites were deemed to be dense type 1 mm. The drilling sequence for the 4.7 implant was
(Misch) bone 8- to 10-mm thick. All implants were 2.0, 2.8, and finally 3.4 mm. For the 5.7 implant, the
obtained from the author’s private practice. All sequence was 2.0, 2.8, 3.4, and finally 5.1 mm (Table
1; Figure 1).
implants used were retrieved, clinically failed
The final osteotomy drill was 2.0 mm for the 2.5-
implants or rejected due to intraoperative sizing
mm-diameter implant, 2.8 mm for the 3.7-mm, 3.4
issues. The author places 5–600 implants annually,
mm for the 4.7-mm, and 5.1 for the 5.7-mm implant.
and the implants were collected over several years.
A small hole was drilled with a No. 1/2 round latch
All implants were acid-etch blasted-surfaced 3.7, 4.7, burr approximately 0.5 mm away from each
5.7 3 10 mm (Legacy, Implant Direct Implants, osteotomy rim and 1 mm deep to accept a thermal
Ventura, Calif) or 2.5 3 10 mm (IntraLock, Boca probe for the electronic thermometer thermocou-
Raton, Fla). ple (Omega Engineering, Stamford, Conn; Figures 2
The workflow consisted of the following: and 3). Each bone segment, implant, and implant
drill was immersed in the water bath and allowed to
Secure clean bovine ramus and prepare thickness
accommodate to the water temperature to a
Select and drill the osteotomies for the selected temperature approximating 378C. During testing,
implants 2.5, 3.7, 4.7, 5.7 each bone temperature was recorded as a start
Drill the thermoprobe hole for placement 0.5 mm temperature when the bone temperature was
from osteotomy and 1 mm deep stable in the water bath. Each implant was then
Place bone with thermoprobe and implant in placed in each site and rotated in a handpiece
water bath at 368C and allow to equalize and torque driver until seated into each osteotomy at 12
stabilize rpm to 50 newton centimeters (Ncm). The insertion
TABLE 1 TABLE 3
Temperature increases for the 5.7-mm-diameter Temperature increases for the 3.7-mm-diameter
implants* implants*
Test 5.7-mm Test 3.7-mm
Diameter Start Temp Peak Temp Temp Rise Diameter Start Temp Peak Temp Temp Rise
1 35.1 35.3 0.2 1 37.4 40.0 2.6
2 35.0 35.1 0.1 2 36.7 37.9 1.2
3 34.8 34.9 0.1 3 36.8 37.4 0.6
4 34.8 34.9 0.1 4 36.7 37.1 0.4
5 34.9 35.0 0.1 5 36.7 37.1 0.4
6 34.9 34.9 0.0 6 36.5 36.9 0.4
7 34.8 34.8 0.0 7 36.0 40.8 4.8
8 34.7 34.8 0.1 8 35.8 36.4 0.6
9 34.7 34.8 0.1 9 34.6 36.4 1.8
10 34.7 34.7 0.0 10 34.6 35.8 1.2
11 34.6 34.6 0.0
12 34.5 34.6 0.1 *Average temperature rise ¼ 1.48C.
TABLE 2 TABLE 4
Temperature increases for the 4.7-mm-diameter Temperature increases for the 2.5-mm-diameter
implants* implants*
Test 4.7-mm Test 2.5-mm
Diameter Start Temp Peak Temp Temp Rise Diameter Start Temp Peak Temp Temp Rise
1 38.0 38.4 0.4 1 34.5 50.3 5.8
2 37.0 38.1 1.1 2 37.2 44.6 7.4
3 37.3 40.1 2.8 3 35.4 38.8 3.4
4 36.6 38.4 1.8 4 35.5 42.8 7.2
5 37.4 38.0 0.6 5 35.8 37.6 1.8
6 35.6 36.3 0.7 6 35.0 46.8 11.8
7 35.7 36.2 0.5 7 35.3 39.8 4.5
8 36.0 36.7 0.7 8 34.4 45.3 10.9
9 36.4 37.0 0.6 9 33.5 48.7 15.2
10 36.5 37.0 0.5 10 34.4 52.4 18.0
FIGURES 4 AND 5. FIGURE 4. The temperature increases on a log scale were found to have better differentials. FIGURE 5. The
implant contact with cortical bone generates the most heat during seating.
does not have this property. Cortical bone has a the 5.7-mm-diameter implant that interfaces with
conductivity of approximately 0.16–0.34 watts of the bone is 10 3 2 3 3.14 3 5.7/3 ¼ 17.898 mm2.
heat per meter per degrees Kelvin (W/m/K) and Thus, the surface area of the larger 5.7-mm implant
spongy bone 0.30 W/m/K,6 whereas titanium is is 170% larger than the smaller 3.7-mm-diameter
18.7–22.1 W/m/K5 or approximately 70 times the implant. The surface of the 2.5 implant is 2.28 times
thermal conductivity of bone. Thus, titanium would smaller than the 5.7 implant, and the 5.7 implant
be expected to absorb and conduct heat away from produced less than 1% of the in-bone heat increase.
a bone-titanium interface. The differences between the average temperature
Fourier’s heat conduction equation describes the rises of the implants confirm the predicted outcome
transfer of heat: q ¼ kA(dT/dx), where q is the rate of Fourier’s formula, where the area of the surface
of heat transfer, k is the constant of the conductivity interface is a major influence on frictional heat
of the material expressed as a negative, A is the transfer.
contact area, and dT/dx is the temperature gradient, The differences between the average tempera-
where dT is the temperature difference, that is, the ture rises of the 5.7, 4.7, and 3.7 implants was less
temperature applied versus the temperature of the than 1.58C, but there was a much larger 7.28C
entity, and dx is the distance into the entity.7 Heat difference between the 3.7-mm-diameter implants
conduction will progress from a body of higher and the 2.5-mm-diameter implants This large
temperature to a body of lower temperature. The difference may be explained by the area and
larger the difference in temperature, the faster the volume differences among the implants, which
heat will transfer. The higher the conductivity, the impart heat absorption properties to the physical
faster the heat transfer. The negative describes the sizes of the implants. The volume of a cylinder is pi
direction of the heat transfer to or away from the 3 radius squared 3 height. The large 5.7-mm-
material of higher thermal conductivity. Thus, heat diameter implants had a very small to no temper-
generated at the titanium-bone interface would be ature rise. This may be due to the heat sink effect of
transferred away from the bone and into the the larger surface area. In addition, the volume of
titanium metal and further into the placement titanium would draw any heat generated into the
attachment. In addition, the area of contact will much more thermoconductive metal and then into
enable faster transfer of heat.7 the placement attachment on the drill handpiece.
As the surface area of contact increases, the heat The contact of the mineral portion of the cortical
transfer increases. The heat of the force is then bone to the titanium oxide surface may be the
dissipated over the larger surface area, and the major cause of the friction and thus the generated
body with the greater capacity for heat absorption heat.8 The organic portion is unlikely to generate
then indeed does absorb the heat. This is the heat frictional heat. Heat is generated by friction of the
sink effect. The force and seating torque on all the implant against the bone and compression of the
implants was the same, 50 Ncm at 12 rpm. The 2.5- compressible bone during seating.9 The implant is
mm-diameter implants generated the most in bone generally slightly larger than the osteotomy accord-
insertion heat, have the smallest surface area, and ing to contemporary placement techniques and
would not transfer as much heat as a larger implant. compresses the bone away from the advancing
For the sake of simplicity in these calculations, implant.
the implants are assumed to be cylinders. The The heat generated was apparently not enough
formula for the surface area of a cylinder excluding to make a significant difference as measured by the
the coronal and apical surfaces is as follows: surface thermoprobe, since the titanium has 70 times
area ¼ h 3 2 3 pi 3 r, where h is the length of the greater potential for heat transmittal than bone.
implant, pi is the constant 3.14, and r is the radius of The thermoprobe was located approximately 0.5
the implant (Table 5). The implant ends do not mm from the implant-bone interface. This place-
contact bone, so these surfaces are not considered. ment distance has been used in past work.10 The
The surface area of the 3.7-mm-diameter, 10-mm- temperature at the bone-implant juncture was likely
length implant has an approximate surface area slightly higher than the temperature measured by
that interfaces with the bone of 10 3 2 3 3.14 3 3.7/ the thermoprobe. Since bone has a relatively low
2 ¼ 10.618 mm2. The approximate surface area of conductivity, the heat at the implant-bone interface
may be higher than the heat detected at 0.5 mm difference provides initial stability to enable os-
from that interface. seointegration.8 There is compression of the bone
It is this author’s experience that most clinical during seating of the tapered implant and friction of
implant seating placements are in bone that is less the rough implant surface rotating against the
dense than type 1 (Misch) bone and may generate freshly cut osseous surface. Blood and tissue
even less heat than this study found. Nevertheless, exudate may act as lubricants to lessen any heat
one recent study found temperature increases in generation. In addition, the implant-bone surfaces
nonvital porcine rib bone when placing 4.0-mm may engage significantly only at the peaks of the
implants.11 A total of 288 self-tapping implants implant threads, minimizing the actual contact area.
were placed in osteotomies to 30, 35, and 40 Ncm. The thread pitch may contribute to the speed of
Thermocouples monitored the bone at 5-, 1.5-, and seating and thermal increases but probably not
10-mm depths. The temperature increases were significantly. Osteotomies are generally cylindrical
found to be significant at the cortical level but not or tapered to match the shape of the selected
so at deeper trabecular levels of bone (5- to 10-mm implant. Thus, the implant as it seats has most
deep).11 Lower insertion torques used did not cause contact with the bone at the osteotomy opening,
a significant temperature increase. The thermop- that is, the cortical bone (Figure 4). This minimal
robes were placed 5-mm away from the osteoto- contact may maximize the heat sink effect of the
mies, which may not reveal the temperature metal implant/drill handpiece. A cylindrical implant/
increases that may occur in closer proximity. osteotomy may generate less heat on seating than
Because the greatest friction may occur at the a tapered implant because of probable less surface
cortical bone at the osteotomy opening, the deep contact and contact force.
placed thermoprobes did not demonstrate any The nonvital bovine ramus bone is much denser
significant temperature increases. than vital human bone and was used for its ready
The friction that occurs may be mostly at the availability. The unyielding nature of the bovine
implant cervical–cortical bone interface in all ramus prevented complete seating of most of the
densities of bone (Figure 5). Thus, the temperature implants, and no attempt to ensure complete
rise may be approximately the same in dense or seating was done to ensure consistency of the
less-dense bone sites if the cortical bone thickness trials. This is an in vitro study and makes no
is equal.11 Thus, most of the heat would be pretense as to its importance or clinical relevance.
generated at the ridge crest cortical bone, but the Obviously, seating temperatures will be different in
transfer of this heat to bone is dependent on the vital bone. Further study is needed to clarify this
ability of the implant to absorb that heat. The larger issue. The use of bone taps is clinically done, but
the displacement and thermal conductivity, the less taps were not used to keep the trials consistent.
heat will be transferred to the bone. Taps may indeed alter the generation of heat. A
Higher insertion torques can produce higher subsequent study may be done to ascertain this.
temperature rises in bone.11 Nevertheless, a larger This is beyond the scope of this in vitro study.
displacement of the implant being seated may There has been some discussion in the dental
transmit heat away from the contacting bone. Small implant literature about osseous thermal damage
2.5-mm-diameter implants may thus generate the during dental implant surgery.12–14 There is some
same heat per square millimeter of surface area debate as to at what temperature is vital bone
during insertion, but the smaller surface area and irreversibly damaged.1 The temperature at which
displacement of the implant may not remove as there is thermal damage to vital bone may yet be an
much generated heat that would then be transmit- undetermined parameter for any individual patient
ted to the bone. This may explain informal clinical for a given osseous site. A study by Eriksson et al14
reports of implant failures in the anterior of the found that bone necrosis can occur at 478C that is
mandible where implants were placed in very dense held for 1 minute. Nevertheless, there may be
bone. different temperatures at which osseous cell dam-
Many manufacturers make osteotomy drills that age does or does not occur depending on the bone
are approximately 0.5 mm smaller in diameter than density, blood supply, and the instrumentation that
the diameter of the implant placed in that site. This is used that causes the temperature rise. Except for
the 2.5-mm-diameter implants, the in vitro results tissue exudate may act as lubricants to minimize
herein had a temperature rise below the reported friction in vital bone.
destructive temperature found by Eriksson et al.14 It Time duration may be an issue in that irrepara-
must be stressed that the present tests were done ble bone necrosis may occur after 2 minutes of
in dense nonvital bovine bone with no vascularity. temperature elevation.14,15 The seating heat in-
Vital osseous heating was studied by Eriksson et creases seen in these tests were momentary and
al.14 In their work, vital rabbit bone was heated to would probably not incur serious osseous necrosis
538C for 1 minute. Bone necrosis occurred but was in vital bone. Nevertheless, if the generated heat is
reparable, and at some time during the 3–5 not conducted away from the implant bone
postoperative week, regeneration did take place. In interface by a large displacement or surface area,
addition, a temperature rise of 4.38C may reduce the the generated heat may induce necrosis.
quality of subsequently repaired bone.13,14 Blood The point velocity of the larger-diameter implant
perfusion may be the important parameter here. The is faster than the point velocity of the smaller-
authors reported that perfusion increased with the diameter implant since all insertions were done at
temperature increase but then stopped when 538C 12 rpm. The 5.7-mm-diameter implant rotated at a
was reached. Thus, this debate may be an issue of velocity of 68.4 mm per minute, while the 3.7-mm-
the importance of blood supply, bone physiology, diameter implant rotated at 14.4 mm per minute.
and density. The bone formation enzyme, alkaline According to Fourier’s formula, the higher point
phosphatase, denatures at 538C and thus would be velocity of the larger implant would generate more
inactivated at this temperature.12 Bone healing may heat than the smaller-diameter implant. Even
be impaired. After removal of the thermal source, though more heat may be generated, this appar-
repair and regeneration may occur. In yet another ently does not outweigh the ability of the larger
Eriksson study, heating vital bone to 448C to 478C for implant to remove heat from the interface. To
2 minutes produced bone necrosis.13,15 This necrosis minimize the heat of surface movement velocity, a
was irreparable. The temperature increases found in very slow placement, less than 12 rpm, of the small
the present in vitro study herein were below implant may be indicated, with the implant cooled
previously reported detrimental temperatures and before insertion and/or the osteotomy site cooled.
may not be detrimental to vital bone. The increase Leaving the placement instrument connected to the
with 2.5-mm implants approached or exceeded the small implant may dissipate any generated heat.
necrosis temperature but only momentarily, and Vital osseous sites may not absorb or accept
again, the test bone is nonvital and may generate generated heat as readily as nonvital bone, and this
higher temperature increases than may be seen in may make any temperature increase clinically
vital bone. insignificant.1 That is, vital bone with a blood
Another study compared the heat generated by supply and tissue fluid and blood immersing the
self-drilling screws versus self-tapping screws in osteotomy may not incur a significant frictional
rabbit calvaria.16 The screws were 0.2 or 0.4 mm seating temperature increase. In previously pub-
larger in diameter than the prepared osteotomies. lished work by this author, no temperature increase
The screws were 1.5 or 2.0 mm in diameter and 5- was seen when performing maxillary osteotomies in
mm long. The authors found that bone damage did two living patients without irrigation.1
occur during insertion of nonirrigated self-tapping The implant seating heat generated in this in
screws, but this may not be clinically significant. The vitro testing in dense type 1 (Misch) bovine bone
osteotomy itself will induce bone cell damage.17 was generally found to be below the purported
The dental implants used in the present study were detrimental temperature and duration magnitudes.
self-tapping screw-type implants of much larger A large implant surface area and displacement
diameter and length and were installed in a water produced less of a temperature rise as the implants
bath kept at or near body temperature, which may were seated. Titanium may be 70 times more heat
simulate clinical conditions more closely. The water conductive than bone. Larger-diameter titanium
bath and the much larger implant sizes used in the implants may act as a heat sink that draws away any
present study herein may dissipate any heat and generated heat from the bone-implant interface.
prevent a larger rise in temperature. Blood and Thus, a larger implant, with larger surface area and