Professional Documents
Culture Documents
KP IX 1817
General Information
ep : mean electrophoretic mobility of the two electrophoretic mobility of the solute and the electro-
osmotic flow in the capillary. Coated capillaries can be
analytes (ep 1 (epb epa ))
2 used to increase the separation capacity of those sub-
stances adsorbing on fused-silica surfaces. Using this
Apparatus mode of capillary electrophoresis, the analysis of both
An apparatus for capillary electrophoresis is com- small (Mr < 2000) and large molecules (2000 < Mr
posed of : <100000) can be accomplished. Due to the high effi-
(1) a high-voltage, controllable direct-current power ciency achieved in free solution capillary electrophore-
supply, sis, separation of molecules having only minute differ-
(2) two buffer reservoirs, held at the same level, con- ences in their charge to mass ratio can be effected. This
taining the prescribed anodic and cathodic solutions , separation mode also allows the separation of chiral
(3) two electrode assemblies (the cathode and the an- compounds by addition of chiral selectors to the sepa-
ode), immersed in the buffer reservoirs and connect- ration buffer.
ed to the power supply,
(4) a separation capillary (usually made of fused-silica) Optimization
which, when used with some specific types of detec- Optimization of the separation is a complex pro-
tors, has an optical viewing window aligned with the cess where several separation parameters can play a
detector. The ends of the capillary are placed in the major role. The main factors to be considered in the
buffer reservoirs. The capillary is filled with the so- development of separations are instrumental and elec-
lution prescribed in the monograph, trolytic solution parameters.
(5) a suitable injection system,
(6) a detector able to monitor the amount of substances Instrumental parameters
of interest passing through a segment of the separa- (1) Voltage A Joule heating plot is useful in optimiz-
tion capillary at a given time. It is usually based on ing the applied voltage and column temperature. Sepa-
absorption spectrophotometry (UV and visible) or ration time is inversely proportional to applied voltage.
fluorometry, but conductimetric, amperometric or However, an increase in the voltage used can cause
mass spectrometric detection can be useful for spe- excessive heat production, giving rise to temperature
cific applications. Indirect detection is an alternative and, as a result thereof, viscosity gradients in the buffer
method used to detect non UV-absorbing and non- inside the capillary. This effect causes band broadening
fluorescent compounds, and decreases resolution.
(7) a thermostatic system able to maintain a constant (2) Polarity Electrode polarity can be normal (anode
temperature inside the capillary is recommended to at the inlet and cathode at the outlet) and the electro-
obtain a good separation reproducibility, osmotic flow will move toward the cathode . If the
(8) a recorder and a suitable integrator or a computer. electrode polarity is reversed, the electro-osmotic flow
The definition of the injection process and its au- is away from the outlet and only charged analytes with
tomation are critical for precise quantitative analysis. electro-osmotic mobilities greater than the electro-
Modes of injection include gravity, pressure or vacuum osmotic flow will pass to the outlet.
injection and electrokinetic injection. The amount of (3) Temperature The main effect of temperature is
each sample component introduced electrokinetically observed on buffer viscosity and electrical conductivity,
depends on its electrophoretic mobility, leading to pos- and therefore on migration velocity. In some cases, an
sible discrimination using this injection mode. increase in capillary temperature can cause a confor-
Use the capillary, the buffer solutions, the precon- mational change in proteins, modifying their migration
ditioning method, the sample solution and the migra- time and the efficiency of the separation.
tion conditions prescribed in the monograph of the (4) Capillary The dimensions of the capillary (length
considered substance. The employed electrolytic solu- and internal diameter) contribute to analysis time, effi-
tion is filtered to remove particles and degassed to ciency of separations and load capacity. Increasing
avoid bubble formation that could interfere with the both effective length and total length can decrease the
detection system or interrupt the electrical contact in electric fields (working at constant voltage) which in-
the capillary during the separation run. A rigorous rins- creases migration time. For a given buffer and electric
ing procedure should be developed for each analytical field, heat dissipation, and hence sample band-
method to achieve reproducible migration times of the broadening, depend on the internal diameter of the ca-
solutes. pillary. The latter also affects the detection limit, de-
pending on the sample volume injected and the detec-
1. Capillary Zone Electrophoresis tion system employed. Since the adsorption of the
In capillary zone electrophoresis, analytes are sample components on the capillary wall limits effi-
separated in a capillary containing only buffer without ciency, methods to avoid these interactions should be
any anti-convective medium. With this technique, sep- considered in the development of a separation method.
aration takes place because the different components of In the specific case of proteins, several strategies have
the sample migrate as discrete bands with different been devised to avoid adsorption on the capillary wall.
velocities. The velocity of each band depends on the Some of these strategies (use of extreme pH and ad-
1820 General Information
In isoelectric focusing, the molecules migrate un- (1) Voltage Capillary isoelectric focusing utilizes
der the influence of the electric field, so long as they very high electric fields, 300 V/cm to 1000 V/cm in the
are charged, in a pH gradient generated by ampholytes focusing step.
having pI values in a wide range (poly- (2) Capillary The electro-osmotic flow must be re-
aminocarboxylic acids), dissolved in the separation duced or suppressed depending on the mobilization
buffer. strategy (see above). Coated capillaries tend to reduce
The three basic steps of isoelectric focusing are the electro-osmotic flow.
loading, focusing and mobilization. (3) Solutions The anode buffer reservoir is filled
(1) Loading step: Two methods may be employed: with a solution with a pH lower than the pI of the most
(i) loading in one step: the sample is mixed with acidic ampholyte and the cathode reservoir is filled
ampholytes and introduced into the capillary either with a solution with a pH higher than the pI of the most
by pressure or vacuum; basic ampholyte. Phosphoric acid for the anode and
(ii) sequential loading: a leading buffer, then the sodium hydroxide for the cathode are frequently used.
ampholytes, then the sample mixed with ampholytes, Addition of a polymer, such as methylcellulose, in
again ampholytes alone and finally the terminating the ampholyte solution tends to suppress convective
buffer are introduced into the capillary. The volume forces (if any) and electro-osmotic flow by increasing
of the sample must be small enough not to modify the viscosity. Commercial ampholytes are available
the pH gradient. covering many pH ranges and may be mixed if neces-
(2) Focusing step: When the voltage is applied, sary to obtain an expanded pH range. Broad pH ranges
ampholytes migrate toward the cathode or the anode, are used to estimate the isoelectric point whereas nar-
according to their net charge, thus creating a pH gradi- rower ranges are employed to improve accuracy. Cali-
ent from anode (lower pH) to cathode (higher pH). bration can be done by correlating migration time with
During this step the components to be separated mi- isoelectric point for a series of protein markers.
grate until they reach a pH corresponding to their isoe- During the focusing step precipitation of proteins
lectric point (pI) and the current drops to very low val- at their isoelectric point can be prevented, if necessary,
ues. using buffer additives such as glycerol, surfactants,
(3) Mobilization step: If mobilization is required for urea or zwitterionic buffers. However, depending on
detection, use one of the following methods. Three the concentration, urea denatures proteins.
methods are available:
(i) in the first method, mobilization is accomplished 4. Micellar Electrokinetic Chromatography (MEKC)
during the focusing step under the effect of the elec- In micellar electrokinetic chromatography, separa-
tro-osmotic flow; the electro-osmotic flow must be tion takes place in an electrolyte solution which con-
small enough to allow the focusing of the compo- tains a surfactant at a concentration above the critical
nents; micellar concentration (CMC). The solute molecules
(ii) in the second method, mobilization is accom- are distributed between the aqueous buffer and the
plished by applying positive pressure after the focus- pseudo-stationary phase composed of micelles, accord-
ing step; ing to the partition coefficient of the solute. The tech-
(iii) in the third method, mobilization is achieved after nique can therefore be considered as a hybrid of elec-
the focusing step by adding salts to the cathode res- trophoresis and chromatography. MECK is a technique
ervoir or the anode reservoir (depending on the di- that can be used for the separation of both neutral and
rection chosen for mobilization) in order to alter the charged solutes, maintaining the efficiency, speed and
pH in the capillary when the voltage is applied. As instrumental suitability of capillary electrophoresis.
the pH is changed, the proteins and ampholytes are One of the most widely used surfactants in MEKC is
mobilized in the direction of the reservoir which the anionic surfactant sodium dodecyl sulfate, although
contains the added salts and pass the detector. The other surfactants, for example cationic surfactants such
separation achieved, expressed as ΔpI, depends on as cetyltrimethylammonium salts, are also used.
the pH gradient (dpH/dx), the number of ampholytes The separation mechanism in MECK is as follows.
having different pI values, the molecular diffusion At neutral and alkaline pH, a strong electro-osmotic
coefficient (D), the intensity of the electric field (E) flow is generated and moves the separation buffer ions
and the variation of the electrophoretic mobility of in the direction of the cathode. If sodium dodecyl sul-
the analyte with the pH (-dµ/dpH): fate is employed as the surfactant, the electrophoretic
migration of the anionic micelle is in the opposite di-
D (dpH / dx) rection, towards the anode. As a result, the overall mi-
pI 3 celle migration velocity is slowed down compared to
E (d / dpH ) the bulk flow of the electrolytic solution.
In the case of neutral solutes, since the analyte can
Optimization partition between the micelle and the aqueous buffer,
The main parameters to be considered in the de- and has no electrophoretic mobility, the analyte migra-
velopment of separations are: tion velocity will depend only on the partition coeffi-
cient between the micelle and the aqueous buffer. In
1822 General Information
the electropherogram, the peaks corresponding to each cause excessive heat production that gives rise to tem-
uncharged solute are always between that of the perature gradients and viscosity gradients of the buffer
electroosmotic flow marker and that of the micelle (the in the cross-section of the capillary. This effect can be
time elapsed between these two peaks is called the sep- significant with high conductivity buffers such as those
aration window). For electrically charged solutes, the containing micelles. Poor heat dissipation causes band
migration velocity depends on both the partition coef- broadening and decreases resolution.
ficient of the solute between the micelle and the aque- (2) Temperature Variations in capillary temperature
ous buffer, and on the electrophoretic mobility of the affect the partition coefficient of the solute between the
solute in the absence of micelle. buffer and the micelles, the critical micellar concentra-
Since the mechanism in MEKC of neutral or tion and the viscosity of the buffer. These parameters
weakly ionized solutes is essentially chromatographic, contribute to the migration time of the solutes. The use
migration of the solute and resolution can be rational- of a good cooling system improves the reproducibility
ized in terms of the retention factor of the solute (k’), of the migration time for the solutes.
also referred to as mass distribution ratio (Dm), which (3) Capillary As in capillary zone electrophoresis,
is the ratio of the number of moles of solute in the mi- the dimensions of the capillary (length and internal
celle to those in the mobile phase. diameter) contribute to analysis time and efficiency of
separations. Increasing both effective length and total
t R t0 V length can decrease the electric fields (working at con-
k' K S stant voltage), increase migration time and improve the
tR VM
t0 1 separation efficiency. The internal diameter controls
tmc heat dissipation (for a given buffer and electric field)
and consequently the sample zone diffusion.
tR: migration time of the solute ,
t0: analysis time of an unretained solute (determined Electrolytic solution parameters
by injecting an electro-osmotic flow marker (1) Surfactant type and concentration The type of
which does not enter the micelle, for instance surfactant, in the same way as the stationary phase in
methanol), chromatography, affects the resolution since it modi-
tmc: micelle migration time (measured by injecting a fies separation selectivity. Also, the log k’ of a neutral
micelle marker, such as Sudan III, which mi- compound increases linearly with the concentration of
grates while continuously associated in the mi- surfactant in the mobile phase.
celle), Since resolution in MEKC reaches a maximum when k’
K: partition coefficient of the solute, approaches the value of t m t0 , modifying the concen-
VS: volume of the micellar phase, tration of surfactant in the mobile phase changes the
VM: volume of the mobile phase. resolution obtained.
(2) Buffer pH Although pH does not modify the
Likewise, the resolution between two closely migrating partition coefficient of non-ionized solutes, it can mod-
solutes (RS) is given by the following equation: ify the electroosmotic flow in uncoated capillaries. A
decrease in the buffer pH decreases the electro-osmotic
t
1 ( 0 ) flow and therefore increases the resolution of the neu-
N 1 k 'b tmc tral solutes in MEKC, resulting in a longer analysis
RS
4 k 'b 1 1 ( t0 ) k ' time.
a (3) Organic solvents To improve MEKC separation
tmc
of hydrophobic compounds, organic modifiers, such as
methanol, propanol, acetonitrile, etc., can be added to
N: number of theoretical plates for one of the so-
the electrolytic solution. The addition of these modifi-
lutes,
ers usually decreases migration time and the selectivity
: selectivity,
of the separation. Since the addition of organic modifi-
k’a, k’b: retention factors for both solutes, respec-
ers affects the critical micellar concentration, a given
tively (k’b > k’a)
surfactant concentration can be used only within a cer-
tain percentage of organic modifier before the
Similar, but not identical, equations give k’ and RS
micellization is inhibited or adversely affected, result-
values for electrically charged solutes.
ing in the absence of micelles and, therefore, in the
absence of partition. The dissociation of micelles in the
Optimization
presence of a high content of organic solvent does not
The main parameters to be considered in the de-
always mean that the separation will no longer be pos-
velopment of separations by MEKC are instrumental
sible; in some cases the hydrophobic interaction be-
and electrolytic solution parameters.
tween the ionic surfactant monomer and the neutral
solutes forms solvophilic complexes that can be sepa-
Instrumental parameters
rated electrophoretically.
(1) Voltage Separation time is inversely proportional
to applied voltage. However, an increase in voltage can
KP X 1823
(4) Additives for chiral separations For the separa- Apparent Number of Theoretical Plates
tion of enantiomers using MEKC, a chiral selector is
included in the micellar system, either covalently 2
t
bound to the surfactant or added to the micellar separa- N 5.54 R
tion electrolyte. Micelles that have a moiety with chiral wh
discrimination properties include salts of N- tR: migration time or distance along the baseline
dodecanoyl-L -amino acids, bile salts, etc. Chiral reso- from the point of injection to the perpendicular
lution can also be achieved using chiral discriminators, dropped from the maximum of the peak corre-
such as cyclodextrins, added to the electrolytic solu- sponding to the component,
tions which contain micellized achiral surfactants. wh: width of the peak at half -height.
(5) Other additives Several strategies can be carried
out to modify selectivity, by adding chemicals to the Resolution
buffer. The addition of several types of cyclodextrins The resolution (RS) between peaks of similar
to the buffer can also be used to reduce the interaction height of two components may be calculated using the
of hydrophobic solutes with the micelle, thus increas- following expression:
ing the selectivity for this type of compound.
The addition of substances, capable of modifying
1.18 (tR 2 t R1 )
solute-micelle interactions by adsorption on the latter, RS t R 2 t R1
is used to improve the selectivity of the separations in wh1 wh 2
MEKC. These additives may be a second surfactant tR1, tR2: migration times or distances along the base-
(ionic or non-ionic) which gives rise to mixed micelles line from the point of injection to the perpendicu-
or metallic cations which dissolve in the micelle and lars dropped from the maxima of two adjacent
form co-ordination complexes with the solutes. peaks,
wh1, wh2: peak widths at half -height.
Quantification When appropriate, the resolution may be calculat-
Peak areas must be divided by the corresponding ed by measuring the height of the valley (Hv) between
migration time to give the corrected area in order to: two partly resolved peaks in a standard preparation and
(1) compensate for the shift in migration time from run the height of the smaller peak (Hp) and calculating the
to run, thus reducing the variation of the response, peak-to-valley ratio:
(2) compensate for the different responses of sample
constituents with different migration times. Hp
Where an internal standard is used, verify that no p/v
Hv
peak of the substance to be examined is masked by that
of the internal standard.
Symmetry Factor
Calculations The symmetry factor (AS) of a peak may be calcu-
From the values obtained, calculate the content of lated using the following expression:
the component or components being examined. When
prescribed, the percentage content of one or more w0.05
AS
components of the sample to be examined is calculated 2d
by determining the corrected area(s) of the peak(s) as a w0.05: width of the peak at 1/20 of the peak height,
percentage of the total of the corrected areas of all d: distance between the perpendicular dropped from
peaks, excluding those due to solvents or any added the peak maximum and the leading edge of the
reagents. The use of an automatic integration system peak at 1/20 of the peak height .
(integrator or data acquisition and processing system)
is recommended. Tests for area repeatability (standard deviation of
areas or of the area/migration-time ratio) and for mi-
System Suitability gration time repeatability (standard deviation of migra-
In order to check the behavior of the capillary tion time) are introduced as suitability parameters. Mi-
electrophoresis system, system suitability parameters gration time repeatability provides a test for the suita-
are used. The choice of these parameters depends on bility of the capillary washing procedures. An alterna-
the mode of capillary electrophoresis used. They are: tive practice to avoid the lack of repeatability of the
retention factor (k’, only for micellar electrokinetic migration time is to use migration time relative to an
chromatography), apparent number of theoretical plates internal standard.
(N), symmetry factor (AS) and resolution (RS). In previ- A test for the verification of the signal-to-noise ra-
ous sections, the theoretical expressions for N and RS tio for a standard preparation (or the determination of
have been described, but more practical equations that the limit of quantification) may also be useful for the
allow these parameters to be calculated from the determination of related substances.
electropherograms are given below.
Signal-to-noise Ratio
1824 General Information
The detection limit and quantification limit corre- read the unsettled apparent volume to the nearest grad-
spond to signal-to-noise ratios of 3 and 10 respectively. uated unit and calculate the bulk density ρB by the fol-
The signal-to-noise ratio (S/N) is calculated using the lowing formula:
following expression:
M
B
2 H V0
S/N
h
H: height of the peak corresponding to the compo- B : Bulk density by constant mass method
nent concerned, in the electropherogram obtained
with the prescribed reference solution, measured (g/mL)
from the maximum of the peak to the extrapolat- M : Mass of powder sample (g)
ed baseline of the signal observed over a distance V0 : Apparent volume of powder sample (mL)
equal to twenty times the width at half height,
h: range of the background in an electropherogram Record the average of 3 determinations and regard
obtained after injection of a blank, observed over the average value as the bulk density according to the
a distance equal to twenty times the width at the constant mass method. If a 30 g sample is too large to
half height of the peak in the electropherogram determine, adjust the mass of sample so as to provide
obtained with the prescribed reference solution an apparent volume of 60-100 mL.
and, if possible, situated equally around the place
where this peak would be found . Method 2 Constant volume method
Unless otherwise specified, pass a quantity of
sample sufficient to complete the test through a sieve
2. Determination of Bulk and No. 16 (1000 µm) to break up agglomerates that may
have formed during storage. Allow an excess of sample
Tapped Densities powder to pour into the measuring vessel having the
volume of V and mass of M0. Then, carefully scrape
Determination of Bulk and Tapped Densities is a excess powder from the top of the vessel using the
method to determine the bulk densities of powdered edge of a slide glass or other tool by smoothly moving
drugs under loose and tapped packing conditions, re- across it. Remove any material from the sides of the
spectively. Loose packing is defined as the state ob- vessel, and determine the total mass Mt and calculate
tained by pouring a powder sample into a vessel with- the bulk density ρB by the following formula:
out any consolidation, and tapped packing is defined as
the state obtained when the vessel containing the pow- (M t M 0 )
der sample is to be repeatedly dropped a specified dis- B
V
tance at a constant drop rate until the apparent volume
of sample in the vessel becomes almost constant. The
ρB: Bulk density by constant volume method (g/mL)
bulk density is expressed in mass per unit apparent
MT: Total mass of powder and measuring vessel (g)
volume of powder (g/mL). Because the bulk density is
M0: Mass of measuring vessel (g)
one of the measures of packing properties, compressi-
V: Volume of measuring vessel (mL)
bility and flow properties, and is dependent on the ‘his-
tory’ of the powder, it is essential to document the bulk
Record the average of 3 determinations and regard
density to specify how the determination was made.
the average value as the bulk density according to the
constant volume method.
Bulk density
The bulk density is an apparent density obtained
by pouring a powder sample into a vessel without any
consolidation. The determination of bulk density is
achieved by measuring the apparent volume of a pow-
der sample having a known mass in a graduated cylin-
der (Method 1) or by measuring the mass of powder in
a vessel having a known volume (Method 2).
disinfection and sterilization treatment must be chosen Microorganisms are killed by putting the object in boil-
appropriately for each application. Generally, the fol- ing water. This method is used for a product which
lowing methods are to be used singly or in combination may be denatured by the moist heat method. As a rule,
after appropriate optimization of operation procedures the product is put in boiling water for 15 minutes or
and conditions, in accordance with the kind and the more.
degree of the contaminating microorganisms and the (iii) Intermittent method
nature of the items to which the methods are applied. Microorganisms are killed by heating for 30 to 60
The validation of sterilization in accordance with minutes repeatedly, three to five times, once a day in
Terminal Sterilization and Sterilization Indicators is water at 80 to 100 C or in steam. This method is used
required when the methods are applied to the manufac- for a product which may be denatured by the moist
turing process of drug products. heat method. There is another method called the low
temperature intermittent method with repeated heating
1. Disinfection methods at 60 to 80 C. During the intermission periods be-
These methods are used to reduce the number of living tween heating or warming, a suitable temperature for
microorganisms, but do not always remove or kill all the growth of microorganisms of 20 C or higher, must
microorganisms present. Generally, disinfection is be maintained.
classified into chemical disinfection with the use of (iv) Ultraviolet method
chemical drugs (disinfectants) and physical disinfec- As a rule, microorganisms are killed by irradiation with
tion with the use of moist heat, ultraviolet light, and ultraviolet rays as a wavelength of around 254 nm.
other agents. This method is used for products which are resistant to
ultraviolet rays, such as smooth-surfaced articles, facil-
1.1. Chemical disinfection ities, and equipment, or water and air. This method
Microorganisms are killed with chemical drugs. The does not suffer from the occurrence of resistance,
killing effect and mechanisms of a chemical drug differ which is observed in chemical disinfection, and shows
depending on the type, applied concentration, action a killing effect on bacteria, fungi, and viruses. It must
temperature, and action time of the chemical drug used, be taken into consideration that direct ultraviolet irra-
the degree of contamination on the object to be disin- diation of the human body can injure the eyes and skin.
fected, and the series and state (e.g., vegetative bacteria
or spore bacteria) of microorganisms. 2. Sterilization methods
Therefore, in applying the method, full consideration is 2.1. Heating methods
required of the sterility and permissible storage period In these methods, the heating time before the tempera-
of the prepared chemical drug, the possibility of re- ture or pressure reaches the prescribed value differs
sistance of microorganisms at the site of application, according to the properties of the product, the size of
and the effect of residual chemical drug on the product. the container, and the conditions. The duration of heat-
In selecting a suitable chemical drug, the following ing in conducting these methods is counted from the
items should be considered in relation to the intended time when all the parts of the product have reached the
use. prescribed temperature.
1) The antimicrobial spectrum (i) Moist heat method
2) Action time for killing microorganisms Microorganisms are killed in saturated steam at a suit-
3) Action durability able temperature and pressure. This method is general-
4) Effect of the presence of proteins ly used for heat-stable substances, such as glass, porce-
5) Influence on the human body lain, metal, rubber, plastics, paper, and fiber, as well as
6) Solubility in water heat-stable liquids, such as water, culture media, rea-
7) Influence on the object to be disinfected gents, test solutions, liquid samples, etc. As a rule, one
8) Odor of the following conditions is used.
9) Convenience of use
10) Easy disposability 115 ~ 118 C for 30 minutes
11) Influence on the environment at disposal 121 ~ 124 C for 15 minutes
12) Frequency of occurrence of resistance 126 ~ 129 C for 10 minutes
1.2. Physical disinfection (ii) Dry-heat method
Microorganisms are killed without a chemical drug. Microorganisms are killed in dry-heated air. This
(i) Steam flow method method is generally used for heat-stable substances,
Microorganisms are killed by direct application of such as glass, porcelain, and metal, as well as heat-
steam. This method is used for a product which may be stable products, such as mineral oils, fats and oils,
denatured by the moist heat method. As a rule, the powder samples, etc. This method is generally con-
product is kept in flowing steam at 100 C for 30 to ducted in the way of direct heating by gas or electricity
60 minutes. or circulating heated air. As a rule, one of the following
(ii) Boiling method conditions is used.
KP X 1827
b. Delayed release dosage forms (also known as en- rotary basket methods is recommended for cap-
teric coated dosage forms) sules, a dosage form that is not ease to sink, or
A delayed release dosage form or enteric coated for slowly dissolving dosage forms. Although
dosage form is a dosage form in which the dissolu- dissolutions from tablets are typically tested
tion of the drug is delayed for a certain period after with the paddle method, the rotary basket meth-
the administration so that the drug is not released od, rather than the paddle method, may be more
in the acidic condition of the stomach by means of suitable in case where the disintegrated debris
composing the product with specialized composi- are settled down at the bottom of the flask there-
tions or specialized methods. Pharmacokinetic cri- by causing a delay in the dissolution. Also disso-
teria of the dosage form are such that the dissolu- lution apparatuses for dissolution testing is rec-
tion profile of the dosage form is comparable to its ommended to accomplish the testing of Suitabil-
conventional release dosage form counterpart ex- ity test for dissolution apparatus (see Appendix 1)
cept that the release occurs after a certain time. for an every certain period of time.
c. Prolonged release dosage forms (also known as
extended release dosage forms)
A prolonged release dosage form or extended re-
lease dosage form is a dosage form in which the
release of the drug substance is prolonged, com-
pared with that from the conventional release dos-
age form of the same route of administration, and
the frequency of administration is reduced com-
pared with that in the conventional forms. These Apparatus 1 Apparatus 2
dosage forms are typically manufactured by means (basket) (paddle)
of specialized compositions and/or specialized
methods.
o In case where these buffer solutions in- medium is not the rate limiting step in the
teract with a drug to create undesirable dissolution rate by maintaining a sink con-
problem, a buffer solution prepared from dition in the test.
citrate or phosphate may be alternatively b) Use of surfactant
used. It is advisable to use a sufficient volume of
d) pH 6.0 Dissolution medium dissolution medium in the test when surfac-
Use disodium phosphate and citric acid tant is added to the medium to enhance the
buffer solution (pH 6.0) of the Korean solubility of a poorly water-soluble drug.
Pharmacopoeia. c) Ease of assay
e) pH 6.8 Dissolution medium For a dissolution test with a drug product
Use the disintegration 2nd fluid. having a trace amount of drug, it is recom-
f) Simulated gastric fluid mended to use the possible minimum in
Use simulated gastric fluid TS of USP. volume when the concentration of the dis-
g) Simulated intestinal fluid solved drug may be lower than the limit of
Use simulated intestinal fluid TS of USP. quantification thereby creating an assay
h) Neutral and basic dissolution medium problem.
Although the dissolution medium, having d) Temperature of test medium
the pH of 1.2, 4.0, 4.5, 6.8 or 7.4, is gener- In general, maintain the temperature of the
ally used, the medium may be set at a dif- medium at 37 ± 5 °C for the dissolution test
ferent pH and have a different composition, of oral solid dosage forms.
if necessary because of the property of the iv. Dissolution testing of drug product containing
dosage form (e.g., the stability and/or the poorly soluble drug
solubility of the drug) or of a specific exper- 1) Determination of impact of medium pH on
imental purpose, and the dissolution is test- dissolution rate
ed. Examples of the special conditions in- Conduct a series of dissolution tests with var-
clude the cases when a particular drug is the ious media such as water, pH 1.2 dissolution
most stable in basic dissolution media, medium, pH 4.0/4.5 dissolution medium and
when basic condition allows a discriminat- pH 6.8 dissolution medium. In these dissolu-
ing analysis or when a given drug has a suf- tion studies, use faster the rotational speed
ficient solubility in basic media. When a (for example, the rotational speed of 120 rpm
dissolution is tested for a product in basic for the rotary basket method and the rotational
condition, coupled with the HPLC assay us- speed of 75 rpm for the paddle method). Sam-
ing C18 column, it may be necessary to neu- ple a portion of the medium at suitable times
tralize or sufficiently dilute the test solution after the initiation of the test, calculate the rate
before the introduction to the column be- of dissolution for the drug and graphically
cause the packing material tends to be un- present the results. Based on these results, the
stable at high pHs. dissolution medium is selected for subsequent
3) Volume of dissolution medium tests. In case where the dissolution rate of the
Although the volume of dissolution medium is drug is below the range between 70 and 85%
typically 900 mL in the test, it may be adjust- of the drug, select a medium that shows the
ed to 500-1 L, depending on the testing condi- most rapid dissolution rate and consider addi-
tion. Dissolution test flask ranging from 150 tion of surfactants to the medium.
mL to over 1 L in size is currently available. 2) Selection of suitable surfactant
General considerations for setting the volume Interaction between drug and surfactant is
of dissolution medium are as follows; governed by the physicochemical properties
a) Solubility of drug of the drug and the surfactant. Therefore, for a
Volume of dissolution medium is sufficient given drug, a various surfactants, e.g., non-
enough to maintain a sink condition during ionic, anionic and cationic surfactants, have to
the testing. In general, a sink condition is a be screened for the selection. In this case, the
condition where at least three times (gener- concentration is recommended to set initially
ally 5-10 times) the volume for the com- at 2% and the suitability is tested for surfac-
plete dissolution of a given amount of drug tants such as;
is available. Therefore, under such condi- o Anionic surfactant: Sodium lauryl
tion, the concentration of the drug should sulfate (SLS)
always be less than 30% of the intrinsic sol- o Cationic surfactant: Cetyltrimethyl
ubility when the drug product is completely ammonium bromide (CTAB)
dissolved in the volume of medium used in o Non-ionic surfactant: Polysorbate 80,
the test. In the dissolution test, it is neces- 40 and 20, lauryl dimethylamine N-
sary to set the dissolution specification such oxide (LDAO)
that the process of the dissolution of drug to 3) Determination of surfactant concentration
KP X 1831
Use the lowest possible concentration of the vi. Two stage dissolution testing for gelatin cap-
surfactant that shows a sufficient dissolution sules
rate within the specified time of the comple- For the case of the dissolution test with hard- or
tion of the test (i.e., in 2 hours or 6 hours). To soft-gelatin capsules, a two-stage dissolution
accomplish this, study the dissolution with a testing is recommended. The first stage dissolu-
gradual decrease in the surfactant concentra- tion is tested using typical dissolution medium
tion such as from the initial 2%, then 1.5, 1.0, as described in the previous section. When the
0.75, 0.5 and finally to 0.25%. dissolution rate is too slow to be tested properly,
v. Dissolution testing for delayed release dosage proceed to the second stage dissolution test with
forms an addition of enzyme to the test medium. Pep-
To confirm the function of enteric coating or the sin or pancreatin is typically used as the diges-
diffusion control of the drug by coating, conduct tive enzyme for the test, and the suitable enzyme
a series of dissolution tests first in an acidic is selected depending on the pH of the medium.
condition (i.e., acid stage) and then in buffer (i.e., When water or a dissolution medium having the
buffer stage). pH of less than 6.8 is used in the test, pepsin is
As an example for a dissolution test for a de- added with the activity of pepsin being not more
layed release dosage form in the official com- than 750000 units per 1 L of the dissolution me-
pendium, dissolution may be tested as directed dium. In this case, the amount of pepsin addition
in the procedure for delayed release dosage is not more than 3.2 g per 1 L of the medium.
forms or first in 0.1 mol/L hydrochloric acid so- For a dissolution medium having the pH not less
lution, an acidic condition, and, then, in pH 6.8 than 6.8, pancreatin is used as the enzyme with
phosphate buffer in the buffer condition. For the the activity of pancreatin being not more than
dissolution testing via the two-stage approach, 1750 units per 1 L of the dissolution medium. In
there are next ways; this case, the amount of pancreatin addition is
1) Dissolution is first tested for 2 hours in 750 not more than 50 mg per 1 L of the medium. In
mL of 0.1 mol/L hydrochloric acid solution at case where water is used as the dissolution me-
37 ± 5 °C. After 2 hours, a portion of the dis- dium at the first stage, 0.1 mol/L hydrochloric
solution medium is sampled for the quantifi- acid solution containing pepsin or pH 6.8 phos-
cation of the drug in the medium. To the phate buffer containing pancreatin may be used
above mentioned 0.1 mol/L hydrochloric acid as the second stage medium. However, if a high
dissolution medium, add 250 mL of 0.2 mol/L concentration of surfactant, such as sodium
sodium phosphate solution, previously laury sulfate, is used in the first stage medium, a
warmed to 37 ± 5 °C. If necessary, the pH of care should be taken in the second stage test be-
the medium may be adjusted with 2 mol/L hy- cause the presence of surfactant may cause de-
drochloric acid or with 1 mol/L sodium hy- naturation of the enzymes in the medium.
droxide, and the dissolution further tested (the
adjustment of the pH should be carried out 5. Dissolution Specification
within not more than 5 minutes of the addition a. Conventional release dosage forms
of 0.2 mol/L sodium phosphate solution). There are two approaches in setting the dissolution
2) Dissolution is first tested for 2 hours in 1 L specification for conventional release dosage
of 0.1 mol/L hydrochloric acid solution at 37 forms.
± 5 °C. After 2 hours, a portion of the dissolu- i. Singe-point specifications: This approach is typ-
tion medium is sampled and the remaining ically applicable to drug products containing
medium is discarded. To the dissolution flask, highly soluble drug substances
transfer 1 L of pH 6.8 phosphate buffer and (Biopharmaceutics Classification System(BCS)
perform the Dissolution Test (prepared by Class I and III) in setting the dissolution specifi-
adding 750 mL of 0.1 mol/L hydrochloric acid cation for the purpose of routine quality control.
solution to 250 mL of 0.2 mol/L sodium phos- (e.g., ‘Perform the test according to the next, the
phate solution, and, if necessary, adjusting the dissolution rate of Acetaminophen in 30 minutes
pH to 6.8 by the addition of 2 mol/L hydro- should be not less than 80%.’) The single-point
chloric acid or of 1 mol/L sodium hydroxide), specification may be set for the dissolution test
previously warmed to 37 ± 5 °C. Another way of a drug product in which not less than 80% of
is to transfer the paddle or the rotary basket, the drug is dissolved within 20-30 minutes, even
containing the sample, from the apparatus of in the presence of the lag time (typically found
the first dissolution study with 0.1 mol/L hy- in film coated formulations and capsules), of the
drochloric acid solution to a new dissolution initiation of the test. It is not necessary to at-
flask containing pH 6.8 phosphate buffer and tempt to construct in vitro-in vivo correlation for
to continue to study the dissolution in the new such product in the development stage.
set up. ii. Two-point specifications: This approach is typi-
cally applicable to drug products containing
1832 General Information
section 6-b-iv)-1) and consult to the condi- of the drug is set as the initial point (1-2
tions in the section 5-b. In this case, use the hours), a point of about 50% dissolution of the
media of pH 1.2 and pH 6.8 as the test media. drug is set as the middle point, and a point of
3) Prolonged release dosage forms about 80% dissolution of the drug or a point
Select at least three time points for sampling when the rate of dissolution does not change
such that a point of about 20-30% dissolution any more is set as a the final point. The rates
of the drug is set as the initial point (1-2 of dissolution for the initial point and the
hours), a point of about 50% dissolution of the middle point are set to ‘the average rate of
drug is set as the middle point and a point of dissolution rate in ± 15%’ and the rate for the
about 80% dissolution of the drug is set as the final point being set to ‘the average rate of
final point. dissolution in – 10%’.
c. The second stage (the dissolution test) 3) Established test the dissolution according to
i. Proceed with the test conditions selected based the preliminary specification using the refer-
on the preliminary test and perform the test ence drug and verify the validity of the speci-
ii. Number of test specimen fication, including the analytical procedure.
Proceed with the test specimens from three lots
having 12 test specimens per each lot and per- 7. Preparation of validation data for dissolution test
form the test. a. Dissolution medium
iii. Dissolution specification Briefly summarize the reason for the selection of
1) Conventional release and delayed release the dissolution medium.
dosage forms b. Dissolution test procedure
a) Time span of test: For conventional release i. Describe the procedure for the dissolution test
dosage forms, use not more than 1 hour as and analysis according to the following items.
the time span of the test. For delayed re- o Dissolution apparatus
lease dosage forms, use not more than 2 o Preparation of the standard solution
hours in the acid stage and not more than 1 o Preparation of the test solution: Sampling
hour in the buffer stage. time of the dissolution solution, dilution, fil-
b) Setting specification of dissolution tration (if there is a filtration procedure in-
Set the specification of the dissolution for volved, present the data for the recovery
the test product at the value of about 10% evaluation)
less than the average rate of dissolution at o Analytical procedure for the quantification
the time point that reaches an almost plat- in the dissolution solution (e.g., HPLC etc.)
eau in the graphical representation of the re- o Formula for the calculation
sults for the dissolution test in the lot that o Dissolution specification
shows the medium rate of dissolution ii. Submit chromatograms or spectra of the blank
among the three lots. test solution, the standard solution and the test
Single-point specification solution.
This specification is applicable to drug c. Validation
product containing a highly soluble drug Describe the validation data for the dissolution test
substance (e.g., BCS Class I or III). Set the and the analytical procedure.
specification as the lower boundary of the i. Data related to the stability of the test solution
dissolution rate at a time point that shows during the analysis
the rate in the range of 70-85%. ii. Analytical procedure
Two-point specification o Perform the test according to the ‘Guideline
This specification is applicable to drug of Validation of Analytical Procedures for
product containing gently dissoluble drug Pharmaceuticals’.
substance (e.g., BCS Class II). 15 Minutes - Range
is generally used as the first point and 30, : Demonstrate linearity, accuracy and preci-
45 or 60 minutes as the second point for the sion within the specification of the dissolution ±
specification. For the drug product in which 20%.
a rapid dissolution may affect the efficacy (For example, for a delayed release dosage
or the adverse reaction, or which has a nar- form having the dissolution specification of
row therapeutic index (see the Attachment 2 not more than 20% at 1 hour, not less than
of the Minimum Requirements for Bioe- 90% at 9 hours, submit the data demonstrat-
quivalence Test), the specification is set us- ing the accuracy, precision and linearity in
ing the two-point testing (if necessary, set the range of 0-110% of the labeled amount).
both upper and lower boundary values).
2) Prolonged release dosage forms Appendix I
Select at least three time points for sampling
such that a point of about 20-30% dissolution Suitability test for dissolution apparatus
1834 General Information
Suitability test has to be conducted to determine concentration and use the standard solutions for
whether a frequently used dissolution apparatus can be the analysis. In case where the test solution is fil-
properly used and the test is carried out by an adequate tered, test for the potential adsorption of salicylic
test procedure. It is recommended that the suitability acid to the filter medium.
test be conducted typically twice a year. In addition,
the test has to be performed when the dissolution appa- Table 1. Dissolution specification of the salicylic acid
ratus is moved or a significant change has been made standard tablet (lot number O)
to the apparatus. Furthermore, the suitability test is Type of dissolution ap- Rate of dissolution at 30
required when the test method is modified from the paratus minutes of the dissolu-
rotary basket method to the paddle method and vice tion test (%)
versa. The suitability of a dissolution apparatus for the Apparatus 1 (the rotary 23-29
rotary basket method or for the paddle method is tested basket method)
using two standard calibrators (USP dissolution cali- Apparatus 2 (the paddle 17-26
brator: www.usp.org). The test is conducted using one method)
tablet of the non-dissolvable salicylic acid standard
tablet and one tablet of the dissolvable prednisone 2. Suitability test for dissolution apparatus using the
standard tablet. For the case of the salicylic acid tablet, 10 mg prednisone standard tablet (lot number N)
50 mmol/L phosphate buffer, pH 7.4 is used as the dis- a. Use 500 mL of water as the dissolution medi-
solution medium and, for the case of the prednisone um.
tablet, water is used as the medium. b. To deaerate the dissolution medium, heat the
1. Suitability test for dissolution apparatus using the solution to 41 °C while stirring, and filter the
300 mg salicylic acid standard tablet (lot number solution under vacuum using the filter paper
O) of the pore size of 0.45 µm. During the filtra-
a. Use 900 mL of 50 mmol/L phosphate buffer tion procedure, continue stirring the filtrate.
as the dissolution medium. In this case, the pH After the filtration procedure, cap the flask
of the buffer is adjusted to 7.4 ± 0.05 in the and continue the stirring for 5 additional
room temperature. minutes under vacuum.
b. To deaerate the dissolution medium, heat the c. Transfer the dissolution medium to the flask
solution to 41 °C while stirring, and filter the of the dissolution apparatus. During the trans-
solution under vacuum using the filter paper fer, a special care should be taken not to trap
of the pore size of 0.45 µm. During the filtra- air bubbles in the medium. Allow the dissolu-
tion procedure, continue stirring the filtrate. tion medium to reach an equilibrium tempera-
After the filtration procedure, cap the flask ture of 37 °C (it is not recommended that ro-
and continue the stirring for 5 additional tate the paddle to facilitate the medium to
minutes under vacuum. achieve the equilibrium).
c. Transfer the dissolution medium to the flask d. When the temperature of the test medium
of the dissolution apparatus. During the trans- reaches at 37 °C, initiate the dissolution test at
fer, a special care should be taken not to trap the rotational speed of 50 rpm.
air bubbles in the medium. Allow the dissolu- e. After 30 minutes of the initiation of the test,
tion medium to reach an equilibrium tempera- sample a portion of the test medium, filter and
ture of 37 °C (it is not recommended that ro- determine the absorbance at 242 nm and cal-
tate the paddle to facilitate the medium to culate the rate of dissolution (%) of salicylic
achieve the equilibrium). acid. If analytically necessary, the filtrate may
d. When the temperature of the test medium be diluted with the dissolution medium.
reaches at 37 °C, initiate the dissolution test f. If dissolution rate(%) is within the next table,
using the rotational speed of 100 rpm. it meets the requirements of the Suitability
e. After 30 minutes of the initiation of the test, test for dissolution apparatus.
sample a portion of the test medium, filter and Note)
determine the absorbance at 296 nm and cal- Dissolve a portion of Prednisone RS in ethanol to
culate the rate of dissolution (%) of salicylic render the concentration of not more than 5%, di-
acid. If analytically necessary, the filtrate may lute the solution with water to obtain a series of
be diluted with the dissolution medium. the standard solutions of known concentration and
f. If the rate of dissolution is within the range use the standard solutions for the analysis. In case
specified in the table below, the dissolution where the test solution is filtered, test for the po-
apparatus is regarded to be suitable. tential adsorption of prednisone to the filter medi-
Note) um.
Dissolve a portion of Salicylic acid RS in ethanol
to render the concentration of not more than 1%, Table 2. Dissolution specification of the prednisone
dilute the solution with the dissolution medium to standard tablet (lot number N)
obtain a series of the standard solutions of known Type of dissolution ap- Rate of dissolution at 30
KP X 1835
made of wire helix and various products, but including - Monochromator compartment: a monochromator
sinker and prolong sinker, may be used as specified in such as diffraction grating, filter, or interference filter
the Dissolution Test. is used.
- Detector compartment: it consists of a detector and
Types of Specification a signal processing system.
sinker - Display unit: it consists of display, recorder, etc.
Basket type sinker (2.6 in radius × and mathematical pretreatment on a spectrum is possi-
1.7 cm) ble.
Specificity is an ability to analyze selectively the nal-to-noise ratio is lowered by 2 times by the first-
analyte, despite of probable existence of interfering order derivative operation and by 4 times by the se-
substances. Generally, a tablet contains lots of sub- cond-order derivative operation. Nonetheless, the sta-
stances in addition to the active ingredients. For exam- bility of recent instruments is excellent and this effect
ple, in the Ambroxol Tablet, the content of the princi- is not a big problem.
pal ingredient, Ambroxol, is 12.5% as 30 mg out of For the construction of calibration model, the use of
210 mg of total mass, and the remaining part, 87.5% of more than one wavelength for the analysis can remove
the total mass may interfere with the analysis of the effect caused by the substances other than the ac-
Ambroxol. These interfering substances are impurities, tive ingredients. An example of chemometric methods
degredation products, water, residual solvents, etc. and is multiple linear regression, which uses more than one
a major part of exceipients. The influence of these sub- wavelength. On the contrary, the methods which use
stances can be reduced by applying chemometric ana- information of a whole range or a certain range of
lytical methods such as the principal component analy- wavelengths are principal component regression, par-
sis. The specificity can be proved by the following tial least squres regression, and so on.
methods. The analysis of the principal components with con-
First, acquire spectra of the active ingredients and all tents of less than 5% of the tablet is expected to be
of other related substances. If the characteristic wave- problematic. Any method for increasing the specificity
lengths of the active ingredients do not overlap with by minimizing or removing the interference comprising
the wavelengths of peaks from other substances when 95% of the tablet may be studied further.
the original spectra without any pretreatment are exam-
ined, the method can be the easiest one to get the speci- 2. Linearity
ficity in analyzing the active ingredients. But when this The linearity of analytical process means the ability
kind of trend is not found with the original untreated to get the result directly propotional to the concentra-
spectra, mathematical pretreatment methods or tion (amount) of analyte in the sample. The linear re-
chemometric resolution algorithms can be applied to gression analysis is performed in the range of concen-
get the specificity by removing the interferences math- tration predictable by the calibration curve of near in-
ematically. frared spectroscopy. Usually, while the calibration
Second, by adding (fortifying the concentration) curve is obtained from the ratios of peak area in the
probable interfering substances to a mixture before range of 0% to 150% in the methods of liquid chroma-
tableting or powdered tablet, the interfering effect on tography, the result of near infrared spectroscopy gives
the analysis is identified. In other words, it should be the linear relationship between a set of reference values
examined whether the wavelength showing variation in the narrower range of 90% to 110% found by the
by the concentration of the additives overlaps with the conventional methods and a set of estimated values
characteristic wavelengths of the active ingredients. found by near infrared spectroscopy. This linear rela-
Third, in many cases, the contents of the principal tionship is evaluated by suitable statistical methods
ingredient are in the range of not less than 95% and not such as calculating the regression line by the least
more than 105% of the labeled amounts. When the squares method. Data such as correlation coefficient of
linear relationship between the actual values and the the regression line, y-intercept, slope and sum of squres
estimated values by near infrared spectroscopy of the of residuals are essential items of the documentation
concentrations outside the allowed range of the labeled for analyzing the regression data.
amounts such as 90% and 110% can be obtained, this
can be the good method to confirm the specificity. 3. Range
Fourth, the most reliable method approaching the Range is the interval between the maximum and the
specificity is the identification and the qualification of minimum of concentrations used for measuring lineari-
substances. The identification of substances is the ty, precision and accuracy during the analytical process.
comparison of the spectra acquired with the test sam- The range of analysis is set to 2 times the allowed
ples, with the spectra acquired with the (powdered) range, for there are lots of difficulties in analyzing tab-
library test samples having variations caused during the lets nondestructive and in collecting test samples for
production process. The qualification of substances is calibration. For example, the allowed range of contents
the comparison of the mean spectrum of the test sam- of the principal ingredients is given to be 5%, consid-
ples, with the library spectra. By these procedures, the ering the degradation of the principal ingredients dur-
placebos can be identified. ing the effective period of use and the error of analyti-
The second derivative operation to spectral data has cal methods. In this case, the minimum range in ana-
several advantages for the analysis. The differentiation lyzing the principal ingredients is 10% of the labeled
between the peaks of the active ingredients and proba- amount. The more accurate calibration be made if the
ble interfering substances can be achieved, and the tablets are collected so that the range is widen to be
background perturbation can be removed. In addition between 80% and 120% of the labeled amount. In re-
to this, peaks can be sharpened, and any variation ality, the range suitable for the purpose of analysis is
caused by the physical states of tablets can be removed, preferred, considering the ranges of 80% to 120% for
resulting in the increased resolution. Usually, the sig- the Assay of the drug substances or the preparations,
1838 General Information
70% to 130% for the Uniformity of Dosage Units and performing the whole procedure 6 times in a short pe-
20% of the range of specification for the Dissolution riod time, as long as there is no thermal degradation of
Test. the samples with concentration equivalent to 100% of
the tested concentration. The sample homogeneity and
4. Accuracy the surface homogeneity should be taken into more
The accuracy of analytical process is the measure consideration especially when the contents of the prin-
expressing the similarity between the values found by cipal ingredients are low. In this case, the transmittance
the conventional analytical method (with proven accu- is better than the reflectance, for the area of transmit-
racy) and the estimated values found by near infrared tance is wider than that of reflectance. Meanwhile, the
spectroscopy. The accuracy should be proved in the transmittance may introduce more noise when the tab-
whole range specified by the analytical method. The lets are thick and the light intensity reaching the detec-
accuracy is a parameter evaluated only after establish- tor is low.
ing precision, linearity and specificity. The representative factors of variation which need
The accuracy can be evaluated by at least 9 meas- examination in within-laboratory reproducibility are
urements in the specified range. For example, the re- the date of experiment, the experimenter, and the in-
sults measured by manipulating 3 levels of concentra- strument used and so on. The inter-laboratory repro-
tion and 3 repeatitions of each concentration level. ducibility is evaluated when the analytical method is
Three minimum levels of concentrations are selected as needed to be standardized for the collaborative experi-
the mean value and both of the maximum and the min- ments between laboratories. If the inter-laboratory re-
imum. The accuracy is expressed as the recovery, when producibility is proven, the within-laboratory reproduc-
the samples spiked with a known level of the analyte ibility is not needed to be validated, but usually, the
are quantitated, or it is examined by comparison be- within-laboratory reproducibility is mainly evaluated,
tween the reference value and the mean value of meas- for the inter-laboratory reproducibility is hard to be
urements in the confidence interval, when the compari- achieved.
son with the true value or the certified or agreed value
is made. In the quantitative analysis by near infrared III. Qualification of NIR Spectrophotometers
spectroscopy, the accuracy is expressed as the standard
error of prediction found with the validation set. The purpose of qualification of NIR spectrophotom-
The calibration set should consist of the samples eters is to verify the suitability of the instrument for the
with the concentrations of the active ingredients pro- intended use by comparing with the specification of the
duced in the batch process and it should include the instrument. The qualification procedure includes De-
whole range of the sample. The selected samples sign Qualification (DQ), Installation Qualification (IQ),
should show the uniform distribution of probability, Operational Qualification (OQ) and Performance Qual-
not the normal distribution. The validation set should ification (PQ).
consist of the samples with the same composition of
the active ingredients as in the calibration set. 1. Design Qualification (DQ)
The set of samples for the parallel test are selected The design qualification provides the evidence
from the independent production batch and the analysis whether the instrument is properly designed to be oper-
is validated by the time. Therefore, the accuracy of the ational for the intended use. The various influencing
analysis can be examined using the set completely in- factors of the instrument are tested to verify whether
dependent from the calibration set and the validation they correspond to the specifications of the instrument.
set of samples. This test is performed once a month At least, it should be verified whether the factors corre-
with independent set of samples after development of spond the manufacturer’s specifications of the instru-
the calibration model. ment.
The test set is selected manually or by software the concentration range is wider (80% to 120%), for
method so that the composition of the analytes is uni- the amount of the active ingredients can be varied
formly distributed. from the first stage of the sample preparation.
The test set can be selected on the basis of the spec- Fourth, the active ingredients and the excipients are
tral variation by software method. added to the powdered tablets and then the mixture
Whichever method is used, the care should be taken is tableted. But in this case, the total mass of the tab-
to have uniform distribution by the concentration range. let is different from that of the products and it may
When the distribution is not uniform (for example, fail in terms of specificity.
concentrations at the center are used), there should be a Fifth, the model comprising all of the variation fac-
proper explanation. tor of the production process by using the global cal-
ibration set, which includes all of the existing varia-
The sample size measurable by NIR spectroscopy is tion factors to make the calibration models. This
significantly smaller than those by the conventional method consists of 4 steps: laboratory composite
method. This is not because of the sample measure- (prepared to have active ingredients with the concen-
ment compartment but because of the sample area il- tration level of 15% of the labeled amount), prepa-
luminated by the near infrared light. The infrared ration of granules, tableting of core tablet, and coat-
measurement can detect the inhomogeneity even with ing the tablets.
the mass of microgram level and a proper measurement
method is needed to accommodate this variation. This 4. Acquisition of sample spectra
property can be an advantage useful for testing the The transmittance is applied to liquids, diluted or
Uniformity of Dosage Units, but in many applications, undiluted and solids in the solution. The cells with
averaging the signal from a larger area is needed. transmittance pathlength of 0.5 to 4 mm or dip-probes
Therefore, to get the averaged information, the sample are used and the compensation for background inter-
is moved transversely or rotated during the data acqui- ferences should be made. On the contrary, the reflec-
sition. tance and the diffused reflectance are applied mainly to
Among the various preparations, in cases of the tab- solids and the test is done to the sample placed in a
lets, there are two cases: the case of using the products proper apparatus. When the fiber-optics are used, re-
as they are and the case of varying ratios of the active producible spectra can be acquired by fixing the sam-
ingredients to the excipients. ple properly. As in the transmittance, compensation for
background interferences should be made.
(1) The case of using the products as they are In cases of solid particles, the physical differences
First, this can be applied only when the transmit- such as particle size, shape, compressibility, etc. can be
tance follows the Beer-Lambert’s law and it also ap- influential, so the particles are manipulated to be as
plied to other tablets of the same composition with small and uniform in size as possible. Moreover, to
different masses. make the effect of the absorbed water and the residual
Second, this can be applied when the samples are solvent constant, the samples are dried for a certain
collected from the normal production and each tablet length of time to have the same condition of the ab-
has the concentration range needed for the usage. sorbed water and the residual solvent of the samples.
When the samples are analyzed by the NIR spectrosco-
(2) The case of varying ratios of the active ingredi- py, the NIR spectra are influenced a lot by the ab-
ents to the excipients sorbed water and the temperature, so it is desirable to
First, the calibration test set is obtained by mixing construct the environment of constant humidity and
the products and the concentration expansion sam- constant temperature so that the experiment can be
ples. Usually, the products lie between 95% and 105% done in the same environment. Moreover, the differ-
of the labeled amount, so to expand this range, tab- ence in the polymorphism of the sample or in the de-
lets with expanded concentration range are synthe- gree of crystalization also can be influential to the
sized in the laboratory. This concentration expansion quantitative analysis, so the case should be taken.
samples have different ratios of the active ingredi- After selecting a proper sample measurement meth-
ents to the excipients and the ranges of between 90% od among various sample measurement apparatus by
and 110% of the labeled amount or between 85% considering the property, the shape, and the size of the
and 115% of the labeled amount are used. sample, the precision of the spectra acquired by repeat-
Second, the concentration is adjusted by adding the ed measurement is evaluated. When the measurement
active ingredients and the excipients to the powdered is repeated 6 times, it meets the requirement if the
products. If the active ingredients are low compared maximum value of the relative standard deviations of
to the total mass of the tablet, the better calibration whole range of wavelengths is not more than 1.0%.
results can be acquired by adding the active ingredi-
ents to the powdered tablets. 5. Construction of calibration model
Third, the concentration of the active ingredients is For the development of a calibration model, a proper
varied using the sample synthesized in the laboratory. mathematical preptreatment is needed to be done, if
This is similar to the use of the powdered tablet, but necessary. The pretreatment procedure is an important
KP X 1841
step in chemometric analysis of the NIR data. The pre- converting the amount (mg) of the analyte to 100 is not
treatment is defined as a procedure in which the spec- exceeding the error range of the conventional method
tral shape is enhanced by mathematical treatment of the of analysis, the use of the model is recommended.
NIR data or unwanted variation is removed prior to the
development of a calibration model. A proper pre- 7. Validation of model performance
treatment method can be selected by testing the spec- Not just to evaluate simply the model, the perfor-
tral data prior to the data modeling. For example, vari- mance of the model is needed to be validated periodi-
ous methods of pretreatment are used in parallel and cally to use that model continuously in the industrial
evaluated and the best one is selected among them. fields. The methods for validating model performance
There are various methods of pretreatment. For ex- can be classified into two methods, which are the
ample, they are normalization, smoothing, baseline method of using the check samples and the method of
correction, derivatives, mean centering, variance scal- comparing with the reference analytical method. First,
ing, autoscaling and so on. to use the check samples, the check samples should be
After selecting a proper method of pretreatment, re- stable over time and the short-term and long-term accu-
gression analysis on the NIR data is done by applying racy of the model should be evaluated with these sam-
various quantitative algorithms. The calibration is a ples. Second, to use the method of comparing with the
procedure by which a mathematical model is con- reference analytical method, the data by the NIR spec-
structed between the instrumental responses and the troscopy and the reference data are acquired periodical-
properties of the sample (usually concentration). As ly for n months (n=1,2,3, ) over n batches and a
mentioned previously, the major algorithms for calibra- paired t-test with the confidence level of 95% is per-
tion are multiple linear regression (MLR), principal formed. When there is significant difference, the quan-
component regression (PCR), partial least squares re- titative model should be reconstructed.
gression (PLSR) and so on. The prediction is a proce-
dure of predicting the properties of the samples from * Refer to the Appendix II for the validation of the
the instrumental signals using the developed model. In model.
a broad sense, there are two different approaches to * Refer to the Appendix III for the usual quantitative
develop the calibration model and they are the analysis and the maintenance of the quantitative model.
univariate analysis and the multivariate analysis. The
univariate analysis is the most commonly used method V. Qualitative Analysis
in the conventional analysis, which relates the single
signal of the analytical instrument is related to the con- The near infrared spectroscopy can be used for the
centration of the single ingredient. This method is not identification of substances and the qualitative analysis.
commonly used in the NIR spectroscopy. In the NIR
spectroscopy, many signals from the analytical instru- Identification : It is used when the chemical identifi-
ment are related to the various properties of the sam- cation of the substance is needed.
ples and the calibration model is constructed by the Qualitative analysis : After chemical identification
multivariate analytical methods. of the substance, the suitability of the sample to the
model of substances is measured. The model is devel-
* For constructing a quantitative model, refer to the oped from the samples representing various infor-
Appendix I. The Appendix I explains mainly MLR, mation on the variation and it includes water, particle
PCR and PLSR, which are used generally in the NIR size, solvent and other chemical and physical
spectroscopy. informations.
Both of the identification and the qualitative analysis
6. Validation of calibration model enable the discrimination of substances in the library.
The quantitative model can be validated internally or Typical qualitative application of the NIR spectroscopy
externally. Independent validation test set is used to get consists of the following procedures.
the information on the predictability of the developed
model. The accuracy and the precision of the NIR • Feasibility test
spectroscopy is compared with those of the reference • Selection of the samples
analytical method. Standard error of calibration (SEC) • Acquisition of sample spectra
and standard error of prediction (SEP) are used to • Construction of library
evaluate the quality of the quantitative model. A con- • Validation of library
ventional factor of model evaluation, the correlation • Routine use
coefficient of regression is also used in the NIR spec- • Maintenance of library
troscopy, but it is not as important as in the conven-
tional univariate analysis. 1. Feasibility test
When the content of a tablet is analyzed, the ratio (%) The feasibility test is performed as a first step, prior
of the amount of analyte to the mass of the tablet is to the development of model. For example, optimal
used as the value of concentration. When the calculated method of sample measurement, amount of the sample,
SEP value derived as the relative error obtained by the minimum number of scans for effective analysis are
1842 General Information
examined. A prior knowledge on the organization of sorbed water and the temperature, so it is desirable to
samples in the library and the molecular structure is construct the environment of constant humidity and
useful in the analysis with the NIR spectroscopy. And constant temperature so that the experiment can be
when the spectra of the representative samples are ac- done in the same environment. Moreover, the differ-
quired, it is checked whether the second derivative ence in the polymorphism of the sample or in the de-
spectra of the samples are different from each other. gree of crystalization also can be influential to the
quantitative analysis, so the case should be taken.
2. Selection of the samples After selecting a proper sample measurement meth-
The selection of the samples is an important proce- od among various sample measurement apparatus by
dure for the successful qualitative analysis. The sample considering the property, the shape, and the size of the
set for developing the library and the independent set sample, the precision of the spectra acquired by repeat-
for the purpose of validation are needed. All of the ed measurement is evaluated. When the measurement
samples used for building and validating the library are is repeated 6 times, it meets the requirement if the
needed to be authenticated. The level of authentication maximum value of the relative standard deviations of
of the sample is varied by the usage and the database whole range of wavelengths is not more than 1.0%.
consists of the samples showing various source of vari-
ation. 4. Construction of Library
The samples from the different batches are collected, The following procedures are used to develop the li-
reflecting the changes in ingredients, suppliers, pro- brary.
cesses, and storage conditions over a certain period of
time. If the chemical and physical stability is proven • Definition of the purpose of developing the library
over the storage period, the samples can be collected • Selection of the sample set for developing the li-
regardless of the storage status. brary
The number of batches to be collected is dependent • Data display
on the complexity of analysis and the substances to be • Selection of the sample set for validating the library
analyzed should include the typical variation of the • Data pretreatment/conversion
system. The number of batches for the qualitative anal- • Construction of the library
ysis is greater than that for the identification. There are • Setting the threshold
many kinds of the sample measurement apparatus, such
as cups, vials, fiber-optics and custom-made apparatus. 1) Definition of the purpose of developing the library
The selection of the apparatus is dependent on the us- Prior to the development of the library, it is im-
er’s needs and the validation of the apparatus is defined portant to set the effective range of the library accord-
in the stage of the design qualification (DQ). The sam- ing to the intended use. This applies to the identifica-
ple measurement apparatus is the source of latent varia- tion and the qualitative analysis of the substances and it
tions during the measurement. Therefore, it should be includes checking the chemical similarity of the groups
validated continuously in terms of reproducibility, if to be analyzed and the number of groups.
possible.
2) Selection of the sample set for developing the li-
3. Acquisition of Sample Spectra brary
The transmittance is applied to liquids, diluted or For the development of the library, the variation
undiluted and solids in the solution. The cells with caused by the following factors should be considered.
transmittance pathlength of 0.5 to 4 mm or dip-probes The factors, especially important for the library for the
are used and the compensation for background inter- qualitative analysis are as follows:
ferences should be made. On the contrary, the reflec-
tance and the diffused reflectance are applied mainly to • Moisture
solids and the test is done to the sample placed in a • Particle size
proper apparatus. When the fiber-optics are used, re- • Residual solvents
producible spectra can be acquired by fixing the sam- • Degradation products
ple properly. As in the transmittance, compensation for • Compositional change of formulated product
background interferences should be made. • Other chemical/physical properties
In cases of solid particles, the physical differences • Time
such as particle size, shape, compressibility, etc. can be • Alternative sources of material
influential, so the particles are manipulated to be as • Retained samples
small and uniform in size as possible. Moreover, to • Temperature
make the effect of the absorbed water and the residual • Operator
solvent constant, the samples are dried for a certain • Presentation, e.g. insertion of tube
length of time to have the same condition of the ab- • Between-instrument variation
sorbed water and the residual solvent of the samples. • Others
When the samples are analyzed by the NIR spectrosco-
py, the NIR spectra are influenced a lot by the ab-
KP X 1843
The range of consideration over these factors is de- unwanted effects or in bringing a small but important
pendent on the intended use and the needed ability of difference into relief.
classification. Like any other analytical techniques, the NIR spec-
troscopy cannot differentiate all of the substance
3) Data display groups, especially very similar groups. In this case, two
The data display checks visually the sample showing groups are combined to one group or any other control
strange spectral features and identifies outliers. If pos- method is used to perform the identification and quali-
sible, outliers should be identified and if there is rea- tative analysis.
sonable analytical explanation, they can be removed There are many algorithms such as correlation, soft
and in that case, the clear reason should be documented. independent modeling of class analogy (SIMCA),
Mahalanobis distance, SMV and so on. The user se-
4) Selection of the sample set for validating the li- lects a proper algorithm with the consideration of the
brary effective range of the library. When the discrimination
There can be a situation, in which the representative of the substances is each, it is recommended to use the
samples are needed to be selected from the large group. simplest algorithm. For example, when the purpose is
In the simple situation, they can be selected manually, only the identification of the substance, the physical
but in more complex situation, they should be selected factors are not needed to be considered, so the second
after the determination of substance groups using a derivatives and the wavelength correlation algorithm
proper selection tool (for example, principal compo- can be selected.
nent analysis, cluster analysis, etc.).
The number of samples for each of the substance 7) Setting the threshold
groups is dependent on the qualitative algorithm and The internal validation is done with the value set by
the state of the system (how accurately the boundaries the software itself or recommended by the manufactur-
of groups needed to be set). er. The threshold of the library can be changed when
the library is validated internally or the sample is eval-
5) Data pretreatment/conversion uated externally.
Data are needed to be pretreated mathematically to
simplify the spectra more. For example, the algorithms 5. Validation of Library
of derivatives and scatter correction can remove the The purpose of validation of an analytical procedure
background variation caused by the physical difference. is to check whether the analysis is suitable for the in-
The original untreated spectra are used when the in- tended use. Upon this purpose, the factors influencing
formation due to the physical status is important. the needed validation should be determined.
Mathematical transformation of spectra data can be
an artifact and can cause loss of important information, 1) Internal validation
so this should be taken into consideration. The algo- The performance of the library should be evaluated
rithms of data pretreatment and conversion should be in the course of constructing any kind of spectral data-
understood correctly and the theoretical basis always base. This is based on the samples constructing the
should be provided whenever any conversion is per- library (whether these samples can be differenctiated).
formed. The internal validation is done by the software and the
detailed procedure may be different by the used soft-
6) Construction of the library ware, but the basic procedure is as follows:
The structure of the library is dependent on the limi-
tation of the software and the user’s needs. In the sim- • The spectra used to construct the library are vali-
plest case, all of the substances are combined into one dated by a proper method (correlation or distance
library. On the contrary, it can be divided into sub- method)
libraries to secure the needed level of the specificity. • It is checked whether the distribution of the sub-
The mathematical transformation procedure may be stances in the library overlaps.
same for all of the substance groups in the principal • The cross-validation is used for the construction of
library. The transformation may be same within the the library.
sub-library but different between the sub-libraries. For
example, this situation is met when the grade of the 2) External validation
lactose is to be determined after the sample is identi- When the internal validation is successful, the per-
fied as lactose from the library of excipients. The spec- formance of the database is validated using certified
tral range can be all of the available wavelengths or can samples not used for the database development.
be limited to cetain wavelengths. The range of wave-
lengths can be limited when a certain measurement Reproducibility: This is not commonly applied
apparatus is used or unnecessary spectral information method for the identification of substances. In the re-
is needed to be removed (outside the dynamic range or producibility test for the qualitative analysis, the sub-
regions of high noise level). An analytical method of stances included in the library should be reliably dif-
partitioning the wavelength range is useful in removing
1844 General Information
ferentiated from the substances outside the library, To add new substances into the library, the test set of
using the threshold value. substances is selected to satisfy the details of sample
Robustness: This category is dependent on the appli- selection parameters and the library is evaluated again
cation and the sample selection technique and the ef- for the validation of the specificity, continuously.
fect of delicate variation in the normal operating condi-
tion of the analysis is tested. The use of the design of 3) Fixing the groups in the library
experiment can maximize the effect of the information Sometimes, the sample set is fixed in the cases as
on the analyte. The following items are considered: follows:
• The effect of environmental conditions (tempera- • Change in physical properties of the substance
ture, humidity) of the analysis • Change in supplier
• The effect of temperature on the sample • Inclusion of the wider range
• The location of sample
• The depth of probe and the degree of compres- In each case, new samples are certified by the meth-
sion/filling of the substance od other than the NIR spectroscopy, before inserting
• Other effect of the sample measurement apparatus them in the model and the library is evaluated again for
• The effect of replacement of parts (lamp) the validation of the specificity, continuously.
• The effect of pretreatment and parameters of the al-
gorithm used to construct the library (gap/segment * Refer to the Appendix 5 when there is any problem
of the derivative operation, distance threshols, etc.) in applying the library for the qualitative analysis.
Appendix III. Routine procedure of quantitative analysis and Maintenance of the quantitative model.
1848 General Information
Class 3 solvents: Solvents with low toxic potential permitted daily exposure to acetonitrile is 4.1 mg/day;
Solvents with low toxic potential to man; thus, the option 1 limit is 410 ppm.
no health-based exposure limit is needed. The maximum administered daily mass of a drug
Class 3 solvents have PDEs of 50 mg or more product is 5.0 g, and the drug product contains two
per day. excipients. The composition of the drug product and
the calculated maximum content of residual acetonitrile
3.2 Methods for Establishing Exposure Limits are given in the following table.
The method used to establish permitted daily expo-
sures for residual solvents is presented in Appendix 3. Compo- Amount in Acetonitrile Daily
Summaries of the toxicity data that were used to estab- nent formulation content exposure
lish limits are published in Pharmeuropa, Vol. 9, No. 1,
Supplement, April 1997. Drug
0.3 g 800 ppm 0.24 mg
substance
3.3 Options for Setting Limits of Class 2 Solvents Excipient 1 0.9 g 400 ppm 0.36 mg
Two options are available when setting limits for
Excipient 2 3.8 g 800 ppm 3.04 mg
Class 2 solvents.
Drug
5.0 g 728 ppm 3.64 mg
Option 1: The concentration limits in ppm stated in Product
Table 2 can be used. They were calculated us-
ing equation (1) below by assuming a product Excipient 1 meets the Option 1 limit, but the drug
mass of 10 g administered daily. substance, excipient 2, and drug product do not meet
the Option 1 limit. Nevertheless, the product meets the
Concentration (ppm) = 1000 PDE (1) Option 2 limit of 4.1 mg per day and thus conforms to
dose the recommendations in this guideline.
Consider another example using acetonitrile as re-
Here, PDE is given in terms of mg/day and dose is sidual solvent. The maximum administered daily mass
given in g/day. of a drug product is 5.0 g, and the drug product con-
These limits are considered acceptable for all sub- tains two excipients. The composition of the drug
stances, excipients, or products. Therefore this option product and the calculated maximum content of residu-
may be applied if the daily dose is not known or fixed. al acetonitrile are given in the following table.
If all excipients and drug substances in a formulation
meet the limits given in Option 1, then these compo- Compo- Amount in Acetonitrile Daily
nents may be used in any proportion. No further calcu- nent formulation content exposure
lation is necessary provided the daily dose does not
Drug
exceed 10 g. Products that are administered in doses 0.3 g 800 ppm 0.24 mg
substance
greater than 10 g per day should be considered under
Option 2. Excipient 1 0.9 g 2,000 ppm 1.80 mg
Excipient 2 3.8 g 800 ppm 3.04 mg
Option 2: It is not considered necessary for each com-
Drug
ponent of the drug product to comply with the 5.0 g 1,016 ppm 5.08 mg
Product
limits given in Option 1. The PDE in terms of
mg/day as stated in Table 2 can be used with
In this example, the product meets neither the Op-
the known maximum daily dose and equation
tion 1 nor the Option 2 limit according to this summa-
(1) above to determine the concentration of re-
tion. The manufacturer could test the drug product to
sidual solvent allowed in drug product. Such
determine if the formulation process reduced the level
limits are considered acceptable provided that
of acetonitrile. If the level of acetonitrile was not re-
it has been demonstrated that the residual sol-
duced during formulation to the allowed limit, then the
vent has been reduced to the practical mini-
manufacturer of the drug product should take other
mum. The limits should be realistic in relation
steps to reduce the amount of acetonitrile in the drug
to analytical precision, manufacturing capabil-
product. If all of these steps fail to reduce the level of
ity, reasonable variation in the manufacturing
residual solvent, in exceptional cases the manufacturer
process, and the limits should reflect contem-
could provide a summary of efforts made to reduce the
porary manufacturing standards.
solvent level to meet the guideline value, and provide a
risk-benefit analysis to support allowing the product to
Option 2 may be applied by adding the amounts of a
be utilized with residual solvent at a higher level.
residual solvent present in each of the components of
the drug product. The sum of the amounts of solvent
3.4 Analytical Procedures
per day should be less than that given by the PDE.
Residual solvents are typically determined using
Consider an example of the use of Option 1 and Op-
chromatographic techniques such as gas chromatog-
tion 2 applied to acetonitrile in a drug product. The
raphy. Harmonized procedures for determining levels
1852 General Information
4.3 Solvents with Low Toxic Potential Furthermore, for the Class 2 or Class 3 solvent is used
Solvents in Class 3 (shown in Table 3) may be re- prior to the last step of a manufacturing process of drug
garded as less toxic and of lower risk to human health. substances, it is not necessary to establish the specifi-
Class 3 includes no solvent known as a human health cation of the residual solvents of the drug substances,
hazard at levels normally accepted in pharmaceuticals. in case of complying with the requirements of the re-
However, there are no long-term toxicity or carcino- sidual solvents that are not remain in the final active
genicity studies for many of the solvents in Class 3. drug substances.
Available data indicate that they are less toxic in acute (Examples of the recommended data) the content of 3
or short-term studies and negative in genotoxicity stud- consecutive batches of residual solvents that is not
ies. It is considered that amounts of these residual sol- more than acceptable concentration limit and the data
vents of 50 mg per day or less (corresponding to 5000 of detection limit. When Class 2 solvents are used, they
ppm or 0.5% under Option 1) would be acceptable should be controlled either in a suitable intermediate or
without justification. Higher amounts may also be ac- in the management data of final active substances de-
ceptable provided they are realistic in relation to manu- pending on the batch scale.
facturing capability and good manufacturing practice
(GMP). 5. GLOSSARY
Genotoxic Carcinogens: Carcinogens which pro-
Table 3. Class 3 solvents, which should be limited by duce cancer by affecting genes or chromosomes.
GMP or other quality-based requirements, in LOEL: Abbreviation for lowest-observed effect
pharmaceutical products. level.
Acetic acid Heptane
Lowest-Observed Effect Level: The lowest dose
Acetone Isobutyl acetate
of substance in a study or group of studies that
Anisole Isopropyl acetate
produces biologically significant increases in fre-
1-Butanol Methyl acetate
quency or severity of any effects in the exposed
2-Butanol 3-Methyl-1-butanol
humans or animals.
Butyl acetate Methylethyl ketone
Modifying Factor: A factor determined by profes-
tert-Butylmethyl ether Methylisobutyl ketone
sional judgment of a toxicologist and applied to
Dimethyl sulfoxide 2-Methyl-1-propanol
bioassay data to relate that data safely to humans.
Ethanol Pentane
Ethyl acetate 1-Pentanol Neurotoxicity: The ability of a substance to cause
adverse effects on the nervous system.
Ethyl ether 1-Propanol
Ethyl formate 2-Propanol NOEL: Abbreviation for no-observed-effect level.
Formic acid Propyl acetate No-Observed-Effect Level: The highest dose of
substance at which there are no biologically sig-
4.4 Solvents for which No Adequate Toxicological nificant increases in frequency or severity of any
Data was Found effects in the exposed humans or animals.
The following solvents (Table 4) may be of inter- PDE: Abbreviation for permitted daily exposure.
est to manufacturers of excipients, drug substances, or Permitted Daily Exposure: The maximum ac-
drug products. However, no adequate toxicological ceptable intake per day of residual solvent in
data on which to base a PDE was found. Manufacturers pharmaceutical products.
should supply justification for residual levels of these Reversible Toxicity: The occurrence of harmful
solvents in pharmaceutical products. effects that are caused by a substance and which
disappear after exposure to the substance ends.
Table 4. Solvents for which no adequate toxicologi- Strongly Suspected Human Carcinogen: A sub-
cal data was found. stance for which there is no epidemiological evi-
dence of carcinogenesis but there are positive
Methylisopropyl ke- genotoxicity data and clear evidence of carcino-
1,1-Diethoxypropane genesis in rodents.
tone
1,1-Dimethoxymethane Methyltetrahydrofuran Teratogenicity: The occurrence of structural mal-
formations in a developing fetus when a sub-
2,2-Dimethoxypropane Petroleum ether
stance is administered during pregnancy.
Isooctane Trichloroacetic acid
Acetonitrile Class 2
Anisole Methoxybenzene Class 3
Benzene Benzol Class 1
1-Butanol n-Butyl alcohol Class 3
Butan-1-ol
2-Butanol sec-Butyl alcohol; Butan-2-ol Class 3
Butyl acetate Acetic acid butyl ester Class 3
tert-Butylmethyl ether 2-Methoxy-2-methyl- propane Class 3
Carbon tetrachloride Tetrachloromethane Class 1
Chlorobenzene Class 2
Chloroform Trichloromethane Class 2
Cumene Isopropylbenzene; (1-Methyl)ethylbenzene Class 2
Cyclohexane Hexamethylene Class 2
1,2-Dichloroethane sym-Dichloroethane; Ethylene dichloride; Ethylene chloride Class 1
1,1-Dichloroethene 1,1-Dichloroethylene; Vinylidene chloride Class 1
1,2-Dichloroethene 1,2-Dichloroethylene; Acetylene dichloride Class 2
Dichloromethane Methylene chloride Class 2
1,2-Dimethoxyethane Ethyleneglycol dimethyl ether; Monoglyme; Dimethyl Cellosolve Class 2
N,N- DMA Class 2
Dimethylacetamide
N,N- DMF Class 2
Dimethylformamide
Dimethyl sulfoxide Methylsulfinylmethane; Methyl sulfoxide; DMSO Class 3
1,4-Dioxane p-Dioxane; [1,4]Dioxane Class 2
Ethanol Ethyl alcohol Class 3
2-Ethoxyethanol Cellosolve Class 2
Ethyl acetate Acetic acid ethyl ester Class 3
Ethyleneglycol 1,2-Dihydroxyethane; 1,2-Ethanediol Class 2
Ethyl ether Diethyl ether; Ethoxyethane; 1,1’-Oxybisethane Class 3
Ethyl formate Formic acid ethyl ester Class 3
Formamide Methanamide Class 2
Formic acid Class 3
Heptane n-Heptane Class 3
Hexane n-Hexane Class 2
Isobutyl acetate Acetic acid isobutyl ester Class 3
Isopropyl acetate Acetic acid isopropyl ester Class 3
Methanol Methyl alcohol Class 2
2-Methoxyethanol Methyl Cellosolve Class 2
Methyl acetate Acetic acid methyl ester Class 3
3-Methyl-1-butanol Isoamyl alcohol; Isopentyl alcohol; 3-Methylbutan-1-ol Class 3
Methylbutyl ketone 2-Hexanone; Hexan-2-one Class 2
Methylcyclohexane Cyclohexylmethane Class 2
Methylethyl ketone 2-Butanone; MEK; Butan-2-one Class 3
Methylisobutyl ketone 4-Methylpentan-2-one; 4-Methyl-2-pentanone; MIBK Class 3
2-Methyl-1-propanol Isobutyl alcohol; 2-Methylpropan-1-ol Class 3
N-Methylpyrrolidone 1-Methylpyrrolidin-2-one; 1-Methyl-2-pyrrolidinone Class 2
Nitromethane Class 2
Pentane n-Pentane Class 3
1-Pentanol Amyl alcohol; Pentan-1-ol; Pentyl alcohol Class 3
1-Propanol Propan-1-ol; Propyl alcohol Class 3
2-Propanol Propan-2-ol; Isopropyl alcohol Class 3
Propyl acetate Acetic acid propyl ester Class 3
Pyridine Class 2
Sulfolane Tetrahydrothiophene 1,1-dioxide Class 2
Tetrahydrofuran Tetramethylene oxide; Oxacyclopentane Class 2
Tetralin 1,2,3,4-Tetrahydro-naphthalene Class 2
Toluene Methylbenzene Class 2
1,1,1-Trichloroethane Methylchloroform Class 1
KP X 1855
Pharmeuropa, Vol. 9, No. 1, Supplement, April 1997, The degree of revalidation required depends on the
page S9. nature of the changes.
In addition, validation methods, other than those
1
n P 00 x 10 atm x 15 8 0 mg mol described in this guideline, may be used, although
the appropriateness of the alternative method has
RT 0 08 L atm 1 mol 1 x 8
15 mg to be justified.
1 8 mg L
5L
III. GLOSSARY
The relationship 1000 L = 1 m3 is used to convert to 1. Analytical Procedure
mg/ m3. The analytical procedure refers to the way of per-
forming the analysis. It should describe in detail
the steps necessary to perform each analytical test.
The analytical procedure includes but is not lim-
7. Guideline of Validation of ited to: the analyte, the sample, the reference
Analytical Procedures for standard, the standard reagents, the use of analyti-
cal instruments, the use of the calibration curve,
Pharmaceuticals the use of formulae for the calculation in the iden-
tification test, purity test or assay.
I. OBJECTIVE - The term ‘validation of analytical proce-
The purpose of this guideline is to provide the de- dure’ is defined as the process of confirming
tailed guidance on how to conduct the validation of that the analytical procedure for a quality
the analytical procedures necessary for application control test of pharmaceuticals is suitable for
(approval) for manufacture · import of pharmaceu- its intended use.
ticals · quasi-drugs according to notifications of - Identification Test refers to ensure the iden-
the Korean Food and Drug Administration, such as tity of an analyte, and generally compare the
‘Regulation of Pharmaceuticals Approval, Notifi- physicochemical characteristics (spectrum,
cation and Review’, and for quality control. information from chromatographic methods,
and chemical reactivity, etc) of a sample with
II. INTRODUCTION those of a reference material.
The objective of validation of an analytical pro- - Purity Test is to ensure that all the analytical
cedure for pharmaceuticals is to demonstrate that procedures performed allow an accurate
the procedure, applied in the quality control of the statement of the content of impurities of an
pharmaceutical, is suitable for the intended pur- analyte, i.e. related substances test, heavy
pose. metals, residual solvents content, etc. Testing
The purpose is to direct how to establish relevant for impurities can be either a quantification
validation parameters for each analytical procedure. test or a limit test for the impurity in a sample.
The scope of analytical procedures that are sub- - Assay (Content or Potency) is to provide an
jected to the present guideline is as follows: exact result which allows an accurate state-
1) Identification tests in the specification for ment on the content or potency of the analyte
pharmaceuticals in a sample. The assay represents a quantita-
2) Purity tests: quantitative tests and limit tests tive measurement of the major component(s)
for the impurity in a sample or other selected component(s) (i.e., stabilizer,
3) Quantification tests: assay for the active additive) in the drug substance or in the drug
component of drug substance or drug prod- product. The validation principles are also
uct, for other selected component(s) in drug applicable to assay for dissolution test.
products, for Uniformity of Dosage Units
test or for dissolution test 2. Specificity
The objective of the analytical procedure should Specificity is the ability to assess unequivocally
be clear since the procedure will govern the valida- the analyte in the presence of components, e.g.,
tion characteristics which need to be evaluated. impurities, degradation products, matrix, etc,
Typical validation parameters include specificity, which may be expected to be present. Lack of
accuracy, precision, detection limit, quantification specificity of an individual analytical procedure
limit, linearity, range, and robustness. Appropriate may be compensated by other supporting analyti-
validation parameters are selected and evaluated cal procedure(s).
depending on the objective of a given analytical
procedure. 3. Accuracy
Furthermore, revalidation may be necessary when The accuracy of an analytical procedure express-
there are change in the synthesis of the drug sub- es the closeness of agreement between the value
stance, change in the composition of the finished which is accepted either as a conventional true
product, and change in the analytical procedure. value or an accepted reference value and the value
1858 Monographs, Part II
Critical separations in chromatography should the range (see section 3) of the analytical proce-
be investigated at an appropriate level to demon- dure. It may be demonstrated directly on the drug
strate that the components are adequately sepa- substance (by dilution of a standard stock solution)
rated. For critical separations, specificity can be and/or separate weighings of synthetic mixtures of
demonstrated by the resolution of the two com- the drug product components, using the proposed
ponents which elute closest to each other. procedure. The latter aspect can be studied during
In cases where a non-specific assay is used, oth- investigation of the range.
er supporting analytical procedures should be Linearity should be evaluated by visual inspec-
used to demonstrate overall specificity. For ex- tion of a plot of signals as a function of analyte
ample, where a titration is adopted to assay the concentration or content. If there is a linear rela-
drug substance for release, the combination of the tionship, test results should be evaluated by appro-
assay and a suitable test for impurities can be priate statistical methods, for example, by calcula-
used. tion of a regression line by the method of least
The approach is similar for both assay and im- squares. In some cases, to obtain linearity between
purity tests. assays and sample concentrations, the test data
may need to be subjected to a mathematical trans-
(1) Impurities are available formation prior to the regression analysis. Data
For the assay, this should involve demonstration from the regression line itself may be helpful to
of the discrimination of the analyte in the pres- provide mathematical estimates of the degree of
ence of impurities and/or excipients; practically, linearity.
this can be done by spiking pure substances (drug The correlation coefficient, y-intercept, slope of
substance or drug product) with appropriate lev- the regression line and residual sum of squares
els of impurities and/or excipients and demon- should be submitted. A plot of the data should be
strating that the assay result is unaffected by the included. In addition, an analysis of the deviation
presence of these materials (by comparison with of the actual data points from the regression line
the assay result obtained on unspiked samples). may also be helpful for evaluating linearity.
For the impurity test, the discrimination may be Some analytical procedures, such as immunoas-
established by spiking drug substance or drug says, do not demonstrate linearity after any trans-
product with appropriate levels of impurities and formation. In this case, the analytical response
demonstrating the separation of these impurities should be described by an appropriate function of
individually and/or from other components in the the concentration (amountly or empiraclly) of the
sample matrix. concentration (or content) of an analyte in a sam-
ple.
(2) Impurities are not available For the establishment of linearity, a minimum of
If impurity or degradation products standards 5 concentrations is recommended. Other approach-
are unavailable, specificity may be demonstrated es should be justified.
by comparing the test results of samples contain-
ing impurities or degradation products to a se- 3. Range
cond well-characterized procedure. In this ap- The specified range is normally derived from lin-
proach, analytical procedure having known vali- earity studies and depends on the intended applica-
dation parameter includes pharmacopoeial meth- tion of the procedure. It is established by confirm-
od, other validated analytical procedure e.g.: ing that the analytical procedure provides an ac-
pharmacopoeial method of other validated analyt- ceptable degree of linearity, accuracy and precision
ical procedure (independent procedure). As ap- when applied to samples containing amounts of
propriate, this should include samples stored un- analyte within or at the extremes of the specified
der relevant stress conditions light, heat, humidity, range of the analytical procedure.
acid/base hydrolysis and oxidation. The following minimum specified ranges should
be considered:
- For the assay, the two results should be com-
pared. A. Assay of a drug substance or a drug product
- For the impurity tests, the impurity profiles Normally from 80 to 120 percent of the test con-
should be compared. centration.
B. Content uniformity
Peak purity tests may be useful to show that the Covering a minimum of 70 to 130 percent of the
analyte chromatographic peak is not attributable test concentration, unless a wider more appropriate
to more than one component (e.g., diode array, range, based on the nature of the dosage form such
mass spectrometry). as metered dose inhalers, is justified
C. Dissolution testing
2. Linearity ±20 % over the specified range indicated in disso-
A linear relationship should be evaluated across lution specification of the Specifications and Test
1860 Monographs, Part II
procedure, for instance, for inclusion of procedures A specific calibration curve should be stud-
in pharmacopoeias. These data are not part of the ied using samples containing an analyte in the
marketing authorization dossier. range of DL. The residual standard deviation
of a regression line or the standard deviation
D. Recommended Data of y-intercepts of regression lines may be
The standard deviation, relative standard deviation used as the standard deviation.
(coefficient of variation) and confidence interval
should be reported for each type of precision in- D. Recommended Data
vestigated. The detection limit and the method used for de-
termining the detection limit should be presented.
6. Detection limit If DL is determined based on visual evaluation or
Several approaches for determining the detection based on signal-to-noise ratio, the presentation of
limit are possible, depending on whether the pro- the relevant chromatograms is considered accepta-
cedure is a non-instrumental or instrumental. Ap- ble for justification.
proaches other than those listed below may be ac- In cases where an estimated value for the detec-
ceptable. tion limit is obtained by calculation or extrapola-
tion, this estimate may subsequently be validated
A. Based on Visual Evaluation by the independent analysis of a suitable number
Visual evaluation may be used for non- of samples known to be near or prepared at the de-
instrumental methods but may also be used with tection limit.
instrumental methods.
The detection limit is determined by the analysis 7. Quantification limit
of samples with known concentrations of analyte Several approaches for determining the quantifi-
and by establishing the minimum level at which cation limit are possible, depending on whether the
the analyte can be reliably detected. procedure is non-instrumental or instrumental.
Approaches other than those listed below may be
B. Based on Signal-to-Noise acceptable.
This approach can only be applied to analytical
procedures which exhibit baseline noise. Determi- A. Based on Visual Evaluation
nation of the signal-to-noise ratio is performed by Visual evaluation may be used for non-
comparing measured signals from samples with instrumental methods but may also be used with
known low concentrations of analyte with those of instrumental methods.
blank samples and establishing the minimum con- The quantification limit is generally determined
centration at which the analyte can be reliably de- by the analysis of samples with known concentra-
tected. A signal-to-noise ratio between 3 or 2:1 is tions of analyte and by establishing the minimum
generally considered acceptable for estimating the level at which the analyte can be quantified with
detection limit. acceptable accuracy and precision.
The slope S may be estimated from the calibra- C. Based on the Standard Deviation of the Re-
tion curve of the analyte. The estimate of σ sponse and the Slope
may be carried out in a variety of ways for ex- The quantification limit (QL) may be expressed
ample: as:
(1) Based on the Standard Deviation of the
Blank QL = 10 * σ / S
Measurement of the magnitude of analytical
background response is performed by analyz- where
ing an appropriate number of blank samples σ = the standard deviation of the response
and calculating the standard deviation of the- S = the slope of the calibration curve
se responses.
(2) Based on the Calibration Curve The slope S may be estimated from the calibra-
1862 Monographs, Part II
timated by determining the minimum pI difference The time required for completion of focusing in
(ΔpI), which is necessary to separate 2 neighboring thin-layer polyacrylamide gels is determined by plac-
bands: ing a colored protein (e.g. hemoglobin) at different
positions on the gel surface and by applying the elec-
D (dpH / dx) tric field: the steady state is reached when all applica-
R pI 3 tions give an identical band pattern. In some protocols
E (d / dpH ) the completion of the focusing is indicated by the time
elapsed after the sample application. The IEF gel can
D: Diffusion coefficient of the protein be used as an identity test when the migration pattern
dpH/dx: pH gradient on the gel is compared to a suitable standard prepara-
E: Intensity of the electric field, in volts per centi- tion and IEF calibration proteins, the IEF gel can be
meter used as a limit test when the density of a band on IEF
-dµ/dpH: Variation of the solute mobility with the is compared subjectively with the density of bands
pH in the region close to the pI appearing in a standard preparation, or it can be used
as a quantitative test when the density is measured
ASSAY using a densitometer or similar instrumentation to
Type of
Testing for - dissolu- determine the relative concentration of protein in the
analytical
impurities tion bands subject to validation.
procedure
(meas-
Identifi-
uremen Apparatus
cation
quan t only) An apparatus for IEF consists of :
characteris- - content
tific limit A controllable generator for constant potential, current
tics / po-
ation and power. Potentials of 2500 V have been used and
tency are considered optimal under a given set of operating
Accuracy - + - + conditions. Supply of up to 30 W of constant power is
Precision recommended,
Repeata- - + - + A rigid plastic IEF chamber that contains a cooled
bility - +(1) - +(1) plate, of suitable material, to support the gel,
Interm.Pr + + + + A plastic cover with platinum electrodes that are con-
ecision - -(3) + - nected to the gel by means of paper wicks of suitable
Specifici- - + - - width , length and thickness , impregnated with solu-
ty(2) - + - + tions of anodic and cathodic electrolytes .
Detection - + - +
Limit Isoelectric Focusing in Polyacrylamide Gels: De-
Quantifica- tailed Procedure
tion Limit The following method is a detailed description of
Linearity an IEF procedure in thick polyacrylamide slab gels,
Range which is used unless otherwise stated in the mono-
graph.
Since D and -dµ/dpH for a given protein cannot
be altered, the separation can be improved by using a Preparation of the Gels
narrower pH range and by increasing R, the intensity Mould The mould (see Figure) is composed of
of the electric field. Resolution between protein bands a glass plate (A) on which a polyester film (B) is
on an IEF gel prepared with carrier ampholytes can be placed to facilitate handling of the gel, one or more
quite good. Improvements in resolution may be spacers (C), a second glass plate (D) and clamps to
achieved by using immobilized pH gradients where hold the structure together .
the buffering species, which are analogous to carrier 7.5% Polyacrylamide gel Dissolve 29.1 g of
ampholytes, are copolymerized within the gel matrix. acrylamide and 0.9 g of N,N’-methylenebisacrylamide
Proteins exhibiting pIs differing by as little as 0.02 pH in 100 mL of water. To 2.5 volumes of this solution,
units may be resolved using a gel prepared with carrier add the mixture of ampholytes specified in the mono-
ampholytes while immobilized pH gradients can re- graph and dilute to 10 volumes with water. Mix care-
solve proteins differing by approximately 0.001 pH fully and degas the solution.
units.
Practical Aspects
Special attention must be paid to sample charac-
teristics and/or preparation. Having salt in the sample
can be problematic and it is best to prepare the sample,
if possible, in de-ionized water or 2% ampholytes,
using dialysis or gel filtration if necessary.
1864 Monographs, Part II
at opposite ends of the gel may not be identical. py. Optical microscopy is particularly useful for char-
A phenomenon known as ‘cathodic drift’, where acterizing particles that are not spherical. This method
the pH gradient decays over time, may occur if a gel is may also serve as a base for the calibration of faster
focused too long. Although not well understood, and more routine methods that may be developed.
electroendoosmosis and absorption of carbon dioxide Apparatus - Use a microscope that is stable and pro-
may be factors that lead to cathodic drift. Cathodic tected from vibration. The microscope magnification
drift is observed as focused protein migrating off the (product of the objective magnification, ocular magni-
cathode end of the gel. Immobilized pH gradients may fication, and additional magnifying components) must
be used to address this problem. be sufficient to allow adequate characterization of the
Efficient cooling (approximately 4 °C) of the bed smallest particles to be classified in the test specimen.
that the gel lies on during focusing is important. High The greatest numerical aperture of the objective
field strengths used during isoelectric focusing can should be sought for each magnification range. Polar-
lead to overheating and affect the quality of the fo- izing filters may be used in conjunction with suitable
cused gel. analyzers and retardation plates. Color filters of rela-
tively narrow spectral transmission should be used
Reagents and Solutions with achromatic objectives and are preferable with
Fixing solution for isoelectric focusing in poly- apochromats and are required for appropriate color
acrylamide gel Dissolve 35 g of 5-sulfosalicylic rendition in photomicrography. Condensers corrected
acid dihydrate and 100 g of trichloroacetic acid in for at least spherical aberration should be used in the
water to make 1000 mL. microscope substage and with the lamp. The numeri-
Coomassie staining TS Dissolve 125 mg of cal aperture of the substage condenser should match
coomassie brilliant blue R-250 in 100 mL of a mixture that of the objective under the condition of use, in the
of water, methanol and acetic acid (100) (5:4:1), and other words this is affected by the actual aperture of
filter. the condenser diaphragm and the presence of immer-
Destaining solution A mixture of water, metha- sion oils.
nol and acetic acid (100) (5:4:1). Adjustment - The precise alignment of all elements
of the optical system and proper focusing are essential.
The focusing of the elements should be done in ac-
cordance with the recommendations of the microscope
9. Particle Size manufacturer. Critical axial alignment is recommend-
Determination ed.
Illumination - A requirement for good illumination
Particle Size Determination is a method to deter- is a uniform and adjustable intensity of light over the
mine directly or indirectly morphological appearance, entire field of view; Kohler illumination is preferred.
shape, size and its distribution of powdered pharma- With colored particles, choose the color of the filters
ceutical drugs and excipients to examine their used so as to control the contrast and detail of the im-
micromeritic properties. Optical microscopy or analyt- age.
ical sieving method is used according to the measuring
purpose and the properties of the test specimen. Visual Characterization - The magnification and
numerical aperture should be sufficiently high to allow
Method 1. Optical Microscopy adequate resolution of the images of the particles to be
The optical microscopy is used to observe the mor- characterized. Determine the actual magnification
phological appearance and shape of individual particle using a calibrated stage micrometer to calibrate an
either directly with the naked eye or by using a micro- ocular micrometer. Errors can be minimized if the
scopic photograph, in order to measure the particle magnification is sufficient that the image of the parti-
size. The particle size distribution can also be deter- cle is at least 10 ocular divisions. Each objective must
mined by this method. It is also possible with this be calibrated separately. The ocular scale, the stage
method to measure the size of the individual particle micrometer scale and the ocular scale should be
even then different kinds of particles mingle if they aligned to calibrate. In this way, a precise determina-
are optically distinguishable. Data processing tech- tion of the distance between ocular stage divisions can
niques, such as image analysis, can be useful for de- be made.
termining the particle size distribution. When the particle size is measured, an ocular mi-
This method for particle characterization can gener- crometer is inserted at the position of the ocular dia-
ally be applied to particles not less than 1 μm. The phragm, and a calibrated stage micrometer is placed at
lower limit is imposed by the resolving power of the the center of the microscope stage and fixed in place.
microscope. The upper limit is less definite and is de- The ocular is attached to the lens barrel and adjusted
termined by the increased difficulty associated with to the focus point of the stage micrometer scale. Then,
the characterization of large particles. Various alterna- the distance between the scales of the two microme-
tive techniques are available for particle characteriza- ters is determined, and the sample size equivalent 1
tion, outside the applicable range of optical microsco- division of the ocular scale is calculated using the fol-
1866 Monographs, Part II
The nest of sieves is subjected to a standardized peri- imen of 10 to 25 g is available, smaller diameter test
od of agitation, and then the weight of material re- sieves conforming to the same mesh specifications
tained on each sieve is accurately determined. The test (table 1) may be substituted, but the endpoint must be
gives the weight percentage of powder in each sieve re-determined. The use of test samples having a small-
size range. er mass (e. g. down to 5 g) may be needed. For mate-
This sieving process for estimating the particle size rials with low apparent particle density, or for materi-
distribution of a single pharmaceutical powder is gen- als mainly comprising particles with a highly iso-
erally intended for use where at least 80 % of the par- diametrical shape, specimen weights below 5 g for a
ticles are larger than 75 m. The size parameter in- 200 mm screen may be necessary to avoid excessive
volved in determining particle size distribution by blocking of the sieve. During validation of a particular
analytical sieving is the length of the side of the min- sieve analysis method, it is expected that the problem
imum square aperture through which the particle will of sieve blocking will have been addressed.
pass. If the test material is prone to picking up or losing
significant amounts of water with varying humidity,
TEST SIEVES the test must be carried out in an appropriately con-
Unless otherwise specified in the monograph, use the trolled environment. Similarly, if the test material is
sieves listed in the Table 1. known to develop an electrostatic charge, careful ob-
Sieves are selected to cover the entire range of parti- servation must be made to ensure that such charging is
cle sizes present in the test specimen. A nest of sieves not influencing the analysis. An antistatic agent, such
having a 2 progression of the area of the sieve as colloidal silicon dioxide and/or aluminum oxide,
openings is recommended. The nest of sieves is as- may be added at a 0.5 percent (m/m) level to minimize
sembled with the coarsest screen at the top and the this effect. If both of the above effects cannot elimi-
finest at the bottom. Use micrometers or millimeters in nated, an alternative particle-sizing technique must be
denoting test sieve openings. [Note - Mesh numbers selected.
are provided in the table for conversion purposes only.] Agitation Methods - Several different sieve and
Test sieves are made from stainless steel or, less pref- powder agitation devices are commercially available,
erably, from brass or other suitable non-reactive wire. all of which may be used to perform sieve analyses.
However, the different methods of agitation may give
Calibration and recalibration of test sieves is in ac- different results for sieve analyses and endpoint de-
cordance with the most current edition of ISO 3310- terminations because of the different types and magni-
12). Sieves should be carefully examined for gross tude of the forces acting on the individual particles
distortions and fractures, especially at their screen under test. Methods using mechanical agitation or
frame joints, before use. Sieves may be calibrated op- electromagnetic agitation, and that can include either a
tically to estimate the average opening size, and open- vertical oscillation or a horizontal circular motion, or
ing variability, of the sieve mesh. Alternatively, for tapping or a combination of both tapping and horizon-
the valuation of the effective opening of test sieves in tal circular motion are available. Entrainment of the
the size range of 212 to 850 m, Standard Glass particles in an air stream may also be used. The results
Spheres are available. Unless otherwise specified in must indicate which agitation method was used and
the individual monograph, perform the sieve analysis the agitation parameters used (if they can be varied).
at controlled room temperature and at ambient relative End point Determination - The test sieving analy-
humidity. sis is complete when the weight on any of the test
Cleaning Test Sieves - Ideally, test sieves should sieves does not change by more than 5 % or 0.1 g (10 %
be cleaned using only an air jet or a liquid stream. If in the case of 76 mm sieves) of the previous weight on
some apertures remain blocked by test particles, care- that sieve. If less than 5 % of the total specimen
ful gentle brushing maybe used as a last resort. weight is present on a given sieve, the endpoint for
Test Specimen - If the test specimen weight is not that sieve is increased to a weight change of not more
given in the monograph for a particular material, use a than 20 % of the previous weight on that sieve. If
test specimen having a weight between 25 and 100 g, more than 50 % of the total specimen weight is found
depending on the bulk density of the material, and test on any one sieve, unless this is indicated in the mono-
sieves having a 200 mm diameter. For 76 mm sieves graph, the test should be repeated, but with the addi-
the amount of material that can be accommodated is tion to the sieve nest of a more coarse sieve intermedi-
approximately 1/7th that which can be accommodated ate between that carrying the excessive weight and the
on a 200 mm sieve. Determine the most appropriate next coarsest sieve in the original nest, i.e., addition of
weight for a given material by test sieving accurately the ISO series sieve omitted from the nest of sieves.
weighed specimens of different weights, such as 25,
50, and 100 g, for the same time period on a mechani- SIEVING METHODS
cal shaker. [Note - If the test results are similar for the
25 g and 50 g specimens, but the 100 g specimen Mechanical agitation
shows a lower percentage through the finest sieve, the Dry Sieving Method - Tare each test sieve to the
100-g specimen size is too large.] Where only a spec- nearest 0.1 g. Place an accurately weighed quantity of
KP X 1869
test specimen on the top (coarsest) sieve, and replace In the sonic sifting method a nest of sieves is used,
the lid. Agitate the nest of sieves for 5 minutes. Then and the test specimen is carried in a vertically oscillat-
carefully remove each from the nest without loss of ing column of air that lifts the specimen and then car-
material. If there is some fine powder on the down ries it back against the mesh openings at a given num-
surface of each sieve, take if off by the brush gently, ber of pulses per minute. It may be necessary to lower
and combine it with the sieve fraction retained on each the sample amount to 5 g, when sonic shifting is em-
next down sieve. Reweigh each sieve, and determine ployed.
the weight of material on each sieve. Determine the The air jet sieving and sonic sieving methods may be
weight of material in the collecting pan in a similar useful for powders or granules when mechanical siev-
manner. Reassemble the nest of sieves, and agitate for ing techniques are incapable of giving a meaningful
5 minutes. Remove and weigh each sieve as previous- analysis.
ly described. Repeat these steps until the endpoint These methods are highly dependent upon proper
criteria are met (see Endpoint Determination under dispersion of the powder in the air current. This re-
Test Sieves). Upon completion of the analysis, recon- quirement may be hard to achieve if the method is
cile the weights of material. Total losses must not ex- used at the lower end of the sieving range (i. e., below
ceed 5 % of the weight of the original test specimen. 75m), when the particles tend to be more cohesive,
Repeat the analysis with a fresh specimen, but using and especially if there is any tendency for the material
a single sieving time equal to that of the combined to develop an electrostatic charge. For the above rea-
times used above. Confirm that this sieving time con- sons endpoint determination is particularly critical,
forms to the requirements for endpoint determination. and it is very important to confirm that the oversize
When this endpoint has been validated for a specific material comprises single particles and is not com-
material, then a single fixed time of sieving may be posed of aggregates.
used for future analyses, providing the particle size
distribution falls within normal variation. INTERPRETATION
If there is evidence that the particles retained on any The raw data must include the weight of test speci-
sieve are aggregates rather than single particles, the men, the total sieving time, and the precise sieving
use of mechanical dry sieving is unlikely to give good methodology and the set values for any variable pa-
reproducibility, a different particle size analysis meth- rameters, in addition to the weights retained on the
od should be used. individual sieves and in the pan. It may be convenient
to convert the raw data into a cumulative weight dis-
Air Entrainment Methods tribution in terms of a cumulative weight undersize,
Air jet and Sonic Shifter Sieving - Different types the range of sieves used should include a sieve
of commercial equipment that use a moving air cur- through which all the material passes. If there is evi-
rent are available for sieving. A system that uses a dence on any of the test sieves that the material re-
single sieve at a time is referred to as air jet sieving. It maining on it is composed of aggregates formed dur-
uses the same general sieving methodology as that ing the sieving process, the analysis is invalid.
described under the Dry Sieving Method, but with a
1)
standardized air jet replacing the normal agitation Additional information on particle size measurement,
mechanism. It requires sequential analyses on individ- sample size, and data analysis is available, for exam-
ual sieves starting with the finest sieve to obtain a par- ple, in ISO 9276.
2)
ticle size distribution. Air jet sieving often includes the International Organization for Standardization (ISO)
use of finer test sieves than used in ordinary dry siev- Specification ISO 3310-1; Test sieves-Technical re-
ing. This technique is more suitable where only over- quirements and testing
size or undersize fractions are needed.
4.50 mm
4.00 mm 4.00 mm 4.00 mm 5 4000 4000 4.7
3.55 mm
3.35 mm 6 5.5
3.15 mm
2.80 mm 2.80 mm 2.80 mm 7 2800 2800 6.5
2.50 mm
2.36 mm 8 7.5
2.24 mm
2.00 mm 2.00 mm 2.00 mm 10 2000 2000 8.6
1.80 mm
1.70 mm 12 10
1.60 mm
1.40 mm 1.40 mm 1.40 mm 14 1400 1400 12
1.25 mm
1.18 mm 16 14
1.12 mm
1.00 mm 1.00 mm 1.00 mm 18 1000 1000 16
900 m
850 m 20 18
800 m
710 m 710 m 710 m 25 710 710 22
630 m
600 m 30 26
560 m
500 m 500 m 500 m 35 500 500 30
450 m
425 m 40 36
400 m
355 m 355 m 355 m 45 355 355 42
315 m
300 m 50 50
280 m
250 m 250 m 250 m 60 250 250 60
224 m
212 m 70 70
200 m
180 m 180 m 180 m 80 180 180 83
160 m
150 m 100 100
140 m
125 m 125 m 125 m 120 125 125 119
112 m
106 m 140 140
100 m
90 m 90 m 90 m 170 90 90 166
80 m
75 m 200 200
71 m
63 m 63 m 63 m 230 63 63 235
56 m
53 m 270 282
50 m
45 m 45 m 45 m 325 45 45 330
40 m
38 m 38 391
KP X 1871
10. Powder Particle Density out between 15 and 30 C, and the temperature must
not vary by more than 2 C during the course of
Determination measurement.
Firstly, weigh the mass of the test cell and record it.
The Powder Particle Density Determination is a After weighing out the amount of the sample as de-
method to determine particle density of powdered scribed in the individual monograph and placing it in
pharmaceutical drugs or raw materials of drugs, and the test cell, seal the cell in the pycnometer. Secondly,
the gas displacement pycnometer is generally used. introduce the measurement gas (helium) into the test
The powder density by this method is determined with cell, in order to remove volatile contaminants in the
an assumption that the volume of the gas displaced by powder. If necessary, keep the sample powder under
the powder in a closed system is equal to the volume reduced pressure to remove the volatile contaminants
of the powder. The bulk density at loose packing or in advance and use it as the test sample for measure-
the tapped density at tapping express the apparent ment.
densities of the powder, since interparticular void vol- Open the valve which connects the reference cell
ume of the powder is considered to be a part of the with the test cell, confirm with manometer that the
volume of the powder. On the contrary, the pressure inside the system is stable, and then read the
pycnometric particle density expresses the powder system reference pressure (Pr). Secondly, close the
density nearly equal to the crystal density, since the valve that connects to the two cells, and introduce the
volume of the powder, that is deducted with void vol- measurement gas into the test cell to achieve positive
ume of open pores accessible to gas, is counted. pressure. Confirm with the manometer that the pres-
Powder particle density is expressed in mass per sure inside is stable, and then read the initial pressure
unit volume (kg/m3), and generally expressed in g/cm3. (Pi). Open the valve to connect test cell with the refer-
ence cell. After confirming that the indicator of the
Apparatus manometer is stable, read the final pressure (Pf). Cal-
The schematic diagram of particle density appa- culate the sample volume (Vs) with the following
ratus for gas displacement pycnometric measurement equation.
is shown in Figure. The apparatus consists of a test
cell in which the sample is placed, a reference cell and Vr
Vs Vc
a manometer. Pi Pr
1
Generally, helium is used as the measurement gas. Pf Pr
The apparatus has to be equipped with a system capa-
ble of pressuring the test cell to the defined pressure
Vr: Reference cell volume (cm3)
through the manometer.
Vc: Test cell volume (cm3)
Calibration of apparatus The volumes of the
Vs: Sample volume (cm3)
test cell (Vc) and the reference cell (Vr) must be accu-
Pi: Initial pressure (kPa)
rately determined to the nearest 0.001 cm3. In order to
Pf: Final pressure (kPa)
assure accuracy of the results of volume obtained,
Pr: Reference pressure (kPa)
calibration of the apparatus is carried out as follows
using a calibration ball of known volume for particle
Repeat the measurement sequence for the same
density measurement. The final pressure (Pf) are de-
powder sample until consecutive measurements of the
termined for the initial empty test cell followed by the
sample volume agree to within 0.5 %, and calculate
test cell placed with the calibration ball for particle
the mean of sample volumes (Vs). Finally, unload the
density measurement in accordance with the proce-
test cell, weigh the mass of test cell, and calculate the
dures, and Vc and Vr are calculated using the equation
final sample mass by deducing the empty cell mass
described in the section of Procedure. Calculation can
from the test cell mass. The powder particle density ρ
be made taking into account that the sample volume
is calculated by the following equation.
(Vs) is zero in the first run.
m
Vr: Reference cell volume (cm3)
Vc: Test cell volume (cm3) Vs
Vs: Sample volume (cm3)
M : Manometer (cm3) ρ: Powder particle density (g/ cm3)
m: Final sample mass (g)
Figure. Schematic diagram of a gas Vs: Sample volume (cm3)
displacement pycnometer used to
determine powder particle density
Procedure
The measurement of the particle density is carried
1872 Monographs, Part II
sion into the container aseptically or the volume of the miscibility between the liquid medium and semisolid
product in each container is too small to be tested, ointments or oils in which test microorganisms were
transfer aseptically a sufficient volume of the product inoculated. Sometimes, these agents also serve to inac-
into each of alternative sterile containers, and mix the tivate or neutralize many of the most commonly used
inoculum. When the product is not sterile, incubate preservatives.
additional containers containing the uninoculated
product as controls and calculate their viable cell 4. Interpretation
counts (the viable counts of bacteria and those of Interpret the preservative efficacy of the product
yeasts and moulds). A sterile syringe, spatula or glass according to Table 1. When the results described in
rod may be used to mix the cell suspension uniformly Table 1 are obtained, the product examined is consid-
in the product. The volume of the suspension mixed in ered to be effectively preserved. There is a strong pos-
the product must not exceed 1/100 of the volume of sibility of massive microbial contamination having
the product. Generally, the cell suspension is inoculat- occurred when microorganisms other than the inocu-
ed and mixed so that the concentration of viable cells lated ones are found in the sterile product to be exam-
is 105 to 106 cells per mL or per gram of the product. ined, and caution is required in the test procedures
Incubate these inoculated containers at 20 C to 25 C and/or the control of the manufacturing process of the
with protection from light, and calculate the viable cell product. When the contamination level in a non-sterile
count of 1 mL or 1 g of the product taken at 0, 14 and product to be examined exceeds the microbial enu-
28 days subsequent to inoculation. Record any marked meration limit specified in “Microbial
changes (e.g., changes in color or the development of
a bad odor) when observed in the mixed samples dur- Table 1. Interpretation criteria by product category
ing this time. Such changes should be considered
when assessing the preservative efficacy of the prod- Interpretation criteria
Prod Microor-
uct concerned. Express sequential changes in the via- After 14 days After 28 days
cate ganisms
ble counts as percentages, with the count at the start of
the test taken as 100. Titration of the viable cell counts 0.1 % of in- Same or less
is based, in principle, on the Pour Plate Methods in Cat- Bacteria oculum count than level after
“Microbial Limit Test”. In this case, confirm whether egory or less 14 days
any antimicrobial substance is present in the test spec- IA Yeasts Same or less Same or less
imen. If a confirmed antimicrobial substance needs to /moulds than inoculum than inoculum
be eliminated, incorporate an effective inactivator of 1count
% of inocu- Same or less
count
the substance in the buffer solution or liquid medium Cat- Bacteria lum count or than level after
to be used for dilution of the test specimen, as well as egory less 14 days
in the agar plate count medium. However, it is neces- IB Yeasts Same or less Same or less
sary to confirm that the inactivator has no effect on the /moulds than inoculum than inoculum
growth of the microorganisms. When the occurrence count
10 % of inocu- count
Same or less
of the preservative or the product itself affects titration Bacteria lum count or than level after
Cat-
of the viable cell count and there is no suitable less 14 days
egory
inactivator available, calculate the viable cell counts
IC Yeasts Same or less Same or less
by the Membrane Filtration method in “Microbial
/moulds than inoculum than inoculum
Limit Test”.
Bacteria countor less
Same count
Same or less
3.2. Category II products
The procedures are the same as those described for Cat- than inoculum than inoculum
Category I products. However, special procedures and egory count count
considerations are required for both uniform disper- II Yeasts Same or less Same or less
sion of the test microorganism in the product and titra- /moulds than inoculum than inoculum
tion of viable cell counts in the samples. countPharmaceutical
Attributes of Non-sterile count
Products” in
For semisolid ointment bases, heat the sample to General Information, caution is also required in the
45 C to 50 C until it becomes oily, add the cell sus- test procedures and/or the control of the manufactur-
pension and disperse the inoculum uniformly with a ing process of the product.
sterile glass rod or spatula. Surfactants may also be
added to achieve uniform dispersion, but it is neces- 5. Culture media
sary to confirm that the surfactant added has no effect Culture media and buffer solutions used for Pre-
on survival or growth of the test microorganisms and servatives-Effectiveness Tests are described below.
that it does not potentiate the preservative efficacy of Other media may be used if they have similar nutritive
the product. For titration of the viable cell count, a ingredients and selective and growth-promoting prop-
surfactant or an emulsifier may be added to disperse erties for the microorganisms to be tested.
the product uniformly in the buffer solution or liquid Soybean Casein Digest Agar Medium
medium. For instance, sorbitan monooleate, Casein peptone 15 g
polysorbate 80 or lecithin may be added to improve Soybean peptone 5.0 g
1874 Monographs, Part II
adsorption (Method II). However, under certain cir- samples, other outgassing methods such as the desorp-
cumstances described below, it may be acceptable to tion-adsorption cycling method may be employed.
determine the specific surface area of a powder from a The standard technique is the adsorption of nitro-
single value of Va measured at a single value of P/P0 gen at liquid nitrogen temperature.
such as 0.30, using the following equation for calcu- For powders of low specific area (< 0.2 m2/g) the
lating Vm: proportion adsorbed is low. In such cases the use of
krypton at liquid nitrogen temperature is preferred
P because the low vapor pressure exerted by this gas
Vm Va (1 ) (3)
greatly reduces error.
P0
All gases used must be free from moisture.
Accurately weigh a quantity of the test powder
The single-point method may be employed di- such that the total surface of the sample is at least 1 m2
rectly for a series of powder samples of a given mate- when the adsorbate is nitrogen and 0.5 m2 when the
rial for which the material constant C is much greater adsorbate is krypton. Lower quantities of sample may
than 1. Close similarity between the single-point val- be used after appropriate validation.
ues and multiple-point values suggests that 1/C ap- Adsorption of gas should be measured either by
proaches zero. The error on Vm can be reduced by us- Method I or Method II.
ing the multiple-point method to evaluate C for one of
the samples of the series on which the material con- Method I: the dynamic flow method
stant C is expected to be large. Then Vm is calculated In the dynamic flow method (see Fig. 1), the rec-
from the single value of Va measured at a single value ommended adsorbate gas is dry nitrogen or krypton,
of P/P0 by the equation: while helium is employed as a diluents gas, which is
not adsorbed under the recommended conditions. A
P 1 C 1 P minimum of 3 mixtures of the appropriate adsorbate
Vm Va 0 1 (4)
P C C P0 with helium are required within the P/P0 range 0.05 to
0.30.
Sample preparation
Before the specific surface area of the sample can
be determined, it is necessary to remove gases that
may have become physically adsorbed onto the sur-
face during storage and handling. If outgassing is not
achieved, the specific surface area may be reduced or
may be variable because an intermediate area of the
surface is covered with molecules of the previously
adsorbed gases or vapors. The outgassing conditions
are critical for obtaining the required precision and
accuracy of specific surface area measurements on
pharmaceuticals because of the sensitivity of the sur-
face of the materials. The outgassing conditions must
be demonstrated to yield reproducible BET plots, a
A: Flow control valve
constant weight of test powder, and no detectable
B: Differential flow controller
physical or chemical changes in the test powder.
C: On-off valve
The outgassing conditions defined by the temper-
D: Gas inlet
ature, pressure and time should be so chosen that the
E: O ring seals
original surface of the solid is reproduced as closely as
F: Cold trap
possible.
G: Thermal equilibration tube
Outgassing of many substances is often achieved
H: Detector
by applying a vacuum, by purging the sample in a
I: Digital display
flowing stream of a non-reactive, dry gas, or by apply-
J: Calibration septum
ing a desorption-adsorption cycling method. In either
K: Sample cell
case, elevated temperatures are sometimes applied to
L: Self seals quick connection
increase the rate at which the contaminants leave the
M: Short path ballast
surface. Caution should be exercised when outgassing
N: Detector
powder samples using elevated temperatures to avoid
O: Path selection valve
affecting the nature of the surface and the integrity of
P: Long pass ballast
the sample.
Q: Flow meter
If heating is employed, the recommended tem-
R: Outgassing station
perature and time of outgassing are as low as possible
S: Diffusion baffle
to achieve reproducible measurement of specific sur-
T: Vent
face area in an acceptable time. For heating sensitive
1876 Monographs, Part II
Fig. 1. Schematic diagram of the dynamic flow Admit a small amount of dry nitrogen into the sample
method apparatus tube to prevent contamination of the clean surface of
sample, remove the sample tube, insert the stopper,
The gas detector-integrator should provide a sig- and weigh it. Calculate the weight of the sample. At-
nal that is approximately proportional to the volume of tach the sample tube to the volumetric apparatus. Cau-
the gas passing through it under defined conditions of tiously evacuate the sample down to the specified
temperature and pressure. For this purpose, a thermal pressure (e.g. between 2 Pa and 10 Pa).
conductivity detector with an electronic integrator is If the principle of operation of the instrument requires
one among various suitable types. A minimum of 3 the determination of the dead volume in the sample
data points within the recommended range of 0.05 to tube, this procedure is carried out at this point, fol-
0.30 for P/P0 is to be determined. lowed by evacuation of the sample. Raise a Dewar
A known mixture of the gases, usually nitrogen vessel containing liquid nitrogen up to a defined point
and helium, is passed through a thermal conductivity on the sample cell. Admit a sufficient volume of
cell and then to a recording potentiometer. adsorbate gas to give the lowest desired relative pres-
Immerse the sample cell in liquid nitrogen, then sure. Measure the volume adsorbed, Va. For multipoint
the sample adsorbs nitrogen from the mobile phase. measurements, repeat the measurement of Va at suc-
This unbalances the thermal conductivity cell, and a cessively higher P/P0 values. When nitrogen is used as
pulse is generated on a recorder chart. the adsorbate gas, P/P0 values of 0.10, 0.20, and 0.30
Remove from the coolant; this gives a desorption are often suitable.
peak equal in area and in the opposite direction to the
adsorption peak. Reference materials
Since this is better defined than the adsorption Periodically verify the functioning of the apparatus
peak, it is the one used for the determination. using appropriate reference materials of known sur-
To effect the calibration, inject a known quantity face area, such as α-alumina for specific surface area
of adsorbate into the system, sufficient to give a peak determination, which should have a specific surface
of similar magnitude to the desorption peak and obtain area similar to that the sample to be examined.
the proportion of gas volume per unit peak area.
Method II: the volumetric method
In the volumetric method (see Figure 2), the rec-
ommended adsorbate gas is nitrogen. And it is admit-
13. Terminal Sterilization and
ted into the evacuated space above the previously out- Sterilization Indicators
gassed powder sample to give a defined equilibrium
pressure, P, of the gas. The use of a diluent gas, such Sterilization is a process whereby the killing or
as helium, is therefore unnecessary, although helium removal of all forms of viable microorganisms in sub-
may be employed for other purposes, such as to meas- stances is accomplished. It is achieved by terminal
ure the dead volume. sterilization or a filtration method. For substances to
which terminal sterilization can be applied, an appro-
priate sterilization method should be selected in ac-
cordance with the properties of the product, including
the packaging, after full consideration of the ad-
vantages and disadvantages of each sterilization meth-
od, from the heat method, irradiation method and gas
method. After installation of the sterilizer (including
design and development of the sterilization process),
validation is required to confirm that the sterilization
process is properly performing its designed function,
under conditions of loading and unloading of the
product, on the basis of sufficient scientific evidence.
After the process has been validated and the steriliza-
tion of the product commenced, the process must be
A: Vacuum gauge controlled correctly, and qualification tests of the
B: Nitrogen reservoir equipment and procedures must be performed regular-
C: Helium reservoir ly. The bioburden per product, prior to terminal sterili-
D: Vapor pressure manometer zation, must be evaluated periodically or on the basis
E: Vacuum and air of batches. Refer to the ISO standard (ISO 11737-1)
F: Cold traps and vacuum pumps relevant to bioburden estimation. For a substance to
which terminal sterilization can be applied, generally
Fig. 2. Schematic diagram of the volumetric method use sterilization conditions such that a sterility assur-
apparatus ance level of less than 10-6 can be obtained. The prop-
KP X 1877
erty of the sterilization should be judged by employing Therefore, in routine sterilization process control,
an appropriate sterilization process control, with the it is required to monitor continuously the tem-
use of a suitable sterilization indicator, and if neces- perature and exposure time, and they should be
sary, based on the result of the sterility test. included in the specifications of the sterilizer.
The filtration procedure is used for the sterilization of 2.2 Irradiation method
a liquid product, to which terminal sterilization cannot Microorganisms are directly killed by ionizing radia-
be applied. Concerning the disinfection and/or sterili- tion, or by the heat generated by microwave radiation.
zation necessary for processing equipment and areas (i) Radiation method
of pharmaceutical products, and performing microbio- Ionizing radiations which may be used are gam-
logical tests specified in the monographs, see Disin- ma () rays emitted from a radioisotope such as 60C,
fection and Sterilization Methods. an electron beam and bremsstrahlung (X rays) gener-
ated from an electron accelerator. Although any pro-
1. Definitions cedure can be applied to thermally unstable products
The definitions of the terms used in this text are with no radioactivity residue, it is necessary to consid-
as follows. er the possibility of material degradation. Although a
Terminal sterilization: A process whereby a prod- 25 kGy dose is traditionally used as a sterilization
uct is sterilized in its final container or packaging, and dose, there are some ways to calculate the dose as
which permits the measurement and evaluation of follows: the bioburden of the substance to be sterilized
quantifiable microbial lethality. is measured and the sterilization dose is calculated
Product: A generic term used to describe raw ma- based on the mean bioburden and the standard re-
terials, intermediate products, and finished products, sistance distribution (Method I in ISO 11137), the
to be sterilized. dose is calculated from the fraction positive infor-
Bioburden: Numbers and types of viable micro- mation from a sterility test in which representative
organisms in a product to be sterilized. product samples are exposed to a substerilizing dose
Sterility assurance level (SAL): Probability of a (Method 2 in ISO 11137), or the dose is calculated
viable microorganism being present in a product unit based on the bioburden and D value of the most re-
after exposure to the proper sterilization process, ex- sistant microorganisms (Log method) (see 5-3). In the
pressed as 10-n. case of the radiation sterilization procedure, factors
Integrity test: A non-destructive test which is which may affect the sterilization include dose (ab-
used to predict the functional performance of a filter sorbed dose) and exposure time. Therefore, in ray
instead of the microorganism challenge test. sterilization process control, it is required to determine
D value: The value which shows the exposure the dose (the absorbed dose) at appropriate intervals
time (decimal reduction time) or absorbed dose (deci- and to monitor continuously the exposure time in
mal reduction dose) required to cause a 1-logarithm or terms of the operation parameters (the conveyer speed
90 % reduction in the population of test microorgan- and the cycle time). The dose control mechanism
isms under stated exposure conditions. should be included in the specifications of the steriliz-
Sterilization indicator: Indicators used to monitor er. In the case of electron beam or bremsstrahlung
the sterilization process, as an index of sterility, in- irradiation, it is required to monitor the acceleration
cluding biological indicators (BI), chemical indicators voltage, the beam current and beam scanning width
(CI), dosimeters and the like. besides the above-mentioned items.
(ii) Microwave method
2. Sterilization Microorganisms are killed by the heat generated
2.1 Heat method by microwave radiation, usually at the frequency of
In the heat method, microorganisms are killed by heat- 2450 50 MHz. This method is applied to liquids or
ing. water-rich products in sealed containers. Since a glass
(i) Moist heat method or plastic container may be destroyed or deformed due
Microorganisms are killed in saturated steam un- to the rise of the inner pressure, the containers must be
der pressure. In this method, factors which may certified to be able to withstand the heat and the inner
affect the sterilization include temperature, steam pressure generated during microwave sterilization.
pressure and exposure time. Therefore, in routine Leakage of electromagnetic radiation must be at a
sterilization process control, it is required to sufficiently low level to cause no harm to humans and
monitor continuously the temperature, steam no interference with radio communications and the
pressure and exposure time, and they should be like. In this method, factors which may affect the steri-
included in the specifications of the sterilizer. lization include temperature, processing time and mi-
(ii) Dry heat method crowave output power. Therefore, in routine steriliza-
Microorganisms are killed in dry heated air. This tion process control, it is required to monitor continu-
method is usually conducted in a batch-type dry ously the temperature, time and the microwave output
heat sterilizer or a funnel-type dry heat sterilizer. power, and they should be included in the specifica-
In this method, factors which may affect the ster- tions of the sterilizer.
ilization include temperature and exposure time. 2.3 Gas Method
1878 Monographs, Part II
Ethylene oxide (EO) is widely used as a steriliza- Table 1. Typical examples of biological indicator
tion gas. Since EO gas has an explosive nature, a 10 – Representative micro-
Sterilization Strain name
30 % mixture with carbon dioxide is commonly used. organisms*
Also, as EO gas is a strong alkylating agent, it cannot ATCC 7953,
be applied to the products which are likely to react NBRC 13737,
with or absorb it. Furthermore, because EO gas is tox- JCM 9488,
ic, the residual concentration of EO gas and other sec- Moist heat Geobacillus KCTC 2107
ondary generated toxic gases in products sterilized method stearothermophilus ATCC 12980,
with EO gas must be reduced to less than the safe lev- NBRC 12550,
els thereof by means of aeration and the like before JCM 2501,
the product is shipped. In this method, factors which KCTC 1752
may affect the sterilization include temperature, gas ATCC 9372,
concentration (pressure), humidity and exposure time. Dry heat
Bacillus atrophaeus NBRC 13721
Therefore, in routine sterilization process control, it is method
KCTC 1022
required to monitor continuously the temperature, gas
ATCC 9372,
concentration (pressure), humidity and exposure time,
Gas method Bacillus atrophaeus NBRC 13721
and they should be included in the specifications of
KCTC 1022
the sterilizer.
* In addition to these microorganisms, other microor-
3. Filtration method ganisms with the greatest resistance to the sterilization
Microorganisms are removed by using a steriliz- procedure concerned, found in the bioburden, can be
ing filter made of an appropriate material. However, used as the biological indicator.
this method is not intended for microorganisms small-
er than bacteria, Generally, a sterilizing filter chal- 4.1.1 D value of BI
lenged with more than 107 microorganisms of a strain Methods for determination of the D value include
Brevundimonas diminuta (ATCC 1914, NBRC 14213, the survival curve method and the fraction negative
JCM 2428), cultured under the appropriate conditions, method (Stumbo, Murphy & Cochran method, Lim-
per square centimeter of effective filter area should ited Spearman-Karber method and the like). In using
provide a sterile effluent. In this method, factors marketed BIs, it is usually unnecessary to determine
which may affect the sterilization include pressure, the D value before use if the D value indicated on the
flow rate, filter unit characteristics and the like. In label has been determined by a standardized biological
routine filtration process control, it is required to per- indicator evaluation resistometer (BIER) under strictly
form integrity tests of the sterilizing filter after each prescribed conditions in accordance with ISO 11138-1.
filtration process (also prior to the filtration process, if It is acceptable that the D value indicated on the label
necessary). shows a scattering of not more than 30 seconds.
4.1.2 Setting up procedure of BI
4. Sterilization Indicators (i) In the case of dry materials
4.1 Biological Indicator (BI) A dry type BI is placed at predetermined cold
A BI is prepared from specific microorganisms spots in the product to be sterilized or a suitable prod-
resistant to the specified sterilization process and is uct showing an equivalent effect in the sterilization.
used to develop and/or validate a sterilization process. The BIs are usually primary packaged in the same way
The dry type BI is classified into two kinds. In one, as the product, including a secondary packaging, if
bacterial spores are added to a carrier such as filter applicable.
paper, glass or plastic and then the carrier are dried (ii) In the case of wet materials
and packaged. In the other, bacterial spore are added Spores are suspended as the BI in the same solu-
to representative units of the product to be sterilized or tion as the product or in an appropriate similar solu-
to simulated products. Packaging materials of the BI tion, and should be placed at cold spots in the sterilizer.
should show good heat penetration in dry heat sterili- 4.1.3 Culture conditions of BI
zation and good gas or steam penetration in ethylene Soybean casein digest medium is generally used.
oxide and moist heat sterilizations. It should be con- General culture conditions are at 55 ~ 60 ℃ for 7 days
firmed that any carrier does not affect the D value of in the case of G. stearothermophilus and at 30 ~ 35 ℃
the spores. In the case of a liquid product, the spores for 7 days in the case of B. atrophaeus.
may be suspended in the same solution as the product
or in a solution showing an equivalent effect in the 4.2 Chemical indicator (CI)
sterilization of biological indicator. However, when CI is an indicator which shows a color change of
the spores are suspended in liquid, it is necessary to a substance applied to a paper slip, etc. as a result of
ensure that the resistance characteristics of the spores physical and/or chemical change due to exposure to
are not affected due to germination. Typical examples heat, gas or radiation. The CI can be classified into
of biological indicator are shown in Table 1. three types. The first is employed to identify whether
or not sterilization has already been implemented, the
KP X 1879
second is employed to control the sterilization process extensive bioburden estimation is considered as the
(for example, its color changes after sterilization for a maximum bioburden count, and the sterilization time
sufficient time), and the third is the Bowie & Dick (or radiation dose) is calculated with the bioburden
type used to evaluate the effectiveness of air removal count based on an objective sterility assurance level.
during the pre-vacuum phase of the pre-vacuum steri- When this procedure is used, it is required to deter-
lization cycle. mine the resistance of the bioburden to the steriliza-
4.3 Dosimeter tion as well as the bioburden count in the product be-
In the radiation (-ray) method, the sterilization ing sterilized. If a more resistant microorganism than
effect depends on the absorbed radiation dose, so the the BI spore is found in the bioburden estimation, it
sterilization process control is mainly performed by should be used as the BI.
measuring the dose. A dosimeter is installed at a posi-
tion corresponding to the minimum dose region of an N0
exposed container or a position where the dose is in a Sterilization time(or radiation dose) D log( )
N
known relation to that in the above region. Measure-
ment should be done for each radiation batch. If there D : D value of the BI
are many containers in the same batch, dosimeters N: Sterility assurance level
should be employed so that more than one dosimeter N0: Maximum bioburden count in the product
is always installed at the effective radiation section of 5.4 Absolute bioburden method
the irradiation chamber. It should be noted that dosim- The sterilization conditions are determined by
eters may be affected by environmental conditions employing the D value of the most resistant microor-
(temperature, humidity, ultraviolet light, time to read- ganism found in the product or environment by the
ing, etc.) before and during irradiation. Practical do- resistant estimation and being based on the bioburden
simeters for –ray and bremsstrahlung sterilization count in the product. Generally, a count of mean
include the dyed polymethylmethacrylate dosimeter, bioburden added three times of its standard deviation
clear polymethymethacrylate dosimeter, ceric-cerous obtained through the extensive bioburden estimation is
sulfate dosimeter, alanine-EPR dosimeter and the like. employed as the bioburden count. When this proce-
A dosimeter for gamma radiation cannot generally be dure is used, it is required to make frequent counting
used for sterilization process control by the electron and resistance determination of microorganisms in
beam energy less than 3 MeV. Dosimeters for electron daily bioburden estimation
beam sterilization include the cellulose acetate dosim-
eter, radiochromic film dosimeter and the like. A prac-
tical dosimeter must be calibrated against an appropri-
ate national or international standard dosimetry system.