Professional Documents
Culture Documents
Pharmaceuticals
Determination of Nitrosamine
Impurities Using the Ultivo Triple
Quadrupole LC/MS
Detection of regulated genotoxic impurities from the
drug manufacturing process
Authors Abstract
Chander Mani and
Saikat Banerjee
Recent findings show that some Angiotensin II receptor blocker (ARB) drug
Agilent Technologies, Inc.
products contain carcinogenic nitrosamine impurities.1 As a result, many of these
products were recalled, demonstrating a need for analytical methods capable of
detecting problematic nitrosamine impurities above target concentration levels.
This Application Note describes how the sensitive Agilent Ultivo LC/TQ based
method works, not only for the six nitrosamines listed by the U.S. Food and Drug
Administration (FDA), but also for the simultaneous detection of six other potentially
mutagenic nitrosamine impurities.
Introduction Nitrosamine Compound Chemical Structure
CH3
The announcement for the recall of NDMA N O
ARB medicines Valsartan, Losartan, H3C N
and Irbesartan made N-Nitroso
impurities a focus for regulatory N
N
O
NDEA
agencies including the FDA and the
European Medicines Agency (EMA).
Nitrosamine impurities are byproducts O
NMBA O N
produced in trace amounts during the N OH
manufacturing processes of these
H3C N O
medicines. These impurities are
N
classified as probable carcinogens (that NEIPA
H3C
is, potentially genotoxic impurities). H3C
The LC/MS-based method described
in this Application Note was performed
N
on an Ultivo triple quadrupole LC/MS. NDIPA
It presents a comprehensive analysis N
O
of 12 nitrosamine compounds at very
low detection limits. All nitrosamine
O
impurities are of small molecular weight, N
NDBA
ranging from 74 to 158. The compounds
N
included:
• N-nitrosodimethylamine (NDMA)
• N-nitrosodiethylamine (NDEA) N
NMEA N O
• N-nitroso-4-methyl-4-aminobutyric
acid (NMBA)
• N-nitrosoethylisopropylamine O
N
(NEIPA)
NPyR
N
• N-nitrosodiisopropylamine (NDIPA)
• N-nitrosodibutylamine (NDBA)
• N-nitrosoethylmethylamine (NMEA) O
NPIP
• N-nitrosopyrrolidine (NPyR) N N
• N-nitrosopiperidine (NPIP)
• N-methyl-N-nitrosoaniline (NMPhA) O
N
• N-isopropylmethyl nitrosamine NMPhA
N
(NMIPA)
• N-tert-butyl-N-ethylnitrosamine
N
NMIPA O N
N-tert-butyl-N-ethylnitrosamine N O
N
2
Experimental LC configuration and parameters
Table 1. UHPLC configuration and settings.
Chemicals and reagents
Parameter Value
All 12 nitrosamine standards used for
Agilent 1290 Infinity II high speed pump (G7120A)
the study were arranged from PS3 Labs Instruments Agilent 1290 Infinity II multisampler (G7167B)
LLP, Hyderabad, TS, India. LC/MS grade Agilent 1290 Infinity II multicolumn thermostat (G7116B)
solvents such as methanol and water Needle Wash Methanol: Water/ 80:20
Analytical Column Agilent InfinityLab Poroshell HPH C18, 2.1 × 100 mm 1.9 µm (p/n 695675-702)
Data analysis Column Temperature 40 °C
Data were acquired and analyzed Mobile Phase A 0.2 % formic acid in water
using Agilent MassHunter software Mobile Phase B Methanol
version 1.1 for Ultivo. MS/MS transitions Flow Rate 0.4 mL/min
were obtained and optimized using Time (min) %B
the Agilent MassHunter Acquisition 0 1
2 13
Optimizer software to determine optimal Gradient
5.5 80
precursor and product ions, fragmentor 8 95
10 95
voltages, and collision energies upon
Stop Time 10 minutes
injection of a neat solution (of each
Post Time 3 minutes
individual compound at a concentration
level of 1,000 ng/mL, 1 µL injection
volume in flow injection mode). Triple quadrupole mass spectrometer configuration and parameters
Table 2. MS configuration and source settings.
Parameter Value
Instrument Agilent Ultivo triple quadrupole mass spectrometer
3
MS/MS compound information for analytes
Table 3. Detailed MRM settings in dynamic MRM mode in Ultivo.
4
Results and discussion Table 4. Result summary of the Agilent Ultivo TQ. Data include signal-to-noise (S/N), calculated LOQ,
coefficient of regression, and calibration curve fit. All standards used 1/x weighted calibration curve.
The calibration concentrations ranged
Agilent Ultivo TQ
from 0.05 to 100 ng/mL with specific
LOD LOD LOQ LOQ Linearity Range
details mentioned in Table 4. R2 values Compound (ng/mL) (S/N) (ng/mL) (S/N) R2 (ng/mL)
were greater than 0.996 for all the NDMA 0.05 9.74 0.1 13.49 0.997 0.05 to 100
analytes, displaying linear responses NDEA 0.025 22.68 0.05 53.82 0.998 0.025 to 100
throughout the concentration range. NMBA 0.05 23.88 0.1 31.18 0.998 0.05 to 100
Figure 1 shows the representative NEIPA 0.025 57.37 0.05 69.67 0.999 0.025 to 100
overlaid dMRM chromatogram from NDIPA 0.025 15.72 0.05 45.18 0.998 0.025 to 100
Ultivo TQ showing elution of all NDBA 0.05 225.18 0.1 323.11 0.999 0.025 to 100
12 nitrosamine compounds at 10 ng/mL. NMEA 0.075 7.84 0.1 12.10 0.999 0.1 to 100
* S/N was calculated using the Auto-RMS algorithm, noise reference selected as sample using
Agilent MassHunter Quantitaive 10 software.
×102
1.08
1.04
NEIPA
1.00
0.96
0.92 NDBA
0.88
0.84
NMIPA
0.80
0.76
0.72
0.68
0.64
0.60
Counts (%)
0.56
0.52 NPyR
NPIP
0.48 NMEA N-tert-butyl-N-ethylnitrosamine
0.44
0.40
0.36
0.32
0.28
0.24 NDEA
0.20
0.16
NDIPA
0.12
0.08 NDMA NMPhA
0.04
NMBA
0
0.6 0.8 1.0 1.2 1.4 1.6 1.8 2.0 2.2 2.4 2.6 2.8 3.0 3.2 3.4 3.6 3.8 4.0 4.2 4.4 4.6 4.8 5.0 5.2 5.4 5.6 5.8 6.0 6.2 6.4
Acquisition time (min)
5
Accuracy and reproducibility provided in Table 4, demonstrating
Calibration curves for each of the reproducibility across all calibration
12 compounds demonstrated an points with CV <15%. Table 5 shows a
accuracy rate within 20% of the detailed summary of accuracies and
expected concentration limit at the reproducibility of results at different
limit of quantitation (LOQ). Detection concentration levels. Figure 2 shows
limits, quantitation limits, and the representative calibration curves
calculated statistics are shown for generated from the Ultivo LC/TQ system.
the measured linear calibration range
Table 5. Accuracy and reproducibility for different concentration levels in Agilent Ultivo LC/TQ (n = 3).
0.25 102.47 2.79 98.53 4.77 95.53 4.31 101.43 5.16 101.6 3.59 105.53 3.60
0.5 100.00 2.68 98.53 2.15 105.67 1.15 95.30 3.15 95.30 1.06 95.77 2.07
1.0 104.27 4.70 99.47 1.22 105.67 1.80 98.47 4.59 97.63 3.88 95.77 2.39
0.25 105.1 5.64 103.63 6.66 101.43 4.79 106.43 6.59 107.5 7.97 106.47 5.27
0.5 91.87 6.20 96.20 3.19 93.07 5.83 84.40 3.35 90.50 5.61 96.77 0.26
1.0 93.80 5.39 99.97 2.36 102.67 0.79 108.03 2.25 100.23 2.94 94.83 1.10
* For NPIP, the first concentration is 0.15 ng/mL, not 0.1 ng/mL.
6
×104
NDMA
3.5 y = 393.189287 * x + 9.190576
3.0 R2 = 0.9975
Responses
2.5
2.0
1.5
1.0
0.5
0
×105
NDEA
1.2 y = 1275.476759 * x + 11.425911
1.0 R2 = 0.9981
Responses
0.8
0.6
0.4
0.2
0
×104
1.8 NMBA
1.6 y = 179.496263 * x + 5.672510
1.4 R2 = 0.9990
Responses
1.2
1.0
0.8
0.6
0.4
0.2
0
×105
NEIPA
5 y = 5617.307181 * x + 22.999383
R2 = 0.9991
Responses
4
3
2
1
0
×105
2.4 NDIPA
y = 2330.350728 * x - 6.678439
2.0
R2 = 0.9990
Responses
1.6
1.2
0.8
0.4
0
×105
NDBA
4 y = 4561.913169 * x + 175.317543
Responses
R2 = 0.9994
3
2
1
0
0 10 20 30 40 50 60 70 80 90 100
Concentration (ng/mL)
Figure 2. Representative calibration curves from the Agilent Ultivo LC/TQ for
compounds observed throughout the chromatogram. All calibration curves
used a 1/x weighting factor.
7
Conclusion References Acknowledgement
The Agilent Ultivo triple quadrupole 1. https://www.fda.gov/drugs/ We sincerely acknowledge and thank
LC/MS can analyze nitrosamine drug-safety-and-availability/fda- PS3 Labs LLP, Hyderabad, TS, India for
impurities at the low concentration levels updates-and-press-announcements- providing us nitrosamine standards.
demanded by regulatory requirements. angiotensin-ii-receptor-blocker-arb-
This Application Note demonstrates recalls-valsartan-losartan
the sensitivity of the Ultivo LC/TQ in 2. FDA guidance document:
the detection of certain nitrosamine Development and validation of
impurities at low concentration a RapidFire-MS/MS method for
levels. This method can be used in screening of nitrosamine impurities.
the quantitation of these impurities
in different ARB drug products with 3. FDA guidance document: Liquid
some alterations in chromatographic chromatography-high resolution
conditions, according to the elution mass spectrometry (LC-HRMS)
pattern of the drug product. method for the determination of six
nitrosamine impurities in ARB drugs.
www.agilent.com/chem