You are on page 1of 10

Hemodialysis International 2007; 11:S39–S48

Beyond ultrapure hemodialysis: A necessary


and achievable goal

Carl M. KJELLSTRAND,1,2,3 Per KJELLSTRAND4,5


1
Loyola University Chicago, Chicago, Illinois; 2SUNY at Brooklyn, Brooklyn, New York; 3Karolinska Institute,
Stockholm, Sweden; 4Lund’s Universitet, Lund, Sweden; 5Senior Scientific Advisor, Gambro AB, Lund,
Sweden

Abstract
Survival of chronic hemodialysis patients is worse than that of many patients with cancers or severe
infections. An important cause of chronic inflammation is impurities infused into patients during
dialysis. Definitions of dialysis purity have been narrow and focused on metals in dialysate water and
on bacterial contaminants. There is no standard for priming fluids or toxins released directly into
blood from inside the extracorporeal circuit. We propose a much broader standard of dialysis purity
that also includes phthalate metabolites, bisphenols, spalled particles, and other contaminants from
dialysis machines, filters, and bloodlines. Standards must include new methods for measuring bac-
teriological contaminants in addition to colony-forming units and endotoxin determinations. These
include the sensitive silkworm larva plasma test that detects peptidoglycan that is missed by endo-
toxin tests and standards for newly detected small molecular bacterial detritus. Current levels for
‘‘standard’’ bacteriological contaminants are woefully inadequate and should be increased. New
standards for contamination with plasticizers and spallation are also necessary. Studies with ultra-
pure dialysis have shown almost immediate patient benefits with increased well-being and stabil-
ization of the cardiovascular system during and between dialyses. Intermediate effects include lower
C-reactive protein levels, better response to erythropoietin, increased appetite, and improved nu-
trition. Over the years, amyloidosis and carpal tunnel syndrome have become less common and
cardiovascular deaths have decreased. Standards for dialysis purity must be sharpened and expand-
ed and this becomes even more urgent with daily and long nightly hemodialysis. All contaminants
received by patients, whether biological, chemical, or physical, must be considered.

Key words: Hemodialysis, ultrapure, dialysis systems

INTRODUCTION AND PERSPECTIVE failure, the procedure usually was only necessary for 2 or
3 weeks. With the arrival of chronic hemodialysis in
When dialysis was introduced into clinical medicine in
1960, the need for dialysis was prolonged abruptly from
the mid-1940s, scant attention was paid to the purity of
days to years. Since then, membranes made from syn-
the dialysate. As the dialysis membranes in use then were
thetic materials have come into use that may allow pas-
impermeable to molecules above a molecular weight of
sage of molecules up to 20,000 da. An unavoidable
1500 da, the membrane protected patients from biologi- compromise is that as toxins are removed from blood,
cal contaminants and as dialysis was only for acute renal
any similarly sized contaminants in the dialysate can pass
into the patient.
On average, most patients in the United States begin-
Correspondence to: C. M. Kjellstrand, MD, PhD, FACP, ning dialysis today rarely survive more than 4 years.1
FRCP(C), 965 Creekside Dr, College Place, WA 99324, With an average of 150 dialyses per year, each for about
U.S.A. 3 hr, patients will have only about 600 dialyses over
E-mail: carl.kjellstrand@gmail.com 4 years for a total of about 1800 hr. With short daily

r 2007 The Authors. Journal compilation r 2007 International Society for Hemodialysis S39
Kjellstrand and Kjellstrand

and long nightly hemodialysis, things are different now: turn are thought to contribute to malnutrition, amyloi-
survival is likely to be better and patients will live longer; dosis, and arteriosclerosis. This has been described as the
the extreme rarity of surviving for more than 40 years MIA-syndrome of dialysis: malnutrition-inflammation-ar-
may become more common.2 Patients dialyzing 6 times a teriosclerosis.5 These relationships have become of great
week now will undergo more than 300 dialyses a year and interest to investigators. A recent Gateway search cross-
be on dialysis for 900–1000 hr if on short daily dialysis referencing inflammation and dialysis over the last few
and for some 2500 hr if on long nightly dialysis. Assum- years yielded 1986 articles; cross-referencing oxidative
ing that only 0.5 g of plasticizers, rubber, and spalled stress with dialysis resulted in 811 articles.
particles are infused at each treatment, this results in a
load of at least 150 g/year and if a patient survives for 25
years some 4 kg.
DEFINITIONS OF PURITY
Most considerations of purity in dialysis have focused Definitions or standards of purity in dialysis almost ex-
on metals, small organic and inorganic contaminants clusively refer to the bacteriological purity of the dialy-
such as chloramines in water, or bacterial growth and sate. This review will not consider metals and small water
endotoxin in dialysate, all on the outside of the dialysis purification compounds such as chloramines. Purity is
membrane. However, there is also the direct interaction defined by endotoxin levels (endotoxin units or EU) and
between blood and surfaces in the dialyzer and blood the number of colony-forming units (CFU) in the dialy-
tubing. The dialysis system has been described as a ‘‘bio- sate. The present standards are shown in Table 1. These
reactor’’ where a patient’s blood interacts with foreign differ considerably from country to country; in the United
surfaces and is exposed to contaminants in the dialysate, States, the allowed number of CFU is 200,000 times and
some of which enter the body of the patient.3 Thus, dia- the endotoxin level is 200 times those in Japan. As meth-
lysis purity and biocompatibility are described generally ods of measuring impurities improve, standards will
as effects of the membrane and the bacterial purity of the change. Thus, any of the current or proposed standards
dialysate.4 are arbitrary and will probably become more stringent
There are many other sources of contaminants in the with the discovery that even minute amounts of contam-
dialysis system, several from the inside. These include im- ination may lead to long-term disease in dialysis patients.
purities from intravenous fluids used to prime and rinse We are of the opinion that the standard should be no CFU
the circuit and for volume administration, and plasticiz- and no measurable level of endotoxin in dialysate. Purity
ers, glues, fillings, bonding materials, spalled particles, should be ‘‘beyond ultrapure.’’
and other substances leached from the dialyzer and blood There are presently no standards for plasticizers or
tubing that are infused into patients’ blood. Plasticizers, spallation.
particularly phthalates in blood tubing, curing materials
used in membrane fabrication, chemicals used in dialyzer TYPES OF IMPURITY
reuse, spalled particles from blood tubing sets and pump
segments, potting compounds in the headers of hollow
fiber dialyzers, and phenols from dialyzer casings, have
Bacteriological
all been associated with ill effects in dialysis patients. Only CFU and endotoxin levels are considered today, but
These contaminants and their effects will increase as the the limulus amebocyte lysate assay detects only the lipo-
number of dialyses per week doubles and as patient sur- polysaccharide endotoxin of Gram-negative bacteria and
vival is prolonged. perhaps b-glycan from fungi. Only the silkworm larva
The clinical consequences of an ‘‘impure dialysis sys- plasma test detects the peptidoglycan of the cell wall
tem’’ have been thought to contribute to inflammation of Gram-positive bacteria.6 Recent studies indicate that
and oxidative stress in hemodialysis patients and these in there are small bacterial fragments or metabolites of

Table 1 Standards of dialysate purity

U.S.A. Proposed U.S.A. EURO ST EURO UP Japan


CFU/mL o2000 o200 o100 o0.1 o0.001
Endotoxin EU/mL No limit o2.0 o0.2 o0.03 o0.01
EURO = Europe; ST =standard; UP =ultrapure.

S40 Hemodialysis International 2007; 11:S39–S48


Beyond ultrapure hemodialysis

Gram-negative bacteria with molecular weights between this to be of the order of 13 to 42 mg for each dialysis, and
1500 and 6000 da that are not detected by the limulus the amount seems proportional to the fat content of the
test. Such fragments easily pass current dialysis mem- patient’s blood.10 This translates to a potential burden of
branes and stimulate inflammatory responses in mono- more than 30 g over 5 years in patient on conventional
cytes such as increased IL-6 formation.7 hemodialysis. While DEHP is relatively nontoxic in cell-
There are also no standards for contamination by vir- culture studies, it is transformed in the body into other
uses. However, hepatitis C virus has been suspected of molecules, particularly mono-(2-ethylhexyl) phthalate
passing through dialysis membranes, perhaps in unde- (MEHP), which is highly toxic. When this is added to
tectable small cracks, and so cross-infecting dialysis pa- cell-culture media in concentrations relevant to human
tients.8 This indicates that patients should have dedicated exposure, almost no cells remain viable after 3 hr of
machines. Dialysis machines are not particularly expen- exposure.11 The potentially harmful effects in humans
sive, but the costs of hospitalization and medical care are. are hotly debated, some calling for an immediate halt to
the use of PVC in medical devices, while others state that:
Plasticizers ‘‘to question the safety of PVC is a major waste of time,
resources and human potential.’’12–14
Several harmful nonbiological substances have been Phthalates induce IL-1 production in vitro and their
shown to be released from components in the dialysis potential toxic effects include necrotizing dermatitis, he-
system. A workshop 30 years ago implicated rubberizers patic inflammation, endocrine disturbances, and carcino-
and metals from the tubing and dialyzer membranes.9 genesis. Phthalates are excreted in the urine, and the
Recently, most interest has focused on problems from plas- blood level of phthalates in both conventional hemodial-
tics used in modern dialysis. These are listed in Table 2. ysis and peritoneal dialysis patients is 4–20 times that of
Phthalates, used for softening polyvinylchloride (PVC), normal controls.15–17 Because of the very large exposure
are those most commonly used today and almost all rele- to phthalates with long-term hemodialysis and especially
vant studies are from Germany and Japan. Fairly large with more frequent hemodialysis, we believe that PVC
quantities of di-(2-ethylhexyl)-phthalate (DEHP) are in- should be eliminated from hemodialysis devices.
fused from blood-tubing sets and from bags containing Other potentially toxic additives in dialysis are bisphe-
intravenous fluids. Kambia and associates have estimated nol-A and its metabolites. These are eluted from the
polycarbonate shell and polysulfone membrane of some
dialyzers.18–20 Again, almost all studies are from Japan.
Table 2 Contaminants of plasticizers, sterilizing agents, and These compounds are present in the blood of dialysis
reuse chemicals
patients in quantities 20 times that in normal controls.21
Compounds in polyvinylchloride The toxic effects of bisphenols include hematopoietic
Di-(2-ethylhexyl) phthalate (DEHP) damage, sperm changes, and, perhaps most alarming, a
Mono(2-ethylhexyl) phthalate (MEHP) defect in gene transcription.22 This finding and the car-
2-Ethylhexane (2-EH) cinogenesis described for phthalates are of particular con-
Phthalic acid (PA) cern in dialysis patients who have an incidence of cancer
4-Hepatanone (4-H) 7 times that of the general population.23
Bisphenol-A The well-known first-use syndrome is related to eth-
And various metabolites
ylene oxide used in dialyzer sterilization.24 Glutaralde-
Ethylene oxide
Chemical for reuse hyde, peracetic acid, and formaldehyde used in chemical
Peracetic acid reuse procedures of dialyzers are other poisonous and
Formaldehyde oxidizing substances that are released into patients. Dur-
Glutaraldehyde ing reuse, these agents bind to materials in the dialyzer,
Unknown reactive contaminants cannot be completely rinsed out, and are slowly released
Glues and infused into patients during the subsequent treat-
Epoxy compounds ment.25 Formaldehyde and glutaraldehyde have a strong
Bonding cross-linking capacity for nucleic acids and proteins and
Filling so have been used as fixatives for biological tissues in
Potting both light and electron microscopy. The cross-linking re-
Cleaning
actions are the mechanism for killing microbes and are
Products from incomplete bonding and hardening
irreversible and so both substances have the potential for

Hemodialysis International 2007; 11:S39–S48 S41


Kjellstrand and Kjellstrand

the build-up of irreversible long-term damage even at in the U. S. Pharmacopoeia, 23, general chapter 788. This
concentrations that yield no acute effects. At present, standard allows up to 12,000 particles o10 mm in size
there are no reliable methods for the prediction of such and 2000 particles o25 mm in size per liter to be infused.
long-term damage and so no way to protect patients other A patient on parenteral nutrition could be infused with
than to keep such substances completely away from their 24,000 small and 4000 larger particles each night, and
equipment.26 would have approximately 0.175 billion small and 30
New, unexpected and unknown compounds may be million larger particles infused over a 20-year period of
introduced in dialysis manufacturing, some with imme- treatment. This amount appears excessive, but the poten-
diate lethal results. In a recent episode, more than 50 tial load of spalled particles for a patient on daily hemo-
patients died worldwide when treated with dialyzers in dialysis dialyzing some 300 times a year and expected to
which a Freon-like compound had been introduced dur- live possibly 10 to 20 years and possibly experiencing up
ing manufacturing.27 Such dramatic episodes are only the to 3000 to 6000 dialyses will be many times greater.
tip of the iceberg, and lower but daily infusions of sub-
stances undoubtedly are damaging to long-term dialysis MEASURING CONTAMINANTS
patients. Very reactive substances from incomplete bond-
ing and hardening of 2-compound adhesives, bonds, fill- A summary of the methods for measuring contaminants is
ers, and glues used in dialysis material fabrication are also shown in Table 3.
infused into patients as it is very unlikely that a predia-
lysis rinse with 0.5 to 2 L of saline will completely wash Bacteriological
these undesirable substances out. Bacteriological contaminants are studied by incubating
All medical devices and fluids have to be sterilized so the suspected sample in or on various growth media. If
that no living microbes enter the patient. Killing microbes living microbes are present, they will eventually grow and
can be achieved in several ways, including either breaking form colonies. There are considerable differences in the
or cross-linking the molecules in the living organism, a number of CFUs detected, depending on the procedure
‘‘stick in the wheel’’ approach. Nucleic acids and proteins and what media and growth times are used.41,42 Endo-
are the main targets of these processes. There are 3 main toxin content is measured by the limulus amebocyte ly-
methods for sterilization: heat, radiation, and reactive sate test or by chromogenic methods that are constantly
chemicals. These means may be combined, while ma- being made more sensitive; the lower limit of detection is
nipulation of pH may also be used to enhance killing ca- now in the neighborhood of 0.005 EU/mL or 1 ng/mL. As
pacity. These methods not only damage the molecules of discussed previously, peptidoglycan is not detected by
the microbes but also affect other molecules in the plas- these methods but is detected with the silkworm larva
tics and the fluids. A serious problem with these ap- plasma test. No commercial test is available for detection
proaches is that the molecular nature of many of the of smaller bacterial fragments at this time.
reactive fragments that develop is unknown. These frag-
ments may not be detected by currently available meth- Plasticizers
ods, and there is no way of predicting their effects on
patients. Degradation of glucose during heat sterilization Phthalate metabolites and bisphenols can be quantified
of fluids for parenteral nutrition and peritoneal dialysis is by several different physical methods, none easy or read-
an example of how complex a seemingly simple chain of ily available commercially.
events can be.28–32 The low sensitivity of existing animal
models means that such models have not proved to be Table 3 Ways to detect and measure contaminants
very useful.33,34 Physical: Chemical finger printing
Spalled particles consisting of PVC, polyurethane, and 1. UV spectro-photometry
silastic, particularly from the pump segments in the ex- 2. NIR ‘‘near infra red’’ spectro-photometry
tracorporeal circuit, have been found in the lungs, skin, 3. IR ‘‘infrared’’ spectro-photometry
spleen, and liver of patients and have been associated 4. HPLC-UV-DAD (diode array detector)
with signs of hepatic irritation and fibrosis and hypercal- Cell testing: Biological fingerprinting
cemia.35–40 We could not find a single article on spalla- Clinical: Clinical fingerprinting
tion in hemodialysis published after 1992. There is 1. Quantitate ill-effects reported by patients
2. Temperature, vital signs
presently no standard for spallation in dialysis, but there
3. Oxidant and inflammatory markers
is a standard for large volume intravenous fluid infusion

S42 Hemodialysis International 2007; 11:S39–S48


Beyond ultrapure hemodialysis

There are 3 non-specific ways to quantify contamin- within 5 years and thus long-term observations are not
ants, physical/chemical, biological, and clinical finger- possible. Long-term problems including amyloidosis and
printing, and all can give estimates of the sum of its complications and cardiovascular problems that de-
contaminants and their ill effects. In physical fingerprint- velop over time are discussed later. However, common
ing, curves of the output from high-performance liquid sense dictates that a membrane that does not activate the
chromatography (HPLC) are made from dialysate and clotting and complement cascades is to be preferred over
predialysis and postdialysis blood samples and the peaks a membrane that causes violent biological activations in a
and valleys compared among different dialysis systems. patient’s blood.
This is similar to early studies of ‘‘middle molecule’’ toxins
in dialysis.43 Only when the chemical nature of the con-
taminant is known and it is available in a pure form
REMOVING IMPURITIES
(which is usually not the case) can the amount of con- There are 4 principal methods of removing contaminants:
tamination be estimated. In biological fingerprinting, the filters, absorbers, chemical cleaning, and physical clean-
fluids are added to cell cultures and damage is estimated ing using large volumes and hot water. The methods used
by cell viability or enzyme release.10–12,33,34 Clinical fin- differ on the outside and inside of the extracorporeal cir-
gerprinting refers to studies of the ill effects reported by cuit. Filters have been used successfully to remove bac-
dialysis patients such as episodes of hypotension, ar- teria and endotoxin down to undetectable limits from
rhythmias, cramps, headache, backache, itching, nausea, dialysate,50 but with the discovery of small bacterial tox-
and vomiting, together with measuring clinical effects ins, there may be a need to decrease pore size in order to
such as fall in blood pressure and changes in pulse rate remove them.7 Recently, an absorber to bind endotoxin
and body temperature. Studies of the effect of different has been described that can reduce the endotoxin content
qualities of dialysate on inflammatory and oxidant stress of both dialysate and blood, but no clinical testing has
markers in patients also belong in this category. been reported.51 Filters and absorbers can be used to
Spallation is studied with in vitro techniques that in- produce a ‘‘beyond ultrapure’’ dialysate but they will not
volve placing a filter or Coulter counter on the venous produce an ultrapure dialysis system.
return line from the dialyzer and counting the number of Phthalates and bisphenols, small molecules with a mo-
particles collected. There are no methods for the study of lecular weight of o1000 da, will pass through most filters
spallation in vivo. and are not bound to specific binding sites. It is also im-
portant to be aware that everything added to the extra-
corporeal circuit may be released and so introducing
BIOCOMPATIBLE MEMBRANES more filters or absorbers may reduce one form of impur-
Complement activation and leucopenia in dialysis was ity but add another. While reuse with chemicals probably
first described more than 30 years ago and was found to reduces both plasticizers and spallation because of the
be highly dependent on the nature of the membrane.44 volumes of fluids used in the washout, the chemicals
Cellophane was the least biocompatible membrane as de- themselves can add toxicity. Moreover, these chemicals
fined by complement activation and leucopenia; polysul- often penetrate poorly into the organic debris left in a
fone, polyamide, and polyacrylonitrile membranes were used dialyzer where bacteria may hide. The best method
the most biocompatible, with other membranes in be- to clean the inside of the extracorporeal circuit appears to
tween. Other biological effects, such as elastase release be thorough physical washing with rinse volumes much
and thrombocytopenias, do not necessarily occur in tan- larger than the usual 0.5 to 2 L of saline, followed by
dem with the more dramatic complement-leucopenia re- polishing with hot water. Water has the advantage of
sponse and there is considerable intrapatient variability in being nontoxic, heat has the advantage that it penetrates
the responses.45,46 Oddly, there are no detectable acute everywhere, and both are easy to dispose of.
clinical ill effects in patients paralleling the dramatic com- Two modern dialysis systems designed to produce be-
plement activation-leucopenia changes in the blood.47,48 yond ultrapure dialysis are the German Geniuss system
More recent data have shown no survival advantage (Fresenius Medical Care Germany, Bod Homburg, Ger-
with open, biocompatible membranes over others.49 many) and the American Aksys PHDs system (Lincoln-
However, any subtle beneficial effects may have been shire, IL, U.S.A.). Both use a tank of dialysate. In the
drowned out by the effect of present-day short, intermit- Genius system, an ultra-sterile, filtered bicarbonate dial-
tent dialysis with its poor patient survival. More than 60% ysate is constantly exposed to the bactericidal effect of
of hemodialysis patients in the United States are dead an ultraviolet light emanating from a light rod in the

Hemodialysis International 2007; 11:S39–S48 S43


Kjellstrand and Kjellstrand

dialysate tank.52,53 In the Aksys system, the dialyzer and Table 4 Comparison of symptoms, as percent of dialyses
tubing set are part of the dialysis apparatus and are with symptom and changes in vital signs as patient changed
changed only monthly. After each dialysis, the entire sys- from conventionally clean dialysis to beyond ultrapure
tem, both inside and outside the extracorporeal circuit, is dialysis
scrubbed with 40 L of water and heat-sanitized with 8 L
Conventional Ultrapure
of water at 75 1C.54–56 Both these systems have under-
N=2819 N=1636 p
gone extensive clinical testing and appear to have brought
about improved patient outcomes. To us, the physical Hypotension 14% 9% o0.0001
Arrhythmia 2% 0.1% o0.0001
process of water and heat cleaning of both the inside and
Headache 7% 4% 0.004
outside of the extracorporeal circuit, together with ex-
Cramps 5% 3% 0.016
tended nonchemical reuse, appears to be the best method Nausea 2% 1% 0.012
presently available to clean out bacterial, plasticizers, and Backache 1.2% 0.2% 0.002
physical spallation contaminants simultaneously. Postdialysis PR 490 35% 24% o0.0001
Fall in systolic BP 17  21 8  21 o0.0001
mmHg
BENEFITS OF BEYOND ULTRAPURE Fall in diastolic BP 6  12 2  12 o0.0001
DIALYSIS mmHg
Increase in PR/min 2  13 0  12 o0.0001
Immediate beneficial effects
The symptoms declined by 50% to 95%, and the decline in blood
Immediate beneficial effects of ultrapure dialysate, usually pressures was much smaller when patients were dialyzed on the
seen within days or weeks, appear to include more stable ultrapure system.
cardiovascular status during dialysis with less hypoten-
sion and arrhythmias and an improved general well-being levels were significantly lower, and at 3 months, while
with fewer episodes of headache, backache, nausea, and hemoglobin levels were equal between the 2 groups, the
other clinical symptoms.55,57–61 In more than 4400 dial- need for erythropoietin was 26% lower in patients using
yses using 6 dialyses a week and the same patients, dia- the ultrapure dialysate.62,63 Bonforte et al., followed 32
lysis with conventional machines was compared with patients for 9 months and reported almost identical find-
slightly longer dialysis using the Aksys PHD. Cleanliness ings.64 Schiffel et al., randomly assigned 30 patients
was measured as the occurrence of any bacterial growth to conventional or ultrapure hemodialysis. During a 24-
and determination of endotoxin levels. In approximately month follow-up, they found lower levels of inflamma-
400 prime samples, conventional machines failed to meet tory markers and better preserved residual renal function
the intravenous fluid standard (o0.5 EU/mL) in a third in those dialyzed with ultrapure dialysate.65,66 In a fur-
of the samples and the old European standard (o0.25 ther study with 12 months of follow-up, inflammatory
EU/mL) in two-thirds of the samples. The PHD showed markers were significantly lower in patients on ultrapure
no detectable endotoxin when tested to the limit of the dialysate while dry body weight, mid-arm muscle cir-
method (o0.005 EU/mL). The patients reported signifi- cumference, serum albumin concentration, growth factor
cantly further improvement in well-being when using the 1, leptin, and protein catabolic rate increased. All these
beyond ultrapure dialysis system and vital signs were factors did not change in patients using conventional
better preserved. In multivariate analysis, the significant dialysate.65,66
factor associated with the improvements was the purity
of the dialysis, not differences in the time or the speed of Long-term benefits
dialysis. The results are summarized in Table 4.
Over decades, patients dialyzed with ultrapure dialysate
have fewer episodes of carpal tunnel syndrome, amyloi-
Intermediate effects dosis, and dialysis arthropathy. Cardiovascular mortality
Intermediate effects take months to detect and include a is also reported to be lower in patients on ultrapure
better response to erythropoietin, lower C-reactive pro- dialysate when compared with patients dialyzed against
tein (CRP) levels, and increased appetite with improved regular dialysate.67–69 Baz et al., followed 226 patients
nutrition. Sitter and coworkers randomized 30 hemodi- for up to 16 years. In 39 of those treated with ultrapure
alysis patients to either conventional or ultrapure dialy- dialysate, none had carpal tunnel symptoms at 10 years
sate. In the group using ultrapure dialysis, CRP and IL-6 compared with 40% of 103 patients treated with conven-

S44 Hemodialysis International 2007; 11:S39–S48


Beyond ultrapure hemodialysis

tional dialysate.67 Similarly, Lonnemann and Koch com- 6. Minimal spallation: much less than the U.S. Pharma-
pared the incidence of carpal tunnel syndrome at 13 years copoeia limits or o1000 particles of o10 mm in size
in patients treated with conventional dialysate (70%) with and o200 particles of o25 mm in size during each
those treated with filtered ultrapure dialysate (30%) and dialysis.
with the Genius machine (15%).68 7. Dedicated machine for each patient.
At least 3 groups have observed less cardiovascular
morbidity in patients treated with ultrapure dialysate
All current standards for purity of dialysis are woefully
when compared with patients on conventional equip-
inadequate as they mainly refer to fluids outside the
ment.52,70–72 In a 3-year investigation of 60 patients,
blood circuit. It is meaningless to use dialysate contain-
Lederer et al., found that CRP levels were significantly
ing o0.005 EU/mL endotoxin and then use prime and
lower in those on the ultrapure system and there were 11
infusion fluids that contain 0.5 EU/mL and high concen-
cardiovascular events in the 38 patients by conventional
trations of phthalate. All contaminants received by
dialysis vs. 1 in 22 patients treated with ultrapure dial-
patients, whether biological, chemical, or physical, need
ysate (po0.0001).72
consideration.
Some may scoff at our definitions and consider beyond
CONCLUSIONS ultrapure dialysis as economically impossible. However,
(1) Beyond ultrapure dialysis means ultrapure dialysate several clinics have used this approach successfully for
on the outside of the dialyzer and also ultrapure decades. The savings from lower morbidity will exceed
conditions inside the blood circuit. the moderate cost.
(2) The benefits of ultrapure dialysate appear to include
a frequent decrease in symptoms during dialysis. Manuscript received October 2006; revised October 2006.
Intermediate effects manifesting themselves within
months include less inflammatory STIMULI, less
need for erythropoietin, better-preserved renal func- REFERENCES
tion, and improved nutrition. Over decades, amy- 1 U.S. Renal Data System. USRDS 2005 Annual Data Re-
loidosis, carpal tunnel syndrome, and perhaps port: Atlas of End-Stage Renal Disease in the United
cardiovascular events are considerably less common. States, National Institutes of Health, National Institute of
(3) The long-term effects of spallation and plasticizer Diabetes and Digestive and Kidney Diseases, Bethesda,
contamination from dialyzers, blood tubing, intra- MD, 2005.
venous fluids, and bags are not known, but many 2 Kjellstrand C, Buoncristiani U, Ting G, et al. Survival in
common ill effects suffered by dialysis patients such 415 patients treated by short daily hemodialysis for 948
patient years rivals renal transplant survival [abstract].
as skin disease, hepatic failure, endocrine dysfunc-
Hemodial Int. 2006; 10:118.
tion, hematopoietic damage, and marked increase in 3 Canaud B. Lecture at the Annual Dialysis Conference,
cancer may be related to such contaminants. February 2004, San Antonio, Texas, U.S.A., 2004.
We define beyond ultrapure, biocompatible hemodialysis 4 Ledebo I. Ultrapure dialysis fluid: Improving conven-
as follows: tional and daily dialysis. Hemodial Int. 2004; 8:159–166.
5 Qureshi AR, Alvestrand A, Divino-Filho JC, et al. In-
1. Bacteriologically ultrapure dialysate: endotoxin levels flammation, malnutrition, and cardiac disease as predic-
o0.0005 U/mL, CFU o 1/1000 mL, and silkworm tors of mortality in hemodialysis patients. J Am Soc
larva plasma test negative. Nephrol. 2002; 13(Suppl 1):S28–S36.
2. Toxicologically ultrapure dialysate: no PVC, bisphe- 6 Tsuchida K, Takemoto Y, Yamagami S, et al. Detection of
nols, trace metals, or reactive compounds. peptidoglycan and endotoxin in dialysate, using silk-
3. Biocompatible membrane in the dialyzer and bio- worm larvae plasma and limulus amebocyte lysate meth-
ods. Nephron. 1997; 75:438–443.
compatible plastics in the blood tubing and
7 Schindler R, Beck W, Deppisch R, et al. Short bacterial
pump segments and no complement activation-leu- DNA fragments: Detection in dialysate and induction of
copenia. cytokines. J Am Soc Nephrol. 2004; 15:3207–3214.
4. No PVC anywhere in the system and so no intraven- 8 Savey A, Simon F, Izopet J, Lepoutre A, Fabry J, Desenc-
ous fluids from PVC-containing bags should be used. los JC. A large nosocomial outbreak of hepatitis C virus
5. Ultraclean dialyzer and blood tubing: no toxic sub- infections at a hemodialysis center. Infect Control Hosp
stances (PVC, bisphenols, reactive compounds). Epidemiol. 2005; 6:752–760.

Hemodialysis International 2007; 11:S39–S48 S45


Kjellstrand and Kjellstrand

9 Kjellstrand CM, Alfrey AC, Eaton JW, et al. Toxicity of 23 Matas AJ, Simmons RL, Kjellstrand CM, Buselmeier TJ,
materials and medications used in dialysis. Trans Am Soc Najarian JS. Increased incidence of malignancy during
Artif Intern Organs. 1978; 24:764–769. chronic renal failure. Lancet. 1975; 1:883–886.
10 Kambia K, Dine T, Azar R, Gresisser B, Luyckx M, 24 Daugirdas JT, Ing TS. First-use reactions during hemo-
Brunet C. Comparative study of the leachability of dialysis: A definition of subtypes. Kidney Int. 1998;
DEHP and of trioctyl trimetallite (TOTM) from 24(Suppl):S37–S43.
hemodialysis tubing. Int J Pharm. 2001; 229:139– 25 Kjellstrand C. The search for scientific truth versus cor-
136. porate interests. Semin Dial. 1993; 6:281–283.
11 Kambia K, Gresisser B, Dine T, Dupin-Spriet T, Luyckx 26 Clemedson C, Barile FA, Chesné C, et al. MEIC evalu-
M, Brunet C. Evaluation of direct cytotoxicity of trioc- ation of acute systemic toxicity: Part VII. Prediction
tyltrimellitate (TOTM), DHEP on hepatocytes. Int J Artif of human toxicity by results from testing the first 30
Organs. 2004; 27:971–978. reference chemicals with 27 further in vitro assays.
12 Tickner JA, Schettler T, Guidotti T, McCally M, Rossi M. Part VIII. Multivariate partial least squares evaluation,
Health risks posed by use of Di-2-ethylhexyl phthalate including the selection of a battery of cell line tests
(DEHP) in PVC medical devices: A critical review. Am J with good prediction of human acute lethal peak blood
Int Med. 2001; 39:100–111. concentrations for 50 chemicals. ATLA. 2000; 28:
13 Ash SR. The importance of polyvinyl chloride in medical 161–200.
devices and the physician’s role in the choice of materi- 27 Bosch X. Baxter withdraws dialyzer after 12 patients die.
als. ASAIO J. 2003; 49:233–236. BMJ. 2001; 323:529.
14 Yakubovich M, Vienken J. Is there a need for plasticizer- 28 Kjellstrand P, Erixon M, Wieslander A, Lindén T, Mar-
free biomaterials in dialysis therapy? Med Device Technol. tinson E. Temperature: The single most important factor
2000; 11:12–3, 18–21. for degradation of glucose fluids during storage. Perit
15 Mettang T, Thomas S, Kiefer T, et al. Uraemic pruritus Dial Int. 2004; 24:385–391.
and exposure to di(2-ethylhexyl) phthalate (DEHP) in 29 Linden T, Cohen A, Deppisch R, Kjellstrand P, Wieslan-
haemodialysis patients. Nephrol Dial Transplant. 1996; der A. 3,4-Dideoxyglucosone-3-ene (3,4-DGE): A cyto-
11:2439–2443. toxic glucose degradation product in fluids for peritoneal
16 Mettang T, Fischer FP, Dunst R, Kuhlmann U, Retten- dialysis. Kidney Int. 2002; 62:697–703.
meier AW. Plasticizers in renal failure: Aspects of meta- 30 Martinson E, Wieslander A, Kjellstrand PT, Boberg U.
bolism and toxicity. Perit Dial Int. 1997; 17(Suppl 2): Toxicity of heat sterilized peritoneal dialysis fluids is
S31–S36. derived from degradation of glucose. ASAIO J. 1992;
17 Akingbemi BT, Ge R, Klinefelter GR, Zirkin BR, Hardy 38:M370–M372.
MP. Phthalate-induced Leydig cell hyperplasia is associ- 31 Erixon M, Wieslander A, Linden T, et al. Take care in
ated with multiple endocrine disturbances. Proc Natl how you store your PD fluids: Actual temperature de-
Acad Sci USA. 2004; 101:775–780. termines the balance between reactive and non-reactive
18 Yamasaki H, Nagake Y, Makino H. Determinations of bi- GDPs. Perit Dial Int. 2005; 25:583–590.
sphenol A in effluents of hemodialyzers. Nephron. 2001; 32 Wieslander A, Nordin MK, Kjellstrand PT, Boberg U.
88:376–378. Toxicity of peritoneal dialysis fluids on cultured fibro-
19 Haishima Y, Hayashi Y, Yagami T, Nakamura T. Elution of blasts L-929. Kidney Int. 1991; 40:77–79.
bisphenol-A from hemodialyzers consisting of polycar- 33 Kjellstrand PT, Lilliehorn G, Rydhög G. Toxicity testing
bonate and polysulfone resins. J Biomed Mater Res. 2001; of polymer materials for dialysis equipment: Is there any
58:209–215. need for in vivo testing. Cell Biol Toxicol. 1994; 10:137–
20 Hayashi Y, Matsuda R, Haishima Y, Yagami T, Nakamura 142.
A. Validation of HPLC and GC-MS systems for 34 Kjellstrand P, Lindqvist E, Nilsson-Thorell K. Toxicity
bisphenol-A leached from hemodialyzers on the testing of polymer materials for dialysis equipment:
basis of FUMI theory. J Pharm Biomed Anal. 2002; 28: Reconsidering in vivo testing. ATLA. 2000; 28:
421–429. 495–502.
21 Sugimura K, Naganuma T, Kakiya Y, Okada C, Sugimura 35 Leong AS, Disney AP, Gove DW. Spallation and migration
T, Kishimoto T. Endocrine-disrupting chemicals in of silicone from blood-pump tubing in patients on he-
CAPD dialysate and effluent. Blood Purif. 2001; 19: modialysis. N Engl J Med. 1982; 306:135–140.
21–23. 36 Altmann P, Dodd S, Williams A, Marsh F, Cunningham J.
22 Singleton DW, Feng Y, Chen Y, et al. Bisphenol-A and Silicone-induced hypercalcaemia in haemodialysis pa-
estradiol exert novel gene regulation in human MCF-7 tients. Nephrol Dial Transplant. 1987; 2:26–29.
derived breast cancer cells. Mol Cell Endocrinol. 2004; 37 Vanholder R. Biocompatibility issues in hemodialysis.
221:47–55. Clin Mater. 1992; 10:87–133.

S46 Hemodialysis International 2007; 11:S39–S48


Beyond ultrapure hemodialysis

38 Laohapand T, Osman EM, Morley AR, Ward MK, Kerr 53 Fassbinder W. Experience with the GENIUS hemodialy-
DN. Accumulation of silicone elastomer in regular sis system. Kidney Blood Press Res. 2003; 26:96–99.
dialysis. Proc Eur Dial Transplant Assoc. 1983; 19: 54 Twardowski ZJ. PHD: The technological solution for
143–152. daily hemodialysis? Nephrol Dial Transplant. 2003;
39 Bommer J, Ritz E. Spallation of dialysis materials— 18:19–23.
problems and perspectives. Nephron. 1985; 39:285– 55 Kjellstrand CM, Blagg CR, Twardowski Z, Bower J. Ap-
289. proaches to quotidian dialysis: The Aksys personal he-
40 Barron D, Harbottle S, Hoenich NA, Morley AR, Apple- modialysis system. Semin Dial. 2004; 17:151–153.
ton D, McCabe JF. Particle spallation induced by blood 56 Kenley R. Efficacy of heat disinfection with the Aksys
pumps in hemodialysis tubing sets. Artif Organs. 1986; personal hemodialysis system. Home Hemodial Int. 1997;
10:226–235. 1:36–40.
41 Ward RA. Dialysis water as a determinant of the ad- 57 Blagg C, Twardowski Z, Bower J, Kjellstrand C. Cardio-
equacy of dialysis. Semin Nephrol. 2005; 25:102–111. vascular instability and dialysate purity [abstract]. ASAIO
42 Ledebo I, Nystrand R. Defining the microbiological qual- J. 2003; 49:190.
ity of dialysis fluid. Artif Organs. 1999; 23:37–43. 58 Blagg CR, Bower JD, Twardowski ZJ, Kjellstrand CM.
43 Schoots AC, Mikkers FE, Claessens HA, De Smet R, Van Well-being during dialysis with biocompatible high flux,
Landschoot N, Ringoir SM. Characterization of uremic high efficiency dialyzers and ultra-pure dialysate [ab-
‘‘middle molecular’’ fractions by gas chromatography, stract]. J Am Soc Nephrol. 2001; 2:371A.
mass spectrometry, isotachophoresis, and liquid chroma- 59 Kjellstrand C, Twardowski ZJ, Bower J, Blagg CR. Hot
tography. Clin Chem. 1982; 28:45–49. water reuse of dialysis filters gives smoother dialysis than
44 Henderson LW, Miller ME, Hamilton RW, Norman ME. single use or chemical reuse. J Am Soc Nephrol. 2003;
Hemodialysis leukopenia and polymorph random mo- 14:502A.
bility-a possible correlation. J Lab Clin Med. 1975; 60 Kjellstrand C, Blagg CR, Twardowski ZJ, Bower J. Car-
85:191–197. diovascular stability during dialysis—relative influence
45 Kjellstrand P, Jacobson SH, Skroder R, et al. Leukocyte of dialysis purity, clearance and ultrafiltration speed [ab-
overshoot: A new sign of bioincompatibility in fast he- stract]. ASAIO J. 2004; 50:177.
modialysis. ASAIO Trans. 1990; 36:M314–M316. 61 Kjellstrand CM, Blagg CR, Young B, Bower J, Twardowski
46 Skroeder R, Kjellstrand P, Holmquist SH, Kjellstrand C, ZJ. 106 patient-years experience with the Aksys PHD
Jacobson S. Individual differences in biocompatibility system for quotidian home hemodialysis [abstract].
responses to hemodialysis. Int J Artif Organs. 1994; Hemodial Int. 2005; 9:80–81.
17:521–530. 62 Ouseph R, Jacobs AA, Ward RA. The use of ultrafiltered
47 Skröder R, Jacobson SH, Lins LE, Kjellstrand CM. dialysate reduces inflammation in hemodialysis patients.
Bio-compatibility of dialysis membranes is of no J Am Soc Nephrol. 2003; 14:30A.
importance for objective or subjective symptoms 63 Sitter T, Bergner A, Schiffl H. Dialysate related cytokine
during or after hemodialysis. Trans ASAIO. 1990; 36: induction and response to recombinant human erythro-
637–639. poietin in haemodialysis patients. Nephrol Dial Trans-
48 Bergamo Collaborative Dialysis Study Group. Acute in- plant. 2000; 15:1207–1211.
tradialytic well-being: Results of a clinical trial compar- 64 Bonforte G, Grillo P, Zerbi S, Surian M. Improvement of
ing polysulfone with cuprophan. Kidney Int. 1991; anemia in hemodialysis patients treated by hemodiafil-
40:714–719. tration with high-volume on-line-prepared substitution
49 Eknoyan G, Beck GJ, Cheung AK, et al. Hemodialysis fluid. Blood Purif. 2002; 20:357–363.
(HEMO) Study Group. Effect of dialysis dose and mem- 65 Schiffl H, Lang SM, Stratakis D, Fischer R. Effects of
brane flux in maintenance hemodialysis. N Engl J Med. ultrapure dialysis fluid on nutritional status and inflam-
2002; 347:2010–2019. matory parameters. Nephrol Dial Transplant. 2001;
50 Martin K, Laydet E, Canaud B. Design and technical 16:1863–1869.
adjustment of a water treatment system: 15 years 66 Schiffl H, Lang SM, Fischer R. Ultrapure dialysis fluid
of experience. Adv Ren Replace Ther. 2003; 10: slows loss of residual renal function in new dialysis pa-
122–132. tients. Nephrol Dial Transplant. 2002; 17:1814–1818.
51 Szathmary S, Hegyi E, Amoureux MC, et al. Character- 67 Baz M, Durand C, Ragon A, et al. Using ultrapure water
ization of the DialGuard device for endotoxin removal in in hemodialysis delays carpal tunnel syndrome. Int J Artif
hemodialysis. Blood Purif. 2004; 22:409–415. Organ. 1991; 14:681–685.
52 Kleophas W, Haastert B, Backus G, Hilgers P, Westhoff A, 68 Schwalbe S, Holzhauer M, Schaeffer J, Galanski M, Koch
van Endert G. Long-term experience with an ultrapure K-M, Floege J. b2-microglobulin associated amyloidosis:
individual dialysis fluid with a batch type machine. A vanishing complication of long-term hemodialysis?
Nephrol Dial Transplant. 1998; 13:3118–3125. Kidney Int. 1997; 52:1077–1083.

Hemodialysis International 2007; 11:S39–S48 S47


Kjellstrand and Kjellstrand

69 Lonnemann G, Koch KM. Beta(2)-microglobulin amyloi- 71 Wanner C, Zimmermann J, Schwedler S, Metzger T. In-
dosis: Effects of ultrapure dialysate and type of dialyzer flammation and cardiovascular risk in dialysis patients.
membrane. J Am Soc Nephrol. 2002; 13(Suppl 1): Kidney Int. 2002; 80(Suppl):S99–S102.
S72–S77. 72 Lederer SR, Schiffl H. Ultrapure dialysis fluid lowers the
70 Zimmermann J, Herrlinger S, Pruy A, Metzger T, Wanner C. cardiovascular morbidity in patients on maintenance he-
Inflammation enhances cardiovascular risk and mortality modialysis by reducing continuous microinflammation.
in hemodialysis patients. Kidney Int. 1999; 55:648–658. Nephron. 2002; 91:452–455.

S48 Hemodialysis International 2007; 11:S39–S48

You might also like