You are on page 1of 1

MCVSIX ENTERPRISES

Blk1 Lot 1 Laarlin Village, Barangay Sampaloc, Apalit Pampanga


Mobile No.: Globe: +63 917 855 0825 / Sun: +63 922 831 6674 / Office: (045) 901 0073
Email: mcvsix.enterprise@gmail.com

PERFORMANCE CALIBRATION REPORT

CUSTOMER: SMARTAID LABORATORY

INSTRUMENT: SNIBE CLIA ANALYZER MAGLUMI 800

SERIAL NO: 230200321500301

DATE PERFORMED: SEPTEMBER 27, 2019

Machine performance is verified using the manufacturer’s recommendation in


validating the functionality of the instrument.

Components Status/Criteria Results Remarks

1. Sampling 400ul 400ul Calibrated

2. Chamber 400ul 400ul Calibrated

3. Wash 400ul 400ul Calibrated

4. Refrigeration 12oC - 15oC 13.2oC Calibrated

5. Incubation 36.7oC - 37.3oC 37.06oC Calibrated

Performance Criteria Results Remarks


1. BGW RLU 100-1500 RLU 1080 Ok
2. BGW CV% <10% 6.8% Ok
3. Background 140000-230000 198042 Ok
4. Lightcheck 400000-650000 529655 Ok
5. LC CV% <10% 1.8% Ok

General Remarks: The unit is calibrated and in good working condition.

Verified by: Johnmar P. Aquino


Service Engineer