EQUIPMENT CALIBRATION CERTIFICATION


Customer Name : SmartAid Laboratory
Address : 203-204 158 Frisco Place cor. Del Monte, Roosevelt Ave.
Quezon City

Medical Equipment : Agappe


Model : MispaCount
Serial/Mfg. Nos. : 320816-000129
Date Performed : September 27, 2020
Next Calibration Schedule : September 27, 2021

The unit’s calibration performance was checked using Biorad Control (lot# 77761, 77762,
77763) All Passed.

Calibration Data
Lot #77883
High Criteria Result Units Remarks
WBC 19.7 ±2.4 19.9 K/µL Passed
RBC 4.84 ±0.4 4.84 M/µL Passed
HGB 16.3 ±0.8 16.39 g/dL Passed
HCT 42.5 ±6.2 45.2 % Passed
PLT 503 ±42 499 K/µL Passed
MCV 87.9 ±7.1 92.0 fL Passed
MPV 10.0 ±1.6 10.12 fL Passed
MCH 33.7 ±3.5 33.76 pg Passed
MCHC 38.3 ±6.1 38.3 g/dL Passed
RDW 13.7 ±4.6 12.9 pg Passed
Lot #77882
Normal Criteria Result Units Remarks
WBC 8.9 ±0.8 8.9 K/µL Passed
RBC 4.03 ±0.4 4.3 M/µL Passed
HGB 13.0 ±0.6 13.5 g/dL Passed
HCT 33.8 ±4.8 33.8 % Passed
PLT 237 ±32 247 K/µL Passed
MCV 84.0 ±6.8 89.3 fL Passed
MPV 9.9 ±1.8 10.25 fL Passed
MCH 32.3 ±3.2 31.7 pg Passed
MCHC 38.5 ± 8.5 37.12 g/dL Passed
RDW 14.5 ±4.2 12.9 pg Passed

Lot #77881
Low Criteria Result Units Remarks
WBC 3.2 ±0.5 3.2 K/µL Passed
RBC 2.29 ±0.3 2.31 M/µL Passed
HGB 6.9 ±0.6 6.9 g/dL Passed
HCT 18.3 ±4.0 19.9 % Passed
PLT 60 ±18 68 K/µL Passed
MCV 79.9 ±7 83.2 fL Passed
MPV 10.1 ±2.1 10.4 fL Passed
MCH 30.3 ±4.0 30.3 pg Passed
MCHC 37.9 ±6.5 37.9 g/dL Passed
RDW 15.6 ±5.0 12.9 pg Passed

Machine performance is verified using the manufacturer’s recommendation in validating the


functionality of the instrument.

Results: Haematology unit passed the tests based on the insert given by the
manufacturer.

Remarks:
Units are evaluated to be in good operational condition.
No major motor repair is required as actual performance still conforms to the standards set by the
manufacturer. System is up to date.

Performed By:

Johnmar P. Aquino
Service Engineer