Professional Documents
Culture Documents
Check-in
Initial Settings
X-ray detector is beyond 12 cm
CURRENT meter is lit (2nd operation panel), select EMISSION
fill in the log and record the Time in
Top Panel
DISPLAY MODE—
NORMAL engaged, CONTRAST and BRIGHTNESS fully
counterclockwise
IMAGE SELECTOR—INV off, SEI engaged, select Analog (under
AUX)
AFD—off
DZM—DUAL/ZOOM: off
PCD—PCD disengaged, select internal
SRT—SCAN ROTATION off and zero, TILTING CORRECT fully
counterclockwise
Stage Controls
iii. wait until the nitrogen flow stops, gently pull the front
cover away from the column until you have a clear access to
the stage
iv. slide the specimen holder onto the dovetail of the holder
mount on the stage
v. check to see if the o-ring on the chamber and mating
surface on the front cover are clean
vi. gently push the cover in to close the chamber and make
sure you do not pinch the silver cable on the right
vii. hold the chamber close and press the SPEC button on the
VACUUM SYSTEM control panel beneath the column to
evacuate specimen chamber
viii. the system is ready when the emission meter light comes
back on
Optional: for the 4-place specimen adapter, adjust the X
and/or Y stage controls by ± 5 units to bring one of the
specimen stubs under the electron beam
Column Conditions
1. adjust specimen WD based on the need for depth of focus or
resolution
Caution: DO NOT position specimen closer than 10 mm to the
final lens
2. adjust COARSE FOCUS so WD (see central LED display on
2nd Operation Panel) approximates the actual distance between
the specimen and final lens
3. select desired accelerating voltage (kV)
4. with GUN BIAS on MANUAL, adjust bias as posted
Caution: NEVER use bias 0; when in doubt, use the higher of the
two biases
Filament Saturation
1. set MAGNIFICATION to 200×
2. set PROBE CURRENT with coarse knob to 1 × 10-10 A (see central
LED display on 2nd Operation Panel)
3. SCAN MODE—engage RDC and LSP
4. SCAN SPEED—engage SR
5. DISPLAY MODE—set CONTRAST to ~1 o’clock and
increase BRIGHTNESS until a line profile is just visible on the CRT
6. adjust line scan profile to lower 1/3 of CRT with SE
IMAGE BRIGHTNESS
7. adjust SE IMAGE CONTRAST to 7 on dial (can adjust it later in
step 12 below to make the LSP easier to read)
8. set DETECTOR switch to SEI
9. press ACCEL VOLTAGE ON button
10. adjust metal stop on FILAMENT knob to 1 o’clock
11. watch the CURRENT METER and slowly (~ 5 s/div)
turn FILAMENT knob clockwise until the needle just registers
12. while observing line scan profile (LSP) on the display,
continue turning FILAMENT knob clockwise until LSP rises slightly
(~ 11 o'clock), reach the first peak (~ 12 o'clock)
Note: If the LSP peaks are too high or too noisy, reduce SE
IMAGE CONTRAST; otherwise increase to make the LSP easier to
read.
13. sharpen peaks of line scan profile
with COARSE and FINE FOCUS and, if necessary, adjust stage
controls to move sample beneath beam to create a line scan
profile with some amplitude
14. continue turning FILAMENT knob clockwise, the line scan
profile will fall and then rise again (~ 12:15) to a persistent
current point or saturation (~12:30)
Note: The saturation point is where you can turn back
the FILAMENT knob by just a tad and the LSP will drop.
15. beam current should be < 100 µA; if > 100 µA, turn
down FILAMENT, increase BIAS to reduce beam current and
resaturate
16. optimize height of LSP with GUN ALIGNMENT TILT knobs,
(X&Y) but DO NOT ADJUST GUN ALIGNMENT SHIFT KNOBS
17. for routine SE imaging, lower FILAMENT CURRENT below
saturation slightly (the effective saturation point); but DO NOT
reduce current for energy dispersive spectroscopy (EDS) or
backscattered imaging (BEI)
18. adjust METAL STOP of FILAMENT knob to touch pointer of
knob
19. lower SE IMAGE CONTRAST until all contrast LED’s are off
and adjust SE IMAGE BRIGHTNESS until 1st of the upper red trio
of LED’s is lit
640×480 80
1280×960 160
2560×1920 640
Appendix
Add spatial calibration
o DSG generated TIFF does not have any provision for
metadata e.g., display magnification or spatial calibration
o if the display magnification is known, spatial calibration can
be added manually or via simple macros in ImageJ
a macro that asks for the display magnification and
then sets the scale
another macro that sets the scale after parsing the
filename encoded with the display magnification
yet another macro that sets the scale after parsing the
filename encoded with the display magnification
recursively in a folder; beware that the macro will add
metadata and overwrite the original files
Stage tilt
o range decreases with larger stub size and shorter working
distance
o increase in tilt brings the top left edge of the specimen
toward the observer, example below shows the tilt range of
a 12 mm stub at 25× and WD 39 mm
0° tilt
15° tilt
30° tilt
45° tilt
60° tilt
75° tilt