You are on page 1of 8

The usefulness of a laser Doppler in the

measurement of toe blood pressures


Jurgen C. de Graaff, MD,a Dirk Th. Ubbink, MD, PhD,a Dink A. Legemate,
PhD, MD,a Rob J. de Haan, PhD,b and Michael J. H. M. Jacobs, MD, PhD,a
Amsterdam, The Netherlands

Objective: The purpose of this study was to evaluate the clinical value and reproducibil-
ity of laser Doppler (LD) versus photoplethysmography (PPG) in the measurement of
the systolic toe blood pressure.
Methods: Toe blood pressure was measured in 60 patients in different stages of periph-
eral vascular disease with simultaneous digital sampling of PPG and two LD signals,
each with a different filter setting (3 second [LD3] and 0.03 second [LD0.03]), and cuff
pressure. These measurements were repeated after 1 week. The signals were analyzed
with previous results ignored. The agreement of the PPG and LD pressures and repro-
ducibility after 1 week were assessed by calculating the intraclass correlation coefficient
(ICC). The agreement variation across the range of pressure values was visually explored
by means of difference plots.
Results: In 19 legs with a very low pressure only LD could adequately measure the pres-
sure, whereas PPG did not. The ICCs between PPG and LD3 and LD0.03 were 0.95 or
more. The ICCs of the 1-week reproducibility of the PPG, LD3, and LD0.03 pressures
were 0.92, 0.88, and 0.86, respectively. The variation was equally distributed across the
range of pressures in all three methods.
Conclusion: LD is a reliable alternative to PPG to measure toe blood pressures.
Furthermore, LD is able to measure low pressures, which is relevant in the assessment
of the presence of critical ischemia. (J Vasc Surg 2000;32:1172-9.)

Toe systolic blood pressure measurements play a diction of the need for amputation.7 The cutoff
prominent role in the diagnosis of critical leg value used for the diagnosis of critical leg ischemia as
ischemia, especially in diabetic patients, in whom the defined in the European Consensus8 is low (30 mm
ankle pressure is not always reliable because of media Hg). Therefore, the technique must be highly accu-
sclerosis.1,2 The toe pressure is also important in rate when used in the low-pressure range.
patients with extensive wounds in whom the ankle The principle of toe blood pressure is based on
cuffs or Doppler probes cannot be applied. In the method of arm blood pressure measurements as
patients with severe arterial insufficiency, toe pres- initially described by Riva-Rocci9; a pneumatic cuff
sure is correlated with the severity of the overall surrounding the limb is inflated to a pressure suffi-
occlusive process,3 the need for vascular interven- cient to stop blood flow.10 During slow deflation,
tion,4,5 the healing potential of ulcers,6 and the pre- the onset of resumption of blood flow distal to the
cuff is defined as the systolic pressure. Nowadays, toe
From the Departments of Vascular Surgerya and Clinical blood pressure is usually measured by either strain
Epidemiology and Biostatistics,b Academic Medical Center. gauge11 or photoplethysmography (PPG).12 Both
Competition of interest: nil. methods can be used in the alternating current (AC)
Supported by The Netherlands Heart Foundation, grant number
and direct current (DC) mode. In the AC mode the
96-113.
Reprint requests: Dirk Th. Ubbink, Department of Vascular resumption of blood is detected by the return of the
Surgery, Academic Medical Center, Meibergdreef 9, PO Box pulsatile signal, which is reduced, or even absent, in
22700, 1100 DE Amsterdam, The Netherlands (e-mail: patients with multiple occlusions as in patients with
D.Ubbink@amc.uva.nl). critical leg ischemia. In the absence of detectable
Copyright © 2000 by The Society for Vascular Surgery and The
pulsations, the plethysmograph can be DC coupled.
American Association for Vascular Surgery, a Chapter of the
International Society for Cardiovascular Surgery. The toe is exsanguinated with a tourniquet before
0741-5214/2000/$12.00 + 0 24/1/108009 inflation of the occluding cuff. Resumption of arte-
doi:10.1067/mva.2000.108009 rial inflow is then indicated by an upward shift from
1172
JOURNAL OF VASCULAR SURGERY
Volume 32, Number 6 de Graaff et al 1173

Fig 1. Scheme of probe application on the big toe and signal analysis. All three signals are plotted in
one graph for simplification, although signals were analyzed independently. The arrows represent the
points at which the systolic pressures were read.

the baseline.13 However, both methods have their Table I. Patient characteristics (n = 60)
shortcomings. The strain gauge is difficult to posi-
Smoking 52%
tion at short digits, and the element might exert Diabetes mellitus 36%
some pressure on the toe, which influences mea- History of cerebrovascular accident 17%
surement accuracy at low pressures,14 whereas the or temporary ischemic attack
History of cardiac failure 41%
PPG in the DC mode is very sensitive to movement Hypertension 48%
artifacts and not commonly used. Hypercholesterolemia 41%
Many methods, such as clearance of radioiso- Presence of peripheral pulsations 69% (of 112 legs)
(dorsalis pedis or posterior tibial
topes,15 pink flush reappearance,3 audio PPG,16 pulse artery)
oximetry,17 and impedance sphygmography18 have
been used for the measurement of toe pressures with
varying success. More recently, the laser Doppler (LD)
technique has been suggested for this purpose.14,19 Furthermore, LD tended to be more sensitive than
LD has also been used as an alternative technique for strain gauge at low pressures. However, the LD tech-
PPG in the measurement of skin perfusion pres- nique has not yet been compared with PPG, which is
sures.20,21 LD has theoretical advantages over PPG, mostly used nowadays, in the same population.
because it is able to detect the minute microcirculato- In this study we compared two techniques (PPG
ry blood perfusion in the skin. PPG is based on and LD) for the measurement of toe blood pressure
changes in the amount of reflected near infrared light measurements as to their use, reliability, and respon-
on pulsatile changes in blood filling of the digit, siveness in patients with different stages of peripher-
whereas LD is based on the change in frequency shift al vascular insufficiency. Additionally, we evaluated
of the reflected light caused by moving blood particles. the 1-week intraobserver reproducibility to evaluate
Furthermore, LD is readily applied without compres- consistency of both methods.
sion, even to short digits. Andersson et al14 compared
LD with the strain gauge method and concluded that Patients and Methods
the LD technique is an alternative and is complemen- Patients. Sixty patients (37 men, 23 women;
tary to the strain gauge method in damaged or ulcer- mean age, 66 years; SD, 10 years) referred to the vas-
ating toes and feet. Beinder et al19 used LD in combi- cular laboratory for the assessment of the
nation with a specially manufactured transparent plas- ankle/brachial index (ABI) consented to participate
tic capsule for positioning the LD probe and cuff to in this study. Patient characteristics are listed in Table
the digit. The toe pressures obtained were in good I. In eight patients only one toe could be investigat-
agreement with data obtained by means of the strain ed, because in four patients the toe or foot was ampu-
gauge method measured in another group of patients. tated, while in the remaining four patients it was not
JOURNAL OF VASCULAR SURGERY
1174 de Graaff et al December 2000

Fig 3. Difference between LD3 pressure and LD0.03 pres-


B sure at the first measurement plotted against the mean of
both methods. The regression line of the differences
(drawn line) and 95% limits of agreement (dotted lines) of
the differences are presented as horizontal lines. The ver-
tical dashed line represents the commonly accepted cutoff
value for critical leg ischemia (< 30 mm Hg). The open
squares represent the pressures that could only be mea-
sured with LD.

in the supine position, which implies that, depending


on the size and position of the foot, the toe was at or
just above heart level during the measurement, which
is the usual position for these measurements. The
legs were covered with sheets or clothing to prevent
cooling. Because of these preparations, we refrained
Fig 2. Difference between pressure measured with PPG from local heating of the toe, which is sometimes
and with the LD setting that detects slow changes in the advocated in PPG studies. The toe pressures were
blood flow (LD3) (A) and the LD setting that could trace simultaneously measured with PPG and LD.
heartbeats (LD0.03) (B) plotted against the mean pressure. The PPG was AC coupled (filter settings, 0.4-2.3
The regression line of the differences (drawn line) and Hz). A PPG consists of an infrared light-emitting
95% limits of agreement (dotted lines) of the differences diode and a photosensor mounted adjacent to one
are presented as horizontal lines. The open squares repre- another on a small probe. Blood, which is more
sent the pressures that can only be measured with LD. opaque than surrounding tissue, attenuates light in
proportion to its content in the tissue. The speed dif-
ferences in the flow causing variations in the amount
of blood content in the digit result in a pulse con-
possible to measure the toe blood pressure with both tour, which is normally presented on a graph.
techniques at the same time because of large wounds LD fluxmetry is a simple, noninvasive sensitive
(n = 3) or the shortness of the toe (n = 1). The technique used to assess cutaneous blood perfu-
remaining 112 legs belonged to different stages of sion.22-25 In the LD instrument, light with a wave-
peripheral vascular disease; there were 20 legs without length of 780 nm is conducted through optical
complaints, 54 with intermittent claudication, and 38 fibers to an area of the skin of about 0.5 mm2 where
with rest pain, ulcers, or gangrene. Patients were accli- it penetrates the skin to a depth of 1 to 1.5 mm and
matized in a temperature-controlled environment (23 is partly reflected. This light, if backscattered by
± 1°C) during a resting period of at least 15 minutes. moving objects (principally erythrocytes), under-
Measurements. Measurements were performed goes a frequency shift, which is proportional to the
JOURNAL OF VASCULAR SURGERY
Volume 32, Number 6 de Graaff et al 1175

Table II. Mean blood pressures


Toe pressure mm Hg (SD) AP ABI
n PPG LD3 LD0.03 mm Hg (SD) % (SD)

Without complaints 20 100 (32) 101 (34) 98 (31) 153 (31) 99 (15)
Intermittent claudication 54 69 (28) 70 (26) 69 (27) 108 (31) 72 (20)
Rest pain, ulcers or gangrene 38 20 (21) 34 (14) 33 (15) 73 (26) 48 (19)
Overall 112 58 (40) 63 (34) 62 (34) 104 (40) 69 (26)

Mean toe blood pressure (measured with PPG, LD3, LD0.03), AP, and ABI in the legs measured at the first measurement.

Table III. Difference between PPG and LD pressures


n Mean difference SD of the difference Limits of agreement P value ICC

PPG vs LD3 (mm Hg) 93 –0.8 9.5 –19.7/18.1 .43 0.96


PPG vs LD0.03 (mm Hg) 91 –0.9 9.8 –20.4/18.6 .36 0.95
LD3 vs LD0.03 (mm Hg) 110 –0.6 8.6 –17.8/16.6 .47 0.97

Comparison between LD3, LD0.03, and PPG at the first measurement with mean difference and SD of the differences and limits of agree-
ment, P value of the Student t test for paired samples, and ICC. The limits of agreement reflect the interval within 95% of the differ-
ences in which the pressures lie, defined as the mean difference ± 2 SD of the difference. Unmeasurable pressures were excluded.

velocity, number, and direction of moving objects notch for the probe (Double-Stick Discs, 3M Health
(LD flux) and is expressed in volts. Two filter set- Care, Borken, Germany) and a probe holder (PF
tings of the partly output signal (time constants) of 104, Perimed) for the LD probes (Fig 1). The probe
the LD instrument were used: 3 seconds (LD3) for holder is a small plastic holder (diameter, 15 mm) in
the detection of slow changes in blood flow (which which the LD probe fits exactly and in which the tip
can be compared with the DC mode), and 0.03 sec- of the probe is just above the skin (approximately
onds (LD0.03), which results in a signal that could 0.1 mm Hg). The exact distance from the skin is
trace heartbeats (which can be compared with the achieved without compressing the skin by means of
AC mode). the probe holder and adhesive tape.
Blood pressure measurements. The ankle systolic Our conventional instrument to measure periph-
pressures in the dorsal pedal and posterior tibial arter- eral pressures (PV-lab, Stöpler, EDI, Burbank, Calif)
ies at the level of the ankle were measured using an 8- was modified so that the cuff pressure, PPG, and LD
MHz Doppler probe and a cuff (12 cm) around the signals could be sampled on-line and analyzed off-
lower leg just above the ankle. The highest pressure line by means of a data acquisition system
was considered to represent the highest perfusion (AcqKnowledge III and MP 100WSW, Biopac
pressure. The ABI was calculated by dividing the ankle System, Inc, Santa Barbara, Calif). The cuff was
pressures by the highest of the left and right brachial rapidly inflated up to a pressure at least 30 mm Hg
blood pressure, measured by means of an automatic higher than that at which the pulsations in the PPG
blood pressure monitor (Criticon, Dinamap Plus, and LD0.03 signal disappeared. Initially, the LD
Tampa, Fla) and expressed as a percentage. showed an increase in flux after which the signal
For the toe blood pressure, a cuff with a width declined to the baseline (Fig 1). After the LD signal
depending on the diameter of the hallux (1.5, 2.5, had reached a stable baseline (after approximately 3
or 3.3 cm disposable cuffs, Hokanson, Bellevue, seconds), the cuff was slowly deflated with a speed of
Wash) was wrapped around the base of the toe. The approximately 3 mm Hg/s. The systolic pressure
cuff width closest to 120% of the diameter of the was defined as the pressure at which the (pulsatile)
hallux was chosen.26 The PPG and two standard LD signal reappeared or rose gradually from the baseline
probes (PF 408, Perimed, Järfälla, Sweden) con- (Fig 1).
nected to the LD instrument (Periflux 4001, One single measurement of the toe blood pres-
Perimed) were attached to the apex of the big toe sure (with all three probes at the same time) was per-
(PPG at the center and LD probes at both sites of formed each session. The measurements were repeat-
the PPG) with double-sided adhesion tape with a ed after 1 week to establish the 1-week intraobserver
JOURNAL OF VASCULAR SURGERY
1176 de Graaff et al December 2000

A reproducibility. All measurements and analyses were


performed by one investigator (JdG), while previous
results were ignored as much as possible.
Statistical analysis. The differences between
PPG versus LD3 and PPG versus LD0.03 pressures
were evaluated by the Student t test for paired sam-
ples. Because we do not know the true pressure, the
average of the LD3 and PPG and the average of
LD0.03 and PPG were considered to be the best esti-
mates of the true pressures. The differences between
LD and PPG were also calculated with limits of
agreement and intraclass correlation coefficients
(ICCs). The limits of agreement were defined as the
mean difference ± 2 SD of the difference, according
to Bland and Altman.27,28 These limits reflect the
interval in which 95% of the differences in pressures
lie. A plot was made of the difference between the
B two measurements against their mean to check
whether the measurement error was independent of
the magnitude of pressure. The regression lines
(with β coefficient, indicating the direction of the
deviation and its P value) indicate whether the vari-
ability in the pressures is independent of the magni-
tude of the pressures. The ICC (two-way random
effects model) was calculated according to the
method as described by Deyo et al.29 The ICC not
only assesses the strength of linear correlation
between two measurements, such as the Pearson
correlation, but also detects systematic error.30
Thus, if one measurement is systematically higher or
lower than the other, the ICC is correspondingly
reduced. The same methods were used to assess the
C 1-week intraobserver reproducibility and the com-
parison between LD3 and LD0.03.

RESULTS
The pressure could not be measured with PPG in
19 toes because no pulsatile signal could be detect-
ed (taken as 0 mm Hg), whereas all of these toe
pressures could be measured with LD3 (mean, 24 ±
14 mm Hg; range, 8-64 mm Hg; number of pres-
sures ≥ 30, 5), and 17 of these with LD0.03 (mean,
21 ± 12 mm Hg; range, 0-48 mm Hg; number of
pressures ≥ 30, 5). Of these patients, one presented
with intermittent claudication, 11 with rest pain,
and 7 with ulcers or gangrene. The mean ankle pres-
sure (AP) and ABI were 64 ± 32 mm Hg (range 0-
1 40 mm Hg) and 41% ± 21% (range, 0%-104%),
Fig 4. Difference between the first and second measure-
respectively. The incidence of diabetes mellitus
ment of PPG pressure (A), LD3 pressure (B), and LD0.03
pressure (C) plotted against the mean of the first and sec- among these patients did not differ significantly
ond measurement. The regression line of the differences from the total patient group (39% vs 36%, respec-
(drawn line) and 95% limits of agreement (dotted lines) of tively). Strikingly, in all but one limb, no peripheral
the differences are presented as horizontal lines. pulsations could be palpated. The pressure of 64
JOURNAL OF VASCULAR SURGERY
Volume 32, Number 6 de Graaff et al 1177

Table IV. One-week reproducibility


Mean Mean Mean SD of the Limits of
n = 86 legs t = 0 (SD) t = 1 (SD) difference differences agreement P value ICC

BP (mm Hg) 151 (23) 145 (19) 7.2 16.1 –25/39 < .01 0.72
AP (mm Hg) 107 (38) 101 (38) 5.9 20.4 –34/46 .01 0.85
ABI (%) 68 (23) 67 (25) 0.7 11.3 –22/23 .57 0.89
PPG (mm Hg) 59 (39) 55 (37) 3.3 16.6 –29/36 .04 0.92
LD3 (mm Hg) 62 (34) 58 (31) 3.8 15.6 –27/34 .03 0.88
LD0.03 (mm Hg) 62 (33) 57 (32) 4.9 14.7 –24/34 .03 0.86

Mean BP, AP, ABI, and toe (PPG, LD3, and LD0.03) pressure with SD, mean difference with SD and limits of agreement, P value of the
Student t test for paired samples, and ICC of the 1-week reproducibility of each technique.
BP, Brachial pressure; t = 0, at the frist measurement; t = 1, 1 week later.

mm Hg, undetected by means of PPG but detected (Table IV). The ICCs indicate a substantial repro-
by LD, was found in a patient with CREST syn- ducibility for all three methods (≥ 0.85). In Fig 4 the
drome with severe microangiopathy without severe differences between the first and the second PPG
arterial occlusions. and LD pressures are plotted against the mean of the
Application to the toe of the LD probe using the first and second measurement. The error of the mea-
probe holders was very simple. Except for one very surement was found to be independent of the mag-
short big toe, the two LD probes could be applied nitude of the pressure values (β coefficient Fig 4, A
next to the PPG probe on the big toe in all patients = .1, P = .42; β coefficient Fig 4, B = .15, P = .12; β
investigated. The LD reading was quite stable and coefficient Fig 4, C = .05, P = .67).
less influenced by movement artifacts than the PPG
signal. The signal-noise ratio was found to be con- DISCUSSION
siderably lower with the LD than with the PPG sig- This study shows that LD is a reliable and clini-
nals. The blood pressures per patient subgroup are cally useful technique for the measurement of systolic
presented in Table II. toe blood pressures. Moreover, LD is more respon-
PPG versus LD. Histograms depicting the dif- sive in detecting the reappearance of skin flow at very
ferences between PPG and LD pressures showed a low perfusion pressures than PPG in the AC mode.
normal distribution (graphs not shown). Table III Andersson et al14 already showed that LD seems
presents the mean difference, the SD of the differ- to be more sensitive than the strain gauge method in
ences, the limits of agreements, and the ICC the low-pressure range for measuring blood pressures.
between PPG pressure and LD pressure. The mean Although also critically reduced pressures could be
differences were small and statistically not signifi- measured with PPG (in our study as low as 15 mm
cant. The ICCs showed almost perfect agreement (≥ Hg), this study showed that LD was able to measure
0.95) between the PPG pressure and LD pressure toe blood pressures in 19 toes, which the PPG in the
(Table III). AC mode could not; these pressures varied between 6
In Fig 2 the differences between PPG and LD and 64 mm Hg. This pressure range is important,
pressures are plotted against the mean of both mea- because it is around the cutoff value for critical leg
surements. The regression lines in both figures indi- ischemia as proposed by the European Consensus (30
cate that the variability in the pressure scores showed mm Hg). This means that the decision for invasive
a small relation with only the magnitude of the therapy could be influenced by the method used. A
LD0.03 pressure values (β coefficient Fig 2, A = .15, low value (eg, 0 mm Hg) would suggest that invasive
P = .14; β coefficient Fig 2, B = .28, P < .01). There treatment is required, whereas a higher value would
was no significant difference between LD3 and advocate initial conservative therapy.
LD0.03 (Table III; β coefficient Fig 3 = .05, P = .6). The LD technique appears to be most valuable in
Reproducibility. For the analysis of the repro- the absence of peripheral pulsations, and in patients
ducibility, only 86 legs were eligible because the with low ankle and toe pressures. On theoretical
other patients were treated within that week. There grounds, LD is probably more sensitive than PPG
was a small but significantly lower brachial, toe, and because LD is based on the registration of a fre-
ankle pressure measured after 1 week, whereas the quency shift (Doppler effect) of red light by the
ABI remained stable during the 1-week time interval moving erythrocytes, whereas PPG is based on a dif-
JOURNAL OF VASCULAR SURGERY
1178 de Graaff et al December 2000

ference in the amount of reflected light by the (LD0.03) is easier with pulsatile flux at higher pres-
absorption of near infrared light by the presence of sures, and the more DC (plethysmographic) mode
erythrocytes in the underlying tissue. Thus, LD is (LD3) is preferred to detect flux increase from base-
able to detect very small blood displacements and is line at lower pressures not showing any pulsatile
very sensitive in detecting the resumption of micro- flux. A further improvement of the LD and PPG
circulatory blood flow in the toe, as opposed to the techniques might be the use of heated probes or the
more crude detection of blood filling by PPG. application of local heat, because the hyperemia
Therefore, we believe that LD has an advantage over increases the total skin perfusion, which pronounces
PPG in establishing critical limb ischemia because reappearance of the skin perfusion.
the absence of PPG pulsatility, which has even been Analysis of the reproducibility indicates that all
suggested to predict wound healing,31 does not have three methods are comparable and reproducible
to be concomitant with a critically reduced perfusion (ICC > 0.85). Insight in the reproducibility of a
pressure. On the other hand, LD has not been com- technique, especially the limits of agreement, is
pared with the PPG in the DC mode, which is the important when interpreting the results of measure-
most accurate method of detecting toe blood pres- ments of the same person within time (eg, in the
sure using PPG.13 But the DC mode appears to be evaluation of therapy or progress of a disease). The
commonly used less frequently, probably because limits of agreement of the ABI are somewhat above
this mode is very sensitive to movement artifacts and 0.2, which is in agreement with the results of Fisher
the toe has to be exsanguinated before inflation of et al,33 but is more than the 0.15, which is general-
the cuff. Moreover, LD has advantages over the ly reported.10,34 This somewhat larger variation is
PPG in DC mode, because the LD is hardly influ- mainly caused by the way of calculating the variabil-
enced by movement artifacts and exsanguination of ity. The significantly lower ankle pressure is due to a
the toe is not required. lower systemic pressure after 1 week, probably
Another advantage of LD is that the LD probe because of reassurance of the patient during the sec-
can easily be applied to short digits, especially when ond measurement, which is represented by a lower
much smaller probes (5 mm) or the transparent brachial pressure after 1 week.
probe holder as described by Beinder et al19 are In conclusion, LD is a responsive and reliable
used. Thus, in practice even digit II, III, or IV might instrument to measure toe blood pressure as com-
be used for toe blood pressure measurements, when pared with routinely used techniques. We recom-
the big toe is not available because of ulcerations or mend the use of LD in situations in which no signal
after amputation. Furthermore, toe pressures mea- can be detected with PPG and a decision is required
sured with LD are found to be as reproducible as as to the presence of critical leg ischemia.
pressures measured with PPG. The coefficients of
REFERENCES
repeatability are in agreement with other techniques
1. Vincent DG, Salles-Cunha SX, Bernhard VM, Towne JB.
such as strain gauge plethysmography.32
Noninvasive assessment of toe systolic pressures with special ref-
On the other hand, the disadvantage of the LD erence to diabetes mellitus. J Cardiovasc Surg 1983;24:22-8.
method is that three devices (LD, cuff inflator, and 2. Goss DE, de TJ, Roberts VC, Flynn MD, Edmonds, ME, et
data acquisition system) are needed instead of one, al. Raised ankle/brachial pressure index in insulin-treated
because the LD instrument is not routinely integrated diabetic patients. Diabet Med 1989;6:576-8.
3. Carter SA, Lezack JD. Digital systolic pressures in the lower
in the peripheral vascular equipment, which increases
limb in arterial disease. Circulation 1971;43:905-14.
the costs. A data acquisition system may not be 4. Ubbink DT, Tulevski II, den HD, Koelemay MJ, Legemate
required, but in our experience, the first pulsations are DA, Jacobs MJ. The value of non-invasive techniques for the
often missed during on-line determination with our assessment of critical limb ischaemia. Eur J Vasc Endovasc
conventional laboratory instrument (resulting in a Surg 1997;13:296-300.
lower pressure), whereas this will not occur when the 5. Vayssariat M, Gouny P, Cheynel C, Gaitz JP, Baudot N,
Nussaume O. Haemodynamics of patients with severe lower
analysis is performed off-line. limb arterial disease: the critical aspects of critical ischaemia.
The filter time for the LD signal did not seem to Eur J Vasc Endovasc Surg 1997;14:284-9.
influence the results of the pressure measurement, 6. Ramsey DE, Manke DA, Sumner DS. Toe blood pressure: a
although the longer time constant (LD3) tends to be valuable adjunct to ankle pressure measurement for assessing
more sensitive. Both filter settings methods (LD3 peripheral arterial disease. J Cardiovasc Surg 1983;24:43-8.
7. Paaske WP, Tonnesen KH. Prognostic significance of distal
and LD0.03) yield comparative results as to PPG, and blood pressure measurements in patients with severe
the agreement between both methods and PPG is ischaemia. Scand J Thorac Cardiovasc Surg 1980;14:105-8.
almost perfect (ICC > 0.95). The faster mode 8. Anonymous. Second European Consensus Document on
JOURNAL OF VASCULAR SURGERY
Volume 32, Number 6 de Graaff et al 1179

chronic critical leg ischemia. Eur J Vasc Surg 1992;6 Suppl 22. Stern MD, Lappe DL, Bowen PD, Chimosky JE, Holloway
A:1-32. GA, Jr, et al. Continuous measurement of tissue blood flow by
9. Riva-Rocci S. Un nuovo sfingmomanometro. Gaz Med laser-Doppler spectroscopy. Am J Physiol 1977;232:H441-8
Torino 1896;47:981. 23. Watkins D, Holloway GA, Jr. An instrument to measure cuta-
10. Carter SA. Role of pressure measurements. In: Bernstein EF, neous blood flow using the Doppler shift of laser light. IEEE
editor. Vascular diagnosis. 4th ed. St Louis: Mosby; 1994. p. Trans Biomed Eng 1978;25:28-33.
486-512. 24. Stern MD. In vivo evaluation of microcirculation by coherent
11. Strandness DE Jr, Bell JW. Diagnosis and objective evalua- light scattering. Nature 1975;254:56-8.
tion of digital systolic blood pressure by strain-gauge tech- 25. Ubbink DT, Jacobs MJ, Tangelder GJ, Slaaf DW, Reneman
nique. Ann Surg 1965;161 Suppl 4:3. RS. Posturally induced microvascular constriction in patients
12. Yao JS, Flinn WR, Bergan JJ. Noninvasive vascular diagnos- with different stages of leg ischaemia: effect of local skin heat-
tic testing: techniques and clinical applications. Prog ing. Clin Sci 1991;81:43-9.
Cardiovasc Dis 1984;26:459-94. 26. Gundersen J. Segmental measurements of systolic blood
13. Bone GE, Pomajzl MJ. Toe blood pressure by photoplethys- pressure in the extremities including the thumb and the great
mography: an index of healing in forefoot amputation. toe. Acta Chir Scand Suppl 1972;426:1-90.
Surgery 1981;89:569-74. 27. Bland JM, Altman DG. Statistical methods for assessing
14. Andersson S, Linderholm H, Rinnstrom O, Burlin L. A laser agreement between two methods of clinical measurement.
Doppler technique for measuring distal blood-pressure: a Lancet 1986;1:307-10.
comparison with conventional strain-gauge technique. Clin 28. Bland JM, Altman DG. Comparing methods of measure-
Physiol 1986;6:329-35. ment: why plotting difference against standard method is
15. Lassen NA, Tvedegaard E, Jeppesen FI, Nielsen PE, Bell, misleading. Lancet 1995;346:1085-7.
Gundersen J. Distal blood pressure measurement in occlusive 29. Deyo RA, Diehr P, Patrick DL. Reproducibility and respon-
arterial disease, strain gauge compared to Xenon-133. siveness of health status measures: statistics and strategies for
Angiology 1972;23:211-7. evaluation. Control Clin Trials 1991;12:142S-58S.
16. Fronek A, Blazek V, Curran B. Toe pressure determination by 30. Kirshner B, Guyatt G. A methodological framework for
audiophotoplethysmography. J Vasc Surg 1994;20:267-70. assessing health indices. J Chronic Dis 1985;38:27-36.
17. Samuelsson P, Blohme G, Fowelin J, Eriksson JW. A new 31. Lee BY, Trainor FS, Kavner D, Crisologo JA, Shaw WW,
non-invasive method using pulse oximetry for the assessment Madden JL. Assessment of the healing potentials of ulcers of
of arterial toe pressure. Clin Physiol 1996;16:463-7. the skin by photoplethysmography. Surg Gynecol Obstet
18. Beyenal-Ogmen B. Bioelectrical impedance sphygmography 1979;148:233-9.
for hydrostatic toe pressure measurement: a new non-invasive 32. Nielsen PE. Digital blood pressure in patients with peripher-
method to assess limb ischaemia. Vasa 1999;28:15-8. al arterial disease. Scand J Clin Lab Invest 1976;36:731-7.
19. Beinder E, Hoffmann U, Franzeck UK, Huch A, Huch R, 33. Fisher CM, Burnett A, Makeham V, Kidd J, Glasson M,
Bollinger A. Laser Doppler technique for the measurement of Harris JP. Variation in measurement of ankle-brachial pressure
digital and segmental systolic blood pressure. Vasa 1992;21: index in routine clinical practice. J Vasc Surg 1996;24:871-5.
15-21. 34. Baker JD, Dix DE. Variability of Doppler ankle pressures with
20. Malvezzi L, Castronuovo JJ Jr, Swayne LC, Cone D, Trivino arterial occlusive disease: an evaluation of ankle index and
JZ. The correlation between three methods of skin perfusion brachial-ankle pressure gradient. Surgery 1981;89:134-7.
pressure measurement: radionuclide washout, laser Doppler
flow, and photoplethysmography. J Vasc Surg 1992;15:823-9.
21. Castronuovo JJ Jr, Adera HM, Smiell JM, Price RM. Skin
perfusion pressure measurement is valuable in the diagnosis
of critical limb ischemia. J Vasc Surg 1997;26:629-37. Submitted Dec 17, 1999; accepted Mar 14, 2000.

You might also like