You are on page 1of 29

Downloaded from http://jicru.oxfordjournals.

org/ at Universite Laval on July 9, 2016


ICRU REPORT 29

ON RADIATION UNITS
INTERNATIONAL COMMISSION

AND MEASUREMENTS
ICRU REPORT 29

Dose Specification for


Reporting External Beam
Therapy with Photons

Downloaded from http://jicru.oxfordjournals.org/ at Universite Laval on July 9, 2016


and Electrons

Issued April 1 , 1978

INTERNATIONAL COMMISSION ON RADIATION


UNITS AND MEASUREMENTS
7910 WOODMONT AVENUE
WASHINGTON, D.C. 20014
U.S.A.
THE INTERNATIONAL COMMISSION
ON RADIATION UNITS AND MEASUREMENTS
INDIVIDUALS PARTICIPATING IN THE PREPARATION OF THIS REPORT
Commission Membership During Preparation of Commission Sponsor
This Report A. W AMBERSIE
H. 0. WYCKOFF, Chairman UCL, Cliniques Universitaires St. Luc
A. ALLISY, Vice Chairman Brussels, Belgium
K. LIDEN, Secretary
R. S. CASWELL
H. J. DUNSTER Report Committee
P. EDHOLM
T. LANDBERG, Chairman
J. R. GREENING University Hospital
D.HARDER Lund, Sweden
P. HARPER P.ALMOND

Downloaded from http://jicru.oxfordjournals.org/ at Universite Laval on July 9, 2016


A. KELLERER M. D. Anderson Hospital
H. H. Rossi Houston, U.S.A.
J.M. V. BURGERS
W. K. SINCLAIR
Antoni van Leeuwenhoek Ziekenhuis
A. TSUYA Amsterdam, Netherlands
A. W AMBERSIE M. BUSCH
L. S. TAYLOR, Honorary Chairman and Member Strahlenklinik der Gesamthochschule
Emeritus Essen, Germany
C. A. JOSLIN
Current Commission Membership Cookridge Hospital
H. 0. WYCKOFF, Chairman Leeds, United Kingdom
.; A. ALLISY, Vice Chairman J.P. PAUNIER
,; G. E. D. ADAMS Hopital Cantonal
-,, R. CASWELL Geneva, Switzerland
.I G. COWPER

./ P. EDHOLM
.I J. R. GREENING
D.HARDER
Consultants Report Committee
/ A. KELLERER
1 H. H. Rossi M. COHEN
J W. K. SINCLAIR McGill University
J. VAN DER SCHOOT Montreal, Canada
A. W AMBERSIE A. DUTREIX
L. S. TAYLOR, Honorary Chairman and Member Institut Gustave-Roussy
Emeritus Villejuif, France
T. R. MOLLER
Technical Secretary University Hospital
W.ROGERNEY Lund, Sweden
The Commission wishes to express its appreciation to the individuals, and their organizations, involved in the
preparation of this report for the time and effort they devoted to this task and to the organizations with which
they are affiliated.

Copyright © International Commission on


Radiation Units and Measurements 1978

Library of Congress Catalog Card Number 78-59476


International Standard Book Number 0-913394-23-8

(For detailed information on the availability of this and other ICRU Reports see page 19
Preface

Scope oflCRU Activities a current problem may seem advisable. Generally


speaking, however, the Commission feels that action
The International Commission on Radiation Units based on expediency is inadvisable from a long-term
and Measurements (ICRU), since its inception in 1925, viewpoint; it endeavors to base its decisions on the
has had as its principal objective the development of long-range advantages to be expected.

Downloaded from http://jicru.oxfordjournals.org/ at Universite Laval on July 9, 2016


internationally acceptable recommendations regard- The ICRU invites and welcomes constructive com-
ing: ments and suggestions regarding its recommendations
(1) Quantities and units ofradiation and radioactiv- and reports. These may be transmitted to the Chair-
ity, man.
(2) Procedures suitable for the measurement and Current Program
application of these quantities in clinical radiology and
radio biology, The Commission has divided its field of interest into
(3) Physical data needed in the application of these twelve technical areas and has assigned one or more
procedures, the use of which tends to assure uniform- members of the Commission the responsibility for
ity in reporting. identification of potential topics for new ICRU activ-
The Commission also considers and makes similar ities in each area. A body of consultants has been
types of recommendations for the radiation protection constituted for each technical area to advise the Com-
field. In this connection, its work is carried out in close mission on the need for ICRU recommendations re-
cooperation with the International Commission on lating to the technical area and on the means for
Radiological Protection (ICRP). meeting an identified need. Each area is reviewed
periodically by its sponsors and consultants. Recom-
mendations of such groups for new reports are then
Policy reviewed by the Commission and a priority assijroed.
The Technical areas are:
The ICRU endeavors to collect and evaluate the Radiation Therapy
latest data and information pertine.nt to the problems Radiation Diagnosis
of radiation measurement and dosimetry and to rec- Nuclear Medicine
ommend the most acceptable values and techniques Radiobiology
Radioactivity
for current use. · Radiation Physics-X Rays, Gamma Rays and Electrons
The Commission's recommendations are kept under Radiation Physics-Neutrons and Heavy Particles
continual review in order to keep abreast of the rapidly Radiation Protection·
expanding uses of radiation. Radiation Chemistry
The ICRU feels it is the responsibility of national Values of Factors- W, S, etc.
Theoretical Aspects
organizations to introduce their own detailed technical Quantities and Units
procedures for the development and maintenance of
standards. However, it urges that all countries adhere The actual preparation of ICRU reports is carried
as closely as possible to the internationally recom- out by ICRU report committees. One or more Com-
mended basic concepts of radiation quantities and mission members serve as sponsors to each committee
units. and provide close liaison with the Commission. The
The Commission feels that its responsibility lies in currently active report committees are:
developing a system of quantities and units having the Average Energy Reqmred to Produce an Ion Pair
widest possible range of applicability. Situations may C;.. and CE
arise from time to time when an expedient solution of Computer Uses in Radiotherapy
iv Preface

Definitions and Terminology for Computed Tomography measurements and, in turn, the WHO assists in the
Dose Specification for Reporting Intracavitary and Inter- world-wide dissemination of the Commission's rec-
stitial Tht:rapy ommendations. In 1960 the ICRU entered into con-
Dosimetry of Pulsed Radiation
sultative status with the International Atomic Energy
Fundamental Quantities and Units
High-Energy Electron Beam Dosimetry Agency. The Commission has a formal relationship
Measurement of Low-Level Radioactivity in Humans with the United Nations Scientific Committee on the
Methods of Assessment of Dose in Tracer Investigations Effects of Atomic Radiation (UNSCEAR), whereby
Microdosimetry ICRU observers are invited to attend UNSCEAR
Photographic Dosimetry in External Beam Therapy meetings. The Commission and the International Or-
Radiobiological Dosimetry
Scanning ganization for Standardization (ISO) informally ex-
Stopping Power change notifications of meetings and the ICRU is
formally designated for liaison with two of the ISO
ICRU Reports Technical Committees. The ICRU also corresponds
and exchanges final reports with the following orga-
In 1962 the ICRU, in recognition of the fact that its nizations:
triennial reports were becoming too extensive and in
some cases too specialized to justify single-volume Bureau International des Poids et Mesures
publication, initiated the publication of a series of Commission of the European Communities

Downloaded from http://jicru.oxfordjournals.org/ at Universite Laval on July 9, 2016


reports, each dealing with a limited range of topics. Council for International Organizations of Medical Sci-
ences
This series was initiated with the publication of six Food and Agriculture Organization
reports: International Council of Scientific Unions
International Electrotechnical Commission
ICRU Report lOa, Radiation Quantities anrl. Units International Labor Office
ICRU Report lOb, Physical Aspects of Irradiation
International Radiation Protection Association
ICRU Report lOc, Radioactivity
International Union of Pure and Applied Physics
ICRU Report lOd, Clinical Dosimetry
United Nations Educational, Scientific and Cultural Or-
ICRU Report lOe, Radiobiological Dosimetry
ganization
ICRU Report !Of, Methods of Evaluating Radiological
Equipment and Materials
The Commission has found its relationship with all
These reports were published, as had been many of of these organizations fruitful and of substantial bene-
the previous reports of the Commission, by the United fit to the ICRU program. Relations with these other
States Government Printing Office as Handbooks of international bodies do not affect the basic affiliation
th.e _National Bureau of Standards. qf the ICRU with the International Society of Radiol-
In 1967 the Commission determined that in the ogy.
future the recommendations formulated by the ICRU
would be published by the Commission itself. This
report is published by the ICRU pursuant to this Operating Funds
policy. With the exception of ICRU Report lOa, the
other reports of the "10" series have continuing valid- In the early days of its existence, the ICRU operated
ity and, since, except in the case of ICRU Report lOe, essentially on a voluntary basis, with the travel and
none of the reports now in preparation is designed o!)erating costs being borne by the parent organiza-
specifically to supersede them, they will remain avail- tions of the participants. (Only token assistance was
able until the material is essentially obsolete. All fu- originally available from the International Society of
ture reports of the Commission, however, will be pub- Radiology.) Recogruzing the impracticability of con-
lished under the ICRU's own auspices. Information tinuing this mode of operation on an indefinite basis,
about the availability of ICRU Reports is given on operating funds were sought from various sources.
page 18. Financial support has been received from the follow-
ing organizations:
ICRU's Relationships With Other Organizations
B.A.T. Cigaretten-Fabriken GMBH
In addition to its close relationship with the Inter- Commission of the European Communities
national Commission on Radiological Protection, the Council for International Organizations of Medical
ICRU has developed relationships with other organi- Sciences
zations interested in the problems of radiation quan- Eastman Kodak Company
tities, units and measurements. Since 1955, the ICRU E. I. duPont de Nemours and Company
Ford Foundation
has had an official relationship with the World Health General Electric Company
Organization (WHO) whereby the ICRU is looked to International Atomic Energy Agency
for primary guidance in matters of radiation units and International Radiation Protection Association
Preface • v

International Society of Radiology In recognition of the fact that its work is made
Japan Industries Association of Radiation Apparatus possible by the generous support provided by these
John och Augusta Perssons stiftelse organizations, the Commission expresses its deep ap-
National Cancer Institute of the U.S. Department of
Health, Education and Welfare preciation.
N.V. Philips Gloeilampenfabrieken HAROLD 0. WYCKOFF,
Picker Corporation Chairman, ICRU
Radiological Society of North America Washington, D.C.
Rockefeller Foundation 1March1978
Siemens Corporation
Society of Nuclear Medicine
Statens laegevidenskabelige Forskningsrad
U.S. Bureau of Radiological Health of the Food and
Drug Administration

Downloaded from http://jicru.oxfordjournals.org/ at Universite Laval on July 9, 2016


Contents
Preface ................................................... . iii
1. Introduction ........ . 1
2. Definitions of Terms and Concepts Currently Used in Ra-
diotherapy . . . . . . ....... . 3
2.1 Volumes . 3

Downloaded from http://jicru.oxfordjournals.org/ at Universite Laval on July 9, 2016


2.1.1 Aim of Therapy .. 3
2.1.2 Target Volume .. 3
2.1.3 Treatment Volume .. . 4
2.1.4 Irradiated Volume ............ . 4
2.1.5 Organs at Risk . 4
2.2 Absorbed Dose Pattern ................ . 4
2.2.1 Representation of a Spatial Dose Distribution by a Set
of Planar Dose Distributions ................... . 4
2.2.2 Maximum Target Absorbed Dose .. . 5
2.2.3 Minimum Target Absorbed Dose . . ........... . 5
2.2.4 Mean Target Absorbed Dose ...... . 5
2.2.5 Median Target Absorbed Dose . . . ...... . 5
2.2.6 Modal Target Absorbed Dose . . . ....... . 5
2.2. 7 Hot Spots . . . . . . . . . . . . . . . . . ...... . 5
3. Recommendations for Reporting Absorbed Dose in Exter-
nal Beam Therapy 10
3.1 Introduction ........... . 10
3.2 Description of Technique .. 10
3.3 Specification of Target Absorbed Dose ................. . 11
3.3.1 Stationary Beams of Photons . . . . . . ........ . 11
3.3.2 Moving Beam Therapy . . . ...... . 13
3.3.3 Electron Therapy . . . . . . . . . . . . . . . . ..... . 14
3.3.4 Complex Treatments: General Recommendations 14
3.3.5 Additional Recommendations ....... . 14
4. Factors Influencing the Biological Effect ....... . 17
4.1 Radiation Quality . . . . . . . . ........ . 17
4.2 Time-Dose Pattern . . . . . . . . . . ....... . 17
References. . . . . ........ . 18
ICRU Reports . . ........ . ................ . 19
Index ............ . 21

vi
Dose Specification for Reporting External
Beam Therapy with Photons and Electrons

1. Introduction

When a patient undergoes a course of radiotherapy, purpose is to define important volumes, areas and
the radiotherapist normally records the radiation absorbed dose patterns and to recommend methods

Downloaded from http://jicru.oxfordjournals.org/ at Universite Laval on July 9, 2016


doses delivered at various points in the irradiated for specifying the absorbed dose in reports of treat-
tissues, including both diseased and healthy tissues. ments with external radiation beams. It is hoped,
This record serves a number of purposes: therefore, that this report will be regarded as a
a. to enable the radiotherapist to maintain his treat- step-albeit a very important step-towards the
ment policy and improve it in the light of experi- achievement of adequate reports of radiation treat-
ence; ment.
b. to enable the radiotherapist to combine the results In principle, reports of radiotherapeutic procedures
of his treatments with those of his departmental should be as complete as possible and should contain
colleagues; adequate and explicit information on the patient and
c. to enable other radiotherapists to benefit from the his disease, the physical parameters and irradiation
department's experience; technique, the overall treatment time and the frac-
d. to enable the results of the department's treatments tionation scheme. The results of treatment can be
to be meaningfully compared with those of other meaningfully interpreted only if all these factors can
centers. be correlated; in particular only if the parameters of
It will be noted that all of these functions, except the irradiation, including the distribution of absorbed
the first, are intended to facilitate communication with dose in space and time, can be significantly correlated
others. In fact, there is little purpose in recording with the clinical and pathological extent of the disease.
absorbed doses if the data cannot be interpreted by Up to now such correlation has been very difficult to
other workers in the field. However, this obvious state- establish, because data relating to both the extent of
ment is far from being realized in practice. Most ra- the disease and to the absorbed dose distribution
diotherapists and physicists are so used to the style (corrected for the individual patient anatomy) have
and conventions used in their own departments that been inadequate. It is expected that the rapid devel-
they would be shocked to learn that their treatment opment of new techniques for the acquisition of pa-
reports were ambiguous, or even incomprehensible, to tient data, such as computed tomography and ultra-
other people. Unfortunately, there is substantial evi- sonography, will greatly improve this situation.
dence that more often than not this is, indeed, the While a complete description of the data relating to
case. It is rare for a description of a treatment to be each patient, as just indicated, is clearly desirable, in
sufficiently explicit and detailed to enable the treat- practice the amount of information that can be re-
ment to be repeated elsewhere without having re- ported in many situations, e.g. in a published paper, is
course to the center of origin for further information. limited. Furthermore, complete information for each
An earlier ICRU report (ICRU, 1973) described the patient, including evaluation of the extent of the dis-
measurement of absorbed dose in a phantom irradi- ease and a full individual dose distribution, is not
ated by a single beam of X or gamma rays. A second always available. One is therefore faced with the prob-
ICRU report (ICRU, 1976) described the determina- lem of selecting a minimum of information for report-
tion of absorbed dose in a patient irradiated by beams ing; such information will be the most relevant for
of X or gamma rays in radiotherapy procedures, and assessing the results of treatment.
defined a number of terms of importance in these The need for selecting is highlighted by the growing
procedures. The present report is the third in a series use of computers for recording patient data and treat-
dealing with dosimetric problems in radiotherapy. Its ment procedures. The information recorded in these
1
2 1. Introduction

systems is usually restricted, but the data should be may well find that the recommendations in this report
valid and unequivocal. A subsidiary purpose of the do not conform exactly with his present system of
present report, in addition to the specification of treat- recording. Nevertheless, there are substantial advan-
ment volumes and absorbed doses, is therefore to tages in adopting a common method of reporting. A
make recommendations on the minimum require- statement in any individual report that the system
ments for reporting external beam therapy. These used conforms to the recommendations of this report
recommendations are intended to be applicable to will remove the need for a definition of terms and a
most clinical situations, past or present, and to most detailed explanation of the meaning of statements
radiotherapy centers. In some situations factors addi- relating to the absorbed dose and other factors. At the
tional to those listed may be considered to have clinical very least, a comparison of any existing system of
implications and should therefore be reported as well. reporting with that recommended here may help to
The preliminary drafts of this report were widely reveal inadequacies in the local procedures. It is hoped
circulated among radiotherapists and radiation phys- that this report will encourage radiotherapy centers to
icists in several countries, but it will be no surprise review their treatment reporting systems and that, in
that it has not been possible to reach a complete doing so, they will give careful consideration to adopt-
consensus of opinion. Any individual radiotherapist ing the present recommendations.

Downloaded from http://jicru.oxfordjournals.org/ at Universite Laval on July 9, 2016


2. Definitions of Terms and Concepts Currently Used in Radiotherapy

Note: Terms and concepts defined in this report are The delineation of the target volume will require
in addition to those defined in previous ICRU reports such considerations as the local invasive capacity of
(e.g., ICRU Report No. 23 (1973), Measurement of the tumor and its potential to spread to regional lymph
Absorbed Dose in a Phantom Irradiated by a Single nodes. Consideration needs to be given to the presence
Beam of X or Gamma Rays, and ICRU Report No. 24 of any specially radiosensitive normal tissue (organs
(1976), Determination of Absorbed Dose in a Patient at risk) as well as to other factors such as the general
Irradiated by Beams of X or Gamma Rays in Radio- condition of the patient.
therapy Procedures). For any given situation there may be more than one
target volume.
Physical treatment planning is dependent on the
2.1 Volumes delineation of the target volume(s) and the prescrip-
tion of the target absorbed dose. These two factors
2.1.1 Aim of Therapy constitute the medical decision which must precede
the determination of the dose distribution in the pa-
A. Curative Treatment of Malignant Disease. In tient. (In the past this sequence has often been re-

Downloaded from http://jicru.oxfordjournals.org/ at Universite Laval on July 9, 2016


the curative treatment of malignant disease anatomic versed and the target volume defined in terms of the
tumor limits may or may not be demonstrated. When dose distribution, for example, the volume enclosed by
demonstrated, the location and extent of the tumor a particular isodose surface. This procedure is not
volume may be determined by means of clinical ex- recommended).
amination, roentgenologic, radioisotopic, ultrasonic, The target volume(s) must always be described,
and microscopic techniques. When the tumor has been independently of the dose distribution, in terms of the
previously removed (e.g., by mastectomy or hysterec- patient's anatomy and topography, and the physical
tomy) the remaining tissues may contain occult dis- dimensions given. When a dose distribution in one or
ease, the limits of which can not be demonstrated. more anatomical sections is presented, the target vol-
When planning treatment the volume to be treated ume (or area in a particular section) should be clearly
to a curative absorbed dose level has to include not indicated on the diagram. If the whole body section
only the demonstrated tumor but also its presumed constitutes the target volume, this fact should be
occult spread. indicated on the isodose chart and stated in any ac-
B. Palliative Treatment of Malignant Disease. The companying description.
palliative treatment of malignant disease may include
all or only part of the demonstrated tumor(s) (e.g., The size and shape of a target volume may change
irradiation of the spine for a painful deposit in a case during a course of treatment (e.g., shrinkage of a
of widespread metastases). mediastinal lymphoma), necessitating replanning.
C. Non-malignant Diseases. The radiotherapy of Note: For external beam therapy the following pa-
non-malignant conditions may or may not include all rameters should be taken into account when describ-
of the affected tissues (e.g., irradiation of a painful ing the target volume:
joint in ankylosing spondylitis). a. expected movements (e.g., caused by breathing) of
those tissues which contain the target volume rel-
ative to anatomic reference points (e.g., skin mark-
2.1.2 Target Volume ings, suprasternal notch),
b. expected variation in shape and size of the target
The target volume contains those tissues that are volume during a course of treatment (e.g., urinary
to be irradiated to a specified absorbed dose according bladder, stomach),
to a specified time-dose pattern. For curative treat- c. inaccuracies or variation in treatment set-up during
ment the target volume consists of the demonstrated the course of treatment.
tumor(s), if present, and any other tissue with pre- An example of target volumes in a patient with a
sumed tumor. carcinoma of the breast is given in Figure 2.1.
3
4 2. Definitions of Terms and Co11cepts

tissue tolerance. The significant absorbed dose level


can be expressed as absorbed dose in percentage (e.g.
50%) of the specified target absorbed dose (see below)
(e.g., Figure 3.5).
The irradiated volume, as well as the treatment
Ox volume, will depend on the treatment technique used.
TRANSVERSE
SECTION
2.1.5 Organs at Risk

a Organs at risk are specially radiosensitive organs in


or near the target volume whose presence influence
treatment planning and/ or prescribed dose.

2.2 Absorbed Dose Pattern

Ox 2.2.1 Representation of a Spatial Dose Distri-

Downloaded from http://jicru.oxfordjournals.org/ at Universite Laval on July 9, 2016


bution by a Set of Planar Dose Distributions

A uniform dose distribution can rarely be achieved


in a target volume, and it is necessary to evaluate the
spatial absorbed dose distribution.
Such an evaluation could be made, in principle, by
b considering the dose distribution in a set of parallel
planar sections sufficiently close to each other (the
distance between two sections being equal to the dis-
tance between the "lattice points", see Section 2.2.4).
Fig. 2.1. Radiotherapy for cancer of the breast as a sole method
However, for practical reasons only a limited num-
of treatment, or combined with surgery. Example of a target volume ber of sections can be evaluated. These sections may
which includes not only the breast (in unoperated patients) and the be selected in the following way (Figure 2.2). The part
chest wall but also the internal mammary, the supraclavicular, and of the patient containing the target volume and rele-
the· axillary lymph nodes. The target volume appears as a target vant anatomic structures is considered as a stack of
area (8ihaded area) in the transverse section (b). In practice, as
shown in (a) several target volumes have to be identified, which are
several parallel slices, the thickness of the slices being
irradiated with different beams, which partly overlap (From Abba- chosen so that in each slice the following conditions
tucci et al., 1972.) are fulfilled:
a. no important variations occur in the external con-
2.1.3 Treatment Volume tour,
b. no important variations occur in the topography of
Because of limitations in treatment techniques it is the relevant internal structures: size, shape and
impossible to administer. the prescribed absorbed dose location of the target volume, organs at risk, heter-
exclusively to the target volume. In general the volume ogeneities, etc.,
receiving at least the same absorbed dose as any part c. no important variations are expected in the dose
of the target volume can not be made to coincide with distribution that are relevant to the treatment plan.
the target volume but will be larger and often of a For each slice a section is chosen on which the extreme
simpler shape. borders of the target volume, the organs at risk, the
The treatment volume is the volume enclosed by an tissue heterogeneities and the reference points in that
isodose surface, the value of which is the minimum slice are projected perpendicularly. The section then
target absorbed dose (see Section 2.2.3). In some cases displays all the relevant structures and the part of the
the treatment volume may be considerably larger than target volume which is located within this slice now
the target volume, as shown in Figure 3.2. appears as a target area.
In many simple situations consideration is given
only to one section, as illustrated in Figure 2.3 for a
2.1.4 Irradiated Volume patient with a carcinoma of the urinary bladder.
The following definitions (Sections 2.2.2-2.2. 7) ap-
The irradiated volume is that volume, larger than ply to dose calculations in a section. They are clinically
the treatment volume, which receives an absorbed relevant only if they can be assumed to represent the
dose which is considered significant in relation to corresponding spatial situation.
2.2 Absorbed Dose Pattern • • • 5
2.2.2 Maximum Target Absorbed Dose (Dr.max) The mean target absorbed dose is then calculated
as the mean of the absorbed dose values in these
The maximum target absorbed dose is the highest lattice points and can be expressed by the equation:
absorbed dose in the target area that can be regarded
as "clinically meaningful". The latter term implies
that at least a minimum area is irradiated to the dose
level designated as "maximum". The minimum area
recommended for this purpose is 2 cm2 , unless the
whole target area is less than 4 cm2 , in which case a where N is the number oflattice points, i is the column
minimum area of 1 cm2 should be taken to define the index in this lattice, j is the row index, and Di,i is the
maximum target absorbed dose. absorbed dose at the lattice point i,j located inside the
The value 2 cm2 is based on two considerations. target area AT.
First, 2 cm2 represents approximately the smallest
area for which the absorbed dose can be calculated
with confidence, either manually or with a computer;
second, the maximum target absorbed dose is often
related to the limiting effects of treatment such as

Downloaded from http://jicru.oxfordjournals.org/ at Universite Laval on July 9, 2016


tissue tolerance and the smallest volume of tissue to 2.2.5 Median Target Absorbed Dose (Dr.median)
which these effects apply is considered to be that
volume whose section is at least 2 cm2 • The median target absorbed dose is the central
Within an isodose curve enclosing an area of 2 cm 2 value among the set of values of the absorbed dose at
the dose at a point may be even higher, but it is all lattice points in the target area, when arranged
recommended that such hot points be ignored in des- according to magnitude.
ignating the value of the maximum target absorbed
dose.

2.2.6 Modal Target Absorbed Dose (Dr,moda1)


The modal target absorbed dose is the absorbed
dose that occurs most frequently at lattice points in
the target area. Its value may be influenced by the
choice of method for its calculation (e.g., spacing of
2.2.3 Minimum Target Absorbed Dose (Dr.min) lattice points). Exceptionally, in a particular patient,
more than one modal target absorbed dose may be
The minimum target absorbed dose is the lowest
found.
absorbed dose in the target area. No area limit is
An example of a computerized calculation of the
recommended when reporting minimum target ab-
different above-mentioned target absorbed doses in
sorbed dose.
one section of a patient is given in Figure 2.4.

2.2.4 Mean Target Absorbed Dose (Dr.mean) 2.2. 7 Hot Spots

For the determination of the mean, as well as of the In many situations tissues outside the target area
median and modal target absorbed dose, it is necessary will receive a relatively high absorbed dose. A hot spot
to calculate the dose at each of a large number of is an area which receives an absorbed dose higher than
discrete points (lattice points), uniformly distributed 100% of the specified target absorbed dose (see Sec-
in the target area. tion 3.3). However, as in the case of the maximum
target absorbed dose (see Section 2.2.2), a hot spot is
considered clinically meaningful only if the corres-
ponding isodose curve encloses an area of at least 2
cm2 in a section.
6 2. Definitions of Terms and Concepts

a
--·-·---~ ---
-·-· -·-· -·-·-4-<M:· . . ·;/J-· -·- -- ·-·--
)---·---·--
II
© I

-·-· ---·-· -·-g · -·-·-·-·-·-·-·-III

~ II

~
- - -·- - - )' , .. ··-\-·-·---·-·-·-·-IV

-·-·cL . L. ~----- v

®
.. III
( I \?<'/ .

(''-----..____,__ "

@
(' \

Downloaded from http://jicru.oxfordjournals.org/ at Universite Laval on July 9, 2016


L" . . l
..
IV
---·-·-·-·--1 7 .A)) }---·-·--·--·-·- I

II
.•....
-·- -~ ~vr·-·-·-· -·-·-·-·-· III

CJ . .
f~o··
.. ..
v

·. ...: n··.... ·:·


-·-·-· - · - -,_!~/-!_-'g"-'"'--<t_,\·
Ii_ ..,: .-1
-·- ·- --·-·-·-·-· IV
..... ~O~ ..... .

"
ti /"x'
)~·.... ·\ r·-----·-- -·- -·- v

I \
Fig. 2.2(a). Example of a multi-section dose plan for the curative treatment of a carcinoma of the nasopharynx (demonstrated tumor=
target A, indicated by black) and its presumed spread to the regional lymph nodes in the neck and supraclavicular fossae (subclinical disease
= target B, indicated by the dotted line (Figure 2.2(a)) or honeycombed area (Figure 2.2(b), page 7). The prescribed absorbed dose for target
A is higher than that for target B. The organs at risk are the spinal cord, brafu stem, and to some extent the eyes. When calculating the
absorbed dose distribution, correction for tissue heterogeneity (lung tissue and bone) may be applied. The preliminary suggestion for treatment
technique is a multifield arrangement with lateral fields towards target A and one anterior and two oblique posterior fields towards target B,
in order to keep the dose to organs at risk as low as possible. After having defined optimal patient positioning with respect to suggested
treatment technique, representative sections will be produced. In this case, 5 sections (I- V) were needed for the calculation of the absorbed
dose pattern because of variation in shape and size of the two target volumes, outline of the patient and treatment technique. Each section
represents one slice of the patient. The extreme borders of the target volume, organs at risk, tissue heterogeneities and reference points in that
slice have been projected perpendicularly on to the section, thus defining the target area and other relevant structures ut;.the section.

Fig. 2.2(b). (Opposite page) The accepted dose plan appears in b. The schematic figure (top left) indicates (Arabic numerals) arrangement
of the fields. (Note: beams 2, 4, and 6 are actually directed towards the right side of the patient.) The weighting of the beams and the absorbed
dose distribution in each of the five sections have been normalized to peak absorbed dose in beam I (Modified from: Landberg and Svahn-
Tapper, 1976.)
b

5•7

II

Downloaded from http://jicru.oxfordjournals.org/ at Universite Laval on July 9, 2016


''
3

>

,I
I

I
I

:~ 2

Field Radiation SSD Field size Weight Wedge Compensator


cm x cm 5cm
60Co gamma rays 70 20 x 15 100 60 cm lead centrolly
2,3 60Co gamma rays 70 8 x 18 50 45°
4,5 8 MV X-rays 100 5 x 6 50 15°
6,7 8 MV X-rays 100 5 x 3 30

Absorbed dose (per cent of


peak absorbed dose in beam 1)

Target A Target B Brain stem, spinal


cord
Max.-Min. Max.-Min. Max.
Section I 135-125 55
II 140-120 140- 40 40
III 140-110 55
IV 145-130 40
v 130-100 45

7
8 2. Definitions of Terms and Concepts

Frontal projection (A.P. radiograph)

Downloaded from http://jicru.oxfordjournals.org/ at Universite Laval on July 9, 2016


Sagittal (or lateral) projec- lransverse (or cross) section
tion (lateral ~adiograph)
Fig. 2.3. Example of a target in a patient with localized cancer of the urinary bladder and presumed occult spread. The transverse section
is considered to be representative of the whole target volume and the relevant normal tissues. For the graphical construction of the transverse
section two radiographs are used that represent two perpendicular projections of the anatomy of interest. The border of the target is indicated
by the dotted line. The target appears in the two radiographs and is indicated as an area in the section. It includes the demonstrated tumor
(indicated in black). the whole bladder and the regional lymph nodes (the iliac and the obturator groups) as well as connecting lymphatics.

Fig. 2.4(a). (Opposite page) Computerized calculation of the


absorbed dose distribution (8 MV x rays) for a patient with carci-
noma of the esophagus, treated only in the prone position. (a)
Transverse section of the patient. The calculation of the absorbed
dose includes correction for tissue heterogeneity, the "effective Fig. 2.4(b). (Opposite page) For the same patient a histogram
density" of the lung tissue being considered to be 500 kg m - 3 and is given to demonstrate the distribution of absorbed dose in the
that of bone to be 1300 kg m- 3 • The border of the target area is target area. The computation of the histogram gives sizes of areas
indicated by the thick line. (one lattice point representing an area of 3 mm X 3 mm= 0.09 cm2)
The display gives the distribution of the absorbed dose and also the for 21 equally large intervals of absorbed dose percentage values,
position of the maximum (4) and the minimum<•> target absorbed ranging from 102% to 93%. The maximum target absorbed dose
dose. In this special case, the weighting of the beams (peak absorbed considered to be meaningful and to be used for reporting (see
dose 50%, 45%, and 45%, respectively), was chosen to give an Section 2.2.2) is 100%. The modal target absorbed dose is 99%, and
absorbed dose of 100% at the point of intersection of the central the median 98%. The computation also gives the mean target
axes of the three beams. absorbed dose (98%) (Modified from: Moller et al., 1976.)
2.2 Absorbed Dose Pattern 9
a

SIN DX

Downloaded from http://jicru.oxfordjournals.org/ at Universite Laval on July 9, 2016


Field SSD Size
cm cm x cm

123~·76911
76111122
9928·76
2-1-llJGI
H.S.
u u
5 cm
... /
VENTR
2
3
100
100
100
9 x 14
8 x 14
8 x 14

Area
7 (=Nx TARGET ABSORBED DOSE
b 0.09)
MINIMUM MODAL MAXIMUM
4.0
'
3.2

2.4 -

1.6

0.8

O.O -t----,.---,.-----r-..u.L........,...._LU-L.U.1~..u....---. Relative absorbed


76 81 86 91 96 101 106 dose ( % )
3. Recommendations for Reporting Absorbed Doses in External Beam Therapy

3.1 Introduction 3.2 Description of Technique

The absorbed dose distribution is usually not uni- Necessary information:


form in the target volume (see Section 2.2.1). However, a. Radiation quality
for the purpose of treatment reporting a nominal I. for conventional (100-300 kV) or low energy
absorbed dose, which will be called target absorbed (< 100 kV) x rays, the accelerating potential
dose, has to be selected. The use of the expression (kV) and HVL,
"tumor dose" is no longer recommended. 2. for gamma rays, the radionuclide (element and
The maximum and minimum target absorbed doses mass number),
alone or together can not as a rule be used for reporting 3. for high energy x rays, the equivalent accel-
since they are not always representative of the dose erating potential (MV) and the type of ma-
distribution. Although local tumor control depends on chine,
the minimum dose to the malignant cell population, 4. for electron beams, the energy (Me V) and type
the use of the minimum target absorbed dose alone of machine.
can not be recommended, since difficulties in deter- b. The number and arrangement of the beams. In-

Downloaded from http://jicru.oxfordjournals.org/ at Universite Laval on July 9, 2016


mining the extent of the malignancy may reduce its dication of SSD or SAD. The description of the
clinical significance and further may lead to ambiguity beam arrangement of external beam therapy
in its definition. should always be done in relation to the patient.
The mean, median and modal target absorbed doses c. Field sizes: geometrical field sizes which usually
can not be generally recommended since they usually correspond to the 50% isodose curve. For fixed
require a complete computerized dose plan, which may SSD, the field sizes are usually given at the skin;
be available only for a limited number of patients for fixed SAD (isocentric arrangement) the field
and/ or hospitals. sizes are given at the isocenter. For any field, its
The absorbed dose selected for reporting should be size in the planned section should be given first.
chosen to be representative of the dose distribution in d. Beam modification devices (wedges,. shielding
the target volume, and its calculation should, if possi- blocks, etc.).
ble, not necessitate special computation facilities. e. Beam weighting. State whether this is defined in
It should be stressed that target absorbed dose, as terms of the ratio of peak (applied) absorbed
specified in the tollowing paragraphs, represents a doses for the single beams, or in terms of the
minimum requirement for reporting, often carrying ratio of doses delivered at a defined point in the
limited radiobiological and clinical information. target, such as the isocenter or the "specification
Finally, the specification of target absorbed dose for point" (see Section 3.3.4). To avoid ambiguity,
reporting depends on the treatment technique, and its the meaning of beam weighting should always be
significance can only be interpreted when information made clear.
on the irradiation technique has been given. f. Corrections for tissue heterogeneity (state
whether performed or not).
g. Patient positioning.

Optional information:
h. Absorbed dose distribution (isodose pattern).
i. "In vivo" absorbed dose measurements.

10
3.3 Specification of Target Absorbed Dose 11

Fields 1 and 2
soco
3.3 Specification of Target Absorbed Dose 70cm SSD

3.3.1 Stationary Beams of Photons

a. For a single beam the point at which the target


absorbed dose should be stated should be on the Ox
central axis at the center of the target area
(Figure 3.1).
b. For two opposed coaxial equally weighted beams
the point at which the target absorbed dose
should be stated should be on the central axis
midway between the beam entrances (Figure
!..I !ewl w
3.2). 2 5cm
c. For two opposed coaxial unequally weighted
beams the point at which the target absorbed Fig. 3.2. Treatment of a bronchogenic carcinoma including

Downloaded from http://jicru.oxfordjournals.org/ at Universite Laval on July 9, 2016


dose should be stated should be on the central treatment of the mediastinum with coaxial opposed equally
weighted 60 Co beams. No correction for tissue heterogeneity. The
axis at the center of the target area (Figure 3.3). point at which the target absorbed dose should be stated (e) is
d. For any other arrangement of two or more inter- located on the central axis, midway between the beam entrances.
secting beams the point at which the target ab- • = demonstrated tumor; !Bl =· target area. In this patient the
sorbed dose should be stated should be at the treatment volume is delineated by the 100% isodose curve.
intersection of the central axes of the beams
(Figures 2.4, 3.4, and 3.5).

Weight = 1 .O

Fields 1 and 2
so co
70 cm SSD

so co
70 cm SSD
100 %
Ox

Ox

!wwl - lwwl
2 5cm
Weight = 0.5

Fig. 3.3. Treatment of a malignant thymoma with two coaxial


opposed unequally weighted beams. The weighting factors are 1.00
5cm
and 0.50 for the anterior and the posterior beams, respectively. No
Fig. 3.1. Single beam therapy with 60 Co of the internal mam- correction for tissue heterogeneity.
mary lymph nodes. No correction for tissue heterogeneity. The The point at which the target absorbed dose should be stated
point at which the target absorbed dose should be stated (e) is (e) is located at the central axis at the center of the target volume.
located on the central axis at the center of the target area. ~ = In this particular case, the relative target absorbed dose to be stated
target area. is 103%. • = demonstrated tumor; ~ = target area.
--'7'-1----------t=SECTION 3

Percentage value
60Co
SAD = 80 cm 60 10
Field No. Size Weighting ------ 70 20 1.
(cm x cm factors BO 30
17 x 20 0.37 .... ·············- 90 40
2 17 x 20 0.37 -------·- 95 50
3 10 x 20 0.13 100
4 10 x 20 0.13

10an
SECTION 2 ............. .......
.__,.~ ~ ~· _,,

Downloaded from http://jicru.oxfordjournals.org/ at Universite Laval on July 9, 2016


Ox

3-·--

2.
SECTION 1
I SECTION 3

60
Fig. 3.4. Irradiation of the pelvis with Co "box technique". Variation in patient outline at different levels necessitated multisection dose
calculations. The figure shows computerized calculation of the absorbed dose distribution in the principal plane (section 2) and in 2 parallel
planes 7 cm above (section 1) and below (section 3). In each section the border of the target area is indicated by the thick dotted line. The
point at which the target absorbed dose should be stated is located at the intersection of the central axes of the 4 beams. The absorbed doses
in the different sections are expressed in percentage of that dose level in the principal plane. The weighting factors indicated in the table
correspond to the contribution of each beam to the target absorbed dose. The target absorbed dose to be reported for section 2 is 100%. For
sections 1 and 3, the absorbed doses at the intersections of the beam axes are 87% and 94%, respectively.

SAD Size
cm cm x cm

100 10 x 10
Fig. 3.5. Treatment of a lesion of the tongue and the floor of
100 8xl0 the mouth (the border of the target area being indicated by the
thick dotted line) with two wedged 8 MV x-ray beams. The point at
which the target absorbed dose should be stated is located at the
intersection of the central axes. The absorbed dose level at this
point is defined as 100%, and is also indicated by the symbol B.
Almost the whole target area is confined within the 95% isodose
curve. The 50% isodose curve is also shown.
The left parotid and sub-maxillary glands were considered to be
organs at risk, and their positions are indicated in the section by the
broken line. The spinal cord receives less than 10%. Note that the
two beams have different wedge filters and also different weighting
factors (Modified from: Van der Laarse, 1976.)

o-------z-------.. -------f,.------it-------o-------z-------i.-------f.-------11-------
3.3 Specification of Target Absorbed Dose 13
3.3.2 Moving Beam Therapy
Ox
For complete rotation the point at which the target
absorbed dose should be stated should be in the prin-
cipal plane at the center of rotation. The same rec-
ommendation applies when the arc of rotation is less Percentage value

than 360° but at least 270° (Figure 3.6). 100 60 10


COBALT 60
ARC THERAPY
For smaller arcs (Figure 3. 7) the target absorbed 105 70 20
SAD = 75 cm
110 80 30 2
dose should be stated in the principal plane, first, at 115 90 40
6.0 x 14.0 cm

ARC = 100
the center of rotation and second, at the center of the 95 50

target area. The irradiation parameters are usually


chosen in such a way that the absorbed dose in the 0 5cm

center of the target area is close to the maximum


target absorbed dose. Fig. 3.7. Moving beam therapy with a relatively small arc
(100°): treatment of a rectal tumour. For the chosen arc, the
maximum absorbed dose is not obtained at the axis of rotation, and
for reporting purposes it is recommended that the absorbed dose be
indicated at the center of rotation and at the cenier of the target

Downloaded from http://jicru.oxfordjournals.org/ at Universite Laval on July 9, 2016


area. In this particular patient the absorbed dose at the center
of the target area is 117% of the absorbed dose at the center of
rotation. The border of the target area is indicated by the thick
dotted line.

o-----z------•------•----•-----u-------l-------•------b------M-------u-------2------•-------b-------8
I
I
I
SAD Field size
I
2
I
Ox cm cm x cm
!
7 x 16
2'l,,b'
~ 100
l!
i.2 22
l 2
2 'l 2<r'22
2
z
2
,• 'l
l l
ll
2 ,,

2
ll I
2 I
>222 I
2 I
2
2
2
2
2
2
I 2 2222222<
222222 <2 2
< 2
l< 2
2 U22 2
2 PZZZZ~ 22'/t' 2 2?22222 2

0----2------•-------•----a-----u---- -t-------•-------b-------11-------v--- -i-----•-------b-------e-------u------2----


I

Fig. 3.6. Treatment of an esophageal lesion by moving beam technique (300°) with 8 MV x-rays. The point at which the target absorbed
dose should be stated is located at the isocenter. The relative absorbed dose at that point is 100%. The target area is located within the 95%
isodose curve. When correcting for tissue heterogeneity the lung was assumed to have a bulk density of 350 kg m- 3 • The spinal cord receives
an absorbed dose ranging from 40% to 20% of that at the isocenter. The border of the target area is indicated by the thick dotted line.
14 3. Recommendations for Reporting Absorbed Doses

3.3.3 Electron Therapy 3.3.4 Complex Treatments: General Recom-


mendations
The energy of the electron herun is generally chosen
so that an isodose curve of at least 80% of the peak There are some situations where the above-men-
absorbed dose encloses the target area. The stated tioned recommendations may not apply. Such are for
target absorbed dose should be the maximum target example
absorbed dose; the absorbed dose in the center of the a. coplanar beruns whose central rays do not inter-
target area is usually close to this value (Figure 3.8). sect at one point (Figures 3.9, 3.10, and 3.11),
or intersect at a point outside the patient
(Figure 2. 2 (b) .IV).
b. non-coplanar berun treatments (Figure 3.9),
c. opposed non-coaxial beams (Figures 3.9 and
3.11),
a
d. different target volumes (Figure 3.9),
Electrons 15 MeV e. modification in the treatment technique.
110 cm SSD
100 %
It is recommended that one or more "specification
6cmx14cm points" be defined for the above-mentioned situations,

Downloaded from http://jicru.oxfordjournals.org/ at Universite Laval on July 9, 2016


each having a meaningful relation to the target volume
and/ or the irradiation beams. The proper choice of
one or more points is facilitated by an isodose presen-
Ox tation. When reporting, information should be given
on the position of the specification point(s). The stated
target absorbed dose should be expressed as a per-
centage of the absorbed dose at one or more of these
points (Figure 3.12). Where more than one target
lwol lwol .....
volume is involved, more than one specification point
Scm is usually necessary (Figure 3.9). When treatment has
Fig. 3.8(a). Electron treatment (15 MeV) of the internal mam- to he modified, new specification point(s) may have to
mary lymph nodes. No correction for tissue heterogeneity. The he chosen with corresponding absorbed dose calcula-
target absorbed dose to be reported is the maximum absorbed dose tions being carried out.
in the central beam (100%). Because of the flatness of the electron
depth dose curve in the region of its maximum, 100% may be found
Note: The situations mentioned in Sections 3.3.1, 3.3.2,
at many points which can not be displayed by an isodose curve. and 3.3.3 can, if preferred, he dealt with ·as complex
Therefore, 100% is not indicated in the figure. In the example chosen treatments using specification points.
here, an absorbed dose close to 100% is found at the center of the For these treatment plans, even if complete ab-
target area.
sorbed dose distributions are not computed for each
particular patient, it is assumed that a basic or model
dose distribution is available, either from the srune
b
radiotherapy center or from a collection (such as an
Electrons 15 MeV atlas; IAEA, 1978) produced elsewhere.
110 cm SSD
100 % 3.3.5 Additional Recommendations
If the absorbed dose in the target volume varies by
more than ± 10% of the stated target absorbed dose,
Ox then the maximum (Section 2.2.2) and the minimum
(Section 2.2.3) target absorbed dose should also be
given. Multiplane calculations may he necessary.
Any hot spot should be reported (see Section 2.2.7).
The value of the absorbed dose of the hot spot to be
reported should be that of the isodose curve that
encloses an area of 2 cm2, not the absolute maximum
- J.+wl lww!
Scm
value.
Fig. 3.8(b). Same as Figure 3.8(a), but corrected for tissue
heterogeneity, the "effective lung density" considered to be 500 kg
For organs at risk the maximum absorbed dose
m.-3
should be given and the runount of the organ involved
stated (e.g., IO cm of the spinal cord, the upper half of
the left kidney).
Any other additional information considered to
carry relevant radiohiological and/or clinical infor-
mation should be reported.
3.3 Specification of Target Absorbed Dose 15
Note I: In an isodose plan the value of each isodose Note III: The transverse sections in this report are
curve should, if possible, be written at the higher shown as viewed from the foot end of the patient. The
absorbed dose side of the curve. right and/or left sides of the sections are indicated by
Note II: The isodose plans shown in this report were Dx and Sin, respectively.
generated in clinical routine work in several different Note IV: The figures used in this report were selected
centers and were only slightly edited for simplicity in to illustrate and clarify the definitions and recommen-
layJout. They show different ways of presenting dose- dations given in the text. It is not the intention of the
distributions. It is not felt that recommendations authors of the report to recommend particular irradia-
should be given at this time for presentation of isodose tion techniques. Of course, the examples were chosen
distributions, but the reading of such a presentation is to be as close as possible to the clinical situations to
facilitated if the isodose curve through the specifica- be illustrated.
tion point (see Section 3.3.4) is given the value 100%.

60Co
Field SSD Size Wedge Beam weight
I cm cm x cm %
70 22 x 22 - 100

Downloaded from http://jicru.oxfordjournals.org/ at Universite Laval on July 9, 2016


1
2 70 22 x 22 100
3 70 5 x 5 15° 40
4 70 5 x 5 15° 40
5 70 12 x 5 - 180
6 70 12 x 5 - 180

20
III
30 30


II

_. I.I IM
5 cm
Fig. 3.9. Dose plan in 3 sections for the treatment of medulloblastoma with two target volumes (a/the demonstrated cerebellar tumor,
and b/the whole subdural space and the ventricular .system) with coplanar beams whose central rays do not intersect at one point, non-
coplanar beams, and opposed noncoaxial beams. The weighting of the beams was chosen to give the total absorbed dose in the two target
volumes with the same number of fractions. The distribution of the absorbed dose is calculated for first, the demonstrated tumor and the
cranial subdural space down to the second cervical vertebra in a frontal section (I), second, for the spinal subdural space in a median sagittal
section (II) from the second cervical vertebra to the second sacral vertebra, and thirdly in a transverse section (III) through the kidneys. The
cranial subdural space is treated with two opposed and the spinal subdural space with two adjacent 6()Co beams to a target absorbed dose of
30 Gy. The cerebellar tumor is further treated with two smaller opposed 6()Co beams to a total of 40 Gy. For reporting, three different
specification points are chosen for the two target volumes, namely one (e) for the demonstrated cerebellar tumor and two C• and•) for the
whole subdural space and the ventricular system. (Modified from: Moller et al., 1976.)
16 3. Recommendations for Reporting Absorbed Doses

Field Radiation SSD Field size Weight (reference= Field Radiation SSD Field size Weight (reference =
cm cm x cm = speci ficotion point) cm cm x cm = specification point)
25 MV X-rays 100 60co gamma
18 x 14 0.22 rays 80 28 x 8 l.2
2 60Co gamma rays 80 27 x 14 l.20 2 25 MV X-rays 100 17 x 8 0.22
3 9 MeV electrons 100 8 x 14 0.12

5 o•

Downloaded from http://jicru.oxfordjournals.org/ at Universite Laval on July 9, 2016


Fig. 3.10. Treatment of a thyroid carcinoma with mediastinal
involvement with coplanar beams whose central rays do not inter- Fig. 3.11. Treatment of the upper one-third of the esophagus
sect at one point. The purpose of treatment was to deliver 50 Gy to with opposed non-coaxial beams. A dorsal 25 MV x-ray beam was
the upper mediastinum at mid-diameter (specification point, indi- added to the ventral 60Co beam in order to avoid overdosage to the
cated by the symbol ®. The 25 MV x-ray beam was necessary tr spinal cord. Sufficiently high absorbed dose is delivered by the
boost the absorbed dose to the caudal part of the target area and dorsal 25 MV x-ray beam to provide 50 Gy to the upper mediastinum
the 9 MeV electron beam to boost the absorbed dose to the cranial at mid-diameter (specification point, indicated by the symbol ®).
part of the target area. It was decided to keep the spinal cord The border of the target area is indicated by the dotted line. The
absorbed dose below 42 Gy. The border of the target area is figures at the isodose lines indicate absolute absorbed dose values.
indicated by the dotted line. The figures at the isodose curves
indicate absolute absorbed dose.

Fields 1 and 2
BO co
130 cm SSD Beam flattening•
contour comp. filter

0
ro

lww! lawaal !awl


5 cm

Fig. 3.12. Dose plan for the sagittal section in mantle treatment, showing isodose distribution and absorbed dose pattern along the
hypopharynx and the esophagus. The relatively short diameter in the neck region is compensated for by an individual copper filter which has
been constructed to correspond to the lined area. This contour compensating filter is placed at the collimator in addition to the general beam
flattening filter used for this technique. The specification point (e) is located on the central axis midway between the beams. The absorbed
dose at this point is 150% of peak absorbed dose at the central axis of either beam. In this particular patient the minimum target absorbed dose
was considered to be of particular importance, being located at the most cranial and most caudal part of the target area <•>
and being 90% of
the absorbed dose at the specification point. IW = target area.
4. Factors Influencing the Biological Effect

4.1 Radiation Quality 4.2 Time-Dose Pattern

Radiation quality(ies) and the respective absorbed The reporting of absorbed dose for external beam
dose(s) should be clearly stated (see Section 3.2.a). therapy should be accompanied by information on the
For comparison purposes, 60Co gamma rays should time-dose pattern including at least the number of
be taken as the reference radiation. Conversion factors fractions and the overall time (in days). The first and
recommended in radiation therapy to take into ac- the last day of treatment are included in the reported
count RBE differences are given in Table 4-1. They overall time.
agree with most of the published data relevant to Unless otherwise stated, it is understood that the
radiation therapy, which indicate a higher RBE value fractionation is regular (5 fractions week- 1 ) with equal

Downloaded from http://jicru.oxfordjournals.org/ at Universite Laval on July 9, 2016


of conventional x rays. Absorbed dose values of con- fractions, all fields being irradiated at each session.
ventional x rays should then be multiplied by 1.18 The position and length of any gap or interruption
when their effects are compared with those of high must be given as well as any change in dose per
energy radiation. fraction.
Unless otherwise stated, it is assumed that the ab- Recording of target absorbed dose rate is useful and
sorbed doses are given by either 60Co gamma rays, is recommended for values below 0.1 Gy min- 1 at any
high energy x rays or high energy electrons. part of the target volume.

TABLE 4-l-Conversion factors recommended in radiation therapy to take into account differences in RBE
Radiation Quality Conversion Factors References
Conventionala (100-300 kV) x rays 1.18 ICRP (1963);
Krohmer (1965);
Wambersie and Dutreix (1971)
1.00 ICRP '(1963); Sinclair (1962);
Wambersie and Dutreix (1971);
Wambersie et al. (1973)
~2 MVxrays 1.00 ICRP (1963); Sinclair (1962);
Sinclair and Kohn (1964);
Wambersie and Dutreix (1971)
Electron beams 1.00 ICRP (1963); Hettinger et al. (1965);
(energy 1-50 Me V) Kim et al. (1968); Sinclair and Kohn (1964);
Wambersie et al. (1966);
Wambersie and Dutreix (1971)
a Note: For low energy (<100 kV) x rays the same conversion factor (1.18) can be recommended for most

practical situations, although some experimental data suggest increase of RBE with decreasing energy (ICRP,
1963; Krohmer, 1965; Wambersie and Dutreix, 1971).

17
References

ABBATUCCI, J. s., QUINT, R., BLOQUEL, J., ROUSSEL, Radiat. Res. 24, 54 7.
A. and URBAJTEL, M. (1972). Techniques de Tele- LANDBERG, T., and SVAHN-TAPPER, G. (1976). "En-
cobaltherapie radicale (L'Expansion Scientifique bloc irradiation of tumours of the head and neck
Fran~aise, Paris). and their lymphatics. I. Technique and dosimetry,"
HETTINGER, G., BERGMAN, S., and 0STERBERG, S. Acta Radiol. Ther. Phys. Biol. 15, 129.
(1965). "The relative biological efficiency (RBE) of MOLLER, T. R., NORDBERG, u. B., GUSTAFSSON, T.,
30 MeV electrons on haploid yeast," Biophysik 2, JoHNSSON, J. E., LANDBERG, T. G., and SvAHN-
276. TAPPER, G., (1976). "Planning, control, and docu-
IAEA (1978). International Atomic Energy Agency, mentation of exernal beam therapy," Acta Radiol.
Atlas of Radiation Dose Distributions, Vol. VI, Suppl. 353.
Treatment Plans for Cobalt-60 Therapy, Mitchell, SINCLAIR, W. K. (1962). "The relative biological effec-

Downloaded from http://jicru.oxfordjournals.org/ at Universite Laval on July 9, 2016


J S. Cohen, M., and Snelling, M., Eds. (Interna- tiveness of22 MeVp X-rays, Cobalt-60 gamma rays,
tional Atomic Energy Agency, Vienna). [In press] and 200 KeVp X rays," Radiat. Res. 16, 336.
ICRP (1963). International Commission on Radiolog- SINCLAIR, W. K. and KOHN, H. I. (1964). "The relative
ical Protection, "Report of the RBE Committee to biological effectiveness of high-energy photons and
the International Commissions on Radiological Pro- electrons," Radiology 82, 800.
tection and on Radiological Units and Measure- VAN DER LAARSE, R. (1976). "Pseudo optimization of
ments," Health Phys. 9, 357. radiotherapy treatment planning," Brit. J. Radiol.
ICRU (1973). International Commission on Radiation 49, 450.
Units and Measurements, Measurement of Ab- W AMBERSIE, A., DUTREIX, A., DUTREIX, J ., LEL·
sorbed Dose in a Phantom Irradiated by a Single LOUCH, J., MousTACCHI, E., and TuBIANA, M.
Beam of X or Gamma Rays, ICRU Report 23 (In- (1966). "Efficacite biologique relative d'un faisceau
ternational Commission on Radiation Units and d' electrons de 20 MeV en fonction de la profondeur,"
Measurements, Washington). Int. J. Radiat. Biol. 10, 261.
ICRU (1976). International Commission on Radiation WAMBERSIE, A., and DUTREIX, A. (1971). "ProblemeE
Units and Measurements, Determination of Ab- dosimetriques poses par la determinations de l'EBR
sorbed Dose in a Patient Irradiated by Beams of X dans un large domaine d'energie (electrons de 15 ii
or Gamma Rays in Radiotherapy Proce,dures, 34 MeV, photons de 55 kV a 20 MV)," page 261 ir
ICRU Report 24 (International Commission on Ra- Biophysical Aspects of Radiation Quality, IAE..A
diation Units and Measurements, Washington). Publication STl/PUB/286, (International Atomic
KIM, J. H., PINKERTON, A., and LAUGHLIN, J. S. Energy Agency, Vienna).
(1968). "Studies on the biological effectiveness of WAMBERSIE, A., PRIGNOT, M., and GuEULETTE, J.
high-energy electron beams at different depths in (1973). "A propose du remplacement du radium par
tissue absorbing material," Radiat. Res. 33, 419. le cesium 137 en curietherapie gynecologique," J.
KROHMER, J. s. (1965). "RBE and quality of electro- Radiol. Electro!. 54, 261.
magnetic radiation at depths in a water phantom,"

18
ICRU Reports

ICRU Reports are distributed by the ICRU Publications' office. Infor-


mation on prices and how to order may be obtained from:

ICRU Publications
P.O. Box 30165
Washington, D.C. 20014
U.S.A.

Downloaded from http://jicru.oxfordjournals.org/ at Universite Laval on July 9, 2016


The extant ICRU Reports are listed below.

ICRU Title
Report No.
lOb Physical Aspects of Irradiation (1964)
lOc Radioactivity (1963)
lOd Clinical Dosimetry (1963)
lOe Radiobiological Dosimetry (1963)
lOf Methods of Evaluating Radiological Equipment and
Materials (1963)
12 Certification of Standardized Radioactive Sources
(1968)
13 Neutron Fluence, Neutron Spectra and Kerma (1969)
14 Radiation Dosimetry: X Rays and Gamma Rays with
Maximum Photon Energies Between 0.6 and 50
MeV (1969)
15 Cameras for Image Intensifier Fluorography (1969)
16 Linear Energy Transfer (1970)
17 Radiation Dosimetry: X Rays Generated at Poten-
tials of 5 to 150 kV (1970)
18 Specification of High Activity Gamma-Ray Sources
(1970)
19 Radiation Quantities and Units (1971)
19S Dose Equivalent [Supplement to ICRU Report 19]
(1973)
20 Radiation Protection Instrumentation and Its Appli-
cation (1971)
21 Radiation Dosimetry: Electrons with Initial Energies
Between 1 and 50 Me V (1972)
22 Measurement of Low-Level Radioactivity (1972)
23 Measurement of Absorbed Dose in a Phantom Irra-
diated by a Single Beam of X or Gamma Rays
(1973)
24 Determination of Absorbed Dose in a Patient Irra-
diated by Beams of X or Gamma Rays in Radio-
therapy Procedures (1976)
25 Conceptual Basis for the Determination of Dose
Equivalent (1976)
19
20 · · · ICRU Reports

26 Neutron Dosimetry for Biology and Medicine (1977)


27 An International Neutron Dosimetry Intercompari-
son (1977)
28 Basic Aspects of High Energy Particle Interactions
and Radiation Dosimetry (1978)
29 Dose Specification for Reporting External Beam
Therapy with Photons and Electrons (1978)

Binders for ICRU Reports are available. Each binder will accommodate
from six to eight reports. The binders carry the identification, "ICRU
Reports," and come with label holders which permit the user to attach
labels showing the Reports contained in each binder.
The following bound sets of ICRU Reports are also available:
Volume I. ICRU Reports lOb. lOc. lOd. lOe, lOf
Volume II. ICRU Reports 12, 13. 14, 15, 16, 17, 18, 19, 19S, 20
Volume III. ICRU Reports 21, 22, 23, 24, 25, 26
(Titles of the individual Reports contained in each volume are given in

Downloaded from http://jicru.oxfordjournals.org/ at Universite Laval on July 9, 2016


the list of Reports set out above.)

The following ICRU Reports were superseded by subsequent Reports


and are now out of print:
ICR U Report No. Title and Reference•
1 Discussion on International Units and Standards for
X-ray Work. Brit. J. Radiol. 23, 64 (1927).
2 International X-ray Unit of Intensity, Brit. J. Radiol.
(new series) 1, 363 (1928).
3 Report of Committee on Standardization of X-ray
Measurements, Radiology 22, 289 (1934).
4 Recommendations of the International Committee for
Radiological Units, Radiology 23, 580 (1934).
5 Recommendationlj of the International Committee for
Radiological Units, Radiology 29, 634 (1937).
6 Recommendations of the International Commission
on Radiological Protection and of the Interna-
tional Commission on Radiological Units, National
Bureau of Standards Handbook 4 7 (U.S. Govern-
ment Printing Office, Washington, D.C., 1951).
7 Recommendations of the International Commission
for Radiological Units, Radiology 62, 106 (1954).
8 Report of the International Commission on Radiol-
ogical Units and Measurements (ICRU) 1956, Na-
tional Bureau of Standards Handbook 62 (U.S. Gov-
ernment Printing Office, Washington, D.C., 1957).
9 Report of the International Commission on Radiol-
ogical Units and Measurements (ICRU) 1959, Na-
tional Bureau of Standards Handbook 78 (U.S. Gov-
ernment Printing Office, Washington, D.C., 1961).
lOa Radiation Quantities and Units, National Bureau of
Standards Handbook 84 (U.S. Government Printing
Office, Washington, D.C., 1962).
11 Radiation Quantities and Units (International Com-
mission on Radiation Units and Measurements,
Washington, D.C., 1968).
• References given are in English. Some of the Reports were also published in other
languages.
Index

Aim of therapy, 3 Specification of target absorbed dose, 4, 11, 13-14


Curative treatment of malignant disease, 3 Complex treatments: general recommendations, 14
Palliative treatment of malignant disease, 3 Electron therapy, 14
Treatment of non-malignant disease, 3 Moving beam therapy, 13
Absorbed dose pattern, 4, 10, 14, 15 Stationary beams of photons, 11
Isodose plan, 10, 15 Variation in target absorbed dose, 4, 14
Representation of a spatial by a set of planar dose distributions, Specification point, 14
4, 14 Stationary beams of photons, specification of target absorbed dose

Downloaded from http://jicru.oxfordjournals.org/ at Universite Laval on July 9, 2016


Beam(s), 10 in, 11
Arrangement of, 10 Intersecting beams, 11
Modifying devices, 10 Opposed, equally weighted beams, 11
Weighting, 10 Opposed, unequally weighted beams, 11
Complex treatments, general recommendations, for specification of Single beam, 11
target absorbed dose in, 14 Target absorbed dose, 3-5, 14
Computers, 1, 5, 10 Maximum, 5, 10
for calculation of absorbed dose pattern, 1, 5, 10 Mean, 5, 10
for obtaining patient anatomy, 1 Median, 5, 10
for recording patient data and treatment procedures, 1 Minimum, 5, 10
Dose rate, 17 Modal, 5, 10
Electron therapy, specification of target absorbed dose in, 14 To be specified for reporting, 10
Fields sizes, 10 Variation in, 4,J4
For SAD, 10 Target area, 3, 5
For SSD, 10 Target volume, 3
Hot spots, 5, 14 Contents of, 3
In vivo dose measurements, 10 Delineation of, 3
Irradiated volume, 4 Description of, 3
Lattice (points), 4, 5 Number of target volumes, 3
In three-dimensional dose calculations, 4 Variation of during treatment, 3
In two-dimensional dose calculations, 5 Time-dose pattern, 17
Moving beam therapy, specification of target absorbed dose in, 13 Gap, 17
Occult disease, 3 Number of fractions, 17
Organs at risk, 4, 14 Overall time, 17
Planning sections, 4, 15 Tissue heterogeneity, 10
Construction of, 4 Treatment technique, 10, 17
Contents of, 4 Beam geometry, 10
Orientation of, 15 Beam modification devices, 10
Selection of, 4 Beam weighting, 10
Radiation quality, 10, 17 Field sizes, 10
RBE, 17 Patient positioning, 10
Recording of absorbed dose delivered, 1 Radiation quality, 10, 17
Computers in, 1 Treatment volume, 4
Purpose of, 1 Tumor volume, 3

21
Downloaded from http://jicru.oxfordjournals.org/ at Universite Laval on July 9, 2016

You might also like