Humalyzer 3000: Specifications & More
Humalyzer 3000: Specifications & More
HumaLyzer 3000
>> System Specifications
>> Accessories
>> Consumables
Sample / Control / Sample type Serum, plasma, urine, CSF (depending on test)
Calibrator Handling Sample vessels All vessels with flow cell
10 mm square cuvettes
12 mm round tubes replacing the flow cell
Sampling volume Sample + reagents >250 µl
Pre-Dilution Manual
Post-Dilution Manual
Reaction Unit / Incubation Positions for incubation 18 positions external incubator incl.
Reaction volume (minimum) 250 µl
Reaction volume (maximum) 1000 µl
Incubation temperature 37 °C
Incubation time 5-999 s (in flow cell)
Reading Optical system Interference filters
Readings Mono- or bichromatic
Light source Halogen lamp (10W)
Spectral range 330-670 nm
Wavelengths pre-installed 340, 405, 505, 545, 580, 630 nm + 2 x optional
Max. # of wavelengths installed 8
Wavelength error (accuracy) ± 3 nm
Wavelength bandwidth (precision) Half bandwith < 10 nm
Detector Gallium-Arsenide-Phosphide photodiode
Absorbance range (linearity) -0.5 to 3.5 OD (Flow cell)
-0.5 to 2.5 OD (cuvette)
Accuracy detector < 0.5% (at 1 OD) to < 3% (at 3 OD)
Signal drift < 0.003 OD / h
Additional Accessories
Cat. No..
Disposable Macro Cuvettes Flow Cell Cleaner
External Keyboard 16700/11
Cat.No. 18052 Cat.No. 18222
Lab Timer 16700/12
Spanish Language (EPROM) 16700/31S