Professional Documents
Culture Documents
*Centre for International Child Health, University of Melbourne Department of Paediatrics, MCRI, Royal
Children’s Hospital, Parkville, Victoria, Australia, {School of Medicine & Health Sciences, University of
Papua New Guinea, {Ashdown Consultants, United Kingdom, 1International Union Against Tuberculosis &
Lung Disease, **Medical Research Council Laboratories, The Gambia and {{AirSep Corporation, Buffalo,
New York, USA
Abstract Hypoxaemia is a common problem causing child deaths in developing countries, but the cost-effective
ways to address hypoxaemia are ignored by current global strategies. Improving oxygen supplies and the detection
of hypoxaemia has been shown to reduce death rates from childhood pneumonia by up to 35%, and to be cheaper
per life saved than other effective initiatives such as conjugate pneumococcal vaccines. Oxygen concentrators
provide the cheapest and most consistent source of oxygen in health facilities where power supplies are reliable. To
implement and sustain oxygen concentrators requires strengthening of health systems, with clinicians, teachers,
administrators and technicians working together. Programmes built around the use of pulse oximetry and oxygen
concentrators are an entry point for improving quality of care, and are a unique example of successful integration of
appropriate technology into clinical care. This paper is a practical and up-to-date guide for all involved in
purchasing, using and maintaining oygen concentrators in developing countries.
following the introduction of pulse oximetry with unreliable power supplies are dis-
to detect hypoxaemia and oxygen concen- cussed. Some of the published experiences
trators as a reliable source of oxygen.5 in different countries are outlined in the
Oxygen concentrators were first devel- Appendix.
oped for military use in the 1950s. In
developed countries in the 1970s, they
began to be used to provide long-term home How Does an Oxygen Concentrator
oxygen therapy for adults with chronic lung Work?
disease. Their use has been extended over
the past 35 years; concentrators are now Air is 21% oxygen, 78% nitrogen and 1%
successfully supplying oxygen needs in other gases. Oxygen concentrators take air
hospitals in developing countries through- from the environment and separate the
out the world, including Egypt,6 Malawi,7 oxygen and nitrogen by means of a
Papua New Guinea,8,9 The Gambia,10 pressure swing adsorption (PSA) process.
Nigeria11 and Nepal.12 Concentrators have Room air is passed through a sieve bed or
been used successfully to supply oxygen to column filled with a regenerative, synthetic
anaesthetic machines.13–15 zeolite molecular sieve which is a beaded,
This article reviews oxygen concentrators inert, ceramic material. This sieve material
and their use in paediatric care in develop- allows the oxygen to pass through freely
ing countries: how they work, the various while the nitrogen is retained under pres-
types which are available, installation and sure. The cycle alternates typically between
maintenance requirements, difficulties two sieve beds (some concentrators have
encountered when using concentrators, more sieve beds), allowing one bed to make
and what has been necessary to sustain their oxygen while the other is depressurised,
use over many years. Some solutions to the freeing the nitrogen to exit the system
problem of using concentrators in hospitals through the exhaust muffler. The oxygen
FIG. 1. Delivery of oxygen from a portable oxygen concentrator to two children at once.
Oxygen concentrators 89
TEXTBOX 1. Key technical points and specifications for oxygen concentrators suited for use in a district hospital ward.
N The concentrator should achieve .85% oxygen concentration at a flow rate of up to 10 L/min.
N The concentrator should operate at a voltage and frequency suitable for the local power supply; this differs
between countries.
N For energy efficiency the power requirements should be close to 350 W for units providing 5 L/min, 410 W for
8-L/min units, and 600 W for 10-L/min units.
N The concentrator should have one or two outlets with individual flow controls and flow indicators.
N Outlet pressure should be no less than 55 kPa for units providing 5 L/min and 138 kPa for 8- and 10-L/min
units.
N Weight should not exceed 25 kg.
N An hour meter should record total hours of unit operation.
N Maximum operating altitude should be not less than 2000 m, with not less than 85% oxygen concentration at
maximum flow.
N Maximum operating temperature should be not less than 40uC.
N Maximum operating humidity should be not less than 95% relative humidity.
N A list of all spare or replacement parts and their costs for 40,000 hours of operation (e.g. compressor, sieve beds
and valve spares kits) should be provided.
N The concentrator should comply with ISO 8359:199620 and IEC 60601-121 and carry a CE marking.
N The concentrator should be fitted with an oxygen monitor which gives an audible/visual alarm when the product
gas is below 82% oxygen concentration
N A user manual intended for hospital use and a service manual with a troubleshooting guide should be provided.
N There should be a 60-month parts warranty at a defined cost.
N The unit should include either a 4-way flow splitter, together with all nozzles and blanking plugs, which can
deliver flows of 0.5, 1.0 and 2.0 L/min (Fig. 4), OR a flow meter stand (Fig. 2). Each flow meter should be
continuously adjustable with an accurate low flow scale acceptable for patient care, such as from 0.1 to 2 L/min (or
higher flow as appropriate).
90 T. Duke et al.
and an inverter. Their capital costs are alternative source of power, and they are
high but they are cheap to run and can expensive.
be cost-effective if properly designed and
maintained.
Installation and Maintenance of
Bedside Oxygen Concentrators
Limitations of Oxygen Concentrators
Oxygen concentrators are usually supplied
Absence of a reliable power supply is a with user manuals and maintenance infor-
common reason for failure of oxygen con- mation explaining how the apparatus works,
centrators. In the Solomon Islands, limita- its limits of performance and what regular
tions in power meant that effective use of maintenance is required. There will be
oxygen concentrators was possible only in instructions on how to unpack and install
hospitals in the major centres.15 In Sierra the concentrator. It is important to check
Leone, where concentrators were the only that the voltage shown on the packing list is
oxygen source, supply was frequently inter- correct for the power supply and that the
rupted by lack of mains power and by the plug fits the mains power socket.
high cost of fuel for generators.16 A concentrator should be positioned close
In The Gambia, because of power limita- to a mains power outlet in a cool part of the
tions, only two of 12 hospitals surveyed were ward, not in direct sunlight, with a good air
suitable for concentrators; in these settings, supply. The ward should be well ventilated
oxygen cylinders were preferred.10 In one and there must be good air circulation
remote rural hospital in The Gambia, around the concentrator itself: clearance
concentrators were run using solar power.17 on all sides should be in accordance with
However, there are important limitations to
the manufacturer’s instructions. The con-
current solar technology. Recent estimates
centrator should be in the shade and at least
are that the capital cost of solar panels is
1.5 m away from any source of heat.
$25,000, plus the cost of batteries, charger
and inverter, merely to run a concentrator
which costs less than $1000.
The use of batteries to supply 24-hour Dividing Flow to Multiple Patients
power to a concentrator from only 4 hours
of mains electricity supply is currently being There are two ways to divide the flow from a
explored (Bradley et al., unpublished data, bedside concentrator to multiple patients. In
personal communication, David Peel). This the past, most concentrators used a flow
would be sufficient for the majority of splitter, which allows a fixed flow of oxygen
district hospitals throughout the world, (0.5 or 1 L/min) through fixed-size orifices.
which typically have some mains power. If not used properly, however, they are liable
Many details of a battery system are still to to deliver extremely low flows to patients.
be evaluated but initial testing is encoura- If a flow splitter is to be used, it must be
ging. ensured that all four ports have either
A universal power supply (UPS) can be oxygen tubing or a blanking plug so that
useful in settings where power surges are oxygen does not escape through unused
common and interruptions of power brief. nozzles. It is important that a blanking plug
However, it is important to note that UPS be applied to any unused port of the flow
provides power for only a few minutes after splitter to avoid wasting oxygen and to
mains power goes off. UPS may buy time to ensure that the correct flow is delivered.
switch to an alternative power source, such At least one manufacturer of concentra-
as a generator, but is not in itself an effective tors now has a flow meter stand as an
Oxygen concentrators 91
alternative to a flow splitter. This has the With the oxygen-delivery system in place,
advantage of being more familiar to clinical the power supply cable should be connected
staff who are used to using flow meters on to the concentrator and then plugged into
cylinders. The flow is simply dialled up for the main power socket. An extension cable
each patient, as long as the aggregate flow should not be used. After the concentrator is
does not exceed the flow from the concen- switched on, it is normal for a continuous
trator. A concentrator can deliver oxygen to alarm to sound for up to 1 minute. If there is
one or more patients requiring low flow at a no power supply or other problems arise, an
time if it is equipped with a flow meter alarm will sound; in such cases, the user
stand (Fig. 2). Table 1 shows the equip- manual should be consulted.
ment needed to administer oxygen from an Once the concentrator is running, the flow
oxygen concentrator to up to five patients. rate should be adjusted to the required L/min
flow range by turning the control knob of the
TABLE 1. Equipment for the administration of oxygen flow meter anti-clockwise. Flow-rate mark-
to multiple patients from an oxygen concentrator.
ings are read at the centre of the ball. In flow
Description Quantity meters equipped with backlines, the proper
viewing angle is achieved when the two lines
Oxygen concentrator 1 appear as one. It can take up to 5 minutes for
Flow meter stand or 1 the concentrator to stabilize at or above the
flow splitter specified minimal performance of 85% O2.
Nozzles of 0.5 and 1 4 each
L/min if using flow spitter
There is no harm in using the unit while it is
Blanking plugs if using 3 building concentration, and most units reach
flow splitter therapeutic levels in less than 2 minutes.
Plastic tubing, 5-mm Up to 15 m64
internal diameter
Conduit to mount plastic Up to 15 m64
tubing on wall Testing After Installation
Non-crush plastic oxygen 8m
delivery tubing After all delivery tubing has been attached,
Prongs (or catheters) 4 the concentrator should be tested for flow
Back-up cylinder with 1
regulator and flow controller delivery and monitored for the oxygen
concentration produced.
92 T. Duke et al.
The flow is tested (a) by submerging the sing the tube under water. The difference in
oxygen catheter in water, or (b) by using an bubbling rate should be observed when the
in-line flow indicator. Assessing flow under flow is adjusted to between 0.5 and 1 L/min
water is a simple bedside test. Nasal prongs (Fig. 3). In-line indicators test flow from
are attached to each tubing outlet and the each concentrator orifice, but are rarely
flow from the prongs is checked by immer- used.
FIG. 4. An oxygen concentration status indicator or oxygen monitor. Note: The lights indicate adequate or below
normal oxygen concentration; a flow meter is shown to the left.
Oxygen concentrators 93
TABLE 2. Causes of low oxygen concentration delivered by a concentrator (as indicated by the oxygen monitor)
and their remedies.
Cause Remedy
the sieve beds and less oxygen produced. oxygen and argon to pass through freely.
When there is a measurable decrease in the As well as nitrogen, the sieve material has an
unit’s oxygen production or the operating attraction for water molecules. Water in the
system pressure, the compressor will require feed air is contained within a water zone of
rebuilding or replacement. the sieve bed and, on depressurisation,
The condition of the compressor’s bear- is released along with the nitrogen as part
ings also determines sound level. There are of the purge gas. A controlled amount of
four bearings in the compressor which allow sieve is intentionally exposed and allowed
the inner components to rotate. As the to ‘contaminate’ with water molecules form-
bearings become worn, the noise level ing the water zone. In a well designed
increases noticeably and service is required. and properly operating PSA system, the
water zone remains constant even when
exposed to high humidity and ambient
Valves temperatures.
Sieve material within the sieve bed can
The valves, or valving system, control the
become contaminated in a number of ways.
PSA process within the oxygen concentra-
A leak in the system, especially on the
tor. They control the pathway of the
product side, allows water molecules in
compressed air which feeds and pressurises
room air to come in contact with and
one sieve bed while the other sieve bed is
contaminate the sieve material. The sieve
allowed to depressurise and purge the
material needs to be tightly packed and
nitrogen through the exhaust muffler.
contained within the sieve bed. Most
Many types of valves (4-way, 3-way, 2-
manufacturers use a spring-loaded piston
way, rotary, spool and sleeve, diaphragm,
design which keeps pressure on the sieve to
and poppet) are found within different
prevent it from shifting. If the material
models of oxygen concentrators. Some
moves or migrates, the contaminated sieve
valves can be opened and serviced, but
in the water zone will move throughout
others are sealed. The type of valve in the
the bed.
concentrator will directly influence the
If the alternating PSA cycle of feeding
need for filtration. Owing to particulates,
and purging the beds is interrupted,
dust, dirt, smoke and surface corrosion
the beds can become contaminated. A
caused by humidity, highly sensitive valves
failure to cycle could be caused by a
with extremely small tolerances are very
defective valve or printed circuit board
susceptible to sticking. They therefore
(PCB) which controls the valves. This could
require a high degree of filtration. Two-
result in one bed being fed room air
way poppet valves have been designed and
continuously. This introduces more water
demonstrated to operate reliably in very
molecules than the water zone can contain,
humid climates with elevated tempera-
expanding the water zone into the active
tures. It is important that the oxygen
part of the bed. Care should be taken to
concentrator has a valving system proven
thoroughly leak-test a unit and repair any
to be suitable to the conditions in which the
leaks which can result in sieve bed failure.
unit is to be operated. To detect leaks, a solution of soapy water
should be sprayed or applied to all fittings
and connections, from the air compressor to
Sieve beds
the oxygen outlet.
The sieve beds hold the molecular sieve Argon, along with the oxygen, passes
material where gas separation takes place. freely through the sieve material and is
The material is regenerative and stores allowed to concentrate, so the maximum
nitrogen under pressure while allowing oxygen concentration obtainable will be
Oxygen concentrators 97
95.5%, the remainder being argon and a on historical reliability data. Many manu-
small amount of nitrogen. facturers provide technical training, either at
their facility or on-site.
To be able to properly support the oxygen
Printed circuit board (PCB)
concentrators in a facility, an inventory of
The PCB is the electronic control for spare parts needs to be maintained and
operating the valve(s) and alarm system. If should include the major components,
it fails, the unit may not cycle properly or components which wear and most, if
may not operate at all. Some systems have a not all, electrical components. Items to
lighting and diagnostic system on the PCB consider include compressors, compressor
to aid troubleshooting. mounts, sieve beds, valves, PCBs, on/off
Table 3 provides a guide to troubleshoot- power switches, power cords, hour meters,
ing some of the more common problems circuit breakers, fuses, cabinet fans and
with oxygen concentrators. Consult the ser- tubing, fittings and filters. This inventory
vice manual for additional troubleshooting of parts should be adjusted according to
suggestions. the number of concentrators being sup-
Regular contact with the manufacturer’s ported. This is where the manufacturer can
technical support department should be be of great assistance in recommending
maintained. This can be done via email quantities and parts commonly used.
and will allow receipt of updates on equip- These parts should be included in the initial
ment and service manuals. Technical sup- purchase contract. Concentrators will not
port offers troubleshooting assistance and run without regular maintenance and repla-
recommends spare parts inventories, based cement of parts.
98 T. Duke et al.
pneumonia dying in the hospitals after the and very dusty conditions. The cost of
system was introduced was 35% lower than treating one patient for 1 year using the
before (risk ratio 0.65, 95% confidence concentrator was 27% that of using cylin-
interval 0.53–0.80, p,0.0001).5 The cost of ders. Greater savings could have been
the programme was estimated to be $51/child achieved by using a flow splitter on the
treated, $50 per DALY averted, and $1670/ concentrator in order to treat several
additional life saved. The programme has patients at once. The use of the concen-
now been extended to 17 hospitals. trator was limited by interruptions to the
power supply. During power cuts, a portable
generator was used to power the concen-
Malawi trator.
A large national Child Lung Health programme
Malawi was one of the first countries with Nepal
limited resources to implement a national
Oxygen concentrators at high altitude
oxygen programme, beginning in 2002. This
was part of a highly successful National Child Oxygen demands are high at Kunde
Lung Health Project. Case fatality rates from Hospital, located near Mount Everest,
pneumonia have consistently fallen since the 3900 m above sea level. Oxygen is required
introduction of the programme. In 2007, an for childbirth, neonatal resuscitation, surgi-
evaluation of 15 hospitals using oxygen cal procedures and management of cardio-
concentrators showed that this technology pulmonary illness and altitude sickness (e.g.
had been successful in reducing hospital high altitude pulmonary or cerebral
mortality for pneumonia. Many units of one oedema). It is a 10-day walk from the
particular model which claimed to comply nearest road, and transport of cylinders is
with international standards were not work- difficult. In 1997, oxygen concentrators
ing. This emphasised that cheaper equip- were installed and connected to power
ment is not always the most cost-effective in supplied by a hydro-electric system with a
the long term, and that manufacturers claims petrol generator as back-up.12 An additional
need to be tested in the field. One challenge two concentrators are kept as back-up. No
was high staff turnover which made it equipment failures occurred over 3 years
difficult to sustain adequate staff skill levels. and the concentrators have replaced oxygen
Another challenge to optimal success was cylinders and portable hyperbaric chambers.
user fees; making hospital care free for
children in one hospital markedly increased
demand and access compared with hospitals The Gambia
where fees were charged.7 Donations and options where power is limited
In 2000, a teaching hospital in The Gambia
Nigeria received a donation of oxygen concentra-
tors.14 Unfortunately, all the equipment
Oxygen concentrators in a neonatal unit
very quickly stopped working. The main
Oxygen concentrators were introduced to a problem was lack of compatibility of fre-
neonatal unit in Nigeria in 1993 to over- quency: machines constructed for use with
come the limitations of cylinder oxygen.11 A 110 V 60 Hz electrical supply did not
model that met international standards was function with the 230 V 50 Hz local elec-
purchased and installed. It ran for 18 hours trical supply. The concentrators were sec-
a day for 3 years without breaking down, ond-hand and, although they had been
despite high daily temperatures (30–32uC) serviced before donation, there was not the
Oxygen concentrators 101
expertise in The Gambia to maintain the In remote district hospitals without reg-
equipment and the donor did not assess ular electricity, oxygen concentrators have
whether the electrical frequency was com- a limited role.10 In one such hospital in
patible. This emphasises the need for better The Gambia, oxygen concentrators were
planning, management and local participa- run successfully on solar power. Despite
tion around donations of equipment to a high initial outlay, running costs were
developing countries. It is also important small. Such a set-up requires a high level of
to ensure that any new equipment, either technical expertise.17 In The Gambia, the
given or bought, is supported by sufficient use of batteries to supply 24-hour power
technical expertise. Following this experi- to a concentrator from only 4 hours of
ence, the hospital organised a committee to mains electricity supply is currently being
oversee all donations, working with a non- developed (B. Bradley, et al., unpublished
government organization partner. data, personal communication, David Peel).