B-2000 Infant Incubator

STANDARD COMPOSITION:

●Servo Control Mode: air temp, skin temp;


● Instant Parameter Displaying: set temp, air temp ,skin temp, heater power rate;
● Audible & Visual Alarms: power failure, sensor failure, higher & lower temp, air& skin temp
override, fan failure, system failure;
● >37℃ Preset Mode: power failure reset memory, alarm silence, keypad lock, self-test function;
● Hood with dual slope: 6 hand-ports; left & right cabinets, top-shelf;
● RS-232 Output Connector (Use for network system ).

SPECIFICATION
POWER
Model B-2000-1E...............................................................................................................................110 / 120 Vac, 50Hz / 60Hz
Model B-2000-2E........................................................................................................................... ....220 / 240 Vac, 50Hz / 60Hz
Power Input ...................................................................................................................................................................... ≤400VA
Refer to IEC60601-1, this equipment belongs to ClassⅠ, Type BF. Its operation is continuous.
ALARM
Power Alarm
Power Failure……..Activates if primary power to the Incubator fails or the power wire is accidentally disconnected from the wall receptacle.
Fan Alarm
Fan Failure.................................................................................................................. ..................................Activates if Fan motor malfunctions.
Sensor Alarm
Air Sensor Failure............................................................................................................................Air Sensor malfunction, refer unit to service.
Sensor Out of Position........................................Activated when Sensor is not in correct position to perform calibration or continue operation.
Skin Sensor Disconnected.....................Activated during Skin Control Mode when Skin Sensor is disconnected from Temperature Controller.
Skin Sensor Failure........................................................................................................................ ......Activated when Skin Sensor malfunctions.
Deviation Alarm
High Air Temperature.......................................................Activated if displayed Air Temperature is 1.5℃ above set point.
Low Air Temperature........................................................Activated if displayed Air Temperature is 2.5℃ below set point.
High Skin Temperature...................................................Activated if displayed Skin Temperature is 1.0℃ above set point.
Low Skin Temperature.....................................................Activated if displayed Skin Temperature is 1.0℃ below set point.
Over Temperature Alarm
Over Air Temperature......................................................Activated when displayed Air Temperature was approach 38℃ for set temperature
≤37, ℃ or approach 40℃ for set temperature >37℃ in Air Mode.
Over Skin Temperature..................................Activated when displayed Skin Temperature was approach 38℃ for set temperature ≤37℃, or
approach 39℃ for set temperature >37℃ in Skin Mode.
System Alarm
System Failure ..................................................................................Activated if Temperature Control System malfunctions.

TEMPERATURE CONTROL RANGE AND CORRELATIVE SPECIFICATIONS


Air Temperature........................................20℃~37℃, for set temperature ≤37℃, 37℃~39℃, for set temperature >37℃
Skin Temperature......................................20℃~37℃, for set temperature ≤37℃, 37℃~38℃, for set temperature >37℃
Temperature Rise Time ....................................................................................................... .............................................. ≤30min
Temperature Variability...................................................................................................... ................................................. ±0.5℃
Temperature Uniformity...................................................................................................................................................... ... ≤1℃
Skin Sensor Precision.......................................................................................................................................................... ≤0.3℃
ENVIRONMENT REQUIREMENT
Environment Temperature
Operation Temperature Range.............................................................................................................................+20℃ ~ +30℃
Store Temperature Range......................................................................................................................................-40 ℃~ +55℃
Environment Humidity
Operation Humidity Range............................................................................................................................................30%~75%
Store Humidity Range......................................................................................................... ..........................................10%~100%
Atmospheric Pressure Range................................................................................................... ............................ 700hPa~1060hPa
OTHER SPECIFICATION
Noise Level Within Hood Environment.................................................................... ≤55dB(Environment noise below 45dB)
Carbon Dioxide (CO2) Level within the Hood.......................Less than 0.5% when a mixture of CO2 in air is delivered at 750ml/min at a
point 10cm above the center of the mattress.
Air Velocity Over Mattress..................................................................................................................................less than 0.35m/s
Infant Bed Max Load................................................................................................................................................. ..................10kg
Transfusion Shelf Max Load........................................................................................................................................................2kg
Cabinet Max Load ............................................................................................................ ........................................................... 6kg
FUNCTION OF “ ” SILENCE KEY
Make the alarm to be silence for 10 minutes by press this key, following:
Deviation Alarm
Sensor Alarm
Fan Alarm
FACTORY DEFAULT SETTINGS
Temperature Control Mode..................................................................................................... ..................Air Temperature Mode
Air Set Temperature............................................................................................................................................................ .........32℃
Skin Set Temperature...................................................................................................................................................................34℃
Temperature Units............................................................................................................ ................................................................℃