STANDARD COMPOSITION:
SPECIFICATION
POWER
Model B-2000-1E...............................................................................................................................110 / 120 Vac, 50Hz / 60Hz
Model B-2000-2E........................................................................................................................... ....220 / 240 Vac, 50Hz / 60Hz
Power Input ...................................................................................................................................................................... ≤400VA
Refer to IEC60601-1, this equipment belongs to ClassⅠ, Type BF. Its operation is continuous.
ALARM
Power Alarm
Power Failure……..Activates if primary power to the Incubator fails or the power wire is accidentally disconnected from the wall receptacle.
Fan Alarm
Fan Failure.................................................................................................................. ..................................Activates if Fan motor malfunctions.
Sensor Alarm
Air Sensor Failure............................................................................................................................Air Sensor malfunction, refer unit to service.
Sensor Out of Position........................................Activated when Sensor is not in correct position to perform calibration or continue operation.
Skin Sensor Disconnected.....................Activated during Skin Control Mode when Skin Sensor is disconnected from Temperature Controller.
Skin Sensor Failure........................................................................................................................ ......Activated when Skin Sensor malfunctions.
Deviation Alarm
High Air Temperature.......................................................Activated if displayed Air Temperature is 1.5℃ above set point.
Low Air Temperature........................................................Activated if displayed Air Temperature is 2.5℃ below set point.
High Skin Temperature...................................................Activated if displayed Skin Temperature is 1.0℃ above set point.
Low Skin Temperature.....................................................Activated if displayed Skin Temperature is 1.0℃ below set point.
Over Temperature Alarm
Over Air Temperature......................................................Activated when displayed Air Temperature was approach 38℃ for set temperature
≤37, ℃ or approach 40℃ for set temperature >37℃ in Air Mode.
Over Skin Temperature..................................Activated when displayed Skin Temperature was approach 38℃ for set temperature ≤37℃, or
approach 39℃ for set temperature >37℃ in Skin Mode.
System Alarm
System Failure ..................................................................................Activated if Temperature Control System malfunctions.