Professional Documents
Culture Documents
Brand : GE
Manufacture : China
Features
• Amorphous silicon non-tiled digital flat panel detector
• Patient-side technique controls with digital display and touch screen (e.g. mA, kVp, mAs)
• Excellent image quality and dose efficiency
• Z-shaped arm positioner with Duo-speed and motorized vertical tube/detector alignment
• Fast preview image
• Multi-resolution image processing capability
• Tissue Equalization
• DICOM 3.0
• Post-acquisition reprocessing for multiple views from single exposure
• Electronic angle meter
• Advanced image post processing tools
• Removable grid
• Self-check for system condition
• Multi language user interface
• AEC/3 ion chambers
Technical Specification
Digital Detector
• Detector nominal image area : 41 cm x 41 cm
• Grey scale : 14 bits
Positioner
• Variable SID : Variable SID distance movement with detents at SID distances of 100 cm and 180 cm
• Arm rotation : from +135° to -30° with digital indication on electronic angle meters
• Digital FP detector housing angulations : +/-45°
• Vertical movement : Duo-speed motorized vertical movement allows fast and easy patient positioning
• Ionization chamber : Three-field ionization chamber
• Arm Design : The Z shaped-arm design accommodates multiple digital imaging
Stretcher Table
Length 216 cm
Width 64 cm
Height from floor 70 cm
Table weight 55 kg
Patient capacity 220 kg
Workstation Specifications
• CPU : Intel Xeon 2.53G
• Memory : 6G memory / 512M Graphics
• External media : DVD RW
• Hard Disk : 250G
Workstation Features
• Worklist management
• Patient information management more than 600 User programmable APR
• Radiographic exposure control kV, mA, mAs
• Digital image / CR, film image selection
• Tissue Equalization
• Image Reprocessing
• EMI image enhancement
• Image transfer and storage.
• Dicom 3.0 connectivity
HF Generator
• Generator Type : This system is available with a high frequency generator.
• Normal Output Power: 50kW
• KV : 40 to 150 in 1 KV increments
• mAs : 0.50- 625 (mAs is 512 or less with AEC) in Renard scale.
• mA : mA selection consists of the following: 10-630 in Renard scale
• AEC : Yes
Tube Specifications
•Focal Spot Sizes : 0.6 mm/1.2 mm
•Heat Storage Capacity: -Anode 111,000 Joules, (150,000 heat units)