DT03000-R004-E5

**************************** SupportWare Manual **********************************

This software is a freeware.


We are not liable for any damage caused through use of this software.
Please use it on the user's responsibility.
If you use a USB-serial conversion adapter, we recommend that you purchase it from us
(The adapter is not needed when SURFCOM FLEX is used).
This software can be used on the following models.
・SURFCOM FLEX series
・SURFCOM 130/480
・HandySurf series

Main functions which can be used on each model are as follows.


(Some functions are subject to use restrictions depending on the model.)
Model Main Function
・Create inspection certificates
・Save measurement data files
SURFCOM FLEX series
・Back up measurement conditions
・Perform measurement
・Create Inspection certificates
SURFCOM 130/480 ・Save measurement data files
・Perform measurement
・Create inspection certificates
HandySurf series ・Save measurement data files
・Perform measurement

Before using the product, carefully read <Precautions on operation>, [Operation environment],
and [Operation method] below.

1
---------------------------------------------------------------------------------------------------------------------------------------

<Precautions on operation>
1. If you have not installed Microsoft Excel, this tool will not start.
2. When selecting files on a window, this tool can only read data files that were output by the
measuring instrument or SupportWare.
3. For any parameters for which calculation cannot be made, messages such as “Rz = No peak”
are displayed/printed on SURFCOM, while “Rz = 0.000 μm” is displayed/printed on the
SupportWare.
4. The name of curves may be different to the ones on the measuring instrument.
5. When using this software on HandySurf, you cannot paste load or amplitude distribution curves
in the inspection certificate.
6. When using this software on HandySurf, if you want to perform the measurement data text file
creation or inspection certificate creation after you measured in SURFCOM operation screen,
please be sure to set the digital output by the side of a measurement machine to "MANUAL".
(Please turn "OFF" No.7 of DIP switch 2.)
7. When using this software on SURFCOM 130/480 version 6.06 or earlier, please note the
following.
・It is impossible to paste BC Graph and ADC Graph in the inspection certificate.
・When creating an inspection certificate or file from saved data, set the output type to “Type 1” on
the RS-232-C condition setting window of the measuring instrument.
・It is Regarding parameters related to BC Graph (tp, Pmr, Pmr2, Pδc, Rmr, Rmr2, Rδc, Wmr,
Wmr2, and Wδc), parameter numbers (such as “1” for Pmr (1)) are different on SURFCOM
130/480 and the SupportWare. On the SupportWare, the number for the first selected
parameter is “0”, followed by serial numbers for the succeeding parameters (Example: when
three parameters, Rmr(2), Rme(4) and Rmr(5), are selected on SURFCOM 130/480, they are
numbered as Rmr_0, Rmr_1, and Rmr_2), respectively on the SupportWare).
・If you want to perform the measurement data text file creation or inspection certificate creation
after you measured in SURFCOM operation screen, please be sure to set the data output of
auto setting screen by the side of a measurement machine to "OFF".
8. When using this software on SURFCOM 130/480 version 6.07 or earlier, please note the
following.
・If you have changed the measurement conditions after the measurement or the reading of the
measurement data, when you create an inspection certificate or a measurement data text file
using RS232C communication, the changed conditions are output.
9. In the following software version of each measuring machine, it is necessary to use it in
combination with SupportWare Ver.1.5 or later.
・HandySurf series Ver.1.14 or later
・SURFCOM FLEX series Ver.1.29 or later
・SURFCOM 130/480 Ver.6.08 (3.408 / 4.408 / 5.408) or later

2
---------------------------------------------------------------------------------------------------------------------------------------
[Operation environment]
In order to use this software, the combination of Windows and Excel is required as shown in the
table below (O mark).

Microsoft Excel
OS ╲ Excel
2003 2007 2010 2013
XP(SP2,3) O O O -
Windows

Vista(SP1,2) O O O -
7 O O O -
8 - - O O
8.1 - - O O

The language of Windows and Excel needs to be the same. When the language is not same, it is
necessary to match a language setup of Excel with the language of Windows by [Office
Language Pack] (available for purchase from Microsoft).
When using the software on SURFCOM 130/480, in Microsoft Excel2003 and Microsoft
Excel2007, a maximum of 32,000 shape curves can be displayed/printed. Any curves in excess
of 32,000 cannot be displayed/printed.
---------------------------------------------------------------------------------------------------------------------------------------

[Operation method]
1. Turn on PC and start Windows.
2. Connect the measuring instrument and PC to be used with a communication cable.
3. Turn on the measuring instrument.
4. Open [SupportWare] from the shortcut on the desktop or the Start menu.
5. The main window appears. If you are using the selected model and [Current status] becomes
“Connected”, you can start to use. If you are using a different model from the selected one,match
the model.
6. If “Disconnected” is shown under [Current status], open the communication setting window and
set the port number checked in paragraph 7, and also set the same communication conditions as
those for the measuring instrument. Once [Current status] becomes “Connected”, you can start
to use.

3
<Steps to confirm COM port numbers on PC>

1. If you are using Windows XP, left click [Stat] → [Setting] → [Control Panel].
Then, change the screen to the classic display and double-click the system icon.
2. When the Windows system properties screen appears, click “Device Manager” of the
hardware tag.

3. Check the port number (“X” for COM X) from the port of the device manager window.

*When you use other OS, the steps you take is the same.

4
How to use this software on SURFCOM FLEX
(For details about each window, see page 9 and after.)

1. Create an inspection certificate using USB communication


1) To create an inspection certificate from measurement results currently displayed:
・ Open the inspection certificate output setting window and make output-related settings.
・ Make sure that measurement results are shown in the data processing section, and click
the [Inspection certificate simple creation] button in the Main window.

2) To create an inspection certificate from measurement data saved in the internal memory:
・ Open the inspection certificate output setting window, make output-related settings, and
select [SURFCOM data] under file selection.
・ When the data numbers for measurement data saved in the internal memory are
displayed, select the data number to be output.
・ Click the [Inspection certificate creation] button on the right bottom of the window.

2. Create an inspection certificate from measurement data files saved in a USB memory:
⇒Text files should contain “measurement conditions”.
1) To create an inspection certificate from text files saved in a USB memory:
・ Open the inspection certificate output setting window, make output-related settings, and
select [Text file] under file selection.
・ Designate “\\SFLEX\MEASDATA_T” in the USB memory as the reference folder.
・ When the file names of the measurement data saved in this folder are displayed, select
the data file name to be output.
・ Click the [Inspection certificate creation] button on the right bottom of the window.

2) To create an inspection certificate from text files saved by the SupportWare:


・ Open the inspection certificate output setting window, make output-related settings, and
select [Text file] under file selection.
・ Designate the folder containing text files as the reference folder.
・ When the file names of measurement data saved in this folder are displayed, select the
data file name to be output.
・ Click the [Inspection certificate creation] button on the right bottom of the window.

5
3. Back up measurement conditions:
1) To back up measurement conditions:
→Up to five measurement conditions saved in the internal memory can be saved in PC.
・ Display the measurement condition export window and click “condition data”(01~05) you
want to save.
・ Set the destination folder and destination file name and click the [Export] button.

2) To restore measurement conditions from the backup file:


→The “condition data” saved in PC can be set in the internal memory (01~05).
・ Display the measurement condition import window and select the file containing
measurement conditions that you want to set in the internal memory.
・ Click the condition data number at the destination.
・ Click the [Import] button.

4. Save measurement data files in PC:


1) To save measurement data saved in the internal memory as measurement data files:
・ Open the measurement data text file creating window and select output items.
・ When the data numbers for measurement data saved in the internal memory are
displayed in the “SURFCOM data” section, click the data number you want to save in PC.
・ Set the destination folder and destination file name and click the [Measurement data text
file creation] button.

5. Perform measurement:
1) To set the course of operation from measurement to the creation of an inspection certificate:
・ Open the SURFCOM operation window.
・ For saving measurement results currently displayed in the internal memory prior to
measurement, use the “Saving the displayed data” function.
・ Set post-measurement operation. Return, measurement data text file creation, and
inspection certificate creation can be performed automatically.
・ Make settings for data saving.
・ Click the [Execution] button.

6
How to Use this software on SURFCOM 130/480
(For details about each window, see page 9 and after.)

1. Create an inspection certificate using RS232C communication


1) To create an inspection certificate from measurement results currently displayed:
・ Open the inspection certificate output setting window and make output-related settings.
・ Make sure that measurement results are shown in the data processing section, and click
the [Inspection certificate creation] button in the Main window.

2. Create an inspection certificate from measurement data files saved in a PC card


→Text files should contain “measurement conditions”.
1) To create an inspection certificate from text files saved in a PC card:
・ Open the inspection certificate output setting window, make output-related settings, and
select [Text file] under file selection.
・ Designate the PC card as the reference folder.
・ When the file names of measurement data saved in this folder are displayed, select the
file name to be output.
・ Click the [Inspection certificate simple creation] button on the right bottom of the window.

2) To create an inspection certificate from text files saved by the SupportWare:


・ Open the inspection certificate output setting window, make output-related settings, and
select [Text file] under file selection.
・ Designate the folder containing text files as the reference folder.
・ When the file names of measurement data saved in this folder are displayed, select the
data file name to be output.
・ Click the [Inspection certificate creation] button on the right bottom of the screen.

3. Save measurement data files in PC


1) To save measurement results currently displayed as measurement data files:
・ Open the measurement data text file creating window and select output items.
・ Set the destination folder and destination file name and click the [Measurement data text
file creation] button.

4. Perform measurement
1) To set the course of operation from measurement to the creation of an inspection certificate:
・ Open the SURFCOM operation window.
・ Set post-measurement operation. Return, measurement data text file creation, and
inspection certificate creation can be performed automatically.
・ Click the [Execution] button.

7
How to Use this software on HandySurf
(For details about each window, see page 9 and after.)

1. Create an inspection certificate using RS232C communication


1) To create an inspection certificate from measurement results currently displayed:
・ Open the inspection certificate output setting window and make output-related settings.
・ Make sure that measurement results are shown in the data processing section, and click
the [Inspection certificate simple creation] button in the Main window.

2. Create an inspection certificate from measurement data files saved in PC


→Text files should contain “measurement conditions”.
1) To create an inspection certificate from text files saved by the SupportWare:
・ Open the inspection certificate output setting window, make output-related settings, and
select [Text file] under file selection.
・ Designate the folder containing text files as the reference folder.
・ When the file names of measurement data saved in this folder are displayed, select the
data file name to be output.
・ Click the [Inspection certificate creation] button on the right bottom of the window.

3. Save measurement data files in PC


1) To save measurement results currently displayed as measurement data files;
・ Open the measurement data text file creating window and select output items.
・ Set the destination folder and destination file name and click the [Measurement data text
file creation] button.

4. Perform measurement
1) To set the course of operation from measurement to the creation of an inspection certificate:
・ Open the SURFCOM operation window.
・ Set post-measurement operation. Measurement data text file creation, and
inspection certificate creation can be performed automatically.
・ Click [Execution] button.

8
Main window

* When not connected to the measuring instrument, functions are limited to the [Inspection certificate
output setting].

1. Inspection certificate creation

You can set up here the data concerning the Inspection certificate creation and create the
Inspection certificate.

1) Inspection certificate
You can input basic data to be described in the Inspection certificate to output. Strings of
characters typed in the Text box shall be described in the Inspection certificate as they are.

2) Output data
Displayed here are data concerning output of the Inspection certificate. It is applicable to
input respective data in the [Inspection certificate output setting].

9
3) Inspection certificate output setting
Press it to display a window to setup data output in the Inspection certificate and output the
Inspection certificate itself. As for the detail, refer to the [Inspection certificate output setting
window] in this document.

4) Inspection certificate simple creation


An inspection certificate is created based on the data currently displayed on the measuring
instrument.

2. Measurement condition data operation

It is applicable to import and export the Measurement condition of the SURFCOM FLEX.
This function cannot be used when connected to S130/480 or HandySurf.

1) Measurement condition import


Used to import the Measurement condition in the internal memory of the SURFCOM FLEX
from the Measurement condition file.
As for the detail, refer to the [Measurement condition import window] in this document.

2) Measurement condition export


Used to create the Measurement condition file from the internal memory of the
SURFCOM FLEX.
It is applicable to import again the Measurement condition in the internal memory of the
SURFCOM FLEX from such a created Measurement condition file. As for the detail, refer to
the [Measurement condition export window] in this document.

3. Measurement data text file creation

A measurement data file can be created.

1) File creation from saved data


A measurement data file is created from measurement data saved in the measuring
instrument. It is applicable to create the Inspection certificate from such a created
Measurement data file. As for the detail, refer to the [Measurement data text file creation
window] in this document.

10
4. Communication

Communication with the measuring instrument can be established and communication


conditions can be set.

1) Power Supply
SURFCOM FLEX : The current battery is displayed as described below.
Battery is full.

Battery is short.

Battery has no remaining power.


Connect the AC power supply.
Not connected to the SURFCOM FLEX.

S130/480 : It is not displayed.


HandySurf : [Power off] shall be displayed. If you press this button, you turn off the
power of HandySurf.

2) Current status
Press this button to switch between “Connected” and “Disconnect” with the measuring
instrument.
Button display shall change every time you press it as shown below.

・ When this button is pressed while connected with the measuring instrument,
communication is disconnected.
・ When this button is pressed while disconnected with the measuring instrument,
communication resumes.

Make sure that the measuring instrument is ON prior before connecting with the instrument.

3) Communication setting
Settings for communication with the measuring instrument can be made. For details, refer to
[Communication setting window] in this document.

5. SURFCOM operation

Press this button to open the window to operate the measuring instrument. For details, refer to
[Measuring instrument operation window] in this document.
11
6. Select the Model

Select the model which is connected with PC. Make sure that the model connected with PC is
always selected. Depending on the model, there are some differences as shown below.
SURFCOM FLEX S130/480 HandySurf
Measurement available unavailable unavailable
condition data
operation
Power supply Displayed Not displayed Not displayed
display
Measuring Selectable Selectable Selectable
instrument data (displayed data only) (displayed data only)
selection
Auto Return Can be set Can be set Setting impossible
(by detector)
Return available available unavailable

When creating an inspection certificate by reading a measurement data text file, select the
model which was used for the creation of that file. Otherwise, an inspection certificate may not
be created properly since measurement conditions and the information of output curves are
different depending on the model.

7. Exit

Press this to exit the SupportWare.

12
Inspection certificate output setting

1. Press the [Inspection certificate output setting] button in the Main window.

2. Inspection certificate output setting window is displayed.

3. On the [File Dialog], select whether you do only output setting, or choose either Measurement
result file or the Measurement data to be created by the Inspection certificate.
・ If you select [Setting only], all the output items become applicable to select without
displaying any list. Under this state it is not applicable to create an Inspection certificate.
Press the [Apply] button to save only the previous result. Refer to 13 below as for the
[Apply] button.
・ If you select [Text file], the text file list appears. Press Browse button to select a folder that
saves the Measurement data text file. Upon selected the folder, files saved in that folder
are displayed in the list. Choose the data from the list.
・ If you select [SURFCOM data], a measurement data list appears when connected with
SURFCOM FLEX. Choose a measurement data from the list. When connected with other
models, the data currently displayed on the measuring instrument is shown as selected.
* [Text data] cannot be read unless measurement conditions are contained.
* [SURFCOM data] can be selected only when communication with the measuring instrument
has been established.

13
4. Upon selected the data, data for the Measurement data are displayed in the [Saving data] of the
[Inspection certificate data]. Note also that the Output curve shall change pursuant to the
[Measurement type] and [Standard]. The output curve also changes according to the model
currently selected by the SupportWare.

5. In the [Inspection certificate] column of the [Inspection certificate data], it is applicable to set
basic data to be described in the Inspection certificate to be output. Strings of characters typed in
the Textbox shall be described in the Inspection certificate as they are.

6. In the [Output], you are requested to designate items to output in the Inspection certificate.
Item(s) applicable to designate shall change subject to selected Measurement data. Note further
any item not found in the Measurement data will not be output in the Inspection certificate even it
was designated.
When using SURFCOM 130/480, please be the same as the settings of print output item of
SURFCOM 130/480.
In addition, when you want to output the section parameter values using SURFCOM FLEX,
please choose the "section" in the output item of SURFCOM FLEX.

7. Under [Graph scale setting], the scale of the graph to be shown on the inspection certificate is
set.
Settings can be made for the following items.
・ Data maximum value standard: The scale is automatically adjusted so that the maximum
value of measurement data can be shown on the graph.
・ Definition of scale interval: Any amount of division can be set in μm or μinch.

14
8. In the [Inspection certificate format], you are requested to designate a format for the Inspection
certificate to output. Selected format shall determine the type and number of the graph
applicable to output. Note that graph having a less priority curve will not appear if you have
selected too much curves from the graph display area. You are requested to adjust the output
curve referring to the image displayed under the Format selection. Principles shall be;

・ In the horizontally long graph area, one of the following curves is shown in order of priority
as shown below. If S130/480 is selected by the SupportWare and model-specific
information is not contained in measurement data, only P and R curves can be output.

Profile Curve
Roughness Curve
Filtered Waviness Curve
Waviness Profile Curve
Rolling Circle Waviness Profile Curve
Filtered Rolling Circle Waviness Curve
P Curve
R Curve

・ The square graph area displays either [BAC] or [ADF] in order of priority as follows.

BAC (P)
BAC (R)
BAC (WC) (in SURFCOM 130/480, displayed as BAC(W) )
ADF (P)
ADF (R)
ADF (WC) (in SURFCOM 130/480, displayed as ADF(W) )

9. Designate a destination to save the Inspection certificate. Press the [Browse] button in the
[Inspection certificate save destination] to select a folder to save the certificate.

15
10. Set the name on the Inspection certificate to create. At first, Please determine whether to
manage by "Year, month, date, and time(Y/M/D/H/M/S)" or "Triple digits (000-999)".
After selecting either, Type in a file name in the Text box next to the [File name]. Shown below is
the name to be created. Note that the extension shall vary depending on the version of the
Microsoft Excel you use.
<When you manage by Year, month, date, and time>
Name put in the Text box + Year, month, date, and time of creation of the Inspection
certificate + extension
<When you manage by Triple digits>
Name put in the Text box + Triple digits + extension
Triple digits can be set optionally in range of 000-999. Moreover, at the time of file
preservation, the increment (+1) is performed.

*As of June 2013, shown below are extensions corresponding to respective version of the
Microsoft Excel.
・ “.xls” for Excel 2003
・ “.xlsx” for versions thereafter

11. Put a check on the [Auto Print] to automatically print the Inspection certificate upon creation.
Printing shall be executed according to the setting of the Microsoft Excel.

12. Press the [Create certificate] to create the Inspection certificate according to the setting.

13. Press [Apply] button to save currently displayed [Inspection certificate], [Output Curve],
[Inspection certificate format] and [Inspection certificate save destination] as previous values.
Those values shall appear on the display as initial values when you start the system next time.
They shall also serve as output setting when executed the [Inspection certificate simple
creation].

14. Press [Cancel] button to close the window.

16
Measurement data text file creation window

1. Press [File creation from saved data] on the Main window.

2. Measurement data text file creation window appears.

3. Select the File type. “Normal (txt)”should be usually selected. When outputting only the profile
curve data in particular, select “Only Profile Curve (tx1)”. With respect to tx1 format, see【tx1
Format File Structure】.
4. When connected with SURFCOM FLEX, data saved in the measuring instrument is displayed on
the list. Then, select the number for data you want to output as a text file. If connected with an
instrument other than SURFCOM FLEX, data displayed on the instrument is shown as selected.
Nothing is displayed on the list.

5. Data of selected Measurement data shall appear in the [Saving data].

6. Check those saved data and select items to output in the Text file from [Output]. You can not put
a check on any item that is not available in the currently selected data. Item that can be put a
check shall be output. (This is effective only when “Normal (txt)” is selected.)

7. Designate a destination to save the Measurement data text file. Press the Browse button on the
[Measurement data text file save destination] and select a folder to save the file.

17
8. Set the name on the Measurement data text file to create. At first, Please determine whether to
manage by "Year, month, date, and time(Y/M/D/H/M/S)" or "Triple digits (000-999)".
After selecting either, Type in a file name in the Text box next to the [File name]. Shown below is
the name to be created.
<When you manage by Year, month, date, and time>
Name put in the Text box + Year, month, date and time of creation of the file +
extension(.txt/.tx1)
<When you manage by Triple digits>
Name put in the Text box + Triple digits + extension(.txt/.tx1)
Triple digits can be set optionally in range of 000-999. Moreover, at the time of file
preservation, the increment (+1) is performed.

9. Press the [Measurement data text file creation] to create the file in designated folder.

10. Press [Apply] button to save currently displayed [Output] and [Measurement data text file save
destination] as previous values. Those values shall appear on the display as initial values when
you start the system next time. Here, if you want to do only setting of the data creation, left click
the [Apply] button and then [Cancel] button to close the window.
【tx1 Format File Structure】
In tx1 format, only the point sequence data of the profile curve is output. The file structure of x1
format is described below.

12.00000<CR><LF> ‥‥‥Evaluation length (Note.1)


32000<CR><LF> ‥‥‥Data points
125.4412<CR><LF> ‥‥‥Z coordinate values continue for the number of data points.
125.3832<CR><LF> The unit is “μm” or “μ” ” The number of decimal places is fixed at four
・ whether in the mm mode or in the inch mode.
・ (Note.1)
123.4455<CR><LF> Evaluation length is expressed in mm or inch. The number of decimal
places is fixed at five in the mm mode and seven in the inch mode.

18
Measurement condition import window

1. Press the [Measurement condition import] button on the Main window.

2. Measurement condition import window appears.

3. On the [Selection of the import data], select a Measurement condition file to import in the
SURFCOM FLEX. Press the Browse button to select the SURFCOM FLEX Measurement
condition file.

4. On the [Designation of the destination to import], designate the import destination in the
SURFCOM FLEX. Select a data number in the destination from the list of the [Condition data].
Asterisk (*) put after the number shows that the data are found.

5. Data of selected condition data are displayed in the [Saving data]. Check those data and press
the [Import] button to execute import.

6. Press the [Cancel] button to close the Measurement condition import window.

19
Measurement condition export window

1. Press the [Measurement condition export] button on the Main window.

2. Measurement condition export window appears.

3. On the [Selection of the export data], select Measurement condition data to export in the
SURFCOM FLEX. Asterisk (*) put after the number shows that the data are found.

4. Data of selected Measurement condition data are displayed in the [Saving data]. Check those
data and press the Browse button in the [Measurement condition file save destination] and
designate a save destination folder for the Measurement condition data.

5. Now put a name on the Measurement condition file to create. Type in a file name in the Text box
next to the [File name]. Shown below is the name to be created.

Name put in the Text box + Year, month, date and time of creation of the file + extension (.CD)

6. Press the [Export] button to create the Measurement condition file in designated folder.

7. Press the [Cancel] button to close the Measurement condition export window.

20
Communication setting window

1. Press the [Communication setting] button on the Main window.

2. Communication setting window appears.

3. Make communication settings under [Communication condition setting]. Settings should be the
same as those for the measuring instrument to establish proper communication. Make settings
while referring to those for the instrument.
Timeout sets the censored time data communication. If a timeout occurs during communications
errors, please change to a longer value.

4. Press [Apply] button to apply the setting.

5. Press [OK] button to close the Communication setting window upon applied the setting.

6. Press [Cancel] button to close the Communication setting window without applying the setting.

21
SURFCOM operation

1. Press the [SURFCOM operation] button on the main window.

2. The SURFCOM operation window appears.

3. When connected with SURFCOM FLEX, under [Before measurement], data currently displayed
on SURFCOM FLEX is saved in the designated number for the internal memory in SURFCOM
FLEX.
Asterisk (*) put after the number shows a place that has data. Designate a number of
destination and left click the Save button. When connected with other instruments,
[Before measurement] cannot be used.

4. On the [After Measurement], it is applicable to designate operation to execute after


measurement.
1) Following items are available in the [Operation].
・ Put a check on the [Auto return], Auto Return begins after measurement (this can be
performed only on SURFCOM FLEX – 50A and S130/480 version 6.07 or later).
・ Put a check on the [Measurement data text file creation] to create Measurement data text
file after measurement. Left click the [Measurement data text file creation setting] button
to set respective output items.
・ Put a check on the [Inspection certificate creation] button to create a certificate after
measurement. Left click the [Inspection certificate output setting] button to set the output
condition item and Inspection certificate format.

22
2) When connected with SURFCOM FLEX, under [Data save], designate a destination to save
data in SURFCOM FLEX. When connected with other instruments, [Data save] cannot be
used.
・ Select the [Auto assignment save] to automatically assign an empty place in the data of
the SURFCOM FLEX. Otherwise “00” (Not saved) shall be selected if no empty place was
found.
・ Select [Designated save] to designate a place to save the Measurement data in the
SURFCOM. Asterisk (*) shall appear in a place where there found data in the
SURFCOM FLEX.
・ Select [Not saved] not to save the Measurement data.

5. Press [Execution] button to execute measurement according to the set conditions.

6. Press [Return] button to execute return operation.

7. The following window appears on the SupportWare while you are operating the measuring
instrument by pressing the [Execution] or [Return] button. Press the [Stop] button on this window
to stop the instrument.

8. Press the [Cancel] button to close the measuring instrument operation window.

*******************************************************************************************************************
Feb. 2014 Tokyo Seimitsu Co., Ltd.

23