1. The calibration switch is located under the battery cover which is under the scale. Remove the
screw and the calibration cover inside the battery compartment. While pressing the switch,
turn on the power then you will go into the calibration mode.
2. The display will show flash “CALMod” three times and then “C.SPAN”.
4. The display will show “UNLOAd”. With nothing on the platter press the “SET” key. If you want
to cancel this span calibration, press the “ZERO” key.
5. Press the “SET” key, then the display shows “StAbLE”. Note that the scale should be stable.
6. When the display shows “LOAd”, place the full capacity of weight on the platter then
press the “SET” key.
7. The display shows “StAbLE” and then “End” then span calibration is finished. Remove the
weight.
A) If the scale is not 100% accurate, turn the scale off and repeat steps 3 to 9.
B) If the scale is 100% accurate, turn the scale off, replace the calibration plate and
screw, and replace the battery cover.