You are on page 1of 2

WORKLOAD FOR RADIOTHERAPY FACILITIES

1. Workload(W) for HDR (Ir-192/Co-60)


Parameters for calculation Ir-192 HDR Co-60 HDR
Maximum Activity 10 Ci 2 Ci
Maximum No. of Patients 8 8
treated per day
Average treatment time per 6 min 8 min
patient
Treatment days per week 5 5
Exposure Rate constant 0.50 1.33
(R/hr.Ci at 1meter )
Weekly machine ON time 4 hrs/wk 6 hrs/wk
Weekly workload (W) 20 cGy/wk 16 cGy/wk
(cGy/Wk)at 1 meter from
source

2. ​Workload (W) for Telecobalt


400cGy/pt. x 70 pt./day x 5day/week = 1.4 x10​5 ​cGy/week ~​1.5x10​5​ cGy/week at 1m

3. Workload (W) for Accelerators


i) Low energy (≤10 MV)

300 cGy /patient x 70 p/day x 5 d/week ​ ​= ​1.0 x 10 5​ ​cGy/week at 1 m

ii) Dual Energy Accelerator ( 6 & 15MV/18MV)

a) Workload for 15 MV/18 MV​ ​component=​ (300cGy /pt. x 35 pt./d x 5 d/week)


​ ​= 0.5 x 10 5​ ​cGy/week at 1 m

b). Workload for 6 MV component ​ ​= 1.0 x 10 5​ ​cGy/week at 1 m

4. Workload (W) for Cyberknife


Workload (W)= ​12.5 Gy/pt. x16 pt./dx5d/wk x (80/100)​2​ =​ 6.4x10​4​ cGy/wk at 1m

Workload for leakage (W​L​) = IMRT factor xW=15x6.4x10​5​ = 9.6x10​5​ cGy/wk at 1m

5. Workload (W) for Tomotherapy

Workload for Leakage (W​L​) ​= 3​ 2 pt. /day x 7 min/pt. x 5 day/wk x (85 / 100)​2​ x 860 cGy/min ​≈ 7.0 x 10​5​ cGy/wk at 1 m

Primary Workload (Wpri) = ​Leakage Workload (WL) / IMRT factor


​= 7.0 x 10​5​ cGy/wk / 16 ​≈ 5 x 10​4​ cGy/wk at 1 m