You are on page 1of 11

CLINICAL RESEARCH

"SF%FOUBM$PMPS.FBTVSJOH%FWJDFT
$*&$PNQMJBOU
Karl Martin Lehmann, DMD
Assistant Professor, Department of Prosthetic Dentistry,
University Medical Center, Mainz, Germany

Alessandro Devigus, DMD


Private Practice, Bülach, Switzerland

Christopher Igiel, DMD


Assistant Professor, Department of Prosthetic Dentistry,
University Medical Center, Mainz, Germany

Michael Weyhrauch,
Assistant Professor, Department of Prosthetic Dentistry,
University Medical Center, Mainz, Germany

Irene Schmidtmann,
Statistician, Institute for Medical Biostatistics, Epidemiology, and Informatics,
University Medical Center, Mainz, Germany

Stefan Wentaschek, DMD MSc


Assistant Professor, Department of Prosthetic Dentistry,
University Medical Center, Mainz, Germany

Herbert Scheller, DMD, PhD


Medical Director, Department of Prosthetic Dentistry,
University Medical Center, Mainz, Germany

Corresponding author: Karl Martin Lehmann


Assistant Professor, Department of Prosthetic Dentistry, University Medical Center, Augustusplatz 2, 55131 Mainz, Germany.

Tel: +49-6131-17-2871, Fax: +49-6131-17-5517. E-mail: karl.lehmann@unimedizin-mainz.de

324
THE EUROPEAN JOURNAL OF ESTHETIC DENTISTRY
70-6.&t/6.#&3t"656./
LEHMANN ET AL

Abstract 0.9999) for the L*C*h° color coordinates.


The regression lines of the L* and C*
Objectives: This in vitro study evaluated coordinates for device A were steeper
the accordance of dental color meas- than those of the CIE-compliant refer-
urement devices with a Commission In- ence system; the regression lines for
ternationale de l’Eclairage (CIE)-compli- devices B, C, and D were nearly paral-
ant reference system, by comparing the lel to those of the reference system, but
CIE lightness, chroma, and hue (L*C*h°) with an offset. The regression lines of
color coordinates of ceramic samples. the h° coordinates for all devices were
Methods: Four color measurement de- almost parallel to those of the reference
vices: Vita Easyshade® Advance (A); system, with slopes near 1. Excluding
DeguDent Shadepilot™ (B); X-Rite® the L* and h° coordinate measure-
Shadevision™ (C); and Crystaleye Olym- ments of device A, the measurements
pus (D), were compared with a CIE-com- with the devices exhibited deviations
pliant reference system by recording from the reference system that were
the L*C*h° color coordinates of ceramic greater than those expected by chance
samples matching the tooth colors of (P < 0.0002).
the Vita Linearguide 3D-Master®, under Conclusions: The dental color measure-
standardized test conditions. Differenc- ment devices assessed here offered
es between regression lines for the den- excellent reproducibility, but showed
tal color measurement device data and significant deviations from the CIE-com-
regression lines for the CIE-compliant pliant reference system regarding the
reference system data were evaluated. L*C*h° color coordinates.
Results: All devices offered high intra-
class correlation coefficients (0.9771– (Eur J Esthet Dent 2012;7:324–333)

325
THE EUROPEAN JOURNAL OF ESTHETIC DENTISTRY
70-6.&t/6.#&3t46..&3
70-6.&t/6.#&3t"656./
CLINICAL RESEARCH

Introduction order systems (eg, L*a*b* and L*C*h°)


established by the Commission Interna-
The reproducibility and accuracy of the tionale de l’Eclairage (CIE).18 Although
detection of color information and its their reproducibility has been frequently
reproduction represent a challenge for investigated, the dental color measuring
both clinicians and technicians in re- devices have rarely been compared with
storative dentistry.1-7 Traditionally, tooth a CIE-compliant reference system.13
color was visually defined with shade The purpose of this study was to
guides that were used for tooth color compare the L*C*h° color coordinates
determination and to aid communica- obtained using dental color measure-
tion between clinicians and technicians. ment devices with those obtained using
However, shade selection is a subjective a CIE-compliant reference system. The
procedure and is affected by many fac- null hypothesis was that the L*, C*, and
tors, including lighting conditions and h° color coordinates of the tested color
human perception.1,8-10 Although visual measurement devices are in accord-
shade matching has improved through ance with those of a CIE-compliant ref-
the use of shade tabs that are matched erence system.
with teeth or restorations in the order
of lightness, chroma, and hue,11 visual
shade selection still results in frequent Materials and methods
errors in appearance and matching res-
torations to natural teeth. Furthermore, Test conditions
communication with a laboratory tech-
nician through the use of inadequate Illumination lighting conditions (Just
shade guides adds a further source of Normlicht, Weilheim an der Teck, Ger-
error, owing to the lack of shade guide many) were set at 6500 K and 2000 Lux.
standardization by manufacturers.3 Daylight was excluded. A single trained
Color measurement devices of dif- operator recorded the L*C*h° color co-
ferent types, including spectrophotom- ordinates of each ceramic sample five
eters, colorimeters, and digital imaging times, with each of the color measure-
instruments, have been used clinically ment devices and with the CIE-compli-
for many years.12-17 These devices al- ant reference system.
low more reproducible and accurate
determinations of color coordinates and Color measurement devices
the quantification of color differences
between natural teeth and dental resto- The devices tested were Vita Easy-
rations. Thus, dental shade matching in- shade® Advance (A), DeguDent Shade-
struments have the potential to increase pilot™ (B), X-Rite® Shadevision™ (C), and
the success of color matching, commu- Olympus Crystaleye (D). The devices
nication, and reproduction.3 These elec- were used according to the manufactur-
tronic color measurement devices have ers’ instructions. The data obtained with
the ability to determine color information these devices are shown in Table 1.
in concordance with the majority of color

326
THE EUROPEAN JOURNAL OF ESTHETIC DENTISTRY
70-6.&t/6.#&3t"656./
LEHMANN ET AL
Data obtained with the color measurement devices: A: Easyshade Advance (Vita); B: Shadepilot (DeguDent); C: Shadevision

CIE-compliant reference system


Launch
date

2003

2011

2006

2001

2007
The CIE-compliant reference system was
a UV-Vis spectrophotometer equipped
with an integrated sphere (Evolution
resolution
Spectral

600®; Thermo Scientific, Madison, WI,


USA). The spectral resolution was set at
(nm)

N/A

N/A
2.0 nm over the entire range of wave-
2.0

25

10

lengths, from 380 nm to 780 nm. The


ment region

scanning speed was set at 60 nm/min.


measurement

measurement
Measure-

The room temperature was held constant


Complete

Complete

Complete at 25°C. Prior to the measurements, the


tooth

tooth

tooth
Spot

Spot

spectrophotometer was prepared for


30 mins to allow thermal equilibration.
The calibration was executed using a
ment range

diffuse 99% reflectance white standard


Measure-

380–780

400–700

400–720

(Labsphere, North Sutton, NH, USA).


(nm)

The recorded spectra were the basis for


N/A

N/A

calculating the CIE L*C*h° coordinates,


and OptLab-SPX 2.5 software (Ascanis,
Searchlight
White LED

illuminator

Überlingen, Germany) was used. Rec-


Tungsten
source

alibration was performed after each


Light

lamp

LED

LED
D65

measurement.
Two reflection
Calibration

Ceramic samples
standards
Reflection

Reflection

Reflection
standard

standard

standard

standard
Ceramic

Ceramic samples matching the tooth


colors of the Vita Linearguide 3D Mas-
ter® (feldspathic ceramic VM13 by VITA
Zahnfabrik; H. Rauter & Co, Bad Säck-
Operation

photometer

photometer

photometer

photometer
colorimeter

ingen, Germany) were used to per-


Imaging,
Spectro-

Spectro-

Spectro-

Spectro-
(X-Rite); and D: Crystaleye (Olympus).

mode

form successive measurements, with


repositioning. The samples were disc
shaped (diameter: 12 ± 0.5 mm; thick-
Evolution 600

ness: 2 ± 0.5 mm) and were stored in


Shadevision
Easyshade

DeguDent,
Shadepilot

Crystaleye

a light-tight box at 25°C. A specimen


Olympus,
Advance
Thermo,

X-Rite,

holder covered with a diffuse 99% re-


Vita,

flectance white standard reflectance


coating was used to fix the ceramic
Reference

samples.
Device
Table 1

D
A

327
THE EUROPEAN JOURNAL OF ESTHETIC DENTISTRY
70-6.&t/6.#&3t"656./
CLINICAL RESEARCH

Statistical analysis (x-axis), together with the corresponding


regression lines. The marked bisecting
Data were imported into Statistical Pack- line indicates complete accordance of
age for the Social Sciences (SPSS) 17.0 the color measurement device and the
(Munich, Germany) and Statistical Ana- CIE-compliant reference system. The
lysis System (SAS) 9.2 (SAS Institute, intraclass correlation coefficients and
2002–2008, Cary, NC, USA). Variance values for the slopes and intercepts of
components attributable to differences the regression lines and corresponding
in the color of the ceramic samples and 95% confidence intervals are shown in
to measurement errors for each type of Table 2. The mean deviations between
measurement were calculated fitting lin- the color measurement devices and the
ear mixed models with SAS PROC MIXED CIE-compliant reference system and
(SAS Institute). The mean differences the corresponding 95% confidence in-
between the measurements obtained by tervals are shown in Table 3. Only the
the electronic color measurement devic- mean values are shown, as all of the
es and those obtained by the CIE-com- tested measurement devices exhibited
pliant reference system were calculated. high intraclass correlation coefficients
We also determined 95% confidence in- of 0.9771 to 0.9999 for the L*C*h° color
tervals (CI) for these differences taking coordinates.
into account repeated measurements.
To assess the equality of the electronic Lightness
color measurements and the spectro-
photometric reference measurements, The regression lines of the lightness (L*)
t tests were used. The CIE L*C*h° color coordinates for the color measuring de-
coordinates were used for further regres- vices B to D were nearly parallel to that
sion analysis. The association between of the CIE-compliant reference system,
the color measurement devices tested with mean slopes of 1.02 to 1.14 (Fig 1).
and the CIE-compliant reference system The slope of the linear regression line for
was calculated through linear regres- device A (slope = 1.80; 95% CI; 1.59–
sion using SAS PROC REG. As repeated 2.01) was steeper than the slopes for the
measurements on the same specimen CIE-compliant reference system and the
by the same device differed very little, other three devices. Nevertheless, de-
we fitted the regression line to the means vice A showed the smallest mean devia-
of the repeated measurements. tion (-0.04; 95% CI; -2.10–2.01), because
device A underestimated low L* values
and overestimated high L* values. This
Results deviation was within a random variation
(P = 0.9669). The B and C devices gave
The L*C*h° color coordinates deter- regression lines almost parallel to that
mined with the dental color measurement of the CIE-compliant reference system
devices (y-axis) were plotted relative to (B: slope = 1.14; 95% CI; 1.11–1.17; C:
those obtained with the CIE-compliant slope = 1.02; 95% CI; 0.98–1.07), al-
spectrophotometric reference system though each had a different offset from

328
THE EUROPEAN JOURNAL OF ESTHETIC DENTISTRY
70-6.&t/6.#&3t"656./
LEHMANN ET AL

Table 2 Regression lines and regression coefficients of the color measurement devices compared with
the CIE-compliant reference system

Interval Intercept,

Interval Intercept,
Color coordinate

95% Confidence

95% Confidence

95% Confidence

95% Confidence
Interval slope,

Interval slope,
Measurement

upper bound

upper bound
lower bound

lower bound
correlation
coefficient
Intraclass

Intercept
device

Slope
L* A 0.9997 -54.09 -68.23 -39.94 1.80 1.59 2.01

L* B 0.9999 -6.91 -9.05 -4.77 1.14 1.11 1.17

L* C 0.9989 4.39 -9.52 7.26 1.02 0.98 1.07

L* D 0.9973 -9.54 1.52 -3.79 1.09 1.00 1.17

C* A 0.9998 7.07 1.71 17.00 1.67 1.27 2.07

C* B 0.9998 5.77 0.83 2.06 1.11 1.08 114.00

C* C 0.9986 1.49 -0.75 1.79 1.06 0.99 1.12

C* D 0.9951 0.52 1.56 9.60 0.81 0.60 1.02

h° A 0.9999 7.21 -13.11 27.53 0.91 0.64 1.18

h° B 0.9975 8.14 0.12 16.16 0.94 0.83 1.04

h° C 0.9886 8.50 -2.46 19.45 0.87 0.73 1.02

h° D 0.9771 28.94 11.01 46.87 0.66 0.42 0.89

Table 3 Mean deviations of the color measurement devices from the CIE-compliant reference system.
Color coordinate

95% Confidence

95% Confidence
Mean deviation

limit for mean

limit for mean


Measurement

upper bound
lower bound
deviation,

deviation,

t value
device

Pr > |t|

L* A -0.04 -2.10 2.01 -0.04 0.9669

L* B 2.40 2.06 2.75 13.96 <0.0001

L* C 6.02 5.79 6.26 50.77 <0.0001

L* D -3.52 -4.02 -3.02 -14.00 <0.0001

C* A 21.78 19.39 24.18 18.01 <0.0001

C* B 3.44 3.18 3.71 25.52 <0.0001

C* C 1.54 1.19 1.89 8.77 <0.0001

C* D 2.12 1.01 3.23 3.79 0.0002

h° A 0.27 -0.72 1.27 0.55 0.5859

h° B 3.24 2.84 3.64 16.04 <0.0001

h° C -1.17 -1.73 -0.61 -4.13 0.0001

h° D 2.91 1.91 3.91 5.75 <0.0001

329
THE EUROPEAN JOURNAL OF ESTHETIC DENTISTRY
70-6.&t/6.#&3t"656./
CLINICAL RESEARCH

95 55
(A) (A)
(B) (B)
90 50
(C)
(C)
(D)
(D) 45
85

mean of C* coordinates
mean of L* coordinates

40
80
35
75
30
70
25
65
20
60
15

55
10

50 5

45 0
55 60 65 70 75 80 85 0 5 10 15 20 25 30

mean of L* coordinates of CIE mean of C* coordinates of CIE


compliant reference system compliant reference system

Fig 1 CIE L* coordinates of the color measure- Fig 2 CIE C* coordinates of the color measure-
ment devices A to D compared with the CIE-com- ment devices A to D compared with the CIE-com-
pliant reference system. pliant reference system.

a line of complete accordance with the for the CIE-compliant reference sys-
reference system. Devices B and C tem, and their slopes were similar to
showed positive mean deviations (B: the respective slopes of the L* coor-
2.40; 95% CI; 2.06−2.75 and C: 6.02; dinate lines (A: slope = 1.67; 95% CI;
95% CI; -5.79–6.26). The regression line 1.27–2.07; B: slope = 1.11; 95% CI;
of device D was also nearly parallel to, 1.08–1.14; C: slope = 1.06; 95% CI;
and offset below, that of the reference 0.99–1.12; D: slope = 0.81; 95% CI;
system (D: mean deviation = -3.52; 95% 0.60–1.02). Color measuring device C
CI; -4.02 to -3.02). With the exception showed the smallest mean deviation:
of device A (P = 0.9669), the deviations (1.54; 95% CI; 1.19–1.89) compared
of the devices with regard to lightness with the other tested devices. The de-
were greater than those expected by viations of the chroma measurements
chance (P < 0.0001). for all devices were greater than those
expected by chance (P < 0.0001).
Chroma
Hue
The regression lines of the chroma (C*)
coordinates for all of the tested devic- Compared with the slope of the regres-
es (Fig 2) were located above the line sion line of the hue (h°) coordinates for

330
THE EUROPEAN JOURNAL OF ESTHETIC DENTISTRY
70-6.&t/6.#&3t"656./
LEHMANN ET AL

Discussion
90
(A)
(B) Electronic devices present new per-
(C)
(D) spectives in color matching.19 In partic-
85
mean of h° coordinates

ular, the strength of these systems lies in


their ability to quantify color information
80
by recording color coordinates (ie, CIE
coordinates), which define color space.
75
Across the CIE L*a*b* color space, dif-
ferences approximately represent the
70
same scope as humans’ color percep-
tion, thereby improving the interpreta-
65 tion of color measurements. In addition,
65 70 75 80 85
color measurement systems have the
mean of h° coordinates of CIE ability to define even small color differ-
compliant reference system
ences between two objects, ie dental
restorations and natural teeth. At this

Fig 3 CIE h° coordinates of the color measure-


level of color measurement, instruments
ment devices A to D compared with the CIE-com- can be judged based on their repeata-
pliant reference system. bility and accuracy. In the present inves-
tigation, the color measurement devices
displayed excellent repeatability for the
L*C*h° color coordinates (ICC > 0.9771)
and thus offer clinically acceptable re-
the CIE-compliant reference system peatability.
(slope = 1), the slopes of the h° linear Based on the nature of color science,
regression lines for all of the devices the true colors of the ceramic speci-
(Fig 3) were lower (A: slope = 0.91; mens used here are unknown. However,
95% CI; 0.64–1.18; B: slope = 0.94; comparisons between the color meas-
95% CI; 0.83–1.04; C: slope = 0.87; urement devices and a CIE-compliant
95% CI; 0.73–1.02; D: slope = 0.66; reference system, which presents the
95% CI; 0.42–0.89). Device A exhibited nearest approach to the true values,
the highest intraclass correlation coef- may allow an estimation regarding their
ficient (ICC = 0.9999) (Table 2) and accuracy.13 This is an important consid-
the lowest mean deviation (0.27; 95% eration, because for color coordinates
CI; −0.72 to 1.27) with respect to the to be useful for both defining colors and
CIE-compliant reference system (Table quantifying color differences, the meas-
3). The deviation for device A was urement system must be highly repro-
within random variation (P = 0.5859), ducible and accurate. Beyond this, ac-
whereas the deviations for devices B curately determined color coordinates
(P < 0.0001), C (P = 0.0001), and D allow comparisons among the results of
(P < 0.0001) were higher than those different color measurement systems.
expected by chance. This is advantageous given that different

331
THE EUROPEAN JOURNAL OF ESTHETIC DENTISTRY
70-6.&t/6.#&3t"656./
CLINICAL RESEARCH

color measurement systems are increas- devices to match CIE-compliant refer-


ingly used in both color matching and ence systems, perhaps by modification
reproduction processes; thus, accurate of the calibration process or transforma-
measurements allowing comparisons tion of the recorded data.
could improve communication between The null hypothesis that the L*C*h°
clinicians and technicians.1,13,19-21 How- color coordinates determined with the
ever, all of the measuring devices in the color measurement devices are in ac-
present study differed to various degrees cordance with those obtained with a
from the CIE-compliant reference system CIE-compliant spectrophotometric ref-
with regard to the L*C*h° color coordi- erence system was rejected.
nates. No devices in this study exhibited
regression lines for the L*, C*, and h° co-
ordinates that corresponded with those Conclusion
for the CIE-compliant reference system
with respect to slope and intercept. Within the limitations of this in vitro study,
In summary, dental color measure- the results indicate that the color meas-
ment devices should exhibit both high urement devices under investigation
reproducibility and accuracy, to allow have excellent reproducibility, but the re-
recorded color coordinates to be com- corded L*, C*, and h° color coordinates
pared among different color measur- clearly deviated from those obtained
ing systems and to provide an accurate with the CIE-compliant reference sys-
evaluation of color differences between tem. The color coordinates recorded by
two objects. It is suggested that manu- different devices are currently not com-
facturers adjust their color measurement parable.

References 5. Sproull RC. Color matching


in dentistry. Part II. Practical
and communication in re-
storative dentistry. Br Dent J
applications of the organiza- 2005;199:43–49; discussion
1. Chu S, Devigus A, Mieleszko tion of color. J Prosthet Dent 33.
A. Fundamentals of Color. 1973;29:556–566. 10. Ahn JS, Lee YK. Difference in
Quintessenz, 2005. 6. Sproull RC. Color matching the translucency of all-ceram-
2. Douglas RD, Brewer JD. in dentistry. Part III. Color ics by the illuminant. Dent
Variability of porcelain color control. J Prosthet Dent Mater 2008;24:1539–1544.
reproduction by commercial 1974;31:146–154. 11. Corcodel N, Rammelsberg
laboratories. J Prosthet Dent 7. Devigus A, Lombardi G. P, Moldovan O, Dreyhaupt J,
2003;90:339–346. Shading Vita YZ substruc- Hassel AJ. Effect of exter-
3. Seghi RR, Johnston WM, tures: influence on value and nal light conditions during
O’Brien WJ. Spectropho- chroma, part I. Int J Comput matching of tooth color:
tometric analysis of color Dent 2004;7:293–301. an intraindividual com-
differences between porce- 8. Donahue JL, Goodkind RJ, parison. Int J Prosthodont
lain systems. J Prosthet Dent Schwabacher WB, Aeppli DP. 2009;22:75–77.
1986;56:35–40. Shade color discrimination 12. Klemetti E, Matela AM, Haag
4. Sproull RC. Color matching by men and women. J Pros- P, Kononen M. Shade selec-
in dentistry. Part I. The three- thet Dent 1991;65:699–703. tion performed by novice
dimensional nature of color. J 9. Jarad FD, Russell MD, Moss dental professionals and
Prosthet Dent 1973;29:416– BW. The use of digital im- colorimeter. J Oral Rehabil
424. aging for colour matching 2006;33:31–35.

332
THE EUROPEAN JOURNAL OF ESTHETIC DENTISTRY
70-6.&t/6.#&3t"656./
LEHMANN ET AL

13. Lehmann KM, Igiel C, 16. Tung FF, Goldstein GR, Jang 19. Chu SJ, Trushkowsky RD,
Schmidtmann I, Scheller S, Hittelman E. The repeat- Paravina RD. Dental color
H. Four color-measuring ability of an intraoral dental matching instruments and
devices compared with a colorimeter. J Prosthet Dent systems. Review of clinical
spectrophotometric refer- 2002;88(6):585–590. and research aspects. J
ence system. J Dent 2010;38 17. Smith RN, Collins LZ, Naeeni Dent 38 Suppl 2:e2–16.
Suppl 2:e65–70. M, Joiner A, Philpotts CJ, 20. Khurana R Tredwin CJ,
14. Dozic A, Kleverlaan CJ, El- Hopkinson I, et al. The in Weisbloom M, Moles DR.
Zohairy A, Feilzer AJ, Khash- vitro and in vivo validation A clinical evaluation of the
ayar G. Performance of five of a mobile non-contact individual repeatability of
commercially available tooth camera-based digital im- three commercially available
color-measuring devices. J aging system for tooth colour measuring devices.
Prosthodont 2007;16:93–100. colour measurement. J Dent Br Dent J 2007;203:675–680.
15. Kim-Pusateri S, Brewer JD, 2008;36(Suppl 1):S15–20. 21. Ghinea R, Pérez MM, Herrera
Davis EL, Wee AG. Reliability 18. CIE Technical Report. Color- LJ, Rivas MJ, Yebra A, Para-
and accuracy of four dental imetry, ed 3. Vienna: Bureau vina RD. Color difference
shade-matching devices. J Central de la CIE, 2004. thresholds in dental ceram-
Prosthet Dent 2009;101:193– ics. J Dent 2010;38(Suppl
199. 2):e57–64.

333
THE EUROPEAN JOURNAL OF ESTHETIC DENTISTRY
70-6.&t/6.#&3t"656./
Copyright of European Journal of Esthetic Dentistry is the property of Quintessence Publishing Company Inc.
and its content may not be copied or emailed to multiple sites or posted to a listserv without the copyright
holder's express written permission. However, users may print, download, or email articles for individual use.

You might also like