Pousett B, Lizcano A, Raschke S.U. An investigation of the structural strength of transtibial sockets fabricated using conventional methods and rapid
prototyping techniques. Canadian Prosthetics & Orthotics Journal. 2019; Volume2, Issue1, No.2. https://doi.org/10.33137/cpoj.v2i1.31008
https://jps.library.utoronto.ca/index.php/cpoj/index
RESEARCH ARTICLE
*CORRESPONDING AUTHOR
Brittany Pousett, BSc, MSc, Certified Prosthetist,
Head of Research at Barber Prosthetics Clinic, 540 SE Marine Dr, Vancouver, British Colombia V5X 2T4, Canada.
Email: brittany@barberprosthetics.com
DOI: https://doi.org/10.33137/cpoj.v2i1.31008
Pousett B, Lizcano A, Raschke S.U. An investigation of the structural strength of transtibial sockets fabricated using conventional methods and 1
rapid prototyping techniques. Canadian Prosthetics & Orthotics Journal. 2019; Volume2, Issue1, No.2. https://doi.org/10.33137/cpoj.v2i1.31008
https://jps.library.utoronto.ca/index.php/cpoj/index
INTRODUCTION
A prosthetic socket is the interface connecting a applied the static portion of the standard and tested
person’s limb to the prosthetic components they use 10 transtibial sockets for P5 at Condition II. They
to interact with the environment. Typically, sockets compared five reinforcement materials and two
are manufactured from a plaster mold of a person’s resins using a 4-hole distal attachment system and
limb which is modified to create an optimized shape.1 found that all 10 sockets failed the ISO 10328
The socket is fabricated over the mold using standard, breaking at the attachment plate.7
materials including thermoplastics and laminated Graebner & Current5 investigated the strength of
composites. 3D scanning systems are an alternate different socket attachment methods for composite
method to digitize the patient’s limb and modify the sockets by applying the same portion of the standard
shape.1 Often, the optimized shape is milled by a as above. They found most attachment methods
computer numerically controlled (CNC) milling tested passed that aspect of the standard, especially
machine and the socket is fabricated using traditional when carbon reinforcement was used.5 Finally,
methods.1 MacKinnon8 used the same jig to perform the same
test on three different socket attachment methods for
As technology advances, the question emerges: ‘Is
thermoplastic sockets. He found two of the methods
this hybrid combination of digital scanning and
passed that portion of the standard when reinforced
design technology with traditional manufacturing
with fiberglass cast.8
methods the best approach?’. Rapid Prototyping
(RP) offers a time efficient way of turning the digital Gerschutz et al.9 took a different approach and
design into a physical socket. RP involves sectioning applied the static part of the ISO Standard 10328 to
the digital 3D socket design into thin slices, and evaluate sockets made in a variety of facilities. For
sending it to a 3D printer that builds the shape layer forces applied at Condition II for P6 (a mass being
by layer.1 Over the past three decades, several further above 100 kg), they found most check
groups have begun to create prosthetic sockets sockets and definitive laminated sockets and all
using rapid prototyping techniques.2,3,4 copolymer sockets failed the standard.9 These
studies show that there is a lot of variability in sockets
The Prosthetist is responsible for choosing
fabrication techniques and attachment methods that
fabrication techniques that provide adequate
result in sockets passing or failing this portion of the
strength and safety to their patients while maximizing
standard.
function.5 Currently, their decisions are not grounded
on an evidence-based foundation as minimal The goal of this project was to evaluate how 3D
evidence is available. printed sockets compare to traditionally fabricated
sockets made out of thermoplastics and laminated
Furthermore, the evaluation of prosthetic sockets is
composites. This was done by applying the static
not subject to any specific standard. ISO 10328:
portion of the ISO 10328 standard for a variety of
Prosthetics–Structural testing of lower-limb
weight limits and load both early and late in the
prostheses is the test standard that is most
stance phase following the same testing protocol as
commonly used to test prosthetic sockets.5 ISO
these previous authors to allow for comparison.5, 7, 8
10328 includes both static and cyclic strength tests
applied in two different loading conditions, Condition METHODOLOGY
I: instant of maximum loading occurring early in the
stance phase of walking, and Condition II: instant of Socket Fabrication & Alignment
maximum loading occurring late in the stance phase
This study chose to evaluate the strength of two
of walking, for three different weight limits, P3 body
different types of total surface bearing sockets, those
mass below 60 kg, P4 body mass below 80 kg, and
made for cushion liners attached to a 5R1 block
P5 body mass above 100 kg.6
(“cushion sockets”) and those made for a locking
Previously, this standard has been used in to liner attached distally via a 4-hole pattern lock
evaluate the strengths of different socket attachment (“locking sockets”) (FIGURE 1). It tested three
methods as this is frequently the point of failure in different fabrication materials, thermoplastic,
transtibial prostheses.5 Current, Kogleg & Barth7 laminated composited and 3D printed Polylactic Acid
Pousett B, Lizcano A, Raschke S.U. An investigation of the structural strength of transtibial sockets fabricated using conventional methods and 1
rapid prototyping techniques. Canadian Prosthetics & Orthotics Journal. 2019; Volume2, Issue1, No.2. https://doi.org/10.33137/cpoj.v2i1.31008
https://jps.library.utoronto.ca/index.php/cpoj/index
6 thermoplastic
3 tested for Condition I (early
2 for more information.
stance), one at each of P3, P4 and P5
sockets
3 tested for Condition II (late stance),
one at each of P3, P4 and P5
composite sockets
3 tested for Condition I (early
6 laminated
stance), one at each of P3, P4 and P5
6 3D printed PLA
3 tested for Condition I (early
stance), one at each of P3, P4 and P5
sockets
3 tested for Condition II (late stance),
one at each of P3, P4 and P5
36 sockets
6 thermoplastic
3 tested for Condition I (early
stance), one at each of P3, P4 and P5
sockets
FIGURE 1: The two types of total surface bearing sockets used; 3 tested for Condition II (late
stance), one at each of P3, P4 and P5
the left one was made for use with locking liners with a 4-hole 18 sockets for cushion liners with
pattern lock distally and the right one was designed for use with a distal attachment block
cushion liners and was attached to a distal attachment block and
composite sockets
reinforced with fiberglass wrap. 3 tested for Condition I (early
6 laminated
stance), one at each of P3, P4 and P5
The structural test model was manufactured from a 3 tested for Condition II (late
cast of an 80 kg male with a unilateral transtibial stance), one at each of P3, P4 and P5
surface bearing shape that was slightly conical, 3 tested for Condition II (late
allowing multiple sockets to be removed without stance), one at each of P3, P4 and P5
patella tendon to the distal end and 32 cm in The sockets were identically aligned in a Vertical
circumference around the patella tendon. One Alignment Jig (Hosmer, Fillauer, Chattanooga,
physical mold was fabricated identical to the modified United States) using the model patient’s alignment,
cast while the other physical mold had the Fillauer which was 5 degrees of flexion and 2 degrees of
cylindrical lock dummy (Chattanooga, United States) abduction. This alignment was done similarly to
shape incorporated into the bottom of the shape. previous studies, which do not follow the ISO 10328
Each mold was digitized using a Spectra Scanner recommendation that the alignment be set in the
(Vorum, Vancouver, Canada) and converted to a 3D “worst condition”.5,6 This decision was made to
print file by Additive O&P (Charlotte, United States). standardize the process using a realistic alignment
18 identically shaped sockets were fabricated from for the chosen model shape as this bench alignment
each of the models; 6 out of each different type of is repeatable whereas the specifics of what makes a
material using an identical process for each material worse case condition is unspecified and is
type. All sockets were fabricated at Barber inconsistent with previous studies. The alignment
Prosthetics Clinic by a Registered Prosthetic chosen will allow future tests to be compared to the
Technician. See TABLE 1 for detailed fabrication socket test done in this study. A 5R1 attachment
information. block was used for sockets made for cushion liners,
Pousett B, Lizcano A, Raschke S.U. An investigation of the structural strength of transtibial sockets fabricated using conventional methods and 2
rapid prototyping techniques. Canadian Prosthetics & Orthotics Journal. 2019; Volume2, Issue1, No.2. https://doi.org/10.33137/cpoj.v2i1.31008
https://jps.library.utoronto.ca/index.php/cpoj/index
as it has been reported to be the most commonly called the offsets (TABLE 2).6 A jig was fabricated for
used socket attachment methods in Canada.8 The these conditions, allowing easy and consistent setup
sockets for the locking liners were attached using the of the socket fixture for each test done. The vertical
distal 4-hole pattern in the lock mechanism. All load was applied using two 19 mm hitch balls
sockets were then attached to an Ottobock adapted to the top and bottom lever of the Tinius
(Duderstadt, Germany) titanium pyramid (5R54), a Olsen universal testing machine. To evenly distribute
23.2 cm aluminum pylon with a titanium connector the load through the socket a high-density urethane
(2R37) and a titanium tube clamp (4R52), all torqued resin (Smooth-CastTM 380, Smooth-On, Macungie,
to manufacturer’s specifications. United States) mold of the limb was made. A steel
rod was molded into the urethane to generate a
TABLE 1. Processes used to produce sockets. Every effort was
made to ensure an identical process was followed for each socket better grip between the top jig and the limb mold,
of the same material. using a 5/8 bolt. The setup can be seen in FIGURE 3.
Laminated TABLE 2. The offset values for the top and bottom load application
Method Thermoplastic 3D Printed
Composite points for all conditions and levels. The forward direction is
12 mm Orfitrans ½ oz. equivalent to anterior/posterior on the socket and the outward
Stiff (a transparent, Dacron Felt, direction is equivalent to medial/lateral on the socket.
Material
Condition II
Condition II
Condition II
Condition I
Condition I
Condition I
Print socket using
direction
Offset
Laminated fused Deposition Reference
Blister-formed
Process for Cushion
Completed
Set lock dummy in Print socket using
1 stage
standard alignment fused Deposition
lamination*3
Sockets
Blister-formed Modelling
Drilled 4
socket on Rockstock Max
holes
Drilled 4 holes V3 Printer with lock
distally for
distally for lock dummy included
lock
installation distally.
installation
*1
Layup: 1 layer 1/2oz dacron felt, 2 layers nyglass, 1 layer carbon
cloth from proximal trimline to 1½” distal to posterior brim trim line,
2 layers nyglass.
*2
Layup: 1 layer nyglass, 1 layer carbon clock from proximal
socket trimline to 1½” distal to posterior brim trimline and from 2”
proximal to the distal end to the distal end of the block, 3 layers of
nyglass.
*3
1 layer 1/2oz darcron felt, 2 layers nyglass, 1 layer carbon cloth
from proximal socket trimline to 1½” distal to posterior brim trimline
and from 2” proximal to the distal end to the distal end of the
socket, 2 x nyglass, 1 layer carbon cloth (as described above), 4
layers nyglass.
Test Step-Up
Pousett B, Lizcano A, Raschke S.U. An investigation of the structural strength of transtibial sockets fabricated using conventional methods and 3
rapid prototyping techniques. Canadian Prosthetics & Orthotics Journal. 2019; Volume2, Issue1, No.2. https://doi.org/10.33137/cpoj.v2i1.31008
https://jps.library.utoronto.ca/index.php/cpoj/index
Condition
Condition
Condition
Condition
Condition
Test
procedure Cushion Sockets attached distally via a 5R1
II
II
II
I
Settling test All 9 sockets passed the ultimate strength test for
force (N) 1024 920 944 828 736 638
both Condition I and II (FIGURE 4A&B), however the
Proof test
force (N) 2240 2013 2065 1811 1610 1395 failure levels and methods varied. For Condition I,
thermoplastic sockets yielded, laminated sockets
Ultimate
static test 4480 4025 4130 3623 3220 2790 cracked up the posterior wall and 3D printed socket
force (N)
broke circumferentially above the ScotchcastTM
(FIGURE 5). For condition II, 2/3 set-ups for each of the
materials failed because the pylon bent and yielded,
For the proof test, the settling test force was applied
often while the socket was left intact (FIGURE 5).
for 30 seconds before it was removed and the set up
rested at zero load for 30 seconds. The test force
was then smoothly increased at a rate between 100-
250 N/S to the proof test force for 30 seconds before
it was removed. All load times were recorded with a
stopwatch.
Pousett B, Lizcano A, Raschke S.U. An investigation of the structural strength of transtibial sockets fabricated using conventional methods and 4
rapid prototyping techniques. Canadian Prosthetics & Orthotics Journal. 2019; Volume2, Issue1, No.2. https://doi.org/10.33137/cpoj.v2i1.31008
https://jps.library.utoronto.ca/index.php/cpoj/index
A: Condition I
Weight (g)
Percentile
to Weight
Force (N)
Strength
Material
Failure
Socket
(%)
Laminated
13132 429 30.61 100 Crack - posterior wall
Composite
Laminated
12113 450 26.92 88 Crack - posterior wall
Composite
Material yield -
Thermoplastic 12566 709 17.72 58
anterior proximal gap
Material yield -
Thermoplastic 11608 738 15.73 51
anterior proximal gap
Material yield -
Thermoplastic 11264 734 15.35 50
anterior proximal gap
Circumferential
3D Printed PLA 7001 541 12.94 42 break above
ScotchcastTM
FIGURE 5. Typical failure methods for sockets made for cushion Circumferential
liners for the following conditions: (A) thermoplastic sockets at 3D Printed PLA 6725 542 12.41 41 break above
ScotchcastTM
Condition I (B) laminated composite sockets at Condition I (C) 3D
Printed PLA sockets at Condition I (D) sockets of all material types Circumferential
3D Printed PLA 5107 544 9.39 31 break above
at Condition II. ScotchcastTM
withstood approximately 71% of the force. For Thermoplastic 5958 733 8.13 51 Attachment
Condition II, the strength to weight ratio is less 3D Printed PLA 4355 547 7.96 50
Pylon, socket crack -
posterior
relevant due to the failure methods being in the
Circumferential
modular components. 3D Printed PLA 4143 543 7.63 48 break above
ScotchcastTM
Locking sockets attached distally via a 4-hole Thermoplastic 4434 733 6.05 38 Pylon
Pousett B, Lizcano A, Raschke S.U. An investigation of the structural strength of transtibial sockets fabricated using conventional methods and 5
rapid prototyping techniques. Canadian Prosthetics & Orthotics Journal. 2019; Volume2, Issue1, No.2. https://doi.org/10.33137/cpoj.v2i1.31008
https://jps.library.utoronto.ca/index.php/cpoj/index
A: Condition I
Weight (g)
Percentile
to Weight
Force (N)
Strength
Material
Failure
Socket
(%)
Laminated
11730 256 45.82 100 Lock mechanism
Composite
Laminated
10058 243 41.39 90 Pyramid adaptor
Composite
Laminated
10364 252 41.13 90 Crack - posterior wall
Composite
Circumferential
3D Printed PLA 10925 368 29.69 65
break - distal end
Circumferential
3D Printed PLA 9355 366 25.56 56
break - distal end
Circumferential
3D Printed PLA 9197 364 25.27 55
break - distal end
Material yield -
Thermoplastic 7650 446 17.15 37
proximal anterior gap
Material yield -
Thermoplastic 6091 431 14.13 31
proximal anterior gap
Material yield -
Thermoplastic 5241 423 12.39 27
proximal anterior gap
FIGURE 6. A (TOP): For locking sockets at Condition I (early
stance phase), all of the set ups failed above the ultimate strength B: Condition II
test values specified in ISO 10328; B (Bottom): For locking Laminated Material yield – lock
3526 249 14.16 100
sockets at Condition II (late stance phase), all the set ups failed Composite broken
below the ultimate strength test values specified in ISO 10328. Laminated Material yield – lock
3278 253 12.96 91
Composite broken
Laminated Material yield – lock
2818 252 11.18 79
Composite broken
Material yield around
Thermoplastic 3000 426 7.04 50
lock
Material yield around
Thermoplastic 2763 437 6.32 45
lock
Material yield around
Thermoplastic 2853 460 6.20 44
lock
Circumferential
3D Printed PLA 2243 366 6.13 43 break around distal
end
Circumferential
3D Printed PLA 2189 367 5.96 42 break around distal
end
Circumferential
3D Printed PLA 2020 365 5.53 39 break around distal
end
Pousett B, Lizcano A, Raschke S.U. An investigation of the structural strength of transtibial sockets fabricated using conventional methods and 6
rapid prototyping techniques. Canadian Prosthetics & Orthotics Journal. 2019; Volume2, Issue1, No.2. https://doi.org/10.33137/cpoj.v2i1.31008
https://jps.library.utoronto.ca/index.php/cpoj/index
100% of sockets passed at Condition I while 50% of sockets do not often break over the typical lifetime of
passed at Condition II. Results showed a statistically a prosthesis. In this study, laminated composite
significant strong association between test condition sockets had the highest strength to weight ratio and
and ultimate strength test results (Fisher’s Exact withstood the highest force. This was especially true
p=0.001; φ=0.577, p=0.001) for Condition I (at early stance) where the sockets
failed at approximately 3 times the ISO standard.
When comparing the ultimate strength test results
This strength is dependent on many factor as
based on manufacturing methods or weight
discussed below.9 Two other studies evaluated the
classification there were no statistically significant
strength of laminated composite sockets, for
differences, however, it is worth noting that 25% of
Condition II for people weighing over 100 kg, using a
sockets failed for each manufacturing method, all for
similar experimental set up. The first study found
locking liners at Condition II. Further testing may
their sockets failed between 1836 – 3160 N with the
produce more decisive results.
lamination failing at the pyramid attachment point.7
DISCUSSION The second study found that for socket reinforced
with carbon weighing between 616 – 795g, failure
This study evaluates the static strength of sockets occurred between 4247– 5663 N.5 Different material
made using a variety of fabrication techniques, lay-ups and socket attachment methods were found
including 3D printing, laminated composites and to increase the strength of laminated composite
thermoplastics. It employs the methodology used by socket.5,7,9 This study also concluded that modular
previous studies to test prosthetic sockets, outlined components began to fail above 5400 N of force.5
in ISO 10328. This study expended beyond this
The sockets tested at Condition II in the current
methodology as it looked at forces in both early
study, weighed between 410–450 g and broke
stance phase and late stance phase, which previous
between 4384 and 6505 N. This is approximately
studies do not do.
double the load reported by the first study and similar
Sockets Made for Cushion Liners to results in the second study, despite sockets in this
study weighing much less. Reasons for this include
Thermoplastic Sockets: material selection, layer order, laminating protocol
Thermoplastic sockets are used as diagnostic and socket attachment methods used. This is to be
sockets. The transparency of this material allows for expected, as studies have reported a large variation
visual inspection of the socket environment to guide in socket strength depending on who manufactures
the prosthetist in adjusting the socket shape. it.9 Findings of the second study were supported by
Thermoplastic sockets are heavier and have less this study which found that set-ups failed at the
strength than laminated composite sockets and, modular components; either because the distal
when tested to failure did not break catastrophically. attachment screw sheared or the pylon yielded. The
A study conducted by Mackinnon8 found that current study indicates that for forces applied at
thermoplastic sockets attached using resin and Condition II, an average force of 4800 N resulted in
ScotchcastTM to a 5R1 block failure at Condition II failure of the modular components. Further testing is
occurred at 4792 N. These results are comparable to required due to the small sample size.
the current study which found thermoplastic sockets 3D Printed Sockets:
using the same attachment methods failed, on
average, at 4910 N, but with the socket tested to P5 3D printing technology has been identified as having
failing at 5958 N. This increased strength could be the potential to benefit the production of prosthetic
from a variety of factors such as using a different sockets.3,4,10,11 For example, in the current study, the
socket shape, differences in plastic thickness, or 3D printed sockets took 9 hours and 9 minutes to
differences in the height and thickness of the print but required much less active time from a
reinforcement material. technician than traditional manufacturing methods.
While 3D printing allows for rapid prototyping of
Laminated Composite Sockets: custom designs, decreased manufacturing times and
Definitive sockets are made from laminated increased opportunities for collaboration, the main
composites. In daily clinical practice, laminated limitation continues to be the lack of standardization
Pousett B, Lizcano A, Raschke S.U. An investigation of the structural strength of transtibial sockets fabricated using conventional methods and 7
rapid prototyping techniques. Canadian Prosthetics & Orthotics Journal. 2019; Volume2, Issue1, No.2. https://doi.org/10.33137/cpoj.v2i1.31008
https://jps.library.utoronto.ca/index.php/cpoj/index
and regulation which may place patients at risk of Inherent in rapid prototyping is the adjustability and
receiving unsafe devices.3,4,12 One way to evaluate flexibility in the manufacturing methods – there are
the safety of 3D printed sockets is to explore how infinite designs, material choices and print settings
they compare strength wise to other fabrication that can be adjusted to influence the final product.
methods available. As in conventional manufacturing methods, this
variability will largely influence how strong sockets
For sockets made for use with cushion liners, this
are.9 More work is required to evaluate these
preliminary testing provides evidence that 3D printed
different parameters and give guidance to which
sockets are strong enough statically to be used with
choices result in better outcomes.
patients as they passed the standards for all weight
limits. However, when tested to failure, they failed at Sockets Made for Locking Liners
approximately half of the force of traditional
This study presents some preliminary evidence that
manufacturing methods. This is hypothesized to be
the use of cylindrical locks in prostheses should be
as a result of the material properties and manufac-
re-considered. Regardless of the manufacturing
turing process. While traditional manufacturing
methods used, the sockets with locks did not pass
methods involve either a solid sheet of plastic, which
the ISO standard for forces applied at Condition II,
yields before it breaks, or layers of sheets of
and in all cases the material around the lock either
reinforcement materials, often braided or weaved for
yielded or cracked. Unless modifications are done to
strength, 3D printing deposits material in layers, thus
relieve the stress concentration from this point or
making it inherently weaker. When force was
include additional reinforcement, these sockets may
applied, it appeared that the 3D printed material
fail when patients are using them. Alternatively, other
sheared between layers. There are many factors in
lock mechanisms may be an option as they result in
the printing process and design that may be able to
different distal socket shapes which may have less
increase the strength of these sockets such as by
concentrated stress points and may withstand higher
changing print orientation, infill pattern, adding
forces. This preliminary evidence supports that 3D
corrugations, changing material types or using a
printed sockets should not be used to create sockets
different type of printer. Subsequent work completed
with distal cylindrical locks.
by Campbell et al. provides preliminary support that
for sockets made for cushion liners 10% changes in Limitations
infill percentage does not affect the strength.13
Balancing the production of clinically-relevant and
For Condition I (early in stance phase), the sockets scientifically sound evidence with the feasibility of
failed well above the ISO standard, by cracking completing the research leads to several limitations
circumferentially about the ScotchcastTM reinforce- which need to be addressed. First of all, sockets are
ment. This indicates that the force is being concen- not subject to ISO 10328 testing. However, as the
trated there, which could be decreased using other components in lower limb prostheses are
different manufacturing methods described above. subject to this standard and as several previous
For Condition II (late in stance phase), the modular studies5,7,9 employed this methodology, it is
components failed before the sockets failed. Modular reasonable to use 10328 as an evaluation tool for
components are regularly used in clinical practice socket strength. The 2006 version of ISO 10328 was
without negative consequences. It is likely that since used as the 2016 version was not yet released at the
they are breaking before sockets are breaking, 3D beginning of testing.
sockets will survive the impact put on them statically.
The protocol outlined in the standard was followed as
Another issue raised is that 3D printed sockets break closely as possible, however several changes were
catastrophically, while other manufactured materials made in order to allow the results to be compared to
yield or tear more slowly. This catastrophic breaking other studies and to make it feasible to conduct in a
may present dangers to patients who could be timely and cost-efficient manner. At this time point,
injured in this process. Additional work is required to only the static portion of the structural tests were
further investigate this issue and determine if this conducted due to the length of time required to
drawback can be avoided, as well as to see how this cyclical testing. However, plans are in place to
material acts when going through cyclical testing. continue work on cyclical testing after addressing
Pousett B, Lizcano A, Raschke S.U. An investigation of the structural strength of transtibial sockets fabricated using conventional methods and 8
rapid prototyping techniques. Canadian Prosthetics & Orthotics Journal. 2019; Volume2, Issue1, No.2. https://doi.org/10.33137/cpoj.v2i1.31008
https://jps.library.utoronto.ca/index.php/cpoj/index
some of the limitations uncovered through this study. larger models with worse-case scenario alignment
When selecting a model, the standard does not and larger sample sizes. There is also an endless
specify a suitable size or shape, so a model was combination of material choices, design options and
chosen that fulfilled technical limitations and resulted 3D print parameters that can be explored. More
in a more expedient process. The model chosen was specifically, the distal attachment could be
slightly smaller than previous studies, but it is a strengthened, particularly in sockets made for
realistic mold, copied from a patient’s everyday locking liners, to extend the use of 3D printing to
prosthesis, and it was consistent across all samples. locking liners. Also, if design or material options
The alignment for the model was also taken from the could eliminate the catastrophic nature of the 3D
patient’s every day alignment, which is a fairly neutral printing failure, patient safety would be significantly
bench alignment. The standard outlines using a enhanced and the adoption of this technology would
worse-case alignment, but as this is not precisely be more widely accepted.
defined, and as all previous studies used a standard
Beyond static testing, cyclical testing of 3D printed
bench alignment, a standard bench alignment was
sockets must also be done to complete the testing
used here too. This allows for comparison with other
palate. Until information is known on how this
studies and consistency between samples.
material performs over time, clinicians cannot be
Finally, due to financial, time and resource confident that this manufacturing method will meet
constraints, two modifications were made to the the demands of ambulation. Future work should
testing protocol. First, only one sample was tested for focus on expanding the static testing that has been
each weight limit and condition. While the standard done to cyclical tests in order to present a more
recommends testing a minimum of two samples of complete picture of how this technology will work for
each condition, choosing only one sample allowed patients.
testing at both early stance and at late stance which
In addition to strength, there are many other factors
had not been completed in any previous study on
that can be explored including the personnel and
socket strength. This resulted in new findings and
material costs of using 3D printing over other
directions for future research to be uncovered.
manufacturing methods, the ease of fabrication,
Second, the wait time between the test force and the
quality and consistency of devices fabricated, and
ultimate strength test force was reduced. This
the methods of introduction of this method into
significantly reduced the testing time required,
clinical practice.
allowing for more samples to be tested.
Pousett B, Lizcano A, Raschke S.U. An investigation of the structural strength of transtibial sockets fabricated using conventional methods and 9
rapid prototyping techniques. Canadian Prosthetics & Orthotics Journal. 2019; Volume2, Issue1, No.2. https://doi.org/10.33137/cpoj.v2i1.31008
https://jps.library.utoronto.ca/index.php/cpoj/index
Pousett B, Lizcano A, Raschke S.U. An investigation of the structural strength of transtibial sockets fabricated using conventional methods and 10
rapid prototyping techniques. Canadian Prosthetics & Orthotics Journal. 2019; Volume2, Issue1, No.2. https://doi.org/10.33137/cpoj.v2i1.31008