Professional Documents
Culture Documents
1 (2019) Page | 1
_________________________________________________________________________________________________________________________
ABSTRACT
The purpose of the experiment was to determine the copper(II) concentration of an unknown solution by
applying concepts involving spectrophotometry. Specifically, UV-Vis spectrophotometry was employed in
the quantitative analysis, and the Shimadzu UVmini-1240 single-beam spectrophotometer was the device
used to measure absorbance. A deep blue colored complex was formed from the addition of concentrated
ammonia to the standard copper(II) solutions, and a calibration curve was drawn from the obtained
absorbance values. The linear equation of the calibration curve was y = 0.0004x - 0.0049, showing that
Beer’s law was obeyed. In the sample analysis, the absorbance values of the unknown fell within the
range reported by the standards, and the average concentration of the stock sample copper(II) was
224.75 ppm. Overall, the experiment was successful in applying spectrophotometry in determining the
concentration of an unknown copper(II) solution.
_________________________________________________________________________________________________________________________
UV-Vis spectrophotometry involves measuring the The objective of the experiment was to determine
transmission properties of substances across the the copper(II) concentration of an unknown
ultraviolet-visible spectral region (Harris, 2010), solution using Beer’s Law (2), which shows that
and thus, it has broad applications in both the absorbance, A, is directly related to the
qualitative and quantitative analysis. UV-Vis concentration of the light-absorbing species in the
spectrophotometry may be used to determine the solution (Harris, 2010).
identity of an unknown substance. Analyte
concentration may also be determined using Beer’s A=ε bc (2)
Law, which describes the linear relationship
between absorbance and concentration. Molar absorptivity, ε, describes how well a
substance absorbs light at a particular wavelength,
Applications of UV-Vis spectrophotometry include and it has units of ppm−1 ∙ cm−1. The width of the
heavy metal tracing and the analysis of proteins in cuvette that contains the sample is the pathlength.
clinical samples. In the experiment, however, UV- The pathlength, b, is expressed in cm, while the
Vis spectrophotometry was utilized in the concentration, c, is in units of ppm. Hence, the
quantitative analysis of copper(II) solutions, and a absorbance value is unitless (Harris, 2010).
single-beam spectrophotometer was the device
used to measure the transmission properties of the
sample.
The absorbance of the working standard The UV-Vis spectrophotometer used in the
copper(II) solution with the highest concentration experiment was the Shimadzu UVmini-1240
was measured against the reagent blank from 300 single-beam spectrophotometer. The components
nm to 700 nm using a UV-Vis spectrophotometer. are shown in figure 1. The UV-Vis
The analytical wavelength was then determined. spectrophotometer is used to identify various
organic, inorganic, and biochemical species given
2.3 Preparation of the Calibration Curve that they absorb ultraviolet and visible radiation
(Skoog, West, Holler, & Crouch, 2013).
The absorbance of the remaining working
standard solutions was measured at the analytical
wavelength obtained previously.
Figure 2. Block diagram a double-beam UV-Vis products with different absorbance values from
Spectrophotometer. the analyte. In terms of instrumental limitations,
3.3 Interpretation of Results when stray light from outside the instrument is
introduced, a decrease in absorbance value is
Measurements were made at the wavelength of observed. If the cells holding the sample and the
maximum absorbance, λmax, where the sensitivity reagent blank are of unequal pathlengths, a
of the analysis is highest, and the curve is flat, so deviation in Beer's law would also be observed.
there is only little variation in absorbance (Harris, Mismatched cells cause an intercept in the
2010). The absorbance of the standard copper(II) calibration curve, and the new equation is shown
solutions was measured at the previously obtained in (3), where k is the intercept.
analytical wavelength, as shown in table 1.
A=ε bc +k (3)
Table 1. Calibration Curve Data
Volume of Concentration Absorbance The deviation of the y-intercept from the
Working of Standard theoretical value is the difference of the intercept
Standard Cu(II), ppm from zero. In the experiment, the y-intercept of the
Solution, mL calibration curve is y = -0.0049, which shows the
2.00 100 ppm 0.034 occurrence of mismatched cells. This also indicates
4.00 200 ppm 0.069 that the reagent blank solution did not account for
6.00 300 ppm 0.109 all interferences.
8.00 400 ppm 0.144
10.00 500 ppm 0.185 In the sample analysis, the absorbance of the
unknown falls within the region covered by the
The calibration curve is shown in figure 2. The standards, as shown in table 2. The average
equation of the line of plot concentration of Cu(II) concentration of the stock sample copper(II) was
vs. absorbance was y = 0.0004x - 0.0049. The 224.75 ppm
positive slope indicates that the concentration is
directly proportional to the absorbance, and Beer’s Table 2. Sample Analysis
law is obeyed. However, the measure of linearity is Trial Absorbance Concentration
R2=0.9993. Though the value is still close to one, of stock sample
the relationship between the concentration and Cu(II), ppm
the absorbance is still not entirely linear. 1 0.089 234.75 ppm
2 0.083 219.75 ppm
3 0.084 222.25 ppm
0.2
Average 0.085 224.75 ppm
0.18
0.16
0.14 3.4 Sources of Errors
Absorbance
0.12
0.1 The presence of fingerprints or dirt on the non-
0.08 frosted side of the cuvette may cause false
0.06
0.04 absorbance readings. Mismatched cell placement
0.02 may also be a possible source of error. The failure
0 of the reagent blank to compensate for
0 100 200 300 400 500 600 interferences results in high absorbance readings.
Concentration of Standard Cu(II), ppm Thus the cuvette should be cleaned and
appropriately positioned in the cuvette holder to
Figure 2. Calibration curve: Concentration of Cu minimize errors. In operating the UV-Vis
(II) vs. Absorbance. spectrophotometer, the compartments must also
be tightly closed to avoid stray light, which results
The linearity of Beer’s law is challenged by certain in low absorbance readings. The multiple errors
real, chemical, and instrumental limitations. In mentioned could be the cause of the occurrence of
terms of real limitations, Beer’s law can only be a y-intercept and the non-linearity of the
observed in dilute solutions with less than 0.01M calibration curve.
because high concentrations cause deviations in
absorptivity due to the electrostatic interactions
between molecules in close proximity. In terms of
chemical limitations, the solutions must not
experience shifts in equilibria that may result in
S. Diaz / Chemistry 26.1 (2019) Page | 4
References
4. Conclusion and Recommendations
[1] Christian, G.D., Dasgupta, P.K., & Schug, K.A.
The measured absorbance values correlate with (2014). Analytical chemistry (7th ed.). Hoboken, NJ:
the concentration of the standard copper(II) John Wiley & Sons.
solutions. Though the calibration curve obtained
was not entirely linear and produced a y-intercept, [2] Harris, D.C. (2010). Quantitative chemical
the results still agree with Beer's Law. The non- analysis (8th ed.). New York, NY: W. H. Freeman.
linearity may only be attributed to human errors
such as failure to clean the cuvettes properly and [3] McMurry, J. (2016). Organic chemistry (9th ed.).
failure of the prepared reagent blank to account for Boston, MA: Cengage Learning.
impurities. Overall, the experiment was successful
in applying spectrophotometry in determining the [4] Petrucci, R.H., Herring, F.G., Madura, J.D., &
concentration of an unknown copper(II) solution. Bissonnette, C. (2017). General chemistry:
The extensive applicability of UV-Vis Principles and modern applications (11th ed.).
spectrophotometry to identify various inorganic, Toronto, CA: Pearson.
organic, and biochemical species (Christian,
Dasgupta, & Schug, 2014), with much accuracy and [5] Silberberg, M., & Amateis, P. (2015). Chemistry:
ease due to modern instrumentation, are what The molecular nature of matter and change (7th
makes this technique advantageous in quantitative ed.). New York, NY: McGraw-Hill Education.
chemical analysis.
[6] Skoog, D.A., West, D.M., Holler, F.J., & Crouch,
S.R. (2013). Fundamentals of analytical chemistry
(9th ed.). Boston, MA: Cengage Learning.
S. Diaz / Chemistry 26.1 (2019) Page | 5
Appendix
A. UV-Vis Spectrophotometer
Operating Procedures
1. The cuvette must be rinsed properly with distilled water. In the analysis of the sample, rinse the
cuvette at least three times with a small amount of the sample solution.
2. In most spectrophotometers, radiation passes through the clear (non-frosted) side of the cell.
Therefore, you should handle only the frosted portion of the cell when filling and rinsing with solution
and when placing it into the cuvette holder.
3. Be sure to wipe off any liquid on the outside of the cuvette as much as possible, including the entire
bottom half.
S. Diaz / Chemistry 26.1 (2019) Page | 6
B. Calculations
M wstd V wstd
M=
V sol' n
y = 0.0004x - 0.0049
y +0.0049
x=
0.0004
Trial 1
0.089+0.0049
x= =234.75 ppm
0.0004
Trial 2
0.083+0.0049
x= =219.75 ppm
0.0004
Trial 3
0.084+ 0.0049
x= =222.25 ppm
0.0004
Average
0.085+0.0049
x= =224.75 ppm
0.0004
S. Diaz / Chemistry 26.1 (2019) Page | 7
C. Raw Data
Absorption Spectrum
Wavelength at maximum absorption (λmax): 621 nm
Calibration Curve
Concentration of working standard Cu(II) solution: 2500 ppm
Sample Analysis
Volume of stock sample solution: 50 mL