EVITA INSITU OXY Dissolved Oxygen Transmitter CSI Specification
EVITA INSITU OXY Dissolved Oxygen Transmitter CSI Specification
PART 1 GENERAL
A. The method of measuring dissolved oxygen will be with replaceable Clark sensor.
1.3 Alternates
A. Performance Requirements
1. Measurement range: 0 to 50 mg/L
2. Accuracy: ±0.5% of full scale
3. Response time: 22 seconds
1.5 Certifications
A. Operational Criteria
1. Sample temperature: 0 to 40 °C
2. Operating temperature: -40 to 60 °C
1.7 Warranty
A. Scheduled maintenance:
1. Replace sensor every 2 years
2. Calibrate every 4 months
B. Unscheduled maintenance: Cleaning as needed based on environmental conditions.
PART 2 PRODUCTS
2.1 Manufacturer
Date SECTION 13400
Project Number MEASUREMENT AND CONTROL INSTRUMENTATION
Project Name Page 2
2.3 Equipment
A. Ion-selective membrane is antibacterial.
B. Transmitter enclosure is PBT/PC, rated at IP68
C. Stainless steel mounting bracket with pivot function.
D. Power supply needed: 12 to 30 Vdc
2.4 Components
A. Standard equipment
1. Transmitter
2. Universal signal converter (USC)
3. Mounting bracket
B. Included replacement parts
1. Membrane cartridge
C. Dimensions
1. Model 4100: 240 mm diameter
2. Models 4150 and 3150
a. Diameter: 50 mm
b. Length: 180 mm
D. Weight
1. Model 4100: 2 kg
2. Models 4150 and 3150: 1 kg
2.5 Accessories
A. Telecommunications module
B. Moisture barrier paste
PART 3 EXECUTION
3.1 Preparation
3.2 Installation
A. Contractor will install the sensor in strict accordance with the manufacturer’s instructions and
recommendation.
B. Manufacturer’s representative will include a half-day of start-up service by a factory-trained
technician, if requested.
Date SECTION 13400
Project Number MEASUREMENT AND CONTROL INSTRUMENTATION
Project Name Page 3
END OF SECTION