You are on page 1of 6

RAMAN SPECTROSCOPY / Near-Infrared 105

Near-Infrared
C Ross, Brock University, St Catharines, ON, Canada absorbed, and some is scattered. Most of the scat-
K T Carron, University of Wyoming, Laramie, WY, USA tered light has the same wavelength as the inci-
& 2005, Elsevier Ltd. All Rights Reserved. dent light, an effect known as Rayleigh scattering.
However, a small fraction of the scattered light
(B1 out of 108 photons) is shifted in wavelength,
Introduction resulting in inelastic scattering or Raman scattering.
In the Raman scattering, the differences in energy
When radiation passes through a transparent medi- between the incident and the scattered photons are
um, the species present scatter a fraction of the beam quantized, corresponding to the energy differences
in all directions. C.V. Raman discovered that incident within the vibrational or the rotational levels of the
light is inelastically scattered from a sample and molecules.
shifted in frequency by the energy of its characteristic In Raman spectroscopy, the wavelength shift is
molecular vibrations, a phenomenon known as Ra- defined as the difference in wavenumber, Dn, (cm  1)
man scattering. Today, laser photons over a wide between the observed radiation and that of the
range of frequencies from the near-ultraviolet (near- source. Two types of Raman scattering are present,
UV) to the near-infrared (NIR) region are used in Stokes and anti-Stokes lines. Stokes lines are found at
Raman scattering studies. By choosing wavelengths wavenumbers that are lower than the Rayleigh peak
which excite appropriate electronic transitions, res- while anti-Stokes peaks occur at wavenumbers grea-
onance Raman studies of selected components of a ter than the wavenumber of the laser excitation
sample or parts of a molecule can be performed. source. Anti-Stokes lines are generally considerably
Over the past few years, the range of excitation less intense than the corresponding Stokes lines. Flu-
wavelengths has been extended to the NIR region in orescence may seriously interfere with the observat-
which background fluorescence is reduced and pho- ion of Stokes shifts but not with anti-Stokes, thus
toinduced degradation of the sample is diminished. with fluorescing samples, anti-Stokes signals may
The NIR region of the spectrum extends from the sometimes be more useful despite their lower in-
upper wavelength end of the visible region at tensities. With highly fluorescing samples, it is more
B770–2500 nm (13 000–4000 cm  1). For many years, common to use a different excitation source, cir-
NIR excitation of Raman spectra was not conside- cumventing the production of fluorescence. This
red viable. However, now high-intensity NIR diode strategy frequently involves moving from the use of
lasers are easily available, making this region attrac- a visible laser source as an excitation source to one
tive for compact, low cost Raman instrumentation. lasing in the NIR region.
Furthermore, the development of low noise, high Figure 1 depicts an energy-level diagram showing
quantum efficiency, multichannel detectors (charge the sources of Raman and Rayleigh scattering. When
coupled device (CCD) arrays) combined with high- a photon having energy n0 collides with a molecule,
throughput single-stage spectrographs used in com- two different events can take place. If the scattered
bination with holographic laser rejection filters, have photon has the same energy as the exciting photon,
led to high-sensitivity Raman spectrometers. the Rayleigh scattering can be observed. The middle
Classical, spontaneous Raman scattering is a pow- arrow depicts the changes that produce Rayleigh
erful analytical tool that allows for the investigation scattering. No energy is lost in Rayleigh scattering,
of the qualitative and quantitative composition of thus the collisions between the photons and the mol-
biological, pharmaceutical, and environmental sam- ecule are termed elastic.
ples. The following discussion of NIR-Raman spec- The energy changes that produce Stokes and anti-
troscopy will begin with a general review of Raman Stokes emission are depicted on the left and right,
spectroscopy, followed by a description of NIR-Ra- respectively. The two differ from the Rayleigh radi-
man, with further discussion about instrumentation ation by frequencies corresponding to 7Dn, the
and applications of the NIR-Raman technique. energy difference between the first vibrational level
and the ground state. In Stokes scattering, the mol-
General Aspects of Raman ecule loses energy  Dn, reaching the vibrational
level n ¼ 1 and not returning to its vibrational state
Spectroscopy n ¼ 0. In this case, the scattered photon has energy
When a beam of monochromatic light is incident on n0  n1, where n1 is the frequency of the vibration. In
a sample, some of the light is transmitted, some is anti-Stokes scattering, the molecule originally in the
106 RAMAN SPECTROSCOPY / Near-Infrared

Electronic excited states

Virtual excited states

Optical field incident


on a molecule

=2

 = 1 Vibrational levels
∆ ∆
=0
0(cm−1) 0− ∆ 0 0+∆
Stokes Rayleigh Anti-Stokes

Figure 1 Energy level schematic of Raman spectroscopy.

n ¼ 1 state, is first promoted to a virtual excited state, orange are visible. Raman may be considered in this
then successively returns to the ground vibrational context as the inelastic equivalent of Rayleigh scat-
state, n ¼ 0. The energy of the scattered photon is tering and follows the same rule.
n0 þ n1 and the spectral line is shifted to a lower This theory predicts that Raman spectroscopy
wavelength and a higher frequency than the incident should be optimally performed in the near-UV not
radiation. However, since the majority of the mole- the NIR region. For example, excitation using
cules are in the ground vibrational state at room the blue line of an Ar þ laser at 454 nm will produce
temperature, the Stokes lines is stronger compared to essentially nine times more intense Raman scattering
the anti-Stokes lines. than 785 nm excitation. Yet in recent years, there has
been an overwhelming trend toward using NIR ex-
citation. This trend stems from a single interference
NIR-Raman Spectroscopy source and frustration in Raman spectroscopy: flu-
orescence. Fluorescence occurs when the excitation is
The Difference in Excitation
able to promote an electron into an excited electronic
Many books on Raman spectroscopy were written state. When impure samples or pure samples with
before NIR lasers became a viable excitation source. low energy transitions are placed in a Raman spec-
Traditionally, the relative magnitude of Raman spec- trometer, the spectrum recorded using visible excita-
troscopy is explained using the n4 factor, which in- tion often shows a broad fluorescence background,
dicates that the Raman intensity for normal Raman frequently devoid of any Raman features. A good
scattering varies with the fourth power of the rule-of-thumb is that the number of electronic states
observed frequency, an event dependent on the ex- decreases with energy. Thus NIR excitation is often
citation laser frequency. However, since Raman is preferred since it will produce high quality Raman
too weak to be seen by the naked eye without the aid spectra devoid of any fluorescence interference unat-
of filters, it is difficult to find good analogies for tea- tainable with a higher energy visible excitation source.
ching. The n4 process provides the explanation for A simple property of any spectrum can provide a
several common phenomena, the blue sky and red more quantitative picture to the problem of fluores-
sunset. Rayleigh proved that elastic scattering of light cence in Raman spectroscopy. The signal-to-noise
(Rayleigh scattering) is responsible for the blue sky. ratio (S/N) describes the quality of the spectrum. A
When the sky is observed during the day, the vision is high S/N is desirable for viewing a spectrum for both
overwhelmed by light scattered by atmospheric mol- qualitative analysis and for concentration determi-
ecules. Since high frequencies scatter more and blue nation with quantitative analysis. Obviously, the S/N
light is the highest frequency of the visible spectrum, can be increased two ways: an increase in signal or a
the sky appears blue. At sunset just the opposite oc- decrease in noise. For example, if the Raman spec-
curs and all the high frequencies are scattered away trum of cyclohexane is taken from a sample and has
to the west. Thus, only the low frequencies of red and been filtered to remove any particulates, the S/N ratio
RAMAN SPECTROSCOPY / Near-Infrared 107

will improve linearly with increasing laser power (for Consider the simple process of weighing coins on a
relatively low laser powers) or as the square root balance. Assume the weight of three coins needs to be
of the integration time. If one takes a spectrum of determined using a balance. The common laboratory
cyclohexane with a fluorescent impurity, the S/N balance has a standard error of e while a very good
will not increase linearly with laser power. Longer analytical balance has an e of 0.0001 g. Thus, if
integration times will be limited by the dynamic weighing 0.00170.0001 g or even 1070.0001 g, the
range of the detector. This originates from the de- error is not dependent on the mass being weighed.
pendency of the noise on the signal in spectroscopy. The three coins individually will weigh m17
When visible photons are detected, they produce a 0.0001 g, m270.0001 g, and m370.0001 g. The ac-
noise equivalent to the variance of the signal. If a curacy can be improved by weighing the coins in
sample has a background fluorescence that increases pairs. In this case, m1 þ m2 ¼ m1270.0001 g, m1 þ
with the laser power or integration time, noise will m3 ¼ m1370.0001 g, and m2 þ m3 ¼ m2370.0001 g.
also increase. This leads to the fundamental question The three weights of each pair can be subtracted to
of whether Raman spectroscopy is a true zero-base- find the individual weights. For example, m1 ¼
2(m12 þ m13  m23)7eT. The total error, eT, can be
1
line technique.
The answer is that Raman spectroscopy rarely fits found from the propagation errors in the three meas-
that description of zero-baseline. Most samples pos- urements. It is {[12(e1 þ e2  e3)]2}1/2. If the errors are
sess a background due to either inherent or impu- random, the equation can be simplified to m1 ¼
rity fluorescence. If NIR excitation circumvents or [14(e21 þ e22 þ e23)]1/2 and as stated above, the balance
strongly reduces the background, the reduction in has an error of e regardless of the mass. This leads to
intensity due to the n4 dependence of the Raman m1 ¼ [14(3e2)]1/2 or 31/2/2e or 87% of e. In the actual
signal becomes irrelevant compared to the noise pro- measurement, the weight of m1 would be m17
duced by a large fluorescence background if visible 0.000087. In general, the more coins that are
excitation is used. weighed, the better the result. If N coins were
A second problem arises from the shape of the weighed, the error would be reduced by N1/2/(N  1),
fluorescent background. If the background were or for a 1000 coins, the error would be reduced to
linear, it could be easily subtracted to produce a 3% of its original value. This may be perceived of as
high-quality spectrum, albeit noisy due to the noise the basis of Fourier Transform Raman spectroscopy
produced by the high background. However, in (FT-Raman spectroscopy). In FT-Raman, the Raman
addition to adding noise to the spectrum, the fluore- scattered light can be considered as waves that are
scence creates broad features in the background that allowed to interfere with each other to produce an
are often very difficult to remove. This leads to interference pattern (interferogram) that can be
serious difficulties in quantitative analysis because mathematically converted by an inverse FT into a
the Raman signal above the baseline is proportional spectrum. The benefit, called Fellgett’s or Multiplex
to the concentration, not the signal above a zero- Advantage, arises from putting the whole spectrum
baseline. It also can lead to small frequency shifts in at once onto the detector. This is exactly analogous
qualitative Raman spectroscopy. NIR-Raman spec- to placing the many coins on the balance at once to
troscopy often leads to higher quality spectra due to find the weight of an individual coin.
the minimization of background signals from the The controversy arises from the fact that optical
sample. detectors are not always like an analytical balance.
As the energy of a photon decreases, the prominent
detector noise becomes the thermal excitation of
The Difference in Detection
electrons in the detector. When this noise source
The S/N is a function of the excitation in Raman dominates, the detector is similar to the analytical
spectroscopy and as already discussed above, Raman balance. When the energy of the photon is large, the
spectroscopy can strongly benefit from using NIR thermal noise becomes insignificant. This transition
excitation. The detection method also affects the S/N occurs in the NIR region. The noise from high-energy
and is still perhaps a point of controversy among photons arises from their quantum nature. If
Raman spectroscopists. The controversy stems from individual photons were counted for 1-s intervals,
the quasiphoton nature of NIR light. One may prefer in one interval, 100 may be counted, in the next 95,
to consider the light particle in nature and count in the next 105, and so on. These differences occur
the photons or one may prefer to consider the light because the photons are random events and may
as a wave. To explain this duality and controversy, a happen more often during one period than the next.
freshman chemistry experiment can be used as an Statistics dictate that the error or standard deviation
illustration. of photons counted is the square root of the number
108 RAMAN SPECTROSCOPY / Near-Infrared

of photons being counted or as stated earlier, the Iodoanisole


signal is equal to the variance in the signal.
Using the weight of the coins, the error of the bal- 785nm
ance can be calculated. The same calculations can be
made as described above, except the error over two
weighings is 21/2e. Calculations show that for m1, the
error is 7(3/2)1/2e or 1.22e. The error has increased
instead of decreasing. What was once an advantage
for one detector is now a disadvantage for another.

Intensity
The result is two camps of Raman spectroscopists.
When a sample is very fluorescent, even in the far-red,
it is best to move further into the IR and use
633nm
FT-Raman spectroscopy. If fluorescence can be elimi-
nated or reduced, conventional NIR-excited dispersive
Raman spectroscopy provides the best option.

Instrumentation
For many years, NIR excitation of Raman spectra 600 700 800 900 1000
Wavenumbers (cm−1)
was not considered viable; however, things changed
with several key instrumentation developments. Figure 2 Effect of different laser wavelengths, 633 nm (He/Ne
CCDs now allow parallel detection of Raman spec- laser) and 785 nm (diode laser), on the reduction of fluorescence
in the analysis of iodoanisole.
tra with high quantum efficiency and low noise. Vol-
ume-holographic filters and gratings improve overall
throughput and provide elastic light rejection for fluorescence. The sample was iodoanisole and most
Raman spectrographs. In addition, economical, long- of recorded signal using visible excitation arises from
life laser diodes provide sufficient power for Raman the fluorescence of that compound. The lower curve
applications and optical fiber probes allow remote was obtained with conventional Raman equipment
chemical analysis by Raman scattering. using the 633-nm line from a He/Ne laser for exci-
tation. The upper curve is for the same sample re-
Laser Source
corded with a spectrometer equipped with a diode
The sources used in modern Raman spectroscopy are laser that emitted at 785 nm. Note the absence of
almost always lasers because their high intensity is fluorescence background signal.
necessary to produce Raman scattering of sufficient Two basic cavity configurations are used to con-
intensity to be measured with reasonable signal- struct and stabilize NIR-Raman laser sources. The
to-noise ratio. Argon ion lasers operate at 488 first design, the Littrow configuration, collimates
or 514.5 nm, krypton lasers operate at 530.0 or light from the laser diode and sends it to the grating
647.1 nm, and helium/neon (He/Ne) lasers operate at at a specific angle of incidence. The first-order dif-
632.8 nm. In NIR-Raman spectroscopy, the laser fracted light is sent directly back to the laser to
sources used are the diode laser (782 or 830 nm) provide optical feedback across a narrow wavelength
and Nd/YAG (1064 nm). NIR sources have two ma- range. The output beam is the light that is reflected
jor advantages over shorter wavelength lasers. The from the grating, thus the output is at an angle that is
first is that they can be operated at much higher approximately twice the angle of incidence from the
power (typically up to 50 W) without causing photo- optical axis. Because of its simple design, the exter-
decomposition of the sample. The second is that they nal-cavity length can be quite short, resulting in wide
are not energetic enough to populate a significant spacing of the external-cavity mode. This spacing
number of fluorescence producing excited electronic facilitates single-frequency operation where the goal
energy states in most molecules. Consequently, fluo- is to have only one external-cavity mode fall within
rescence is generally much less intense or nonexistent the bandwidth of the diffraction grating. The band-
with these lasers. The Nd/YAG line at 1064 nm is width of the grating is relatively large because the
particularly effective in eliminating fluorescence. The Littrow configuration uses a single-pass geometry.
two lines of the diode array laser at 782 and 830 nm In the second design, the Littman configuration,
also markedly reduce fluorescence in most cases. the collimated laser light strikes the grating near the
Figure 2 provides an example where the diode grazing incidence so that the diffracted order does
laser source completely eliminates background not return to the laser directly. Instead, the diffracted
RAMAN SPECTROSCOPY / Near-Infrared 109

light is reflected by a mirror, diffracted by the grating array with n pixels is equivalent to n spectrometers
a second time and then returned as optical feedback accumulating separately or one spectrometer wor-
to the laser diode. The advantage of the double-pass king n times.
geometry is that the grating bandwidth is less than
half of what it is in the Littrow case but the external
cavity length tends to be longer. In practice, the Litt-
man configuration is usually selected when narrower-
Applications
frequency operation is desired. NIR-Raman spectroscopy has been used for a number
of applications and is particularly useful for biologi-
cal and biomedical uses. Fluorescence has been a
Spectrometer
limiting factor for much Raman analysis of biologi-
A Raman spectrophotometer analyzes the radiation cal samples, particularly whole-cell or whole-tissue
scattered by molecules when they are illuminated samples. NIR excitation reduces interference from
with monochromatic exciting radiation. Currently, fluorescence and decreases photoinduced degradation
most Raman spectrometers are either FT instruments of the sample, enabling researchers to obtain spectra
equipped with cooled germanium transducers or for a variety of biomaterials and living cells.
multichannel instruments based upon CCDs. These Interest in both in vivo and in vitro use of NIR-
transducers, in contrast to photomultiplier tubes, are Raman spectroscopy of tissues for diagnostics con-
sensitive to radiation at 785 nm produced by diode tinues to grow. NIR-Raman has been evaluated as a
lasers, which provide Raman excitation of many diagnostic tool for cervical precancers. Although
compounds without significant fluorescence. CCDs spectra could not be collected in vivo due to long
are not sensitive to the 1064-nm radiation from an integration times, NIR-Raman was found to have
Nd/YAG laser. specificity that was superior to standard method-
Figure 3 is a schematic representation of a typical ologies, such as coloscopy and cytology, for differ-
NIR-Raman dispersing spectrometer with a CCD in entiating squamous intraepithelial lesions (SIL) from
the Littman configuration. The source is a diode laser non-SIL. Several research groups have described
and filter system that yields radiation at 785 nm. This NIR-Raman instrumentation with CCD detection
beam is focused on the end of the excitation filter by for biological tissue analysis, both in vivo and in
means of a lens, and the Raman emission is trans- vitro. In a specific study, an 810-nm excitation
mitted through the collimating lens. A diffraction wavelength was used for in vitro laboratory analysis
grating disperses the radiation and reflects it onto a of aorta tissues. Subsequent work showed that flu-
CCD that is made up of a large number of pixels. orescence interference in tissue Raman spectra could
These pixels store the charge produced by the light be reduced even further using 830-nm excitation
quanta. Since the spectral information seen by the without compromising CCD sensitivity. The useful-
individual pixels is accumulated simultaneously, an ness of NIR-Raman coupled with CCD detection for

Focusing lens

Grating CCD

Lens Beamsplitter Lens


Sample

Bandpass filter Mirror


Collimating lens
½ Waveplate

Diode laser @ 785 nm


Mirror

Figure 3 Typical Raman dispersing spectrometer with a CCD in the Littman configuration. (Courtesy of DeltaNu, Laramie, WY.)
110 RAMAN SPECTROSCOPY / Surface-Enhanced

studying cancerous changes in the colon, urinary drastically reduced at NIR wavelengths. In addition
bladder, breast, and soft tissue sarcomas has also to food processing, NIR-Raman has been used for
been examined. determination of food quality as the main com-
NIR-Raman has found applications in the phar- ponents of food (carbohydrates, proteins, and lipids)
maceutical industry. McCreery and colleagues re- all show characteristic Raman lines using NIR
ported the use of the technique for identification of detection.
pharmaceuticals inside amber vials. Even with the The pairing of SERS with NIR detection has
signal attenuation through the glass, adequate spec- proved to be a powerful DNA detection technique.
tra were obtained for determination of vial content NIR–SERS was reported to provide excellent dis-
with 1–60 s integration times. Using a library of crimination against fluorescent interference and was
spectra, identification of the pharmaceuticals in the nonresonant with most molecules. This allowed
vials was performed and identification was found to greater excitation intensities without photobleaching
display accuracy between 88% and 96%. This work or destruction of the analyte. In one study, Kneipp
demonstrated the potential of NIR-Raman for online and colleagues used this technique in rapid DNA se-
process monitoring. quencing to detect single-molecule DNA bases or
The environmental community has also found uses nucleotides. NIR–SERS provided a method for de-
for NIR-Raman analysis. A continuous method was tection and identification of the single DNA base,
developed by Weissenbacher and colleagues to detect adenine, without any additional labeling. This study
trace organic pollutants using flow injection analysis reported trace detection levels of adenine and AMP
and surface enhanced Raman spectroscopy (SERS). with well-resolved spectra.
This method uses NIR excitation and FT-SERS de-
tection to detect parts per million of pesticides in See also: Raman Spectroscopy: Instrumentation;
aqueous solutions. In this study, the authors describe Surface-Enhanced.
the simultaneous detection of two pesticides, carbe-
nazim and metazachlorine.
A variety of industrial processes benefit from the Further Reading
use of NIR-Raman spectroscopy. In a study exa-
Hanlon EB, Manoharan R, Koo TW, et al. (2000) Pros-
mining unleaded petroleum gasoline, with excitation
pects for in vivo Raman spectroscopy. Physical and
at 514 nm from an argon laser, no Raman features
Medical Biology 45: R1–R59.
were observed above the strong fluorescence. With Isola N, Stokes DL, and Vo-Dinh T (1998) Analytical
He/Ne laser excitation at 633 nm, several Raman Chemistry 70: 1352–1356.
features appeared above the background. With exci- Keller S, Lochte T, Dippel B, and Schrader B (1993) Qual-
tation at 785 or 852 nm, a good quality Raman ity control of food with near-infrared excited Raman
spectrum was obtained. In the textile industry, NIR- spectroscopy. Fresenius Journal of Analytical Chemistry
Raman was used to examine ready-made textiles for 346: 863–867.
the discrimination of the different raw materials. Kneipp K, Kneipp H, Itzkan I, Dasari R, and Feld M
NIR-Raman has been introduced for use into the (1999) Ultrasensitive chemical analysis by Raman spec-
food processing industry. In the production of oils troscopy. Chemical Reviews 99: 2957–2975.
McCreery RL, Horn AJ, Spencer J, and Jefferson E (1998)
and fats, the determination of the amount of unsat-
Journal of Pharmaceutical Science 87: 1–8.
uration, such as cis and trans isomers, can be Mahadevan-Jansen A, Mitchell MF, Ramanujam N, et al.
important for food processing and food labeling. (1998) Photochemistry and Photobiology 68: 123–132.
NIR-Raman was reported to measure the total Mulvaney SP and Keating CD (2000) Raman spectroscopy.
unsaturation and the cis and trans isomers online Analytical Chemistry 72: 145R–157R.
during the production process. Fluorescence inter- Weissenbacher N, Lendl B, Frank J, et al. (1997) Journal of
ference with visible excitation of many fats was Molecular Structure 410–411: 539–542.

Surface-Enhanced
R E Littleford, D Graham, and W E Smith, University Introduction
of Strathclyde, Glasgow, UK
I Khan, Imperial College, London, UK
Surface enhanced Raman scattering (SERS) is a sen-
& 2005, Elsevier Ltd. All Rights Reserved. sitive spectroscopic technique for the detection and

You might also like