You are on page 1of 2

QUICK GUIDE

ERBECRYO® 2
This quick guide is no substitute for briefing and user manual. This quick guide does not include any safety notes. Make yourself familiar
! with the safety notes in the user manuals.

Necessary operating steps


1 Opening the gas bottle 5 Checking the function and seal of the instrument

VWHULOHZDWHURU
1D&OVROXWLRQ
DSSUR[Ɠ&

1. Immerse the instrument tip to a depth of approx. 5 cm.


1. Open the bottle valve. 2. If bubbles develop during immersion, eliminate the bubbles by
2. Check the seal of all connections. No gas may escape. moving the instrument tip back and forth quickly in the water.
3. Activate ERBECRYO 2 for approx. 5 seconds.
2 Switch on the device At a sufficient level of freezing performance, a clearly visible, homoge-
neous ball of ice must form at the tip of the instrument. No gas bubbles
should escape.
Never use an instrument if the freezing performance is inadequate or
if the instrument is not leak-tight. If an instrument is not leak-tight,
there is a risk of a gas embolism.

6 Freezing
3 Connecting the instrument

4 Checking the reusability of the instrument 1. Position the instrument.


2. Press the footswitch. The Freeze mode icon is shown on the dis-
play.
3. Carry out the required application.
4. Note the timer setting with the optical and acoustic signals in each
case.

7 Thawing

 Release the footswitch.

CRYOSURGERY 80114-351 10.15


8 Closing the gas bottle after use 11 Changing the timer setting

Optional operating steps


9 Select program
The effect and timer are saved in a program.

• No Timer: The freeze interval is not measured.


• Count Up: The freeze interval is measured in increments of one sec-
ond
• Count Up & Beep: The freeze interval is measured in increments of
one second. A signal sounds once every second.
• Count Down & Beep: Enables selection of a countdown interval. A
signal sounds once the countdown has been completed.

10 Changing the effect

By selecting the effect you can influence the temperature of the probe
tip. The greater the effect, the colder the probe tip will be.

© Erbe Elektromedizin GmbH Hotline +49 7071 755 123 www.erbe-med.com