̭͈Ρ΅νιϋΠ͉ͅȂུࢊβȜΐ̳̞̀ͦ͘͘͜܄ȃ
Caution Refer to the Read Me First: Safety and Note National Instruments recommends against
Electromagnetic Compatibility document at switching active RF signals. As a relay actuates, the
ni.com/manuals for important safety and channel is momentarily unterminated. Some RF
compliance information. sources can be damaged by reflections if their
outputs are not properly terminated. Refer to your
About These Specifications RF source documentation for more information.
Specifications characterize the warranted performance of the
instrument under the following operating conditions: RF Performance Characteristics
• Chassis is powered on Values in parentheses are typical.
• Calibration adjustment cycle maintained Insertion loss
• 50 Ω termination connected to unused I/O front panel ≤2.4 GHz.....................................<4.1 dB (<3.4 dB)
connectors ≤6 GHz........................................<5.8 dB (<5.1 dB)
Typical Specifications are specifications met by the majority ≤6.6 GHz.....................................<7.0 dB (<6.1 dB)
of the instruments under the stated operating conditions.
Typical specifications are not warranted.
VSWR
0
1.5
where IL represents insertion loss in dB 1.4
1.3
T represents the temperature at which the property is 1.2
being measured in °C 1.1
1
T0 represents the reference temperature in °C 0.01 1 2 3 4 5 6 7
α represents insertion loss temperature coefficient in Frequency (GHz)
ppm/°C
Isolation (dB)
–40
≤2.4 GHz.....................................>70 dB (>84 dB)
–60
≤6 GHz........................................>61 dB (>74 dB)
–80
≤6.6 GHz.....................................>59 dB (>72 dB)
–100
CH–CH Isolation
–120
≤2.4 GHz.....................................>69 dB (>90 dB) 0.01 1 2 3 4 5 6 7
≤6 GHz........................................>58 dB (>74 dB) Frequency (GHz)
≤6.6 GHz.....................................>53 dB (>71 dB)
–2
–3 Note Measurements performed with two 10 dBm
–4 input tones = 1 MHz apart. This specification is
–5 based on both experimental and calculated data.
–6
–7
0.01 1 2 3 4 5 6 7
Frequency (GHz)
Note The NI 2543 can recognize trigger pulse Shock and Vibration
widths less than 150 ns by disabling digital filtering. Operational Shock ...........................30 g peak, half-sine,
For information about disabling digital filtering, 11 ms pulse
refer to the NI Switches Help. (Tested in accordance
with IEC 60068-2-27.
Front panel input voltage Test profile developed in
Minimum .................................... –0.5 V accordance with
MIL-PRF-28800F.)
VL maximum .............................. +0.7 V
VH minimum............................... +2.0 V Random Vibration
Nominal ...................................... +3.3 V Operating ....................................5 to 500 Hz, 0.3 grms
Maximum ................................... +5.5 V Nonoperating ..............................5 to 500 Hz, 2.4 grms
(Tested in accordance
Output trigger
with IEC 60068-2-64.
Destinations ................................ PXI trigger lines 0–7, Nonoperating test profile
Front panel exceeds the requirements
Pulse width ................................. Programmable of MIL-PRF-28800F,
(1 μs to 62 μs) Class 3.)
Physical Characteristics
Switch type ..................................... FET
Part
Accessory Length Number
NI PXI-2543 NI PXIe-2543
SMA 100, SMA male to SMA 0.15 m 763443-01 6.6GHz Terminated 6.6GHz Terminated
Solid State Dual 4x1 Solid State Dual 4x1
male flexible cable 50Ω Multiplexer 50Ω Multiplexer
0.45 m 763444-01
Diagrams
Figure 4 shows the NI 2543 power-on state. COM B COM B
COM A COM A
k0
CH 0A k4 CH 2B CH 2B
50 Ω CH 2A
CH 2A
k1
CH 1A k6
50 Ω k7
k2 COM A CH 3B CH 3B
50 Ω CH 3A CH 3A
CH 2A k5
50 Ω
k3
TRIG IN TRIG OUT TRIG IN TRIG OUT
CH 3A
50 Ω
ESD ESD
k8
SENSITIVE SENSITIVE
CH 0B k12 ALL PORTS ALL PORTS
50 Ω +30 dBm MAX +30 dBm MAX
k9
CH 1B k14
50 Ω k15
k10 COM B
50 Ω
CH 2B k13
50 Ω
k11
CH 3B Figure 5. NI 2543 Front Panel Connector
50 Ω
CE Compliance
This product meets the essential requirements of applicable
European Directives as follows:
• 2006/95/EC; Low-Voltage Directive (safety)
• 2004/108/EC; Electromagnetic Compatibility Directive
(EMC)